CMi Intoxilyzer S-D2 Operator's Manual

S-D2
OPERATOR’S MANUAL
650132 Lev A
Operator’s
316 E. 9th St. Owensboro, KY 42303 866-835-0690 Fax: 270 -685-6268 www.alcoholtest.com
S-D2
Manual
No part of this work covered by the copyrights hereon may
be reproduced or copied in any form or by any means—
graphical, electronic, mechanical, including photocopying, taping
or information storage and retrieval systems—without
© 2002 by CMI, Inc.
written permission of CMI, Inc.
INTOXILYZER® S-D2 Operator’s Manual
INTRODUCTION . . . . . . . . . . . . . . . . . . . 1
PRINCIPLES OF OPERA TION . . . . . . . . . 2
• Instrument Features . . . . . . . . . . . . . 3
Mouthpiece
Sampling Port “SET” Button “READ” Button Alcohol Level Display Calibration Control Breath Sampling Lights Battery Leather Case
SUBJECT BREATH TEST . . . . . . . . . . . . 7
• Ambient T emperature . . . . . . . . . . . . 7
• Preliminary . . . . . . . . . . . . . . . . . . . . 8
• Detailed Procedure . . . . . . . . . . . . . . 8
Ready Check
SET Attach Mouthpiece Instruct the Subject Take Sample Observe Display Discard Mouthpiece Reset and Wait
CALIBRA TION REQUIREMENTS: USE OF
THE DRY GAS STANDARD . . . . . . . . . . 14
TABLE OF
CONTENTS
INTOXILYZER® S-D2 Operator’s Manual
• The Alcohol Standard . . . . . . . . . . . . 14
• The Gas Standard . . . . . . . . . . . . . . . 15
• Using a Dry Gas Standard
at High Altitude . . . . . . . . . . . . . . . . 15
• Altitude Correction Chart . . . . . . . . . . 16
• Use of a Wet-Bath Simulator . . . . . . . 17
CALIBRA TION CHECK PROCEDURE . . . 18
• Detailed Procedure . . . . . . . . . . . . . . 18
Ready Check
SET Sample the Dry Gas Standard Read the Display Reset and Wait
CALIBRATION ADJUSTMENT . . . . . . . . . 21
• The Calibration Control . . . . . . . . . . . 21
• The Dry Gas Calibration Value . . . . . . 21
• Detailed Procedure . . . . . . . . . . . . . . 22
Preliminary
Ready Check SET Sample the Dry Gas Standard Adjust the Display to the Value of the Dry Gas Standard Reset and Wait
ROUTINE FIELD SERVICE CHECKS . . . . 27
• Battery Check & Replacement . . . . . . 27
• Sampling System Check . . . . . . . . . 28
• Breath Sampling Light Check . . . . . . . 29
• Test Fuel Cell Switch . . . . . . . . . . . . 31
• Fuel Cell Replacement . . . . . . . . . . . 31
POINTS TO REMEMBER . . . . . . . . . . . . . 32
• Test Procedure . . . . . . . . . . . . . . . . . 32
• Mouthpiece . . . . . . . . . . . . . . . . . . . . 33
• Smoking . . . . . . . . . . . . . . . . . . . . . . 33
• Food and Drink . . . . . . . . . . . . . . . . . 33
• Storage Between Tests . . . . . . . . . . . 33
• Radio Transmitters . . . . . . . . . . . . . . 34
• Physical Shock . . . . . . . . . . . . . . . . 34
INTOXILYZER® S-D2 SPECIFICA TIONS . . 35
INTOXILYZER® S-D2 Operator’s Manual
workplace safety applications. The S-D2 is just one of the range of instruments manufactured by CMI, Inc. for these purposes.
The Intoxilyzer® S-D2 is accurate and reliable, allowing a complete breath test procedure to be conducted in about one minute.
INTRODUCTION
The Intoxilyzer® S-D2 represents a well established concept in breath alcohol testing analysis. It is used worldwide in law enforce­ment, transportation and
This manual describes the operation, maintenance, calibration check, and calibration adjustment of the S-D2. This manual should be read completely and fully understood by each operator prior to testing a subject. It is further recommended that operators practice the breath testing process before giving an actual “in the field” test.
The information contained in this manual concerning operational procedures, protocols, accuracy verification schedules and techniques is presented solely for the purpose of establishing a general knowledge of the instrument and its operation. CMI recognizes that many law enforcement customers have agencies responsible for specifying instrument procedures, protocols and techniques based upon their individual needs, current laws and pre-existing administrative rules. Therefore, if any such procedure, protocol, schedule or technique described in this manual conflicts with the rules set forth by a recognized
1
governing agency, CMI defers to those authorized entities.
INTOXILYZER® S-D2 Operator’s Manual
PRINCIPLES OF
OPERATION
The Intoxilyzer® S-D2 uses an electrochemical fuel cell, containing two platinum electrodes, to detect and measure the concentration of alcohol vapor in expired breath. When breath is drawn into this fuel cell, by means of the sampling system, a small voltage is generated in proportion to its breath alcohol concentration. This fuel cell is fed to an electronic amplifier and displayed on a digital meter (liquid crystal).
The S-D2 incorporates two breath sampling lights, controlled by an interlinked pressure switch and timer system. The sampling lights show the operator if the subject is blowing correctly , and when he has provided a suitable sample of breath for analysis.
The instrument is simple to operate and may be used as often as required, provided that a suitable delay is allowed between successive tests. This time delay allows the fuel cell to clear itself of alcohol and prevents the possibility of additive readings. If no alcohol is present in a test, a second test may be analyzed immediately , since the fuel cell voltage is already at zero. Unless the breath alcohol level of the subject is very high,
2
INTOXILYZER® S-D2 Operator’s Manual
the instrument will generally be clear enough to receive and analyze the second sample in less than two minutes.
INSTRUMENT FEA TURES
1) Mouthpiece This is attached to the sampling port. For
hygenic reasons, mouthpieces are supplied separately packed and are disposable. A new mouthpiece must be used for each breath test. This minimizes health concerns and prevents cross-sample alcohol contamination.
2) Sampling Port This forms the entrance to the fuel cell detector.
When inserted into the small hole in the side of the mouthpiece, it allows a small portion of breath to be drawn into the instrument for analysis.
3) ‘SET’ Button This button forms part of the sampling system.
When fully depressed, the button locks to set the instrument ready for sampling. When the ‘SET’ button rises, the sample to be analyzed is drawn directly into the fuel cell detector.
4) ‘READ’ Button This button has two functions:
1) to release the ‘SET’ button and take
the sample,
2) to switch on the amplifier and display
systems
3
INTOXILYZER® S-D2 Operator’s Manual
2
1
4
3
5
7
8
INSTRUMENT FEATURES
1 Mouthpiece 2 Sampling Port
3 ‘SET’ Button
6
9
4 ‘READ’ Button 5 Alcohol Level Display 6 Calibration Control (on side) 7 Breath Sampling Lights 8 Battery (on bottom) 9 Leather Case
4
INTOXILYZER® S-D2 Operator’s Manual
The ‘READ’ button is spring loaded. Momentary depression will take the breath sample. Constant pressure is required to switch on the amplifier and display systems.
5) Alcohol Level Display This three digit, liquid crystal display shows
the breath alcohol concentration of the subject. The display has built-in illumination and is activated by depressing the ‘READ’ button.
If the figure on the left shows ‘L ’ when the ‘READ’ button is depressed, the battery needs to be replaced.
6) Calibration Control This screw control, located on the right side of the
case, is used for periodic calibration adjustments with either a dry gas standard or wet bath simulator.
7) Breath Sampling Lights Light ‘A’ illuminates to indicate that the subject
is blowing hard enough to obtain a proper breath sample. Light ‘B’ illuminates when the subject has blown long enough and indicates when the ‘READ’ button should be pressed. In other words, light ‘B’ illuminates when the subject has provided a suitable sample for breath analysis.
8) Battery The battery is located directly behind the sliding
base on the bottom of the S-D2. It powers the amplifier, digit al display and sampling lights and
5
should be replaced when the letter ‘L’ appears on the left side of the display .
9) Leather Case The S-D2 is supplied in a leather protective case.
The unit should be kept inside the case at all times, except when calibrating, changing the battery or during maintenance.
INTOXILYZER® S-D2 Operator’s Manual
6
INTOXILYZER® S-D2 Operator’s Manual
basic steps. These should be understood and followed to insure maximum efficiency of operation.
OPERATION CHECK LIST
SUBJECT
BREATH
TEST
The operating sequence for testing a subject with the Intoxilyer® S-D2 is simple, consisting of the following
1. Ready Check
2. Set
3. Attach Mouthpiece
4. Instruct the Subject
5. Take Sample
6. Observe Reading
7. Remove and Discard Mouthpiece
8. Reset and Wait
AMBIENT TEMPERA TURE
The Intoxilyzer 32-104° degrees Fahrenheit (0-40° Celsius) temperature range. Keeping the instrument within this range insures minimum condensation of alcohol and water from the
®
S-D2 is designed primarily for use in the
7
INTOXILYZER® S-D2 Operator’s Manual
breath and permits both accurate and rapid breath alcohol measurements.
It is recommended that in very cold weather the instrument is stored in a pocket (preferably inside a coat) and returned there after use.
PRELIMINARY
Ask the subject when he/she last took anything by mouth. Some foods and even “non-alcoholic” drinks may contain traces of alcohol, which the subject may later claim affected the result of the test through a “mouth alcohol” effect. To prevent this, wherever possible, insure that a delay of about 20 minutes has elapsed since the subject took anything by mouth—even medicines which may contain alcohol.
Do not even allow the subject a glass of water prior to the test since this will cool the mouth and dilute the saliva, temporarily reducing the amount of alcohol in the breath, and, consequently , the instrument reading.
Insure that no radio transmitter is currently being used in the immediate vicinity of the test.
DET AILED PROCEDURE
This section describes in detail how the Intoxilyzer
®
S-D2 is used in a breath test and examines what is actually happening at each stage of the operation. These instructions are also given in the form of a checklist on the card which accompanies the instrument as well as in the “Introduction” section of this operator’s manual.
1) READY CHECK The instrument should first be checked to insure it is
ready to receive a sample. This is accomplished by
8
INTOXILYZER® S-D2 Operator’s Manual
checking to insure the fuel cell is discharged and free of alcohol from any previous sample. Elimination of alcohol from the fuel cell should take no longer than two minutes—except in unusual cases—depending on how much alcohol was actually present in the last sample.
To conduct a READY CHECK, depress the ‘READ’ button and hold it down for at least ten seconds. This will release the ‘SET’ button and switch on the amplifier and display circuits. Observe the display; it should not exceed .002 after ten seconds. If the display does not show .002 or less during the ten second READY CHECK, the fuel cell may have traces of residual alcohol from a previous sample. If this occurs after a previous test, depress the ‘SET’ button to lock and wait one minute before repeating the READY CHECK. If the left digit shows ‘L’, replace the battery.
To complete the READY CHECK, depress the ‘READ’ button for at least ten seconds.
2) SET Once the READY CHECK is complete, the sampling
system must now be SET to prepare it to draw a breath sample into the fuel cell for analysis. Press down the ‘SET’ button until it locks. This pushes down the diaphragm and holds it against a spring-loaded catch. This action also places a short-circuit across the fuel cell,
9
which accelerates its inter-sample recovery time.
Loading...
+ 28 hidden pages