Cole-Parmer PT-T-353, PT-T-4502, PT-T-2002, PT-T-753, PT-T-6002 Instruction Manual

...
1
User Manual
Cole-Parmer Symmetry
PA-T and PT-T Balance Series
01-16-17 REV03
2
3
Contents
1. GENERAL INFORMATION ..................................................................................................................................... 6
1.1 INTENDED USE ................................................................................................................................................................... 6
1.2 PRECAUTIONS ................................................................................................................................................................... 6
1.3 SUPERVISION OVER METROLOGICAL PARAMETERS .......................................................................................................... 6
1.4 WARRANTY CONDITIONS ................................................................................................................................................... 6
2. UNPACKING AND INSTALLATION ....................................................................................................................... 7
2.1 PLACE OF USE AND ASSEMBLING ....................................................................................................................................... 7
2.2 STANDARD DELIVERY COMPONENTS LIST.......................................................................................................................... 7
2.3 UNPACKING ....................................................................................................................................................................... 7
2.4 BALANCE ASSEMBLY .......................................................................................................................................................... 8
2.5 BALANCE LEVELING .......................................................................................................................................................... 14
2.6 POWERING THE DEVICE ................................................................................................................................................... 14
3. BALANCE CONTROL ........................................................................................................................................... 15
3.1 BALANCE KEYBOARD ........................................................................................................................................................ 15
3.2 BALANCE HOME SCREEN.................................................................................................................................................. 15
3.3 ENTERING BALANCE MENU ............................................................................................................................................. 16
3.4 SCREEN SCROLLING .......................................................................................................................................................... 16
3.5 SOFT KEYS LIST ................................................................................................................................................................. 17
4. WEIGHING MODE .................................................................................................................................................. 17
4.1 GOOD WEIGHING PRACTICE ............................................................................................................................................ 18
4.2 UNITS ............................................................................................................................................................................... 18
4.3 USER-DEFINED UNIT ........................................................................................................................................................ 19
4.4 BALANCE ZEROING ........................................................................................................................................................... 20
4.5 BALANCE TARING ............................................................................................................................................................. 21
4.6 WEIGH MODE SETTINGS .................................................................................................................................................. 22
5. ADJUSTMENT ........................................................................................................................................................ 24
5.1 INTERNAL AUTOMATIC ADJUSTMENT ............................................................................................................................. 24
5.2 MANUAL INTERNAL ADJUSTMENT................................................................................................................................... 25
5.3 EXTERNAL ADJUSTMENT .................................................................................................................................................. 25
5.4 ADJUSTMENT TEST .......................................................................................................................................................... 25
6. WORKING MODES- GENERAL INFORMATION ............................................................................................... 26
6.1 RUNNING WORKING MODE ............................................................................................................................................. 27
6.2 QUICK ACCESS KEYS FOR EACH WORKING MODE ............................................................................................................ 27
7. PARTS COUNTING ................................................................................................................................................ 29
7.1 SUPPLEMENTARY SETTINGS FOR PARTS COUNTING MODE ............................................................................................ 30
7.2 PARTS COUNTING QUICK ACCESS KEYS ......................................................................................................................... 31
7.3 PARTS COUNTING PROCEDURE ....................................................................................................................................... 31
7.4 SETTING REFERENCE BY ENTERING DETERMINED PART MASS ........................................................................................ 32
7.5 SETTING REFERENCE MASS BY DETERMINING PART MASS .............................................................................................. 32
7.6 ACQUIRING PART MASS FROM A DATABASE ................................................................................................................... 34
4
8. CHECKWEIGHING ................................................................................................................................................. 35
8.1 SUPPLEMENTARY SETTINGS FOR CHECKWEIGHING MODE ............................................................................................. 35
8.2 CHECKWEIGHING QUICK ACCESS KEYS .......................................................................................................................... 35
8.3 MAKING USE OF CHECKWEIGHING THRESHOLDS ............................................................................................................ 35
9. DOSING ................................................................................................................................................................... 37
9.1. SUPPLEMENTARY SETTINGS FOR DOSING MODE ........................................................................................................... 37
9.2 DOSING QUICK ACCESS KEYS ......................................................................................................................................... 37
9.3 MAKING USE OF PRODUCTS DATABASE FOR DOSING OPERATION .................................................................................. 38
10. PERCENT WEIGHING ....................................................................................................................................... 39
10.1. SUPPLEMENTARY SETTINGS FOR PERCENT WEIGHING ................................................................................................ 39
10.2 PERCENT WEIGHING QUICK ACCESS KEYS ................................................................................................................... 40
10.2 COMPARISON OF SAMPLE OF REFERENCE MASS ........................................................................................................... 40
11. DENSITY OF SOLIDS ........................................................................................................................................ 42
11.1 SUPPLEMENTARY SETTINGS FOR SOLIDS DENSITY MODE.............................................................................................. 43
11.2. SOLIDS DENSITY QUICK ACCESS KEYS ......................................................................................................................... 43
11.3 SOLIDS DENSITY DETERMINATION ................................................................................................................................. 43
12. DENSITY OF LIQUIDS ...................................................................................................................................... 45
12.1 SUPPLEMENTARY SETTINGS FOR LIQUIDS DENSITY MODE ............................................................................................ 46
12.2 LIQUIDS DENSITY QUICK ACCESS KEYS ........................................................................................................................ 46
12.3 LIQUIDS DENSITY DETERMINATION ............................................................................................................................... 46
13. ANIMAL WEIGHING .......................................................................................................................................... 48
13.1 SUPPLEMENTARY SETTINGS FOR ANIMAL WEIGHING MODE ........................................................................................ 49
13.2. ANIMAL WEIGHING QUICK ACCESS KEYS ................................................................................................................... 50
13.3 ANIMAL WEIGHING PROCEDURE ................................................................................................................................... 50
14. STATISTICS ....................................................................................................................................................... 51
14.1. SUPPLEMENTARY SETTINGS FOR STATISTICS MODE ..................................................................................................... 51
14.2. STATISTICS QUICK ACCESS KEYS ................................................................................................................................. 51
14.3 PARAMETERS FOR SERIES OF MEASUREMENT .............................................................................................................. 51
15. PEAK HOLD ....................................................................................................................................................... 53
15.1 SUPLEMENTARY SETTINGS FOR PEAK HOLD MODE ....................................................................................................... 53
15.2. PEAK HOLD QUICK ACCESS KEYS ................................................................................................................................. 53
15.3 MEANS OF OPERATION .................................................................................................................................................. 54
16. FORMULATIONS ............................................................................................................................................... 54
16.1 SUPPLEMENTARY SETTINGS FOR FORMULATIONS MODE ............................................................................................. 55
16.2. FORMULATIONS QUICK ACCESS KEYS ........................................................................................................................ 55
16.3 ADDING FORMULATION TO FORMULATIONS DATABASE .............................................................................................. 56
16.4 USING FORMULATIONS FOR WEIGHING ........................................................................................................................ 58
16.5 PERFORMED FORMULATIONS REPORT .......................................................................................................................... 63
17. DATABASES ...................................................................................................................................................... 64
17.1 DATABASE CONNECTED OPERATIONS ........................................................................................................................... 64
5
17.2 PRODUCTS ..................................................................................................................................................................... 66
17.3 OPERATORS ................................................................................................................................................................... 66
17.4 PACKAGING .................................................................................................................................................................... 67
17.5 CUSTOMERS ................................................................................................................................................................... 67
17.6 FORMULATIONS............................................................................................................................................................. 68
17.7 FORMULATIONS REPORTS ............................................................................................................................................. 68
17.8 AMBIENT CONDITIONS .................................................................................................................................................. 68
17.9 WEIGHING RECORDS ..................................................................................................................................................... 69
17.10 ALIBI ............................................................................................................................................................................. 69
18. COMMUNICATION............................................................................................................................................. 70
18.1 RS 232 PORTS SETTING................................................................................................................................................... 70
18.2 ETHERNET PORT SETTINGS ............................................................................................................................................ 71
18.3 WI-FI PORT SETTINGS ..................................................................................................................................................... 71
18.4 USB PORT ....................................................................................................................................................................... 74
19. PERIPHERAL DEVICES ................................................................................................................................... 78
19.1 COMPUTER .................................................................................................................................................................... 79
19.2 PRINTER ......................................................................................................................................................................... 79
19.3 SAVE MEASURMENT DATA ONTO USB FLASHDRIVE ...................................................................................................... 80
20. USER ................................................................................................................................................................... 81
21. PRINT MODE ...................................................................................................................................................... 82
21.1 ADJUSTMENT PRINTOUT REPORT .................................................................................................................................. 83
21.2 ADJUSTMENT PRINTOUT REPORT .................................................................................................................................. 84
22. PROXIMITY SENSORS ..................................................................................................................................... 87
23. ADVANCED FEATURES ................................................................................................................................... 88
23.1 BUTTON, LABELS AND TEXT FIELDS CONFIGURATION ................................................................................................... 88
23.2 LABELS ........................................................................................................................................................................... 91
23.3 TEXT FIELDS .................................................................................................................................................................... 94
23.4 BAR GRAPHS .................................................................................................................................................................. 96
23.5 QUICK ACCESS KEYS CUSTOMIZATION ........................................................................................................................... 98
23.6 WEIGHING UNIT ACCESSIBILITY ..................................................................................................................................... 99
23.7 START UNIT SELECTION ................................................................................................................................................ 100
23.8 BALANCE SETTINGS ...................................................................................................................................................... 100
24. MAITENANCE .................................................................................................................................................. 104
25. ACCESSORIES ................................................................................................................................................ 107
25.1 DENSITY DETERMINATION KIT ..................................................................................................................................... 107
26. APPENDIX ........................................................................................................................................................ 110
26.1 ADVANCED SERIES BALANCES ...................................................................................................................................... 110
26.2 DIMENSIONS ................................................................................................................................................................ 111
26.3 CONNECTORS .............................................................................................................................................................. 112
26.5 ERROR MESSAGES ........................................................................................................................................................ 113
26.6 WARRANTY CARD ........................................................................................................................................................ 113
6
1. GENERAL INFORMATION
1.1 INTENDED USE
The balances are designed to provide accurate measurement of weighed loads, performed under laboratory conditions.
1.2 PRECAUTIONS
Prior to first use, it is highly recommended to carefully read this User Manual, and operate
the balance as intended.
Do not operate the touch panel using sharp-edged tools (knife, screwdriver, etc.). While loading the balance make sure that load is placed in the very center of the weighing
pan.
Make sure the load does not exceed instrument’s measuring range (maximum capacity). Do not leave heavy loads on the weighing pan for a long period of time. In case of failure, immediately unplug the instrument. Balances to be decommissioned, should be decommissioned in accordance with valid legal
regulations.
Do not use the balance is areas endangered with explosion. The balance is not designed to
operate in EX zones.
1.3 SUPERVISION OVER METROLOGICAL PARAMETERS
Metrological parameters of a balance need to be checked by an authorized user. Inspection frequency is qualified by the ambient conditions in which a balance is used, processes carried out and adopted quality management system.
1.4 WARRANTY CONDITIONS
A. Cole Parmer will exchange, replace or repair the existing balance for any damage that
appears to be faulty by production or by construction within the 5-year warranty period.
B. Warranty is voided if:
mechanical defects caused by inappropriate use:
defects of thermal and chemical origin,
defects caused by lightning, overvoltage in the power network
defects caused by water damage
or other random event
overloading the mechanical measuring system installing another version of the operating system utilizing the balance contrary to its intended use repairs carried out by non-authorized service centers removing or destroying protective stickers which secure the balance’s housing against
unauthorized access
C. Warranty card must be filled out for warranty to be valid.
7
2. UNPACKING AND INSTALLATION
2.1 PLACE OF USE AND ASSEMBLING
The balance should be stored and used in locations free of vibrations and shakes, free of air
movement and dust.
Ambient air temperature should not exceed the range of: +10 °C ÷ +40 °C. Ambient relative humidity should not exceed 80%. During balance operation, ambient temperature in the weighing room should not change
rapidly.
The balance should be located on a stable wall console desk or a stable working table which
is not affected by vibrations and distant from heat sources.
Take special precaution when weighing magnetic objects, as part of the balance is a strong
magnet. Should such loads be weighed, use under-pan weighing option, which removes the
weighed load from area influenced by the balance’s magnet. The hook for under-pan weighing is installed in balance’s base.
Keep all package element should your device be transported in the future. Remember that
only original packaging can be used for shipping purposes. Prior to packing, uncouple any cables, remove any separable components (weighing pan, shields, inserts). Pack the device components into an original packaging. The original packaging protects the equipment against potential damage during transportation.
2.2 STANDARD DELIVERY COMPONENTS LIST
Balance and components shown in Section 2.4 depending on balance model Warranty Card USB
o User Manual o Balance USB Driver o RLAB Software o RLAB Software Manual
2.3 UNPACKING
To unpack the system, follow the diagram below-
8
2.4 BALANCE ASSEMBLY
Model: PA-T-225 Components:
Installation:
1) Remove the transport lock– gently press the transport lock and turn it accordingly to <OPEN>, keep the transport lock should your balance be transported in the future.
2) Check grounding spring to insure it is in the appropriate location. Make sure that the grounding spring juts slightly out of the hole.
9
3) Install components following diagram below:
i. Bottom insert
ii. Centering ring [embossment side up] iii. Weighing pan iv. Draft shield
Model: PA-T-1224, PA-T-224 Componets:
Installation:
1) Remove the transport lock– gently press the transport lock and turn it accordingly to <OPEN>, keep the transport lock should your balance be transported in the future.
10
2) Check grounding spring to insure it is in the appropriate location. Make sure that the grounding spring juts slightly out of the hole.
3) Install components following the diagram below:
i. Bottom insert
ii. Centering ring [embossment side up]
iii. Weighing pan
11
Model: PT-T-252, PT-T-753 Components:
Installation:
1) Remove transport lock, keep the transport lock should your balance be transported in the future.
2) Check that the grounding spring is in the correct location, the spring juts slightly out of the hole.
3) Install components following the diagram below:
i. Rubber feet (grounding foot can be in any location on the balance)
ii. Weighing pan
iii. Glass draft shield
12
Model: PT-T-802, PT-T-2002, PT-T-4502 Components:
Installation:
1) Remove transport lock, keep the transport lock should your balance be transported in the future.
2) Check that the grounding spring is in the correct location, the spring juts slightly out of the hole.
3) Install components following the diagram below:
i. Rubber feet (grounding foot can be in any location on the balance)
ii. Weighing pan
13
Model: PT-T-6002 Components:
Installation:
1) Remove transport lock, keep the transport lock should your balance be transported in the future.
2) Check that the grounding spring is in the correct location, the spring juts slightly out of the hole.
3) Install components following the diagram below:
i. Draft shield
ii. Rubber feet (grounding foot can be in any location on the balance)
iii. Weighing pan
14
2.5 BALANCE LEVELING
It is necessary to level the balance prior to plugging it in. To level the balance, turn its feet until the air bubble is in the center position.
The balance should firmly rest on a surface, each of the feet must be supported.
2.6 POWERING THE DEVICE
Before plugging in your balance, it is imperative to wait until the balance reaches thermal stabilization.
For balances that were stored in much lower temperatures (e.g. during winter period), thermal stabilization period will take at least 4 hours for Precision balances, and 8 hours for Analytical balances.
Balance should be plugged in only with the power adapter that comes standard with the
model. Nominal power supply of the power adapter (specified on the power adapter data plate) should be compatible to the power supply.
Plug the balance in – connect the power adapter to the socket, next connect its connector to
port located at the back of the balance housing.
Press button located in the top right hand corner of the key pad.
Remember to start the balance with no load on the weighing pan
Wait until the start-up procedure is completed. The home screen of balance software is
displayed automatically.
If the indication is different than zero, please press button
15
3. BALANCE CONTROL
3.1 BALANCE KEYBOARD
3.2 BALANCE HOME SCREEN
16
3.3 ENTERING BALANCE MENU
Operation of the balance menu is intuitive and easy to use. The touch panel makes the software operation easy. Pressing a function key, a soft key or an area on the display initiates an assigned function or process
To enter balance main menu, press <PARAMETERS> button. Clicking any button within this section, or clicking any button with a parameter name, results in a change of color. This is for signaling purposes. If a given area has any function or action assigned, then it is performed automatically upon clicking (e.g. adjustment procedure), respectively a window with parameters or a list of appropriate settings is displayed.
3.4 SCREEN SCROLLING
There are two methods for scrolling the screen of parameters window. The first one requires pressing, holding down and scrolling up or down the scrollbar located on the left. The second one requires pressing, holding down and scrolling up or down on any point of the displayed window.
17
3.5 SOFT KEYS LIST
Press to enter the main menu.
Press to clear the editing field.
Press to scroll the menu “up”, or “down”.
Press to enable / disable an on­screen keyboard.
Press to confirm changes.
Press to export databases (key active upon plugging a USB flash drive).
Press to resign form introducing function modifications.
Press to import databases (key active upon plugging a USB flash drive).
Press to add a new record to a database.
Press to select variables, out of the list, for a printout.
Press to search for a record in a database.
Press to move (exit) one level up.
Press to print out a record from a database.
4. WEIGHING MODE
Load an object on the balance weighing pan. The stabilization of weighing result is indicated by a stability marker visible on the left side of balance display, read the measurement result.
18
4.1 GOOD WEIGHING PRACTICE
To ensure long lasting use of a balance with correct and reliable measurements of weighed loads, follow the procedures below:
Start the balance with no load on the weighing pan. Load the weighing pan carefully and avoid dropping it:
Place weighed load in the center of the weighing pan:
Avoid side loading, in particular side shocks:
4.2 UNITS
To change the weighing unit, press the weighing unit icon. This brings up the available unit’s list, choose new unit from the list displayed.
19
Units List:
Unit
Denotation
Unit
Denotation
gram
[g]
Taele China
[tlc]
milligram
[mg]
Momme
[mom]
kilogram
[kg]
Grain
[gr]
carat
[ct]
Newton
[N]
pound
[lb]
Tical
[ti]
ounce
[oz]
baht
[baht]
ounce Troy
[ozt]
tola
[tola]
pennyweight
[dwt]
mesghal
[msg]
Taele Hongkong
[tlh]
User unit 1
[u1]
Taele Singapore
[tls]
User unit 2
[u2]
Taele Taiwan
[tlt]
4.3 USER-DEFINED UNIT
You may create two custom user-defined units. Displayed value of a user-defined unit is a multiplication of measured mass value and a coefficient specified for the user-defined unit. The units
can be freely named with use of 3 characters’ maximum. By default, the names are displayed as
[u1] user unit 1, and [u2] user unit 2.
20
4.4 BALANCE ZEROING
Zeroing is a function allowing to zero mass indication.
The Zeroing process is an equivalent for determining new zero point, recognized by the balance as precise zero. Zeroing is possible only for stable status of display indication.
CAUTION
Zeroing the display indication is possible only within ±2% range of instrument’s maximum
capacity. If the zeroed value is above ±2% of the maximum capacity, then the balance indicates a respective error message.
21
4.5 BALANCE TARING
Taring is a function allowing to determine net weight of a measured object. To determine net weight
of the object, place object’s container (packaging) on the weighing pan, and on stabilization of
measurement result press key.
The balance enables assigning tare value to a database-stored product. Using this option, the software automatically uploads data on tare value for a product upon its selection from the database.
CAUTION
Taring negative values is impossible. On taring negative values, the balance responds with an error message. In such case, zero balance indication and repeat taring procedure.
Manual tare determination
Press quick access key
An on-screen numeric keyboard is displayed
Enter tare value and press key
The balance returns to the weighing mode, and the display indicates entered tare value with
minus ‘–‘ sign
Deleting tare
The tare value indicated on balance display can be deleted by pressing key on balance overlay.
Procedure 1 - on taking the tared load off the weighing pan
Press key.
The NET marker is deleted, and new zero point of the balance is determined.
22
Procedure 2 – with tared load on the weighing pan
Press key. The NET marker is deleted, and new zero point of the balance is determined. When tare value exceeds 2% of the maximum capacity, respective message is displayed to
inform a user about the fact.
Selecting tare value out of TARE DATABASE Procedure:
While in optional mode, press button located in a top right hand corner of the mass
display.
Wait for a respective window to open, next select <PACKAGING>. Window with a list of tare values recorded into tare database opens. Select the packaging that is to be used. The balance returns to the weighing mode, and the display indicates selected tare value
with a minus ‘–‘ sign.
OR
While in optional mode, press button (if displayed on the screen), Window with a list of tare values recorded into tare database opens. Select the packaging that is to be used. The balance returns to the weighing mode, and the display indicates selected tare value
with a minus ‘–‘ sign.
4.6 WEIGH MODE SETTINGS
The balance allows setup of operating parameters (filters, value release and autozero function, deleting the last digit and other settings) separately for each working mode.
Procedure:
23
Filter level setting
Filter settings adjustment depends on the working environment. For the best possible conditions the filter can work in a very fast mode (V.FAST value for Filter parameter); however, if the conditions are poor (shakes, drafts), the filter should be set to slow or very slow option (SLOW or V. SLOW value for Filter parameter). The effectiveness of the filter is different throughout the weighing range.
The filter works slower when “approaching” the weighed mass, it works more quickly for weighed
mass within the set range of the filter (parameter for setting filter range is accessible only from the service menu – the user does not have any access to it).
Depending on the filter, the weighing time is shorter (V.FAST and FAST) or longer (SLOW and V. SLOW).
CAUTION! The higher filter level, the longer the weighing time.
Value release
Since ambient conditions at a workplace vary, it is necessary to determine the value release parameter that are best for your working environment, parameter options are: FAST.+REL., FAST or RELIABLE. Depending on the selected option, weighing time is either shorter or longer.
Autozero function
The balance features an autozero function (Auto). This function automatically controls and corrects the zero reading. When Autozero is enabled, it compares balance readings at declared time interval e.g. 1s, if weighing pan is unloaded and display indication is close to zero. If results vary less than declared AUTOZERO range e.g. one division, balance zeroes automatically, marker of stable
measurement result , and precise zero marker are displayed. If AUTOZERO function is enabled, then each weighing process starts from precise zero point.
There are, however, some instances when this function can be a disturbing factor for the measuring process; e.g. very slow placing of a load on the weighing pan (load adding). Here, zero reading correction can also correct the actual reading of loaded mass.
24
Last digit display
Function enables displaying the last digit of decimal place for a weighing result. There are three available options:
Always: all digits visible Never: last digit is not displayed When stable: last digit is displayed only for a stable weighing result
Balance ambient conditions
Parameter relating to ambient and environmental conditions in which the balance operates. There are two options: STABLE and UNSTABLE. Selecting STABLE mode makes the balance work much faster, i.e. weighing takes much less time than for UNSBABLE mode. If the ambient conditions are unstable it is recommended to use UNSTABLE mode. By default, the parameter is set to STABLE option.
5. ADJUSTMENT
To ensure the highest weighing accuracy, it is recommended to periodically introduce a corrective factor of indications to balance memory, the said factor must be referred to a mass standard. In other words, balance adjustment shall be performed from time to time.
Adjustment should be carried out:
Before the beginning of weighing procedure
If long breaks between following measuring series occur
If temperature inside the balance changes more than: 1°C or 2°C for Analytical balances or 2°C
for Precision series balances.
Types of adjustment:
Internal automatic adjustment
Manual internal adjustment
Adjustment with an external weight of declared mass which cannot be modified or of any mass,
but not lower than 30% of maximum range.
5.1 INTERNAL AUTOMATIC ADJUSTMENT
This menu is for declaring a value initiating an automatic internal adjustment. Accessible options:
None – automatic adjustment disabled Time – adjustment takes place in time intervals declared in menu <Automatic adjustment
time>. The time interval is declared in hours and ranges between 0.5 and 12 hours.
Temperature – adjustment is triggered by temperature change only Both – adjustment is triggered by both, temperature changes and time.
CAUTION! Parameter settings can be modified only for balances that are not a subject to conformity assessment (verification).
25
5.2 MANUAL INTERNAL ADJUSTMENT
Internal adjustment is carried out with of an internal adjustment weight. button, when pressed, automatically triggers adjustment process. Upon adjustment process completion, respective message informing about process end and about its status is displayed.
CAUTION! Adjustment procedure requires stable environmental conditions (no air drafts or ground vibrations). The process must be carried out with an empty weighing pan.
5.3 EXTERNAL ADJUSTMENT
External adjustment is carried out by means of an external mass standard of specified accuracy class and weight. Both, accuracy class and mass standard weight depend on balance type and max capacity. The process takes semi-automatic form; successive stages are signaled with prompts.
Process stages:
Enter <Adjustment> submenu, next select “External adjustment” option, “Remove weight” prompt is displayed.
Take the weight off the weighing pan and press button. Whereas balance
determines start mass, “Adjustment; Please wait…” prompt is displayed,
Upon completed start mass determination procedure “Put weight …” prompt is displayed
along with particular mass standard value.
Put the required weigh on a pan and press button, Upon completed procedure “Remove weight” prompt is displayed. Take the weight off the weighing pan, wait for <Adjustment> window to be displayed
again.
5.4 ADJUSTMENT TEST
<Adjustment test> function enables comparing the result of an internal automatic adjustment with the value recorded in balance factory parameters. The comparison is used for determining balance sensitivity drift over time.
26
6. WORKING MODES- GENERAL INFORMATION
The balance features the following working modes:
Weighing
Means of operation: weight of a load is determined through an indirect measurement. A balance measures gravitational force which attracts the load. An obtained result is processed to a digital format and displayed in a form of measurement result. More information found in Section 4.
Parts Counting
Means of operation: based on a determined mass of a single part it is possible to count another part, if the mass of the single part is determined with sufficient accuracy, and that the following parts are equal in mass. More information found in Section 7.
Check Weighing
Means of operation: control of sample mass with applied thresholds. A user should specify the value of min threshold <LO> and max threshold <HI>. More information found in Section 8.
Dosing
Means of operation: a user should specify sample’s target mass to be obtained by
pouring. More information found in Section 9.
Percent Weighing
Means of operation: control of percent ratio of a sample in relation to a standard (reference). Obtained data provides percent ratio on how test sample differs from the accepted standard (reference). More information found in Section 10.
Density of solids
Means of operation: based on Archimedes principle, a balance determines density of solids. The mode requires an optional density determination kit. More information found in Section 11.
Density of liquids
Means of operation: based on Archimedes principle, a balance determines density of liquids. The mode requires an optional density determination kit. More information found in Section 12.
Animal Weighing
Means of operation: mass measurement takes place with application of filters dampening animal moves on a weighing pan, thus enabling obtaining a correct measurement result. More information found in Section 13.
Statistics
Means of operation: carried out measurements are used to calculate statistical data, such as Min, Max, deviation, etc. More information found in Section 14.
Peak Hold
Means of operation: max temporary indication occurring in course of the weighing process is hold on a display. More information found in Section 15.
Formulations
Means of operation: by mixing specified ingredients you can obtain particular mixture, in order to program given formulation you have to specify weight of particular ingredients. More information found in Section 16.
27
6.1 RUNNING WORKING MODE
To run working mode other than the one currently operating:
Press pictogram of currently used working mode, the pictogram is in the top left hand
corner
Wait for the available working modes list to be displayed Select the working mode you need to operate.
6.2 QUICK ACCESS KEYS FOR EACH WORKING MODE
Key
Function
Modes featuring the key
Accept/Print
All modes
Print header
All modes
Print footer
All modes
Zero
All modes
Tare
All modes
Change unit
All modes (Parts Counting and Percent Weighing modes excluded)
Select unit
All modes (Parts Counting and Percent Weighing modes excluded)
Parameters
All modes
Databases
All modes
28
User
All modes
Product
All modes
Customer
All modes
Packaging
All modes
Adjustment
All modes Variable 1
All modes
Variable 2
All modes
Variable 3
All modes
Hide/show last digit
All modes (Parts Counting and Percent Weighing modes excluded)
Working mode parameters
All modes
Set part mass
Parts Counting mode exclusively
Determine part mass
Parts Counting mode exclusively
Determine mass using 5 parts
Parts Counting mode exclusively
Determine mass using 10 parts
Parts Counting mode exclusively
Determine mass using 20 parts
Parts Counting mode exclusively
Determine mass using 50 parts
Parts Counting mode exclusively
Determine mass using 100 parts
Parts Counting mode exclusively
Thresholds
Check weighing mode exclusively
29
Min threshold
Check weighing mode exclusively Max threshold
Check weighing mode exclusively
Target value
Dosing mode exclusively
Set reference mass
Percent Weighing mode exclusively
Determine reference mass (set as 100%)
Percent Weighing mode exclusively
Start
Solids and Liquids Density modes and Formulations mode
Results
Statistics mode exclusively
Finish
Statistics mode exclusively
Finish
Peak Hold mode exclusively
Formulation
Formulations mode exclusively
7. PARTS COUNTING
< Parts counting> is a working mode which enables determination of quantity of small parts with
equal weight. Determined mass of a single part is used for the counting procedure. The single part mass may be:
Determined from a reference quantity of parts subjected to weighing, Acquired from database of products, Entered manually as a numeric value.
Procedure for activation of Parts Counting mode
Select Parts counting> mode, home screen featuring parts counting pictogram in a top bar is automatically displayed.
30
Info field provides the following information:
Gross weight Tare weight Net weight Part mass Product User
7.1 SUPPLEMENTARY SETTINGS FOR PARTS COUNTING MODE
The supplementary settings enable adjusting the working mode to your needs and requirements.
ACAI, Automatic Accuracy Correction, provides the following options:
YES, single part mass is updated NO, single part mass is not updated
Means of operation of ACAI function:
1. Number of pieces (on adding) on balance weighing pan has to be greater than before.
2. Number of pieces (on adding) on balance weighing pan must be less than twice the amount displayed before adding the pieces.
3. Current quantity of parts must be within 0,3 tolerance of the total value.
4. Measurement result has to be stabilized.
Procedure: