Cole-Parmer 19060-51, 19060-52, 19060-53, 19060-54, 19060-55 Instruction Manual

...
INSTRUCTION MANUAL
Cole-Parmer
19060 Series
EC Meter
Printed in the U.S.A. 03/98
Date:___________________
Serial No.:_____________
Dwg. No.:_______________
Calibrated By:__________
Table of Contents
Shipping checklist continued page 3
Cells, Calibration Solution & Introduction page 3
Intro. continued, Conductivity, Resistivity & TDS page 4
Salinity (19060-10 only), Temperature page 5
Operation, Placement of Instrument, Specifications page 5
Specifications continued page 6
Specifications continued page 7
Power entry module, RS232 & Software diskette page 8
Diskette continued, Analog Recorder Output (19060-10) page 9
Front Panel Controls page 9
Front & Rear Panel Controls, Conductivity Cells page 10
Conductivity Cells cont., Cleaning Cells page 11
Cell Constant page 11
Cell Constant Adjust, Calibration page 12
Making Conductivity, Resistivity & TDS Measurements page 13
Making Salinity Measurements (19060-10 only) page 14
Checking instrument accuracy, Solutions & Literature page 14
Declaration of Conformity for CE mark page 15
1
WARRANTY:
COLE PARMER warrants this product to be free from defects in materials and workmanship for a period of one year from date shipped. Warranty will be allowed whenever possible, however all warranty claims will be reviewed by COLE PARMER.
This warranty shall not apply to fuses, disposable batteries (rechargeable type batteries, pH electrodes, temperature probes and conductivity cells are warranted for 90 days) or any product or part which have been subject to misuse, neglect, tampering, accident or abnormal conditions of operations.
COLE PARMER is pleased to offer suggestions on the use of this product; however, we have no control over its use or intended use. No representation or warranty, whether of merchantability, fitness for any particular purpose is made beyond the repair, replacement or refund of purchase price at the sole discretion of COLE PARMER. In no event shall COLE PARMER be liable for special or consequential damages for injury to person or property which may result from the use of this product. Users shall determine the suitability of this product for its intended applications before using and users shall assume all risks and liability whatsoever in connection therewith regardless of our suggestions as to applications or constructions.
RETURN OF ITEMS:
Authorization should be obtained from our Customer Satisfaction Department (Call - 1-800-323-4340) before returning any item for any reason. When applying for authorization, please include any data regarding the reason the item is being returned. All items must be carefully packed as to prevent shipping damage, and insured against loss and shipping damage. COLE PARMER will not be responsible for any shipping damage. Items being returned without prior authorization may not be accepted.
NOTE:
COLE PARMER reserves the right to make changes in specifications, designs, construction and appearance of our products without notice.
SHIPPING CHECKLIST:
Upon receipt, the instrument should be carefully unpacked and inspected for shipping damage. All materials in the container should be checked against the enclosed packing list. If the instrument has been damaged in transit, you must file a damage claim with the carrier. Please retain all packing material and carton.
MODEL 19060-00 Packing includes the following:
Model 19060-00 Conductivity Meter U.S. Standard Power Cord (removable) Fuses [2] (1/16 amp) for optional 230 VAC operation Ferrite Shield [2] see page 15
2
Instruction Manual
SHIPPING CHECKLIST CONTINUED:
Model 19060-10 Packing includes the following:
Model 19060-10 Conductivity Meter with RS232 & Recorder Output U.S. Standard Power Cord (removable) Fuses [2] (1/16 amp) for optional 230 VAC operation Analog Recorder Connector Software diskette (For PC data logging) Ferrite Shield [2] see page 15 Instruction Manual
CELLS:
Order cells separately Four wire cell with embedded thermistors Cell Constant approximately 10 cm -1
Au cells are best suited for high purity solutions Multipurpose cells are combination dip and flow
19060-51 Dip Cell (Au) with 48” cable 19060-52 Dip Cell (Pt) with 48” cable 19060-53 Multipurpose cell (Au) with 48” cable 19060-54 Multipurpose cell (Pt) with 48” cable 19060-55 Multipurpose cell (Au) with 2 meter cable 19060-56 Multipurpose cell (Pt) with 2 meter cable
01481-11 Platinizing Station for replatinizing (Pt) cells
(does not include ASTM D1125 Platinizing Solution)
CALIBRATION SOLUTION:
See our General Catalog for available Conductivity Calibration Solutions.
INTRODUCTION:
The model 19060-00 is a bench top microprocessor based multifunction meter designed for the measurement of conductivity and temperature of aqueous solutions. This conductivity meter has six ranges, each from 0 to 3300 counts and is displayed on a large sixteen character dot matrix liquid crystal display. All instruments are shipped from the factory for 115 VAC line operation. The power entry module on the rear panel can be easily changed for 230 VAC operation when necessary.
This meter features four front panel controls, one for manual range selection or auto ranging. One for function selection of conductivity, resistivity, TDS, self-test or standby. One for parameter selection of reference temperature, temperature coefficient, cell constant and TDS multiplier. The last front panel control is for the display and setting of the parameters as well as transmit on command (transmit on command for model 19060-10 only).
3
INTRODUCTION CONTINUED:
The model 19060-10 has an additional function of Salinity, RS232 output for a serial printer or PC data display, logging and printing. An analog output of conductance and temperature for strip chart recording.
CONDUCTIVITY:
The basic unit of resistance is the ohm, conductance is the reciprocal of resistance, its basic unit of measurement is the mho (international system of units for mho is Siemens [S]).
In the measurement of aqueous solutions, conductivity is based on the reciprocal of the resistance of a 1-cm cube of material measured between opposite faces. This is a cell constant of 1. Conductivity cells usually consist of two metallic plates of a determined size mounted in a defined area. The cell constant “K” is the length “L” (or distance between the plates) of the conducting path in centimeters divided by the effective cross sectional area “A” of the conducting path in square centimeters (K=L/A). The resistance of conductor is inversely proportional to its cross sectional area and directly proportional to its length.
The model 19060-00, -10 is designed to use a cell with a constant of 10 cm-1. The instrument’s cell constant parameter setting is adjustable from 8.00 to 12.00.
Conduction in aqueous solutions is by ionic movement and increases with temperature. This change is expressed in percent per degree “C”
relative to a reference temperature normally 25.0°C and is commonly called the slope.
The model 19060-00, -10 has Automatic Temperature Compensation for slope correction with reference temperature parameter adjustable from 12 to
28°C, temperature coefficient parameter adjustable from 0 to 12.5% per degree C and displays conductivity in mS or mS.
RESISTIVITY:
The model 19060-00, -10 displays resistivity as a reciprocal of the conductivity. Resistivity’s basic unit of measurement is the ohm.
Model 19060-00, -10 displays resistivity in MW and KW.
TOTAL DISSOLVED SOLIDS:
The Total Dissolved Solids (TDS) in a solution does not change with temperature. Conductivity is sensitive to temperature. The Model 19060-00, -10 automatically calculates the TDS from the conductivity reading by applying the appropriate factor that the user sets or by using the preset parameter. The TDS multiplier is adjustable from 40% to 125% and displays TDS in PPM (Parts P Thousand).
er Million) or PPT (Parts Per
4