CoaguSense Coag-Sense Prothrombin Time PT/INR Self-test User's Manual

Coag-Sense®
Prothrombin Time (PT)/INR
Self-Test User’s Manual
Page 2 For In Vitro Diagnostic Use CoaguSense, Inc.
Caution: Federal law restricts this device to sale by or on the order of a physician.
© Copyright 2013, CoaguSense, Inc. All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system or translated into any language or computer language, in any form or by any means, including, but not limited to, electronic, magnetic, optical, chemical, manual, or otherwise without written permission of CoaguSense.
Coag-Sense is a registered trademark of CoaguSense, Inc.
CoaguSense, Inc. For In Vitro Diagnostic Use Page 3
Coag-Sense® PT/INR Monitoring System
The Coag-Sense PT/INR Monitoring System is a medical device for quantitative Prothrombin Time (PT) and International Normalized Ratio (INR) testing using Coag-Sense Test Strips.
The Coag-Sense system can be used for self-testing of clotting times by patients at home. However, results obtained should not be used to adjust your medication without talking with your doctor or health care professional.
Contacting CoaguSense
If you have any questions or concerns with the Coag-Sense system, please contact CoaguSense Technical Support at:
CoaguSense, Inc. 48377 Fremont Blvd., STE 113 Fremont, CA 94538 Toll Free: 1-866-903-0890
E-Mail: techsupport@coagusense.com
WARNING: Contact your health care provider if you have any questions about your test results and any actions you should take.
Note : The Coag-Sense PT/INR Monitoring System meter is
packaged in a special box. Do not discard this box. Re­use the package to transport the meter or, if directed by Customer Service to return it for testing.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 4 For In Vitro Diagnostic Use CoaguSense, Inc.
Table of Contents
Contacting CoaguSense ...................................................................... 3
Table of Contents ................................................................................. 4
1. About This Manual ..................................................................... 5
2. System Description .................................................................... 6
3. Reordering Information ............................................................ 10
4. Warnings and Precautions ....................................................... 10
5. Hazards and Symbols .............................................................. 15
Directions for Use .............................................................................. 16
6. Operating Conditions ............................................................... 16
7. Power On and Off .................................................................... 17
8. Setting the Time and Date ....................................................... 17
9. Performing a Control Test ........................................................ 19
10. Performing a PT Test ............................................................... 24
11. Collecting a Fingerstick Sample ............................................... 29
Sample Collection and Transfer Methods ............ Error! Bookmark not
defined.
12. Reviewing the Memory ............................................................. 34
13. Control Strips ........................................................................... 35
14. Replacing the Batteries ............................................................ 35
15. Cleaning the Meter ................................................................... 38
16. Troubleshooting ....................................................................... 39
17. Performance Characteristics .................................................... 46
18. Meter Specifications ................................................................. 47
19. Warranty .................................................................................. 48
20. Index ........................................................................................ 50
About This Manual
CoaguSense, Inc. For In Vitro Diagnostic Use Page 5
1. About This Manual
The purpose of the Coag-Sense PT/INR Monitoring System User’s Manual is to help you understand your Coag-Sense system, its parts, and its intended function. It provides you with the information you need to perform PT/INR testing with the Coag-Sense system.
It is important to read this entire User Manual and inserts that came with the disposable Coag-Sense test strips. You must complete proper training on the Coag-Sense system and practice the test with a health care provider before attempting to use this device. The Coag-Sense system should only be used with a doctor’s prescription.
This User Manual has different formats and symbols to distinguish warnings, notes, and meter buttons.
WARNING: This indicates a warning or precaution. Please read and understand all warnings and precautions. They tell you about potential safety hazards and situations that may cause injury. If you have any questions, please ask your doctor or contact Technical Support (USA) at 1-866-903-0890.
Note : Notes provide additional information that is useful or
important. All notes are italicized. Words in BOLD ALL- CAPITALS refer to buttons on the Coag-Sense meter.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 6 For In Vitro Diagnostic Use CoaguSense, Inc.
2. System Description
The Coag-Sense Self-Test PT/INR Monitoring System is used for measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Coag-Sense system is intended for use outside the body by people taking warfarin or other oral anticoagulant (blood thinning) therapy who need to monitor clotting time.
The meter performs a self-test when it is first turned on. If there are any problems with the meter, an error message is shown on
the display. Refer to the “Troubleshooting” section of this manual
or contact Tech Support. A test strip is inserted and heated in the meter prior to sample application.
The strip contains a tiny wheel with spokes that pull the sample into the reaction well. The spokes quickly and completely mix the sample with the clot initiating component of the test strip.
The PT time is determined from when (a) the sample is drawn into the reaction well of the test strip and detected by a beam of light until (b) a clot forms and interrupts a beam of light. The PT result is converted to an INR (International Normalized Ratio) using the calibration data stored in the meter. INR is a mathematical correction of the PT result that adjusts for sensitivity differences among different PT systems.
The meter continues to check every feature of its operation through a system of self-checks.
System Description
CoaguSense, Inc. For In Vitro Diagnostic Use Page 7
Your Coag-Sense Self-Test PT/INR Monitoring System (Catalog number 03P50-01) comes supplied with:
Coag-Sense PT/INR Meter  Coag-Sense Self-Test User’s Manual  Coag-Sense Self-Test Quick Reference Guide  Sample Transfer Tubes  Sample package of Single-use, Auto Sterile Lancets  Four AA 1.5 V Alkaline Batteries (not installed)
You will also need:
Coag-Sense Test Strip (catalog number 03P56-50),
which includes:
50 Patient Test Strips 2 Low Control Strips 2 High Control Strips 1 Control Strip Activation Solution
Gauze or cotton balls  Lancets, 21 gauge
An optional AC Power Adapter can be ordered (catalog number 03P64-01).
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 8 For In Vitro Diagnostic Use CoaguSense, Inc.
The Coag-Sense Meter: Top View
The Coag-Sense Meter: Bottom View
Meter Display
Power On/Off Button
Menu Buttons
Strip Holder
Battery Door
Battery Door Release
System Description
CoaguSense, Inc. For In Vitro Diagnostic Use Page 9
The Coag-Sense Meter: Rear View
The Coag-Sense Test Strip
AC Adapter Jack
Data/Printer Port
Handle
Mixing Wheel and Reaction Well
Bar Code (Faces down when inserted into meter)
Sample Application Area (Green light flashing when ready for sample)
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 10 For In Vitro Diagnostic Use CoaguSense, Inc.
3. Reordering Information
Product
Catalog #
Coag-Sense Self-Test PT/INR System
03P50-01
Test Strip Kit (Qty 50)
03P56-50
Sample Transfer Tubes (Pack of 54)
03P53-54
Lancets (Box of 100)
03P58-02
AC Power Adapter (optional item)
03P64-01
Control Strip Kit -4
03P72-04
4. Warnings and Precautions
General
Patients taking Warfarin (Coumadin) and other oral
blood thinners should consult with their healthcare provider before adjusting their dosage.
Patients should consult with their doctor for the
appropriate INR therapeutic range for them.
Patients who have recently taken or are currently taking
any type of Heparin or Low Molecular Weight Heparin anticoagulant should not use this test system and should consult their doctor.
Test Site and Blood Sample
The quality of the blood sample can become the
weakest link in any PT test. A blood sample of poor quality can produce results of poor quality. Read the
Warnings and Precautions
CoaguSense, Inc. For In Vitro Diagnostic Use Page 11
section on “Collecting a Fingerstick Sample” for more
information.
Use only fresh fingerstick capillary blood for testing. The
blood should only come in contact with the products provided with the Coag-Sense System. Other products may have anti-coagulant agents on their surfaces and result in unreliable test results.
The Coag-Sense PT/INR System is for in vitro diagnostic
use only.
Blood samples must be applied to the test strip
immediately after collection or the blood begins to clot, causing unreliable results.
Squeezing the fingerstick site excessively (milking)
releases “tissue layer” fluid and tissue factors that can cause unreliable results.
The fingerstick site should be washed with warm water
and soap, and then completely dried. The site must be clean of all hand oils/lotions and foreign matter, which may cause unreliable results.
If Alcohol wipes are used, wipe the fingerstick site with a
gauze pad and make sure the site is completely dry. If any alcohol remains (or is re-introduced) on the finger, it may cause unreliable results.
The quality of fingerstick and the sample delivered are
important to the test results. If there is a question
about the sample or sample collection, obtain a new strip, repeat the fingerstick on a different finger, and test again.
If a test requires a repeat, use a different finger for the
fingerstick, since blood may have started to clot on the first finger, which may cause unreliable results.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 12 For In Vitro Diagnostic Use CoaguSense, Inc.
If there is a bubble or an air pocket showing in the blood
sample in the transfer tube, start the test over. Use a new strip and fingerstick (using a different finger and transfer tube) or results may be unreliable.
Note : If you are unable to get a PT/ INR test result contact
Technical Support at 1-866-903-0890, your home testing service provider or your healthcare provider.
Meter
Use only AA 1.5 V alkaline batteries in the Coag-Sense
meter. Rechargeable batteries cannot be used because damage to the meter may result. Do not throw used batteries in the trash. Dispose of properly.
The meter is a delicate instrument, and should be
handled with care. Dropping or other mishandling may cause damage to the meter. If this should occur, call Tech Support.
Do not allow any liquids to spill on the meter. If this
should occur, call Tech Support.
Do not put the meter in liquid. Do not allow liquids to get
into any of the connectors or plugs on the meter.
Only use the method provided in this User’s Manual to
clean the Coag-Sense meter.
Do not move or touch the meter while it is running a test.
Unreliable results may occur.
Store and use the Coag-Sense PT/INR Test System
following the instructions in this manual.
Use only the optional Coag-Sense AC power adapter
with the meter or damage to the meter may result.
Warnings and Precautions
CoaguSense, Inc. For In Vitro Diagnostic Use Page 13
The Coag-Sense PT/INR Test System is designed for
indoor use only.
This equipment generates, uses, and can radiate
radiofrequency (RF) energy. If your meter is not set up and used according to this User Manual, the RF energy may interfere with other devices in the area.
This equipment is tested to meet the limits for medical
devices, which is designed to provide a reasonable protection against harmful interference when the equipment is operated in a home environment. If not installed and used in accordance with these instructions, it may cause harmful interference to other devices in the vicinity. If this equipment does cause harmful interference to other devices, which can be determined by turning the equipment on and off, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device. Increase the separation between the equipment. Connect the equipment to an outlet on a circuit
different from that to which the other devices are connected.
In the unlikely event of an electric power surge (i.e.,
severe static discharge during a thunderstorm), when using the optional AC power adapter, the display screen may go blank. If this occurs, unplug the power supply from the back of your meter, wait 5 seconds and plug it back in. Normal operation should return, but you may have to reset the time and date.
DO NOT OPEN THE METER. Do not attempt to repair
or modify this meter. The Coag-Sense Self-Test PT Meter does not require any periodic maintenance
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 14 For In Vitro Diagnostic Use CoaguSense, Inc.
and there are no user serviceable parts inside. If you have problems, please contact technical support or your health care provider to arrange for service.
Test Strips/Control Strips/Control Strip Activating Solution
The test strips are designed for single use only. Do not
reuse the test strips.
Discard used strips in a puncture resistant, biohazard
waste (SHARPS) container.
PT Test Strips, Control Strips, and Control Strip
Activating Solution are perishable goods with a limited shelf life. Do not use any of these items if the expiration date has passed.
Refer to the package insert that is supplied with each
box of test strips for more information.
Hazards and Symbols
CoaguSense, Inc. For In Vitro Diagnostic Use Page 15
5. Hazards and Symbols
Warning. This indicates a warning or precaution,
requiring special attention
Class II Equipment. The AC Adapter is double insulated
Biological Risks: Disposable items pose biological risks. The strips and fingerstick materials should be disposed in appropriate biohazard waste containers
Electronic device. Dispose of unit and batteries properly
Use by/Expiration Date
Lot number
For In vitro diagnostic use
Storage temperature range
Manufacturer
Single Use Only – Do Not Reuse
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 16 For In Vitro Diagnostic Use CoaguSense, Inc.
Directions for Use
Patients taking Warfarin (Coumadin) and/or other oral blood thinners should consult with their doctor before adjusting their dosage. Consult your doctor for the appropriate PT and INR values for you.
Note : The Coag-Sense PT Test System meter is packaged in
a special box. Do not discard this box. Re-use the package to transport the meter or, if directed by Customer Service to return it for testing.
6. Operating Conditions
To ensure that your Coag-Sense System is working correctly, be sure the following conditions are met:
Be sure that the meter and strip are at room temperature
before use. Room temperature is between 65ºF and 90ºF (18ºC and 32ºC).
Relative humidity should be between 10% and 85%,
without condensation, for testing.
Avoid dropping the meter or treating it roughly.  Use the meter only on a level, stable surface.  Do not move or touch the meter during testing.  Do not place the meter in direct sunlight or high intensity
light.
Setting the Time and Date
CoaguSense, Inc. For In Vitro Diagnostic Use Page 17
7. Power On and Off
Install the AA 1.5 V batteries, which are needed to
maintain the time and date settings. (See “Replacing the Batteries” for more information). The optional AC power adapter can be connected at this time.
Place the meter on a flat, stable surface. To turn the
meter on. Press the POWER button, which is on the topside of the meter, beneath the display. To turn the meter off, press the same POWER button again.
8. Setting the Time and Date
Action
Meter Display
1. If the date and time have not been set before on the meter or the setting has been lost, the display looks like this with blinking characters.
2. The blinking characters may be changed by pressing + to increase or – to decrease. When correct, press NEXT to advance to the next character.
Press NEXT to advance to next character.
12:00PM
1/01/13
+
-
DONE
NEXT
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 18 For In Vitro Diagnostic Use CoaguSense, Inc.
Action
Meter Display
3. Once you have entered the correct date and time, press DONE to exit this function.
4. Once the date and time are set, the display looks like this.
5. If you need to change the date or time in the future, press SET. The meter display shows the set time and date.
6. Adjust the time and date by pressing + to increase or – to decrease. When the time and date are correct, press DONE when finished.
Note : The clock time does not adjust for daylight savings time.
Press DONE when date and time are correct
8:00AM
3/07/13
+
-
DONE
NEXT
Press SET to adjust date and time again (if necessary)
READY
INSERT
STRIP
SET
MEM
OFF
Press DONE when date and time are correct
12:00PM
1/01/013
+
-
DONE
NEXT
Setting the Time and Date
CoaguSense, Inc. For In Vitro Diagnostic Use Page 19
9. Performing a Control Test
Control testing confirms the performance of both the meter and the test strips and should be completed for each new lot of test strips and any other time as indicated by your testing service provider or doctor. Control testing can also be run whenever the PT results are unexpected to make sure that the system is working properly. There are 2 low control strips, 2 high control strips and a control strip activation solution shipped with each test strip kit. Extra controls may be ordered separately.
Follow these steps to perform a test on a low or high control.
Note : The following directions are for running a low control
strip. When this procedure is complete, run a high control strip. The controls may be run in any order. The meter will display and store the results in PT seconds only. The meter does not use or require results from the control strips prior to running a patient test strip.
WARNING: Do not move or touch the meter while it
is running a test. Unreliable results may occur.
Note : If an error message appears, consult the
“Troubleshooting” section of this manual.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 20 For In Vitro Diagnostic Use CoaguSense, Inc.
Action
Meter Display
1. Make sure that the meter is on by pressing the POWER button on the top of the meter.
2. Open a low control package, tearing at the notched end. Remove the strip. Set the package aside.
Note : Make sure that the
expiration date has not passed by checking the date on the front of the control package.
3. Holding the round end, gently push the strip completely into the meter. The strip fits snuggly when pushed all the way toward the back wall of the strip holder.
4. When the strip is correctly inserted, the display looks like this.
Note : If anything other than
READY
INSERT
STRIP
SET
MEM
OFF
LOW CONTROL
DETECTED
Performing a Control Test
CoaguSense, Inc. For In Vitro Diagnostic Use Page 21
Action
Meter Display
this is displayed, refer to the “Troubleshooting” section.
5. The meter warms the strip. The display looks like this. It shows a countdown of the time remaining during the warm-up cycle.
Note : Do not apply the control
activation solution until the warm-up is complete and the meter tells you to do so.
6. The meter beeps once when it is ready for the control strip activation solution. The screen looks like this.
Note : You now have 2 minutes to apply the activation solution to
the control strip.
7. Open the control activation solution and hold at an angle to allow insertion of the transfer tube
8. Holding the transfer tube below bulb insert into control activation solution.
Note : DO NOT SQEEZE THE BULB. Be careful to avoid getting
APPLY
CONTROL
SOLUTION
WARMING
PLEASE WAIT
24
Warming count down in seconds
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 22 For In Vitro Diagnostic Use CoaguSense, Inc.
Action
Meter Display
bubbles in the transfer tube.
9. Rest hand on instrument to steady. Move fingers to flat sides of bulb being sure to cover air hole. Insert tip into sample application well of test strip, touching tip down on strip at flashing green light. Squeeze bulb until solution leaves tube.
Note : Keep pressure on bulb and pull away from strip (avoids back suction)
10. When the control strip activation solution is properly applied and detected the flashing green light will turn off, and the meter display looks like this.
Note : If this screen is not displayed, either not enough solution
was applied or the solution had bubbles in it. Remove the strip. Retest with a new control strip. Do NOT attempt to add more solution to the control strip.
11. When the low control testing is complete, the display
shows “OK” and looks
similar to this.
TESTING
PLEASE
WAIT
LO CONTL OK
PT 12.8
12/18/13 4:01 PM
REMOVE STRIP
Performing a Control Test
CoaguSense, Inc. For In Vitro Diagnostic Use Page 23
Action
Meter Display
Note : To avoid confusing control strip INR results with patient test
strip INR results, the high control test will display the control result in PT seconds only.
Note : If anything other than this is displayed, refer to the
“Troubleshooting” section.
Note : The date and time shown in the display are examples only.
The date and time shown after actual testing is the current date and time.
Note : Remember to repeat this entire procedure with a high
control strip.
12. When high control testing is complete, the display
shows “OK” and looks
similar to this.
Note : If an error message appears, refer to the “Troubleshooting”
section.
13. Once the controls have been successfully tested, remember to throw the control strips into a biohazard (SHARPS) container. You can now proceed to test your blood. If you are not going to test, turn off the meter by pressing the POWER button.
HI CONTL OK
PT 42.1
12/18/13 4:05 PM
REMOVE STRIP
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 24 For In Vitro Diagnostic Use CoaguSense, Inc.
10. Performing a PT Test
WARNING: Place the meter on a stationary, level surface for testing. Do not move the meter or allow it to vibrate during a test. Unreliable results may occur.
Action
Meter Display
1. Make sure that the meter is on by pressing the POWER button on the top of the meter.
Note : If an error message
appears, refer to the
“Troubleshooting”
section.
2. Open a PT test strip package, tearing at the notched end. Remove the strip. Set the package aside.
Note : Make sure that the expiration date has not passed.
3. Holding the round end, gently push the strip completely into the meter in one smooth motion. The strip fits snuggly when pushed all the way toward the back wall of the strip holder.
READY
INSERT
STRIP
SET
MEM
OFF
Reviewing the Memory
CoaguSense, Inc. For In Vitro Diagnostic Use Page 25
Action
Meter Display
4. When the test strip is correctly inserted, the display looks like this.
5. The meter warms the strip. The display looks like this. The meter counts down the time remaining during the warm-up cycle.
Note : Do not apply any test
sample until the warm­up completes and the meter tells you do so.
While the meter is warming up, get ready to perform a
fingerstick. See “Collecting a Fingerstick Sample” in this
manual.
6. When the warm-up completes, the meter beeps once. The screen looks like this.
Note : You now have 2 ½
minutes to apply the sample to the test strip.
PAT.
STRIP
DETECTED
SET
MEM
OFF
APPLY
SAMPLE
SET
MEM
OFF
WARMING
PLEASE WAIT
24
Warming count down in seconds
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 26 For In Vitro Diagnostic Use CoaguSense, Inc.
Action
Meter Display
7. Immediately after collecting the patient sample, place the tip of the sample transfer tube at a 45º angle into the sample well on the test strip where you see the flashing green light. Gently touch the tip down onto the sample well.
8. Slowly squeeze the bulb until the blood leaves the tube being careful not to introduce air bubbles into the sample. Keep pressure on bulb while you pull your hand away to avoid back suction of sample.
9. Discard the sample transfer tube in a biohazard waste container.
10. When the sample is detected, the meter display looks like this.
Note : If this screen is not displayed, either not enough blood
sample was applied or the sample had bubbles in it. Remove the strip and retest with a new strip.
TESTING
PLEASE
WAIT
Reviewing the Memory
CoaguSense, Inc. For In Vitro Diagnostic Use Page 27
Action
Meter Display
11. When testing is complete, the meter beeps once. The results (INR and PT in seconds) are shown on the screen.
Note : If any display other than the one shown is visible (such as
CLOT TIME TOO SHORT or NO CLOT DETECTED), refer
to the “Troubleshooting” section.
12. Record the results. Then remove the test strip. Throw it away in the biohazard container.
Note : Repeat the test if the results seem unusually low or high. If
the results still seem unusual after a second test, contact your doctor or CoaguSense Tech Support.
Turn the meter off by pressing the POWER button when you are finished testing. If left unattended, the meter automatically turns off in a few minutes.
The last 100 test results are stored in memory with the time and date. Refer to “Reviewing the Memory” in this manual for more information.
INR 2.2
PT 22.2
1/18/13 5:50 PM
REMOVE STRIP
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 28 For In Vitro Diagnostic Use CoaguSense, Inc.
WARNING: Unexpected results
An unexpected result may include any result that falls outside the therapeutic target range, or a result that falls inside the target range but is not consistent with your current health status (e.g., you have bleeding or bruising).
What can cause unexpected results:
1. Certain prescription drugs (for example, heparin) and certain over-the-counter medications (for example, antibiotics) can affect the action of oral blood thinners and the INR value.
2. Changes in diet, lifestyle, or taking nutritional supplements such as ginkgo biloba can affect the action of oral blood thinners and the INR value.
3. Liver diseases, congestive heart failure, thyroid dysfunction, Lupus, antiphospholipid antibody syndrome (APS) and other diseases or conditions can affect the action of oral blood thinners and the INR value.
Be sure your doctor is aware of any of these conditions before you begin testing, and any time there are changes in health status or medications after you have begun testing.
What to do when you get an unexpected result:
Follow instructions for re-testing on the Coag-Sense PT/INR meter. For unexpected results, contact Technical support at 1-866-903-0890. Always follow your doctor’s instructions for adjusting your dose of anticoagulant medication, or any other corrective actions.
Reviewing the Memory
CoaguSense, Inc. For In Vitro Diagnostic Use Page 29
11. Collecting a Fingerstick Sample
Tips for a Successful Fingerstick
Make sure that you have all the supplies needed before
you start.
Lancet device Sample Transfer Tubes Sterile alcohol prep pad Gauze square or cotton ball Band-Aid (optional) Puncture resistant biohazard container (SHARPS)
Do not use fingers with tight rings, scars, calluses, or
other features that prevent getting good access to the blood.
Thoroughly clean your hands by washing with soap and
warm water first. The warm water also helps blood circulation. Completely dry your hands. Instead, you may use Isopropyl alcohol wipes and then dry the area with gauze.
For better blood flow, keep your hand below your heart.  One of the middle or index fingers on either hand is
recommended.
Gently squeeze or massage the finger to be lanced, near
the tip. Good circulation can be seen if your fingertip changes to a pinkish shade.
Gently squeeze or massage the finger to be lanced, near
the tip. Good circulation can be seen if your fingertip changes to a pinkish shade.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 30 For In Vitro Diagnostic Use CoaguSense, Inc.
Note : If the fingertip color is not changing, then warm the
fingers carefully by running them under warm water or rubbing them vigorously. Remember to dry them completely or your results may be unreliable.
Always use new unused materials (such as lancet,
gauze, etc.).
If you are being helped when getting a test sample, your
helper should wash their hands with soap and warm water. They should also put on gloves before doing the fingerstick.
Use a 21g 1.8 mm depth single-use auto-disabling
lancet. Smaller gauge/shallow depth lancets (i.e. diabetes 23g lancets) should not be used. Refer to the Lancet device instructions for more information on use.
Lance the fleshy part of the fingertip just slightly left or
right of the center. Press lancet firmly against finger.
Refer to the package inserts for the Lancet device for
more information.
The best test sample is when:
The blood is collected right after the fingerstick and put
into the test strip sample well without delay. If the blood is not collected or tested quickly (within 15 seconds), repeat with a fresh fingerstick and a new strip.
There are no bubbles or air pockets in the tube or
sample.
Reviewing the Memory
CoaguSense, Inc. For In Vitro Diagnostic Use Page 31
Action
Meter Display
1. Wash hands with soap and warm water. Dry completely.
2. Choose a site toward the center of one of the middle fingers to lance.
Note : Avoid the more
sensitive area in the center. Stay away from any calluses or scars.
3. If desired, clean the fingertip with an alcohol wipe using one side for the first cleaning. Use the second side for a final wipe.
4. Dry the fingertip with gauze to remove any excess alcohol.
Note : Residual alcohol will affect results. Be certain that finger is completely dry.
5. Hold the finger tightly between your thumb and index finger, or place hand on a table with the palm facing up. Keep your fingertip from touching anything so that it remains clean and ready.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 32 For In Vitro Diagnostic Use CoaguSense, Inc.
Action
Meter Display
6. Remove the cap from the single use lancet. Place it against the skin. Holding the body of the lancet, push down firmly against the finger to lance the surface of skin.
Note : The blood should flow
freely. If it doesn’t,
gently squeeze the finger to get it started. Lowering your hand and arm to your lap so that the fingertip is below the heart helps the blood drop form.
WARNING: Squeezing the fingerstick site excessively (milking)
releases “tissue layer”
fluid that can cause unreliable results.
Reviewing the Memory
CoaguSense, Inc. For In Vitro Diagnostic Use Page 33
Action
Meter Display
7. When ready to collect the drop of blood, hold the Sample Transfer Tube between your thumb and forefinger below the bulb, being sure not to cover the air hole in bulb. DO NOT SQUEEZE THE BULB.
8. With tube horizontal touch tip to bead of blood. Let capillary action fill capillary portion of tube until blood flow stops. Squeeze finger to produce additional blood if require to completely fill capillary portion of tube.
9. Once you have collected the sample, immediately put it into the sample well on the
test strip. See “Performing a PT Test” section of this
manual.
10. Apply gauze to your fingerstick site with firm pressure until the bleeding stops. Apply a bandage if desired.
11. Throw the lancet and the strip into a biohazard container.
WARNING: If there is a bubble or an air pocket showing in the blood sample in the sample transfer tube, start the test over with a new fingerstick on a different finger.
Note : The minimum amount
must be collected before adding it to the test strip.
Never add more blood to the
test strip.
Capillary portion of tube ends here
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 34 For In Vitro Diagnostic Use CoaguSense, Inc.
12. Reviewing the Memory
The Coag-Sense meter stores up to 100 results along with the date and time, in its memory. When the 100th result is reached, the first result 1 is replaced (written over) with test results for test number 101. This continues with the oldest result being replaced with the most recent. Memory is not lost if there is a break in power for any length of time. Memory cannot be erased.
Action
Meter Display
The memory can be accessed from any mode that displays a MEM button.
1. Press MEM. The meter displays the last two records.Press PREV or NEXT to scroll through the result records.
3. Press DONE when finished.
Note : If any messages are
displayed, such as
CLOT TIME TOO SHORT, or NO CLOT DETECTED refer to the
“Troubleshooting”
section of this manual.
Example of display with two most recent records:
2/07/13 12:56 PM INR 5.2 PT 49.1
4/07/13 12:56 PM INR 2.9 PT 28.8
NEXT
PREV
DONE
Replacing the Batteries
CoaguSense, Inc. For In Vitro Diagnostic Use Page 35
13. Control Strips
Quality control is an important part of PT time testing. Using control strips make sure that you are carrying out the steps of the test correctly. It also ensures that your Coag-Sense System is working properly and that test strip integrity has been maintained. Both a high and low control should be tested before running the first PT test strip from a new box. A control can also be tested if you have unexpected results. There are 2 low control strips, 2 high control strips and control activation solution shipped with each box of test strips. Additional controls may be ordered separately. See “Performing a Control Test” in this manual.
14. Replacing the Batteries
It is not necessary to install batteries in the meter if the optional AC adapter is plugged into a wall socket. However, if no batteries are installed and the meter is unplugged or there is a power outage, the time and date settings go back to the factory settings.
The meter is designed for AA 1.5 V alkaline batteries only. It does not re-charge these alkaline batteries when connected to AC power.
WARNING: Use only AA 1.5 V alkaline batteries in the Coag-Sense meter. Rechargeable batteries should not be used as they can result in damage to the meter.
The AA 1.5 V batteries can last for approximately 300 tests if the meter is operating only on batteries. The batteries can last
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 36 For In Vitro Diagnostic Use CoaguSense, Inc.
through their shelf life if the optional AC power adapter is plugged in.
If the batteries are running low, the meter displays a BATTERY LOW message. The meter can run another one or two additional PT tests, but the batteries should be replaced as soon as possible. You can plug the optional AC power adapter into a wall socket and replace the batteries later.
When the message BATTERY TOO LOW SEE MANUAL appears on the meter display, the meter shuts off after a brief delay. The batteries must be replaced or the meter must be connected to a wall socket with the optional AC power adapter to continue testing. The meter time and date settings are lost and the meter will need to have the date and time reset. See “Setting the Time and Date” in this manual.
Complete the following steps to replace the AA 1.5 V alkaline batteries.
1. Turn the meter upside down.
2. Remove the battery door by pressing on the battery door release.
Replacing the Batteries
CoaguSense, Inc. For In Vitro Diagnostic Use Page 37
3. Remove the old batteries and replace with 4 new standard
1.5V AA alkaline batteries. (The proper direction for battery placement is shown on a figure inside the battery compartment).
4. Replace the battery door.
5. Properly dispose of old
batteries.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 38 For In Vitro Diagnostic Use CoaguSense, Inc.
15. Cleaning the Meter
No maintenance is required other than routine cleaning.
Clean the outside of the Coag-Sense meter with a clean damp non-abrasive cloth. If required, a mild disinfectant (such as a 10% bleach solution or 70% isopropyl alcohol) may be used. Apply the cleaning solution to the cloth to dampen. Then put cloth on the meter. Alcohol prep pads may also be used.
WARNING: Do not put the meter in liquid. Do not allow liquids to get into any of the connectors or plugs on the meter.
WARNING: Do not allow any liquids to spill on the meter. If this should occur, unplug the meter (if plugged in) and call Tech Support.
Troubleshooting
CoaguSense, Inc. For In Vitro Diagnostic Use Page 39
16. Troubleshooting
You may see the following error messages while using the Coag­Sense meter. This section discusses how to resolve most problems that you might encounter. If you have any questions or problems during the troubleshooting process, note the display wording and contact Tech Support at 866-903-0890 or email
support@coagusense.com.
Meter Display
Possible Cause
Solution
REMOVE STRIP
Meter turned off with used strip in it.
If no strip present possible shipment damage.
Remove the strip and begin again.
Call Tech Support
NO SAMPLE DETECTED
Either no sample or not enough sample was applied to the strip within 2 1/2 minutes
after the “Apply Sample” message was
displayed. This can also happen if sample is applied on the strip but outside of the sample application well.
Repeat the entire procedure (including fingerstick on a different finger) with a new strip.
Apply the sample
within 2 1/2 minutes after display of the
“Apply Sample”
message.
Ensure that the
pipette tip/transfer tube touches the sample well before dispensing sample.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 40 For In Vitro Diagnostic Use CoaguSense, Inc.
Meter Display
Possible Cause
Solution
CLOT TIME TOO SHORT
The clotting time was very short and out of testing range (<8 seconds).
An air bubble was detected in the sample.
Repeat the entire procedure (including fingerstick on a different finger) with a new strip.
Visually confirm that no air bubbles are in the sample before applying to test strip.
Gently squeeze the transfer tube bulb until the blood exits being sure not to introduce air bubbles.
If the same message repeats, contact Technical Support.
Troubleshooting
CoaguSense, Inc. For In Vitro Diagnostic Use Page 41
Meter Display
Possible Cause
Solution
NO CLOT DETECTED
The sample clotting time was very long and out of testing range.
There was insufficient sample transferred to the test strip. Possible causes include; improper lancing (21g lancet required), an air bubble in the sample, not allowing sample to completely fill glass portion of transfer tube, or the sample was drawn back into the transfer tube before removing tip from the test strip well..
Visually confirm that no air bubbles are in the sample and that the glass portion of the transfer tube is full before applying to test strip.
Release pressure on the transfer tube bulb only after removing transfer tube from sample application well.
Repeat the entire procedure (including fingerstick) with a new strip. If the same message displays, use an alternative testing method and contact Tech Support.
TEST STRIP EXPIRED SEE MANUAL
The lot of strips has expired.
Date is incorrect.
Use a different lot of strips that has not expired.
Verify the date setting on the meter is current.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 42 For In Vitro Diagnostic Use CoaguSense, Inc.
Meter Display
Possible Cause
Solution
CONTROL OUT OF RANGE
The control strip result is outside of its acceptable range. This may be due to a problem with the shipment/storage of the control strips or the control activation solution. Plasma on control strips have a limited shelf life and the clotting time will change when exposed to temperatures outside the storage range.
Repeat test with another control strip. If the second test is out of range, contact Tech Support.
Control strips should be tested immediately upon receipt of your shipment of new test strips as they have a limited shelf life.
ROOM TEMP INCORRECT SEE
MANUAL
The temperature of the room is either below or above the operating temperature range of the meter.
Move the meter to a place that is within the operating temperature range of the meter (55ºF to 95ºF, 13ºC to 35ºC) and allow meter time to adjust to correct temperature. Repeat testing.
WHEEL PROBLEM
The test strip was inserted at an incorrect angle or speed.
There may be a problem with the wheel on the strip or with the meter
A used strip was inserted.
Reinsert the strip holding the back of the meter steady with one hand while inserting the strip completely using a quick smooth motion with the other hand. If display persists try again with another strip.
If the message displays again, contact Tech Support.
Troubleshooting
CoaguSense, Inc. For In Vitro Diagnostic Use Page 43
Meter Display
Possible Cause
Solution
BAR CODE READ FAILURE
The strip was inserted at an incorrect angle or speed.
There may be a problem with the bar code on the strip or with the meter.
The meter is in direct sunlight or near a high­intensity light source.
Take the strip out and reinsert holding the back of the instrument steady with one hand while inserting the strip completely with the other hand. Insert the strip using a quick smooth motion.
Move the meter indoors into room lighting or away from light source.
If error persists try again with another strip. If the message displays again, contact Tech Support.
HEATER PROBLEM
The meter is too warm or too cold, or there may be a problem with the meter.
Move the meter to a place that is within the operating temperature range of the meter (55ºF to 95ºF, 13ºC to 35ºC) and allow meter time to adjust to correct temperature. Repeat testing.
Turn meter off then on again.
Try again with another strip.
If the display persists contact Tech Support.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 44 For In Vitro Diagnostic Use CoaguSense, Inc.
Meter Display
Possible Cause
Solution
DETECT PROBLEM
There may be a problem with the strip or with the meter.
Take the strip out and reinsert holding the back of the instrument steady with one hand while inserting the strip completely with the other hand. Insert the strip using a quick smooth motion.
Try again with another strip.
If the message persists contact Technical Support.
LIQUID PROBLEM
There may be a problem with the strip or with the meter.
Take the strip out and reinsert holding the back of the instrument steady with one hand while inserting the strip completely with the other hand. Insert the strip using a quick smooth motion.
Try again with another strip.
If the message persists contact Technical Support.
MOTOR PROBLEM
There may be a problem with the strip or with the meter.
Try again with another strip or contact Tech Support.
Troubleshooting
CoaguSense, Inc. For In Vitro Diagnostic Use Page 45
Meter Display
Possible Cause
Solution
BATTERY LOW
The meter batteries need to be replaced.
The meter can complete the current test, as well as one or two more tests. However, the batteries should be replaced as soon as possible.
BATTERY TOO LOW
The meter batteries must be replaced.
Replace batteries to continue with testing.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 46 For In Vitro Diagnostic Use CoaguSense, Inc.
17. Performance Characteristics
Expected Values: Fingerstick whole blood results are reported
in units equivalent to a reference method.
Measuring Range: INR 0.8 to 8.0 Normal Range: The following example represents a common
normal range for the Coag-Sense System. INR: 0.7 to 1.2
PT: 8.0 to 15.0
Warranty
CoaguSense, Inc. For In Vitro Diagnostic Use Page 47
18. Meter Specifications
Operating Temperature
55°F to 95°F (13°C to 35°C)
Operating Humidity
10% to 85% (without condensation)
Storage Temperature
-4°F to 122°F (-20°C to 50°C)
Storage Humidity
10% to 95% (without condensation)
Memory
Capable of storing 100 tests with time and date
Battery
Quantity 4 of 1.5V (AA) alkaline batteries.
AC Input
120 VAC (Use Coag-Sense Adapter Only)
Power Output
6.0V, 2.0A
Blood Sample Size
10-12 µL
Data Port
RS232
Size
3” (7.6 cm) x 6.5” (16.5 cm) x 5.75” (14.5 cm)
Weight
With 4 AA 1.5 V alkaline batteries:
1.2 lb (0.5 kg)
Equipment Classification
Class II with external power supply. Internally powered when operated with
batteries. IPXO rating.
WARNING: Use only the Coag-Sense AC power adapter or damage to the meter may result.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 48 For In Vitro Diagnostic Use CoaguSense, Inc.
19. Warranty
Limited One (1) Year Warranty
Use of the Coag-Sense PT/INR Monitoring System
The Coag-Sense PT/INR Monitoring System is designed for use in monitoring patients on oral anticoagulant therapy. Proper adherence to the instructions in this User Manual and package insert are critical to proper operation. WARNING: Failure to
comply with the User Manual could lead to inaccurate PT/INR results which could lead to incorrect medication dosing which could lead to injury or death.
Limited Warranty
CoaguSense warrants that the Coag-Sense meter is free from all defects in material and workmanship for a period of one (1) year from date of purchase. When the meter is used for the intended purpose and in the appropriate manner, the remedy is repair or replacement at CoaguSense's option. The warranty does not apply to a meter damaged by misuse, alteration or tampering to either the hardware or software. Contact CoaguSense Tech Support at 886-903-0890 for instructions.
THIS WARRANTY APPLIES ONLY TO THE METER. COAGUSENSE'S ENTIRE LIABILITY IN CONNECTION WITH THE METER, REGARDLES OF THE LEGAL OR EQUITABLE BASIS OF ANY CLAIM, IS LIMITED TO THE PURCHASE PRICE OF THE METER. IN NO EVENT SHALL COAGUSENSE, INC. BE LIABLE TO THE PURCHASER FOR ANY INCIDENTAL, CONSEQUEN-TIAL (INCLUDING BUT NOT LIMITED TO LOSS OF INCOME OR PROFITS) SPECIAL, INDIRECT, OR PUNITIVE DAMAGES ARISING FROM OR IN
Warranty
CoaguSense, Inc. For In Vitro Diagnostic Use Page 49
ANY WAY CONNECTED WITH THE PURCHASE OR OPERATION OF THE METER OR ITS PARTS. NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IS IMPLIED FROM THE SALE OF THE COAG-SENSE PT/INR TEST SYSTEM. NO WARRANTY, EXPRESS OR IMPLIED (IF ANY) SHALL EXTEND FOR A LONGER DURATION THAN THE DURATION OF THE EXPRESS WARRANTY STATED ABOVE.
Instructions for Meter or Product Return
Upon review and agreement with CoaguSense Customer Service, you may be directed to return the unit. Should this occur, clean the outside surface using 5% bleach solution or 70% isopropyl alcohol as described in the Maintenance” section. The original packaging may be required for this purpose. If this is not available, a cushioned shipping box must be used to return the meter.
Coag-Sense® Prothrombin Time (PT)/INR Monitoring System Self-Test User’s Manual
Page 50 For In Vitro Diagnostic Use CoaguSense, Inc.
20. Index
A
AC Adapter, 18 AC Power, 36 Air Bubble, 34 Alcohol Wipe, 30
B
Bar Code Read Failure, 44 Batteries, 18, 36 Battery Low, 46 Battery Too Low, 46 Blood Flow, 30 Blood- Minimum Amount, 34
C
Calibration, 14 Class II Equipment, 15 Cleaning the Meter, 39 Clock, 19 Clot Time Too Long, 42 Clot Time Too Short, 41 Collecting a Fingerstick
Sample, 30 Control Out of Range, 43 Control Strip, 36 Control Strip Activation
Solution, 23 Customer Service, 3, 17, 54
D
Date, 18, 19, 28, 35, 36, 37 Detect Problem, 45
E
E-Mail Support, 3, 54
Expected Values, 47
H
Hazards and Symbols, 15 Heater Problem, 44
I
IEC 60601-1, 15 INR, 3, 28 Install Batteries, 36
L
Lancet, 33 Liquid Problem, 45 Low Control Strip, 20
M
Measuring Range, 47 Memory, 28, 35 Meter, 9 Motor Problem, 45
N
No Sample Detected, 40 Normal Range, 47
P
Performing a Control Test, 20 Performing a Patient Test, 25 Prothrombin Time, 3 PT, 3, 28 PT Test Strip Package, 25
R
Remove Strip, 40
CoaguSense, Inc. For In Vitro Diagnostic Use Page 51
Room Temp Incorrect, 43
S
Squeezing the Fingerstick Site,
33 System Description, 6
T
Test Strip, 9
Troubleshooting, 40
U
Unexpected Results, 29
W
Warnings and Precautions, 10 Warranty, 49 Wheel Problem, 43, 44
Technical Support CoaguSense, Inc.
48377 Fremont Blvd., STE 113 Fremont, CA 94538 Toll Free: 1-866-903-0890
E-Mail: techsupport@coagusense.com
CoaguSense, Inc. 48377 Fremont Blvd, STE 113 Fremont, CA 94538 USA Tel: (510) 270-5442
Fax: (510-226-6540
EMERGO EUROPE Molenstraat 15 2513 BH, The Hague The Netherlands
Product Made in U.S.A.
REF Part No. 03P51-01
CSI P/N 100219 Rev D
Loading...