Cole-Parmer PT-T-353, PT-T-2002, PT-T-4502, PT-T-753, PT-T-6002 Quick Start Guide

...
Start up Guide
PA-T and PT-T
Advanced Series Balances
To view a complete user manual scan QR code.
1. CONTENT
Models: PA-T-225
Balance
Models: PA-T-124 | PA-T-224
Balance
Models: PT-T-353 | PT-T-753
Balance
Weighing pan
Weighing pan
Weighing pan
x 1
Draft shield
Draft shield
Draft shield
Centring ring
Bottom insert
Grounding foot
Bottom insert
Power adapter
Foot
x 3
Power adapter
Power adapter
Models: PT-T-802 | PT-T-2002 | PT-T-4502
Balance
Models: PT-T-6002
Balance
Weighing pan
x 1
Weighing pan
x 1
Grounding foot
Draft shield
x 1
Foot
Grounding foot
Power adapter
Foot
Power adapter
2. WORKROOM AND BASIC OPERATION GUIDELINES
o
C
+40 +104oF
o
+10
C
o
F
+50
Place the balance away from heat sources. Avoid exposing the balance to the sunlight.
Operate the device in a room where the temperature ranges between 10–40 °C (50–104 °F) and where the relative humidity is below 80%.
Place the balance on a solid surface to ensure stability. To obtain stable and repeatable weighing results, an anti-vibration table is recommended.
Avoid exposing the balance to a magnetic field.Do not weigh magnetic substances.
Do not place the balance in a hazardous area. Do not weigh explosive materials.
Avoid air drafts and air movements at the workstation.
Make sure that the balance is placed on an even surface.
Do not place the balance on an unstable ground exposed to shocks and vibrations.
3. UNPACKING
1 2 3
TOP
Keep the packaging should a warranty claim or service be required.
4. ACTIVITIES TO BE DONE PRIOR TO OPERATION
4.1. Remove transport lock
Models: BA-T-64 | BA-T-124 | BA-T-214
1
Models: BT-T-203 | BT-T-403 | BT-T-503 | BT-T-802 | BT-T-2002 | BT-T-4002 | BT-T-6001 | BTT-10001 BT-T-20001
2
Keep the transport lock should a warranty claim or service be required.
1 2
Keep the transport lock should a warranty claim or service be required.
4.2. Grounding spring check
Models: BA-T-64 | BA-T-124 | BA-T-214
Check the grounding spring. Make sure that the grounding spring juts slightly out of the hole.
Models: BT-T-203 | BT-T-403 | BT-T-503 | BT-T-802 | BT-T-2002 | BT-T-4002 | BT-T-6001 | BTT-10001 BT-T-20001
Check the grounding spring. Make sure that the grounding spring juts slightly out of the hole.
Grounding foot