circle RC100 Data Sheet

Page 1
Circle Model RC100 System Performance Data Sheet
The RC100 is tested and certied to NSF/ANSI 42, 53 and 58 for the reduction of Aesthetic Chlorine, Taste and Odor, Cyst, VOCs, Fluoride, Pentavalent Arsenic, Barium, Radium 226/228, Cadmium, Hexavalent Chromium, Trivalent Chromium, Lead, Copper, Selenium and TDS as veried and substantiated by test data. The RC100 conforms to NSF/ANSI 372 for low lead compliance.
This system has been tested according to NSF/ANSI 42, 53 and 58 for the reduction of the substances listed below. The concentration of the indicated substances in water entering the system was reduced to a concentration less than or equal to the permissible for water leaving the system, as specied in NSF/ANSI 42, 53 and 58.
Inuent Challenge
Substance
Arsenic (Pentavalent) 0.05 +/- 10% 0.01 98.9
Barium 10.0 +/- 10% 2 98.2
Radium 226/228 25 pCi/L 5 pCi/L 80
Cadmium 0.03 +/- 10% 0.005 98.2
Chromium (Hexavalent) 0.3 +/- 10% 0.1 97.3
Chromium (Trivalent) 0.3 +/- 10% 0.1 97.8
Copper 3.0 +/- 10% 1.3 98.7
Lead 0.15 +/- 10% 0.01 98.2
Selenium 0.10 +/- 10% 0.05 97.3
Fluoride 8.0 +/- 10% 1.5 83.8
TDS 750 +/- 40 187 92.5
Aesthetic Chlorine 2.0 +/- 10 % >/= 50% Reduction 98.4
VOCs 0.300 +/- 10% >/= 95% Reduction 99.8
While testing was performed under laboratory conditions, actual performance may vary.
Rated Capacity 3,500 gal. for Aesthetic Chlorine / 243 gal. for VOC’s
Min-Max operating pressure: 40psi-120psi (275Kpa-827Kpa)
Min-Max feed water temperature: 41°F – 95°F (5°C – 35°C)
Rated Service Flow 0.07 gallons per minute
Product Eciency Rating 30.6%
Concentration
(mg/L unless specied)
Maximum Allowable Product Water
Concentration
(mg/L unless specied)
Percent Reduction
(%)
Do not use with water that is microbiologically unsafe or of unknown quality without adequate disinfection before or after the system.
Refer to the owners manual for specic installation instructions, manufacturer’s limited warranty, user responsibility, and parts and service availability.
The inuent water to the system shall include the following characteristics:
No organic solvents
Chlorine: < 2 mg/L
pH: 7 - 8
Temperature: 41 ~ 95 ºF (5 ~ 35 ºC)
Systems certied for cyst reduction may be used on disinfected waters that may contain lterable cysts.
Iron: < 2 mg/L
Turbidity: < 1 NTU
Hardness: < 1 000 mg/L
Brondell, Inc. P. O. Box 470085 San Francisco CA 94147-0085 888-542-3355 brondell.com support@brondell.com
Page 1 of 3
Page 2
Circle Model RC100 System Performance Data Sheet
For parts and service availability, please contact Brondell at 888-542-3355.
This system has been tested for the treatment of water containing pentavalent arsenic (also know as As(V), As(+5), or arsenate) at concentrations of 0.050 mg/L or less. This system reduces pentavalent arsenic, but may not remove other forms of arsenic. This system is to be used on water supplies containing a detectable free chlorine residual at the system inlet or on water supplies that have been demonstrated to contain only pentavalent arsenic. Treatment with chloramines (combined chlorine) is not sucient to ensure complete conversion of trivalent arsenic to pentavalent arsenic. Please see the Arsenic Facts section of this Performance Data Sheet for further information.
Eciency rating means the percentage of the inuent water to the system that is available to the user as reverse osmosis treated water under operating conditions that approximate typical daily usage.
The product water should be tested every 6 months to ensure that the contaminants are being reduced eectively. For any questions, please contact Brondell toll free at 888-542-3355.
This reverse osmosis system contains a replaceable treatment components, critical for the eective reduction of total dissolved solids and that product water shall be tested periodically to verify that the system is performing properly. Replacement of reverse osmosis component should be with one of identical specications, as dened by the manufacturer, to assure the same eciency and contaminant reduction performance.
The estimated replacement time of lter, which is a consumable part, is not an indication of quality guarantee period, but it means the ideal time of lter replacement. Accordingly, the estimated time of lter replacement may be shortened in case it is used in an area of poor water quality.
Part Number Filter Name Usable period
Sediment Filter 6 months
RF20
RF-40 RO Membrane Filter 24 months
Pre-Carbon Plus Filter 6 months
Post Carbon Block Filter 6 months
* The lter replacement cycle may be reduced for areas with poor water quality or greater use. * The period for the lter exchange is based on an average production of 2.6 gallons of drinking water per day.
ARSENIC FACTS
Arsenic (abbreviated As) is found naturally in some well water. Arsenic in water has no color, taste or odor. It must be measured by a lab test. Public water utilities must have their water tested for arsenic. You can get the results from the water utility. If you have the own well, you can have the water tested. The local health department or the state environmental health agency can provide a list of certied labs. Information about arsenic in water can be found on the Internet at the US Environmental Protection Agency website: www.epa.gov/safewater/arsenic.html There are two forms of arsenic: pentavalent arsenic (also called As(V), As(+5), and arsenate) and trivalent arsenic (also called As(III), As(+3), and arsenite). In well water, arsenic may be pentavalent, trivalent, or a combination of both. Special sampling procedures are needed for a lab to determine what type and how much of each type of arsenic is in the water. Check with the labs in the area to see if they can provide this type of service. Reverse osmosis (RO) water treatment systems do not remove trivalent arsenic from water very well. RO systems are very eective at removing pentavalent arsenic. A free chlorine residual will rapidly convert trivalent arsenic to pentavalent arsenic. Other water treatment chemicals such as ozone and potassium permanganate will also change trivalent arsenic to pentavalent arsenic. A com­bined chlorine residual (also called chloramine) may not convert all the trivalent arsenic. If you get the water from a public water utility, contact the utility to nd out if free chlorine or combined chlorine is used in the water system.
The RC100 system is designed to remove pentavalent arsenic. It will not convert trivalent arsenic to pentavalent arsenic. The system was tested in a lab. Under those conditions, the system reduced 0.050 mg/L pentavalent arsenic to 0.010 mg/L (ppm) (the USEPA standard for drinking water) or less. The performance of the system may be dierent at the installation. Have the treated water tested for arsenic to check if the system is working properly. The RO component of the RC100 system must be replaced every 24 months to ensure the system will continue to remove pen­tavalent arsenic. The component identication and locations where you can purchase the component are listed in the installation/ operation manual.
Brondell, Inc. P. O. Box 470085 San Francisco CA 94147-0085 888-542-3355 brondell.com support@brondell.com
Page 2 of 3
Page 3
Circle Model RC100 System Performance Data Sheet
Volatile Organic Chemicals (VOCs) included by surrogate testing*
Chemical
Drinking water
regulatory level
(MCL/MAC) mg/L
Inuent challenge
1
concentration2 mg/L
Chemical
reduction percent
alachlor 0.002 0.05 > 98 0.001 atrazine 0.003 0.1 > 97 0.003 benzene 0.005 0.081 > 99 0.001 carbofuran 0.04 0.19 > 99 0.001 carbon tetrachloride 0.005 0.078 98 0.0018 chlorobenzene 0.1 0.077 > 99 0.001
Maximum
product water
concentration mg/L
3
3
3
3
3
chloropicrin - 0.015 99 0.0002 2,4-D 0.07 0.11 98 0.0017 dibromochloropropane(DBCP) 0.0002 0.052 > 99 0.00002 o-dichlorobenzene 0.6 0.08 > 99 0.001 p-dichlorobenzene 0.075 0.04 > 98 0.001 1,2-dichloroethane 0.005 0.088 95
5
1,1-dichloroethylene 0.007 0.083 > 99 0.001 cis-1,2-dichloroethylene 0.07 0.17 > 99 0.0005 trans-1,2-dichloroethylene 0.1 0.086 > 99 0.001 1,2-dichloropropane 0.005 0.08 > 99 0.001 cis-1,3-dichloropropylene - 0.079 > 99 0.001
0.0048
3
3
3
3
3
3
dinoseb 0.007 0.17 99 0.0002 endrin 0.002 0.053 99 0.00059 ethylbenzene 0.7 0.088 >99 0.001
3
ethylene dilbromide (EDB) 0.00005 0.044 > 99 0.00002
haloacetonitriles (HAN) bromochloroacetonitrile dibromoacetonitrile dichloroacetonitrile trichloroacetoritrile
-
-
-
-
0.022
0.024
0.0096
0.015
98 98 98 98
0.0005
0.0006
0.0002
0.0003
haloketones (HK): 1,1-dichloro-2-propanone 1,1,1-trichloro-2-propanone
heptachlor (H-34,Heptox) 0.0004 0.08 > 99 0.0004 heptachlor epoxide 0.0002 0.0107 hexachlorobutadiene - 0.044 > 98 0.001
-
-
0.0072
0.0082
99 96
6
98 0.0002
0.0001
0.0003
3
hexachlorocyclopentadiene 0.05 0.06 > 99 0.000002 lindane 0.0002 0.055 > 99 0.00001 methoxychlor 0.04 0.05 > 99 0.0001 pentachlorophenol 0.001 0.096 > 99 0.001 simazine 0.004 0.12 > 97 0.004 styrene 0.1 0.15 > 99 0.0005 1,1,2,2-tetrachloroethane - 0.081 > 99 0.001 tetrachloroethylene 0.005 0.081 > 99 0.001 toluene 1 0.078 > 99 0.001 2,4,5-TP (silvex) 0.05 0.27 99 0.0016 tribromoacetic acid - 0.042 > 98 0.001
3
3
3
3
3
3
1,2,4-trichlorobenzene 0.07 0.16 > 99 0.0005 1,1,1-trichloroethane 0.2 0.084 95 0.0046 1,1,2-trichloroethane 0.005 0.15 > 99 0.0005 trichloroethylene 0.005 0.18 > 99 0.0010
trihalomethanes (includes): chloroform (surrogate chemical) bromoform bromodichloromethane chlorodibromomethane
xylenes (total) 10 0.070 > 99 0.001
0.080 0.300 95 0.015
3
4
3
4
3
5
3
4
4
3
3
3
3
3
3
3
6
3
3
3
3
4
3
4
3
3
* Chloroform was used as the surrogate chemical for VOC reduc tion claims
1 The se harmonized va lues were agreed u pon by represent atives of USEPA and Health C anada for the purp ose of evaluatin g products to the r equirements of th is Standard. 2 Inu ent challenge leve ls are average inu ent concentration s determined in su rrogate qualic ation testing. 3 Max imum product wa ter level was not obs erved but was se t at the detection l imit of the analysi s. 4 Max imum product wa ter level is set at a valu e determined in su rrogate qualic ation testing. 5 Chem ical reductio n percent and maxi mum product wate r level calculate d at chloroform 95% br eakthrough po int as determined i n surrogate quali cation testing . 6 The su rrogate test resu lts for heptach lor epoxide demo nstrated a 98% red uction. Thes e data were used to cal culate an upper occ urrence concentr ation which would p roduce a maximum product w ater level at the MCL .
Brondell, Inc. P. O. Box 470085 San Francisco CA 94147-0085 888-542-3355 brondell.com support@brondell.com
Page 3 of 3
Loading...