Cole Parmer Techne FSC400D, Techne FSC496D Instruction Manual

Sample Concentrator
FSC400D & FSC496D
Instruction Manual
TECH0002 / Version 1.0
Contents
Introduction 3
General Description 3
Important Safety Advice 4
Installation 5
Operation 6
Preparation 6
Processing of Samples 8
Maintenance 8
Cleaning and Care 8 Carriage Friction 8 Replacing Sealing Pad 9 In Case of Contamination 9
Service and Repair 10
Warranty 10 Replacement Parts and Accessories 10
Diagram of Gas Chamber 11
Technical Specication 12
Customer and Technical Support 12
Solvent Evaporation Temperatures 13
2
Thank you for purchasing this Techne product. To get the best performance from the equipment, and for your own safety, please read these instructions carefully before use.
Before discarding the packaging check that all parts are present and correct.
When unpacking the unit, check that the following have been removed from the packing:
v Base board v Column and carriage v Gas chamber v Instruction manual v Warranty card
General Description
The Techne Sample Concentrator has been designed to provide a quick and convenient method of accelerating the concentration by evaporation of solvents from samples prior to analysis. It is designed to work in conjunction with Techne DB100/3, DB200/3, DB-3, and DB-3A Dri-Block® heaters*.
The Sample Concentrator consists of a gas chamber mounted above a Dri-Block containing samples are placed. Hyperdermic needles carry the gas down from the chamber into the test tubes. The samples are heated from below by the heater and the flow of gas directed over the surface of the samples displaces the evaporated solvent from above the liquid’s surface. This significantly increases the rate of sample concentration.
The unit provides accurate and reproducible control over the positioning of the needles. The gas chamber can be removed from its stand to provide easy access to the samples, and it can be replaced quickly. The whole unit is compact enough for convenient use in a fume cupboard.
®
heater into which vessels
®
* The Dri-Block
heater is not provided and must be purchased separately.
Important Safety Advice
Users should be aware of the following safety advice:
v DO NOT use combustible substances near hot objects. v DO NOT use the equipment in hazardous atmospheres. v HIGH TEMPERATURES ARE DANGEROUS as they can cause serious burns to operators and ignite
combustible material. Users should be aware of the following safety advice:
v USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS. v DO NOT apply severe physical stress or use at temperatures above the rated value.
Symbols Dened
WARNING BIOHAZARDHOT
SURFACE
3
Installation
For packing purposes the Sample Concentrator is supplied unassembled in three parts; the base board, the column and carriage, and the gas chamber. The unit is re-assembled as follows:
1. Unscrew the column fastening knob from the column support bracket already fixed to the base board. Locate the column on the two pins in the support bracket. Screw the fastening knob through the rear of the column and tighten firmly.
2. Align the threaded hole on the back of the gas chamber with the mounting on the carriage. Tighten the locking knob to secure the gas chamber in position.
3. Connect the intake nozzle at the rear of the gas chamber to a suitable supply of gas via a pressure reducing valve. The gas pressure should be adjusted as required at the reducing valve but should not exceed 2 psi. Pressure in excess of 2 psi will cause gas to be vented and wasted.
®
4. Place the Techne Dri-Block two spigots on the base board which align with two centrally located screw heads on the underside of the Dri-Block® unit.
Gas chamber
heater onto the base of the Sample Concentrator. It is positioned by means of
Carriage
Fastening knob
Locking knob
Intake nozzle
Column
Fastening knob
Needles
Base board
Location spigots
4