Cole-Parmer PC100 Instruction Manual

PC100 Handheld pH/Cond./TDS/Sal. Meter
Instruction Manual
Table of Contents
1
2. Specifications -------------------------------------------------------------------------------------------------------------------------
3. Instrument Description -------------------------------------------------------------------------------------------------------------
3.1 LCD display----------------------------------------------------------------------------------------------------------------------
3.2 Keypad functions ---------------------------------------------------------------------------------------------------------------
3.3 Meter sockets -----------------------------------------------------------------------------------------------------------------
Reading stable display mode ----------------------------------------------------------------------------------------------
3.4
3.5 Auto power-off --- -----------------------------------------------------------------------------------------------------------------
4. pH measurement ---------------------------------------------------------------------------------------------------------------------
4.1. pH probe information---------------------------------------------------------------------------------------------------------
4.2. pH calibration related information---------------------------------------------------------------------------------------------
4.3. pH Meter calibration -------------------------------------------------------------------------------------------------------------
4.4. Customer-defined calibration ---------------------------------------------------------------------------------------------------
4.5. Solution Measurement -----------------------------------------------------------------------------------------------------------
4.6. pH probe maintenance-----------------------------------------------------------------------------------------------------
2
3
3
4
5
5
5
6
6
6
7
8
9
10
5. mV measurement---------------------------------------------------------------------------------------------------------------------
6. Conductivity measurement - ------------------------------------------------------------------------------------------------------
6.1. Conductivity probe information-------------------------------------------------------------------------------------------
6.2. Conductivity calibration related information---------------------------------------------------------------------------------
6.3. Conductivity Meter calibration--------------------------------------------------------------------------------------------------
6.4. Relationship among TDS, Salinity, and Cond.--------------------------------------------------------------------------
6.5. Customer-defined calibration ---------------------------------------------------------------------------------------------
6.6. Sample measurement------------------------------------------------------------------------------------------------------------
6.7. Conductivity probe maintenance-----------------------------------------------------------------------------------------
7. Parameter setting--------------------------------------------------------------------------------------------------------------------
7.1. Main menu---------------------------------------------------------------------------------------------------------------------
7.2. Submenu -----------------------------------------------------------------------------------------------------------------------
7.3. Submenu of pH parameter setting ---------------------------------------------------------------------------------------
7.4. Submenu of Cond. parameter setting------------------------------------------------------------------------------------
7.5. Basic parameter setting submenu-----------------------------------------------------------------------------------------
11
12
12
12
14
14
15
15
16
17
17
17
19
20
21
8. Meter Kits----------------------------------------------------------------------------------------------------------------------------
9. Warranty------------------------------------------------------------------------------------------------------------------------------
Appendix I: Parameter setting & Restore to factory default setting-------------------------------------------------------
Appendix II: Code Symbol & Abbreviation Glossary--------------------------------------------------------------------------
Appendix III: Self-diagnostic Information ----------------------------------------------------------------------------------------
21
23
24
25
26
1. Introduction
Thank you for purchasing our PC100 Handheld pH/Cond./TDS/Sal. Meter.
This meter is a perfect combination of the advanced electronic technology, sensor technology and
software design, and is the most cost effective handheld electrochemical meter which is suitable for
usage in industrial and mining enterprises, water treatment engineering, environmental protection
industry, etc, especially suitable for application in field.
With built-in microprocessor chip, beautiful appearance and easy to use, this meter has the following
prominent features:
1.1 Basic features
z With built-in microprocessor chip, the meter has intelligent functions like automatic calibration,
automatic temperature compensation, function setting, self-diagnostic information, auto power-off
and low voltage display, etc.
z With advanced digital processing technology, the response speed and accuracy of the meter are
greatly improved. Stable reading display mode is also equipped.
z The meter meets IP67 dust and water proofing rate.
1.2 pH measurement features
z 1-3 point automatic calibration with calibration instruction and automatic checking functions.
z Automatically recognize 6 types of pH buffer solution. Buffer solution selectable: USA series, NIST
series and customer-defined solution.
z The criteria of the reading stability are selectable.
1.3 Conductivity measurement features
z 1- 4 point automatic calibration with calibration instruction and automatic checking functions.
z Automatically recognize 4 types of conductivity standard solution and customer-defined solution can
also be set up.
z With conductivity, TDS and salinity three measurement modes, can switch to display the results.
1
Specifications
2.1 Technical Parameters:
Technical Parameters
pH
mV
Cond.
Measuring Range
0.00 ~ 14.00pH
Resolution 0.1/0.01 pH
Accuracy ±0.01 pH ±1digit
Temperature
Compensation Range
0 ~ 100°CAuto. or Manual
Measuring Range ±1000mV
Resolution 1mV
Accuracy ±0.1% FS ±1digit
Conductivity: 0~200 mS/cm(divided into four ranges):
0~199.9μS/cm ; (200~1999) μS/cm ;
Measuring Range
2.00~19.99mS/cm; 20.0~199.9mS/cm
TDS: (0~100)g/L; Salinity: (0~100)ppt
Resolution 0.1/1μS/cm 0.01/0.1 mS/cm
Accuracy ±1.0% FS ±1digit
Temperature
Compensation Range
0 ~ 80°Cmanual or automatic
Cell constant 0.1 / 1 / 10 cm-1
Tem p.
Measuring Range
Resolution 0.1 °C
0~100 °C
Accuracy ±0.5°C±1digit
1.2. Other technical parameters:
Power AA batteries × 3 (1.5V× 3)
IP rating IP67 water-proof and dust-proof
Dimension & Weight Meter: (86×196×33 )mm / 335g
2
2. Instrument Description
3.1. LCD display
Diagram-1 LCD Display Screen
Diagram-1 LCD Display
1 — Measuring mode icons
2 — Measurements
3 — Prompts
4 —Measurement Units
5 —Temperature unit
6 — pH and Cond. calibration units
7 —Calibration value and prompts
8 — Temperature value and prompts
9 — Temperature compensation state icons
ATC — automatic temperature compensation,
MTC — manual temperature compensation
10 — Calibration indication icon
11 — Stable reading indication icon
12 — Low battery icon, when this icon appears, please renew the battery
3
3.2. Keypad functions
Diagram-2 Keypad panel
3.2.1. Keypad operations: Short press ----- <1.5 seconds , Long press ----- >1.5 seconds.
3.2.2. Turn on the meter: Press to turn on the meter: LCD full display display some parameters
setting content display the last measuring mode before turning off
3.2.3. Turn off the meter
In the measurement mode, press and hold for 2 seconds to turn off the meter.
Note: In the calibration mode or the parameter set-up mode, pressing is invalid. Please
press key to return to the measurement mode, then press to turn off the meter.
Chart -1 Keypad operations and functions
Keypad Operations Functions
Short press In the power-off mode, press this key to turn on the meter
Long press In the measurement mode, press and hold this key for 2 seconds to turn off the meter.
In measurement mode: Press this key to select measurement mode:
Short press
and
Long press
Short press
In the mode of manual temperature compensation (MTC), when press and hold this key,
the temperature value flashes, then press this key to change the temperature value (in
one direction), and press to confirm.
In the parameter set-up mode, press this key to change the serial number of the main
menu and the submenu (in one direction).
In the submenu mode, press this key to change parameters and set-up (in one direction).
In the measurement mode, press this key to enter in the calibration mode
In the calibration mode or the parameter set-up mode, press this key to return to the
measurement mode.
Short press
In the measurement mode, press this key to enter in the parameter set-up main menu.
In the calibration mode, press this key to conduct calibration.
In the parameter set-up mode, press this key to select programs.
4
3.3. Meter Sockets
The meter is with BNC and RCA sockets, which are protected by grey rubber caps. (As showed in
Chart-2)
Chart-2- Meter sockets
Photos Descriptions
BNC socket (Right) — for pH or ORP probe
BNC socket (Left) — for Cond. probe
RCA socket (Middle) — for temperature probe. When both pH probe
and Cond. probe are connected to meter, just plug in temperature
plug of either probe.
3.4. Reading stable display mode
When the measuring value is stable, smiley icon appears on LCD screen,
see Diagram – 3. If the smiley icon does not appear or flash, please do
not get readings or make calibration until the measuring value is
stable. In parameter-setting P1.6, there are 3 criteria for stability:
(Normal), (High) and (Low). The factory default set is “Normal”.
“High” is set for stability for longer time; “Low” is set for stability for shorter
time. User can select suitable stability criteria according to different testing
requirements.
Diagram- 3
3.5. Auto power-off
The meter will be automatically power-off if nothing is done for 20 minutes. In the parameter setting
menu P6.8, user can set up to turn on or turn off this function.
5
4. pH measurement
4.1. pH probe information
The meter is equipped with a 201T-F three-in-one combination pH probe, which has built-in temperature
sensor to realize automatic temperature compensation. The BNC plug of the probe is connected to the
pH socket while the RCA plug is connected to the temperature socket. When the probe is immersed in
the solution, please stir the solution briefly to get rid of the air bulb and in this way, a stable measurement
will be reached fast.
4.2. pH calibration related information
4.2.1. Standard buffer solution
Buffer solution selectable: USA series and NIST series, and also user-defined solution. Please refer to
clause 7.3 and select a proper series in parameter setting menu P1.1. Please see Chart - 3 for the two
series of standard buffer solution.
Chart - 3 pH standard buffer solution series
pH standard buffer solution series
Calibration indication icons
USA series NIST series
Three-point calibration
4.2.2. Three-point calibration mode
The instrument can perform 1-3 point calibration. In three-point calibration mode, the first point
calibration must use 7.00 pH (or 6.86 pH) standard solution, then select other standard solution to
perform the second and the third point calibration. See chart – 4. During the calibration process, the
instrument displays the probe slope of acidity range and alkalinity range respectively.
Chart - 4 Three-point calibration mode
USA standard NIST standard
One-point calibration 7.00 pH 6.86 pH
7.00 pH and
Two-point calibration
4.00 pH or 1.68pH
7.00 pH and 10.01 pH 6.86 pH and 9.18 pH
1.68pH and 4.00 pH 1.68pH and 4.01 pH
7.00 pH 6.86 pH
10.01 pH 9.18 pH
Indication
Icon
Applicable
range
Accuracy
±0.1pH
6.86 pH and
4.01 pH or 1.68pH
Range<7.00pH
Range>7.00pH
Three-point calibration
7.00 pH, 4.00 pH or
1.68pH, 10.01 pH
6.86 pH, 4.01 pH or
1.68pH, 9.18 pH
6
Large Range
4.2.3. Calibration intervals
Calibration intervals depend on the sample, the probe performance, and the required accuracy. For high
accuracy measurements ( ±0.02pH), the meter should be calibrated before taking a measurement. For
general accuracy (±0.1pH), after one time calibration, the meter can be used for approximately one
week or longer.
The meter must be recalibrated in the following situations:
(a) New probe or probe that is unused for a long period of time
(b) After measuring acids solution (pH<2) or alkaline solutions (pH>12)
(c) After measuring solution that contains fluoride or strong organic solution
(d) If the solution’s temperature differs greatly from that of the calibration solution
4.3. pH meter calibration ( take three-point calibration as an example)
4.3.1. Press key to enter into the calibration mode, “CAL 1” blinks at the
top right of LCD screen and “7.00 pH ”blinks at the bottom right of LCD screen,
indicating using pH 7.00 buffer solution to make the 1
st
point calibration.
4.3.2. Rinse pH probe in pure water, allow it to dry, and submerge it in pH7.00
buffer solution. Stir the solution briefly and allow it to stay in the buffer solution
until a stable reading is reached. The scanning and locking process of buffer
solution will be showed at the bottom right corner of LCD screen. Er 2 will be
showed if press key before the value is locked. See chart – 5.
4.3.3. When the meter locks 7.00 pH, stable icon will appear on LCD screen.
Press key to calibrate the meter. End icon appears after calibration is done.
st
The 1
point calibration is finished, meanwhile, the meter’s display will show at
the top right a blinking CAL2, and show at the bottom right a blinking 4.00pH and
10.01pH alternately, indicating using pH4.00 or pH10.01 buffer solution to make
nd
the 2
point calibration.
4.3.4. Take out pH proberinse it in pure water, allow it to dry, and submerge
it in pH4.00 buffer solution. Stir the solution briefly and allow it to stay in the buffer
solution until a stable reading is reached. The meter’s screen will show scanning
and locking process of calibration buffer solution at the bottom right of LCD. When
the meter locks 4.00 pH, stable icon displays on LCD. Press key to
calibrate the meter. End icon and probe slope of acidity range will be displayed
after calibration is done, meanwhile, the meter’s screen will show at the top right
a blinking CAL3, and show at the bottom right blinking 10.01pH, indicating using
pH10.01 buffer solution to make the 3
rd
point calibration.
7
Diagram- 4
Loading...