1. GENERAL INFORMATION ..................................................................................................................................... 7
3. BALANCE CONTROL ........................................................................................................................................... 15
4.6.2 VALUE RELEASE ........................................................................................................................................................ 23
8. PARTS COUNTING ................................................................................................................................................ 30
12. DENSITY OF SOLIDS ........................................................................................................................................ 43
13. DENSITY OF LIQUIDS ...................................................................................................................................... 47
16. PEAK HOLD ....................................................................................................................................................... 54
19.3WI-FIPORTSETTINGS(TA.C AND TT.CSERIES)................................................................................................................. 72
21. USER ................................................................................................................................................................... 82
24. ADVANCED FEATURES ................................................................................................................................... 90
The balances are designed to provide accurate measurement of weighed loads, performed under
laboratory conditions.
1.2 PRECAUTIONS
• Prior to first use, it is highly recommended to carefully read this User Manual, and operate
the balance as intended.
• Do not operate the touch panel using sharp-edged tools (knife, screwdriver, etc.).
• While loading the balance make sure that load is placed in the very center of the weighing
pan.
• Make sure the load does not exceed instrument’s measuring range (maximum capacity).
• Do not leave heavy loads on the weighing pan for a long period of time.
• In case of failure, immediately unplug the instrument.
• Balances to be decommissioned, should be decommissioned in accordance with valid legal
regulations.
• Do not use the balance is areas endangered with explosion. The balance is not designed to
operate in EX zones.
1.3 SUPERVISION OVER METROLOGICAL PARAMETERS
Metrological parameters of a balance need to be checked by an authorized user. Inspection
frequency is qualified by the ambient conditions in which a balance is used, processes carried out
and adopted quality management system.
1.4 WARRANTY CONDITIONS
A. Cole Parmer will exchange, replace or repair the existing balance for any damage that
appears to be faulty by production or by construction within the 5-year warranty period.
B. Warranty is voided if:
•mechanical defects caused by inappropriate use:
• defects of thermal and chemical origin,
• defects caused by lightning, overvoltage in the power network
• defects caused by water damage
• or other random event
• overloading the mechanical measuring system
• installing another version of the operating system
• utilizing the balance contrary to its intended use
• repairs carried out by non-authorized service centers
• removing or destroying protective stickers which secure the balance’s housing against
unauthorized access
C. Warranty card must be filled out for warranty to be valid.
7
2. UNPACKING AND INSTALLATION
2.1 PLACE OF USE AND ASSEMBLING
• The balance should be stored and used in locations free of vibrations and shakes, free of air
movement and dust.
• Ambient air temperature should not exceed the range of: +10 °C ÷ +40 °C.
• Ambient relative humidity should not exceed 80%.
• During balance operation, ambient temperature in the weighing room should not change
rapidly.
• The balance should be located on a stable wall console desk or a stable working table which
is not affected by vibrations and distant from heat sources.
• Take special precaution when weighing magnetic objects, as part of the balance is a strong
magnet. Should such loads be weighed, use under-pan weighing option, which removes the
weighed load from area influenced by the balance’s magnet. The hook for under-pan
weighing is installed in balance’s base.
• Keep all package element should your device be transported in the future. Remember that
only original packaging can be used for shipping purposes. Prior to packing, uncouple any
cables, remove any separable components (weighing pan, shields, inserts). Pack the device
components into an original packaging. The original packaging protects the equipment
against potential damage during transportation.
2.2 STANDARD DELIVERY COMPONENTS LIST
• Balance and components shown in Section 2.4 depending on balance model
• Warranty Ca rd
• USB
o User Manual
o Balance USB Driver
o RLAB Software
o RLAB Software Manual
2.3 UNPACKING
To unpack the system, follow the diagram below-
8
2.4 BALANCE ASSEMBLY
Model: TA-225.C, TA-64
Components:
Installation:
1) Remove the transport lock– gently press the transport lock and turn it accordingly to
<OPEN>, keep the transport lock should your balance be transported in the future.
2) Check grounding spring to insure it is in the appropriate location. Make sure that the grounding
spring juts slightly out of the hole.
3) Install components following diagram below:
i. Bottom insert
ii. Centering ring [embossment side up]
iii. Weighing pan
iv. Draft shield
1) Remove the transport lock– gently press the transport lock and turn it accordingly to
<OPEN>, keep the transport lock should your balance be transported in the future.
2) Check grounding spring to insure it is in the appropriate location. Make sure that the grounding
spring juts slightly out of the hole.
3) Install components following the diagram below:
1) Remove transport lock, keep the transport lock should your balance be transported in the future.
2) Check that the grounding spring is in the correct location, the spring juts slightly out of the hole.
3) Install components following the diagram below:
i. Rubber feet (grounding foot can be in any location on the balance)
ii. Weighing pan
12
Model: TT-6002.C, TT-6002
Components:
Installation:
1) Remove transport lock, keep the transport lock should your balance be transported in the future.
2) Check that the grounding spring is in the correct location, the spring juts slightly out of the hole.
3) Install components following the diagram below:
i. Draft shield
ii. Rubber feet (grounding foot can be in any location on the balance)
iii. Weighing pan
13
Model: TT-10102.C
Components:
Installation:
Install components following the diagram below:
i. Draft shield
ii. Weighing pan
2.5 BALANCE LEVELING
It is necessary to level the balance prior to plugging it in. To level the balance, turn its feet until the
air bubble is in the center position.
The balance should firmly rest on a surface, each of the feet must be supported.
14
2.6 POWERING THE DEVICE
Before plugging in your balance, it is imperative to wait until the balance reaches thermal
stabilization.
For balances that were stored in much lower temperatures (e.g. during winter period), thermal
stabilization period will take at least 4 hours for Precision balances, and 8 hours for Analytical
balances.
• Balance should be plugged in only with the power adapter that comes standard with the
model. Nominal power supply of the power adapter (specified on the power adapter data
plate) should be compatible to the power supply.
• Plug the balance in – connect the power adapter to the socket, next connect its connector to
port located at the back of the balance housing.
• Press button located in the top right hand corner of the key pad.
! Remember to start the balance with no load on the weighing pan
• Wait u ntil the start-up procedure is completed. The home screen of balance software is
displayed automatically.
• If the indication is different than zero, please press button
3. BALANCE CONTROL
3.1 BALANCE KEYBOARD
15
3.2 BALANCE HOME SCREEN
3.3 ENTERING BALANCE MENU
Operation of the balance menu is intuitive and easy to use. The touch panel makes the software
operation easy. Pressing a function key, a soft key or an area on the display initiates an assigned
function or process
To enter balance main menu, press < PARAMETERS> button. Clicking any button within this
section, or clicking any button with a parameter name, results in a change of color. This is for
signaling purposes. If a given area has any function or action assigned, then it is performed
automatically upon clicking (e.g. adjustment procedure), respectively a window with parameters or a
list of appropriate settings is displayed.
16
3.4 SCREEN SCROLLING
Press to enter the main menu.
Press to clear the editing field.
Press to scroll the menu “up”,
Press to enable / disable an on-
Press to confirm changes.
Press to export databases (key
drive).
Press to resign form introducing
Press to import databases (key
Press to add a new record to a
Press to select variables, out of the
Press to search for a record in a
Press to print out a record
There are two methods for scrolling the screen of parameters window. The first one requires
pressing, holding down and scrolling up or down the scrollbar located on the left. The second one
requires pressing, holding down and scrolling up or down on any point of the displayed window.
3.5 SOFT KEYS LIST
or “down”.
function modifications.
database.
database.
from a database.
screen keyboard.
active upon plugging a USB flash
active upon plugging a USB flash
drive).
list, for a printout.
Press to move (exit) one level up.
17
4. WEIGHING MODE
Load an object on the balance weighing pan. The stabilization of weighing result is indicated by a
stability marker visible on t he left side of balance display, read the measurement result.
4.1 GOOD WEIGHING PRACTICE
To ensure long lasting use of a balance with correct and reliable measurements of weighed loads,
follow the procedures below:
• Start the balance with no load on the weighing pan.
• Load the weighing pan carefully and avoid dropping it:
• Place weighed load in the center of the weighing pan:
• Avoid side loading, in particular side shocks:
18
4.2 UNITS
Unit
Denotation
Unit
Denotation
To change the weighing unit, press the weighing unit icon. This brings up the available unit’s list,
choose new unit from the list displayed.
You may create two custom user-defined units. Displayed value of a user-defined unit is a
multiplication of measured mass value and a coefficient specified for the user-defined unit. The units
can be freely named with use of 3 characters’ maximum. By default, the names are displayed as
[u1] – user unit 1, and [u2] – user unit 2.
4.4 BALANCE ZEROING
Zeroing is a function allowing to zero mass indication.
The Zeroing process is an equivalent for determining new zero point, recognized by the balance as
precise zero. Zeroing is possible only for stable status of display indication.
! CAUTION
Zeroing the display indication is possible only within ±2% range of instrument’s maximum
capacity. If the zeroed value is above ±2% of the maximum capacity, then the balance
indicates a respective error message.
20
4.5 BALANCE TARING
Taring is a function allowing to determine net weight of a measured object. To determine net weight
of the object, place object’s container (packaging) on the weighing pan, and on stabilization of
measurement result press key.
The balance enables assigning tare value to a database-stored product. Using this option, the
software automatically uploads data on tare value for a product upon its selection from the
database.
! CAUTION
Taring negative values is impossible. On taring negative values, the balance responds with
an error message. In such case, zero balance indication and repeat taring procedure.
4.5.1 MANUAL TARE DETERMINATION
• Press quick access key
• An on-screen numeric keyboard is displayed
• Enter tare value and press key
• The balance returns to the weighing mode, and the display indicates entered tare value
with minus ‘–‘ sign
4.5.2 DELETING TARE
The tare value indicated on balance display can be deleted by pressing key on
balance overlay.
Procedure 1 - on taking the tared load off the weighing pan
• Press key.
• The NET marker is deleted, and new zero point of the balance is determined.
21
Procedure 2 – with tared load on the weighing pan
• Press key.
• The NET marker is deleted, and new zero point of the balance is determined.
• When tare value exceeds 2% of the maximum capacity, respective message is
displayed to inform a user about the fact.
4.5.3 TARE DATABASE
Procedure-
• While in optional mode, press button located in a top right hand corner of the mass
display.
• Wait for a respective window to open, next select <PACKAGING>.
• Window with a list of tare values recorded into tare database opens.
• Select the packaging that is to be used.
• The balance returns to the weighing mode, and the display indicates selected tare value
with a minus ‘–‘ sign.
OR
• While in optional mode, press button (if displayed on the screen),
• Window with a list of tare values recorded into tare database opens.
• Select the packaging that is to be used.
• The balance returns to the weighing mode, and the display indicates selected tare value
with a minus ‘–‘ sign.
4.6 WEIGH MODE SETTINGS
The balance allows setup of operating parameters (filters, value release and autozero function,
deleting the last digit and other settings) separately for each working mode.
Procedure-
22
4.6.1 FILTER LEVEL SETTINGS
Filter settings adjustment depends on the working environment. For the best possible conditions
the filter can work in a very fast mode (V.FAST value for Filter parameter); however, if the
conditions are poor (shakes, drafts), the filter should be set to slow or very slow option (SLOW
or V. SLOW value for Filter parameter). The effectiveness of the filter is different throughout the
weighing range. The filter works slower when “approaching” the weighed mass, it works more
quickly for weighed mass within the set range of the filter (parameter for setting filter range is
accessible only from the service menu – the user does not have any access to it).
Depending on the filter, the weighing time is shorter (V.FAST and FAST) or longer (SLOW and
V. SLOW).
CAUTION!
The higher filter level, the longer the weighing time.
4.6.2 VALUE RELEASE
Since ambient conditions at a workplace vary, it is necessary to determine the value release
parameter that are best for your working environment, parameter options are: FAST.+REL.,
FAST or RELIABLE. Depending on the selected option, weighing time is either shorter or
longer.
4.6.3 AUTOZERO FUNCTION
The balance features an autozero function (Auto). This function automatically controls and
corrects the zero reading. When Autozero is enabled, it compares balance readings at declared
time interval e.g. 1s, if weighing pan is unloaded and display indication is close to zero. If results
vary less than declared AUTOZERO range e.g. one division, balance zeroes automatically,
marker of stable measurement result , and precise zero marker are displayed.
If AUTOZERO function is enabled, then each weighing process starts from precise zero point.
There are, however, some instances when this function can be a disturbing factor for the
measuring process; e.g. very slow placing of a load on the weighing pan (load adding). Here,
zero reading correction can also correct the actual reading of loaded mass.
23
4.6.4 LAST DIGIT DISPLAY
Function enables displaying the last digit of decimal place for a weighing result. There are three
available options:
• Always: all digits visible
• Never: last digit is not displayed
• When stable: last digit is displayed only for a stable weighing result
4.6.5 BALANCE AMBIENT CONDITIONS
Parameter relating to ambient and environmental conditions in which the balance operates.
There are two options: STABLE and UNSTABLE. Selecting STABLE mode makes the balance
work much faster, i.e. weighing takes much less time than for UNSBABLE mode. If the ambient
conditions are unstable it is recommended to use UNSTABLE mode. By default, the parameter
is set to STABLE option.
5. ADJUSTMENT (TA.C AND TT.C SERIES)
To ensure the highest weighing accuracy, it is recommended to periodically introduce a corrective
factor of indications to balance memory, the said factor must be referred to a mass standard. In other
words, balance adjustment shall be performed from time to time.
Adjustment should be carried out:
• Before the beginning of weighing procedure
• If long breaks between following measuring series occur
• If temperature inside the balance changes more than: 1°C or 2°C for Analytical balances or 2°C
for Precision series balances.
Types of adjustment:
• Automatic adjustment (automatic or manual)
• External adjustment
• User defined adjustment- external weight of declared mass which cannot be modified or of any
mass, but not lower than 30% of maximum range.
5.1 INTERNAL ADJUSTMENT
Internal adjustment can be triggered in two fashions, automatically or manually by the user.
5.1.1 AUTOMATIC ADJUSTEMENT
This menu is for declaring a value initiating an automatic internal adjustment. Accessible
options:
• None – automatic adjustm ent disabled
• Time – adjustment takes place in time intervals declared in menu <Automatic adjustment
time>. The time interval is declared in hours and ranges between 0.5 and 12 hours.
• Temperature – adjustment is triggered by temperature change only
• Both – adjustment is triggered by both, temperature changes and time.
CAUTION!
Parameter settings can be modified only for balances that are not a subject to conformity
assessment (verification).
24
5.1.2 MANUAL INTERNAL ADJUSTMENT
Internal adjustment is carried out with of an internal adjustment weight. button, when
pressed, automatically triggers adjustment process. Upon adjustment process completion,
respective message informing about process end and about its status is displayed.
CAUTION!
Adjustment procedure requires stable environmental conditions (no air drafts or ground
vibrations). The process must be carried out with an empty weighing pan.
5.2 EXTERNAL ADJUSTMENT
External adjustment is carried out by means of an external mass standard of specified accuracy
class and weight. Both, accuracy class and mass standard weight depend on balance type and max
capacity. The process takes semi-automatic form; successive stages are signaled with prompts.
Procedure-
• Enter <Adjustment> submenu, next select “External adjustment” option,
• “Remove weight” prompt is displayed.
• Take the weight off the weighing pan and press button. Whereas balance
determines start mass, “Adjustment; Please wait…” prompt is displayed,
•Upon completed start mass determination procedure “Put weight …” prompt is displayed
along with particular mass standard value.
• Put the required weigh on a pan and press button,
• Upon completed procedure “Remove weight” prompt is displayed.
• Take the weight off the weighing pan, wait for <Adjustment> window to be displayed
again.
5.3 USER ADJUSTEMENT
An external calibration process can be carried out with an external mass standard of defined value
by the user. The calibration mass must be at a value greater than 30% of the balances Max range
Procedure-
• Enter <Adjustment> submenu, next select “User adjustment” option,
• Type in calibration weight to be utilized (must be at a value greater than 30% of
balances max range)
• Press button.
• The balance prompts you to “Remove weight”, with weighing pan clear press button
• Put the required weigh on a pan and press button,
• Upon completed procedure “Remove weight” prompt is displayed.
• Take the weight off the weighing pan, wait for <Adjustment> window to be displayed
again.
25
5.4 ADJUSTM E NT TE S T
<Adjustment test> function enables comparing the result of an internal automatic adjustment with
the value recorded in balance factory parameters. The comparison is used for determining balance
sensitivity drift over time.
6. ADJUSTMENT (TA AND TT SERIES)
To ensure the highest weighing accuracy, it is recommended to periodically introduce a corrective
factor of indications to balance memory, the said factor must be referred to a mass standard. In other
words, balance adjustment shall be performed from time to time.
Adjustment should be carried out:
• Before the beginning of weighing procedure
• If long breaks between following measuring series occur
• If temperature inside the balance changes more than: 1°C or 2°C for Analytical balances or 2°C
for Precision series balances.
Types of adjustment:
• External adjustment
• User defined adjustment- external weight of declared mass which cannot be modified or of any
mass, but not lower than 30% of maximum range.
6.1 EXTERNAL ADJUSTMENT
External adjustment is carried out by means of an external mass standard of specified accuracy
class and weight. Both, accuracy class and mass standard weight depend on balance type and max
capacity. The process takes semi-automatic form; successive stages are signaled with prompts.
Procedure-
• Enter <Adjustment> submenu, next select “External adjustment” option,
• “Remove weight” prompt is displayed.
• Take the weight off the weighing pan and press button. Whereas balance
determines start mass, “Adjustment; Please wait…” prompt is displayed,
•Upon completed start mass determination procedure “Put weight …” prompt is displayed
along with particular mass standard value.
• Put the required weigh on a pan and press button,
• Upon completed procedure “Remove weight” prompt is displayed.
• Take the weight off the weighing pan, wait for <Adjustment> window to be displayed
again.
26
6.2 USER ADJUSTEMENT
Weighing
balance measures gravitational force which attracts the load. An obtained result is
in a form of measurement result. More
Parts Counting
information found in Section 9.
Dosing
Percent Weighing
Means of operation: control of percent ratio of a sample in relation to a standard
Density of solids
Density of liquids
An external calibration process can be carried out with an external mass standard of defined value
by the user. The calibration mass must be at a value greater than 30% of the balances Max range
Procedure-
• Enter <Adjustment> submenu, next select “User adjustment” option,
• Type in calibration weight to be utilized (must be at a value greater than 30% of
balances max range)
• Press button.
• The balance prompts you to “Remove weight”, with weighing pan clear press button
• Put the required weigh on a pan and press button,
• Upon completed procedure “Remove weight” prompt is displayed.
• Take the weight off the weighing pan, wait for <Adjustment> window to be displayed
again.
7. WORKING MODES- GENERAL INFORMATION
The balance features the following working modes:
Means of operation: weight of a load is determined through an indirect measurement. A
processed to a digital format and displayed
information found in Section 4.
Means of operation: based on a determined mass of a single part it is possible to count
another part, if the mass of the single part is determined with sufficient accuracy, and that
the following parts are equal in mass. More information found in Section 8
Check Weighing
Means of operation: control of sample mass with applied thresholds. A user should
specify the value of min threshold <LO> and max threshold <HI>. More
Means of operation: a user should specify sample’s target mass to be obtained by
pouring. More information found in Section 10.
(reference). Obtained data provides percent ratio on how test sample differs from the
accepted standard (reference). More information found in Section 11.
Means of operation: based on Archimedes principle, a balance determines density of
solids. The mode requires an optional density determination kit. More information
found in Section 12.
Means of operation: based on Archimedes principle, a balance determines density of
liquids. The mode requires an optional density determination kit. More information
found in Section 13.
27
Animal Weighing
Means of operation: mass measurement takes place with application of filters dampening
Statistics
Peak Hold
Formulations
Key
Function
Modes featuring the key
animal moves on a weighing pan, thus enabling obtaining a correct measurement result.
More information found in Section 14.
Means of operation: carried out measurements are used to calculate statistical data, such
as Min, Max, deviation, etc. More information found in Section 15.
Means of operation: max temporary indication occurring in course of the weighing
process is hold on a display. More information found in Section 16.
Means of operation: by mixing specified ingredients you can obtain particular mixture, in
order to program given formulation you have to specify weight of particular ingredients.
More information found in Section 17.
7.1 RUNNING WORKING MODE
To run working mode other than the one currently operating:
• Press pictogram of currently used working mode, the pictogram is in the top left hand
corner
• Wait for the available working modes list to be displayed
• Select the working mode you need to operate.
7.2 QUICK ACCESS KEYS FOR EACH WORKING MODE
Accept/Print
All modes
Print header All modes
Print footer
All modes
Zero All modes
28
All modes (Parts Counting and Percent
All modes (Parts Counting and Percent
All modes (Parts Counting and Percent
Tare All modes
Change unit
Select unit
Weighing modes excluded)
Weighing modes excluded)
Parameters All modes
Databases All modes
User All modes
Product
Customer
All modes
All modes
Packaging All modes
Adjustment All modes
Variable 1 All modes
Variable 2 All modes
Variable 3 All modes
Hide/show last digit
Working mode parameters All modes
Set part mass
Determine part mass
Determine mass using 5 parts
Determine mass using 10 parts Parts Counting mode exclusively
Weighing modes excluded)
Parts Counting mode exclusively
Parts Counting mode exclusively
Parts Counting mode exclusively
29
Solids and Liquids Density modes and
Determine mass using 20 parts
Parts Counting mode exclusively
Determine mass using 50 parts
Parts Counting mode exclusively
Determine mass using 100 parts Part s Counting mode exclusively
Thresholds Check weighing mode exclusively
Min threshold Check weighing mode exclusively
Max threshold Check weighing mode exclusively
Target value Dosing mode exclusively
Set reference mass
Determine reference mass (set
as 100%)
Start
Percent Weighing mode
exclusively
Percent Weighing mode
exclusively
Formulations mode
Results St atistics mode exclusively
Finish Statistics mode exclusively
Finish Peak Hold mode exclusively
Formulation Formulations mode exclusively
8. PARTS COUNTING
< Parts counting> is a working mode which enables determination of quantity of small parts with
equal weight. Determined mass of a single part is used for the counting procedure. The single part
mass may be:
• Determined from a reference quantity of parts subjected to weighing,
• Acquired from database of products,
• Entered manually as a numeric value.
30
Loading...
+ 89 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.