have READ this
MANUAL and YOU
UNDERSTAND its
contents. *
These WARNINGS are
included for the health
and safety of the
operator and those in
the immediate vicinity.
*If you are using a Clemco Distributor
Maintenance and Parts Guide, refer to
the orange warnings insert preceding
the Index before continuing with the
enclosed instructions.
Electronic files include a Preface
containing the same important
information as the orange cover.
PREFACE
• Employers are responsible for identifying all job site hazards, educating and training all persons who will operate and
maintain these products, and ensuring that all blast operators and their assistants understand the warnings and
information contained in these instructions relating to safe and proper operation and maintenance of this equipment.
• Serious injury or death can result from failure to comply with all Occupational Safety and Health Administration
(OSHA)regulations and all manufacturer’s instructions.
• This equipment is not intended for use in any area considered hazardous per National Electric Code NFPA 70 2011,
Article 500.
• Read this document and follow all instructions before using this equipment.
OSHA regulations relating to abrasive blasting are contained in the Code of Federal Regulations, Title 29 (29 CFR 1910 General Industry;
1915 Maritime; 1926 Construction). The most pertinent include: 1910.94 Ventilation, 1910.95 Occupational Noise Exposure, 1910.132
Personal Protective Equipment, 1910.133 Eye and Face Protection, 1910.134 Respiratory Protection, 1910.135 Head Protection,
1910.244 (b) Remote Controls. Consult www.osha.gov for complete information.
NOTICE
TO
PURCHASERS
AND USERSOFOUR
PRODUCTS AND THISINFORMATIONALMATERIAL
Clemco proudly provides products for the abrasive blast
industry and is confident that industry professionals will use
their knowledge and expertise for the safe and efficient use of
these products.
The products described in this material, and the information
relating to these products, are intended for knowledgeable,
experienced users.
No representation is intended or made as to: the suitability of
the products described here for any purpose or application, or
to the efficiency, production rate, or useful life of these
products. All estimates regarding production rates or finishes
are the responsibility of the user and must be derived solely
from the user’s experience and expertise, not from information
contained in this material.
It is possible that the products described in this material may
be combined with other products by the user for purposes
determined solely by the user. No representations are
intended or made as to the suitability of or engineering
balance of or compliance with regulations or standard practice
of any such combination of products or components the user
may employ.
Abrasive blast equipment is only one component of an
abrasive blasting job. Other products, such as air
compressors, air filters and receivers, abrasives, scaffolding,
hydraulic work platforms or booms, equipment for lighting,
painting, ventilating, dehumidifying, parts handling, or
specialized respirators or other equipment, even if offered by
Clemco, may have been manufactured or supplied by others.
The information Clemco provides is intended to support the
products Clemco manufactures. Users must contact each
manufacturer and supplier of products used in the blast job for
warnings, information, training, and instruction relating to the
proper and safe use of their equipment.
GENERAL INSTRUCTIONS
This material describes some, but not all, of the major
requirements for safe and productive use of blast machines,
remote controls, respirator systems, and related accessories.
All equipment and accessories must be installed, tested,
operated and maintained only by trained, knowledgeable,
experienced users.
The blast operator and all workers in the vicinity must be
properly protected from all job site hazards including those
hazards generated by blasting.
Work environments involving abrasive blasting present
numerous hazards. Hazards relate to the blast process from
many sources that include, but are not limited to, dust
generated by blasting or from material present on the surface
being blasted. The hazards from toxic materials may include,
but are not limited to, silica, cyanide, arsenic, or other toxins
in the abrasives or in the coatings, such as lead or heavy
metals. Other hazards from toxins include, but are not limited
to, fumes from coating application, carbon monoxide from
engine exhaust, contaminated water, chemicals or asbestos.
In addition, physical hazards that may be present include, but
are not limited to, uneven work surfaces, poor visibility,
excessive noise, and electricity. Employers must identify all
job site hazards and protect workers in accordance with
OSHA regulations.
Never modify Clemco equipment or components or
substitute parts from other manufacturers for any Clemco
components or parts. Any unauthorized modification or
substitution of supplied-air respirator parts violates OSHA
regulations and voids the NIOSH approval.
IMPORTANT
Contact Clemco for free booklets:
Blast Off 2 – Guide to Safe, Productive, and Efficient Abrasive
Blasting, and Abrasive Blasting Safety Practices – Guide to Safe
Abrasive Blasting.
Clemco Industries Corp. One Cable Car Drive Washington MO 63090
Tel: 636 239-4300 — Fax: 800 726-7559
Email: info@clemcoindustries.com
Website: www.clemcoindustries.com
PREFACE
OPERATIONAL INSTRUCTIONS
OPERATOR SAFETY EQUIPMENT
OSHA regulation 1910.134 requires appropriate respiratory protection for blast operators and workers in
the vicinity of blasting. These workers must wear properly-fitted, properly-maintained, NIOSH-approved,
respiratory protection that is suitable for the job site hazards. Blast respirators are to be worn only in
atmospheres not immediately dangerous to life or health from which wearers can escape without use of
the respirator.
The employer must develop and implement a written respiratory protection program with required
worksite- specific procedures and elements for required respirator use. The employer must provide
effective training to employees who are required to use respirators. The training must be comprehensive,
understandable, and recur annually, and more often if necessary.
NEVER use abrasives containing more than one percent crystalline silica. Fatal diseases, such as silicosis,
asbestosis, lead or other poisoning, can result from inhalation of toxic dusts, which include, but are not
limited to, crystalline silica, asbestos, and lead paint. Refer to NIOSH Alert 92-102; and OSHA CPL 03-00-007:
“National Emphasis Program – Crystalline Silica”, in which OSHA describes policies and procedures for
implementing a national emphasis program to identify and reduce or eliminate health hazards from exposure
to crystalline silica. Numerous topics associated with the hazards of crystalline silica in silica blasting sand
can be found on http:// osha.gov/. Clemco urges users of silica blasting sand to visit this website, and read
and heed the information it contains.
Always make sure the breathing air supply (respirator hose) is not connected to plant lines that supply
gases that include, but are not limited to, oxygen, nitrogen, acetylene, or other non-breathable gas. Never
modify or change respirator air line connections without first testing the content of the line for safe
breathing air. Failure to test the line may result in death to the respirator user.
• Breathing air quality must be at least Grade D,
as defined by the Compressed Gas Association
specification G-7.1, per OSHA Regulation 29 CFR
1910.134. When compressed air is the breathing air
source, a Clemco CPF (suitable sorbent bed filter)
should be used. Respirator hose connecting the
respirator to the filter must be NIOSH approved.
Non- approved hose can cause illness from
chemicals employed to manufacture the hose.
• All workers must always wear NIOSH-approved
respirators when any dust is present. Exposure to
dust can occur when handling or loading abrasive,
blasting, cleaning up abrasive, or working in the
vicinity of blasting. Before removing the respirator,
test the air with a monitoring device to ensure it is
safe to breathe.
INDUSTRY ORGANIZATIONS
For additional information, consult:
Occupational Safety and Health Administration (OSHA) - www.osha.gov
Compressed Gas Association (CGA) - www.cganet.com
The Society for Protective Coatings (SSPC) - www.sspc.org
National Association of Corrosion Engineers (NACE) - www.nace.org
American Society for Testing and Materials (ASTM) - www.astm.org
National Institute of Occupational Safety and Health (NIOSH) - www.niosh.gov
American National Standards Institute (ANSI) - www.ansi.org
• Clemco respirators DO NOT remove or protect against
carbon monoxide or any other toxic gas. Monitoring
devices must be used in conjunction with the respirator to
ensure safe breathing air. Always locate compressors and
ambient air pumps where contaminated air will not enter
the air intake.
• Always use Clemco lenses with Clemco respirators;
installing non-approved lenses voids the NIOSH approval.
Respirator lenses are designed to protect the wearer from
rebounding abrasive; they do not protect against flying
objects, heavy high-speed materials, glare, liquids, or
radiation.
PREFACE
BLAST MACHINES AND REMOTE CONTROLS
OSHA regulation 1910.169 describes the necessity of pressure relief valves on compressed air equipment. Do not
operate blast machines with air compressors that are not equipped with properly functioning pressure relief valves.
OSHA regulation 1910.244(b) requires the use of remote controls on blast machines.
Serious injury or death can result from many sources, among them:
Involuntary activation of the remote controls. Never modify or substitute remote control parts; parts are not
compatible among different manufacturers. Welding hose is not suitable for remote control hose. Its ID and material
composition make it unsafe for remote control use.
Exceeding the maximum working pressure. Clemco blast machines are built to ASME-code and carry a ‘U’ or ‘UM’
stamp, and National Board/serial number. Every machine is marked with its maximum working pressure. Never
exceed the maximum working pressure limits of the blast machine.
Uncontrolled blast stream. High-velocity abrasive particles will inflict serious injury. Always point the blast nozzle
in the direction of the blast surface only. Keep unprotected workers out of the blast area.
Welding on the blast machine. Never weld on the blast machine; welding voids the National Board approval and
may affect the dimensional integrity of the vessel.
Moving the blast machine. Never manually move a blast machine containing abrasive, any machine containing
abrasive must be moved with appropriate mechanical lifting equipment.
HOSES, COUPLINGS, AND NOZZLE HOLDERS
The inside diameter (ID) of air hoses, fittings, and connections
should be at least four times larger than the nozzle orifice size. Blast
hose ID should be three to four times the size of the nozzle orifice.
Example: a #6 nozzle (3/8” diameter orifice) calls for 1-1/2” ID blast
hose and 1-1/2” ID or larger compressor hose. All hose runs should be
kept as short as possible and run in as straight a line as possible to
reduce pressure loss.
To install, squarely cut the end of the hose so that it fits snugly
against the coupling or hose end shoulder. Always use the screws
recommended by the manufacturer ensuring that they do not
penetrate the inner wall. Make sure the couplings tightly fit the hose.
Install cotter pins at every connection or use couplings with built-in
lock-springs to prevent disengagement. Install safety cables at all
connections to prevent whipping if hoses disengage or blow out.
MAINTENANCE AND REPAIR
Completely read and follow all service instructions and recommended maintenance intervals. Always shut off compressor and
depressurize blast machine before performing any maintenance. At every service interval, clean all filters, screens, and alarm systems. If
spring-loaded abrasive valves are used, always cage spring before disassembly.
WARRANTY
The following is in lieu of all warranties, express, implied or statutory, and in no event shall seller or its agents, successors, nominees or
assignees, or either, be liable for special or consequential damage arising out of a breach of warranty. This warranty does not apply to any damage or
defect resulting from negligent or improper assembly or use of any item by the buyer or its agent or from alteration or attempted repair by any person
other than an authorized agent of seller. All used, repaired, modified, or altered items are purchased “as is” and with all faults. In no event shall seller
be liable for consequential or incidental damages. The sole and exclusive remedy of buyer for breach of warranty by seller shall be repair or
replacement of defective parts or, at seller’s option, refund of purchase price, as set forth below
:
1. Seller makes no warranty with respect to products used other than
in accordance hereunder.
2. On products seller manufactures, seller warrants that all products are to
be free from defects in workmanship and materials for a
period of one year from date of shipment to buyer, but no warranty is
made that the products are fit for a particular purpose.
3. On products which seller buys and resells pursuant to this order, seller
warrants that the products shall carry the then standard warranties of the
manufacturers thereof, a copy of which shall be made available to the
customer upon request.
4. The use of any sample or model in connection with this order is for
illustrative purposes only and is not to be construed as a warranty that the
product will conform to the sample or model.
5. Seller makes no warranty that the products are delivered free of the
rightful claim of any third party by way of patent infringement or the like.
6. This warranty is conditioned upon seller’s receipt within ten (10)
days after buyer’s discovery of a defect, of a written notice stating in
what specific material respects the product failed to meet this
warranty. If such notice is timely given, seller will, at its option, either
modify the product or part to correct the defect, replace the product or
part with complying products or parts, or refund the amount paid for
the defective product, any one of which will constitute the sole liability
of the seller and a full settlement of all claims. No allowance will be
made for alterations or repairs made by other than those authorized
by seller without prior written consent of seller. Buyer shall afford
seller prompt and reasonable opportunity to inspect the products for
which any claim is made as above stated.
Except as expressly set forth above, all warranties, express, implied
or statutory, including implied warranty of merchantability, are
hereby disclaimed.
PREFACE
r
r
r
as
e
Co
s
2. Breathing Air Compressor for
High Pressure Respirators
or Ambient Air Pump for
Low Pressure Respirators
4. External or
Helmet Mounted
Carbon Monoxide
Monitor /Alarm
3. NIOSH Approved
Supplied-Air Respirator
5. CPF Air Filte
1. Air Compresso
12. Abrasive
11. Appropriately Sized Nozzle
9. Blast Hose
10. Hose Couplings and Safety Cables
6. ASME Code
t Machin
Bl
8. Remote
ntrol
7. Air Line
Moisture
Separato
Make sure all blast operators are properly trained and suitably attired with a blast suit, safety boots, leather gloves, respiratory and hearing protection. Every day before start up,
check all equipment components, including piping, fittings, and hoses, and valves, for leaks, tightness, and wear. Repair or replace as needed. Use the following checklist.
1. PROPERLY-MAINTAINED AIR COMPRESSOR
sufficient volume (cfm) at given pressure for nozzle and other tools. ADD 50%
volume (cfm) reserve to allow for nozzle wear. Use large compressor outlet and
sized to provide
7. AIR LINE FILTER
(moisture separator) installed as close as possible to the
blast machine inlet and sized to match the size of the inlet piping or larger air supply
line. Clean filter and drain often. Damp abrasive causes operational problems.
air hose (at least 4 times the nozzle orifice diameter). For oil-lubricated
compressors, the employer shall use a high- temperature or carbon monoxide
alarm, or both, to monitor carbon monoxide levels. If only high-temperature
alarms are used, the air supply shall be monitored at intervals sufficient to
prevent carbon monoxide in the breathing air from exceeding 10 ppm. Follow
the manufacturer’s checklist and maintenance instructions.
2. BREATHING-AIR COMPRESSOR
(or oil-less ambient air pump)
capable of providing Grade D quality air, located in a dust free area. Read # 1
above.
worn by blast operators, and other workers exposed to blast dust.
Make sure all respirator components are in place — all lenses, inner collar, and
cape. Thoroughly inspect all components for wear. The NIOSH approval (approval
number is listed in the owner’s manual) is for a complete assembly from point of
attachment on the CPF (sorbent bed) filter to the complete respirator. Substitution of
any part voids the NIOSH approval.
4. CARBON MONOXIDE MONITOR/ALARM
installed at the CPF filter or
inside the supplied-air respirator for monitoring for the presence of deadly CO gas
and warning the operator(s) when the CO level reaches an unacceptable level.
When an ambient air pump is used for breathing air, a CO monitor provides a
measure of safety. Read # 1 above.
5. BREATHING-AIR FILTER (OSHA-REQUIRED
sorbent bed filter) for
removal of moisture and particulate matter in the compressed air breathing-air
supply. Monitor the condition of the cartridge and replace when odor is detected or
at 3 month intervals, whichever comes sooner. The breathing air filter does NOT
detect or remove carbon monoxide (CO). Always install a CO monitor/alarm.
6. BLAST MACHINE
(bearing U or UM stamp, National Board Number, and
Maximum Working Pressure) sized to hold a 30-minute abrasive supply. Examine
pop-up valve for alignment. Check piping, fittings, screens, valves for tightness,
leaks, and wear. Always ground the machine to eliminate hazard of static shock.
Install a blast machine screen to keep out foreign objects. Use a blast machine
cover if left outdoors overnight. Never exceed the maximum working pressure of the
8. REMOTE CONTROLS
operating condition. Test and check all components to ensure all parts are present
and fully functional. Use genuine replacement parts. NEVER mix parts from different
manufacturers. Never use welding hose for remote control hose.
9. BLAST HOSE
The ID should be three to four times the size of the nozzle orifice diameter. Blast hose
should be arranged in as straight a line as possible from the blast machine to the
work area, avoiding sharp bends.
10. COUPLINGS AND NOZZLE HOLDERS
be installed with manufacturer recommended screws. Coupling lugs must snap firmly
into locking position. Gasket must always be used to form a positive seal, and cotter
pins must be installed. Replace gasket when wear, softness or distortion is detected.
Check nozzle holder for thread wear; replace at any sign of wear. Install safety cables
at all connections.
11. NOZZLE
orifice size should be checked and nozzle replaced when worn
1/16” from original size. (No. 5 nozzle has 5/16” orifice diameter; replace when it
measures 3/8”). Threads should be inspected daily for wear and nozzle should be
replaced when wear is detected. Always use a nozzle washer.
12. ABRASIVE
must be a material specifically manufactured for blasting. It
should be properly sized for the job. Check material safety data sheet for freesilica, cyanide, arsenic, lead and other toxins and avoid use when these toxic,
harmful substances are present.
SURFACE TO BE BLASTED
Take appropriate protective measures as required by OSHA to ensure the blast
operator, other workers in the vicinity, and any bystanders are properly protected.
should have an inside diameter sized to suit the blast nozzle.
should fit snugly on the hose and
should be examined for hazardous substances.
vessel.
1.0 INTRODUCTION
CMS-1 CARBON MONOXIDE MONITORPage 1
1.1Scope of manual
1.1.1These instructions cover the setup, operation,
maintenance, troubleshooting, and replacement parts for
the CMS-1 carbon monoxide (CO) monitor.
1.1.2The CMS-1 monitor is manufactured by
Invertech Inc., for Clemco Industries Corp. The monitor
is a potential life saving instrument. To assure its
performance, the monitor must be properly installed,
tested, calibrated, and maintained. Before installing and
using the monitor, all personnel involved with the
operation and maintenance of the instrument must read
this entire manual, including the orange cover.
1.2 Safety Alerts
1.2.1Clemco uses safety alert signal words, based on
ANSI Z535.4-2011, to alert the user of a potentially
hazardous situation that may be encountered while
operating this equipment. ANSI's definitions of the signal
words are as follows:
This is the safety alert symbol. It is
used to alert you to potential physical
injury hazards. Obey all safety
messages that follow this symbol to
avoid possible injury or death.
Notice indicates information that is considered
important, but not hazard-related, if not
avoided, could result in property damage.
Caution indicates a hazardous situation that, if
not avoided, could result in minor or moderate
injury.
Danger indicates a hazardous situation that, if
not avoided, will result in death or serious
injury.
No annunciator lights or alarms .................................. 8.3
Either alarm-lights or alarm horn fails ......................... 8.4
ACCESSORIES AND REPLACEMENT PARTS ....... 9.0
1.4 Components
1.4.1The primary components are shown in Figure 1.
The instrumentation is enclosed in a non-metallic case.
The monitor system includes the instrument case,
filter/regulator, ten feet of 1/8" ID x 1/4" OD tubing with
adaptor fittings, a cylinder of 10 PPM (parts per million)
test gas, and calibration connector which includes the
connector valve and tubing.
Pressure Regulator/ Filter
Mounting Flange (2)
Fuse
Case Cover
Reducer Fitting
1.5 Operating Principles
1.5.1 The CMS monitor samples respiratory air from a
breathing-air source. The air to the respirator does not
actually pass through the monitor. This enables one
monitor to detect CO (carbon monoxide) in the entire
breathing-air system, not just in the air to a single
respirator.
When the CMS is correctly installed and
maintained, it monitors the level of carbon
monoxide in the air supply. The instrument is a
monitoring device; it DOES NOT remove or
convert carbon monoxide.
1/8 ID Tube Adaptor Fitting
This end connects to air source
1.5.2 During operation, a small sample of air, referred
to as sample-air, passes through the instrument, and is
continuously monitored by the chemical cell sensor. If
the sample-air supply fails for any reason, such as air
blockage, loss of pressure, or excessively high pressure,
an intermittent alarm horn will sound and the alarm-light
will illuminate yellow to alert the user.
1.5.3The alarm trip-point is preset at 10 parts per
million (PPM), which is the maximum permissible
exposure level (PEL) of CO for Grade-D breathing air in
the USA. Refer to Section 2.5 to change the alarm trip
point to 5 PPM when required by local regulations.
1.5.4In the event CO is detected in concentrations of
10 PPM, a continuous alarm horn sounds, and the "AIR
QUALITY" alarm light illuminates red. The alarms alert
the user to immediately stop blasting, and remove the
respirator as soon as it is safe to do so.
1.6 Annunciator Lights and Audible Alarm
Use and Functions
1.6.1 A horn (audible alarm) is provided on the
exterior of the instrument case, and annunciator lights
(visual alarms) are mounted on the faceplate. The lights
illuminate green when conditions are safe, or yellow or
red when an alarm occurs. The alarm horn sounds in
conjunction with the visual alarm.
Alarm Light:
Green Light - No Audible Horn
Indicates there is correct air flow and pressure to the
sensor. Correct flow is observed on the flow meter;
the flow ball is usually between .5 and .8 SCFH
(standard cubic feet per hour).
Yellow Light - Intermittent Alarm Horn
(Approximately every ten seconds)
Sample-Air Failure: This alarm is due to either low
flow, or high flow in the test chamber. Low flow (flow
ball is low in the flow meter) is usually due to loss of
pressure or plug in the sample-air line, or the
external pressure regulator is set too low. High flow
(flow ball is high in the flow meter) is an indication
that the external pressure regulator is set too high.
Yellow Light - No Audible Horn
Warm-Up Period: During the first minute or two of
operation, the monitor goes through a warm-up
period while the sensor stabilizes. During the warmup period, this alarm occurs, and all other alarms
are non-operational. After the warm-up is complete,
the alarm-light turns green if the air is free of CO, or
turns red if the air is contaminated with CO.
Air Quality light:
All alarm conditions require the immediate
attention of the user. Stop blasting immediately,
and remove the respirator as soon as it is safe
to do. Check the monitor to determine the
cause for alarm.
1.6.2 The alarm lights and the nature of the alarm
horn indicate the condition causing an alarm. Functions
of the alarm annunciators are as follows:
Armed Light: This indicates whether the alarm horn is
armed. Green means the alarm horn is armed; no
light means it is disarmed. NOTE: The horn should
be disarmed only to temporarily silence it during
calibration by the technician.
Do not use the compressed air monitored by
this instrument for breathing without checking
to make sure all three alarm lights are
illuminated green. Failure to heed this warning
can cause death from the inhalation of carbon
monoxide.
Green Light - No Audible Horn
Safe Condition: This means the sample-air gas is
below the permissible exposure level. Do not use
the compressed air for breathing unless the AIR
QUALITY light is illuminated green.
Red Light - Continuous Audible Horn Alarm
Toxic Gas Detected: This alarm occurs when the
sensor detects 10 PPM of CO. NOTE: The alarm
light may be red for a brief time as the sensor
stabilizes after the warm-up.
1.7 Digital Display Window
1.7.1 The digital display shows the level of carbon
monoxide (CO) in parts per million (PPM). The sensor
also responds to hydrogen sulfide and a limited number
of other toxic gases, and may display inflated CO
readings when other toxic gases are detected. NOTE:
The display may be slightly high for a brief period while
the sensor stabilizes after the warm-up.
1.8 Switches, Use and Functions
1.8.1 Run/Calibration Switch: Position the toggle in
the RUN position during the operating mode (all
operations other than the actual calibration test and
calibration process). If the toggle is not in the RUN
position, the instrument will initiate a yellow light and
intermittent alarm, indicating that the sample-air is not
reaching the sensor. Correctly positioning the toggle will
terminate the intermittent alarm.
1.8.2 Arm/Disarm Switch: This switch arms and
disarms the alarm horn. The switch is tripped ON and
OFF by inserting a small probe (a straightened paper
clip works well) through the small access opening in the
faceplate, and pushing gently until the detent is felt. It
may take a couple pushes to trip the switch. The horn is
armed when the ARM light is green. The horn is
disarmed when the light is unlit. NOTE The horn should
be disarmed only to temporarily silence it during
calibration by the technician.
1.9 Optional Accessories
1.9.1 Remote alarm: The 12-volt DC remote alarm is a
high-intensity, red lens strobe light and high-decibel horn.
Unlike external alarms, this alarm does not require an
external power source. Installation instructions are
provided with the alarm.
Its use is recommended in a high-noise area or where the
monitor cannot be placed in a conspicuous location. The
alarm kit includes the alarm, stand, wiring connector, and
50 ft. cable. See Accessories in Section 9.1. With
additional cable, the remote alarm may be placed up to
150 ft. from monitor.
a wall or panel, and mark the mounting hole locations.
Drill appropriately-sized holes, and secure with
fasteners.
DO NOT mount this instrument inside a blast
room. An operator wearing protective clothing
inside a blast room will not hear the alarm horn
when the instrument is in an alarm condition.
The alarm case is not designed to withstand the
continual impact of abrasive that takes place
inside a blast room. DO NOT mount this
instrument inside a paint room or in an
explosive or flammable environment. This
instrument is not rated for explosive
atmospheres. Refer to instructions within the
manual for installing a remote alarm, auxiliary
horns, lights, or safety shutdown devices.
Mounting Flange (2)
Mounting Screws (4)
Reducer Fitting
2.0 SETUP
2.1 Assembling the Monitor, Ref. Figure 2
2.1.1 Using the screws and instructions provided in
the mounting flange packet, attach the mounting flanges
to the instrument case.
2.1.2Apply Teflon thread seal tape to the male
threads of the 1/4" x 1/8" reducer fitting and tubing
elbow, and assemble as shown in Figure 2. Make sure
the regulator is correctly installed so the air flows toward
the instrument. The tubing elbow may be rotated after
the instrument is mounted to enable the tubing to be
connected with minimum bends.
2.2 Mounting the Monitor
2.2.1 The simplest way to mount the CMS-1 is with
the mounting flanges provided, ref. Figure 2. Mount the
monitor in convenient location that is close to the
compressed air source. Refer to Section 2.3.2. The
monitor must be positioned upright and level to ensure
the flow meter responds correctly. Place the monitor on
IN Port
Pressure Regulator/Filter
Tubing Elbow
Figure 2
2.3 Compressed-Air Connections
2.3.1 Air pressure requirements
2.3.2.1 If the compressed-air source pressure is
between 55 psi and 145 psi, no line pressure adjustment
is required. If line pressure is above 145 psi, a regulator
is required in the air line (before the monitor’s regulator)
to reduce pressure to be within the 55 to 145 psi
operating range. Pressure above 145 psi could damage
the monitor.
Do not use pipe joint compound to seal pipe
fittings. Use Teflon tape for pipe-thread sealer.
Although unlikely, there is a possibility the
monitor will respond to gases given off by
some compounds.
2.3.2.1 To ensure accurate and rapid air monitoring, the
instrument must be placed as close to the compressedair source as practical. Ten feet of 1/8" ID tubing with
one 1/8 NPT elbow/tubing adaptor and one 1/8" NPT
straight adaptor are furnished to be used between the
compressed air source and monitor inlet. Do not use
larger diameter tubing or pipe, or longer lengths of
tubing, as it increases the time it takes for the sample air
to reach the monitor.
2.3.2.2 Locate a suitable place to tap into the respirator
compressed air supply. Do not tap into a dead-end air
line. The tap location must be where moving air is
assured. The tap location should be within 10 feet of the
monitor, and the monitor must be placed in a location
that permits continual observance of the visual and
audible alarms.
CMS-1
Power Cord
Wall Mounted
Power Cord
1/8 ID Tubing
1/8 NPT Adaptor
To Respirator
CPF Filter
Compressed-Air Line
CMS-1
1/8 ID Tubing
1/8 NPT Adaptor
Isolation Valve
Compressed-Air Line
The monitor, remote alarm, or an auxiliary
external alarm, must be in a conspicuous place
to ensure that any alarm condition is observed.
Using the dry-contact terminals and a relay, the
monitor may be interlocked with other devices
such as lights horns, or compressor shutdown,
to provide additional safeguard against an
unobserved alarm.
2.3.2.3 Place a tee in the air-supply line, and use a
bushing or other adaptor, to connect the 1/8 NPT tube
adaptor supplied. Install an isolation valve at the source
to enable depressurization for service and to remove the
monitor. Typical setups are shown in Figure 3.
2.3.2.4 The air-supply-line tee should face upward to
prevent water from accumulating in the sample-air line. If
the tap cannot face upward, install a short 1/4 NPT
drop pipe and drain to prevent water from entering the
monitor.
2.3.2.5 Attach the urethane tubing between the air
supply tube-adaptor fitting and monitor inlet-tube adaptor
fitting. Cut-off excess tubing.
Compressed-Air Line
1/8 NPT Adaptor
1/8 ID Tubing
Isolation Valve
Keep this line as
short as possible.
Drop Line
Power Cord
Drain
Figure 3
2.4 Electrical Connections
2.4.1 120-Volt AC supply: CMS-1 monitors are
generally installed indoors and operate on AC power. A
120-volt AC power cord is supplied and can be plugged
into a compatible AC grounded electrical service.
Electrical power may be rewired by a qualified electrician
with conduit that conforms to applicable codes.
220-Volt AC: These monitors are available from the
factory on special order. The only difference between
the 120-Volt and 220-Volt unit is the power cord.
2.4.2 AC to 12-Volt DC field conversion
Note: The following field conversion instructions to 12volt DC is for monitors with Serial Nos. 4446 and higher.
To convert monitors with serial numbers lower than 4446,
request manual No. 23301, Revision D or earlier.
The AC power cord provided with the monitor may be
used, using black as the positive. A user supplied DC, 2wire cord may be used, using red as positive. If a new
12-volt cord is used, splice the positive through the fuse,
and use the spade on the outlet of the fuse to connect to
J5 DC positive post, as shown Figure 4.
2.4.2.1 Make sure that all electrical power to the monitor
is locked-out and tagged-out.
2.4.2.2 Loosen the two case-cover screws and open the
case cover.
Fuse Assembly
Circuit Board
Disconnect white
and black wire from
power supply leads
Lead Wires
Connector Post
Remove Connector
Power Supply
any wires or tubing. NOTE: The screws on one side are
shorter than those on the other side, replace accordingly.
2.4.2.4 Remove the connector with red and black wires
from the plug on the circuit board, as shown in Figure 4.
If there is a chance the monitor will be converted back to
AC power, tie the wires out of the way. If the monitor is to
be permanently converted to 12-volts, pop off the clear
terminal cover and remove the wires from the terminal.
2.4.2.5 Disconnect the white and black wires from the
power supply lead-wires at the connectors, as shown in
Figure 4.
2.4.2.6 Plug the black wire onto the J5 (DC +) post at the
upper edge of the circuit board, as shown in Figure 4.
Plug the white wire onto the J6 (DC-) post.
2.4.2.7 Reposition the faceplate, making sure that all
internal air lines are free of interference, binding, or
kinks, and that all tube connections are secure. Tighten
the faceplate screws finger-tight to ensure they are
threading correctly, and then tighten to barely snug with
a screwdriver.
2.4.2.8 Close and secure the instrument case cover,
tightening the screws to a maximum of 20-inch lbs.
2.4.2.9 Remove the plug from the other end of the cord.
2.4.2.10 Attach a positive (+) terminal connector to the
end of the black wire and a negative (-) connector to the
end of the white wire. Clip the green wire.
Green Ground Wire
J6 (DC )
J5 (DC+)
White wire to DC (J6)
Tie-off or
remove wires
The green ground wire may remain connected or
clipped when using the AC cord for DC
Some items removed or rotated for clarity
Black wire to DC+ (J5)
Figure 4
2.4.2.3 Remove the four faceplate screws, and carefully
remove the faceplate, being careful not to disconnect
Use any connectors that are compatible with a 12volt DC system. Make sure the connectors are
clearly marked positive (+) and negative (-). Attach
the positive connector to the wire leading to the DC+
post on the circuit board, and the negative
connector to the DC- post on the circuit board. The
monitor will not operate if wires are reversed.
2.5 Changing Alarm Trip-Point
2.5.1 The alarm trip-point is preset at 10 parts per
million (PPM), which is the maximum permissible
exposure limit (PEL) of CO for Grade-D breathing air in
the USA. Use the following instructions to change the
trip point to 5 PPM when required by local regulations.
2.5.2 Unplug the monitor from the power supply.
2.5.3 Open the case cover and insert a straightened
paper clip into the "ARM" port in the faceplate. While
CMS-1 CARBON MONOXIDE MONITORPage 7
maintaining pressure on the paper clip, plug the monitor
into the power supply; do not remove the paper clip.
2.5.4Observe the display window, it will show four
lines of information; the bottom line is the trip-point (PPM
10). Within a few seconds the trip-point will change to
PPM 5; when it changes, remove the paper clip. The
trip-point is now set to alarm at 5 PPM CO. Close cover.
2.5.5To change the trip point back to 10 PPM, unplug
the monitor and repeat the process. There are only two
settings, 10 PPM and 5 PPM.
2.6 Connecting External Alarm and Shutdown
Device, Ref. Figure 5.
Note: External alarms require external power. Do not
confuse an external alarm with the optional remote
alarm, which does not require external power.
Instructions for the remote alarm are provided with the
alarm.
2.6.1A dry relay contact is mounted on the circuit board
(Ref. Figure 5) for use with external alarms and shutdown
devices. These contacts are rated at 5 amperes. Use the
contacts to operate relays for external devices.
Dry Relay Contact
3.0 OPERATION
Do not breathe compressed air that this
instrument monitors unless the instrument has
been calibration tested, and if needed,
calibrated per Section 4.0. Using a monitor that
has not been calibration tested can permit
undetected CO to enter the respirator air lines.
Breathing toxic gases can cause death.
3.1 Setup for Operation
3.1.1 Position the Run/Calibration toggle switch to the
RUN position.
3.1.2Open the sample-air isolation valve to supply air
to the monitor.
3.1.3Observe the flow meter ball, and adjust the
pressure regulator until the ball remains between .5 and .8
SCFH.
3.1.4Plug the instrument into a grounded 120-volt AC
power supply. If the monitor was converted to operate
on 12-volt, connect to a 12-volt DC power source. Make
sure the positive and negative leads go to the correct
terminal as noted in Paragraph 2.4.2.10 and the notice
following it. There is no ON/OFF switch on the monitor; it
will begin operation as soon as power is applied.
This contact is for external
alarms only.
NC
COM
External
Power
NO
L2
+
L1
Typical wiring for normally open alarm
Do not connect the optional
remote alarm to this contact.
Auxiliary Alarm
NO
COM
NC
Figure 5
3.2 Warm-Up Period
3.2.1 When electrical power is applied and sample-air
flow set between .5 and .8 SCFH, the monitor goes
through a warm-up period of approximately one minute.
3.2.2During the warm-up period, the alarm horn is
disabled and the ALARM light is yellow.
3.2.3At the end of the warm-up period, the alarm
horn is reactivated, and all the annunciator lights will be
green indicating the following:
ARMED: The alarm horn is armed, to provide an
audible alarm if CO reaches the permissible limit of
10 PPM.
ALARM: There is correct sample-air flow to the
sensor.
AIR QUALITY: The sample air at the sensor does
not exceed the permissible contamination limit.
3.2.4Make sure alarms function per Section 6.4. If the
alarms function as described, proceed to place the
instrument in operating mode, per Section 3.3.
3.2.5 If the instrument goes into any alarm mode (any
annunciator light is any color other than green or any
audible alarm) after the warm-up, re-initiate the warm-up
by terminating and restarting power. If the ARMED light
is not lit, or the ALARM light is yellow with an intermittent
horn, proceed as follows:
ARMED light is not lit: The alarm may be disarmed. Arm
the alarm by inserting a probe (straightened paper clip)
through the small ARM access opening in the faceplate,
pushing gently until the detent is felt. It may take several
pushes to trip the switch.
ALARM light is yellow with an intermittent horn: This
usually means low or high flow through the sensor.
Generally, with the correct flow, the ball will be between
.4 and .9, in some cases slightly higher or lower flow is
required. Increase or decrease pressure at the pressure
regulator in half turn increments until the horn silences.
Approximately one minute later, the ALARM light should
change from yellow to green.
3.2.6A continuous alarm with red AIR QUALITY light
indicates that the sample air exceeds the permissible
contamination level (the digital display confirms the
contamination level), or the instrument requires a
calibration test. Refer to Section 4.0.
3.4 Shutdown
3.4.1 If the instrument is in a facility that has 24-hour
compressed air and electrical power, there is no need to
shut off the monitor. If the compressed air system is
shutdown, the electrical power must also be turned off,
by unplugging the power cord, or disengaging power if
the monitor is permanently wired.
If the compressed-air source is shut down
without shutting off electrical power, the
monitor will go into low-flow alarm, sounding
an intermittent alarm and displaying a yellow
ALARM light.
4.0 CALIBRATION TEST and CALIBRATION
Note: Thoroughly review and understand the calibration
test and calibration process before testing or calibrating
the instrument..
DO NOT use the respirator during the warm-up
period. Alarms are disabled during the warm-up
and do not warn against toxic gases.
3.3 Operating Mode
3.3.1 Following the warm-up period, with the toggle
positioned to RUN, and the sample air flowing through
the unit, the monitor is in full operating mode.
3.3.2Before donning the respirator, make sure the
monitor is in the operating mode; all three alarm-lights
must be lit green, with no audible alarm.
Do not use compressed air monitored by this
instrument for breathing without checking to
make sure the instrument is in full operating
mode, and all the annunciator lights are
illuminated green. Failure to heed this warning
can cause death from the inhalation of carbon
monoxide.
The following calibration instructions are
written for 10 PPM of CO Test Gas. To avoid
confusion, do not test or calibrate the
instrument with any concentration of CO test
gas. Using other concentrations, while
following the calibrating instructions, will place
the monitor out of calibration, and may fail to
alert the user of toxic gases. Breathing toxic
gases can cause death.
The following materials are required for tests and
calibration:
10 PPM test gas ............................. Stock No. 22865
Impurity-free test gas .................... Stock No. 11132
applying test gas of a known concentration to ensure the
instrument responds with an alarm when CO
concentrations exceed the permissible level, and that it
responds with a safe signal when impurity-free air is
applied. Calibration test does not include any adjustment
of the instrument. Calibration (adjustments) should be
done only when calibration testing shows it is necessary.
4.1.3 Calibration process:Calibration means
adjusting the instrument. Calibration should be done
only when the instrument does not respond during
calibration testing, or remains in alarm at startup.
4.2 Stabilization Period
4.2.1 The monitor must be in operation for at least 30
minutes before calibration tests. DO NOT test the
instrument until it has gone through the warm-up period
and has been in operation with sample air flowing
through it (See Section 3.0) for at least 30 minutes.
4.3 Prepare Calibration Connector, Figure 6
4.4 Prepare Monitor for Calibration
4.4.1 Open the instrument case cover to access the
faceplate. Refer to Figure 7 for faceplate-calibration and
testing callouts.
4.4.2Connect the tube end connector to the "Test
Gas" port by aligning the tabs and inserting the tube end
connector into the port, and turn it clockwise to lock.
4.4.3Place the Run/Calibration toggle toward "CAL"
(calibration). The "ALARM" light will immediately turn
yellow, and within a few seconds, the intermittent alarm
horn will sound. Disarming the alarm per Section 1.8.2
will silence the horn during calibration tests.
4.5 Calibration Test
4.5.1 Gently pull the tubing side of the release
connector (refer to Figure 6) to make sure it is correctly
seated in the body. Slowly open the calibration connector
valve to introduce test gas. Test gas is entering the unit
when the flow meter ball rises. If the ball does not rise
when the calibration valve is opened, the test gas cylinder
is probably empty.
4.3.1Make sure the flow-control knob is turned fully
clockwise to the CLOSED position.
4.3.2Thread the calibration connector onto the
cylinder of calibration gas.
Tube End Connector
CLOSE
OPEN
Slide Release
Push in to release
the tubing connector
Flow-Control Knob
Calibration Connector
Test Gas
4.5.2Adjust the flow-control knob until the flow meter
ball remains between .5 and .8 SCFH. The valve is
extremely sensitive. Minor adjustments may be required
to correctly position the ball.
4.5.3Allow test gas to flow through the instrument
until the digital readout stabilizes (about one minute).
4.5.4If the display stabilizes between 9 PPM and 11
PPM (10 PPM plus or minus one), the calibration test is
complete. Return to operating mode per Section 4.7.
4.5.5If the display stabilizes at any value other than
between 9 PPM to 11 PPM, and if the instrument has
been in operation for at least 30 minutes, calibrate the
instrument per Section 4.6. If the instrument has not
been in operation for 30 minutes, return it to operating
mode per Section 4.7, and operate with sample air for at
least 30 additional minutes. Re-test the calibration. If the
display stabilizes at a value other than 10 PPM plus or
minus one, calibrate the instrument per Section 4.6
4.5.6When test/calibration is completed, return the
instrument to operating mode per Section 4.7.
4.6.1Do not calibrate the instrument unless it has gone
through two stabilizing periods and calibration tests, per
Sections 4.2 and 4.5, and only if the digital display is
other than 09, 10, or 11 (10 PPM plus or minus one).
CMS-1 CARBON MONOXIDE MONITORPage 10
4.6.2 If the number in the display is lower than 10
PPM, calibrate through the upright triangle. If the
number is higher than 10 PPM, calibrate through the
inverted triangle. Calibrate by inserting a straightened
paper clip through the small access opening in the
appropriate triangle, pushing gently until the detent is
felt. It takes several gentle pushes to change the display.
Repeatedly and slowly, press the switch until the display
registers 10 PPM.
4.6.3When the display shows 10 PPM, the monitor is
calibrated. Return the instrument to operating mode per
Section 4.7.
4.7 Return Monitor to Operating Mode
4.7.1 Close the connector’s flow-control valve.
4.7.2 Remove the calibration connector from the
monitor by gripping the tube-end connector, push in
lightly and turn it counterclockwise to unlock, and pull
straight out.
4.7.3Place the Run/Calibration toggle switch toward
"RUN". The flow meter ball should rise to .5 to .8 SCFH.
After the warm-up, the "ALARM" light should change
from yellow to green. If the light does not change to
green, increase or decrease flow by adjusting the
pressure regulator as required.
4.7.4 Make sure the "ARM" light is lit. See Sec. 1.7.2.
4.7.5 Close the instrument case cover and tighten the
screws to a maximum of 20-inch-lbs.
4.8.2Follow the steps in Section 4.5 Calibration
Testing, but use impurity-free test gas in place of the 10
PPM gas.
4.8.3. If the monitor is correctly calibrated and in
working order, the display will show 00 PPM. This means
the monitor was accurately reading contaminated air. DO
NOT USE THE COMPRESSED AIR FOR BREATHING.
Do not breathe air that this instrument identifies
as toxic, until the source of contamination is
identified and corrected. Breathing toxic gases
can cause death.
4.8.4 Identify the source of contamination. Do not
overlook the possibility that contaminated air entered the
compressor intake. The contamination could be from
engine or other exhaust entering the intake. If the source
of contamination is temporary, the monitor will return to
a non-alarm (safe) condition after the contamination is
cleared from the compressed air system.
4.8.5If the monitor is out of adjustment or not in good
working order, the display will show other than 00 PPM.
Zero the instrument per Section 5.0.
4.8.5.1 Calibrate the monitor per Section 4.6 and
reapply impurity-free gas. If the display shows other than
00 PPM, the monitor requires service.
4.7.6Remove the calibration connector from the
cylinder of test gas. The test gas cylinder has a positive
seal, whereas the calibration connector valve does not.
If the connector is not removed from the test gas
cylinder, over time, the cylinder will empty.
4.7.7If the tubing needs to be removed from the
calibration connector, press and hold the slide release,
and gently pull the tubing connector out from the release
fitting.
4.7.8 Store all material in a clean, dry area.
4.8 Impurity-Free Air (zero contamination) Test
4.8.1This test should be done whenever the
instrument stays in an alarm condition after it is returned
to the operating mode. This test shows whether the
alarm condition is due to contaminated air, or a
malfunctioning monitor.
Do not use compressed air monitored by this
instrument for breathing unless the instrument
is in good working condition. Using a monitor
that is not in calibration or not working
correctly can permit undetected CO to enter the
breathing-air lines. Breathing toxic gases can
death.
5.0 ZEROING the MONITOR
The following materials are required before zeroing the
monitor:
Impurity-Free Gas ........................... Stock No. 11132
Use the triangular adjustment
ports to calibrate the monitor
using only 10 PPM test gas.
"ARM" Switch
(Arms and disarms alarm horn)
5.1The monitor must be in operation for at least 30
minutes before zeroing. NOTE: If zeroing follows sensor
replacement, repeat the process after two days of
operation.
Flow-ball
RUN
Position toggle toward
CAL (calibration)
CAL
Test Gas Port
Connects from test gas
Zeroing Potentiometer
Use this pot only to zero
the monitor, and only use
impurity-free (zero) gas.
Do not use to calibrate.
Figure 7
5.7Adjust the flow-control knob until the flow meter
ball remains between .5 and .8 SCFH. The valve is
extremely sensitive. Several minor adjustments may be
required to position the ball correctly.
5.2Open the instrument case cover to access the
faceplate. Refer to Figure 7 for callouts.
5.3Attach the calibration connector to the impurity-
free gas, per Section 4.3.
5.4Connect the calibration connector to the "Test
Gas" port on the faceplate, by aligning the tabs and
inserting the tube end connector into the port and turning
it slightly clockwise to lock.
5.5Place the Run/Calibration toggle toward "CAL"
(calibration). The "ALARM" light will immediately turn
yellow, and within a few seconds the intermittent alarm
horn will sound. Disarming the alarm per Section 1.8.2
will silence the horn during the adjustment.
5.6Gently pull the tubing side of the release
connector, to make sure it is correctly seated in the
body. Slowly open the calibration connector valve to
introduce test gas. Test gas is entering the unit when the
flow meter ball rises. If the ball does not rise when the
calibration valve is opened, the test gas cylinder is
probably empty.
5.8Allow test gas to flow through the instrument
until the digital readout stabilizes (about one minute).
5.9Using a miniature screwdriver inserted through
the zeroing potentiometer opening, adjust the
potentiometer until the display reads ".00" (decimal,
zero, zero) NOTE: The readout responds slowly to the
potentiometer adjustment. Turn it slowly when making
adjustments.
5.10After the display shows ".00", slowly turn the
potentiometer clockwise until the decimal disappears.
5.11Return the instrument to operating mode per
Section 4.7.
calibrate the instrument subjects it to external
contamination. Take care not to introduce contaminants
into the instrument when the case is open.
6.1.2Clean the exterior of the case with a cloth
soaked in a solution of water and mild detergent. Do not
clean with solvent cleaners.
6.2 Calibration Test Schedule
6.2.1 Test the calibration when it is initially setup and
daily for the first week. Check it once a week for the first
month. Check it at least once a month thereafter. Refer
to Section 4.0 for test procedure.
6.3 Calibration Schedule
6.3.1Avoid the urge to calibrate the instrument.
Calibrate only when the calibration tests show it is
required. Refer to Section 4.0.
calibrate the monitor and large spans (drifting) in the
display from one calibration to another.
7.1.2The following materials are required to replace
the sensor:
10 PPM test gas .......................... Stock No. 22865
Impurity-free gas .......................... Stock No. 11132
7.1.3Open the instrument case and remove the four
faceplate-screws. NOTE: The screws on the left are
shorter than those on the right, replace accordingly.
7.1.4Carefully lift the faceplate to expose the
instrumentation, using care not to disconnect any wires
or tubing.
7.1.5Locate the clear-plastic sensor housing, shown
in Figure 8, and remove the mounting screws. The
screws are easily removed Do not press hard against
the screw-heads; excessive pressure could damage the
circuit board.
7.1.6 Carefully lift the sensor housing out of the way.
6.4 Alarm Test
6.4.1 Although uncommon, alarm-lights and horns do
fail. Check their function before each use by placing the
Run/Calibration toggle toward "CAL". The "ALARM" light
will immediately turn yellow, and within a few seconds,
the intermittent alarm horn will sound. Placing the toggle
in "RUN" position returns the alarms to their normal
operation. Never use the respirator without first
verifying that the monitor is in the operating mode;
all three annunciator lights must be green, with no
audible alarm.
7.0 SERVICE MAINTENANCE
NOTE: Do not attempt to repair the instrument or
replace any item that is not listed in this section, or
that requires parts not shown in Section 9.0. Contact
a Clemco distributor for authorization to return the
instrument for evaluation or service.
7.1 Sensor Replacement
7.1.1 Sensor life depends on several factors, but in
most cases, the sensor should last two to three years.
Signs that the sensor is failing is the frequent need to
7.1.7 Pull straight up to remove the sensor.
7.1.8 Discard the old sensor to prevent reuse.
7.1.9 Handle the new sensor with care. Remove the
shorting spring from the pins, as shown in Figure 8, and
then plug the sensor into the sockets.
7.1.10 Place the sensor housing over the sensor. Align
the mounting screw holes, and alternately tighten the
screws until snug. Do not push hard on the screw
heads when tightening. Do not over-tighten.
7.1.11 Replace the faceplate, making sure that all
internal air lines are free of interference, binding or
kinks, and that all tube connections and wire
connections are secure. Tighten the faceplate screws
finger-tight to ensure they are threading correctly, and
then tighten them barely snug with a screwdriver.
7.1.12 Apply sample air and power per Section 3.0.
Run the monitor in operating mode for at least four hours
before calibrating.
7.1.13 Check the display to verify it reads zero (00). If
not, adjust per Section 5.0.
7.1.15 Close and secure the instrument case cover,
tightening the screws to a maximum of 20-inch-lbs.
Pin Sockets
Circuit Board
Sensor
Mounting Screws
Remove shorting spring
from pins before installing
new sensor.
Sensor Housing
Figure 8
8.0 TROUBLESHOOTING
Shorting electrical components can result in
serious electrical shocks, or can damage
equipment. All electrical troubleshooting must
be performed by a qualified electrician.
8.1 Instrument will not calibrate: If the display
does not stabilize, or respond to calibration
adjustments.
8.1.1 Make sure the Run/Calibration toggle switch is
set toward "CAL".
8.1.2Check flow meter. If the flow ball does not rise,
the connector’s slide valve may be off, or calibration
cylinder may be empty.
8.1.3 Replace sensor.
8.2 Intermittent alarm: A pressure switch
monitors pressure of sample-air delivered to the
sensor. If pressure at the switch (which is monitored
by flow through the flow meter) drops below
minimum requirement, or rises above maximum, the
monitor initiates an intermittent alarm.
7.1.16 Return the monitor to service and record sensor
replacement date.
7.1.17 Follow calibration test schedule per Section 6.2.
7.2 Fuse Replacement
7.2.1 The fuse cap is located on the side of the
instrument case. Remove the cap to access the fuse.
Purchase a 1-amp fuse locally and keep spares on
hand.
8.2.1Before doing pressure tests, make sure the
Run/Calibration toggle switch is toward "RUN". If the
toggle is not in the "RUN" position, the instrument will
initiate an intermittent alarm, indicating that the sampleair is not reaching the sensor. Correctly positioning the
toggle will terminate the intermittent alarm.
8.2.2Make sure the sample-air is between 55 psi and
145 psi. If the sample-air is higher than 55 psi, the
restriction is internal. Proceed as follows:
8.2.3Check flow through the flow meter. If flow ball is
below .4 SCFH or above .9 SCFH, adjust the pressure
regulator accordingly, until the flow ball remains between
.5 and .8 SCFH. In some cases, slightly higher or lower
flow is required. Slowly increase and decrease pressure
at the pressure regulator. If the regulator pressure was
set outside the limits, the alarm will disengage as soon
as the pressure is corrected.
8.2.4Check internal instrumentation tubing for breaks,
kinks, or disconnection. If tubing has come loose,
reconnecting it will re-establish function and the monitor
should operate correctly.
8.2.5 Inspect purple restrictor orifice for blockage;
orifice is extremely small .004". Replace the orifice if it is
blocked.
Tubing coming loose can be the result of
supplying the instrument with pressure much
higher than operating pressure. Refer to
Section 2.3.1. If this type of failure occurs, the
maintenance service technician should inspect
the instrument as soon as possible.
8.3 No annunciator lights or alarms. This
condition indicates a loss of electrical power, as it is
unlikely that audible and visual alarms will fail at the
same time.
8.3.1 Make sure the electrical power cord is
connected to the appropriate power source.
8.3.2Make sure the fuse, located in the case, is not
blown.
8.3.3 Make sure that the power supply is ON.
8.3.4 For 12-volt system, make sure the battery is fully
charged and that the charging system is operational.
Make sure the positive and negative terminals are
connected correctly.
8.3.5Check for faulty power supply, or loose plug
connection on circuit board.
8.4 Either alarm-lights or alarm horn fails.
8.4.1Check for loose plug connections on circuit
board.
8.4.2To test the alarm, remove the plug connection
on the circuit board from the suspect alarm, and apply
external 12-volt power to the alarm plug. If the alarm
fails to activate, replace the alarm.
8.4.3To test the circuit board, remove the plug
connection from the faulty alarm. Use a voltmeter to
check voltage across circuit board pins. When testing an
alarm horn, switch the Run/Calibration Switch to the
"CAL" (calibration) position. This will cause an
intermittent alarm, and should register on the voltmeter
each time the alarm activates. A monitor with faulty
circuit board should be returned for service.