When used correctly, these units pose no health risk. However, these units
can deliver dangerous levels of electricity and are to be operated only by
qualified personnel following the guidelines laid out in this instruction manual.
Anyone intending to use this equipment should read the complete manual
thoroughly. The unit must never be used without the safety lid correctly in
position. The unit should not be used if there is any sign of damage to the
external tank or lid.
These units comply with the following European directives:
2006/95/CE Low Voltage Directive and 2014/30/UE (official Title 2004/108/EC)
EMC Electromagnetic Compatibility
By virtue of the following harmonised standards:
BS EN IEC 61010-1: 2010 Safety Testing of Lab Equipment
BS EN IEC 61326-1:2013 EMC Electro Magnetic Compatibility
15/11/2018 Page 4
Packing List
Each multiSUB unit includes a tank, wired electrodes, lid and the following
items:
SKU Tray Tray Dams Combs Loading Guides Cables
MSMINI7 MS7-UV7, 7 x
7cm (W x L)
MSMINI10 MS7-UV10, 7
x 10cm (W x
L)
MSMINIDUO MS7-UV7,
MS7-UV10
MSMIDI7 MS10-UV7, 10
x 7cm (W x L)
MSMIDI10 MS10-UV10,
10 x 10cm (W
x L)
MSMIDIDUO MS10-UV7,
MS10-UV10
MS7-UVDAM,
Pack of 2
MS7-UVDAM,
Pack of 2
MS7-UVDAM,
Pack of 2
MS10-UVDAM,
Pack of 2
MS10-UVDAM,
Pack of 2
MS10-UVDAM,
Pack of 2
2 x MS7-8-1,
1mm, 8 sample
2 x MS7-8-1,
1mm, 8 sample
2 x MS7-8-1,
1mm, 8 sample
2 x MS10-16-1,
1mm, 16
sample
2 x MS10-16-1,
1mm, 16
sample
2 x MS10-16-1,
1mm, 16
sample
MS7-LG – Strips,
MS7-WP – Platform
MS7-LG – Strips,
MS7-WP – Platform
MS7-LG – Strips,
MS7-WP – Platform
MS10-LG – Strips,
MS10-WP –
Platform
MS10-LG – Strips,
MS10-WP –
Platform
MS10-LG – Strips,
MS10-WP –
Platform
CSLCAB
CSLCAB
CSLCAB
CSLCAB
CSLCAB
CSLCAB
MSCHOICE7 MS15-UV7, 15
x 7cm (W x L)
MSCHOICE10 MS15-UV10,
15 x 10cm (W
x L)
MSCHOICE15 MS15-UV15,
15 x 15cm (W
x L)
MSCHOICETRIOMS15-UV7,
MS15-UV10,
MS15-UV15
MSCHOICEST20MS15-UVST20
15 x 20 cm,
(W x L)
MSCHOICEST25MS15-UVST25
15 x 25 cm,
(W x L)
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
2 x MS15-20-1,
1mm, 20
sample
2 x MS15-20-1,
1mm, 20
sample
2 x MS15-20-1,
1mm, 20
sample
2 x MS15-20-1,
1mm, 20
sample
4 x MS15-28MC1, 1mm thick, 28
sample MC
4 x MS15-28MC1, 1mm thick, 28
sample MC
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
CSLCAB
CSLCAB
CSLCAB
CSLCAB
CSLCAB
CSLCAB
15/11/2018 Page 5
MSMAXI10 MS20-UV10,
20 x 10cm (W
x L)
MS20-UVDAM,
Pack of 2
2 x MS20-20-1,
1mm, 20
sample
MS20-LG – Strips,
MS20-WP –
Platform
CSLCAB
MSMAX15 MS20-UV15,
20 x 15cm (W
x L)
MSMAXI20 MS20-UV20,
20 x 20cm (W
x L)
MSMAXIDUO MS20-UV10,
MS20-UV20
MSMAXI25 MS20-UV25,
20 x 25cm (W
x L)
MS20-UVDAM,
Pack of 2
MS20-UVDAM,
Pack of 2
MS20-UVDAM,
Pack of 2
MS20-UVDAM,
Pack of 2
2 x MS20-20-1,
1mm, 20
sample
2 x MS20-20-1,
1mm, 20
sample
2 x MS20-20-1,
1mm, 20
sample
2 x MS20-20-1,
1mm, 20
sample
Packing List Checked by: ________________________
Date: ________________________
MS20-LG – Strips,
MS20-WP –
Platform
MS20-LG – Strips,
MS20-WP –
Platform
MS20-LG – Strips,
MS20-WP –
Platform
MS20-LG – Strips,
MS20-WP –
Platform
CSLCAB
CSLCAB
CSLCAB
CSLCAB
The packing lists should be referred to as soon as the units are received to
ensure that all components have been included. The unit should be checked
for damage when received.
Cleaver Scientific is liable for all missing or damaged parts / accessories within
7 days after customers have received this instrument package. Please contact
Cleaver Scientific immediately regarding this issue. If no response within such
period is received from the customer, Cleaver Scientific will no longer be liable
for replacement/damaged parts.
Please contact your supplier if there are any problems or missing items.
15/11/2018 Page 6
Specifications
1 Power supply cables 1500V rated with retractable 4mm connectors
2 Safety lid and viewing pane High impact moulded acrylic construction
9 Safety lid thumb locators For safe removal of lid
10 Moulded buffer tank High Impact moulded acrylic construction for chemical
compatibility and shock resistance
15/11/2018 Page 7
Operating Instructions
Further information (including videos) regarding setting up and running the
multiSUB® units can be found at www.cleaverscientific.com
Usage Guidance and restrictions
Maximum altitude 2,000m.
Temperature range between 4°C and 65°C.
Maximum relative humidity 80% for temperatures up to 31OC
decreasing linearly to 50% relative humidity at 40OC.
Not for outdoor Use.
This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.
POLLUTION DEGREE 2, states that: “Normally only non-conductive pollution
occurs.
Occasionally, however, a temporary conductivity caused by condensation
must be expected”.
Setting up the Horizontal Gel Tanks
Fitting Electrodes
Note the position of the lid on the unit. This shows the correct polarity
and the correct orientation of the cables, black is negative and red
positive.
Remove the lid from the unit. Note if the lid is not removed, fitting the
cables may result in un-tightening of the gold plug and damage to the
electrode.
Screw the cables into the tapped holes as fully as possible so that there
is no gap between the lid and the leading edge of the cable fitting.
Refit the lid.
Fitting Loading Guides
These can be fitted to enhance visibility of the wells if desired. They can be
fitted to the white vinyl platform sheet or to the unit itself.
1. Seat the tray in the unit and note the position of the comb grooves. The
samples run black to red but the trays can be used frontward or
backwards so ensure that the comb grooves closest to the black
electrode are marked.
15/11/2018 Page 8
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.