Version 1
OmniPAGE Vertical Electrophoresis Systems
OmniPAGE Mini
Instruction Manual
Catalogue No: CVS10D
CVS10DSYS
CVS10PRE
CVS10DSYS-CU
![](/html/84/846a/846ad49eb4da8816ce3e1d1a2f5c9cfa3ce31271d16defb7660c9f661a9d1a6a/bg2.png)
Page 2
Table of Contents
Page
Section 1 Safety Information 3
1.1 Safety Precaution 3
Section 2 General Information 4
2.1 Introduction 4
2.2 Product Description 4
2.3 Packing Lists 5
2.4 Specifications 6
2.5 Care and Maintenance 7
Section 3 Operating Instructions 8
3.1 Setting up the omniPAGE Gel Tank 8
3.2 Gel casting 8
3.3 omniPAGE module assembly and Sample loading 13
3.4 Gel Running 14
3.5 Gel Removal 14
Section 4 Gel Preparation 14
4.1 Gel Selection 14
4.2 Volumes required per gel 15
4.3 Gel Preparation 16
4.4 Stacking Gel Preparation 16
4.5 Buffer Volume 17
4.6 Gel Running Conditions 17
4.7 Preparation of denatured protein samples for loading 18
4.8 Stock Solutions 19
Section 5 Trouble Shooting Guide 20
Section 6 Product Information 23
6.1 Catalogue numbers and product descriptions 23
6.2 Combs 24
6.3 Catalogue numbers and product descriptions for CSL Related products 26
Section 7 Warranty 27
![](/html/84/846a/846ad49eb4da8816ce3e1d1a2f5c9cfa3ce31271d16defb7660c9f661a9d1a6a/bg3.png)
Page 3
Section 1 Safety Information
1.1 Safety Precaution
When used correctly, these units pose no health risk. However, these units can deliver dangerous levels of
electricity and are to be operated only by qualified personnel following the guidelines laid out in this
instruction manual. Anyone intending to use this equipment should read the complete manual thoroughly.
The unit must never be used without the safety lid correctly in position. The unit should not be used if there is
any sign of damage to the external tank or lid.
Acrylamide is a powerful neurotoxin in solution form. Polymerized gels can contain some unpolymerized
solution and protective gloves and clothing must be worn.
These units comply with the following European directives:
2006/95/CE Low Voltage Directive and 2014/30/UE (official Title 2004/108/EC) EMC Electromagnetic
Compatibility
By virtue of the following harmonised standards:
BS EN IEC 61010-1: 2010 Safety Testing of Lab Equipment
BS EN IEC 61326-1:2013 EMC Electro Magnetic Compatibility
![](/html/84/846a/846ad49eb4da8816ce3e1d1a2f5c9cfa3ce31271d16defb7660c9f661a9d1a6a/bg4.png)
Page 4
Section 2 General Information
2.1 Introduction
Cleaver Scientifics’ omniPAGE range of Vertical Gel Units combines ease of use with high resolution
separations. Four sizes, Mini 10 x 10cm, Mini Wide 20 x 10cm and Maxi 20 x 20cm, VS 30 x 30cm share a host
of common features including a guaranteed leak proof seal required for trouble free, rapid and uncomplicated
gel casting. Utilising a built in gel running module eliminates time consuming transfer of glass plates during
casting, a process which can cause gel damage and misalignment. Glass plates with permanently bonded
spacers guarantee perfect spacer alignment. The glass plate sandwich is then simply inserted between
pressure bars and the new zero screw slide clamps clicked into position. This ensures fast set up times while
even pressure bars and ultra soft seals guarantee leak proof casting. Once the gel has polymerised, the gel
running module is just inserted into the gel tank for electrophoresis.
2.2 Product Description
Cleaver Scientifics’ vertical electrophoresis units comes as a complete package including Gel Tank, Safety Lid,
Gel Casting Inner Module, Glass plates with bonded spacers, Combs, Multipurpose Key, Power Cables and
Cooling pads. Detailed description of all the components is written below.
Injection moulded gel tank and safety lid provides a sealed electrophoresis system
which is compatible with all major types of 8 X 10cm and 10 x 10cm precast gel. All
omniPAGE tanks are equipped with specially designed thumb locators for lifting up
the safety lid easily. All safety lids are designed to accommodate the polarity
design of the unit. They have special tapped holes which connect with the leading
edge of the power cables.
Injection moulded inner module gives a twofold effectiveness for gel casting and
running, thus no need of transferring the glass plates after casting the gel.
2mm thick glass plates for the omniPAGE mini and 4mm thick glass plates for rest
of the range prevents breakage and have bonded spacers for convenience. All our
spacers are colour coded depending upon their thickness.
CVS10KEY or the multipurpose key can be easily used to separate your notched
and plain glass plate to release the gel. The same key can be used to open the
CVS10 clamping doors.
Cleaver Scientifics’ power cables are designed with protective retractable
connectors which are compatible with most power supplies.
Rapid set up cooling packs enhance resolution eliminating the need of a chiller.
Combs with special
gel loading guides
Special combs (Combicombs) are designed with special oval shaped gel loading
guides on the other end of an ordinary comb which can be used.
![](/html/84/846a/846ad49eb4da8816ce3e1d1a2f5c9cfa3ce31271d16defb7660c9f661a9d1a6a/bg5.png)
Page 5
2.3 Packing Lists
CVS10D, CVS10DSYS, CVS10PRE, CVS10DSYS-CU
Each unit includes a tank, lid, internal module, electrodes and the following accessories:
VS10NG – Notched,
Pk/2
VS10PGS1 – Plain with
bonded 1mm spacers,
Pk/2
VS10-DP – Dummy
Plate
2 of VS10-12-1
1mm thick, 12
sample
VS10NG – Notched,
Pk/2
VS10PGS1 – Plain with
bonded 1mm spacers,
Pk/2
VS10-DP – Dummy
Plate
2 of VS10-12-1
1mm thick, 12
sample
VS10DCAST
VS10DCASTM –
Mat
See CVS10DYSYS
CVS10EXCASTER
Packing list checked by ________________________ Date ____________
The packing lists should be referred to as soon as the units are received to ensure that all components have
been included. The unit should be checked for damage when received.
Please contact your supplier if there are any problems or missing items.
![](/html/84/846a/846ad49eb4da8816ce3e1d1a2f5c9cfa3ce31271d16defb7660c9f661a9d1a6a/bg6.png)
Page 6
2.4 Specifications of OmniPAGE Vertical Electrophoresis Units
Plate
Dimensions
Gel Dimensions
(WxL)
26 X 16 X 16cm
(W X D X H)
26 X 16 X 28CM
(W X D X H)
36 X 33 X 18CM
(W X H X D)
80 Samples
20 Samples per Gel
192 Samples
48 Samples per Gel
192 Samples
48 Samples per Gel
300 Samples per
Run
75 Samples per Gel
Combs Available
No. of Teeth
Thickness
1, 5, 8MC, 9, 10,
12, 16MC, 20
0.75, 1, 1.5, 2mm
1, 5, 10, 18MC, 24,
30, 36MC, 48
0.75, 1, 1.5, 2mm
1, 5, 10, 18MC, 24,
30, 36MC, 48
0.75, 1, 1.5, 2mm
1, 2, 4, 28MC,
56MC, 75
0.25, 0.35, 0.5, 1,
1.5, 2.0mm
Environmental
Operating
Conditions
Maximum Altitude
2,000 m
Temperature
Range 4°C - 65°C
Humidity
Upto 80%
Not for outdoor
Use
Maximum Altitude
2,000 m
Temperature
Range 4°C - 65°C
Humidity
Upto 80%
Not for outdoor
Use
Maximum Altitude
2,000 m
Temperature
Range 4°C - 65°C
Humidity
Upto 80%
Not for outdoor
Use
Maximum Altitude
2,000 m
Temperature
Range 4°C - 65°C
Humidity
Upto 80%
Not for outdoor
Use
This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.
POLLUTION DEGREE 2, states that: “Normally only non-conductive pollution occurs.
Occasionally, however, a temporary conductivity caused by condensation must be expected”.
![](/html/84/846a/846ad49eb4da8816ce3e1d1a2f5c9cfa3ce31271d16defb7660c9f661a9d1a6a/bg7.png)
Page 7
2.5 Care and Maintenance
Cleaning Large Format Vertical Units
Units are best cleaned using warm water and a mild detergent. Water at temperatures above 600 C can cause
damage to the unit and components.
The inner module should be thoroughly rinsed with warm water or distilled water to prevent buildup of salts
but care should be taken not to damage the enclosed electrode and vigorous cleaning is not necessary or
advised. Air drying preferably before use.
The units should only be cleaned with the following:
Warm water with a mild concentration of soap or other mild detergent.
Compatible detergents include dishwashing liquid, Hexane and Aliphatic hydrocarbons. The units should not
be left in detergents for more than 30 minutes.
The units should never come into contact with the following cleaning agents, these will cause irreversible
and accumulative damage:
Acetone, Phenol, Chloroform, Carbon tetrachloride, Methanol, Ethanol, Isopropyl alcohol, Alkalis.
RNase Decontamination
This can be performed using the following protocol:Clean the units with a mild detergent as described above.
Wash with 3% hydrogen peroxide (H2O2) for 10 minutes.
Rinsed with 0.1% DEPC- (diethyl pyrocarbonate) treated distilled water.
Caution: DEPC is a suspected carcinogen. Always take the necessary precautions when using. RNaseZAP™
(Ambion) can also be used. Please consult the instructions for use with acrylic gel tanks.
![](/html/84/846a/846ad49eb4da8816ce3e1d1a2f5c9cfa3ce31271d16defb7660c9f661a9d1a6a/bg8.png)
Page 8
Section 3 Operating Instructions
3.1 Setting up the omniPAGE Gel Tank
Note: Before setting up the Gel Tank please ensure that it has been properly cleaned and dried.
1. Note the position of the lid on the unit. This shows the correct polarity and the correct orientation of the
cables, black is negative and red positive.
2. Remove the lid from the unit.
Note: If the lid is not removed, fitting the cables may result in un-tightening of the gold plug and damage
the electrode.
3. Screw the cables into the tapped holes as fully as possible so that there is no gap between the lid and the
leading edge of the cable fitting.
4. Refit the lid and the unit is now ready to be used.
3.2 Gel casting
Cleaning the Glass Plates
Clean a set of glass plates for each gel first with distilled water and then with 70 % ethanol.
One set of glass plates constitutes one notched glass plate and one plain glass plate with bonded spacers.
When using a triple glass plate sandwich, two notched glass plates are required, one set of free spacers and a
set of plain glass plates with bonded spacers. The plain glass plate is positioned outermost, then a notched
glass plate, free spacers and second notched glass plate. Alternatively, accessory notch glass plates with
bonded spacers are available.
Note: All glass plates, gel casting modules, casting base and accessories must be completely dry before the
set – up. Wet components are more likely to miss-align and cause leaks.
![](/html/84/846a/846ad49eb4da8816ce3e1d1a2f5c9cfa3ce31271d16defb7660c9f661a9d1a6a/bg9.png)
Page 9
Glass cassette Assembly
Assemble the glass plates so that the bottom of the glass plates and the spacers are perfectly aligned.
For triple plate sandwiches, the free spacers Need to be perfectly aligned which is best performed using a
small spacer or comb to push the spacers apart. Notched glass plates with bonded spacers do not need
manual alignment.
NOTE: The glass plates with bonded spacers have an arrow in the top of the spacers which are slightly
longer than the glass plate to indicate the top.
Casting Stand Assembly
Position the Slab Gel Insert on a flat surface.
Insert the glass plates into the Slab Gel Insert
between the pressure bar and the blue gasket.
The Slab Gel Insert contains pressure bars
which impart even pressure onto the glass plates
and allow even screw pressure transfer onto the
sealing edge of the glass plate, ensuring complete sealing. Ensure that the pressure bars are adequately open
for the thickness of spacer used. The bar can be opened by loosening the screws or by sliding the clamps.
When using a triple glass plate sandwich, the pressure bars will need to be in the completely open position.
Then fully tighten the pressure bar screws in the order top then bottom.