CLEAN Instruments CON30 Operation Manual

Page 1
CON30
ENT
SETCAL
1. Product Overview
-2-
4. Key panel
-3-
-4-
Operation
Manual
Conductivity tester
CON30 Conductivity / TDS / Salinity tester
1.
2.
3
4.
5.
6.
7.
8.
9.
10.
11.
12.
2
3
3
4
4
5
5
6
7
7
8
8
Product Overview
Precautions
CONFIGURATION and LCD INTRODUCTION
Key Panel
Battery and electrode installation
Measurement Mode
Calibration Mode
Set up
Parameters
Package
Error
GENERAL INFORMATION
····························································
····································································
······················
········································································
··································
·······················································
····························································
··············································································
····································································
··········································································
··················································································
···················································
TABLE OF CONTENTS
⻾䱃حக
Water Analysis Solutions
3. CONFIGURATION and LCD INTRODUCTION
Set up
Data stabilized
Calibration Mode
Low battery indicator
Measurement mode
Auto Lock function
Auto Lock after data stabilized
In calibration
Sensor slope is above 85%
Sensor slope is between 70%-85%
ENT
5. Battery and Electrode installation
Unscrew the screw cap by following the
arrow direction, slightly unplug the
electrode, install the battery(pay attention
to the anode and cathode of the
electrode).
Then install back the electrode and the
screw cap.
Low battery prompt is at the top of the
screen when battery power is low.( You can
refer to the LCD INTRODUCTION). Please
change the same model battery.
CAL
ENT
SET
CAL
PH30
ENT
SET
CAL
PH30
AAA
+
SET
2. Precautions
Floats on water
IP67 rated
Large LCD with backlight Multi-line display for easy reading
Field throw measurement ( Auto Lock Function)
1*1.5V AAA
long battery life
www.cleaninst.com E-mail: CS@cleaninst.com 021-63531465
P H30
P H 3 0
E NT
E N T
SET
S E T
C A L
C A L
PH30PH3
0
ENTENT
SETSET
CALCAL
ENTENT
SETSETCALCA
L
+
-
AAA
Calibration key when in measurement mode Long press to enter into parameter browsing Increment values or scroll through the next options available
Long press to Power ON/OFF Short press to ON/OFF backlight Escape key when in calibration mode and setting up
Data setting Decrement values or scroll through the next options available
Confirm key Press to unlock in Auto lock mode
Short press: Short press means release the key once after pressing. (If there is no mark out below, default it as Short press)
Long press: Long press means press the key for 3 seconds and then release it Hold: Hold means not release the key and accelerate after a certain time. Not release the key
until the value is set as needed.
Key instruction
⻾䱃حக
CLEAN INSTRUMENTS
Thank you for selecting Clean products.
Before using the tester, relevant person must read and understand the contents
of this operation manual. Operation Manual should be kept accessible within
the person who use the testers.
Precise and easy operation, Comfortable holding, all functions operated in one hand.
Large LCD with backlight, Multi-line display for easy reading, Auto-Power Off saves battery after
5 mins non-use. 1*1.5V AAA long battery life.
Floats on water, IP67 rated.
Field throw measurement( Auto Lock Function)
Broad measuring range: 0.0 S/cm -20.00mS/cm,from pure water to waste water applications.
ȕ
ȕ
ȕ
ȕ
ȕ
ȕ
CAL
ENT
SET
CAL
PH30
ENT
SET
CAL
Please wipe up the electorde with clean tissue or cloth before using.
Do not throw or beat ,squash or twist the tester deliberately.
Do not expose to the direct sunlight or put in high humidity or high
temperature environment .
Do not use for long time under water.
CON30
ENT
SETCAL
Page 2
-6-
-7- -8-
8. Set up
Press ” ” to power up. The tester will then enter into measurement mode.
The measurement icon will twinkle if the data is unstabilized. Once stabilized, the top
stabilized icon will be displayed then.
Note: In measurement mode, long press " " to switch between Conductivity / TDS /
Salinity measurement mode.
In measurement mode, press “ ” to enter calibration mode. Prompting icons will be
displayed on the screen.
6. Measurement Mode
7. Calibration Mode
In measurement mode, press “ ” to enter SET UP. Corresponding icons will be
displayed. You can press ” ” and “ ” to select the menu and then press ”
to confirm. Press ” ” or “ ” to set up and confirm by press ” ”. You can
also press “ ” to back to measurement mode.
The Set up menu as follow:
ENT
ENT
CAL
SET
SET
CAL
SET
9. Specification
12. GENERAL INFORMATION
10. Package
Size: L*W*H 210*100*65mm Weight: 350g
Weight
Series
Tester
.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www
Series
Tester
ww.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.cleaninst.com www.c
CLEAN Water Analysis Solutions
www.cleaninst.com
Series Tester
30
PH30 pH Tester ORP30 ORP Tester CON30 Conductivity/TDS/Salinity Tester DO30 Dissolved Oxygen Tester FCL30 Free Chlorine Tester DOZ30 Dissolved ozone Tester
MCON tester body CS3930 electrode
1.5VAAA battery
When ERR occurs, you can check the electrode and the standard solution.
11. Error
Authorization must be obtained from CLEAN Instruments Customer Service Center to issue a RIR number before returning items for any reason. When applying for authorization, please nclude date requiring the reason of return. Instruments must be carefully packed to prevent damage in shipment and insured against possible damage or loss. CLEAN Instruments will not be responsible for any damage resulting from careless or insufficient packing. Warning: Damage as a result of inadequate packaging is the User / distributor’s responsibility. Please follow the guidelines below before transporting.
Use the original packaging material if possible, when transporting back the unit for repair. Otherwise wrap it with bubble pack and use a corrugated box for better protection. Include a brief description of any faults suspected for the convenience of Customer Service Center, if possible . If there are any questions, feel free to contact our Customer Service Center or distributors.
Warranty
Return Of Malfunction Instruments
Guidelines Or Returning Unit For Repair
CLEAN Instruments warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and has not been the result of abuse or misuse within the warranty period, please return to CLEAN Instruments and amendment will be made without any charge. CLEAN Instruments Customer Service Center will determine if product problem is due to deviations or customer abuse. Out of warranty products will be repaired on a charge basis.
Range
Resolution
Aaccuracy
Temp. Range
Working Temperature
Auto Temperature Compensation
Calibration
TDS Range
TDS Coefficient
Salinity Range
Salinity Coefficient
Atmospheric Pressure Compensation
Auto backlight off
Auto power off
Power
Dimensions
Weight
Protection
0.0 S/cm (ppm) - 20.00 mS/cm (ppt)
0.1 S/cm (ppm) - 0.01 mS/cm (ppt)
±2% F.S.
0 - 100.0 °C / 32 - 212 °F
0 - 60.0 °C / 32 - 140 °F
0 - 60ºC
1 or 2 point auto calibrate
(0% zero oxygen or 100% in air)
0.0 mg/L (ppm) - 20.00 g/L (ppt)
0.40 - 1.00 Adjustable (Coefficient:0.5)
0.0 mg/L (ppm) - 13.00 g/L (ppt)
0.65
600 - 1100 mbar
1 minutes
5 minutes without key be pressed
1x1.5V AAA battery
(H×W×D) 185×40×48 mm
95g
IP67
ENT
CAL
SET
ENT
CAL
SET
ENT
CAL
SET
ENT
CAL
SET
ENT
CAL
SET
ENT
CAL
SET
ENT
-5-
P01: Unit of the temperature: You can select between °C and °F.
P02: Temperature base: Range:15-35 °C
P03: Temperature compensation factor: Range:0.0-10.0%
P04: TDS factor
P05: Salinity factor
P06: Auto Lock. You can select “ON” or “OFF” to auto lock the data once stabilized.
P07: Factory defaults. You can select to revert to factory defaults.
Clean the electrode with pure water. Dip the electrode into the standard solution, the
stabilized icon will be displayed at the top of the screen. Press” ” to confirm. The
standard value is at the bottom. You can adjust the value as standard by pressing
”or ” ”. Press” ” to confirm. The slope rate of the calibration will be
displayed and automatically go back to the measurement mode. You can also press
”to skip the calibration and back to the measurement mode.
Long press“ “ to check through the calibrated parameters, Press“ “ and”
to check the Zero offset and the Slope.
ENT
ENT
CAL
CAL
CAL
SET
SET
CAL
SET
Loading...