C&H Technologies EM405D100, EM405-8x106, EM405-8x107 Specification Sheet

C&H Technologies, Inc. 445 Round Rock West Drive Round Rock, TX 78681-5012
(512) 733-2621 FAX (512) 733-2629 www.chtech.com
Sampling Strobe:
Internal: to 5 MHz
Backplane Trigger: to 5 MHz
Source and prescaler software
programmable
Interrupts:
Data Stored
FIFO Half-Full
FIFO Full
Time-Stamp Rollover
Change of State
Level Transition
Bit Pattern
I/O Connectors:
44 pin DSUB
Indicators: Power
Module Access
LAN Status
Temperature:
Operating 0° C to 50° C Storage -40° C to 70° C
Networking
Easy network configuration through web based interface
TCP/IP and UDP/IP supported for M-module communication
Static IP, DHCP, and AutoIP addressing supported
Ethernet 10Base-T or 100Base­TX (Auto Sensing)
Applications
Acquisition of transient signals
Real-time data compression
Logic Analysis
Event counting
Additional Information
User Manuals for this carrier and C&H M-modules can be found on our website at www.chtech.com.
Specifications:
Number of Channels: 32, 64 or 128
Sample Rate: up to 5 Msps
Input Debounce: 0 to 128 ms
Software selectable
Input Specifications:
Threshold Resolution 100mV Threshold Range 0 to 25.6V Threshold Accuracy ±0.5% of FS Max. Input Voltage 48V Input Hysteresis 25mV Input Impedance >100K_ Input Current <300_A
Input Masking:
Allows selection of bits to be monitored
Local Memory:
32K deep (data and 31–bit timestamp)
Input Polarity: Software Selectable
Ethernet Event Detector Instruments
The EM405D100, EM405-8x106 and EM405-8x107 are 32, 64, and 128 channel input modules that sample and selectively store up to 128 bits of data along with a 31 bit time tag at rates up to 5 MSPS. These instruments have the ability to store all data at the specified sample rate or to selectively store input values based upon changes in state on one or more of the inputs. Each version is 1U high with the 32 channel unit designed for bench applications and the 64 and 128 channel version for rack mounting.
Specifications:
Implementation:
All versions use either two (2), four (4) or eight (8) ANSI/VITA Standard 16 Channel M-Modules (MA203), each of which have the block diagram shown above
See the EM405D and EM405-8 on C&H web site for further information on the instrument enclosure mounting and dimensions.
Trigger/Sample Strobe Features:
An external clock may be routed through one mezzanine to a carrier trigger and back to the Event Detector triggers to minimize skew between channels
One mezzanine’s clock may be used as a master in a similar manner
External Trigger Inputs:
32 Channel Unit TTL
64 & 128 Channel Units TTL and LVDS
External Power (32 Channel Unit):
AC Input 100 to 240VAC, 47 to 63Hz with furnished adapter
Alternatively DC powered by 12V, 1A
External Power (64 and 128 Channel Units):
AC Input 85 to 265VAC, 47 to 63Hz
Alternatively DC Powered by 36 to 75V
Software Driver:
Windows DLL
Soft Front Panel Executable
Online Help
LabWindows/CVI Function Panels
ANSI-C Source Code
LabView
Conformance:
These modules will be CE compliant for Emissions, Immunity and Safety
Ordering Information:
32 Channel Event Detector:
Model EM405D100 P/N 11030010-0001
64 Channel Event Detector:
Model EM405-8x106 P/N 11030010-0002
128 Channel Event Detector:
Model EM405-8x107 P/N 11030010-0003
DEBOUNCE
M-MODULE
I/F
TIMER
31-BIT
SAMPLE
AND
STORE
LOGIC
FIFO
RESET
PROGRAM
TRIGGER A OR BEXTCLK
MEMORY
48-BIT
CONTROL
STORE
FRONT
PANEL I/O
WATCH
MASK
CLOCK
10KHz
- 5MHz
INTERRUPT
LOGIC
PROGRAM
CONTROL
RUN
SCLK
TRIGGER A OR B
SW
EXTRUN
SIGNAL
CONDITIONING
ACCESSORY
MODULE
ISOLATED
DC-DC
CONVERTER
INTERRUPT
PRESCALER
TRIGGER A OR B
INPUT
SIGNALS
Loading...