Circutor MPC-20, MPC-50 Instruction Manual

Page 1
STEP AND TOUCH VOLTAGE TESTER
TESTER
MPC-20 (20 A) code 7 70 123 MPC-50 (50 A) code 7 70 075
INSTRUCTION MANUAL
( M98113601-20 / 03A )
(c)
CIRCUTOR S.A.
Page 2
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
2
1.- MPC-20 - MPC-50 .................................................................... 4
1.1 Usual operating procedure 4
1.2 Precautions whilst operating 5
2.- DESCRIPTION OF THE MAIN COMPONENTS ...................................... 6
2.1 Source of regulated voltage 6
2.2 Control and protection circuit 6
2.3 Voltage and current tester 6
2.4 Touch electrodes 6
2.5 Accessories 6
3.- CONTROL AND CONNECTION FEATURES OF THE MPC-20 / MPC-50 .... 7
3.1 Control Components 7
3.2
External connection components 7
3.3
Internal connection components 8
4.- MAIN FEATURES ............................................................................. 10
4.1.- Power supply 10
4.2.- Dimensions, weight:
5.1.1 Attaching the equipment 14
4.3.1 Features as a step and touch voltage tester. MPC-20 TESTER 11
4.3.2 Features as a step and touch voltage tester. MPC-50 TESTER 12
4.4.- Features as an ground resistance tester 13
5.- TESTING STEP OR CONTACT VOLTAGE............................................ 14
5.1 Installation and start-up 14
5.2 Attaching the MPC-20 / MPC-50 14
5.3 Auxiliary ground for current injection 14
5.4 Connecting the current circuit 16
5.5
Connecting the power supply 16
5.6
Setting the maximum default current 17
5.7
Attaching the testing electrodes 18
5.8
Button for testing erratic currents 18
5.9 Adjusting the current and testing the
STEP or TOUCH VOLTAGE 19
5.10
Test results 20
6.- TESTING GROUND RESISTANCE ..................................................... 21
6.1 Attaching and connecting the MPC-20 / MPC-50 22
6.2 Attaching the "MEASURING ROD" 22
6.3 Testing erratic voltages 22
6.4 Testing ground resistance 22
6.5 Test results 23
6.6 Repeating tests 24
Page 3
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
3
CHAPTER II
7.- EXISTING STANDARDS ................................................................... 25
7.1 Abstract and table of voltages 25
7.2 Standards relating to the MPC-20 / MPC-50 26
8.- CORRECTION OF THE READ OUTS .......................................................... 27
9.- MAINTENANCE............................................................................... 29
10.- TECHNICAL SERVICE ..................................................................... 29
Page 4
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
4
MPC-20 & MPC-50
1.- PRELIMINARY WARNINGS:
1.1.- USUAL OPERATING PROCEDURE
The MPC-20 and MPC-50 testers are compact and robust and are
designed to test where large powers are needed for a short period of time:
- MPC-20: up to 12 kVA
- MPC-50: up to 20 kVA
Its microprocessor controlled testing system ensures that tests are made
in approximately 10 seconds.
Its portability means that weight has been kept to a minimum with the
use of minimal loss toroidal transformers.
Heat generated by the transformer windings is continuously eliminated by an in-built ventilation system. It is able to operate indefinitely, making 12 sec. tests at 40 sec. intervals. If for any reason it overheats, a red indicator light will show why and will prevent it from operating until normal temperature is re­established.
Page 5
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
5
1.2.- WARNINGS!
THE CURRENT USED FOR MEASUREMENTS, 20 A with the MPC­20 and 50 A in the MPC-50, IS SUFFICIENT TO CREATE SIGNIFICANT VOLTAGES AROUND THE ELECTRODES, ESPECIALLY THE AUXILIARY ELECTRODES.
PLEASE TAKE ALL NECESSARY PRECAUTIONS TO AVOID
ACCIDENTS.
For example: A 5 Ohms auxiliary electrode creates a voltage of 250 V around itself.
VOLTAGE MAY REACH 600 V AROUND THE CABLE OUTPUT with the MPC-20 and 400 V with the MPC-50.
THE POWER CIRCUIT CONNECTIONS MUST NEVER BE HANDLED WITHOUT HAVING FIRST STOPPED THE EQUIPMENT AND REMOVED THE OPERATING KEY.
BEFORE ANY MEASUREMENTS ARE TAKEN, SIGNS MUST PREVENT PEOPLE FROM PASSING THROUGH OR REMAINING IN THE INJECTED CURRENT AREA. IN THE EVENT OF HIGH GROUND RESISTANCES, VOLTAGES UP TO 600 V MAY BE CREATED, WHICH ARE FATAL TO ANIMALS OR PERSONS NEAR TO THE DISPERSERS DURING TESTS.
Page 6
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
6
2.- DESCRIPTION OF THE MAIN COMPONENTS
2.1.- VOLTAGE SOURCE
MPC-20: voltage source, with 10 regulating scales using a switch and
fine adjustment using a dial. Supplies a regulable voltage between 0 and 600 V and is set for a current of 20 A. It can test to 20 A in installations with a 30 Ohms resistance (ground and auxiliary added).
MPC-50: voltage source, with 8 regulating scales using a switch and fine
adjustment using a dial. Supplies a regulable voltage between 0 and 400 V and is set for a current of 50 A. It can test to 50 A in installations with a 8 Ohms resistance (ground and auxiliary added).
The remaining components are the same in each:
2.2.- CONTROL AND PROTECTION CIRCUIT
Button for testing erratic currents. Two directional selector key to control the current injection (0º and 180º): Line voltage indicator. Over heating indicator. Fuse protection for the operating circuit Protective circuit breakers for all the equipment and for the output circuit.
2.3.- VOLTAGE AND CURRENT TESTER, controlled by micro-processor. It
test and also calculates Vp, Vc and Rt. The alphanumeric liquid crystal display shows preset data, measurements and calculations.
2.4.- TOUCH ELECTRODES, that are cylindrical, 200 square cm at the base
and weigh 250 N, with carrying handles and terminals for connecting the testing cables.
2.5.- ACCESSORIES. 50 metre cables, 10 mm in diameter with terminals for
current injection. 50 metre cables for voltage reading. Bridging cable to join the two touch electrodes.
Page 7
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
7
3.- CONTROL AND CONNECTION COMPONENTS OF THE MPC-20 & MPC-50
3.1.- CONTROL ELEMENTS
1 VOLTAGE OUTPUT SWITCH MPC-20: FROM 1 TO 10 MPC-50: FROM 1 TO 8
Switch
Position
MPC-20 MPC-50
1 0 to 60 V 0 to 50 V 2 0 to 120 V 50 to 100 V 3 20 to 180 V 100 to 150 V 4 180 to 240 V 150 to 200 V 5 240 to 300 V 200 to 250 V 6 300 to 360 V 250 to 300 V 7 360 to 420 V 300 to 350 V 8 420 to 480 V 350 to 400 V 9 480 to 540 V 10 540 to 600 V
2 FINE REGULATION CONTROL (according to previous table)
3 RED TEMPERATURE INDICATOR
4 RED SYSTEM VOLTAGE INDICATING LIGHT
5 ANALOGUE AMMETER FOR ADJUSTMENT AND ASSISTANCE
6 GENERAL AND OUTPUT CIRCUIT BREAKER
7 SELECTOR WITH KEY OPERATE THE TEST
8 LIQUID QUARTZ DISPLAY
9 3 BUTTONS TO PRESET Imax VALUE
10 BUTTON TO SELECT OPERATING MODE (Vp/c or Rt)
11 BUTTON FOR ERRATIC VOLTAGE MEASUREMENTS
999
999999
999
ENTER
ENTERENTER
ENTER MODE
MODEMODE
MODE
3.2.- EXTERNAL CONNECTION COMPONENTS
12 OUTPUT TERMINALS: MAX 50 A, MAX 600 V
13 VOLTAGE INPUT TERMINALS TO THE VOLTMETER
Page 8
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
8
3.3.- INTERNAL CONNECTION COMPONENTS
MPC-20:
30 AND 31 SINGLE PHASE POWER SUPPLY: 230 V T.T TERMINAL WITH OBLIGATORY EARTHING CONNECTION
MPC-50:
30 AND 32 SINGLE PHASE POWER SUPPLY: 400 V 30 AND 31 SINGLE PHASE POWER SUPPLY: 230 V (EXCEPTIONAL) T.T TERMINAL WITH OBLIGATORY EARTHING CONNECTION
SIZE:
Page 9
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
9
Control panel:
999-> 999 ENTER ESC
RT-VPC
0º OFF 180º
(LLAVE)
ERRATICAS
ENTRADA TENSION
Vm
PILOTO RED
PILOTO TERMICO
CONMUTADOR DE SALIDAS 1 AL 8
1
2
0
50
A
MPERIMETRO
INT. GENERAL
DISPLAY
VOLTAGE INPUT
CIRCUIT BREAKER LIGHT
SYSTEM LIGHT
GENERAL SWITCH
AMMETER
OUTPUT 1 TO 8 SWITCH
ERRATICS
(KEY)
Page 10
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
10
4.- MAIN FEATURES
4.1.- POWER SUPPLY
POWER SUPPLY MPC-20 MPC-50
VOLTAGE
FREQUENCY
MAXIMUM CURRENT
CIRCUIT BREAKER
POWER SUPPLY CABLE
MAXIMUM POWER
230 ± 10 % SINGLE PHASE
50 / 60 Hz
54 A
100 A 2 X 10 + 6 T.T. 12 kVA
400 ± 10 % SINGLE PHASE 230 V exception 50 / 60 Hz
57 A to 400 V (95 A to 230 V)
100 A 2 x 16 + 16 T.T. 20 kVA
4.2.- SIZES AND WEIGHT
SIZES: HEIGHT 800 mm WIDTH 550 mm LENGTH 500 mm
WEIGHT: 142 Kg WITHOUT ELECTRODES
ELECTRODE SIZES: HEIGHT 210 mm APPROX. DIAMETER 160 mm APPROX.
WEIGHT: 25.5 Kg EACH
Page 11
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
11
4.3.- TESTING STEP AND CONTACT VOLTAGE
4.3.1.- MPC-20 TESTER
CURRENT 0 ÷ 20 A REGULABLE OVERVOLTAGE 1, 1 In, 1 Min VOLTAGE 0 ÷ 600 V,
10 SCALES AND FINE ADJUSTMENT
MAX. RESISTANCE 30 OHMS (GROUND TESTED + AUXILIARY
GROUND)
RESISTANCE OF HIGHER VALUES
ALLOWS TESTING AT LOWER CURRENTS
METHOD USED INJECTION AND TESTING OF CURRENT
(Ip)
TESTING VOLTAGE IN SURFACE AREA
(Vm)
DEFAULT SETTING Imax CALCULATES: Vp/c=Vm xMax/Ip
ANALOGUE AMMETER  CLASS: 1.5
RANGE: 20 A
DIGITAL AMMETER CONTROLLED BY MICROPROCESSOR
ACCURACY: 0.5 % MEASUREMENTS ± 2
digits
MAX. : 99.9 A RESOLUTION : 10 mA MINIMUM READING : 20 mA
DIGITAL VOLTMETER CONTROLLED BY MICROPROCESSOR
SCALES : 30 V and 600 V SCALE CHANGE : AUTOMATIC ACCURACY: 0.5 % MEASUREMENTS ± 2
digits
MINIMUM READING : 20 mV
IMPEDANCE CUR. : 1000 ohms or 600
kohms (switch)
RESOLUTION : 10 mV
MAX. PROGRAMMABLE CURRENT VALUE
9999 A
SELECTION OPERATING MODE (Vp/c or Rt)
BY BUTTON (MODE) + Rt/Vp-c Selector
DEFAULT CURRENT PRESELECTION
BY BUTTONS
TESTING ERRATIC VOLTAGES
VOLTMETER CONTROL BY BUTTON
CONTROL OF STOP AND START
BY SELECTOR WITH KEY: ON 0°--- OFF --- ON 180º
Page 12
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
12
4.3.2.- TESTING STEP AND CONTACT VOLTAGE: MPC-50 TESTER
CURRENT 0 ÷ 50 A REGULABLE OVERVOLTAGE 1, 1 In, 1 Min VOLTAGE 0 ÷ 400 V,
8 SCALES AND FINE ADJUSTMENT
MAX. RESISTANCE 8 OHMS (GROUND TESTED + AUXILIARY
GROUND) RESISTANCE OF HIGHER VALUES
ALLOWS TESTING AT LOWER CURRENTS
METHOD USED INJECTION AND TESTING OF CURRENT
(Ip)
TESTING VOLTAGE IN SURFACE AREA
(Vm)
DEFAULT SETTING Imax CALCULATES: Vp/c=Vm x Max/Ip
ANALOGUE AMMETER CLASS : 1.5
RANGE : 50 A OVERRANGE : 300 A
DIGITAL AMMETER CONTROLLED BY MICROPROCESSOR
ACCURACY: 0.5 % MEASUREMENTS ± 2
digits
MAX. : 99.9 A  RESOLUTION : 10 mA  MINIMUM READING : 20 mA
DIGITAL VOLTMETER CONTROLLED BY MICROPROCESSOR
SCALES : 30 V and 500 V SCALE CHANGE : AUTOMATIC ACCURACY: 0.5 % MEASUREMENTS ± 2
digits
MINIMUM READING : 20 mV
IMPEDANCE CUR. : 1000 ohms or 600
kohms (switch)
RESOLUTION : 10 mV
MAX PROGRAMMABLE CURRENT VALUE
9999 A
SELECTION WORK MODE (Vp/c or Rt)
BY BUTTON (MODE) + Rt/Vp-c Selector
DEFAULT CURRENT PRESELECTION
BY BUTTONS
TESTING ERRATIC VOLTAGES
VOLTMETER CONTROL BY BUTTON
CONTROL OF STOP AND START
BY SELECTOR WITH KEY
ON 0°-- OFF --- ON 180º
4.4.- TESTING GROUND RESISTANCES
(USING CURRENTS UP TO 20 A or 50 A)
Page 13
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
13
METHOD USED INJECTION AND TESTING OF CURRENT
MPC-20 : 0 ÷ 20 A MPC-50 : 0 ÷ 50 A
TESTING VOLTAGE BETWEEN GROUND AND NEUTRAL AREA
(Vm)
CALCULATES: Rt.=Vm/Ip
ELECTRODES NEEDED GROUND TO MEASURE
DISTANT AUX ELECTRODE
"ROD" ELECTRODE FOR TAKING
VOLTAGE
MAXIMUM ERROR 3 % OF MEASUREMENTS ± 2 digits DIGITAL AMMETER CONTROLLED BY MICROPROCESSOR
ACCURACY: 0.5 % MEASUREMENTS ± 2
digits
MAX. : 99.9 A RESOLUTION : 10 mA MINIMUM READING : 20 mA
DIGITAL VOLTMETER CONTROLLED BY MICROPROCESSOR
RANGES : 30 V and 500 V RANGE CHANGE : AUTOMATIC ACCURACY: 0.5 % MEASUREMENTS ±
2 digits
MINIMUM READING : 20 mV IMPEDANCE CUR. : 600 kohms RESOLUTION : 10 mV
SELECTION WORK MODE (Vp/c or Rt)
BY BUTTON (MODE) + Rt/Vp-c Selector
TESTING ERRATIC VOLTAGES
VOLTMETER CONTROL BY BUTTON
CONTROL OF STOP AND START
BY SELECTOR WITH KEY
ON 0°-- OFF --- ON 180º
Page 14
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
14
5.- TESTING STEP OR CONTACT VOLTAGE
5.3.- INSTALLATION AND START-UP
This manual contains information and warnings that must be followed by the user to ensure the safe operation of the equipment and to maintain it in a safe condition.
Usual operation must not be started until the final assembly of the cables and weights, as recommended in section 1.2.- Connection instructions of this manual.
If the equipment is used in a way not specified by the manufacturer, the protection of the equipment and the people operating it may be compromised.
When it is likely that there has been a loss of safety (e.g. when damage is visible), the equipment must be disconnected from the power supply. In the event please contact a qualified service representative.
5.4.- ATTACHING THE MPC-20/ MPC-5
To test correctly, the MPC-20 or the MPC-50 must be placed 20 or more metres from the ground to be tested. Do not place it in the area occupied by the electrodes in the ground.
5.5.- AUXILIARY GROUND FOR CURRENT INJECTION
There must be auxiliary ground for current injecting 50 or more metres away from the ground to be tested Resistance between the auxiliary ground added to the ground to be measured must not exceed 8 ohms for the MPC-50 and 30 Ohms for the MPC-20.
-----------------------------------------------------------------
Page 15
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
15
DIAGRAM FOR TESTING STEP VOLTAGE
-----------------------------------------------------------------------------------------------------------
DIAGRAM FOR TESTING CONTACT VOLTAGE
GROUND
TO BE
TESTED
1 metre
Tested
voltage
L (minimum 20
metres)
Test intensity lp
AUXILIARY
GROUND
L (minimum 20 metres)
GROUND
TO BE
TESTED
Metal parts
Tested
voltage
L (minimum 20 metres)
Test intensity lp
AUXILIARY
GROUND
Page 16
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
16
5.6.- CONNECTING THE CURRENT CIRCUIT
ENSURE THAT THE SELECTOR (ON-OFF-ON) ON THE
MPC-20 & MPC-50 IS DISCONNECTED AND WITHOUT THE KEY.
Connect the "OUTPUT CABLES": to the ground to be measured and the other to the auxiliary ground.
5.7.- CONNECTING THE POWER SUPPLY
5.7.1.- POWER SUPPLY FOR THE MPC-20:
Connect the supply cable to a 230 V II+TT system or an independent generator with the same VOLTAGE.
5.7.2.- POWER SUPPLY FOR THE MPC-50:
Connect the supply cable to a 400 V II+TT system or an independent generator with the same VOLTAGE (As an exception the MPC-50 may be connected to 230 V).
Page 17
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
17
5.8.- SETTING THE MAXIMUM DEFAULT CURRENT
Once the connections described above have been made and after taking SAFETY MEASURES, testing can be done in the following way:
Place the regulator on the "0%" position.
Place the Voltage switch on position "1".
CONNECT THE MPC SWITCH. The model type and the version of the
setting will appear on the display.
Place the "OPERATING MODE" selector on "V p/c". (Connect the 1000 Ohms resistance switch located on the back of the equipment: voltmeter internal impedance of 1000 Ω).
PRESS the upper right "MODE" button (operating mode) and the display is ready to start testing voltage ("V") or resistance ("Rt"). Use the said button to select "V" on the display.
MAXIMUM CURRENT must be entered as preset in the installation procedure. The value in A of the maximum current must be entered as a testing constant. This value must be supplied by the electricity supply company according to MIE-RAT-19.
Using the 3 buttons above the display, (,  and Enter), SET THE
MAXIMUM DEFAULT CURRENT (maximum 9999):
a.- To enter setting, press () . The first lx digit will flash.
b.- To enter or change this value press ( ) repeatedly: the digit flashing at that time will increase.
c.- When the required value appears on the screen move on to the next digit by pressing (), and so on changing the remaining values.
d.- To enter press (Enter) button.
Page 18
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
18
5.9.- ATTACHING THE TESTING ELECTRODES
- Attach the touch electrodes (25 kg weights) in a suitable place to start
testing:
STEP VOLTAGE
Attach the two touch electrodes 1 metre apart and near to the ground to be tested. Each electrode are joined by a different cable from the voltmeter.
TOUCH VOLTAGE
Attach the two electrodes together 1 metre from the vertical of the tested object, both connected by a single cable from the voltmeter. The other voltmeter cable is connected to the tested object with a crocodile clip.
NUMBER OF TESTS
To ensure that there are no dangerous step or touch voltages, it is necessary to take many measurements within the test area as well as on neighbouring land. To do this, it is useful to have a plan of the installation with distances of 1 metre marked on it. Whilst testing touch voltage, one must be aware of the possible transmission of dangerous voltages away from the ground electrode via long metallic structures (metal fences, pipes, rails etc). Testing must be done further away from these conductive devices.
5.10.- TESTING VOLTAGES DUE TO STATIC CURRENTS
Prior to starting, press the "ERRATIC VOLTAGES" button (keeping it pressed down).
If the Vm value remains at 0 V the presence of erratic voltages and currents will be ruled out. If any value appears, it must be noted to subsequently compare it with the 20 A and 50 A measurements and to make the appropriate numerical corrections). (See section 8 "CORRECTING READ OUTS").
Page 19
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
19
5.11.- ADJUSTING THE CURRENT AND TESTING OF STEP OR TOUCH VOLTAGE
Turn the key in one of the two directions and keep it there.
BEFORE TURNING THE SELECTOR MAKE SURE THAT NO ONE IS
USING THE POWER OR TESTING CIRCUIT. PREVENT PEOPLE REMAINING OR PASSING THROUGH THE AREA WHERE THE CURRENT IS TO BE INJECTED BY USING THE APPROPRIATE SIGNS. IN THE EVENT OF HIGH GROUND RESISTANCES, VOLTAGES UP TO 600 V MAY BE CREATED, WHICH ARE FATAL TO ANIMALS OR PERSONS NEAR TO THE DISPERSERS DURING TESTS.
Adjust the current to 20 A with the MPC-20 and 50 A or slightly higher for the MPC-50:
FINE REGULATOR:
Starting with the switch on position "1", slowly turn the fine regulator towards the right (from 0% to 100%). To find out the current being injected, the needle on the analogue ammeter is the guide at first and then the read out lp. on the digital indicator.
SWITCHING SCALES:
If the nominal current is not reached, even with the regulator on 100%, it will then be necessary to increase the switch regulation by one scale.
Adjustments are done by trial and error between the different regulation scales, starting with the lowest: 1 to 10 on the MPC-20 and from 1 to 8 on the MPC-50. Switching from one scale to another is always done “off line”, i.e. without turning the selector key and placing the regulator in the 0% position.
Voltage testing will be done AUTOMATICALLY at the same time as
the MPC-20 or MPC-50 are adjusted with the operating key. (The voltmeter begins testing when the key is turned and blocks measurements when the key is no longer turned.)
Page 20
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
20
READ TESTED VOLTAGE (V) on the display.
Note: If the read out on the display is less than 30 V the internal scale is automatically changed to 30 V AC.
DISENGAGE THE OPERATION SELECTOR. Once the test has
finished the values remain displayed until the key is turned again, the equipment is switched off or the Imax value is set again.
5.12.- TEST RESULTS (THAT WILL REMAIN DISPLAYED)
Real current Maximum current of the test made 
Im. NN,N I x. NNNN
default theory
Voltage tested during the test
Vm. NN,N V NNNNN
Calculated voltage default theory
(step or touch)
Vm x Ix
STEP VOLTAGE (Vp) : V = -------------------­ Im
Vm x Ix TOUCH VOLTAGE (Vc) : V = ---------------------
Im
Page 21
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
21
6.- TESTING GROUND RESISTANCE
The main advantage of testing ground resistance using high currents is that as well as being measured, the ground is also being tested to see if it is capable of accepting such a current. By doing it this way, one avoids errors of measurement from low currents due to, for example, surface moisture or a highly corroded cable or one with defective lining.
CONNECTION DIAGRAM
TESTED VOLTAGE
TESTING
ROD
TESTING GROUND RESISTANCE
GROUND TO
BE TESTED
AUXILIARY
GROUND
TEST INTENSITY lp
Page 22
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
22
6.3.- ATTACHING AND CONNECTING THE MPC-20 or MPC-50
The "OPERATING MODE" selector must be placed on Rt.
Attaching and connecting the MPC to the system is the same as that
for testing step and touch voltage.
Connecting the current circuit using the two output sockets is also the
same.
6.4.- ATTACHING THE MEASURING ROD
One of the voltmeter cables is connected to the ground to be
measured.
The other voltmeter cable is connected to the GROUND using the
"MEASURING ROD" in an area in between the ground to be measured and the auxiliary ground.
6.5.- TESTING VOLTAGES DUE TO ERRATIC CURRENTS
Prior to starting, press the "ERRATIC VOLTAGES" button (keeping it
pressed down).
If the Vm value remains at 0 V the presence of erratic voltages and
currents will be ruled out. If any value appears, it must be noted to be subsequently compared with the 20 A and 50 A measurements and the appropriate numerical corrections made). (See section 8 "CORRECTING READ OUTS").
6.6.- TESTING GROUND RESISTANCE
If testing has to be done at 20 A (MPC-20) or at 50 A (MPC-50),
adjustments are made according to section 11.3.2.
If testing has to be done with currents below 20 A or 50 A respectively,
the adjustment is also done by trial and error between the different regulation scales starting with the lowest (1, 2, 3, ...). Switching from one scale to another is always done “off line”, i.e. without turning the selector key (according to section 11.3.2.).
Page 23
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
23
Measurements are taken automatically on turning the key (the
voltmeter and the ammeter begin testing on turning the key and block measurements when the key is no longer turned).
BEFORE TURNING THE SELECTOR MAKE SURE THAT NO ONE
IS USING THE POWER OR TESTING CIRCUIT.
READ TESTED VOLTAGE (V) on the display.
Note: If the read out on the display is less than 30 V the internal scale is automatically changed to 30 V AC.
DISENGAGE THE OPERATION SELECTOR Once the test has
finished the values remain displayed until the key is turned again, the equipment is switched off or the equipment is disconnected.
The MPC-5 will automatically calculate resistance and store the
results.
6.7.- MEASUREMENT RESULTS (THAT WILL REMAIN DISPLAYED)
Real current of the test made 
Im. NN,N Rt :
Voltage tested during the test
Vm. NN,N NNN.N
Calculated ground
resistance
Rt.= Vm./Im.
Ensure that the measuring electrode is actually testing the voltage drop in the ground that is being tested which is a consequence of the current injected by the MPC itself. Placing the measuring electrode in the right position is vital. It is important that it is far from the area under the influence of both the measuring electrode and the auxiliary electrode to obtain maximum accuracy.
Page 24
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
24
6.8.- REPEATING TESTS
Ensure that the measuring electrode is actually testing the voltage drop in the ground that is being tested which is a consequence of the current injected by the MPC itself. Placing the measuring electrode in the right position is vital. It is important that it is far from the area under the influence of both the measuring electrode and the auxiliary electrode to obtain maximum accuracy.
It is advised that the measuring rod is moved +10% and then -10% from its starting position along the axis formed by the earthing and the auxiliary electrode separated by a distance L (if its position L=20 metres, it is to be moved 2 metres from its original position). Make a note of the resistance value for each position of the rod. If the values are the same or vary slightly the measurement is correct. On the other hand if it is within the area influenced by the two power slopes. The distance between the ground to be measured and the auxiliary ground has to be increased.
Page 25
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
25
CHAPTER II
7.- EXISTING STANDARDS
7.3.- SUMMARY OF HIGH VOLTAGE REGULATIONS
The Technical Complementary instruction MIE-RAT-13 published in the B.O.E. (Official State Gazette) 183 of 1st August 1984, section 1.1. sets the maximum admissible step and touch voltages for EARTHING INSTALLATIONS.
In sections 8.1 and 8.2 it establishes the requirement to verify that the step and touch voltages are within the admissible limits. It establishes the requirement for periodical checking every 3 years as a maximum. It lays down that the measuring electrodes that simulate feet must have a surface area of 200 square cm each and that they each have a downward force onto the ground of 250 N.
STEP VOLTAGE
-The voltage that can occur between the feet of somebody standing one
step away whilst the installation is operating" (I.E.C. standard 11-8, sec.176).
TOUCH VOLTAGE
-The voltage that a human body may experience whilst in contact with
housings and metallic structures that normally do not carry voltages.
MAXIMUM ADMISSIBLE VOLTAGES
- Maximum admissible step and touch voltages in a certain installation,
(B.O.E. 291 5-12-87), are determined by the following summarised formulae:
10 K K
Vp = -------------- Vc = -----------
t
n t n
Being:
K=72 and n=1 for times less than 0.9 s. K=78.5 and n=0.18 for times between 0.9 and 3 s. t = length of the fault in seconds. For times between 3 and 5 seconds the touch voltage will not exceed 64 V and the step voltage 640 V.
For times above 5 seconds touch voltage will not exceed 50 V and step voltage 500 V.
Page 26
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
26
SUMMARY TABLE OF MAXIMUM VOLTAGES
TIME MAXIMUM STEP
VOLTAGE
VOLTAGE TOUCH
VOLTAGE
Over 5 s 500 V * 50 V From 3 to 5 s 640 V 64 V 2 s 690 V 69 V 1 s 785 V 78.5 V
0.9 s 800 V 80 V
0.7 s 1020 V 102 V
0.5 s 1440 V 144 V
0.2 s 3600 V 360 V
0.1 s or less 7200 V 720 V
* Supposes that the real step voltage on the human body is less than that
measured (shoes, etc.).
7.4.- STANDARDS REFERRING TO THE MPC-20 / MPC-50
MIE-RAT-01 MIE-RAT-13 MIE-RAT-19 BOE 183 1st August 1984 BOE 291 5th December 1987
Page 27
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
27
8.- CORRECTING THE READ OUTS
Results from testing touch or step voltages may be incorrect due to the
existence of erratic, vagabond or parasite currents passing through the ground.
The B.O.E. 291, of 5 of December of 1987 establishes:
1.1.12 "... unless a test method is used that eliminates the effect (in testing) of these parasite currents, e.g. polarity inversion, it will mean that the injected current will be 1% of the current set for the installation and will not be less than 50 A for stations and substations and 5 A for transforming stations."
As a result of currents not indicated by the B.O.E. being used, it will be
necessary to use phase inversion for testing.
It is also found that in practice, the tested erratic voltages are always negligible in terms of those created by injected currents (as long as they conform to the set minimum).
However, in certain cases it will be possible to do a pretest on existing voltages without injecting a current. To do this tests by pressing "Erratic voltages" are done. Only when measurements of erratic voltages are above 10% of measurements with current injection, the following calculation must be made:
Vp =
)( º)
()
VV
Ve
0 180
2
22
2
+
Vp = Step or touch voltage Ve = Voltage tested "ERRATIC VOLTAGES" V0° = Voltage tested "ON 0°" V180° = Voltage tested "ON 180°"
Page 28
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
28
DERIVATION OF THE FORMULA FOR STEP AND CONTACT VOLTAGE
The following equations are derived from the graph:
V Vp Ve ß Ve ß1
222
=+× ( sen ) ( cos )
Vp Ve ß Ve ß2
222
=−× ( sen ) ( cos )
Developing the equations:
Vp Vp Ve ß Ve ß Ve ß12
22 2222
=+×× +× sen sen cos
Vp Vp Ve ß Ve ß Ve ß22
22 2222
=− ×× +× sen sen cos
If the two equations are added, we have to:
V VpVe ß ß VpVe1222 22
22 222 2 22
+= + + = +(sen cos )
Solving Vp: Vp =
() ( )
()
VV
Ve
12
2
22
2
+
V1 = read voltage on injecting current at 0º V2 = read voltage on injecting current at 180º Ve = read voltage without injecting current
Page 29
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
29
9.- MAINTENANCE
The MPC does not require specialised maintenance. As far as possible avoid any adjustment, maintenance or repairs when the equipment is open. If this is unavoidable it must be done by qualified personnel or technicians well informed of the procedure to follow.
Before any modification to the connections, replacement, maintenance or repair, the equipment must be disconnected from the supply.
If any operating fault or equipment protection fault is suspected, the equipment must be taken out of service ensuring that no accidental reconnection may occur. The equipment is designed to have any of its components replaced quickly in the event of breakdown.
10.- TECHNICAL SERVICE
In the event of any equipment failure or any operational queries please contact the technical service of CIRCUTOR S.A.
CIRCUTOR S.A. - After sales service c / Lepanto , 49 08223 - TERRASSA tel - (+34) 93 - 745 29 00 e-mail: central@circutor.es
Page 30
--- MPC 20 & 50 TESTER - M-981136 (vers. BD) --- Page
30
APPENDIX
MPC-50 INTERNAL TERMINAL BOARD
FUSIBLE MANIOBRA
400 V 230 V
CAMBIO DE TENSN
CHANGE OF VOLTAGE:
- Attach the voltage bridge in the required position
- Attach the indicating tag on the external supply cable
OPERATING FUSE
CHANGING VOLTAGE
Loading...