Operating Manual
USB-based
8-channel Thermocouple Input Module
MODEL NO. 18200-40
The 18200-40 is a USB 2.0 full-speed, thermocouple input module that is fully compatible with both USB
1.1 and USB 2.0 ports.
With the 18200-40, you can take measurements from type J, K, R, S, T, N, E, and B thermocouples.
The 18200-40 provides two integrated cold junction compensation (CJC) sensors for thermocouple measurements.
An open thermocouple detection feature lets you detect a broken thermocouple. An on-board microprocessor automatically linearizes the measurement data.
The 18200-40 is a standalone plug-and-play module which draws power from the USB cable. No external power supply is required. All configurable options are software programmable.
The 18200-40 is fully software calibrated.
Cole-Parmer Instrument Company
625 East Bunker Court
Vernon Hills, Illinois 60061-1844
(847) 549-7600
(847) 247-2929 (Fax)
800-323-4340
www.coleparmer.com
e-mail: techinfo@coleparmer.com
Document Revision 1, May, 2006
© Copyright 2006 Measurement Computing Corporation, TCR4
2
Page 3

Table of Contents

Preface
About this Operating Manual...............................................................................................................5
What you will learn from this manual ................................................................................................................5
Conventions in this manual ................................................................................................................................5
Where to find more information.........................................................................................................................5
Chapter 1
Introducing the 18200-40......................................................................................................................6
Overview: 18200-40 features .............................................................................................................................6
18200-40 block diagram.....................................................................................................................................7
Software features................................................................................................................................................7
Connecting a 18200-40 to your computer is easy...............................................................................................8
Connecting a 18200-40 to your computer is easy...............................................................................................8
Chapter 2
Installing the 18200-40..........................................................................................................................9
What comes with your 18200-40 shipment? ......................................................................................................9
Hardware .......................................................................................................................................................................... 9
Additional documentation................................................................................................................................................. 9
Unpacking the 18200-40 ..................................................................................................................................10
Unpacking the 18200-40 ..................................................................................................................................10
Installing the software ......................................................................................................................................10
Installing the hardware .....................................................................................................................................10
Configuring the 18200-40.................................................................................................................................10
Calibrating the 18200-40 ..................................................................................................................................10
Chapter 3
Sensor Connections ...........................................................................................................................11
Screw terminal pin out......................................................................................................................................11
Thermocouple input terminals (C0H/C0L to C7H/C7L) .................................................................................................12
Ground terminals (GND) .................................................................................................................................................12
Power terminals (+5V).....................................................................................................................................................12
Digital terminals (DIO0 to DIO7)....................................................................................................................................12
CJC sensors......................................................................................................................................................................12
Thermocouple connections...............................................................................................................................13
Wiring configuration........................................................................................................................................................13
Digital I/O connections.....................................................................................................................................14
Chapter 4
Functional Details ...............................................................................................................................15
Thermocouple measurements ...........................................................................................................................15
Cold junction compensation (CJC)..................................................................................................................................15
Data linearization.............................................................................................................................................................15
Open-thermocouple detection (OTD) ..............................................................................................................................15
USB connector..................................................................................................................................................16
LED ..................................................................................................................................................................16
Power................................................................................................................................................................16
Chapter 5
Specifications......................................................................................................................................17
Analog input section.........................................................................................................................................17
Channel configurations.....................................................................................................................................17
3
Page 4
18200-40 Operating Manual
Accuracy...........................................................................................................................................................18
Thermocouple measurement accuracy.............................................................................................................................18
Throughput rate ................................................................................................................................................18
Digital input/output...........................................................................................................................................19
Memory ............................................................................................................................................................19
Microcontroller.................................................................................................................................................19
USB +5V voltage .............................................................................................................................................19
Power................................................................................................................................................................19
USB specifications ...........................................................................................................................................20
Environmental ..................................................................................................................................................20
Mechanical .......................................................................................................................................................20
Screw terminal connector type and pin out.......................................................................................................20
Screw terminal pin out.....................................................................................................................................................21
4
Page 5
Preface

About this Operating Manual

What you will learn from this manual

This manual explains how to install, configure, and use the 18200-40 so that you get the most out of its USB­based temperature measurement features.
This manual also refers you to related documents available on our web site, and to technical support resources.

Conventions in this manual

For more information on …
Text presented in a box signifies additional information and helpful hints related to the subject matter you are reading.
Caution! Shaded caution statements present information to help you avoid injuring yourself and others,
damaging your hardware, or losing your data.
<#:#> Angle brackets that enclose numbers separated by a colon signify a range of numbers, such as those assigned
to registers, bit settings, etc.
bold text Bold text is used for the names of objects on the screen, such as buttons, text boxes, and check boxes. For
example:
1. Insert the disk or CD and click the OK button.
italic text Italic text is used for the names of manuals and help topic titles, and to emphasize a word or phrase. For
example: The InstaCal® installation procedure is explained in the Quick Start Guide. Never touch the exposed pins or circuit connections on the board.

Where to find more information

The following electronic documents provide helpful information relevant to the operation of the 18200-40.
! Cole-Parmer's Specifications: 18200-40 (the PDF version of the Electrical Specification Chapter in this
guide) is available on our web site at www.coleparmer.com/catalog/MoreInfo/18200-40-spec.pdf
! Cole-Parmer's Quick Start Guide is available on our web site at
www.coleparmer.com/catalog/MoreInfo/CP-Quick-Start.pdf
! Cole-Parmer's Guide to Signal Connections is available on our web site at
www.coleparmer.com/catalog/MoreInfo/CPsignals.pdf
! Cole-Parmer's Universal Library User's Guide is available on our web site at
www.coleparmer.com/catalog/MoreInfo/sm-ul-user-guide.pdf
! Cole-Parmer's Universal Library Function Reference is available on our web site at
www.coleparmer.com/catalog/MoreInfo/sm-ul-functions.pdf
! Cole-Parmer's Universal Library for LabVIEW
www.coleparmer.com/catalog/MoreInfo/SM-UL-LabVIEW.pdf
18200-40 Operating Manual (this document) is also available on our web site at
www.coleparmer.com/cpusbdaq1/18200-40.pdf
User’s Guide is available on our web site at
.
.
.
.
.
.
.
5
Page 6
Chapter 1

Introducing the 18200-40

Overview: 18200-40 features

This manual contains all of the information you need to connect the 18200-40 to your computer and to the signals you want to measure.
The 18200-40 is a USB 2.0 full-speed, thermocouple input module that is supported under popular Microsoft® Windows® operating systems. The 18200-40 is fully compatible with both USB 1.1 and USB 2.0 ports.
With the 18200-40, you can take measurements from type J, K, R, S, T, N, E, and B thermocouples.
The 18200-40 provides two integrated cold junction compensation (CJC) sensors for thermocouple measurements.
An open thermocouple detection feature lets you detect a broken thermocouple. An on-board microprocessor automatically linearizes the measurement data.
The 18200-40 is a standalone plug-and-play module which draws power from the USB cable. No external power supply is required. All configurable options are software programmable.
The 18200-40 is fully software calibrated.
6
Page 7
18200-40 Operating Manual Introducing the 18200-40

18200-40 block diagram

18200-40 functions are illustrated in the block diagram shown here.
Precision
5V Ref.
DIO
USB
USB
Micro
USB
+5V
I/O
Isolator
Isolated
DC/DC
500 V
Isolation
Barrier
8
Isolated
Micro
Te mp
sensor
(+12)
(-12)
Figure 1-1. 18200-40 functional block diagram
SPI
24-bit A/D
(CH0, CH1)
24-bit A/D
(CH2, CH3)
24-bit A/D
(CH4, CH5)
24-bit A/D
(CH6, CH7)
Input mux.
Input mux.
CJC
CH0-3
Input mux.
Input mux.
CJC
CH4-7
Screw terminalScrew terminalScrew terminal

Software features

For information on the features of InstaCal and the other software included with your 18200-40, refer to the Quick Start Guide that shipped with your device. The Quick Start Guide is also available in PDF at
www.coleparmer.com/catalog/MoreInfo/CP-Quick-Start.pdf
.
7
Page 8
18200-40 Operating Manual Introducing the 18200-40
Connecting a 18200-40 to your computer is easy
Installing a data acquisition device has never been easier.
! The 18200-40 relies upon the Microsoft Human Interface Device (HID) class drivers. The HID class
drivers ship with every copy of Windows that is designed to work with USB ports. We use the Microsoft HID because it is a standard, and its performance delivers full control and maximizes data transfer rates for your 18200-40. No third-party device driver is required.
! The 18200-40 is plug-and-play. There are no jumpers to position, DIP switches to set, or interrupts to
configure.
! You can connect the 18200-40 before or after you install the software, and without powering down your
computer first. When you connect an HID to your system, your computer automatically detects it and configures the necessary software. You can connect and power multiple HID peripherals to your system using a USB hub.
! You can connect your system to various devices using a standard four-wire cable. The USB connector
replaces the serial and parallel port connectors with one standardized plug and port combination.
! You do not need a separate power supply module. The USB automatically delivers the electrical power
required by each peripheral connected to your system.
! Data can flow two ways between a computer and peripheral over USB connections.
8
Page 9

Installing the 18200-40

What comes with your 18200-40 shipment?

The following items are shipped with the 18200-40.

Hardware

The following items should be included with your shipment.
! 18200-40
Chapter 2
! USB cable (2 meter length)

Additional documentation

In addition to this hardware manual, you should also receive the Quick Start Guide (available in PDF at
www.coleparmer.com/catalog/MoreInfo/CP-Quick-Start.pdf
software you received with your 18200-40 and information regarding installation of that software. Please read this booklet completely before installing any software or hardware.
.). This booklet supplies a brief description of the
9
Page 10
18200-40 Operating Manual Installing the 18200-40
Unpacking the 18200-40
As with any electronic device, you should take care while handling to avoid damage from static electricity. Before removing the 18200-40 from its packaging, ground yourself using a wrist strap or by simply touching the computer chassis or other grounded object to eliminate any stored static charge.
If your 18200-40 is damaged, notify Cole-Parmer immediately by phone, fax, or email.
! Phone: 800-323-4340. ! Fax: 847-247-2929 ! Email: techinfo@coleparmer.com

Installing the software

Refer to the Quick Start Guide for instructions on installing the software on the C-P Data Acquisition Software CD. This booklet is available in PDF at www.coleparmer.com/catalog/MoreInfo/CP-Quick-Start.pdf.

Installing the hardware

To connect the 18200-40 to your system, turn your computer on, and connect the USB cable to a USB port on your computer or to an external USB hub that is connected to your computer. The USB cable provides power and communication to the 18200-40.
When you connect the 18200-40 for the first time, a dialog (other Windows versions) opens as the 18200-40 is detected.
When this balloon or dialog closes, the installation is complete. The This indicates that communication is established between the 18200-40 and your computer.
Caution! Do not disconnect any device from the USB bus while the computer is communicating with the
18200-40, or you may lose data and/or your ability to communicate with the 18200-40.
If the LED turns off
If the LED is lit but then turns off, the computer has lost communication with the 18200-40. To restore communication, disconnect the USB cable from the computer, and then reconnect it. This should restore communication, and the LED should turn back on.
Found New Hardware popup balloon (Windows XP) or
USB LED should flash and then remain lit.

Configuring the 18200-40

All hardware configuration options on the 18200-40 are programmable with software. Use InstaCal to set the thermocouple type for each channel pair. Configuration options are stored on the 18200-40 's isolated microcontroller in EEPROM, which is non-volatile memory on the 18200-40 module. Configuration options are loaded on power up. The factory default configuration is Type J thermocouple.
Allow the 18200-40 to warm up for 30 minutes after powering up before taking measurements. This warm up time minimizes thermal drift and achieves the specified rated accuracy of measurements.

Calibrating the 18200-40

The 18200-40 is fully calibrated via InstaCal. Allow a 30 minute warm up before calibrating.
10
Page 11
Chapter 3

Sensor Connections

The 18200-40 supports type types J, K, R, S, T, N, E, and B thermocouples.
Thermocouple selection
The thermocouple type you select will depend on your application needs. Review the temperature ranges and accuracies of each type to determine which is best suited for your application.

Screw terminal pin out

The 18200-40 has four rows of screw terminals — two rows on the top edge of the housing, and two rows on the bottom edge. Each row has 26 connections. Between each bank of screw terminals are two integrated CJC sensors used for thermocouple measurements. Signals are identified in Figure 3-1.
RSVD
27 RSVD
28 GND
29 C7L
30 C7H3132 NC
6
RSVD
RSVD
33 C6L
34 C6H
35 NC
36
37
CJC SensorCJC Sensor
11
10
RSVD
38 GND
39 C5L
40 C5H4142 NC
16
RSVD
43 C4L
44 C4H
45 NC4647 +5V
48 GND
49 DIO7
50 DIO6
51 DIO5
52 DIO4
20
RSVD 1
NC 2
C0H 3
C0L 4
NC 5
C1H 7
C1L 8
GND 9
RSVD
RSVD
NC 12
C2H 13
C2L 14
NC 15
C3H 17
C3L 18
GND 19
+5V 21
GND 22
DIO0 23
DIO1 24
DIO2 25
RSVD
RSVD
RSVD
DIO3 26
Figure 3-1. 18200-40 screw terminal pin numbers
11
Page 12
18200-40 Operating Manual Sensor Connections
Table 3-1. 18200-40 screw terminal descriptions
Pin Signal Name Pin Description Pin Signal Name Pin Description
1 RSVD Reserved, Do Not Use 27 RSVD Reserved, Do Not Use 2 NC Not connected 28 GND Ground
3 C0H CH0 sensor input (+) 29 C7L
4 C0L 5 NC Not connected 31 RSVD Reserved, Do Not Use
6 RSVD Reserved, Do Not Use 32 NC Not connected 7 C1H CH1 sensor input (+) 33 C6L
8 C1L 9 GND Ground 35 NC Not connected
10 RSVD Reserved, Do Not Use 36 RSVD Reserved, Do Not Use
11 RSVD Reserved, Do Not Use 37 RSVD Reserved, Do Not Use 12 NC Not connected 38 GND Ground
13 C2H CH2 sensor input (+) 39 C5L
14 C2L 15 NC Not connected 41 RSVD Reserved, Do Not Use
16 RSVD Reserved, Do Not Use 42 NC Not connected 17 C3H CH3 sensor input (+) 43 C4L
18 C3L 19 GND Ground 45 NC Not connected
20 RSVD Reserved, Do Not Use 46 RSVD Reserved, Do Not Use 21 +5V +5V output 47 +5V +5V output 22 GND Ground 48 GND Ground 23 DIO0 Digital Input/Output 49 DIO7 Digital Input/Output 24 DIO1 Digital Input/Output 50 DIO6 Digital Input/Output 25 DIO2 Digital Input/Output 51 DIO5 Digital Input/Output 26 DIO3 Digital Input/Output 52 DIO4 Digital Input/Output
CJC sensor
CH0 sensor input (
CH1 sensor input (
CH2 sensor input (
CH3 sensor input (
-)
-)
-)
-)
30 C7H CH7 sensor input (+)
34 C6H CH6 sensor input (+)
40 C5H CH5 sensor input (+)
44 C4H CH4 sensor input (+)
CJC sensor
CH7 sensor input (
CH6 sensor input (
CH5 sensor input (
CH4 sensor input (
-)
-)
-)
-)
Use 16 AWG to 30 AWG wire for your signal connections.
Tighten screw terminal connections
When making connections to the screw terminals, be sure to tighten the screw until tight. Simply touching the top of the screw terminal is not sufficient to make a proper connection.

Thermocouple input terminals (C0H/C0L to C7H/C7L)

You can connect up to eight thermocouples to the differential sensor inputs (C0H/C0L to C7H/C7L). The 18200-40 supports type types J, K, R, S, T, N, E, and B thermocouples.

Ground terminals (GND)

The six analog ground terminals (GND) provide a common ground for the input channels and DIO bits and are isolated (500 VDC) from the USB GND.

Power terminals (+5V)

The two +5V terminals are isolated (500 VDC) from the USB +5V.

Digital terminals (DIO0 to DIO7)

You can connect up to eight digital I/O lines to the screw terminals labeled DIO0 to DIO7. Each terminal is software configurable for input or output.

CJC sensors

The 18200-40 has two built in high-resolution temperature sensors. One sensor is located on the right side of the package, and one sensor is located at the left side.
12
Page 13
18200-40 Operating Manual Sensor Connections

Thermocouple connections

A thermocouple consists of two dissimilar metals that are joined together at one end. When the junction of the metals is heated or cooled, a voltage is produced that correlates to temperature.
The 18200-40 makes fully differential thermocouple measurements without the need of ground-referencing resistors. A 32-bit floating point value in either a voltage or temperature format is returned by software. An open thermocouple detection feature is available for each analog input which automatically detects an open or broken thermocouple.
Use InstaCal to select the thermocouple type (J, K, R, S, N. E, and B) and one or more sensor input channels to connect the thermocouple.

Wiring configuration

Connect the thermocouple to the 18200-40 using a differential configuration, as shown in Figure 3-2.
I#+
NC
C#H
C#L
4W##
IC##
C#H
C#L
GND
I#-
Figure 3-2. Typical thermocouple connection
Connect thermocouples to the 18200-40 such that they are floating with respect to GND (pins 9, 19, 28, 38). The 18200-40 GND pins are isolated from earth ground, so connecting thermocouple sensors to voltages referenced to earth ground is permissible as long as the isolation between the GND pins (9, 19, 28, 38) and earth ground is maintained.
When thermocouples are attached to conductive surfaces, the voltage differential between multiple thermocouples must remain within ±1.4 V. For best results, we recommend the use of insulated or ungrounded thermocouples when possible.
Maximum input voltage between analog input and ground
The absolute maximum input voltage between an analog input and the isolated GND pins is ±25 VDC when the 18200-40 is powered on, and ±40 VDC when the 18200-40 is powered off.
If you need to increase the length of your thermocouple, use the same type of thermocouple wires to minimize the error introduced by thermal EMFs.
13
Page 14
18200-40 Operating Manual Sensor Connections

Digital I/O connections

You can connect up to eight digital I/O lines to the screw terminals labeled DIO0 to DIO7. You can configure each digital bit for either input or output. All digital I/O lines are pulled up to +5V with a 47 K ohm resistor (default). You can request the factory to configure the resistor for pull-down to ground if desired.
When you configure the digital bits for input, you can use the 18200-40 digital I/O terminals to detect the state of any TTL-level input. Refer to the schematic shown in Figure 3-3. If you set the switch to the +5V input, DIO0 reads TRUE (1). If you move the switch to GND, DIO0 reads FALSE (0).
DIO0
+5V+GND
Figure 3-3. Schematic showing switch detection by digital channel DIO0
All ground pins on the 18200-40 (pins 9, 19, 28, 38) are isolated from earth ground. If a connection is made to earth ground when using digital I/O and conductive thermocouples, the thermocouples are no longer isolated. In this case, thermocouples must not be connected to any conductive surfaces that may be referenced to earth ground
For more information on digital signal connections
For more information on digital signal connections and digital I/O techniques, refer to the Guide to Signal Connections (available on our web site at www.coleparmer.com/catalog/MoreInfo/CPsignals.pdf.)
14
Page 15
Chapter 4

Functional Details

Thermocouple measurements

A thermocouple consists of two dissimilar metals that are joined together at one end. When the junction of the metals is heated or cooled, a voltage is produced that correlates to temperature.
The 18200-40 hardware level-shifts the thermocouple’s output voltage into the A/D’s common mode input range by applying +2.5 V to the thermocouple’s low side at the C#L input. Always connect thermocouple sensors to the 18200-40 in a floating fashion. Do not attempt to connect the thermocouple low side C#L to GND or to a ground referencing resistor.

Cold junction compensation (CJC)

When you connect the thermocouple sensor leads to the sensor input channel, the dissimilar metals at the 18200-40 terminal blocks produce an additional thermocouple junction. This junction creates a small voltage error term which must be removed from the overall sensor measurement using a cold junction compensation technique. The measured voltage includes both the thermocouple voltage and the cold junction voltage. To compensate for the additional cold junction voltage, the 18200-40 subtracts the cold junction voltage from the thermocouple voltage.
The 18200-40 has two high-resolution temperature sensors that are integrated into the design of the 18200-40. One sensor is located on the right side of the package, and one sensor is located at the left side. The CJC sensors measure the average temperature at the terminal blocks so that the cold junction voltage can be calculated. A software algorithm automatically corrects for the additional thermocouples created at the terminal blocks by subtracting the calculated cold junction voltage from the analog input's thermocouple voltage measurement.
Increasing the thermocouple length
If you need to increase the length of your thermocouple, use the same type of thermocouple wires to minimize the error introduced by thermal EMFs.

Data linearization

After the CJC correction is performed on the measurement data, an on-board microcontroller automatically linearizes the thermocouple measurement data using National Institute of Standards and Technology (NIST) linearization coefficients for the selected thermocouple type.
The measurement data is then output as a 32-bit floating point value in the configured format (voltage or temperature).

Open-thermocouple detection (OTD)

The 18200-40 is equipped with an open-thermocouple detection for each analog input channel. With OTD, any open-circuit or short-circuit condition at the thermocouple sensor is detected by the software. An open channel is detected by driving the input voltage to a negative value outside the range of any thermocouple output. The software recognizes this as an invalid reading and flags the appropriate channel. The software continues to sample all channels when OTD is detected.
Input leakage current
With open-thermocouple detection enabled, 105 nA (max.) of input leakage current is injected into the thermocouple. This current can cause an error voltage to develop across the lead resistance of the thermocouple that is indistinguishable from the thermocouple voltage you are measuring. You can estimate this error voltage with the following formula:
error voltage = resistance of the thermocouple x 105 nA
To reduce the error, reduce the length of the thermocouple to lower its resistance, or lower the AWG of the wire by using a wire with a larger diameter. With open-thermocouple detection disabled, 30 nA (max.) of input leakage current is injected into the thermocouple.
15
Page 16
18200-40 Operating Manual Functional Details

USB connector

The USB connector provides +5V power and communication. No external power supply is required.
LED
The LED indicates the communication status of the 18200-40. It uses up to 5 mA of current. Table 4-2 defines the function of the 18200-40 LED.
Table 4-2. LED Illumination
LED Illumination
Steady green The 18200-40 is connected to a computer or external USB hub.
Pulsing green Data is being transferred.
Indication
Upon connection, the LED should flash three times and then remain lit (indicates a successful installation).

Power

The two +5V terminals are isolated (500 VDC) from the USB +5V.
Caution! Each +5V terminal is an output. Do not connect to an external power supply or you may damage
the 18200-40 and possibly the computer.
16
Page 17

Specifications

Typical for 25 °C unless otherwise specified. Specifications in italic text are guaranteed by design.

Analog input section

Table 1. Generic analog input specifications
Parameter Conditions Specification
A/D converters Four dual 24-bit, Sigma-Delta type Number of channels 8 differential
Input isolation 500 VDC minimum between field wiring and
USB interface
Channel configuration Thermocouple sensor type
Differential input voltage range
Absolute maximum input voltage
Input impedance 5 Gigohm, min. Input leakage current Open thermocouple detect enabled 105 nA max.
Thermocouple ±0.080 V
±C0x through ±C7x relative to GND (pins 9,19,28,38)
±25 V power on, ±40 V power off.
Chapter 5
Normal mode rejection ratio f Common mode rejection ratio
Resolution 24 bits No missing codes 24 bits Input coupling DC Warm-up time 30 minutes min. Open thermocouple detect Automatically enabled when the channel pair
CJC sensor accuracy
= 60 Hz 90 dB min.
IN
f
= 50 Hz/60 Hz 100 dB min.
IN
is configured for thermocouple sensors. The maximum open detection time is 3 seconds.
15 °C to 35 °C ±0.25 °C typ.,
±0.5 °C max.
0 °C to 70 °C -1.0 to +0.5 °C max

Channel configurations

Table 2. Channel configuration specifications
Sensor Category Conditions Specification
Thermocouple J, K, S, R, B, E, T, or N 8 differential channels
Note 1: Channel configuration information is stored in the EEPROM of the isolated microcontroller by
the firmware whenever any item is modified. Modification is performed by commands issued over USB from an external application, and the configuration is made non-volatile through the use of the EEPROM.
Note 2: The factory default configuration is Type J.
17
Page 18
18200-40 Operating Manual Specifications

Accuracy

Thermocouple measurement accuracy

Table 3. Thermocouple accuracy specifications, including CJC measurement error
Sensor Type Maximum error Typical error Temperature range
±1.499 °C ±0.507 °C -210 to 0 °C J
±0.643 °C ±0.312 °C 0 to 1200 °C
±1.761 °C ±0.538 °C -210 to 0 °C K
±0.691 °C ±0.345 °C 0 to 1372 °C
±2.491°C ±0.648 °C -50 to 250 °C S
±1.841 °C ±0.399 °C 250 to 1768.1 °C
±2.653 °C ±0.650 °C -50 to 250 °C R
±1.070 °C ±0.358 °C 250 to 1768.1 °C
±1.779 °C ±0.581 °C 250 to 700 °C B
±0.912 °C ±0.369 °C 700 to 1820 °C
±1.471 °C ±0.462 °C -200 to 0 °C E
±0.639 °C ±0.245 °C 0 to 1000 °C
±1.717 °C ±0.514 °C -200 to 0 °C T
±0.713 °C ±0.256 °C 0 to 600 °C
±1.969 °C ±0.502 °C -200 to 0 °C N
±0.769 °C ±0.272 °C 0 to 1300 °C
Note 3: Thermocouple specifications include linearization, cold-junction compensation and system noise.
These specs are for one year, or 3000 operating hours, whichever comes first and for operation of the 18200-40 between 15 °C and 35 °C. For measurements outside this range, add ±0.5 degree to the maximum error shown. There are CJC sensors on each side of the module. The accuracy listed above assumes the screw terminals are at the same temperature as the CJC sensor. Errors shown do not include inherent thermocouple error. Please contact your thermocouple supplier for details on the actual thermocouple error.
Note 4: Thermocouples must be connected to the 18200-40 such that they are floating with respect to
GND (pins 9, 19, 28, 38). The 18200-40 GND pins are isolated from earth ground, so connecting thermocouple sensors to voltages referenced to earth ground is permissible as long as the isolation between the GND pins and earth ground is maintained.
Note 5: When thermocouples are attached to conductive surfaces, the voltage differential between
multiple thermocouples must remain within ±1.4V. For best results we recommend the use of ungrounded or insulated thermocouples when possible.

Throughput rate

Table 4. Throughput rate specifications
Number of input channels Maximum throughput
1 2 Samples/second 2 2 S/s on each channel, 4 S/s total 3 2 S/s on each channel, 6 S/s total 4 2 S/s on each channel, 8 S/s total 5 2 S/s on each channel, 10 S/s total 6 2 S/s on each channel, 12 S/s total 7 2 S/s on each channel, 14 S/s total 8 2 S/s on each channel, 16 S/s total
Note 6: The analog inputs are configured to run continuously. Each channel is sampled twice per second.
The maximum latency between when a sample is acquired and the temperature data is provided by the USB unit is approximately 0.5 seconds.
18
Page 19
18200-40 Operating Manual Specifications

Digital input/output

Table 5. Digital input/output specifications
Digital type CMOS Number of I/O 8 (DIO0 through DIO7) Configuration Independently configured for input or output.
Power on reset is input mode.
Pull up/pull-down configuration All pins pulled up to +5 V via 47 K resistors (default). Pull down to
ground (GND) also available.
Digital I/O transfer rate (software paced) ! Digital input – 50 port reads or single bit reads per second typ.
! Digital output – 100 port writes or single bit writes per second typ.
Input high voltage 2.0 V min., 5.5 V absolute max. Input low voltage 0.8 V max., -0.5 V absolute min. Output low voltage (IOL = 2.5 mA) 0.7 V max Output high voltage (IOH = -2.5 mA)
Note 7: All ground pins on the 18200-40 (pins 9, 19, 28, 38) are isolated from earth ground. If a
3.8 V min.
connection is made to earth ground when using digital I/O and conductive thermocouples, the thermocouples are no longer isolated. In this case, thermocouples must not be connected to any conductive surfaces that may be referenced to earth ground.

Memory

Table 6. Memory specifications
EEPROM 1,024 bytes isolated micro reserved for sensor configuration
256 bytes USB micro for external application use

Microcontroller

Table 7. Microcontroller specifications
Type Two high performance 8-bit RISC microcontrollers

USB +5V voltage

Table 8. USB +5V voltage specifications
Parameter Conditions Specification
USB +5V (VBUS) input voltage range 4.75 V min. to 5.25 V max.

Power

Table 9. Power specifications
Parameter Conditions Specification
Supply current USB enumeration <100 mA Supply current (Note 8) User +5V output voltage range (terminal block pin 21 and 47) User +5V output current (terminal block pin 21 and pin 47) Isolation Measurement system to PC 500 VDC min.
Note 8: This is the total current requirement for the 18200-40 which includes up to 10 mA for the status
LED.
Continuous mode 70 mA max.
Connected to self-powered hub. (Note 9) 4.75 V min. to
5.25 V max.
Bus-powered and connected to a self-powered hub. (Note 9) 10 mA max.
19
Page 20
18200-40 Operating Manual Specifications
Note 9: Self-Powered Hub refers to a USB hub with an external power supply. Self-powered hubs allow
a connected USB device to draw up to 500 mA.
Root Port Hubs reside in the PC’s USB Host Controller. The USB port(s) on your PC are root port hubs. All externally powered root port hubs (desktop PC’s) provide up to 500 mA of current for a USB device. Battery-powered root port hubs provide 100 mA or 500 mA, depending upon the manufacturer. A laptop PC that is not connected to an external power adapter is an example of a battery-powered root port hub.

USB specifications

Table 10. USB specifications
USB device type USB 2.0 (full-speed) Device compatibility USB 1.1, USB 2.0 Self-powered, 100 mA consumption max
USB cable type A-B cable, UL type AWM 2527 or equivalent. (min 24 AWG VBUS/GND,
min 28 AWG D+/D-)
USB cable length 3 meters max.

Environmental

Table 11. Environmental specifications
Operating temperature range Storage temperature range -40 to 85 ° C Humidity 0 to 90% non-condensing
0 to 70 ° C

Mechanical

Table 12. Mechanical specifications
Dimensions 127 mm (L) x 88.9 mm (W) x 35.56 (H) User connection length 3 meters max.

Screw terminal connector type and pin out

Table 13. Screw terminal connector specifications
Connector type Screw terminal Wire gauge range 16 AWG to 30 AWG
20
Page 21
18200-40 Operating Manual Specifications

Screw terminal pin out

Table 14. Screw terminal pin out
Pin Signal Name Pin Description Pin Signal Name Pin Description
1 RSVD Reserved, Do Not Use 27 RSVD Reserved, Do Not Use 2 NC 28 GND
3 C0H CH0 sensor input (+) 29 C7L
4 C0L 5 NC 31 RSVD Reserved, Do Not Use
6 RSVD Reserved, Do Not Use 32 NC 7 C1H CH1 sensor input (+) 33 C6L
8 C1L 9 GND 35 NC
10 RSVD Reserved, Do Not Use 36 RSVD Reserved, Do Not Use
11 RSVD Reserved, Do Not Use 37 RSVD Reserved, Do Not Use 12 NC 38 GND
13 C2H CH2 sensor input (+) 39 C5L
14 C2L 15 NC 41 RSVD Reserved, Do Not Use
16 RSVD Reserved, Do Not Use 42 NC 17 C3H CH3 sensor input (+) 43 C4L
18 C3L 19 GND 45 NC
20 RSVD Reserved, Do Not Use 46 RSVD Reserved, Do Not Use 21 +5V +5V output 47 +5V +5V output 22 GND 48 GND 23 DIO0 Digital Input/Output 49 DIO7 Digital Input/Output 24 DIO1 Digital Input/Output 50 DIO6 Digital Input/Output 25 DIO2 Digital Input/Output 51 DIO5 Digital Input/Output 26 DIO3 Digital Input/Output 52 DIO4 Digital Input/Output
CJC sensor
CH0 sensor input (
CH1 sensor input (
CH2 sensor input (
CH3 sensor input (
-)
-)
-)
-)
30 C7H CH7 sensor input (+)
34 C6H CH6 sensor input (+)
40 C5H CH5 sensor input (+)
44 C4H CH4 sensor input (+)
CJC sensor
CH7 sensor input (
CH6 sensor input (
CH5 sensor input (
CH4 sensor input (
-)
-)
-)
-)
21
Page 22
Declaration of Conformity
Manufacturer: Measurement Computing Corporation Address: 10 Commerce Way Suite 1008 Norton, MA 02766 USA
Category: Electrical equipment for measurement, control and laboratory use.
Measurement Computing Corporation declares under sole responsibility that the product
18200-40
to which this declaration relates is in conformity with the relevant provisions of the following standards or other documents:
EU EMC Directive 89/336/EEC: Electromagnetic Compatibility, EN 61326 (1997) Amendment 1 (1998)
Emissions: Group 1, Class B
! EN 55011 (1990)/CISPR 11: Radiated and Conducted emissions.
Immunity: EN61326, Annex A
! IEC 1000-4-2 (1995): Electrostatic Discharge immunity, Criteria A. ! IEC 1000-4-3 (1995): Radiated Electromagnetic Field immunity Criteria A. ! IEC 1000-4-8 (1994): Power Frequency Magnetic Field immunity Criteria A.
Power line and I/O tests to IEC 1000-4-4, IEC 1000-4-5, IEC 1000-4-6, and IEC 1000-4-11 were not required. The device is DC powered from an I/O cable which is less than three meters long.
Declaration of Conformity based on tests conducted by Chomerics Test Services, Woburn, MA 01801, USA in April, 2005. Test records are outlined in Chomerics Test Report #EMI4193.05.
We hereby declare that the equipment specified conforms to the above Directives and Standards.
Carl Haapaoja, Director of Quality Assurance
Page 23
HM 18200-40.doc
Cole-Parmer Instrument Company
625 East Bunker Court
Vernon Hills, Illinois 60061-1844
(847) 549-7600
Fax: (847) 247-2929 (Fax)
800-323-4340
www.coleparmer.com
E-mail: techinfo@coleparmer.com
Loading...