When used correctly, these units pose no health risk. However, these units
can deliver dangerous levels of electricity and are to be operated only by
qualified personnel following the guidelines laid out in this instruction manual.
Anyone intending to use this equipment should read the complete manual
thoroughly. The unit must never be used without the safety lid correctly in
position. The unit should not be used if there is any sign of damage to the
external tank or lid.
These units comply with the following European directives:
2006/95/CE Low Voltage Directive and 2014/30/UE (official Title 2004/108/EC)
EMC Electromagnetic Compatibility
By virtue of the following harmonised standards:
BS EN IEC 61010-1: 2010 Safety Testing of Lab Equipment
BS EN IEC 61326-1:2013 EMC Electro Magnetic Compatibility
15/11/2018 Page 4
Page 7
Packing List
Each multiSUB unit includes a tank, wired electrodes, lid and the following
items:
SKU Tray Tray Dams Combs Loading Guides Cables
MSMINI7 MS7-UV7, 7 x
7cm (W x L)
MSMINI10 MS7-UV10, 7
x 10cm (W x
L)
MSMINIDUO MS7-UV7,
MS7-UV10
MSMIDI7 MS10-UV7, 10
x 7cm (W x L)
MSMIDI10 MS10-UV10,
10 x 10cm (W
x L)
MSMIDIDUO MS10-UV7,
MS10-UV10
MS7-UVDAM,
Pack of 2
MS7-UVDAM,
Pack of 2
MS7-UVDAM,
Pack of 2
MS10-UVDAM,
Pack of 2
MS10-UVDAM,
Pack of 2
MS10-UVDAM,
Pack of 2
2 x MS7-8-1,
1mm, 8 sample
2 x MS7-8-1,
1mm, 8 sample
2 x MS7-8-1,
1mm, 8 sample
2 x MS10-16-1,
1mm, 16
sample
2 x MS10-16-1,
1mm, 16
sample
2 x MS10-16-1,
1mm, 16
sample
MS7-LG – Strips,
MS7-WP – Platform
MS7-LG – Strips,
MS7-WP – Platform
MS7-LG – Strips,
MS7-WP – Platform
MS10-LG – Strips,
MS10-WP –
Platform
MS10-LG – Strips,
MS10-WP –
Platform
MS10-LG – Strips,
MS10-WP –
Platform
CSLCAB
CSLCAB
CSLCAB
CSLCAB
CSLCAB
CSLCAB
MSCHOICE7 MS15-UV7, 15
x 7cm (W x L)
MSCHOICE10 MS15-UV10,
15 x 10cm (W
x L)
MSCHOICE15 MS15-UV15,
15 x 15cm (W
x L)
MSCHOICETRIOMS15-UV7,
MS15-UV10,
MS15-UV15
MSCHOICEST20MS15-UVST20
15 x 20 cm,
(W x L)
MSCHOICEST25MS15-UVST25
15 x 25 cm,
(W x L)
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
MS15-UVDAM,
Pack of 2
2 x MS15-20-1,
1mm, 20
sample
2 x MS15-20-1,
1mm, 20
sample
2 x MS15-20-1,
1mm, 20
sample
2 x MS15-20-1,
1mm, 20
sample
4 x MS15-28MC1, 1mm thick, 28
sample MC
4 x MS15-28MC1, 1mm thick, 28
sample MC
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
MS15-LG – Strips,
MS15-WP –
Platform
CSLCAB
CSLCAB
CSLCAB
CSLCAB
CSLCAB
CSLCAB
15/11/2018 Page 5
Page 8
MSMAXI10 MS20-UV10,
20 x 10cm (W
x L)
MS20-UVDAM,
Pack of 2
2 x MS20-20-1,
1mm, 20
sample
MS20-LG – Strips,
MS20-WP –
Platform
CSLCAB
MSMAX15 MS20-UV15,
20 x 15cm (W
x L)
MSMAXI20 MS20-UV20,
20 x 20cm (W
x L)
MSMAXIDUO MS20-UV10,
MS20-UV20
MSMAXI25 MS20-UV25,
20 x 25cm (W
x L)
MS20-UVDAM,
Pack of 2
MS20-UVDAM,
Pack of 2
MS20-UVDAM,
Pack of 2
MS20-UVDAM,
Pack of 2
2 x MS20-20-1,
1mm, 20
sample
2 x MS20-20-1,
1mm, 20
sample
2 x MS20-20-1,
1mm, 20
sample
2 x MS20-20-1,
1mm, 20
sample
Packing List Checked by: ________________________
Date: ________________________
MS20-LG – Strips,
MS20-WP –
Platform
MS20-LG – Strips,
MS20-WP –
Platform
MS20-LG – Strips,
MS20-WP –
Platform
MS20-LG – Strips,
MS20-WP –
Platform
CSLCAB
CSLCAB
CSLCAB
CSLCAB
The packing lists should be referred to as soon as the units are received to
ensure that all components have been included. The unit should be checked
for damage when received.
Cleaver Scientific is liable for all missing or damaged parts / accessories within
7 days after customers have received this instrument package. Please contact
Cleaver Scientific immediately regarding this issue. If no response within such
period is received from the customer, Cleaver Scientific will no longer be liable
for replacement/damaged parts.
Please contact your supplier if there are any problems or missing items.
15/11/2018 Page 6
Page 9
Specifications
1 Power supply cables 1500V rated with retractable 4mm connectors
2 Safety lid and viewing pane High impact moulded acrylic construction
9 Safety lid thumb locators For safe removal of lid
10 Moulded buffer tank High Impact moulded acrylic construction for chemical
compatibility and shock resistance
15/11/2018 Page 7
Page 10
Operating Instructions
Further information (including videos) regarding setting up and running the
multiSUB® units can be found at www.cleaverscientific.com
Usage Guidance and restrictions
Maximum altitude 2,000m.
Temperature range between 4°C and 65°C.
Maximum relative humidity 80% for temperatures up to 31OC
decreasing linearly to 50% relative humidity at 40OC.
Not for outdoor Use.
This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.
POLLUTION DEGREE 2, states that: “Normally only non-conductive pollution
occurs.
Occasionally, however, a temporary conductivity caused by condensation
must be expected”.
Setting up the Horizontal Gel Tanks
Fitting Electrodes
Note the position of the lid on the unit. This shows the correct polarity
and the correct orientation of the cables, black is negative and red
positive.
Remove the lid from the unit. Note if the lid is not removed, fitting the
cables may result in un-tightening of the gold plug and damage to the
electrode.
Screw the cables into the tapped holes as fully as possible so that there
is no gap between the lid and the leading edge of the cable fitting.
Refit the lid.
Fitting Loading Guides
These can be fitted to enhance visibility of the wells if desired. They can be
fitted to the white vinyl platform sheet or to the unit itself.
1. Seat the tray in the unit and note the position of the comb grooves. The
samples run black to red but the trays can be used frontward or
backwards so ensure that the comb grooves closest to the black
electrode are marked.
15/11/2018 Page 8
Page 11
2. Remove the tray.
3. Peel the back off the loading guide and carefully apply the loading
guide directly to the gel platform.
The unit is now ready to be used.
Gel Preparation
The Table below shows the volume of agarose solution required to make the
desired agarose gel for each unit tray size. For a standard 0.7% agarose gel,
add 0.7 grams of agarose to 100 ml of 1x TAE or TBE solution. The same 1X
solution should be used in the tank buffer solution.
multiSUB® Mini
multiSUB® Midi
multiSUB®
Choice
multiSUB®
Choice STRETCH
multiSUB® Maxi
Tray
Gel volume for a 5mm thick
gel
Tray
Gel volume for a 5mm thick
gel
Tray
Gel volume for a 5mm thick
gel
Tray
Gel volume for a 5mm thick
gel
Tray
7 x 7 cm 7 x 10 cm
25 mL 35 mL
10 x 7 cm 10 x 10 cm
35 mL 50 mL
15 x 7 cm15 x 10 cm15 x 15 cm
52.5 mL 75 mL 112.5 mL
15 x 20 cm15 x 25 cm
150 mL187.5 mL
20 x 10 cm 20 x 15 cm 20 x 20 cm
Gel volume for a 5mm thick
gel
100 mL 150 mL 200 mL
1. Add the agarose powder to a conical flask.
2. Add the appropriate amount of 1x TAE or TBE solution from the table
above. To prevent evaporation during the dissolving steps below, the
conical flask should be covered with parafilm.
3. Dissolve the agarose powder by heating the agarose either on a
magnetic hot plate with stirring bar or in a microwave oven. If using the
microwave method, the microwave should be set at around a 400 watt
or medium setting and the flask swirled every minute. The solution
15/11/2018 Page 9
Page 12
should be heated until all crystals are dissolved. This is best viewed
against a light background. Crystals appear as translucent crystals.
These will interfere with sample migration if not completely dissolved.
The gel must be cooled to between 50°C and 60°C degrees before pouring.
Gel Pouring
The multiSUB® range of units allows three different methods of gel casting:
1. Casting Dams
2. Flexicaster
3. Traditional Tape
Casting Dams
1. To fit the casting dams, place one casting dam on the bench with the
groove facing upwards (1). Push the edge of the tray down firmly into
the groove (2). Repeat this for the other side (3). The dams should be
fitted so that there is no gap between the sides of the tray and the
groove in the dams. This will ensure that there is no possibility of gel
leakage.
2. Place the comb(s) in the grooves. Each tray has more than one comb
grove so that multiple combs can be used. Using multiple combs
increases sample number available per gel but decreases run length
and care must be taken to ensure that samples from the first wells do
not migrate into the lanes of the second comb wells.
3. Pour in the agarose carefully so as not to generate bubbles. Any
bubbles that do occur can be smoothed to the edge of the gel and
dispersed using a pipette tip.
4. Allow the agarose to set, ensuring that the gel remains undisturbed.
15/11/2018 Page 10
Page 13
5. Carefully remove the gel casting gates and comb and transfer the gel
including tray to the main tank.
(1)(2)(3)
15/11/2018 Page 11
Page 14
Flexicaster
1. Level the Flexicaster base by adjusting the feet so that the bubble is
exactly central.
2. Insert the desired length tray into the Flexicaster such that one end of
the tray is pushed up and seals against the silicone mat of the
permanent end of the Flexicaster.
3. Position the movable end of the Flexicaster so that the silicone mat is
pushed against the other end of the tray.
4. Turn the cam so that the silicone mat tightly seals against the side of the
tray. Pour in the agarose carefully so as not to generate bubbles. Any
bubbles that do occur can be smoothed to the edge of the gel and
dispersed using a pipette tip.
5. Allow the agarose to set, ensuring that the gel remains undisturbed.
6. Carefully remove the gel casting gates and comb and transfer the gel
including tray to the main tank.
MS7-FC and MS20-FC Flexicasters
15/11/2018 Page 12
Page 15
Tape
1. Autoclave or plastic backed general tape should be used. A length
5cm longer than the width of each end of the tray should be cut. One
length should be placed over one end of the tray and stuck m1cm in
from the tray edge. This should then be folded, and the edges sealed
securely. Repeat for the other end and place onto a level surface for
gel pouring.
2. Place the comb(s) in the grooves. Each tray has more than one comb
grove so that multiple combs can be used. Using multiple combs
increases sample number available per gel but decreases run length
and care must be taken to ensure that samples from the first wells do
not migrate into the lanes of the second comb wells.
3. Pour in the agarose carefully so as not to generate bubbles. Any
bubbles that do occur can be smoothed to the edge of the gel and
dispersed using a pipette tip.
4. Allow the agarose to set, ensuring that the gel remains undisturbed.
5. Carefully remove the gel casting gates and comb and transfer the gel
including tray to the main tank.
Running the Gel
1. Mix the sample to be loaded with sample buffer – see solutions for
common sample buffers. Usually 3ul of sample buffer is adequate but
less may be used with sample volumes of less than 10ul.
2. Fill the unit with buffer until the gel is just flooded with buffer. This will give
the fastest resolution times. For enhanced quality of resolution of
sample, fill the unit to 5mm above the gel.
3. Load the samples into the wells using pipettes. Multi-channel pipettes
can be used for loading samples with MC compatible combs, see
listing in accessories for identification of these.
4. Carefully place the lid on the tank and connect to a power supply.
5. Typically, gels are run at between 90 and 150 volts. However, maximum
voltages are indicated on the serial badge of each unit. It should be
noted that higher voltages generally give faster but poorer quality
sample resolution.
15/11/2018 Page 13
Page 16
Gel Staining and Viewing
The Multi Sub trays allow staining to be performed without removing the gel
from the tray if this is preferred.
1. Transfer the gel to a vessel containing the appropriate volume of 0.5 µg/ml
ethidium bromide stain for 15–30 minutes, see solutions for stock stain
concentration and adjust to the volume used accordingly. The entire gel
should be covered.
NOTE: Ethidium bromide is a suspected carcinogen and the necessary safety
precautions should be taken.
2. De-stain the gel for 10–30 minutes in distilled water again ensuring the gel
is completely immersed.
3. Rinse the gel twice for a couple of seconds with distilled water.
4. Transfer the gel to a UV Transilluminator.
The samples will often appear as brighter, clearer bands when
photographed or viewed using a gel documentation system. However, if the
gel bands are too faint then the staining procedure should be adjusted so
that there is less de-staining. If there is too much background, then the
staining procedure should be adjusted so that there is more de-staining.
Solutions
0.5M EDTA stock (500mL) dissolve in 400 ml distilled water:
93.05g EDTA disodium salt
Fill to 500 ml litre final volume with distilled water
50X TAE stock (1L) dissolve in 750 ml distilled water:
242 g tris base (FW = 121)
57.1 ml glacial acetic acid
100 ml 0.5 M EDTA (pH 8.0).
Fill to 1 litre final volume with distilled water
15/11/2018 Page 14
Page 17
10X TBE stock (1L) dissolve in 750 ml distilled water:
108 g tris base (FW = 121)
55 g boric acid (FW = 61.8)
40 ml 0.5 M EDTA (pH 8.0)
Fill to 1 litre final volume with distilled water
Loading Dye
10x sample buffer stock consists of 50% glycerol, 0.25% bromophenol blue,
and 0.25% xylene cyanole FF in 1x TAE buffer. Only 1–10 ml of the 10x loading
dye should be prepared.
Ethidium Bromide Solution
Add 10 mg of Ethidium Bromide to 1 ml distilled water.
References
1. Sambrook, Fritsch, and Maniatis, Molecular Cloning A Laboratory
Manual, Second Edition, Cold Spring Harbor Laboratory Press, 1989.
2. Current Protocols in Molecular Biology, Greene Publishing Associates
and Wiley-Interscience, 1989.
15/11/2018 Page 15
Page 18
Troubleshooting
Problem Cause Solution
Bands sharp but not enough
bands seen
Band smearing and streaking
Curved line or distortion of bands Bubbles in sample wells
Curved bands, smiles Sample overload Reduce load.
Gel agarose percentage too
high
Incomplete digestion
Agarose has improper
endosmosis
Salt concentration in sample too
high
Excessive power and heating
Sample spilled out of well
Incomplete digestion, nuclease
contamination, bad enzyme
Sample wells cast through the
gel. Sample leaks along bottom
of running surface
Decrease agarose percentage.
Review enzyme activity, digest
further.
Reduce salt concentration to
≤0.1M.
Reduce voltage. See
electrophoresis instructions.
Apply sample carefully. Increase
gel thickness for large sample
volumes. Adjust comb height.
Comb should be placed to 1 to 2
mm above the base of the
running surface.
Remove bubbles prior to
electrophoresis.
Differential relative mobilities
Gels crack
High MW bands sharp; low MW
bands smeared
Ragged bands
Sample spilled out of wells
Unit not levelled
Too high voltage gradient,
especially with low melting
temperature agarose or low gel
strength gels
Gel agarose percentage too low
Sample density incorrect
Sample well deformed
Excessive power or heating
Samples should have proper
density. Apply carefully.
Level unit. Use a steady work
bench.
Reduce voltage. Run gel at lower
temperature.
Increase agarose percentage.
Switch to polyacrylamide.
See sample application
instructions.
Carefully remove comb,
especially from soft gels. Make
sure gel has solidified.
Cooling soft gels aids in comb
removal.
Reduce voltage. See
electrophoresis instructions.
Slanted lanes (bands)
Gel not fully solidified
Comb warped or at an angle
Gel to solidify for at least 3045min.
Check alignment of comb.
15/11/2018 Page 16
Page 19
15/11/2018 Page 17
Page 20
Care and Maintenance
Cleaning Horizontal Units
Units are best cleaned using warm water and a mild detergent. Water at
temperatures above 60°C can cause damage to the unit and components.
The tank should be thoroughly rinsed with warm water or distilled water to
prevent build-up of salts, but care should be taken not to damage the
enclosed electrode and vigorous cleaning is not necessary or advised.
Air drying is preferably before use.
The units should only be cleaned with the following:
Warm water with a mild concentration of soap or other mild detergent.
Compatible detergents include dishwashing liquid, Hexane and Aliphatic
hydrocarbons
The units should not be left to in detergents for more than 30 minutes.
The units should never come into contact with the following cleaning agents,
these will cause irreversible and accumulative damage:
TBE10X1L Cleaver Buffer Tris-Borate-EDTA Running Buffer- 10 x 1L
TBE10X5 Cleaver Buffer Tris-Borate-EDTA Running Buffer- 10 x 5L
TAE50X1L Cleaver Buffer Tris-Borate-EDTA Running Buffer- 50 x 1L
TAE50X5L Cleaver Buffer Tris-Borate-EDTA Running Buffer- 50 x 5L
CSL-MDNA-100BH 100bp DNA ladder, 100 – 300bp, 1 x 500µl vial
CSL-MDNA-100BP 100bp DNA ladder, 100 – 1500bp, 1 x 500µl vial
CSL-MDNA-1KB 1Kb DNA ladder, 250 – 10Kb, 1 x 500µl vial
CSL-MDNA-50BP 50bp DNA ladder, 50 – 1500bp, 1 x 500µl vial
CSL-MDNA-BR Broad Range DNA ladder, 100bp – 10Kb, 1 x 500µl vial
CSL-MDNA-HR High Range DNA ladder, 250bp – 25Kb, 1 x 500µl vial
CSL-LOADDYE 10x Bromophenol Blue Loading Dye, 1mL
CSL-LOADDYE10 10x Bromophenol Blue Loading Dye, 10mL
SAFEVIEW BLUE Light Transilluminator 21 x 21cm
CSLUVTS312 UV Transilluminator, small 21 x 21 cm, 312nm
CSL-GELX4 4mm x 1mm, Gel Cutting Tips, 250/ bag
CSL-GELX4RACK 4mm x 1mm, Gel Cutting Tips, 5 racks of 48
CSL-GELX6.5 6.5mm x 1mm, Gel Cutting Tips, 250/ bag
CSL-GELX6.5RACK 6.5mm x 1mm, Gel Cutting Tips, 5 racks of 48
CSLQSPIN
CV20 Cleaver Pipette - Volume; 2 - 20µl
520, 560 & 580nm filters (OMNIDOC-SYBR, -AF560 & -AF580); and White
Light Table (OMNIDOC-WLT). Requires a PC or laptop
Agarose Tablet 100g, Low EEO (200 x 0.5g tablets, supplies as 20 blister
packs of 10 x 0.5g tablets)
Mini Centrifuge complete with 1.5/2.0 ml rotor, strip tube rotor, 0.5 and
0.4 ml adapters, 230V, Purple lid
15/11/2018 Page 26
Page 29
15/11/2018 Page 27
Page 30
Warranty
The Cleaver Scientific Ltd. (CSL) multiSUB® Horizontal Electrophoresis units
have a warranty against manufacturing and material faults of twelve months
from date of customer receipt.
If any defects occur during this warranty period, CSL will repair or replace the
defective parts free of charge.
This warranty does not cover defects occurring by accident or misuse or
defects caused by improper operation.
Units where repair or modification has been performed by anyone other
than CSL or an appointed distributor or representative are no longer under
warranty from the time the unit was modified.
Units which have accessories or repaired parts not supplied by CSL or its
associated distributors have invalidated warranty.
CSL cannot repair or replace free of charge units where improper solutions or
chemicals have been used. For a list of these please see the Care and
Maintenance subsection.
If a problem does occur, then please contact your supplier or Cleaver
Scientific Ltd:
Cleaver Scientific Ltd.
Unit 41, Somers Road Industrial Estate
Rugby, Warwickshire, CV22 7DH
Tel: +44 (0)1788 565300
Email: info@cleaverscientific.com
15/11/2018 Page 28
Page 31
Page 32