Civco GUS G32-S, GUS G32-E, GUS Disinfection Soak Station Operator's Manual

CIVCO Medical Solutions
102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482 Website: WWW.CIVCO.COM
Model G32-S Model G32-E
Disinfection Soak Stations
Operator’s Manual
®
is a registered trademark of CIVCO Medical Solutions.
GUS
All other trademarks are property of their respective owners.
Copyright © 2018 All Rights Reserved.
Printed in the USA.
GUS Disinfection Soak Station™
Table of Contents
Principle of Operation ........................................................................................ 2
Safety Precautions .............................................................................................. 2
Warning Symbols ................................................................................................ 2
Signal Words ....................................................................................................... 2
General Information ............................................................................................ 3
Installation and Setup .................................................................................................................. 4
Upon Delivery .............................................................................................................................. 4
Installation ................................................................................................................................... 4
Filling the High-Level Disinfection Tray(s)................................................................................... 5
Operation ............................................................................................................. 6
Placing Instruments into the System ........................................................................................... 6
Removing and Rinsing Instruments ............................................................................................ 6
Disposal of Disinfectant ..................................................................................... 6
Built-in Disposal System ............................................................................................................. 7
Filter Life .............................................................................................................. 8
Filter Replacement ...................................................................................................................... 8
Cleaning the System ........................................................................................... 9
Rinse Water ................................................................................................................................. 9
Limited Warranty ............................................................................................... 10
Registration Card .............................................................................................. 10
APPENDIX A: TROUBLESHOOTING ............................................................... 11
APPENDIX B: SITE REQUIREMENTS .............................................................. 12
APPENDIX C: GUS PRODUCT SPECIFI CATIONS .......................................... 13
UL Approval ............................................................................................................................... 13
M0036_F 1 G32-S / G32-E
GUS Disinfection Soak Station™
Principle of Operation
The GUS Disinfection Soak Station provides for the safe soaking of heat sensitive instruments in high-level disinfectant solution. The system’s blower forces air through a carbon filter which removes toxic vapors from the high-level disinfectants.
Do not use the system for any purpose other than its intended use.
Safety Precautions
Disconnect power before replacing filter or when servicing the unit.
Refer to the SDS of the high-level disinfectant that you are using for additional safety precautions.
Warning Symbols
The meaning of the labels and symbols that appear on the packaging and/or the system are as follows:
Signal Words
The following signal words are used throughout this manual:
WARNING Indicates a potentially hazardous situation which, if not avoided, could result in
NOTE Indicates additional helpful information.
Dangerous electrical voltage
Hazardous area
Protective Grounding
death or serious injury.
CAUTION Indicates a potentially hazardous situation which, if not avoided, may
result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
M0036_F 2 G32-S / G32-E
GUS Disinfection Soak Station™
Figure 1: GUS Model G32-E with Tray(s)
Top louver Acrylic door
Main Power
Soak Tray(s)
Built-In Disposal System
External
Rinse Tray(s)
General Information
The GUS Disinfection Soak Station eliminates vapors from high-level disinfectants and contains splashes and spills. Blowers pull fumes into a patented carbon filter, where they are adsorbed and eliminated. Clean air is returned to the room through the top louver panel. These systems are available as Standard Countertop models (see figure1), or as Space Saver models which are designed to fit under an upper wall cabinet.
The GUS
Glutaraldehyde
Ortho-phthalaldehyde (OPA)
Optional filters are available for:
Peracetic Acid
Formaldehyde
WARNING Failure to properly follow the instructions in this manual may result in damage
On/Off Switch and Suction
Hose Connection*
System with the standard filter is designed to remove fumes from:
Hydrogen Peroxide
Hydrogen Peroxide
to the system and/or scopes.
On/Off Switch
Fuse
panel
M0036_F 3 G32-S / G32-E
GUS Disinfection Soak Station™
Installation and Setup
Upon Delivery
Match all items in the package with those listed on your packing list. Inspect each item for damage. Fill tray(s) (if provided) with water and check for leakage.
If any components are damaged or missing, or if you have any questions, call CIVCO Medical Solutions at 1-800-445-6741. Do not use the system.
Installation
Countertop models
1. Place the white polypropylene base on the countertop with the grooves facing up.
Note that the grooves should be on the sides and back.
2. Gently lift the system by the two side straps and place into the grooves on the base.
There must be a clearance of 1” above the unit.
3. Plug the system into a standard wall outlet.
4. Remove the protective covering from the acrylic doors.
5. Open the acrylic doors and place the soak and rinse tray(s)s inside.
Mobile models (Assembly of Cart Required)
1. Remove the cart from the top of the box and place onto a level surface. Assemble as
directed.
2. Gently lift the system out of the packing box and, with someone holding the cart, place the
system on top.
CAUTION It may require two people to lift the system and place it onto the cart.
3. Plug the system into a standard wall outlet.
4. Remove the protective covering from the acrylic doors.
5. Open the acrylic doors and place the soak and rinse tray(s)s inside.
WARNING Extreme caution should be taken if moving the mobile system while it
contains high-level disinfectant.
M0036_F 4 G32-S / G32-E