The installation should comply with all the electrical
requirements indicated in this document. These requirements
should be upgraded if Local Standards are more stringent.
RMS Line current during a X-ray exposure, generator stand-by consumption(W),
the differential sensitivity(mA) and the thermomagnetic breaker should be:
Maximum power line impedance :
COMED Medical Systems Co.,Ltd.
4
Istallation
Table. Magnetic breaker requirement
LINE VOLTAGE SINGLE-PHASE( 37.5kW )
208VAC
273 A
230VAC
240VAC
Minimum kVA required
Stand-by Consumption
Differential Sensitivity
Thermomagnetic Breaker
Below 210VAC requires an auxiliary boost transformer to adequate the line voltage to the
Table. Wire size requirement
(SINGLE-PHASE 1Ø)
Line voltage
208 VAC
230 VAC
240 VAC
(RMS = momentary line current based on 100ms X-ray exposures)
generator input
15m 30m
50 ㎟ 105 ㎟
50 ㎟ 105 ㎟
50 ㎟ 95 ㎟
Maximum kW ×1.2
50% of the RMS line current
Wire size at ;
245 A
232 A
500 W
30 mA
COMED Medical Systems Co.,Ltd.
5
Istallation
4. PACKING
tube assembly stand guide
The generator is shipped in Table to facilitate transport and
installation.
Upon receipt of the X-ray unit and associated equipment,
inspect all shipping containers for signs of damage. If
damage is found, notify the carrier or his agent immediately.
COMED Medical Systems Co.,Ltd.
6
Istallation
5. DIMENSION
1. Tube forward and backward movement range : 185mm
2. X-axis center rotation angle range : over ± 90°
3. Z-axis center rotation angel range : over ± 90°
4. Ttube max upward movement : above ground 1720 mm
5. Tube max downward movement : above ground 650 mm
6. Tube up and down movement range : 1170 mm
7. Tube left and right movement range : 1940 mm
8. Table forward and backward movement range : 200mm
9. Table left and right range : 600mm
Total weight : 430kg
COMED Medical Systems Co.,Ltd.
7
Istallation
6. INSTALLATION
Top table
Braker rubber stopper
1. Put in top table to table.(first, remove rubber stopper and insert rubber stopper after insert
top table)
2.connect braker into table.
Breaker interval : 4 - 6mm
COMED Medical Systems Co.,Ltd.
8
Istallation
When heavy person be on the bed, consider interval of locker and bed (about 5 mm)
And fix with volts and wind tightly.
3.Connect rail to table.
4.Put in stand support to rail.
5.Lift up rail with rail support.
① Rail ②Stand support ③Rail support
6.Connect stand into stand support.
t
Stand support
Weight blance fixing bolts
COMED Medical Systems Co.,Ltd.
9
Istallation
Tube support and stand ware fixed with tape(or wire).
This fixing tape should be remove the last.
Because, wire inside stand would be not tangle.
Tube support must be face below direction(direction of weight balance )
Pic. Stand assembly and inside drawing
Weight balance
A angle B angle
7.fixing stand with six bolts.
Six bolts
Using the six bolts contrast A angle and B angle equaled
And then remove fixing bolt in weight balance.
wire
A angle
Tube support
Fixing tape
Fixing bolt
B angle
COMED Medical Systems Co.,Ltd.
10
Istallation
8.fixing the tube assembly into tube support and cable connection.
Connect cable duct into rail support
COMED Medical Systems Co.,Ltd.
11
Istallation
9.fixing all cover.
Assemble front and back case.
Insert rail-cover into rail and sticking.
(Release the dish-head-bolt and tighten with the rail-cover by the bolt.)
.
COMED Medical Systems Co.,Ltd.
12
Istallation
10.connection HV cable and terminal connection
HV cable Connection
When you are connecting high voltage cable, be careful of its polarity.
*CAUTION: Cable must be connected deep into the hole.
Cathod(-)
Anode(+)
COMED Medical Systems Co.,Ltd.
13
Istallation
T2 OIL INJECTION
T1
CATHOD
GND
* CORE
CORE
No. Symbol Description Remark
1 mA TP+ mA Test Point (+)
2 mA TP- mA Test Point (-)
3 F-L Large Focus of Filament
4 F-COM Common
5 F-S Small Focus of Filament
6 mA +F/B mA Feedback (+)
7 mA –F/B mA Feedback (-)
8 kV +F/B kV Feedback (+)
9 INT Tank Interlock
10 GND Ground(Earth)
mA & mAs Meter
Connection
ANODE
COMED Medical Systems Co.,Ltd.
14
Istallation
11. Board and terminal connection
11.1 I/F Board and terminal connection
COMED Medical Systems Co.,Ltd.
15
Istallation
Terminal connection
1. +24VB : Handel Bar Board Relay Power (+) AND (LAMP Power)
Stand Bulky Driving Magnetic Conductor Power (+) AND (LAMP Power)
2. +24VB GND : Handel Bar Board Relay Power (-)
Stand Bulky Driving Magnetic Conductor Power (-)3.
3. 0VAC : Collimator Power Cable
4. 25VAC : Collimator Power Cable
5. 0VAC : Table Bulky Power Cable (Table Board Power 0VAC com)
6. BKY1 EXP2 : Table Bulky Power Cable
7. 0VAC : Table Bulky Power Cable
8. BKY2 EXP2 : Stand Bulky Power Cable
9. 0VAC : Stand Bulky Power Cable
10. THERM : TUBE Over-Heat Protection Cable
11. THERM : TUBE Over-Heat Protection Cable
12. ROTOR1(W/COM) : Tube Rotor Connection Cable
13. ROTOR2(B/MAIN) : Tube Rotor Connection Cable
14. ROTOR3(G/AUX) : Tube Rotor Connection Cable
15. 110VAC : Table Board Power Cable
16. 0VAC : Table Board Power Cable
17. CASE GND : Ground
18. WARNING 2-1 : PSU Charge ON Signal(OPTION)
19. WARNING 2-3 : PSU Charge ON Signal(OPTION)
20. SMPS +24V : Table Stand Driving Magnetic Conductor Power (+)
21. SMPS -24V : Table Stand Driving Magnetic Conductor Power (-)
COMED Medical Systems Co.,Ltd.
16
Istallation
I/F Board connection
TB501
TB502
CN503
CN502
TB504
TB503
TB502
CN501 CN504
COMED Medical Systems Co.,Ltd.
17
Istallation
TB 504 (BR-944C) TB 503 (BR-944C)
PIN I/O Signal PIN I/O Signal
1 O BUCKY1 SPLY 1 I DOOR
2 O BUCKY1 DR 2 I GND
3 O BUCKY1 EXP1 3 I OTHER DEVICE 1
4 O BUCKY1 EXP2 4 I GND
5 O SPARE OUT1-1 5 I OTHER DEVICE 2
6 I SPARE OUT1-2 6 I GND
7 I SPARE OUT2-1 7 I SPARE IN 1
8 O SPARE OUT2-2 8 I GND
9 O BUCKY2 SPLY 9 I SPARE IN 2
10 O BUCKY2 DR 10 I GND
11 O BUCKY2 EXP1 11 O LIGHT IN USE-1
12 O BUCKY2 EXP2 12 O LIGHT IN USE-2
13 I BUCYK2 OK 13 O WARNING 1-1
14 I BUCYK2 OK (GND) 14 O WARNING 1-2
15 O BUCYK1 OK 15 O WARNING 1-3
16 O BUCYK1 OK (GND) 16 O WARNING 2-1
17 O +24V 17 O WARNING 2-2
18 O 24V GND 18 O WARNING 2-3
TB 502 (BR-944C) TB 501 (BR-944C)
PIN I/O Signal PIN I/O Signal
1 O ROTOR 1 (COM) 1 25 VAC
2 O ROTOR 2 (MAIN) 2 16 VAC
3 O ROTOR 3 (SHIFT) 3 0 VAC
4 O GND 4 220 VAC
5 I TUBE THERMAL 5 110 VAC
6 I TUBE THERMAL 6 0 VAC
7 O 25 VAC (10A) 7
8 O 16 VAC (10A) 8
COMED Medical Systems Co.,Ltd.
18
Istallation
J
J
J
11.2 Handle Bar Board and Table Board terminal connection
HANDLE BAR BOARD
3
TABLE BORAD
TB1
2
TB2
TB3
4
COMED Medical Systems Co.,Ltd.
19
Istallation
J4 J3
PIN I/O Signal PIN I/O Signal
1 O 24 VDC 1 O F/B-
2 O 0 VDC 2 O F/B+
3 I RY1 3 O U/D-
4 I RY2 4 O U/D+
5 I RY3 5 O 90-
6 I RY4 6 O 90+
7 I RY5 7 O ROT-
8 I RY6 8 O ROT+
9 9 O R/L-
10 10 O R/L+
J2 TB1
1 I LOCK- 1 I 110 VAC
2 I LOCK+ 2 I 0 VAC
3 I 0 VDC 3 I BUCKY 2
4 I 24 VDC 4 I BUCKY 2
TB2 TB3
1 O MG 1+ 1 I POS(RTN)
2 O MG 1- 2 O POS
3 O MG 2+ 3 I TRAY(RTN)
4 O MG 2- 4 O TRAY
MG 3+ 5 O 5 I T/C(RTN)
6 O MG 3- 6 O T/C
7 O MG 4+ 7
8 O MG 4- 8
9 O MG 5+ 9 I BUCKY OUT(RTN)
10 O MG 5- 10 O BUCKY OUT
COMED Medical Systems Co.,Ltd.
20
Istallation
7. Installation trouble shooting
7.1 bucky stand trouble shooting
1) locker interval(top cover off)
two bolts release and calibrate..
(doing move bucky device)
2) If x-ray film image moved, you should fix bucky to wall
COMED Medical Systems Co.,Ltd.
first, remove bolts(1) and top cover off
21
Istallation
3)Bucky stand input signal
1. DC+ (locker power)
2. DC- (locker power)
3. START( BKY2 EXP2—moving motor power110VAC )
4. START( 0VAC — moving motor power 110VAC )
5. CASE GND (Ground)
Pic1. Bucky stand wire
(BKY2 EXP2)
(0VAC)
COMED Medical Systems Co.,Ltd.
22
Istallation
1) side bolts off and side cover off
2) front bolts off and front wood cover off
side bolts view front bolts view
COMED Medical Systems Co.,Ltd.
23