Cleaver Scientific TPS150V Instruction Manual

1
TPS150V Power Supply
Instruction manual
www.cleaverscientific.com.com
info@cleaverscientific.com
2
Packing list
- 1x TPS150V
- 1x Power Cord
- 1x Instruction Manual
Signed by:
Date:
Cleaver Scientifc is liable for all missing or damaged parts / accessories within 7 days after customer received this instrument package. Please contact Cleaver Scientific immediately regarding this issue. If no response within such time period from consignee party, that will be consignee party’s whole responsibility.
3
Warning
Cleaver Scientific’s Power Supply has been tested and found to comply with the limits for the CE regulation. It is also RoHS compliant to meet the requirements of the environmental directive. These requirements are designed to provide reasonable protection against harmful interference when the instrument series is operated in a commercial environment. When in use these instruments may generate, use, and radiate radio frequency energy; and if these units are not installed and used in accordance with the instruction manual harmful interference to radio communications may occur. Operation of this instrument series in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at their expense. Changes or modifications not expressly approved by the party responsible for compliance could nullify the user’s authority to operate the equipment. It is strongly recommended for the user to read the following points carefully before operating this equipment.
1. Read and follow the manual instructions carefully.
2. Do not alter the equipment. Failure to follow these directions could result in personal and/or laboratory hazards, as well as invalidate equipment warranty.
3. Use a properly grounded electrical outlet with correct voltage and current handling capacity.
4. Disconnect from power supply before maintenance and servicing. Refer servicing to qualified personnel.
5. Never use this instrument series without having the safety cover correctly in position.
6. Do not use the unit if there is any sign of damage to the external tank or cover. Replace damaged parts.
7. Do not use in the presence of flammable or combustible material; fire or explosion may result. This device contains components which may ignite such materials.
8. Refer maintenance and servicing to qualified personnel.
9. Ensure that the system is connected to electrical service according to local and national electrical codes. Failure to make a proper connection may create fire or shock hazard.
10. Use appropriate materials and operate correctly to avoid possible hazards
4
of explosion, implosion or release of toxic or flammable gases arising from overheated materials.
11. The unit shall be operated only by qualified personnel.
Safety Information
Take all necessary precautions for using any electrical device. Before connecting the electrical supply, check to see if the supply voltage is within the range stated at the rating label, and see to it that the device be seated firmly. Place the unit in a safe and dry location; it must NOT touch its surroundings. Follow the safety precautions for chemicals / dangerous materials. If needed, please contact qualified service representative or info@cleaverscientific.com
Environmental Conditions
Ensure the instrument is installed and operated strictly under the following conditions:
1. Indoor use only
2. ≤95% RH
3. 75 kPa – 106 kPa
4. Altitude must not exceed 2000 meters
5. Ambient to 40°C operating temperature
6. Pollution degree: 2
7. Mains supply voltage fluctuations up to ±10% of the normal voltage
Avoiding Electrical Shock Follow the guidelines below to ensure safe operation of the unit. The NANOPAC Power Supplies has been designed to use insulated wires thus minimizing any potential shock hazard to the user. Cleaver Scientific recommends against the use of uninsulated wires. To avoid electrical shock:
1. In the event of solution spilling on the instrument, it must be dried out for at
least 2 hours and restored to NORMAL CONDITION before each operation.
2. Never connect or disconnect wires loading from the power jacks when the
red indicator light of power switch is on.
3. WAIT at least 5 seconds after stopping a run before handling output leads
or any connected apparatus.
4. ALWAYS make sure that your hands, work area, and instruments are clean
and dry before making any connections or operating the power supply.
5. ONLY connect the power cord to a properly grounded AC outlet.