Clean PH200, PH500 Operation Manual

CLEAN PH200/ PH500 pH/ mV/ ORP/ Ion
Versatile Water Analysis Meter
CLEAN INSTRUMENTS CO. LTD
2F No.140, Zhongxiao St., Zhonghe Dist., New Taipei City ,Taiwan
www.cleaninst.com
E-mail: CS@cleaninst.com
MODE
ESC
CAL INFO
MI
SET
ENT
HOLD
MR
PH200
MODE
ESC
CAL
INFO
MI
SET
ENT
HOLD
MR
PH500
pH/ mV/ ORP/ Ion/ Temp
CLEAN PH200/PH500
Versatile Water Analysis Meter
TABLE OF CONTENTS
01 02 03 04 07 08 08 09 09 09 10 11 12 12 13 13 14 15 15 15 16 16 17 19 20 21 22 23
1 Preface 2 Basic Operation 3 Key Panel 4 Appearance
5.1 Technical Specifications
5.2 Model and Function
5.3 Browse the parameter
5.4 Set UP
5.4.1 pH set up
P01 pH Standard Solution
P02 Manual Temperature setting
P03 Temperature Offset setting
P04 Auto Lock Function setting
P05 Prompt tone setting (optional)
P06 Auto Power Off Setting
P07 Clearing Memory Function
P08 Reverting to Factory Default Setting
5.4.2 ORP set up
5.4.3 Ion set up
P01 Measuring unit setting
P02 Calibration point setting
P03 Ionic valence setting
5.5 Calibration Mode
5.6 Measurement Mode
5.7 Storage of measuring Data
5.8 Browse on Setting Parameters
5.9 Appendix GENERAL INFORMATION
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
01
02
BNC Temperature
+
-
M
SET
ENT
HOLD
MR
PH200
+
-
03
04
1 Battery power
2 S- Stable. If twinkling, means unstable.
3 CAL – Calibration
4. SET- Setup display
5 HOLD- Auto lock
6 °C °F - Temperature
7 ATC - Automatic Temperature Compensation
8 PHmVORPIon ppmmg/L- Measurement modeUnit of Measurement
9 STD - Standard solutions
10 mV - Zero driftmV indicator or Ion sensor slope
11 % - Uncalibrated meter is 0% and “%”twinkling. When sensor slope is 65%, “%”twinkling
Display
4. Appearance
①②③④ ⑤
⑥ ⑦
05
Sensor
Manual Temperature setting
Temperature offset setting
Automatic Temp. Compensation
Prompt tone setting
Electrode slope
Auto lock
Auto off
no stored data
Clear data
Offset
Menu item
Buffer Solution
Buffer kinds
Reset to factory defaults
Save Data
On
Off
No
Yes
Temperature value Over
Temperature value Under
Full Data Storage
Measuring Value Over
Measuring Value Under
Battery
Error
Ion concentration unit selection
Ion calibration Points
Ionic valence
10 of the fourth power: 10000
10 of the fifth power: 100000
Display Character Table
ATC
06
·
·
·
·
·
·
·
·
·
·
·
·
·
·
·
·
One key to switch among pH, mV, ORP, Ion measuring modes.
pH value, mV value, Temperature value with screen display simultaneously, humanized design.
°C °F optional
Four sets with 11 points buffer combination, covering global standards including US, EU, CN, JP.
Two points ORP calibration
Ion concentration measuring range: 0.000-99999mg/L
Large LCD, with high luminance LED backlight. One key to auto calibration: Zero offsetElectrode offsetensure the accuracy
One key to browse the setting parameters, including Electrode zero offset, sensor slope and all
set parameters.
Prompt tone after measurement readings stable(Optional)
HOLD Auto Lock function
Temperature offset adjust
200 sets of measuring data storage
Auto power off if no operations in 10 minutes. (Optional)
2*1.5V 7AAA batteries. Long battery life span
CLEAN PH200/PH500 series products with precise and practical design concept: simple operation,
powerful functions, complete measuring parameters, wide measurement range. Four sets with 11
points buffer combination, one key to calibrate and auto buffer recognition. Clear and easy
understandable operation interface, excellent anti-interference, precise measuring, convenient
to operate, combined with the high luminance LED backlight, CLEAN PH200/PH500 is your
professional choice, a reliable partner in Lab, factory, and routine measuring in school.
CLEAN PH200 / PH500 pH/ mV/ ORP/ Ion/ Temp
Versatile Water Analysis Meter
07
Range
Resolution
Accuracy
Range
Resolution
Accuracy
Range
Resolution
Accuracy
Range
Resolution
Accuracy
Power supply
B1
B2
B3
B4
Environment
Memory
Dimensions
Weight
-2.00 ~ 20.00 pH
0.01 pH
±0.01 pH
-2000~2000 mV
0.1mV ( -999 999mV ), 1 mV ( <-999 mV or >999 mV )
± 1 mV
0.000-99999mg/L, ppm
0.001,0.01,0.1,1 mg/L, ppm
1%+1LSD(1 valence, 2%+1LSD(2 valence), 3%+1LSD(3 valence)
-40.0-125.0℃(-40.0-257.0°F
0.1°C, 0.1°F
+0.2°C, 0.1°F
2*1.5V 7AAA batteries
1.68, 4.01, 7.00, 10.01 (US)
2.00, 4.01, 7.00, 9.21, 11.00 (EU)
1.68, 4.00, 6.86, 9.18, 12.46 (CN)
1.68,4.01, 6.86, 9.18 (JP)
-5 °C60 °CRelative humidity﹤90%
200 sets
94*190*35mm (W*L*H)
250g
pH
ORP
Ion
Temperat
ure
Battery
pH Buffer
Kinds
Others
5.1 Technical Specifications
Transformer 100 ~ 220V
140*210*35mm (W*L*H)
650g
PH500PH200
08
5.3 Browse the parameter
Long press for 3 seconds in measurement mode.Then press to browse the parameters
1. Zero offset ( )
2. Slope ( )
3. Buffer kinds
4. Temperature compensation
5. Prompt tone(ON/OFF)
6. Hold Auto Lock
7. ON/OFF Auto power off
8. Battery power
5.2 Model and Function
pH
Buffer kinds
Manual temperature
Temperature offset
Auto Lock
Prompt tone
Auto power off
Clear the saved data
Revert to factory default
P-01
P-02
P-03
P-04
P-05
P-06
P-07
P-08
Ion
Switch measuring unit
Calibration point
Ionic valence setting
---------
ORP
---------
---------
---------
---------
09
P-01 pH Standard Solution
After power-up, the unit will enter the measurement mode. Factory default is pH measuring
mode. You can switch between the pH measurement mode and the ORP measurement mode
by pressing the Key. Press to enter set up step. Press to enter electrode
calibration mode.
In P-01, press to enter. Press to select among B1,B2,B3,B4 buffer kinds.
Press to confirm.
Please refer to above description steps to set up P-01.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
B1: 1.68, 4.01, 7.00, 10.01 (US)
B2: 2.00, 4.01, 7.00, 9.21, 11.00 (EU)
B3: 1.68, 4.00, 6.86, 9.18, 12.46 (CN)
B4: 1.68,4.01, 6.86, 9.18 (JP)
5.4 Set UP
5.4.1 pH set up
P-02 Manual Temperature setting
After entering P-02, you can set the manual temperature by pressing . Long press
for speedy setting. Confirm your setting by pressing .
You can switch between °C and °F by pressing . Press to confirm. This setting will only
take effect after the meter restarted.
Manual Temperature setting range: -40.0-125.0℃(-40.0-257.0°F)
Please refer to above description steps to set up P-02.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
Note:
1. Please make sure the Temperature Electrode is UNPLUGGED when manual Temperature
compensation. Otherwise, the unit will display “Err” and go back to the set up menu for fresh
operation.
2. It is not working to unplug the Temperature Electrode while setting. You have to set again after
removing the electrode and quit the current mode.
10
P-03 Temperature Offset setting
After entering P-03, you can set the temperature to the appropriate value by pressing
according to the actual temperature. Long press for speedy setting. Confirm your setting
by pressing .
Temperature Offset range: ±10°C.
Please refer to above description steps to set up P-03.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
Note:
1. Please make sure the Temperature Electrode is PLUGGED when Temperature Offset setting.
Otherwise, the unit will display “Err” and go back to the set up menu for fresh operation. The
ATC” will be displayed beneath the main screen the screen normally.
2. It is not working to plug the Temperature Electrode while setting. You have to set again after
plugging in the electrode and quit the current mode.
11
After entering P-04, you can select ON or OFF the Auto Lock Function by pressing . If the
Auto Lock Function is ON, the meter can lock the measured value after the reading had stabilized
and display . Press to unlock.
Please refer to above description steps to set up P-04.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
P-04 Auto Lock Function setting
12
P-05 Prompt tone setting (optional)
After entering P-05, you can select ON or OFF the prompt tone by pressing . If the
prompt tone is ON, you will hear the “Beep” tone after the reading had stabilized when in
calibration.
Please refer to above description steps to set up P-05.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
After entering P-07, you can select NO or YES to not clear or clear the data by pressing .
If select YES, the meter will clear all the saved data.
Please refer to above description steps to set up P-06.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
After entering P-06, you can select ON or OFF the Auto Power Off by pressing . If the
Auto Power Off is ON, the meter will power off if no operation in 10mins.
Please refer to above description steps to set up P-06.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
P-07 Clearing Memory Function
P-06 Auto Power Off Setting
13
P-08 Reverting to Factory Default Setting
14
After entering P-08, you can select NO or YES to not reverting or reverting to factory settings by
pressing . If select YES, the meter will revert all the settings to factory defaults and all
the settings will lost forever. The unit will restart at the same time.
Please refer to above description steps to set up P-08.
You can go to next parameter setting by pressing , or pressing key to quit and
go back to measurement mode.
Note:
1. During setting, you can press key to quit setting whenever necessary.
2. During setting, you can adjust anything twinkling by pressing . If it is the data
twinkling, you can speedy adjust the data by pressing .
Below is only the whole meter setting. Please refer to pH setting for detailed set up steps.
15
5.4.2 ORP set up
5.4.3 Ion set up
P-01 Measuring unit setting
After entering P-01, you can switch between measuring unit mg/L and ppm by pressing .
Confirm your setting by pressing .
Please refer to above description steps to set up P-01.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
P-04 Auto Lock Function setting
P-05 Prompt tone setting (optional)
P-06 Auto Power Off Setting
P-07 Reverting to Factory Default Setting
After power-up, the unit will enter the measurement mode. Factory default is pH measuring
mode. You can switch among the pH measurement mode, ORP measurement mode and the
Ion measuring mode by pressing the Key. In Ion measuring mode, press to enter set
up step. Press to enter calibration step.
After entering P-02, you can select 2 points, 3 points, 4 points, 5 points, 6 points, 7 points for
calibration by pressing . Confirm your setting by pressing .
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
Please refer to above description steps to set up P-02.
After entering P-03, you can select 1 valence,2 valence, 3 valence by pressing . Confirm
your setting by pressing .
Please refer to above description steps to set up P-03.
You can go to next parameter setting by pressing , or pressing key to quit and go
back to measurement mode.
Below is only the whole meter setting. Please refer to pH setting for detailed set up steps.
P-03 Ionic valence setting
16
P-02 Calibration point setting
P-04 Auto Lock Function setting
P-05 Prompt tone setting (optional)
P-06 Auto Power Off Setting
P-07 Reverting to Factory Default Setting
1. Dip the sensor into the already set buffer solution when in pH calibration mode. For details,
please refer to the standard solution setting.
2. Press key to enter pH calibration step. “ ” will be displayed. Dip the sensor into buffer
solution(7.00 or 6.86 pH). When the value is stable and display “ ” together with the twinkling
stops,indicates the first zero point calibration is completed. Screen blanks for 2 seconds for saving
the data. Then the meter will automatically go to second point calibration. Dip the sensor into the
second buffer solution. When the value is stable and display “ ” together with the twinkling stops,
indicates the second point calibration is completed. You can go next to third point calibration or
fourth point calibration. You can also press key go back to measurement mode. (To check
the calibration results: long press for 3 seconds)
Note: During the calibration process, you can also press key to terminate calibration process
and back to measurement mode. Calibrated points are already saved.
If “ Err ” shows when in calibration, could be the following reasons:
1.Wrong buffer solution. First point defaults as Zero point calibration. Please follow the Auto Buffer recognition order.
2.Poor performance of electrode.
3. Stabilized time is more than 30 seconds.
4. After calibration, in measurement mode, if”%” is continuous twinkling, means the sensor slope is below 65%. Please timely maintain the electrode or change electrode. Then re-calibrate.
17
pH calibration
5.5 Calibration Mode
change standard solution
change standard solution
Press key to enter ORP calibration step. “ ” will be displayed.
Dip the sensor into any standard solution which the user already known the mV value in it. When the
value is stable and display “ ”, press keys to adjust the four-digit STD number at the
bottom row. So the value can be the same as the already known mV value. Press to confirm.
First point calibration is completed. You can go to second point calibration or press key to quit
and go back to measurement mode.
Press key in Ion measuring mode to enter the calibration step. “ ” will be displayed.
Dip the sensor into any standard solution which the user already known the concentration value in it.
When the value is stable and display “ ”, press keys to adjust the corresponding
concentration value at the bottom row(0.10, 1,10, 100, 1000, 10
4
, 105mg/L). Press to confirm.
First point calibration is completed. Go to second point calibration. Change the standard solution.
When the value is stable and display “ ”, press keys to adjust the corresponding
concentration value at the bottom row(0.10, 1,10, 100, 1000, 104 , 105mg/L). Press to confirm.
Second point calibration is completed (Factory default is 2 points calibration). Meter will go back
measurement mode after calibration. (To check the calibration results: long press for 3 seconds)
Measuring range: 0.000 mg/L, ppm ~ 99999 mg/L, ppm
Note: Five-element number displayed on screen. The lower in value, the higher in resolution.
Maximum resolution is 3-digit after the decimal point.
Calibration order from low concentration value to high concentration. Otherwise screen will display ERR.
18
ORP calibration
Ion calibration
19
1. You can first set the meter per your request. Otherwise, the setting will be factory defaults.
2. Use tap water or distilled water to clean the pH and the temperature electrode, removing the
adhesive impurities on the surface of electrode.
3. After power on, press to select the operation mode (pH, mV or Ion).
4. Immerse the pH and the temperature electrode in the solution and stir gently. The measuring
mark will twinkle. You can read the value once the data is stabilized and shows the stable
mark . When sensor slope is below 65%, slope will twinkle, please timely maintain the electrode
or change electrode.
1. When the screen shows “ATC” means the temperature sensor is working and shows the actual
temperature upper right corner. You can refer to the P-02 “Temperature Offset setting” if a revision
on the temperature is required.
2.If “ATC” is not shown on the screen means the temperature electrode is disconnected
and is manual temperature compensation. Factory default is 25°C. You can refer to “
Manual temperature setting” to adjust the temperature.
3. The measuring value is stabilized when the screen showed , The is showed if the Auto
Lock Function is on. Press to unlock.
5.6 Measurement Mode
Measuring mode switch [ pH/mV, ORP, Ion, Temp ]
After power-up, the unit will enter the measurement mode. Factory default is pH measuring
mode. You can switch between the pH measurement mode, ORP measurement mode and the
Ion measurement mode by pressing the Key.
Factory default no reading value for Ion measurement mode. Need to first calibrate and then
reading value displayed.
20
Storage of measuring data
Browse on data storage
1.If you need to save the measuring data once it is stabilized, press , the Lot no. will be shown
on the upper right of the screen. Press to save the data. You can also press to not save.
The meter will automatically go back to measurement mode after the data saved. The unit
at most can save 200 sets data. The screen will show “OVER” if more than 200 sets reminds you the
memory card is full.
2.Enter into setting mode P-07 if you need to delete the saved data forever. You can refer to
Clearing memory function
1. Press to enter into the browse on the data storage , the Lot no. will be shown and data
twinkling. Press to check all the saved data. Press to go back to measuring mode.
If showed “NONE” means no data is saved yet.
2. If you quit and re-entering during checking the data, the meter will freeze in the last set of data.
5.7 Storage of measuring data
pH setting parameters browse
Ion setting parameters browse
1. Long press to enter into Setting parameter when in measurement mode. Press
to check through all the Calibration parameters and Setting parameter.
2. Press to quit and go back to measurement mode.
5.8 Browse on Setting Parameters
21
Sensor slope and Zero offset
Ion sensor slope
22
pH Factory defaults
Corresponding interface
P-01
P-02
P-04
P-05
P-06
Item
Standard Solution
Temperature
Auto Lock
Prompt tone setting
Auto power off
Backlight
Factory Defaults
B1: 1.68, 4.01,7.00,10.01 (US)
Manual Temperature Compensation 25 °C
OFF
OFF (No configuration)
ON
OFF
5.9 APPENDIX
23
CLEAN Instruments warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and has not been the result of abuse or misuse within the warranty period, please return to CLEAN Instruments and amendment will be made without any charge. CLEAN Instruments Customer Service Center will determine if product problem is due to deviations or customer abuse. Out of warranty products will be repaired on a charge basis.
Authorization must be obtained from CLEAN Instruments Customer Service Center to issue a RIR number before returning items for any reason. When applying for authorization, please nclude date requiring the reason of return. Instruments must be carefully packed to prevent damage in shipment and insured against possible damage or loss. CLEAN Instruments will not be responsible for any damage resulting from careless or insufficient packing. Warning: Damage as a result of inadequate packaging is the User / distributor’s responsibility. Please follow the guidelines below before transporting.
Use the original packaging materialif possible, when transporting back the unit for repair. Otherwise wrap it with bubble pack and use a corrugated box for better protection. Include a brief description of any faults suspected for the convenience of Customer Service Center, if possible.If there are any questions, feel free to contact our Customer Service Center or distributors.
GENERAL INFORMATION
Warranty
Return Of Malfunction Instruments
Guidelines Or Returning Unit For Repair