The spectrophotometer described in this manual is designed to be used by properly
trained personnel in a suitable equipped laboratory. For the correct and safe use of this
spectrophotometer it is essential that laboratory personnel follow generally accepted safe
procedures in addition to the safety precautions called for in this manual.
The inside of the power supply unit is a hazardous area and its cover should not be
removed under any circumstances. ANY Servicing must be done by an authorized person.
Some of the chemicals you use in the spectrophotometer may be corrosive, flammable,
radioactive, toxic, and/or potentially infective. Care should be taken to follow the normal
laboratory procedures for handling chemicals and samples. Please carefully read the
Safety, Electrical, Warning, Performance and Radio Interference instructions below.
Safety
This spectrophotometer has been designed and tested in accordance with EN 61326-1:
1997 Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use standard (EMC Requirements). The spectrophotometer has been supplied
in a safe condition.
The safety statements in this manual comply with the requirements of the HEALTH AND
SAFETY AT WORK ACT, 1974.
Read the following before installing and using the instrument and its accessories.
Electrical
Before switching on the spectrophotometer, make sure it is set to the voltage of the local
power supply (see Installation).
The main plug shall be inserted in a socket provided with a protective earth contact. The
protective action must not be negated by the use of an extension cord without a protective
conductor.
Warning
Any interruption of the protective conductor inside or outside the spectrophotometer or
disconnection of the protective earth terminal is likely to make the spectrophotometer
dangerous. Intentional interruption is prohibited.
has been impaired, the spectrophotometer shall be made inoperative and be secured
against any unintended operation. NEVER touch or handle the power supply on the Model
2100UV due to the high voltage.
The protection is likely to be impaired if, for example, the apparatus
• Shows visible damage
• Fails to perform the intended measurements
• Has been subjected to prolonged storage under unfavorable conditions
Whenever it is likely that the protection
3
• Has been subjected to severe transport stresses
Performance
Carry out performance checks with particular reference to wavelength and absorbance
accuracy to ensure that the spectrophotometer is working within its specification,
especially when making measurements of an important nature. Performance checks are
detailed in this manual.
Radio Interference
For compliance with the EMC standards referred to in the EC Declaration of Conformity, it
is necessary that only shielded cables supplied by are used when connecting the
instrument to computers and accessories.
Introduction
The Model’s 2100 and 2100UV Spectrophotometers are single beam, general purpose
instruments designed to meet the needs of the conventional laboratory. 2100 and 2100UV
are ideal for various applications, such as: Clinical Chemistry, Biochemistry,
Petrochemistry, Environmental Protection, Food and Beverage Labs, Water and Waste
Water Labs and other fields of quality control.
Featuring a digital display of photometric result, easy operation and wavelength range of
325~1000nm (Model 2100) or 200
ideal for measurements in the visible wavelength region of the electromagnetic spectrum
(ultraviolet and visible for the Model 2100UV).
~1000nm (Model 2100UV), 2100 and 2100UV are
Working Principle
The spectrophotometer consists of five parts:
1) Halogen/DeuteriumLamp to supply the light;
2) A Monochromator to isolate the wavelength of interest and eliminate the unwanted
second order radiation;
3) A Sample Compartment to accommodate the sample solution;
4) A Detector to receive and convert the transmitted light to an electrical signal;
5) A Digital Display to show absorbance or transmittance. Figure-1 illustrates the
relationship between these
Light Source Monochromator Sample Compartment Detector Digital Display
In your spectrophotometer, light from the lamp is focused on the entrance slit of the
monochromator where the collimating mirror directs the beam onto the grating. The grating
parts.
Figure-1 Block Diagram for the Spectrophotometer
100%T
0 A
4
disperses the light beam to produce the spectrum, a portion of which is focused on the exit
slit of the monochromator by a collimating mirror. From here the beam is passed to a
sample compartment through one of the filters, which helps to eliminate unwanted second
order radiation from the diffraction grating. Upon leaving the sample compartment, the
beam is passed to the silicon photodiode detector and causes the detector to produce an
electrical signal that is displayed on the digital display.
Specifications
Table-1 lists the specification for both Model 2100 and 2100UV.
Wavelength Range 325~1000 nm 200~1000 nm
Spectral Bandpass 5 nm
Wavelength Accuracy Better than 2 nm
Wavelength Repeatability Better than 1 nm
Stray Radiant Energy < 0.3% T at 340 and 400 nm
Photometric Range 0~125%T
Photometric Accuracy
Power Requirements
Sample Holder 4-poisiton Cuvette Holder
Dimensions 470W x 400D x 140H (mm)
Net Weight 10 kg (22 lbs) 12 kg (27 lbs)
Table-1 Specifications
Model 2100 Model 2100UV
-0.1~2.5 Abs
0~1999 C (0~1999 Factor)
±0.004 Abs at 0.5 Abs
115/230 V.
±10%, 60/50 Hz switchable
Unpacking Instructions
Carefully unpack the contents and check the materials against the following packing list to
ensure that you have received everything in good condition:
Packing List
Description
Spectrophotometer
Dust Cover
Cuvette
Table-2 Packing List
Quantity
Model 2100 Model 2100V
1 1
1 1
Set of 4, glassSet of 2, quartz; Set of 4, glass
4-position Cuvette Holder (Manual)
Power Cord
User Manual
1 (installed) 1 (installed)
1 1
1 1
5
Installation
1. After carefully unpacking the contents, check the materials with the packing list above to
ensure that you have received everything in good condition.
2. Place the instrument in a suitable location away from direct sunlight. In order to have the
best performance from your instrument, keep it as far as possible from any strong
magnetic or electrical fields or any electrical device that may generate high-frequency
fields. Set the unit up in an area that is free of dust, corrosive gases and strong vibrations.
3. Remove any obstructions or materials that could hinder the flow of air under and around
the instrument.
4. Select either 230V or 125V/115V on the voltage selector switch on the back of the 2100 and 2100UV shown in Figure-2, to match your local main voltage supply.
5. Turn on your 2100 and 2100UV and allow it to warm up for 15 minutes before taking
any readings.
Figure-2 Back of 2100 and 2100UV
2100 and 2100UV Spectrophotometer Operation Panel
Model 2100UV Operation Panel
Figure-3 shows the Model 2100UV and its Components:
•Sample Compartment: Accommodate different cuvette holders. 4-position Cuvette
Holder is preinstalled in the standard package
•
Power Indicator: Indicate whether power is on or off
•
Cuvette Holder Control Knob: Select the sample position of the 4-position Cuvette
Holder by pulling out or pushing in
Operation Panel: has eleven components, see below for details.
•
6
Figure-4 illustrated the detailed information of the Operation Panel:
1. Data Display Screen: Display the data of Transmittance, Absorbance, Concentr ation, and Factor modes.
2. Mode Indicator: The red LED light indicates the operation MODE. Allow the operator to
know the measurement mode currently in use (T--%Transmittance, A--Absorbance, C-Concentration, and F--Factor).
3. 0A/100%T Button: Blank the 2100/2100UV when blank reference solution is in the Sample
Compartment. Every time a new wavelength is selected, press 0A/100%T Button to
confirm. It will automatically blank the 2100/2100UV.
4. MODEButton: There are four modes. T mode is transmittance mode; A mode is
absorbance mode; C mode is to measure unknown sample concentra tion through a
standard solution; F mode is to measure unknown value with a previously determined
factor.
5. PRINT Button: Press to send Data Display Screen and Wavelength Display Screen
results to RS-232C port and printer.
6. Wavelengt h Display Screen: Display the current/desired wavelength.
7. Wavelengt h Control Buttons: Allow user to select the desired wavelength with the ∧ and
∨
(increase and decrease) buttons.
8. INC/DEC ( Concentration/Factor Buttons) and Lamp Control Buttons: Allow user to
increase or decrease the displayed Concentration at C mode or Factor number at F mode.
At T mode or A mode, press INC button will shutdown the halogen lamp (W), press DEC
mode will shutdown the deuterium lamp (D2). To turn on the lamps, just press the
INC/DEC button again.
9. ENT Button: First press at C mode enters reading on Data Display Screen. Other
presses send displayed results to printer. If using Windows-based Application So ftware,
press this button to communicate with the software (refer to the 2100 Series
Spectrophotometer Software User’s Manual Version SB-2.10 for details). If operating in
F Mode, then press to enter factor. This causes the Mode to enter the factor number and
change to C mode.
10. Visible Lamp Indicator (W): In the operating mode of Absorbance or %Transmittance (A
or T), pressing the INC button will turn off/on the lamp.
11. UV Lamp Indicator (D2): In the operating mode of Absorbance or %Transmittance (A or
T), pressing the DEC button will turn off/on the lamp.SinceDeuterium lamp (D2) is an
expensive item, it is recommended to turn off the) when not using ultraviolet light for
measurement.
7
Figure-3 Model 2100UV Spectrophotometer
Figure-4 Operation Panel of Model 2100UV
8
Model 2100 Operation Panel
The Model 2100 has the similar outlook as Model 2100UV. It also has the following
components:
•Sample Compartment: Accommodate different cuvette holders. 4-position Cuvette
Holder is preinstalled in the standard package
•
Power Indicator: Indicate whether power is on or off
•
Cuvette Holder Control Knob: Select the sample position of the 4-position Cuvette
Holder by pulling out or pushing in
Operation Panel: has nine components, see below for details.
•
Figure-5 illustrated the detailed information of the Operation Panel:
1. Data Display Screen: Display the data of Transmittance, Absorbance, Concentration, and Factor modes.
2. Mode Indicator: The red LED light indicates the operation MODE. Allow the operator to
know the measurement mode currently in use (T--%Transmittance, A--Absorbance, C-Concentration, and F--Factor).
3. 0A/100%T Button: Blank the 2100/2100UV when blank reference solution is in the Sample
Compartment. Every time a new wavelength is selected, press 0A/100%T Button to
confirm. It will automatically blank the 2100/2100UV.
4. MODEButton: There are four modes. T mode is transmittance mode; A mode is
absorbance mode; C mode is to measure unknown sample concentra tion through a
standard solution; F mode is to measure unknown value with a previously determined
factor.
5. PRINT Button: Press to send Data Display Screen and Wavelength DisplayScreen
results to RS-232C port and printer.
6. Wavelength Display Screen: Display the current/desired wavelength.
7. Wavelengt h Control Buttons: Allow user to select the desired wavelength with the
∧
and ∨ (increase and decrease) buttons.
8. INC/DEC (Concentration/Factor Buttons): Allo w user to increase or decrease the
displayed Concentration at C mode or Factor number at F mode.
9. ENT Button: First press at C mode enters reading on Data Display Screen. Other
presses send displayed results to printer. If using Windows-based Application So ftware,
press this button to communicate with the software (refer to the 2100 Series
Spectrophotometer Software User’s Manual Version SB-2.10 for details). If operating in
F Mode, then press to enter factor. This causes the Mode to enter the factor number and
change to C mode.
9
Figure-5 Operation Panel of Model 2100
Start-up Self-Calibration
The Model 2100 and 2100UV have self-calibration programs that begin when the
instrument is powered on. The following are the steps taken automatically each time the
instrument is switched on. Each step is displayed on the
Figure-6
Calibration: P1 Process, P2 Process, P3 Process, P4 Process, Show Process, PC CONN
Process, and Normal Start-up Screen.
to Figure 12 and Table-3 show the detailed information of the Start-up Self-
Wavelength DisplayScreen.
Figure-6 P 1 Process
10