CIRCUTOR AR5L User Manual

SUPPLY NETWORK ANALYZER
AR5 and AR5-L
INSTRUCTION MANUAL
( M98151101-03-10B )
(c)
CIRCUTOR S.A.
CONTENTS page
1.- BASIC INSTRUCTIONS ........................................................................ 4
1.1.- Introduction: check the contents of your package................................. 4
1.2.- Safety conditions................................................................................... 4
1.3.- Connection instructions......................................................................... 4
1.4.- Operation instructions ........................................................................... 5
2.- ANALYZER MAIN FEATURES.............................................................. 6
2.1.- Basic features ....................................................................................... 6
2.2.- Other characteristics ............................................................................. 6
3.- KEYBOARD FUNCTIONS..................................................................... 9
4.- Installation AND START-UP .................................................................. 9
4.1.- Steps necessary for the START-UP ..................................................... 9
4.2.- Connection diagram............................................................................ 11
4.3.- Starting the analyzer up ...................................................................... 16
4.4.- Loading a new program ...................................................................... 18
4.5.- Turn the analyzer off ........................................................................... 19
4.6.- Back-light ............................................................................................ 19
4.7.- Choice of the working program ........................................................... 19
4.8.- Recharging the analyzer battery ......................................................... 20
4.9.- Energy saving ..................................................................................... 20
5.- DATA VISUALIZATION ON DISPLAY................................................. 21
5.1.- Base screen........................................................................................ 21
5.1.1.- Screen of instantaneous values.................................................... 21
5.1.2.- Screen of maximum and minimum values ................................... 22
5.2.- Other visualization screens................................................................. 23
5.2.1.- Visualization of 3 parameters in a big size mode.......................... 23
5.2.2.- Bar graphs .................................................................................... 24
5.2.3.- Oscilloscope.................................................................................. 25
5.2.3.1.- Three phases: Voltage - Current............................................. 25
5.2.3.2.- Zoom....................................................................................... 26
5.2.3.3.- Harmonic factorization. ........................................................... 27
5.2.4.- Setup visualization ........................................................................ 29
5.3.- Warning messages ............................................................................. 29
6.- PROGRAMMING THE ANALYZER..................................................... 30
2
6.1.- SETUP menu...................................................................................... 31
6.1.1.- MEASURE .................................................................................... 32
6.1.1.1.- WIRING: Circuit type............................................................... 32
6.1.1.2.- PT/CT : Transformation ratios................................................. 32
6.1.2.- RECORD menu ............................................................................ 32
6.1.2.1.- PERIOD: recording period ...................................................... 33
6.1.2.2.- TRIGGER: Trigger conditions ................................................. 34
6.1.2.3.- NAME: recording file name ..................................................... 37
6.1.2.4.- PARAM: Choosing the parameters to be saved...................... 38
6.1.3.- COMM: Communication parameters............................................. 39
6.1.4.- CLOCK: Internal clock .................................................................. 39
6.1.5.- PASSWORD: Safety setting. ........................................................ 40
6.1.6.- RECALL: Read configuration........................................................ 41
6.2.- DISPLAY menu................................................................................... 42
6.2.1.- BAR.GR. ....................................................................................... 42
6.2.2.- EXPAND ....................................................................................... 42
6.2.3.- CONTRAST: Screen contrast ....................................................... 42
6.2.4.- ANGLE.......................................................................................... 43
6.3.- RUN: data recording process status ................................................... 43
6.4.- FILES Menu........................................................................................ 43
6.4.1.- DIR: Directory ............................................................................... 43
6.4.2.- DELETE: Deleting a file ................................................................ 43
6.4.3.- FORMAT: Formatting the analyzer internal memory..................... 44
6.5.- CLEAR menu: Deleting data............................................................... 44
6.6.- Menu OFF: Enable / Disable Password. ............................................. 45
6.7.- Menu LANGUAGE .............................................................................. 45
7.- ANALYZER COMMUNICATIONS ....................................................... 45
8.- TECHNICAL SPECIFICATIONS.......................................................... 46
9.- SAFETY WARNINGS .......................................................................... 48
10.- MAINTENANCE................................................................................... 49
11.- CHANGING THE BATTERY................................................................ 49
12.- TECHNICAL SERVICE........................................................................ 50
3
1.- BASIC INSTRUCTIONS
This manual is aimed to familiarize de user with the operation of the portable
power analyzers model AR5-L and AR5 in order to get the best from its features.
These analyzers have been built with components incorporating the most advanced technology in microelectronics and offer benchtop features over the market in measurement and recording of electrical magnitudes in industrial power supply networks.
You are kindly requested to carefully read this manual before connecting and powering the analyzer in order to avoid irreversible damage, which might be caused by an improper connection.
1.1.- Introduction: check the contents of your package
After receiving the analyzer, please check the following points:
a) The analyzer model corresponds with your order specifications.
b) After unpacking, check that the instrument has not been damaged in transit
c) The standard set includes the following items:
Kit AR5-L or AR5
1 Power supplier set 100 V a.c. – 240 V a.c. / 12 V d.c.
1 Connection cord between the power supplier set and the main.
1 Connection cable between the analyzer and the power supplier
set.
1 RS-232 communication cable.
4 Voltage leads.
4 Alligator clamps (3 in AR5 model).
1 Instruction Manual.
CD with the PC program
1.2.- Safety conditions
The manual you hold in your hands contains information and warnings about the analyzer that the user should respect in order to guarantee a proper operation of all the instrument functions and keep its safety conditions.
1.3.- Connection instructions
.
Before powering and connecting the analyzer check the following points:
a) Supply voltage: Through and external power supplier set.
Input mains 100 V a.c. – 240 V a.c. / output to AR5/AR5-L 12V d.c.
b) Frequency : 45...65 Hz.
c) Maximum voltage at the voltage measuring circuit:
500 V a.c. Phase-to-neutral (CAT III)
4
d) Current measuring range: according to the Current clamp used
Current clamps Measuring range
CP-2000-200
10 to 2000 A a.c. (switch at 2000)
1 to 200 A a.c. (switch at 200)
CPR-1000 5 to 1000 A a.c.
CPR-500 2,5 to 500 A a.c.
CP-100 (M1-U) 0, 5 to 100 A a.c.
CP-5
25 mA to 5 A a.c.
Flexible clamp Measuring range
C-FLEX 200-2000-20000 100 to 20000 A a.c (scale 20000
A)
10 to 2000 A a.c. (scale 2000 A)
1 to 200 A a.c. (scale 200 A)
NOTE: It is advisable to measure near of full-scale value to get better accuracy.
1.4.- Operation instructions
The analyzer is a programmable instrument so offering diverse operation modes, which can be selected from the available programming menus (6.­PROGRAMMING THE ANALYZER).
Please read carefully the paragraphs involving (4.- Installation AND START­UP 6.- PROGRAMMING THE ANALYZER) in order to select the most suitable operation mode for your requirements.
If the instrument is not used according to
manufacturer’s specifications, the protection of
the instrument can be damaged.
5
2.- ANALYZER MAIN FEATURES
Clamp inputs (.. 2 V a.c.).
Power Supply: 12 V c.c. / Serial input
VL1
VL2
VL3
AUX
*
*
2.1.- Basic features
The AR5 series analyzers are programmable instruments that measure, calculate and store to memory the main parameters of three phase electrical
supply networks.
a.c. Voltage inputs
IL2IL3
IL1 I
N
N
* IN only in AR5-L models
Setup: Analyzer’s setting is completed through a system of spreadable menus
that provide a friendly-use and intuitive configuration process.
Data visualization: By means of a liquid crystal graphical display, 160 x 160 pixels and with backlight ability, the user can view instantaneous, maximum and
minimum values of each parameter at each phase.
Internal battery: permits the user to accomplish with analysis works without
the need of an external auxiliary power supply. The analyzer is equipped with an intelligent battery charging system, which expands the battery life span. For recharge the battery is necessary to connect the power supplier.
Installation: Analyzer is suitable for analysis works over any type of electric networks (single-phase, bi-phase, 3-wire and 4-wire).
Measuring: Measurement of average values of main electrical parameters, as well as, recording of maximum and minimum values. To execute these measurements, the analyzer is equipped with three a.c. voltage inputs and four a.c. current inputs (through current clamps ../ 2 V a.c.).
Data collection: The analyzer has a 1 Mb on-board memory to save into all the parameters measured or calculated by the analyzer, in order to be further retrieved from a PC.
PC software: together with the analyzer, it is also factory supplied a friendly­use, powerful software that permits data downloading from the internal memory to a PC and a further complex analysis of recorded data.
2.2.- Other characteristics
6
- Small size, low weight, portable instrument.
- True R.M.S. measuring system.
- Instantaneous, maximum and minimum values of each discrete electric
parameter.
- Electrical energy meters incorporated
- Harmonic measurement
- Neutral current measurement (only in AR5-L model)
- RS-232 communication to PC
- Ability of setting a recording threshold, so that data is only recorded in
memory when measured values are not within the defined threshold (see
Section 6.1.2.2.-TRIGGER: Trigger conditions).
- Automatic data collection in memory at regular user-definable periods.
Depending on the type of circuit under analysis, following enumerated parameters are measured and can be saved in memory:
NOTE: It’s not allow to save cosϕ data in memory. Only can be read on
display.
4-wire three-phase system:
Parameter Symbol L1 L2 L3 Three-phase value
Phase-to-Neutral voltage V x x x Current A x x x x Neutral current (only in AR5-L model) Frequency Hz x Active power kW x x x x Reactive power L kvarL x x x x Reactive power C kvarC x x x x Apparent power kVA x Power factor PF x x x x Active energy Reactive-energy (inductive) Reactive-energy (capacitive) Voltage harmonics x x x Current harmonics x x x Current Neutral harmonics (only in AR5-L model)
IN x
kW h
kvaLh
kvaCh
x
x x x x x x x x x x x x
7
3-wire three-phase system:
Parameter Symbol L1-L2 L2-L3 L3-L1 Three-phase value
Phase-to-Phase voltage Current Frequency Active power Reactive power L Reactive power C Apparent power Power factor Active energy Reactive-energy (inductive) Reactive-energy (capacitive) Voltage harmonics Current harmonics
V A
Hz
kW
kvarL
kvarC
PF
kW h
kvahL
kvahC
Single-phase system:
Parameter Symbol L1
Voltage Current Frequency Active power Reactive power L Reactive power C Apparent power Power factor Active energy Reactive-energy (inductive) Reactive-energy (capacitive) Voltage harmonics Current harmonics
x x x x x x x x x x x x x x x x x x x x
x x x x x x x x x x x x x x x x x x x x x x x
V A
Hz
kW
kvarL
kvarL /(-C)
PF
kW h
kvahL
kvahC
x x x x x x x x x x x x x
Bi-phase system:
Parameter Symbol L1-N
Voltage Current Neutral current*
Frequency Active power Reactive power L Reactive power C Apparent power Power factor Active energy Reactive-energy (inductive) Reactive-energy (capacitive) Voltage harmonics Current harmonics Neutral current harmonics
*only in AR5-L model
L2-N
Bi-phase
value
L1-L2
V A
IN
Hz
kW
kvarL
kvarC
PF
kW h
kvahL
kvahC
x x x x x x
x
x x x x x x x x x x
x
x x x
x x x x x x x x x x x
x
8
3.- KEYBOARD FUNCTIONS
The analyzer has a 9 buttons keyboard to perform configuration and
control actions of all the instrument options.
- [ON] to turn the analyzer on.
- [OFF] (Quick press) to turn on/off the display back-light.
- [OFF] (5 seconds press) to turn analyzer off.
- [6666], [5555], [4444] & [3333]. To select among several options.
- [SET] to access setting options.
- [ENTER] to validate a setting option or to program some parameters of the visualization screens.
- [ESC] to select different visualization screens or to exit the setting actions.
However, most of keys are double functional: The own instrument
understands the meaning of the order at each case.
4.- Installation AND START-UP
The manual you hold in your hands contains information and warnings that the user should respect in order to guarantee a proper operation of all the instrument functions and keep its safety conditions.
If the instrument is not used according to manufacturer’s
specifications, the protection of the instrument can be damaged. Note that with
the instrument powered on, cover opening or elements removal actions may allow accessing dangerous parts. Therefore, before any adjustment, replacement, maintenance or repairing operation is carried out, the instrument must be disconnected from any power supply source.
When any protection failure is suspected to exist, the instrument must be
immediately put out of service. Contact then with a qualified service representative.
4.1.- Steps necessary for the START-UP
9
For the Star-up of the equipment and to let it ready to begin to register is necessary to make the following steps.
1) Plug the battery, see point 11.-CHANGING THE BATTERY
2) Format the memory, see 6.4.3.-FORMAT: Formatting the analyzer internal memory
3) Set the analyzer in hour, see point 6.1.4.-CLOCK: Internal clock
4) Charge the battery by a period of minimum 14 hours. .
10
4.2.- Connection diagram
IL1
IL2
VL1
VL2
VL3
AUX
A.- CONNECTION DIAGRAM FOR THREE-PHASE - 4-WIRE – SYSTEM
This connection is only available for AR5-L model
( SET ---> SETUP ---> MEASURE ---> WIRING ---> 3Φ 4 WIRES)
L1
L2
L3
N
IL3
11
I
N
N
B.- CONNECTION DIAGRAM FOR THREE-PHASE - 3-WIRE - SYSTEM.
IL1
IL2
VL1
VL2
VL3
AUX
( SET ---> SETUP ---> MEASURE ---> WIRING ---> 3Φ 3 WIRES )
L1
L2
L3
IL3
I
N
N
12
L1
IL1
IL2
VL1
VL2
VL3
AUX
L2
L3
C.- CONNECTION DIAGRAM FOR THREE-PHASE - 3-WIRE (ARON)
( SET ---> SETUP ---> MEASURE ---> WIRING ---> 3 PT – 2 CT
IL3
13
I
N
N
L1
IL1
IL2
VL1
VL2
VL3
AUX
N
D.- CONNECTION DIAGRAM FOR SINGLE-PHASE - SYSTEM
( SET ---> SETUP ---> MEASURE ---> WIRING ---> 1Φ)
IL3
14
I
N
N
L1
IL1
IL2
VL1
VL2
VL3
AUX
L2
N
E.- CONNECTION DIAGRAM FOR BI-PHASE - SYSTEM
This connection is only available for AR5-L model
( SET ---> SETUP ---> MEASURE ---> WIRING ---> 1Φ Split )
IL3
15
I
N
N
4.3.- Starting the analyzer up
Before connecting the instrument to the mains, please consider following points:
5) Supply voltage: 100 V a.c. – 240 V a.c. , 50... 60 Hz.
The instrument must be energized by a supply circuit with protection earth terminal.
6) Maximum input voltage at the voltage measuring circuit:
500 V a.c. phase-neutral (CAT III)
Use always the voltage leads factory-shipped with the instrument.
7) Burden: 15 VA.
8) Operation conditions:
- Operation temperature: 0 ºC to 40 ºC.
- Operation humidity: 80% for temperatures up to 31ºC decreasing linearly to 50% relative humidity at 40ºC.
9) Safety : Designed to meet protection category III 600V as per EN 61010.
10) Current measuring range: according to the current clamp used
Current clamps Measuring range
CP-2000-200 clamp 10 to 2000 A a.c. (scale 2000 A)
1 to 200 A a.c. (scale 200 A)
CPR-1000 clamp 5 to 1000 A a.c.
CPR-500 clamp 2,5 to 500 A a.c.
CP-100 (M1-U) clamp 0´5 to 100 A a.c.
CP-5 clamp 25 mA to 5 A a.c.
Flexible clamp Measuring range
C-FLEX 200-2000-20000 100 to 20000 A a.c. (scale 20000 A)
10 to 2000 A a.c. (scale 2000 A)
1 to 200 A a.c. (scale 200 A)
NOTE: It is always advisable to take measurements at the high range of the
scale to obtain a better accuracy.
To start measurement works with the analyzer:
11) Connect to the main with the factory-shipped cables. Also connect the protection earth terminal to avoid possible disturbances over the analyzer.
12) Connect the voltage leads at each phase of the monitored system, as well as the neutral when it exists.
13) Connect the Current clamps at each phase conductor. Each current phase must coincide with its voltage phase.
14) Respect the connection modes shown at each diagram to correctly achieve powers, PF and energies readouts.
16
Loading...
+ 36 hidden pages