Cole-Parmer C100 Quick Manual

d
t
th
C100 Handheld Cond./TDS/Sal. Meter
Quick Manual
1. Keypad operations:
Keypad Operations Functions
Short press In the power-off mode, press this key to turn on the meter
Long press
In the measurement mode: Press this key to select measurement mode:
In the mode of manual temperature compensation (MTC), when press and hold this key,
Short/long
press
the temperature value flashes, then press this key again to change the temperature
value (only one direction), and press to confirm
In the parameter set-up mode, press this key to change the serial number of the main
menu and the submenu (only one direction)
In the submenu mode, press this key to change parameters and set-up (only one
direction)
Short press
In the measurement mode, press this key to enter in the calibration mode
In the calibration mode or the parameter set-up mode, press this key to return to the
measurement mode
Short press
In the measurement mode, press this key to enter in the parameter set-up main menu
In the calibration mode, press this key to make calibration
In the parameter set-up mode, press this key to select programs
2. Conductivity meter calibration and measurement
Press to turn on
the meter.
Press key to
enter into the
calibration mode.
Submerge the electrode in 84 μS solution, press key when icon appears, the 1st point
calibration is done, the meter enters into the
measurement mode automatically.
Submerge the electrode in 1413 μS solution,
press key when icon appears, the
nd
2
point calibration is done, the meter enters
into the measurement mode automatically.
Submerge the electrode in 12.88mS solution,
The standard solution Range
84 μS/cm 0~200 μS/cm
1413 μS/cm 200~2000 μS/cm
12.88 mS/cm 2~20 mS/cm
111.9 mS/cm 20~200 mS/cm
press key when icon appears, the
3 rd point calibration is done, the meter enters
into the measurement mode automatically.
Submerge the electrode in 111.9mS solution,
press key when icon appears, the
4th point calibration is done, the meter enters
into the measurement mode automatically.
s
1
point calibration
n
2
point
calibration
rd
3
point
calibration
4
point calibration
Sample
test
Notes
1) The meter can perform 1~ 4 point calibration at random
2) Normally perform 1 point calibration with 1413μS standard solution.
3. Parameter setting and factory default setting
Modes Prompts Parameter setting items
P2.1 Select probe constant
Select conductivity
standard solution
Select reference
temperature
Select temperature
compensation
P2.0
Cond.
P2.2
P2.5
P2.6
coefficient
P2.7 Select TDS coefficient
Restore to factory
default setting
P6.0
P2.8
P6.1 Select temperature unit
Basic
Parameters
P6.8
Automatic Power-off
setup
Code and
Abbr.
1.010.00.1
/ °C°F
Restore to
Description
factory default
1.0
StdCUS
Std
15~30 °C 25 °C
0.00~9.99 0.19
0.40~1.00 0.71
NoYes
No
-
OnOff
-
4. Self-diagnostic information
Icons
Wrong conductivity calibration solution
or exceed recognition range of the
meters
Press key when measuring value
is not stable during calibration.
During calibration, the measuring value
is not stable for 3min.
Self-diagnostic information
Check up
1.Check if the conductivity calibration solution is accurate.
2.Check if the connection between meter and probe is good.
3.Check if the probe is failed.
Press key when icon appears
1.Shake the probe to eliminate bubbles in probe head.
2.Replace conductivity probe with a new one.