TS2
Service
Thermocot
Manual
iN
i
|
|
ή
[
il
Eu
ag
페
1
1
EL
Ц
|
CIG
Health
e
Commonweatih
Care
Indu
С.М.
т
000
029
|
729
pi
TS2
Thermocot
Service
Manual
TS2
Thermocot
Index
SPECIFICATION
Notes
to
specification
Warning
Caution
CHECKOUT
Photographs
Cleaning
Photographs
Alarm
TECHNICAL
Routine
Calibration
Air
Photographs
Statements
Table
Maintenance
Flow
Diagram
Statements
PROCEDURE
DESCRIPTION
Procedure
Service
Manual
13,
17,
25
Page
4
5
8
9
10
12
14
18
19
20
23
23
24
25
MECHANICAL
General
Front
Panel
Canopy
Frame
Motor
Assembly
ELECTRONICS
Circuit
Software
Control
Power
Display
Alarm
Circuit
SERVICE
Description
PCB
PCB
Diagram
CENTRES
DRAWINGS
Assembly
Assembly
Assembly
Assembly
Description
PCB
PCB
AND
PARTS
LISTS
47,
26,
28,
30,
32,
48,
27
29
31
33
34
35
42
44
45
46
46
49
50
TS2
Thermocot
Service
Manual
Specifications
4.
1.
Introduction
Control
Temperature
Range
The
TS2
Thermocot
incubator
transported
operates
features
control
safety
See
2.
Overall
Overall
Overall
Weight:
Mattress
Height
3.
from
quick
and a comprehensive
checks
also
the
Physical
Length:
Width:
Height:
above
Power
which
is
within,
various
operation,
and
Notes
Size:
Mattress:
Requirements
is
an
Air
Controlled
designed
or
alarms.
at
Dimensions
to
be
outside, a hospital
supply
voltages
good
temperature
range
the
end
of
this
440mm
235mm
easily
and
of
automatic
section.
830mm
430mm
(17
23kg
305mm
(12"x
infant
It
(33
inch)
(17
inch)
1/2
inch)
(51
575mm
x
22
(9
1/4
lb)
1/2")
inch)
37.0°C
to
30.0°C
(See
note
C
Warm-Up
5.
Start
21.0°C
30.0°C
Standby
6.
Less
(See
7.
Correlation
SetTemp
33.0°C
34.0°C
36.0°C
Temperature
than
Note F at
Accuracy
Digital
Canopy
(See
(SET
at
the
1.0°C
across
the
to
Incubator
Readout
Thermometer
Notes
on
mode
end
of
this
Times
Time
15min
5min
Smin
end
E
toReach
32.0°C
34.0°C
36.0°C
Distribution:
mattress.
of
this
of
Indicators
Temperature:
(Actual) + 0.5°C
end
at
H
and
Digital
section)
Overshoot
&
<2.0°C
section.)
340.80
this
of
Display)
Overshoot
(typical)
<0.2°C
<0.2°C
section)
Voltage
Hz
Max
AMPS
TYPICAL
Standby
On
TYPICAL
Standby
On
AMPS
AMPS
HEATER
240
50 | 60 | DC | DC
ON:
0.7 | 1.2 | 7.1
0.8 | 13 | 8.5
HEATER
0.2 | 04 | 14 | 14
0.3 | 0.4 | 1.7 | 19
|115 | 24|
1|
2|
OFF:
10|
12
20
|104
|119
8.
Fresh
(Fresh
Filter:
Effective
CarbonDioxideConcentratior
9.
<0.25%
CO,
Air
Air
Inlet
3m
"Filtrete“
area
Note
(See
Intake
open) > 10
grade
8700mm?
at
I
the
G0110
(13.5
end
pm
inch?)
this
of
section
TS2
10.
1.
il.
11.
Across
12.
i,
i.
Ш.
iv.
v.
13.
(See
the
Thermocot
Oxygen
With
applied-
With
8
lpm
after
an
ambient
(See
notes
Air
Velocity
mattress
Sound
At
centre
Sameas
8
lpm
Oxygen
Same
as
10
Ipm
Same
as
Alarm
Alarm
level
in
front
(See
Note
Alarms
Alarm
Fresh
5
sounding
Concentration
Fresh
Air
Inlet
maximum
Air
Inlet
Oxygen
minutes
of
B,
J)
-
<0.1
applied,
from
21%
m/s
Levels
of
Mattress
(i)
but
with
applied
(i)
but
with
Oxygen
of
Chart,
applied
(i)
but
with
3
metres
incubator
K)
and
serviceMana
open,
5
Ipm
Oxygen
40%
O,
Closed,
O,
-
55%
O,
minimum
(typical)
52
dbA
(max)
50
dbA
(typical)
Note
52dbA
54dbA
60dbA
>65dbA
L)
(typical)
(typical)
(typical)
NOTES:
A.
This
cannot
incubator
automatic
temperature.
adjusted
requirements
B.
OXYGEN:
achievable
applying
end,
amount
atmosphere.
If
the
maximum
achieved.
cause
some
If
the
of
oxygen
Note
that
Fresh
applied
accumulation
Carbon
is
an
Air
Controlled
be
used
as
ie.
there
control
by
The
within
oxygen
is
dependent
and
Fresh
of
Note
physiological
patients.
Fresh
Air
of
The
air
the
nursing
of
the
oxygen
the
at
on
position
Air
Inlet
40%
oxygen
that
Inlet
concentration
it
is
important
Air
Inlet
when
-
otherwise
Dioxide.
of
the
there
patient's
incubator;
a
Baby
Controlled
is
no
provision
the
infant's
temperature
staff
to
suit
infant.
concentration
incubator,
the
nipple
the
flow
of
leaks
is
left
only
even
this
complications
is
closed,
can
to
oxygen
can
by
at
the
rate
to
the
open,
a
can
be
low
level
with
higher
be
obtained.
only
close
is
being
be
an
expired
ae
一
it
for
must
be
the
level
left
and
the
outside
may
levels
the
一
SOFTWARE
Auto
reset
HARDWARE
Latch,
manual
14.
Humidity
Increase
in
holders
OVERTEMP
at
in
humidity
about
36.0°C
OVERTEMP
reset
by
20%
RH.
adding
(See
water
Note
37.5°C
38.0°С
to
sponges
M)
As
a
guide,
applied,
55%
within
ambient).
will
give
Note
though,
increase
incubator.
the
if
a
flow
the
concentration
5
minutes
Higher
higher
flows,
concentrations.
that
higher
noise
level
of
8
(i.e.
inside
lpm
or
of
oxygen
will
build
34%
above
longer
times,
flows
also
the
up
is
to
TS2
C.
D.
E.
Thermocot
Temperature
Set
The
During
to
available
will
normal
exceed
to
override
testing
of
points.
when
incubator.
Always
preparing
starting
When
overshoot.
Overtemperature
the
of
Temperature
Set
than
more
Temperature
about
after
warm-up
A
minimal
only
recommended
incubator
the
if
service
All
an
Temperatures
calibrated
centre
at
parameters
ambient
of
diameter
discs
The
the
above
points.
disc
Fach
diameter
50mm
diameter
Service
is
operation
37.0°C.
the
However,
Biomedical
a
maximum
this
Overtemperature
check
nurse
to
cold
from
inadvertent
avoid
To
this
for
and
34°C,
higher
a
to
minutes.
20
STANDBY
from
overshoot,
this
that
likely
is
notice.
short
specified
temperature
measured
are
thermocouples
black-painted
and
25mm
of
mounted
are
mattress
in
held
is
rods
plastic
base.
Manual
limited
not
is
it
facility
a
Technician
for
Temperature
Set
the
infant
an
there
alarms,
first
change
then
setting
will
it
and
mode
be
to
are
21°C.
of
attached
brass
thickness
a
horizontally
nominated
the
at
by
place
protruding
37.0°C.
to
possible
is
which
purpose
the
alarm
the
in
be
may
operation
make
warm-up
the
required
if
produce
is
employed
be
pressed
measured
using
to
having
discs
of
1mm
three
from
an
the
no
Set
into
at
the
1mm.
100mm
test
a
à
the
assess
To
Control
temperature,
this
following
Point
the
are
points
the
Temperature
the
and
5
100mm
is
A
mattress
same
the
at
the
of
mattress
CONTROL
selected
mode).
INCUBATOR
Temperature
F.
the
To
concentration
Fresh
Air
to
mattress
point
monitored
The
increases
lpm
this
applied.
at
is,
That
mattress
evaluate
Inlet
Air
administered
is
point
a
and
150mm
sound
sound
the
increases
sound
The
measured
perpendicular
3m
of
incubator.
incubator
the
the
floor.
temperature
can
difference
the
and
temperature
is
points
test
above
surface;
lines
to
TEMPERATURE:
the
100mm
the
100mm
until
level
as
level
at
Both
four
height
drawn
of
each
display
digital
TEMPERATURE:
measured
above
surface.
maximum
will
which
a
open,
at
above
CO,
the
from
apart
stability
inside
level
54dbA
to
the
of
point
the
positioned
are
flow
is
sound
oxygen
a
distribution,
any
at
set
be
between
the
of
each
at
measured.
exact
the
other
above
from
corners.
its
point
at
the
Test
centre
points
test
mid-
the
centre
the
Temperature
SET
(in
Note
in
A
centre
exact
CO,
the
with
occur,
CO,
mixture
4%
rate
a
the
concentration
this
achieved.
is
incubator
the
is
typically
when
Audible
which
the
to
of
750m1/min
of
of
centre
at
is
point
increased.
52dbA,
lpm
10
Alarm
distance
a
at
is
the
of
front
meter
level
above
1.5m
the
of
of
of
in
the
a
At8
but
is
is
and
TS2
Thermocot
L.
There
One
and
readout
automatically
disconnected
can
However
overtemperature
be
initiated
failure
Temperature
temperature
erature
disconnected,
alarm
latch
by
the
are
two
is
in
the
monitors
goes
be
muted.
there
which
Alarm
which
so
that
incubator
Service
Overtemperature
microprocessor
its
own
readings;
over
37.5°C
cause
the
and
sound
is
also a totally
detection
if
there
is a microprocessor
causes
to
the
cannot
it
the
go
up
to
Hardware
will
cause
will
sound a steady
be
muted,
cannot
being
be
tumed
Manual
control
if
it
will
heaters
an
alarm
independent
circuit.
Incubator
38.0°C.
Overtemp-
the
heaters
and
reset
other
off.
Alarms.
system
the
to
be
which
This
At
this
to
Audible
will
than
will
be
The
added
by
sponge
factors:
+
+
e
e
*
amount
to
adding
holders
The
relative
drawn
Whether
The
temperature
the
sponges.
The
Incubator
The
length
to
build
of
moisture
the
circulating
water
is
humidity
in
through
(dry)
of
up.
which
air
in
to
the
sponges
controlled
of
the
the
Fresh
oxygen
Temperature.
time
of
the
allowed
is
being
water
will
the
by
several
air
Air
for
'
in
be
incubator
the
being
Filter.
applied.
added
to
humidity
Note
cause
that
the
both
Overtemperature
heaters
to
be
disconnected.
Alarms
TS2
Thermocot
Service
Manual
Warning
statements
(A
is a possibility
operator.)
WARNING
of a source
or
cause
WARNING
detailed
Incubator
any
withdrawn
WARNING
should
operation
qualified
the
equipment.
WARNING
presence
possible
these
WARNING
in
patient's
minutes.
WARNING
move
When
loss.
WARNING
the
Warning
an
portion
known
the
incubator
Statement
of
Do
not
place
of
radiant
electric
the infant
Complete
All
Do
conditions.
Do
incubator
The
the
Do
radiator.
in
this
into
service.
of
the
from
personnel
be
thoroughly
and
under
medical
risks
not
use
of
flammable
explosion
not
leave
temperature
canopy
infant
it
is
raised
not
leave a patient
while
the
Manual
is
injury
to
the
incubator
heat
Such
to
overheat.
Checkout
before
If
procedure
service.
using
familiar
the
personnel
and
benefits
the
incubator
anaesthetics;
hazard
the
patient
for
long
at
should
in
or
out
there
is
the
used
when
the
patient
in
such
as
exposure
Procedure
placing
the
incubator
it
must
the
incubator
with
direction
familiar
of
using
in
exists
under
unattended
periods.
least
only
of
significant
canopy
Check
every
be
raised
the
incubator.
unattended
is
there
or
the
path
sunlight
may
the
fails
be
its
of
with
the
the
a
the
30
to
heat
raised.
the
as
in
WARNING
does
pressure
measurements
the
concentrations
environment
concentrations
retrolental
below
infants.
WARNING
measured
flowmeter.
should
infant's
WARNING
the
service
incubator
WARNING
the
WARNING
increase
incubator.
WARNING
dependent
air
openings
mattress.
proper
air
build-up.
triggered.
The
concentration
not
predictably
of
oxygen
regulation
40%
If
be
When a system
incubator
If
patient.
The
Proper
circulation.
air
temperature
of
is
fibroplasia.
can
oxygen
with a pressure
The
monitored
head.
should
and
the
to
ensure
an
alarm
administration
the
noise
temperature
upon
Do
at
the
Obstruction
circulation,
An
air
of
inspired
determine
in
the
blood;
are
extremely
inspired
when
necessary.
can
be
is
applied
concentration
infant
is
level
continuous
not
ends
control
circulation
oxygen
an
oxygen
High
increase
Even
dangerous
its
compensated
at a position
fail
alarm
be
removed
shifted
continuity
muted
closely
of
within
control
cover
and
sides
could
loss
of
and
alarm
oxygen
the
partial
blood
gas
important
enriched
oxygen
the
risk
of
concentrations
to
some
flow
should
of
oxygen
near
the
is
activated,
from
to
another
of
care.
monitor
oxygen
unobstructed
air
lead
heat,
carbon
will
the
is
circulation
of
the
to
loss
of
loss
of
dioxide
may
be
for
be
TS2
Thermocot
Service
Manual
Caution
Statements
{A
CAUTION
CAUTION
CAUTION
caution
a
possibility
incandescent
phototherapy)
canopy.
least
ether,
scouring
incubator.
the
CAUTION
of
the
CAUTION
electricity
canopy
CAUTION
cleaning
least
to
cool.
CAUTION
for
cleaning
has
amounts
left
in
particularly
statement
of
Do
not
mount a radiant
Thoroughly
once a week
Do
not
use
alcohols,
compounds
These
surfaces
Do
To
When
10
If
been
to
not
autoclave
plastic
minimise
do
with a dry
disassembling
disconnect
minutes
flammable
make
thoroughly
of
agents
the
incubator
in
damage
light
on
or
clean
organic
ketones
compounds
which
parts
the
not
polish
for
agents
sure
the
presence
is
used
to
source
over
the
or
after
solvents
(e.g.
for
cleaning
they
the
of
the
generation
the
cloth.
the
power
the
heating
that
air
dried.
such
as
can
cause a fire
the
warmer
(including
the
incubator
each
are
mattress
incubator.
incubator
the
have
the
alcohol
of
when
there
equipment.)
or
incubator
at
patient.
including
acetone)
may
incubator
been
Small
oxygen.
or
the
damage
applied.
or
any
of
static
and
allow
elements
used
incubator
or
ether
is
for
at
CAUTION
with
this
readily
burn
on
CAUTION
automatically
the
is
period
prevent
The
every 6 months.
CAUTION
sponge
free
sponges
water
CAUTION
20Kg.
two
either
CAUTION
danger
may
in
Never
other
type
and
violently.
suitable
The
incubator
to
he
taken
the
possible
battery
When
holder
liquid
is
is
spilled
The
It
should
people
end.
The
of
produce
the
incubator.
oil
or
grease
than a substance
of
service.
in
lubricants.
voltage
battery
should
applying
that
at
Thermocot
and
use
fire.
Oils
the
presence
CIG
Health
of
the
checked
is
turned
out
of
service
should
damage
be
water
be
careful
is
not
soaked
least
5mm
mop
it
weighs
only
be
only
by
of
oxygen
Auxiliary
sparks
should
continually
on.
routinely
up
lifted
oxygen
and
internal
that
below
the
increases
equipment
equipment
approved
grease
of
oxygen
Care
can
battery
If
the
incubator
for
an
be
removed
due
to
leaking.
replaced
to
the
humidity
the
level
up
into
the
immediately.
in
excess
by
at
handles
not
be
for
oxidise
will
advise
is
when
extended
to
of
the
lip.
If
of
least
at
the
which
placed
TS2
Thermocot
Service
Manual
Checkout
Procedure
WARNING
while a patient
WARNING
before
fails
it
should
repaired.
Mechanical
WARNING
outlet
1.
Inspect
2.
Examine
3.
Tilt
remove
hinge
4.
Undo
mattress
5.
Remove
the
impeller
place.
the
two
shaft
retaining
6.
Replace
Baffle
with
position
for
Do
not
perform a checkout
is
Complete
putting
any
portion
Disconnect
before
the
the
canopy
the
pins
the
the
two
sensors,
are
Verify
presence
sensors, 8 heater
hole.
all
and
its
four
if
cleanliness
the
the
of
be
removed
Checks
the
disassembling
power
the
incubator
back
mattress
are
screwed
four
screws
tray.
Remove
baffle
the
undamaged
cleanliness
of
rubber
Check
that
screws
are
parts
securely,
Mattress
screws.
extra
humidity
of
all
in
the
incubator.
checkout
incubator
the
checkout
from
power
lead.
for
so
that
tray.
completely
which
the
and
check
two
heaters
and
of
seals
holes
the
two
firmly
tray,
securing
Sponges
is
parts.
procedure
procedure
into
use.
procedure
service
signs
Check
mattress
all
in
including
and
lead
from
the
incubator.
of
damage.
it
is
possible
that
home.
secure
that
the
and
securely
parts.
around
and
the fan
liner
place.
the
must
be
required.
If
it
its
the
the
tray.
liner,
the
in
Check
the
the
latter
in
Check
to
8.
9.
10.
be
mesh
this
centre
head
rotor
When
night
efficiency.
incubator.)
rotor
replace
Close
the
canopy
doors.
They
when
the
catches
otherwise
inside
Check
Check
inlet
Check
filter.
in
place.
In a clean
operating
remain
However
continuous
environment,
effective
The
dust
observed
withstand a force
the
incubator
that
all
for
smooth
rotor.
the
condition
It
should
environment
conditions a new
effective
adverse
operation
life.
filter
will
from
the
or
by
removing
peel
off
the
of
the
screw
which
and
filter
replacing a filter
way
round
(The
Tighten
is
stiff
to
the
cover
and
should
are
screws
operation
be
clean
for
conditions,
will
shorten
darken
incoming
through
blue
rotor;
underneath
retains
assembly.
to
achieve
label
turn
(but
strip.
test
the
two
open
automatically
pressed
without
are
of
the
and
about 6 months.
in a dusty
in
the
the
cover
make
must
the
screw
of
opening.
tight.
of
fresh
and
securely
under
filter
the
colour
air.
This
inlet
rotor
strip
the
sure
optimum
face
not
too
but
20N
the
air
such
filter's
as
colour
protective
plate.
in
is a Philips
complete
into
so
stiff)
porthole
should
from
fresh
inlet
held
normal
will
as
it
takes
To
the
it
is
the
the
that
the
and
air
up
can
do
7.
Close
the
canopy
ensuring
canopy
position
that
can
only
by
deliberately
the
to
latch
be
moved
its
mid-way
is
releasing
secure
from
position
and
the
that
the
catch.
Replacement
5
under
part
pack
includes
filters
number
gaskets
are
available
NWT900900500.
and
blue
in
packs
cover
of
The
strips.
TS2
Thermocot
Service
Manual
Operational
TA
Fs
Fs
Ta
т,
1.
Following
make
place
air
filter.
open
Put
the
position.
unconnected
audible
Turn
power
Put
the
position.
of
all
into
the
light
should
light
should
other
be
off.
There
alarm
Mode
cancelled
should
There
mode.
should
should
been
Push
the
Operational
of
all
displays
that
all
digital
Checks
the
sure
that
inchiding
Put
the
position.
main
ON/OFF
If
the
there
Power
OFF
the
lead.
main
ON/OFF
There
displays
STANDBY
be
be
lights
and
should
to
having
should
repaired.
be
draw
been
by
pushing
come
on.
be
If
there
be
investigated
not
be
used
ON
button
mode.
should
lamps
and
display
above
mechanical
all
parts
the
mattress
fresh
switch
power
lead
should
Fail
alarm.
incubator
switch
should
before
the
mode - the
on
flashing
on
but
there
the
digital
an
intermittent
attention
selected
MUTE.
no
alarms
are
any alarms,
until
to
Again
be
all
segments
are
illuminated.
checks
are
securely
and
the
air
inlet
rotor
to
the
is
still
be a continuous
and
connect
to
the
be
an
automatic
incubator
Standby
and
the
should
display
audible
to
the
Standby
and
this
No
Mute
in
the
Standby
their
and
the
incubator
the
condition
enter
the
an
automatic
initiated.
Check
of
the
in
fresh
in
the
ON
the
ON
test
goes
heater
be
no
should
can
light
cause
has
test
be
Push
display
light
Push
Set
Release
of
the v (down)
Note
selected.
incubator
ON/OFF
fail
and
in
the
that
was
10.
Push
displays
Alarm
11.
Tum
switch.
should
Select
temperature - it
selected
12.
Allow
stabilised
This
blinking.
at
this
alarms.
13.
Select
and
60
minutes.
alarms.
the A (up)
the
Set
should
be
the
button
temperature
the
button
the
Set
temperature
Tum
at
the
switch.
alarm.
Restore
check
that
Operational
the
Set
temperature
selected.
the
TEST
should
should
off
Operational
the
will
temperature.
the
allow
the
incubator
After
come
in
9.
incubator
at
the
be
indicated
Run
maximum
the
Again,
sound
15
on
the
incubator
button
temperature.
off
button.
off
the
button.
be
in
should
momentarily
and
the
continuously
up
in
0.1
increments.
and
check
you
the
power
power
source,
There
should
power
after
incubator
mode
(not
is
All
briefly
lit.
briefly.
at
the
minutes
selected
incubator
turn
Standby
mode
and
be
to
run
Set
by
the
There
temperature,
to
there
should
The
“Actual”
“Set”
light
to
move
the
function
have
to
the
not
at
be a power
15
seconds,
has
resumed
Standby)
the
same
lamps
and
The
Audible
ON/OFF
it
on
again;
mode.
check
the
the
same
until
it
has
temperature.
Heater
for
60
minutes
should
be
37.0°C,
run
for
another
be
no
to
on.
the
its
and
as
Set
as
was
light
no
it
The
digital
display
air
temperature,
on,
and
the
There
should
the
Actual
be
no
should
heater
light
above
alarms.
show
light
incubator
should
the
display.
be
TS2
Thermocot
Service
Manual
Electronics
1.
A
technician
facility
the
Diagnostic
accessible
beneath
shown
to
verify
the
on
Check
can
correct
push-button
through
circuit
the
front
use
the
performance.
the
slot
breakers.
panel
self-diagnostic
Use
which
is
in
the
rear
panel
The
results
digital
display.
are
Diagnostics
For 1 push
2
|
lola
From
any
either
Display
Set
Air
Airtemp
Air
Input
Heater
Heater
Internal
selection
the A (up)
temp
temp © #1
@
temp
Voltage
temp
temp
Battery
(9
Volt
nominal)
Fan
Speed
[Acceptable
Le.20
rps
to
or
1200
rpm
retum
or v (down)
Sensor
#5
Sensor
#2
Sensor
(rectified
#3
Sensor
#4
Sensor
Voltage
(display
range
is
35
rps
to
2100
to
normal
button.
(return
(ambient)
(return
D.C.)
is 4 x
80
rpm]
by
air)
air)
revs/sec)
to
140
pushing
.
TS2
Thermocot
Service
Front
Canopy
PEN:
Manual
Catch
SE
Infant
Restraint
Harness
O
On/Off
“> “>
switch
Fresh
air
mlet
Control
panel
«ad
db
Oxygen
inlet
nipple
チン
TS2
Thermocot
Service
Manual
CAUTION
HIGH
OXYGEN
To
increase
corcentrator.
Oxygen
(via
40%
ar
oxygen
No
oxygen
appty
Te)
at
ripple
at
det
enc:
With
oxyger:
increased
by
closing
inlet
using
analyser
not
close
nel
no
sources
near
applied.
may.
over.
fresh,
ah
trash
oxygen
of
concentrations
be
Mondor
Do
ar
anche
Anion
neubator
Re;
ee
ее
4
」
7
a
-
RR
9
DED.
^
22
ys
の
2%
^
ЕВ:
ЕР
~
ag
ge
9
ON
Le 2 s
%
o
Е
-
~
5
42200
OE
GUS,
~
E
SINS
SØG
2
TR x eee
4
の
ASE
20
a
paro
>
DİM
22
295
ASG).
À
&
TS2
Thermocot
Service
Manual
Cleaning
WARNING
WARNING
CAUTION
CAUTION
CAUTION
Disconnect
outlet
Allow
incubator
handling
Do
ether,
scouring
incubator.
the
surfaces
To
electricity
dry
cloth.
If
for
cleaning,
been
thoroughly
assembly.
Disassembly
1.
Make
sure
before
disassembling
at
least
has
been
the
heaters.
not
use
alcohols,
compounds
These
to
minimise
do
not
flammable
make
Procedure
the
incubator
the
10
switched
organic
ketones
compounds
which
the
polish
agents
sure
air
dried
power
for
lead
the
minutes
off
solvents
(e.g.
acetone)
cleaning
may
they
are
generation
the
canopy
have
been
that
all
before
is
turned
from
its
incubator.
after
the
before
including
or
the
damage
applied.
of
static
with
used
parts
have
re-
off.
a
Check
to
touch.
The
the
7mm
Cleaning
10.
All
removed
only
Do
not
coarse
The
clean.
The
the
two
immersed
wipe
11.
The
brush
antiseptic
Do
not
coarse
that
the
Remove
liner
may
two
retaining
Nut-driver
parts
antiseptic
use
alcohols,
polish.
mattress
Do
not
Mattress
heaters
if
necessary.
clean.
canopy
may
or
cloth
soap
use
alcohols,
polishes.
two
heaters
both
be
removed
screws
is
required.
should
soap
can
be
immerse
Tray,
the
and
be
scrubbed
and
wiped
and
warm
ether,
heaters.
for
are
be
and
warm
ether
or
scrubbed
the
mattress.
Baffle,
the
Liner
Scrub
clean.
water.
ketones
are
not
too
cleaning
removed.
cleaned
ketones,
them
with a soft
water.
and
the
Impeller,
can
all
Use
using
wiped
and
only
or
hot
if
A
or
be
an
2.
Disconnect
outlet.
3.
Open
4.
Undo
tray.
mattress.
5.
Remove
tray.
6.
Remove
7.
Remove
the
power
the
canopy
the 4 screws
Remove
the
the
the
the
sponges
baffle.
impeller
lead
to
its
retaining
tray,
the
from
from
from
fully
screws
the
the
the
back
position.
the
mattress
mattress
motor
power
and
shaft.
the
With a slightly
seals
fan
13.
Remove
and
needs
discoloured a new
Wipe
slightly
14.
With a damp
contains
or
similar
tube
on
the
shaft
to
remove
the
Air
the
filter.
replacing - if
the
rotor
damp
the
technique,
and
its
top
damp
sensors,
any
inlet
Check
it
filter
and
filter
cloth.
cloth,
wipe
filter,
and,
gasket.
cloth
wipe
the
heaters
deposits.
rotor,
the
filter
the
filter
to
is
excessively
is
required.
panel
with
the
area
using a bottle
the
air
entrainment
the
and
see
a
which
rubber
the
panel,
if
it
brush