Cole-Parmer Cond 500 Instruction Manual

Operating Instructions
Cole-Parmer CON 500
Conductivity/TDS/Temperature Controller
625 East Bunker Court, Vernon Hills, Illinois 60061 Tel: 1-847-549-7600 or Toll-free: 1-800-323-4340
Fax: 1-847-247-2949 www.coleparmer.com
CONTENTS page I. INITIAL INSPECTION AND ASSEMBLY 2 II. GENERAL INTRODUCTION 3 III. USING THE CON 500 4 A.FRONT PANEL 4 B.NORMAL MODE DISPLAY 5 C.REAR CONNECTORS 7 D.TURNING ON/OFF THE INSTRUMENT 8 IV. MODEL CON 500 MODES 9
A. MAIN DISPLAY MODE 9 B. CALIBRATION/SETTING MODE 11
a.PASSWORD CHECK PAGE 12 b.USER SETTING PAGE 12 c.CONDUCTIVITY CALIBRATION PAGE 13 d.CONDUCTIVITY CONTROL SETTING PAGE 16 e.CURRENT SETTING PAGE 17 f.TEMP. CONTROL SETTING PAGE 17
V. CONTROLLING THE RELAYS 18
A. RELAY ISOLATION VOLTAGE 18 B. RELAY OUTPUT LOAD 18 C. RELAY ACTION, SETPOINT, HYSTERESIS
MODE &HYSTERESIS VALUE 18 D. CONDUCTIVITY/TDS RELAYS 19 E. TEMPERATURE RELAY 19 VI. 4-20 mA OUTPUT 20 A. ISOLATION VOLTAGE 20 B. OUTPUT LOAD 20 C. CONDUCTIVITY OR TDS LINEAR CURRENT
OUTPUT 20 VII. RS485 INTERFACE OPERATION 21 A. INTRODUCTION 21 B. PREPARING THE METER 21 VIII.ERROR DISPLAYS AND TROUBLESHOOTING 22 IX. SPECIFICATION 23 X. WARRANTY 24 XI. RETURN OF ITEMS 24
1
I. INITIAL INSPECTION AND ASSEMBLY
Carefully unpack the instrument and accessories. Inspect for damages made in shipment. If any damage is found, notify your Cole-Parmer representative immediately. All packing materials should be saved until satisfactory operation is confirmed.
MOUNTING PROCEDURE
1. Make a cutout on any panel, with a thickness of 1/16 in. (1.5 mm)
to 3/8 in. (9.5mm).
2. Remove the mounting assembly from the controller and insert the
controller into the cutout.
3. Replace the mounting bracket assembly onto the controller and
secure the controller to the mounting panel.
PANEL CUTOUT DIMENSIONS
0.04"R(1.0mm) MAX.
0.04
3.66"
93.00mm?
0.00
1.00
0.00
3.66"
93.00mm?
0.04
0.00
1.00
0.00
Warning:
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
2
Cleaning the instrument:
1. Be sure to remove the power before attempting to clean the meter.
2. Use a lint free cloth and clean water or neutral detergent.
3. Wipe the outer surface of the instrument only.
4. Wipe-dry the instrument before powering again.
II. GENERAL INTRODUCTION
The Cole-Parmer CON 500 (Conductivity and Temperature) System, is a rugged microprocessor based instrument assembled in a watertight ¼ DIN case, designed for use in laboratories and process control applications.
The model CON 500 microprocessor allows the user to easily recalibrate the parameters for the probes. The Conductivity system requires only a single calibration. The microprocessor also performs a self-diagnostic routine every time you turn on the unit providing you with basic information about the stability of the instrument.
The system simultaneously displays Conductivity, Temperature, relay status and mA output in one LCD graphic screen. This system uses a two-wire or four-wire type sensor for the conductivity and a precise thermistor for temperature, providing you with accurate readings for all your measurements.
The model CON 500 is equipped with three relays: two programmable High/Low control relays for conductivity/TDS and one programmable High/Low control relay for temperature. All relays are hysteresis driven and configurable to CENTER or EDGE mode. The system also has an isolated 4-20mA analog output, offset and span configurable for Conductivity/TDS display.
The model CON 500 comes with a RS485 interface which can easily let the user log all data with an IBM© PC/AT compatible computer.
3
III. USING THE CON 500
A. FRONT PANEL
1. The [MODE/-] key.
1a. In Normal mode, pressing this Key will change the display to Conductivity or TDS.
1b. In Calibration/Setting mode, pressing this key will move to the next digit of the current active parameter.
1c. In Calibration/Setting mode, pressing this key for 2 seconds will move you back to the previous parameter. Pressing this key for 2 seconds at the first parameter of the page will move you back to last parameter of the previous page.
2. The [CAL/VIEW] key.
2a. During any normal mode, pressing this key will switch to first Calibration/Setting screen.
2b. During Calibration/Setting mode, pressing this key will switch to the next available Calibration/Setting page. Pressing this key at the last User/Calibration page (TEMP. CONTROL SETTING page) will return the user to the Normal mode.
4
3. The [] UP key.
456
During Calibration/Setting mode, pressing this key will increment the current blinking digit of the active parameter.

4. The [] DOWN key.
During Calibration/Setting mode, pressing this key will decrement the blinking digit of the active parameter.
5. The [] ENTER key.
During Calibration/Setting mode, pressing this key will save the current modified parameter and move to the next parameter on the page. If the parameter is the last one on the page then it will move to the first parameter on the next available page.
6. The [ ] LIGHT key.
Pressing this key will turn on or turn off the backlight of the LCD. The backlight will automatically turn off if there is no key activity within two minutes.
7. LCD screen.
B. NORMAL MODE DISPLAY
1
2
7
8
3
9
1. CELL – actual cell constant of the probe computed from the last cell constant calibration. In TDS display it will show the TDS factor instead of the cell constant.
5
2. CURRENT OUT - this will display the actual output of the isolated 4-20 mA output. At POWER-ON this will show “OFF” for about 3 seconds before going to
normal operation. After going to the Calibration/Setting pages when the unit is not password locked this will show ”FROZEN” for about 3 seconds before returning to normal operation.
3. ( ) key icon annunciator - this will be displayed if the Calibration /Setting pages are locked, meaning the user will not be able to change the values unless
the correct 4 digit number has been entered at the PASSWORD CHECK page.
4. H1 or L1 annunciator -
displayed then the relay is ON. (H stands for high action control and L stands for low action control)
this is the status of Conductivity/TDS Relay 1, if this is
5. H2 or L2 annunciator – this is the status of Conductivity/TDS Relay 2, if this is displayed then the relay is ON.
6. H3 or L3 annunciator – this is the status of the Temperature Relay 3, if this is displayed then the Relay is ON.
7. TC- user selected temperature coefficient of the conductivity/ TDS solution.
8. Conductivity or TDS display.
9. Temperature display.
6
C. REAR CONNECTORS
Before wiring the probes, relays, analog output, RS485 and the power cord, be sure that you are connecting to the right terminal as shown above (be sure the ac-plug is not plugged-in). Remember that the unit is ON once the user plugs in the power cord to an AC power supply.
1. Connect the AC line to the rear of the instrument. The model can be used with 115 or 230 VAC 50/60 Hz. Power consumption is 6 watts. Make sure the EARTH connector is connected to the earth lead of the AC power line.
2. Connect the proper load to the output relays. Make sure that the load does not exceed the relay rating, 5 Amp at 115 VAC and 2.5 Amp at 230 VAC.
3. Set the proper load to the 4-20mA output connector. Make sure that the load impedance is less than 500 Ohms.
4. If the user uses a two-wire conductivity probe, the user needs to short terminal D1 with terminal S1 and terminal D2 with terminal S2.
CAUTION: Make sure you connect the AC power cord to the correct AC
terminals. Connecting incorrectly may damage the unit permanently.
7
D. TURNING ON/OFF THE INSTRUMENT
By just plugging the unit to a correct AC voltage, the unit will be ready for use. There is no Power key so unplugging and plugging the unit will turn OFF or turn ON the unit respectively. After the unit is turned on, it will perform some basic self diagnostics and will display “OK” or “BAD”. If you received any “BAD” messages, turn OFF the unit and turn it ON again. (see VIII. ERROR DISPLAYS AND
TROUBLESHOOTING). If the message persists, then you might need to call our Applications Specialists. After the self-diagnostic is complete, the temperature will be displayed on the
lowest part of the screen and you are ready to make conductivity or TDS measurements. Just immerse the probes halfway to the liquid. If possible do not allow the probes to touch any solid object in the solution. There should be no air bubbles around the probes either. Shaking or moving the probes vigorously before recording any measurement will dislodge any bubbles formed in the probes.
PROBE HANDLING AND MAINTENANCE
REMINDER: If the user uses a two-wire conductivity probe, the user needs to
short terminal D1 with terminal S1 and terminal D2 with terminal S2.
Handling the probe
To ensure accurate and repeatable results:
1. The probe (cell) must be clean before making any measurements. If measuring low conductivity solutions, extraordinary cleanliness may be necessary.
2. The probe cell plates must be totally submerged in the solution. The electrode chamber must be free of trapped air or particulates. You may need to tap the probe gently to dislodge any air bubbles.
8
Loading...
+ 18 hidden pages