Cole parmer 8711 User Manual

OPERATION MANUAL
THERMO-HYGROMETER
Model:8903/8904/8906
Model:
8711 8721
o
o
C / F
HOLD
OFF
RESET
MN / MX
PRG
MX / td
MODEL: 8721
o
C / F
RESET
o
OFF
ON
HOLD
MN / MX
MODEL: 8711
FUNCTION KEYS
1) ON key - Power on
2) OFF Key - Power off
3) PRG Key (8721 only)
- Switch to alarm check mode from measuring mode
- Switch to alarm setting mode from alarm check mode
- Save the alarm settings and calibrations
- Switch to temperature and humidity calibration
oo
4) C/ F Key
- Switch between display of C and F
1
oo
5) RESET Key
- Reset Min/Max memory
- Erase alarm settings
- Skip calibration mode
- Switch back to normal measuring mode from alarm check mode or calibration mode
6) HOLD Key
- Hold display
7) MN/MX Key
- Display minimum value of memory
- Display maximum value of memory
- Increase/Decrease the temperature deviation value
- Sleep/Non-Sleep mode switch
8) NX/td Key (8721 only)
- To flash next digit for alarm setting
- To display the next alarm setting value
- To flash next digit for calibration
- Dew Point read out
9) 9V DC power jack
10) Probe socket
11) RS232 output socket
QUICK REFERENCE
A. MEASURING MODE
This is the basic function to measure temperature ( C or F) and relative humidity.
B. CALIBRATION MODE
With our HR33 & HR75 humidity reference, the end users can do the calibration themselves. This feature provides a DO-IT-YOURSELF accuracy maintenance and replacement of the probe.
oo
2
C. LOW/HIGH HUMIDITY CALIBRATION MODE
Two points' calibration is necessary for maintaining the accuracy of the meter.
D. ALARM CHECK MODE
An environment monitoring system is implanted in the meter for special uses. A monitoring range can be set up earlier,the user will be warned for any out-of-range reading.This Mode provides you to check the alarm settings to see if they are under your requirement.
E. ALARM SETTING MODE
Set, adjust and cancel the alarm settings.
F. RS232 OUTPUT To link to the computer gets records. G. SLEEP MODE OPTION We have a Non-sleep mode and a 20
minutes sleep mode. You can switch easily between two.
H. MN/MX MODE Maximum and Minimum monitoring.
You can find out the minimum and maximum values of previous recording simply by pressing MN/MX key sequentially. The screen will display Fig. 1 & Fig. 2. Press it again to go back to normal mode or the meter will return to normal mode in 10 seconds.
RS232
FIG 1
3
MIN
%RH
o
F
RS232
FIG 2
%RH
o
F
MAX
I. HOLD MODE Hold the screen for specific measuring.
Simply press HOLD key when you do the measurement. The screen will display Fig. 3. Press it again to go back to normal mode.
RS232
FIG 3
HOLD
%RH
o
OPERATION MODES
The 8721/8711 provides 6 operation modes, 3 output port, sleep mode switch
oo
and C/ F switch.
A. MEASURING MODE
This is the normal operation mode for measuring relative humidity and temperature in the default unit. When the power is first on, the full LCD display will appear as Fig. 4.
Full display will appear for 1 second, then readings for humidity and temperature, (Big characters is humidity and small characters is temperature) will display. Fig. 5
4
F
FIG 4
td
MAX
HOLD
BAT MIN
RS232
%RH
RS232
LOTEMP
HITEMP
LO%RH
HI%RH
o
o
F
C
FIG 5
%RH
o
F
B. LOW HUMIDITY CALIBRATION MODE (The calibration solution is optional, not included in standard package.)
Two points' calibration is necessary for the accuracy of the meter.
1. Power off. Push ON & PRG at the same time and hold till the LCD displays CAL (Fig. 6). As ON & PRG are released, the C ( or F, depended on
oo
which is the default unit) will display at the right bottom corner (Fig. 7).
FIG 6
FIG 7
o
C
5