C&H Technologies PX454S User Manual

C&H Technologies, Inc. 445 Round Rock West Drive Round Rock, TX 78681-5012
(512) 733-2621 FAX (512) 733-2629 www.chtech.com
Resolution:
10 Hz sampling: 22.5 bits 50/60 Hz sampling: 20 bits 1 KHz sampling: 11 bits
Input Connector: 5W5S DSUB
Airflow (min): 10m3/h
Temperature:
Operating: 0°C to 60°C Storage: -20°C to 70°C
Power: +5V @ 1.1 A with DC/DC
CPCI/PXI Compliance
Complies with PCI spec. 2.0 R3.0 and PCI spec 2.2
5V and 3.3V signaling voltage (VIO) supported
5V only power supply
33MHz PCI data bus
Five trigger lines compliant with PXI Specification 2.1
Form Factor: Size 3U
Applications
DMM Instrument Functions
Factory Test
Ordering Information
Part Number 11030230-0001
Additional Information
User Manuals for C&H carriers and this module can be found on our website at
www.chtech.com
.
Specifications:
General Characteristics:
DC or AC voltage and current
2 or 4 wire resistance measurement
Up to 22 bits accuracy (6.5 digits)
Max. resolution of 50nV, 5nA, 50_Ω
DC Voltage: ±125 mV to ±500V
DC Current: ±12.5mA to ±2.5A
AC V
rms
: 250mV to 250V
AC I
rms
: 25mA to 2.5A
Ohms: 0 to 250Ω to 2.5MΩ
True RMS measurements
Auto calibration
PX454S Digital Multimeter Module
The PX454S is a full function DMM suitable for use on a variety of platforms with the family of C&H carriers. This DMM is based upon a 24 bit Sigma-Delta A/D converter with on-chip notch filter. This technology allows sampling frequencies (or filter frequencies) between 10 Hz and 1 KHz providing varying resolutions. The core measurement section is optically isolated and shielded.
This unit also features a position for adding one single-wide M-Module.
C&H Technologies, Inc. 445 Round Rock West Drive Round Rock, TX 78681-5012
(512) 733-2621 FAX (512) 733-2629 www.chtech.com
DC Voltage Measurement:
Maximum Voltage Input:
HV Input: 500VDC for 500V range
Normal Input: 125VDC for other ranges
Measuring Ranges:
• ±125mV, ±1.25V, ±12.5V, ±125V, ±500V
Measuring Accuracy:
24 hours, 23°C ±1°C: error < 0.025%
90 days, 23°C ±5°C: error < 0.05%
Temp Coefficient, 0 to +55°C: error < 0.01%
Input Impedance:
500V input (HV) 40 MΩ
125V and 12.5V ranges 10 MΩ
1.25V and 125 mV ranges >1 GΩ
Input Current: 10 pA max @ 23°C
CMRR: 100dB @ 50/60 Hz
NMRR: 60 dB @ 50/60 Hz
AC Voltage Measurement:
Maximum Input Voltage:
High Voltage Input: 250V AC
Normal Input: 60V AC
Measuring Ranges:
250mVrms, sine 40 Hz-100KHz: error < 5%
2.5Vrms, sine 40 Hz-100KHz: error < 1%
25Vrms, sine 40 Hz-100KHz: error < 1%
250Vrms (HV in) sine 40 Hz-100KHz: error < 1%
Crest Factor: 4 max
Measuring Accuracy @ 50/60 Hz:
24 hours, 23°C ±1°C: error < 0.5%
90 days, 23°C ±5°C: error < 1.0%
Range Error: < 1%
Input Impedance:
High-voltage input: 40 MΩ
Normal Input: 1 MΩ
Input Current: 10 pA max @ 23°C
CMRR: 100dB @ 50/60 Hz
DC Current Measurement:
Maximum Voltage: 60 V on all ranges
Measuring Ranges: ±12.5mA, ±125mA, ±1.25A, ±2.5A
Measuring Accuracy:
24 hours, 23°C ±1°C: error < 0.025%
90 days, 23°C ±5°C: error < 0.05%
Temperature Coefficient, 0 to +55°C: error < 0.01%
Shunt:
12.5mA and 125mA ranges: 1 Ω
1.25A and 2.5A ranges: 0.1 Ω
AC Current Measurement:
Maximum Voltage: 42 V on all ranges
Measuring Ranges: 25mA, 250mA, 2.5A all RMS
Crest Factor: 4 max
Measuring Accuracy:
24 hours, 23°C ±1°C: error < 0.5%
90 days, 23°C ±5°C: error < 1.0%
Shunt:
25mA and 250mA ranges: 1 Ω
2.5A ranges: 0.1 Ω
Resistance Measurement:
Measuring Ranges: 250_, 2.5K_, 25K_, 250K_, 2.5M_
Measuring Accuracy (All but 2.5M_ range):
24 hours, 23°C ±1°C: error < 0.1%
90 days, 23°C ±5°C: error < 0.2%
Temperature Coefficient, 0 to 55°C: error < 1%
Measuring Accuracy (For 2.5M_ range):
24 hours, 23°C ±1°C: error < 1.0%
90 days, 23°C ±5°C: error < 2.0%
Temperature Coefficient, 0 to 55°C: error < 0.1%
Measuring Modes: 2- or 4-wire
Maximum Measuring Voltage: 2.5V
Loading...