The information in this document is subject to change without notice.
Chroma ATE INC. makes no warranty of any kind with regard to this manual, including, but
not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Chroma ATE INC. shall not be held liable for errors contained herein or direct, indirect, special,
incidental or consequential damages in connection with the furnishing, performance, or use of
this material.
CHROMA ATE INC.
66 Huaya 1st Road, Guishan, Taoyuan 33383, Taiwan
Copyright Notices. Copyright 2008 Chroma ATE INC., all rights reserved. Reproduction,
adaptation, or translation of this document without prior written permission is prohibited,
except as allowed under the copyright laws.
iii
Warranty
All of Chroma’s instruments are warranted against defects in material and workmanship for a
period of one year from date of shipment. Chroma agrees to repair or replace any assembly
or component found to be defective, under normal use during this period. Chroma’s obligation
under this warranty is limited solely to repairing any such instrument, which in Chroma’s sole
opinion proves to be defective within the scope of the warranty when returned to the factory or
to an authorized service center. Purchaser is responsible for the shipping and cost of the
service item to Chroma factory or service center. Shipment should not be made without prior
authorization by Chroma.
This warranty does not apply to any products repaired or altered by persons not authorized by
Chroma, or not in accordance with instructions furnished by Chroma. If the instrument is
defective as a result of misuse, improper repair, or abnormal conditions or operations, repairs
will be billed at cost.
Chroma assumes no responsibility for its product being used in a hazardous or dangerous
manner either alone or in conjunction with other equipment. High voltage used in some
instruments may be dangerous if misused. Special disclaimers apply to these instruments.
Chroma assumes no liability for secondary charges or consequential damages and in any
event, Chroma’s liability for breach of warranty under any contract or otherwise, shall not
exceed the purchase price of the specific instrument shipped and against which a claim is
made.
Any recommendations made by Chroma regarding the use of its products are based upon
tests believed to be reliable; Chroma makes no warranty of the results to be obtained. This
warranty is in lieu of all other warranties, expressed or implied, and no representative or
person is authorized to represent or assume for Chroma any liability in connection with the
sale of our products other than set forth herein.
“O” indicates that the level of the specified chemical substance is less than the threshold level
specified in the standards of SJ/T
“
specified in the standards of SJ/T
Remarks:
Directive 2011/65/EU.
The recycling label shown on the product indicates the Hazardous Substances contained in
the product as the table listed below.
: See <Table 1>.
: See <Table 2>.
<Table 1>
Part Name
PCBA
CHASSIS
ACCESSORY
PACKAGE
Chromium
O O O O O O
O O O O O O
O O O O O O
O O O O O O
Biphenyls/
Group
-11363-2006 and EU Directive 2011/65/EU.
” indicates that the level of the specified chemical substance exceeds the threshold level
-11363-2006 and EU Directive 2011/65/EU.
The CE marking on product is a declaration of product compliance with EU
Disposal
Do not dispose of electrical appliances as unsorted municipal waste, use separate collection
facilities. Contact your local government for information regarding the collection systems
available. If electrical appliances are disposed of in landfills or dumps, hazardous substances
can leak into the groundwater and get into the food chain, damaging your health and
well-being. When replacing old appliances with new one, the retailer is legally obligated to
take back your old appliances for disposal at least for free of charge.
v
<Table 2>
Hazardous Substances
Lead
Mercury
Cadmium
Hexavalent
Polybrominated
Polybromodiphenyl
Ethers
Selected Phthalates
Pb
Hg
Cd
Cr6+
PBB/PBDE
DEHP/BBP/DBP/DIBP
“O” indicates that the level of the specified chemical substance is less than the threshold level
specified in the standards of SJ/T
“
specified in the standards of SJ/T
1.
2.
environment specified in each product’s specification.
Part Name
PCBA
CHASSIS
ACCESSORY
PACKAGE
Chromium
O O O O O
O O O O O
O O O O O
O O O O O O
Biphenyls/
Group
-11363-2006 and EU Directive 2011/65/EU..
” indicates that the level of the specified chemical substance exceeds the threshold level
-11363-2006 and EU Directive 2011/65/EU..
Chroma is not fully transitioned to lead-free solder assembly at this moment; however,
most of the components used are RoHS compliant.
The environment-friendly usage period of the product is assumed under the operating
Disposal
Do not dispose of electrical appliances as unsorted municipal waste, use separate collection
facilities. Contact your local government for information regarding the collection systems
available. If electrical appliances are disposed of in landfills or dumps, hazardous substances
can leak into the groundwater and get into the food chain, damaging your health and
well-being. When replacing old appliances with new one, the retailer is legally obligated to
take back your old appliances for disposal at least for free of charge.
vi
vii
Unpacking for Check & Inspection
Item Name
Qty
Description
USA standard 180 degree straight head power cord with the
length 1.8 meter, 15A
15A Fuse
2
15A SLOW
PHONE CABLE 6P6C to connect multiple devices, length 1
meter
D-SUB-25P-MALE*2 to connection multiple devices, length 1
meter
Single head high voltage terminal + 20kV high voltage cable,
length 3.1 meters
RTN/LOW Cable
Note 2
Single head BNC(MALE)+RG-174, length 3 meters
Quick Start Guide
2
One English version and one Traditional Chinese version.
User’s Manual CD
1
CD for user’s manuals in English and Traditional Chinese
Note
1. When additional item is required, just inform Chroma the item name.
the cable quantity for 10CH model is 10 sets and for 4CH mode is 4 sets.
Before shipment, this instrument was inspected and found to be free of mechanical and
electrical defects. As soon as the instrument is unpacked, inspect for any damage that may
have occurred in transit. Save all packing materials in case the instrument has to be returned.
If damage is found, please file claim with carrier immediately. Do not return the instrument to
Chroma without prior approval.
Standard Package
US Power Cord 1
CANBUS Cable 1
D-SUB Cable 1
High Voltage Cable
2. The cable quantity is varied with the output channel configured, for instance,
Note 2
viii
Hazard Operation Methods
Please see Chapter 3 Precaution before Use in this manual for detail
descriptions of usage notices and operation hazards.
Do not tie up the high voltage cable with RS232, Handler and GPIB control
CAUTION
WARNING
1. Do not touch the testing area when this Hipot Tester is outputting voltage or you may get
electric shock and it may cause death.
Be sure to obey the following:
․ The earth wire must be connected exactly and use a standard power cord.
․ Do not touch the output terminal.
․ Do not touch the test wire that connected to the terminal in test.
․ Do not touch any unit under test.
․ Do not touch any component that connected to output terminal for charge.
․ Do not touch the test unit right after the test is ended or when the output is just turned
off.
2. The electric shock incident may occur when:
․ The earth terminal of Hipot Tester is not connected properly.
․ The insulating gloves are not in use during test.
․ Users touch the test unit right after the test is done.
cables or other low voltage side wires. If so, it could cause the product or
PC to be down or damaged.
ix
Storage, Freight, Maintenance & Cleaning
Storage
When not in use, please pack the device properly and store in a suitable environment.
Freight
Please pack the device carefully before moving it. If any of the original packing material is
missing, please use suitable alternative material and mark it “fragile” and “keep away from
water” to avoid damaging to the product.
This product is a piece of precision test equipment, so please do not drop or hit it.
Maintenance
In case of any malfunction or abnormality, please refer to the manual, or contact our local
distributor for prompt service. Do not touch any parts inside the instrument to avoid any
danger to yourself or damage to the product.
Regular check and calibration is required to ensure the device meets the product specification.
The suggested calibration period every 12 months.
Cleaning
Remove all connected wires and cables on the instrument before cleaning. Use a brush to
clean the dust on it. For internal cleaning, use a low-pressure air gun to vacuum the dust
inside or send it back to the distributors or agents of Chroma for cleaning.
x
Revision History
Date
Version
Revised Sections
Nov. 2008
1.0
Complete this manual.
June 2009
1.1
Add “ACA MEAS.” in the section of “Setting SYSTEM CONFIG” under
(IEEE-488.2).”
Nov. 2009
1.2
Update the values in “Specification” chapter and the figures of
section.
June 2010
1.3
Add two new items in the chapter of “Precaution before Use”.
Dec. 2010
1.4
Update “Material Contents Declaration.”
May 2011
1.5
Add descriptions for 19020-4 & 19022-4 two new models.
Sep. 2011
1.6
Update “Standard Package” and its Note in “Unpacking for Check & Inspection”.
Dec. 2012
1.7
Add “CE Declaration of Comformity”.
Update “Pin Assignment“ in the chapter of “HANDLER Interface”.
Aug. 2013
1.8
Update the following:
chapter of “GPIB/RS232 Interface”
Oct. 2014
1.9
Add PSC Mode to DC, IR and OSC tests procedures and commands.
Aug. 2015
2.0
Add descriptions for new model 19021-4.
Mar. 2016
2.1
Modify the following sections:
chapter of “Calibration Procedure.”
Aug. 2016
2.2
Update “CE Declaration of Conformity”.
Apr. 2017
2.3
Update “Material Contents Declaration” and “CE Declaration of Conformity”.
The following lists the additions, deletions and modifications in this manual at each revision.
the chapter of “Operation.”
Update the commands in the chapter of “GPIB/RS232 Interface
“Example of Using Internal Power Supply” as well as of “Example of Using External Power Supply” in “Example of External Control Circuit”
– “Setting TEST CONTROL”, “Setting SYSTEM CONFIG” and
“Selecting Test Mode” sections in the chapter of “Operation”
– “Commands Summary” and “Command Description” in the
– “Features” in the chapter of “Preface.”
– Leakage current meter spec. and note description in the chapter
of “Specification.”
– “Front Panel”, “Rear Panel” and “Setting TEST CONTROL” in the
chapter of “Operation.”
– “Commands Summary” and “Command Description” in the
chapter of “GPIB/RS232 Interface.”
– “ACA Current Calibration” and “AC ARC Calibration” in the
7.7.1 AC ARC Calibration ................................................................................7-22
7.7.2 DC ARC Calibration ................................................................................7-23
7.8 When Calibration is Done..............................................................................7-23
Preface
1-1
1. Preface
1.1 Product Overview
The 19020/19021/19022 Multi-Channel Sync Hipot Testers are high quality devices specially
designed to test the hipot leakage current and insulation resistance automatically for
electrical and electronic equipment.
For withstand voltage test, the output power of each channel for 19020/19020-4 is AC:
50VA(5kV, 10mA) /DC: 30VA(6kV, 5mA), for 19021/19021-4 is AC: 48VA(6kV, 8mA) and for
19022/19022-4 is DC: 28VA(8kV, 3.5mA). Thus they can be used to perform withstand
voltage tests on electronic, electrical equipment as well as on components.
For insulation resistance test, the range it can show is 0.1MΩ ∼ 50GΩ and the test voltage is
50V~1000V that can be set as desired. (Note: Only 19021/19021-4 can conduct AC
withstand voltage test.)
The Hipot Tester uses a clear display to show all settings, time, current, voltage, resistance
and memory channel no., etc. without the need to memorize the parameters set previously.
The Hipot Tester has equipped the device to identify pass or fail products as well as to output
signals of test result and to remote control other devices. It has GPIB, HANDLER and RS232
interfaces that are of advantage to automatic test system. This Tester equipped with assorted
devices mentioned above is capable of performing highly efficient and accurate tests for
electrical, electronic equipment and components.
1.2 Features
■Sync Withstand Voltage Test
The testers have the world’s first sync withstand voltage test function that a single device
can have 10 channels to output and test simultaneously and maximum 10 devices can
be controlled (master & slave) for 100 channels. They can be grouped for output to avoid
voltage difference from generation due to adjacent test and furthermore to improve the
production efficiency.
■Multiple Tests
The testers are able to select single test function such as AC withstand voltage test, DC
withstand voltage test, Insulation Resistance (IR) test and Open/Short Check (OSC) test
■OSC (Open/Short Check)
The testers have built-in OSC function to check if there is any Open (bad connection) or
Short (UUT shorted) occurred during test. It may cause the Fail product to be identified
as Pass if Open occurs. If Short is detected and resolved early when occurred, it can
reduce the damage to fixture and save the test cost.
The testers have a clearest design for display. All programs for settings such as test
voltage, current state, test readings, test steps and test states are able to be viewed from
the LED display directly.
■High/Low Limit Comparison for Pass/Fail Products
The testers have been designed to do High/Low Limit comparison for the Unit Under
Test (UUT.) This function is available in Withstand Voltage test or Insulation
Resistance test. The low limit comparison for hipot and high limit comparison for
insulation are used to test if there’s any bad connection or loose test wire that causes
misjudgment.
■Remote Control
The HANDLER of the testers is able to extend the control signals for controlling
externally. It usually connects to the control box of automated devices.
■Change Voltage Ramp Time
These testers have a [RAMP] function that can set the time required for voltage rises
from zero to set value.
■Change Voltage Fall Time
These testers have a [FALL] function that can set the time required for voltage falls from
set value to low when the test time ends.
■Auto Switch Low Current Range
The current meter ranges for withstand voltage test in the testers have two ranges, one
is low AC range 0~2.999mA and the other is 3.00 mA ~ maximum AC. There are three
ranges for DC: 0~299.9uA low current range, 0.300mA ~2.999mA current range and
3.00~ maximum AC. If the tested current is low, software can be used to switch the
current range to low range automatically for resolution improvement before the test ends
as need.
The 19021 and 19021-4 hipot testers have 4 AC ranges which are 30uA, 300uA,
3.000mA and 8.0mA. If the tested current is low, it can use software to set if
automatically changing the current range to low range before the test ends, so that the
resolution can be increased.
Specification
2-1
Maximum/Model
19020
19020-4
19021
19021-4
19022
19022-4
Output Channels
10CH
4CH
10CH
4CH
10CH
4CH
AC
5kV,10mA
6kV,8mA
-------
DC
6kV,5mA
-------
8kV,3.5mA
IR
1kV
-------
1kV
Multi-hipot unit
Output channels are defined by model, independent output
Master fixed the frame number on 0
Slave need to select frame number (1~9)
Withstanding Voltage Test
AC: 0.05-Maximum, steps 0.001kV, 50Hz/60Hz ± 0.1%, sine
<Note1>
V-monitor: ± (1% of reading + 0.1% of full scale), 2V
resolution
2 pins connector, pin1 pull-up to digital +V source with
4.7kohm resistor, and pin 2 tied to digital GND.
Ambient Temperature and Relative Humidity
Specifications range
18 to 28°C (64 to 82°F), ≤ 70% RH.
Operable range
0°C to 45°C, 15% to 95% RH@ ≤ 40°C and no
condensation.
Storage range
-10 to 60°C (-14 to 140°F), ≤ 80% RH.
Power Requirement
Line voltage
AC 100V~240V±10%
Frequency
47~63 Hz
Power consumption
Standby: < 250W
With rated load: <1100W
Dimension
428 W x 174 H x 600 D mm
Weight
Approx. 40kg
Safety
Ground bond
Less than 100mΩ at 25Amp, 2sec
Hi-Pot L + N to Earth
Less than 10mA at WVAC 1.5kV, 60Hz, 3sec no flashover
happen (ARC level < 8mA, tested by Chroma 19032)
Insulation L + N to Earth
Greater than 20MΩ at 500V dc, 2 sec.
Line Leakage Current
Less than 3.5mA at Vin max (132V at 120V selected
voltage), normal and reverse.
Note
1. When the testing time is less than 0.3 second, the output voltage specification is
DC ARC Validation point is 1.50kV with a 500kΩ resistor.
Specification
±(4% of setting + 0.2% of full scale)
It is necessary to calibrate again when the high voltage module on the rear
panel is replaced.
2. The accuracy of 30uA range is only valid for capacitive load.
3. AC ARC Validation point is 1.25kV with a 250kΩ resistor.
Precaution before Use
3-1
3. Precaution before Use
The Hipot Tester can output up to 8kV high voltage for external test. Accidents may occur or
even cause death if using this Tester incorrectly or in the wrong way. Thus for safety sake, be
sure to read the precautions in this chapter to avoid any accidents from happening.
1. High voltage module replacement
If users replace or switch the high voltage modules on the rear panel, to ensure the
Tester output is still within the specification, users need to recalibrate the Tester and
ensure the specification.
2. Electric shock
To prevent the incident of electric shock from occurring, it is suggested to wear the
insulated rubber gloves before using the Hipot Tester for electricity related tasks.
3. Grounding
A safety ground terminal is located at the rear of the Tester chassis; please use a proper
tool to ensure it is grounded accurately. If not it would be very dangerous when the
power circuit or the connection cable of any device shorts with ground terminal as the
chassis may contain high voltage. Anyone who touches the device in this case may
cause electric shock. Therefore, it is necessary to connect the safety ground terminal to
earth properly as the arrow shows in Figure 3-1.
Figure 3-1
4. Connecting test cable to LOW terminal
It is necessary to check if the test cable is connected all the time when the Tester is in
use. When connecting a test unit with test cable, connect the RTN/LOW test cable to the
UUT first. When the host RTN/LOW terminal is connected, it is very dangerous if the test
cable on RTN/LOW terminal is not connected correctly or falls as the entire UUT may full
of high voltage.
5. Connecting the test cable to high voltage output terminal
When the RTN/LOW test cable is connected, follow the steps below to connect the high
voltage output cable.
․ Press STOP.
․ Ensure the DANGER indicator is off.
․ Short the test cable of RTN/LOW and high voltage output to make sure there is no
voltage output.
․ Plug in the high voltage test cable to high voltage output terminal.
․ Last connect the RTN/LOW test cable to the unit under test and then connect high
voltage test cable.
6. End the test
When the test is end or the Tester is not in use or is in use but needs to leave it
unattended for a while, it is necessary to toggle the power switch to О (i.e. to shut off the
power) as Figure 3-2 shows.
Figure 3-2
7. Do not touch the hazard areas when the Tester is in test mode
When the Tester is in use, touching the object with high voltage such as UUT, test cable,
probe and output terminal is very dangerous.
8. Ensure the test is done
Sometimes users might need to touch the high voltage objects such as UUT, high
voltage test cable or output terminal etc. due to configuration or test required change. In
that case, please ensure the following:
∗ The power switch is turned off.
∗ The UUT may full of high voltage when completing the Insulation Resistance
test; thus it is necessary to follow the description of item 9 and 10 for
execution.
<<< Warning! Charging when doing insulation resistance test >>>
9. Charging
When doing insulation resistance test, the UUT, capacitor, test cable, probe and output
terminal, even the Tester itself may full of high voltage. The charged voltage may need
some time to discharge completely after turning off the power switch. It is necessary to
Precaution before Use
3-3
e
RC
t /
−
e
RCt /−
e
RCt
/−
follow the instruction described above for actions. Do not touch any area that may cause
electric shock especially when the power is just turned off.
10. Ensure the charged voltage is fully discharged
The time required for fully discharging the voltage depends on the test voltage applied
and the features of UUT. Assuming the high voltage added on the UUT equals the high
voltage added to a 0.01uF capacitor and paralleled to a 100MΩ resistance circuit. When
the test voltage is 1000V, then it requires approximately 3.5 seconds for the voltage that
added to test and on UUT to fall to 30V under after turned off the power. For 500V test
voltage, it requires about 2.8 seconds. Assuming the time constant of a UUT is already
known, the way described above can be used to calculate the time required for voltage
falling to 30V under after powered off by timing the time constant multiple to the time
decreased to 30V under as Figure 3-3 shows.
Calculation: Vo
Ex : 1000V ×
= 0.03
= VIL
= 30V
-t / RC = ln 0.03 ∴ t = 3.5 Sec
Figure 3-3
11. Remote controlling the Tester
The Hipot Tester can be remote controlled generally for high voltage output via external
control signal. When performing it, it is necessary to follow the control guidelines for
safety and precautions.
∗ Do not allow any accidental high voltage output that may cause hazard.
∗ When there is high voltage output from the Tester, do not allow any operator
or other personnel to touch the UUT, test cable or probe output and etc.
12. Turning on or off the power switch
Once the power switch is turned off, it needs to wait for a few seconds to turn it on again.
Do not power it on and off continuously to avoid occurring errors. It is very dangerous to
power it on and off continuously when in high voltage output state in particular. When
turning on or off the power, the high voltage output terminal cannot connect to
any object to avoid the hazard caused by abnormal high voltage output.
13. Other notices
Do not short-circuit the output line, grounding wire, communication cable or other
device’s grounding wire or AC power to avoid charging the entire Tester to dangerous
voltage. To short-circuit the terminal of high voltage output and LOW, it is necessary to
ground the Tester chassis to earth properly.
Be sure to use the fuse with correct specification or it may cause hazard.
CAUTION
User’s Manual
<<< Emergency Case >>>
14. Process for emergency case
To avoid causing bigger hazard when in emergency situations like electric shock, UUT or
Tester burnout, please perform the steps below:
∗ First cutoff the power switch.
∗ Second unplug the power cord.
<<< Resolving Problems>>>
15. Problems occurred
Problems occurred in the following situation are very dangerous. The output terminal
may still have high voltage output even the STOP key is pressed; therefore, users
should be extremely careful when dealing with it.
․ The DANGER LED indicator keeps on when STOP key is pressed.
․ The DANGER LED indicator is on but the voltage meter has no readings.
When the above situation occurs, shut down the power and unplug the AC power cord
immediately. Do not use the device again as failure is awfully hazardous. Please send
the hardware back to Chroma or its distributor for repair service.
16. DANGER Indicator failure
When pressing the START key the voltage meter has readings but the DANGER LED
indicator is still off, it means the indicator may be broken. Please power off the hardware
and replace it with another device, then send the broken one back to Chroma or its
distributor for repair service.
17. AC INPUT used by the Tester is 100V~240V auto switch and the fuse specification is
15A Slow/250V. Also to avoid electric shock the fuse should be changed when the
power cord is not plugged in. When replacing, use a flat screwdriver to pry the fuse
holder inside the power socket and remove the fuse to replace with a new one by
pushing it in gently, and then push the power socket back to its position.
18. Tester is normally operated under AC power
If the selected voltage range for local power supply is unstable, it may cause the device
to work inaccurately or abnormally. Thus, please use appropriate equipment such as a
power supply regulator to convert it to a suitable one.
19. Tester uses a power transformer with 1100VA or above
When the device to be tested draws a great deal of current, the current (about 10amp)
may flow in for more than 10ms before judging for defect item and cutting off the output
current. The same situation may occur before test, thus it is necessary to watch out the
power cord capacity and the connecting cables used for other instruments or devices.
20. Storage
The temperature and humidity for the Tester is 5°C~35°C, 70% RH in normal. The
operation may malfunction if exceeding the range. Do not mount the Tester to a fixed
place in case it needs to be removed. The storage temperature for the Tester is from
-10°C to 60°C, ≤ 80% RH. If it is not in use for a long time, please pack it with its original
Precaution before Use
3-5
The protection ground terminal to prevent electric shock due to leakage
using the Scanner.
A warning label to avoid any improper use due to procedure, application
or other reasons that may cause injury or death to human.
A caution label to avoid any improper use due to procedure, application
the unit under test.
A notice label for important information on procedure, application or other
areas. Please read it clearly.
WARNING
CAUTION
package for storage. For proper test and safety measures, do not place the Tester under
direct sunlight, high temperature, trembling, humid or dusty area.
21. Warming up
The tester is activated when power is on; however, in order to meet the specifications for
accuracy please warm it up for 15 minutes or above.
22. Safety symbols
The terminal may output lethal voltage. Please read all precautions
clearly.
There are detail descriptions in the User’s Manual. Please read it to get
detail information.
to chassis. It is necessary to connect this ground terminal to earth before
or other reasons that may cause damage or bad result to the Scanner or
23. Warning label during test
“DANGER – HIGH VOLTAGE TEST IN PROGRESS, UNAUTHORIZED PERSON KEEP
AWAY.”
24. Common Environment Conditions
(1) Indoor use
(2) Altitude: 2000 m
(3) Temperature: 5°C to 45°C
(4) Humidity: Maximum 80%RH at 31°C decreasing to 50%RH at 40°C
(5) Transient Overvoltage at Mains Supply: 2500V
(6) Pollution Degree: 2
25. Keep test cable away from the panel
Please keep the high voltage cable or the DUT away from the panel at least 30 cm during
operation to avoid the display interference caused by high-voltage discharge.
26. Notices for connecting automated device
(1) The grounding system of the device and the automated station should be connected
together.
(2) Add anti-interference iron core to the high voltage cable and the 2 ends (device
output and DUT) of RTN/LOW test cable with winding at least 1 circle.
(3) The high voltage and RTN/LOW test cable must be separate from the control cable.
(4) The high voltage and RTN/LOW test cable must keep proper distance from the
scanner panel.
4-1
4. Operation
MAIN INDEX
01 . PROGRAM
02 . STORE/RECALL
REMOTE
LOCK
CORR
ERROR
Function Key Area
Message Indication Box
Message Line
Title Bar
4.1 Front Panel
Operation
Figure 4-1
The front panel is divided into several easy-to-use areas. This section introduces each control
item and the information displayed on LCD.
The indicator shows the test result in the channel numbers specified by
and 19020-4 shows the test result of CH1~CH4.
User’s Manual
Buttons
(1) Power Switch : It is the AC power switch for this Hipot Tester. Read Chapter 3
Precaution before Use in this manual carefully before using this
switch.
(2) STOP Key : It is the reset key. When pressed the Hipot Tester will cutoff output
immediately or return to ready-to-test state and clear all judgments.
(3) START Key : It is the test activation key. When pressed, the Hipot Tester is in test
state, which means there is output on test terminal and the judging
functions are activated at the same time.
(4) Cursor Keys : ▲ and ▼keys are used to move the reserved cursor.
(5) TEST Key : Press this key under each major display mode can return to the
“TEST” main screen.
(6) MAIN INDEX Key: Press this key under each major display mode can return to the
“MAIN INDEX” main screen.
(7) SYSTEM Key : Press this key under each major display mode can return to the
“SYSTEM” main screen.
(8) Data Entry Keys/Program Keys
0 . ~ 9 : They arenumeric/character keys for inputting test parameter data
(value or English letters.)
ENTER : It is the input confirmation key for setting test parameters.
CLR : It is the cancel key for clearing the inputted test parameters when
error occurs and entering the new data again.
(9) Function Keys : Different function key descriptions will appear in different screen
and the mapping function keys are at the right of LCD. If the
description is blank, it indicates the mapping function key is invalid.
(10) Indicators
DANGER LED : It is the indicator of test status. When on it means the Tester is
performing test when on. Do not touch the test terminals as they may
contain high voltage output.
PASS LED: It is the indicator for pass items. CH1~CH10 indicators stand for the
test results of CH1~CH10. The rectangular PASS LED is on only
when the test results of CH1~CH10 are all good.
FAIL LED : It is the indicator for fail items. CH1~CH10 indicators stand for the
test results of CH1~CH10. The rectangular FAIL LED is on when one
of the CH1~CH10 test results is bad and will keep on until STOP is
pressed.
the model, for instance, the 19020 shows the test result of CH1~CH10
4-5
4.2 Rear Panel
Operation
Figure 4-2
(1) GPIB Interface:
This is the connector of GPIB interface card under IEEE-488-1978 standard. See
Chapter 6GPIB/RS232 Interface (IEEE-488.2) for detail usage description.
(2) RS232 Interface:
This is the connector of optional RS232 interface card. GPIB and RS232 cannot be used
at the same time.
(3) CAN BUS Interface:
This connector is used for data communication between MASTER and SLAVE.
(4) Terminal Resistance Selector:
This DIP switch is used to set the terminal resistance for CAN BUS interface. It is
necessary to set the DIP switch to ON for the first and last Testers on the CAN BUS
transmission path.
(5) INTER LOCK: High voltage can only be outputted when these two terminals are
short-circuited.
(6) HANDLER Interface:
This is the connector for HANDLER interface. See Chapter 5 HANDLER Interface for
detail usage description.
(7) Internal Communication Interface:
This connector is used for control signal transmission between MASTER and SLAVE.
(8) Fuse Holder:
See Chapter 3 Precaution before Use for detail specification or the label mark on the
rear panel.
It is the safety grounding terminal. Please use an appropriate tool to connect it to earth
properly. If it is not properly grounded, the Tester chassis may contain high voltage when
the power circuit or any device’s cable is shorted with the grounding terminal, and it is
very dangerous as anyone who touches it may cause electric shock incident. Therefore,
the safety ground terminal must be connected to earth properly.
(10) AC Input:
It contains a three-wire AC power socket. The AC power required by the Tester is
supplied by this power socket. The power socket or connecting cable can be interrupt
device.
(11) High Voltage Module:
It is the combination of high voltage output circuits. Be sure the anchor screws are
secured.
(12) High Voltage Output Channel:
It is the high potential terminal for high voltage output. The output terminal that belongs
to high potential output terminal usually has high voltage output. Thus, it is very
dangerous. Do not touch it especially when the DANGER LED is on with high voltage
output.
(13) RTN/LOW Channel:
It is the common test terminal that is the reference terminal for high voltage test. It is the
low potential terminal that almost equals to chassis grounding terminal.
4.3 Notices before Using & Procedure
1. Before plugging in the AC power cord, make sure the power in use matches the power
indicated on the rear panel and the switch is OFF.
2. Read the precautions described in Chapter 3 carefully and keep them in mind before
power-on the Tester.
3. Once the Tester is powered on, it will start self-test. If any abnormal occurs, turn off the
power switch immediately and unplug the power cord.
Operation
4-7
Setting Items
Description
TEST CONTROL
It sets the related parameters for test. See section 4.4.2 for details.
SYSTEM CONFIG
It sets the system related parameters. See section 4.4.3 for
details.
KEY LOCK
It sets the keyboard lock function. See section 4.4.4 for details.
FAIL LOCK
It sets the fail lock function on keyboard. See section 4.4.5 for
details.
CHANGE PASSWORD
It changes the user’s password. See section 4.4.6 for details.
CALIBRATION
It sets the calibration related function. See section 4.4.7 for details.
ERROR LOG
It logs the errors messages generated when connecting with PC.
See section 4.4.8 for details.
ABOUT
It shows the firmware version related description.
SYSTEM
01 . TEST CONTROL
02 . SYSTEM CONFIG
03 . KEY LOCK
04 . FAIL LOCK
05 . CHANGE PASSWORD
06 . CALIBRATION
07 . ERROR LOG
08 . ABOUT
REMOTE
LOCK
CORR
ERROR
4.4 Setting SYSTEM Parameters
4.4.1 Entering SYSTEM Setting Screen
Press SYSTEM in any screen will show the following:
Procedure
1. When title bar shows “SYSTEM”, press ▲, ▼ to move the highlight to the item to be set.
2. Press ENTER to go the sub menu or set the parameter data.
3. Press numeric/character keys or Function Keys to set the parameter data.
4. When error occurs during data input, press CLR to clear it and re-enter. At last, press
ENTER to confirm the parameter data.
The following table lists the setting items of System and their descriptions.
It sets the time duration the beeper
sounds for PASS.
02
AC
FREQUENCY
50, 60Hz
60
It sets the frequency of output voltage
when doing AC withstand voltage test.
08
SOFTWARE
AGC
ON/OFF
ON
It sets the software AGC function to be
on or off.
04
MIN. VOLTAGE
80%
It sets the percentage for the output
voltage to reach the set voltage.
05
WV AUTO
RANGE
ON/OFF
OFF
It sets the range auto change function for
withstand voltage test to be on or off.
06
CH AFTER FAIL
STOP/
CONTINUE
STOP
It sets if the channel stops testing when
FAIL occurs during test.
When it is set to ON, it will judge if the
judge if the
during ramp time execution.
Output
Frame 0 are
ON
It sets the default channel to be on. The
Setting DEF. CHANNELS.
09
SCREEN
ON/OFF
ON
It sets the LCD screen to be on or off
during test.
TIMER /
TEST
1. When EOT set to TIMER, it means
nEOT and PASS_FAIL signals act
User’s Manual
4.4.2 Setting TEST CONTROL
In SYSTEM screen, move the highlight to TEST CONTROL and press ENTER to go to TEST
CONTROL setting screen as shown below:
When in TEST CONTROL screen, press ▲, ▼ to move the highlight to the item to be set for
related setting.
The following table lists the setting items of TEST CONTROL and their descriptions.
OFF,50~95%
current readings exceed the high limit
RAMP
07
JUDGMENT
ON / OFF ON
during ramp time execution.
When it is set to OFF, it won’t
current readings exceed the high limit
DEF.
08
CHANNELS
10 EOT
-------------
channel of
TEST
settings here will become the default of
PROGRAM channel. Details please see
Operation
4-9
immediately after test time ends
after the high voltage discharge ends.
When set to ON, the discharge circuit will
will discharge to safe voltage.
1. When the offset is bigger than the AC
1. The DISCH. Vmin setting is only valid when the EOT is set to TEST.
offset.
TEST CONTROL
FRAME
0 1 2 3 4 5 6 7 8 9
NEXT FRAME
01 02
ON
03
04
OFF
05 06 07 08 09 10
EXIT
Remote
Lock
offset
Error
Please follow the model to set the output channel, for instance, the 19020
“Module Fail” if set otherwise and the test is unable to start.
without waiting for the high voltage
discharge to end.
2. When EOT set to TEST, it means
nEOT and PASS_FAIL signals act
DISCH. Vmin
11
<Prompt 1>
AC OFFSET
12
<Prompt 2>
2. For 12. AC OFFSET in 30uA current measurement range, the
Setting DEF. CHANNELS:
Move the highlight to DEF. CHANNELS and press Function Key [SETUP] to go to DEF.
CHANNELS setting screen as shown below:
ON/OFF OFF
0.001~ 2.5mA 0.001mA
When EOT is set to TIME, the tester will end the testing without
discharge, thus the DISCH. Vmin is invalid.
deduction of OFFSET is always current value = measured current –
discharge to the voltage lower than safe
voltage.
When set to OFF, the discharge circuit
OFFSET setting, current reading=
measured current-offset.
2. When the offset is smaller than AC
OFFSET setting, current reading =
√(measured value)
2
– (offset)
2
CH
Press ▲, ▼ move the highlight to channel to be set and press Function Key [ON] [OFF] to
enable or disable it. Use Function Key [NEXT FRAME] to move the highlight to next FRAME.
can set the output channel to CH1~CH10 while the 19020-4 can only set
the channel to CH1~CH4. The message line on the test screen will show
It sets the transmission baud rate of
RS232 interface.
LINK SETUP
MASTER/SLAVE/SCA
MASTER
It sets the tester to be MASTER or
connecting to A190201/A190202.
LINK ADDRESS
1~9
0
It sets the tester address when linked
when set to MASTER.
ACA MEAS
RMS/GENERAL
RMS
It sets the AC current measurement.
DC ARC RATE
1.0~10.0
1.0
The ARC LIMIT setting will multiply this
MODE test.
1. When LINK SETUP is set to SLAVE, besides the items “02.
selection.
User’s Manual
4.4.3 Setting SYSTEM CONFIG
In SYSTEM screen, move the highlight to SYSYEM CONFIG and ENTER to go to SYSYEM
CONFIG setting screen as shown below:
When in SYSTEM CONFIG screen, press ▲, ▼ to move the highlight to the item desired for
setting the related function.
The following table lists the setting items of SYSYTEM CONFIG and their descriptions.
ADDRESS = 0~30
NNER
SYSTEM CONFIG”, “05. CHANGE PASSWORD”, “06.
CALIBRATION”, “07. ERROR LOG” and “08. ABOUT” in the
SYSTEM screen, the rest items are all invalid.
2. Only the 19020/19020-04 LINK SETUP settings has SCANNER
shows UNINSTALLED if GPIB card is
SLAVE when linked for test.
Be sure to set it to SCANNER when
for test.
Note: The tester address is fixed to 0
rate for RAMP, DWELL time during DC
Operation
4-11
KEY LOCK
LOCK KEY:
USER PASSWORD:■
EXIT
REMOTE
LOCK
CORR
ERROR
KEY LOCK
UNLOCK KEY:
USER PASSWORD:■
EXIT
REMOTE
LOCK
CORR
ERROR
4.4.4 Setting KEY LOCK
The way to set KEY LOCK:
In SYSTEM screen, move the highlight to KEY LOCK and press ENTER to go to KEY LOCK
setting screen as shown below:
1. Enter the PASSWORD when in KEY LOCK screen. (The default is 0 0 0 0.)
2. Press ENTER will prompt a selection window to select if to lock RECALL MEMORY.
Users can use Function Keys [YES], [NO] to select if locking the function of MEMORY
RECALL as well.
3. When KEY LOCK is ON, the LOCK text is reversed to indicate the host is in parameter
protection mode. The “OFFSET”, “GET Cs”, “CLEAR” in [TEST] and “PROGRAM”,
“STORE” in [MAIN INDEX] as well as the “TEST CONTROL”, “SYSTEM CONFIG”,
“FAIL LOCK”, “CHANGE PASSWORD” and “CALIBRATION” in [SYSTEM] are all invalid
for setting.
4. When setting KEY LOCK, if RECALL LOCK ON is selected, the MEMORY RECALL
function is also invalid.
The way to release KEY LOCK:
If the LOCK box is reversed in power on screen, the KEY LOCK function can be released.
Press SYSTEM to select KEY LOCK as shown below:
Enter the PASSWORD and press ENTER, the LOCK box returns to normal indicating the
KEY LOCK is cancelled.
4.4.5 Setting FAIL LOCK
The way to set FAIL LOCK:
In SYSTEM screen, move the highlight to FAIL LOCK and press ENTER to go to FAIL LOCK
setting screen as shown below:
1. Enter the PASSWORD when in FAIL LOCK screen.
2. When FAIL LOCK is ON, the LOCK text is reversed to indicate the host is in parameter
FAIL LOCK mode. The “RECALL”, “OFFSET”, “GET Cs” in [TEST] and “PROGRAM”,
“STORE/RECALL” in [MAIN INDEX] as well as the “TEST CONTROL”, “SYSTEM
CONFIG”, “KEY LOCK”, “CHANGE PASSWORD” and “CALIBRATION” in [SYSTEM]
are all invalid for setting.
3. When FAIL LOCK is set and the test result is FAIL, all keys are invalid except the
Function Key [CLEAR] and STOP in TEST screen. It is necessary to press the Function
Key [CLEAR] to enter the FAIL LOCK password to continue the test.
The way to release FAIL LOCK:
If the LOCK box is reversed in power on screen, the FAIL LOCK function can be released.
Press SYSTEM to select FAIL LOCK as shown below:
Operation
4-13
If the memory has been cleared following the description of “Clear the
PASSWORD will return to initial that is 0000.
FAIL LOCK
UNLOCK:
USER PASSWORD:■
EXIT
REMOTE
LOCK
CORR
ERROR
CHANGE PASSWORD
USER PASSWORD:■
EXIT
REMOTE
LOCK
CORR
ERROR
Enter the PASSWORD and press ENTER, the LOCK box returns to normal indicating the
FAIL LOCK is cancelled.
4.4.6 Changing PASSWORD
Setting password for KEY LOCK:
In SYSTEM screen, move the highlight to CHANGE PASSWORD and press ENTER to go to
CHANGE PASSWORD setting screen as shown below:
1. Enter the PASSWORD (enter 0000 if it hasn’t been set) and press ENTER will prompt a
“NEW PASSWORD” window.
2. Enter the NEW PASSWORD (maximum 10 characters) and press ENTER will prompt a
“CONFIRM PASSWORD” window.
3. Enter the same password again and press ENTER. A message “CHANGE PASSWORD OK!” will appear to indicate the password has been changed. Press Function Key [EXIT]
to quit the Setting Screen.
settings and test procedures in memory” in section 4.4.7, the
Please refer to Chapter 7 Calibration Procedure for related info.
Clear the settings and test
85246
This function will clear all settings and test
procedures in memory and return to factory
default.
CALIBRATION
CAL. PASSWORD︰■
EXIT
REMOTE
LOCK
CORR
ERROR
ERROR LOG
1. +0,No error
EXIT
REMOTE
LOCK
CORR
ERROR
User’s Manual
4.4.7 Setting CALIBRATION
In SYSTEM screen, move the highlight to CALIBRATION and press ENTER to go to
CALIBRATION setting screen as shown below:
When in CALIBRATION screen, press keys to set the related functions.
The following table lists the setting items of CALIBRATION and their descriptions.
procedures in memory
4.4.8 Setting ERROR LOG
In SYSTEM SETUP screen, move the highlight to ERROR LOG and press ENTER to go to
ERROR LOG screen as shown below:
When in ERROR LOG screen it will show the error message generated during connection.
The ERR box will return to normal without reverse when in ERROR LOG screen.
4-15
4.5 Setting PROGRAM
PROGRAM
01.
TEST STEP :
1
UP
02.
TEST MODE :
AC
03.
VOLTAGE :
OFF
DOWN
04.
HIGH LIMIT :
0.500 mA
05.
LOW LIMIT :
OFF
NEW
06.
ARC LIMIT :
OFF
07.
TEST TIME :
3.0 sec
INSERT
08.
RAMP TIME :
OFF
09.
FALL TIME :
OFF
DELETE
10.
CHANNELS :
DEFAULT
EXIT
1-10
REMOTE
LOCK
CORR
ERROR
MAIN INDEX
01 . PROGRAM
02 . STORE/RECALL
REMOTE
LOCK
CORR
ERROR
4.5.1 How to Get in PROGRAM Setting Screen
Press MAIN INDEX in any screen will show the following:
Operation
4.5.2 Setting Program Procedure
1. In MAIN INDEX screen, press ▲, ▼ to move the highlight to [PROGRAM] and press
ENTER to enter into the program setting screen as shown below:
2. When in PROGRAM screen, use Function Keys [NEW] to add new test steps from 1 to
10.
3. Use Function Keys [UP] and [DOWN] to switch to previous or next test step.
4. Press ▲, ▼ to move the highlight to the item to be set and press ENTER to confirm it.
1. When in PROGRAM screen, press ▼ to move the highlight to the following position.
2. Use Function Key [AC], [DC], [IR], [OSC] and [PA] to select the test mode. There are AC
/ DC / IR / OSC /PA available for selection. Different test mode and model number has
different programs for setting.
4.5.4 Description of Parameters
Following explains the parameters set in each test mode.
AC Withstand Voltage (AC) Test Mode (for Model 19020/19020-4/19021/19021-4)
VOLTAGE : It sets the voltage required for AC withstand voltage test.
HIGH LIMIT : It sets the high limit of leakage current.
LOW LIMIT : It sets the low limit of leakage current. The range is smaller than the high
limit of leakage current or OFF.
ARC LIMIT : It sets the ARC limit, 0 means OFF.
TEST TIME : It sets the time test required, 0 means continue test.
RAMP TIME : It sets the time required for ramping to the set voltage, 0 means OFF.
FALL TIME : It sets the time required for falling to low voltage from set, 0 means OFF.
Operation
4-17
Please follow the model to set the output channel, for instance, the 19020
“Module Fail” if set otherwise and the test is unable to start.
PROGRAM
01.
TEST STEP :
1
AC
02.
TEST MODE :
DC
03.
VOLTAGE :
0.050 kV
DC
04.
HIGH LIMIT :
0.500 mA
05.
LOW LIMIT :
OFF
IR
06.
ARC LIMIT :
OFF
07.
TEST TIME :
3.0 sec
OSC
08.
RAMP TIME :
OFF
09.
DWELL TIME :
OFF
PA
10.
FALL TIME :
OFF
11.
CHANNELS :
DEFAULT
EXIT
12.
SCANNER :
NONE
SELECT MODE
REMOTE
LOCK
CORR
ERROR
Please follow the model to set the output channel, for instance, the 19020
the channel to CH1~CH4. The message line on the test screen will show
CHANNELS : It sets the high voltage channel for output. Select Function Key
[DEFAULT] means to use the “DEF. CHANNELS” settings in “TEST
CONTROL.” Select Function Key [SETUP] means to reset the high
voltage channel for output.
SCANNER : It sets the output mode of A190201/A190202 SCANNER. It can select
[ PS ], [ PC ], [ SC ], [ P+SC ] and [ PS+C ]. It is only valid
when the LINK SETUP in the SYSTEM of 19020/19020-4 is set to
SCANNER.
can set the output channel to CH1~CH10 while the 19020-4 can only set
the channel to CH1~CH4. The message line on the test screen will show
DC Withstand Voltage Test Mode (DC) (for Model 19020/19020-4/19022/19022-4)
VOLTAGE : It sets the voltage required for DC withstand voltage test.
HIGH LIMIT : It sets the high limit of leakage current.
LOW LIMIT : It sets the low limit of leakage current. The range is smaller than the high
limit of leakage current or OFF.
ARC LIMIT : It sets the ARC limit, 0 means OFF.
TEST TIME : It sets the time test required, 0 means continue test.
RAMP TIME : It sets the time required for ramping to the set voltage, 0 means OFF.
DWELL TIME : It sets the time required for DWELL, 0 means OFF.
(It does not judge the high and low limit of leakage current during DWELL
TIME but only when the set range is within the high limit.)
FALL TIME : It sets the time required for falling to low voltage from set, 0 means OFF.
CHANNELS : It sets the high voltage channel for output. Select Function Key
[DEFAULT] means to use the “DEF. CHANNELS” settings in “TEST
CONTROL.” Select Function Key [SETUP] means to reset the high
voltage channel for output.
SCANNER : It sets the output mode of A190201/A190202 SCANNER. It can select
[ PS ], [ PC ], [ SC ], [ P+SC ] and [ PS+C ]. It is only valid
when the LINK SETUP in the SYSTEM of 19020/19020-4 is set to
SCANNER.
can set the output channel to CH1~CH10 while the 19020-4 can only set
“Module Fail” if set otherwise and the test is unable to start.
PROGRAM
01.
TEST STEP :
1
AC
02.
TEST MODE :
IR
03.
VOLTAGE :
0.050 kV
DC
04.
LOW LIMIT :
1.0 MΩ
05.
HIGH LIMIT :
OFF
IR
06.
TEST TIME :
3.0 sec
07.
RAMP TIME :
OFF
OSC
08.
FALL TIME :
OFF
09.
RANGE :
AUTO
PA
10.
CHANNELS :
DEFAULT
11.
SCANNER :
NONE
EXIT
SELECT MODE
REMOTE
LOCK
CORR
ERROR
IR Display
Set voltage to 50V~499V
Set voltage to500V~1000V
5mA(2.7~5mA)
0.1MΩ~2.4MΩ
0.1MΩ~7.7MΩ
3mA (0.27~3mA)
0.1MΩ~7.7MΩ
0.1MΩ~24.5MΩ
300uA(27~300uA)
0.1MΩ~24.5MΩ
0.1MΩ~49.9MΩ
50MΩ~245MΩ
30uA(2.7~30uA)
0.1MΩ~49.9MΩ
0.1MΩ~49.9MΩ
0.50GΩ~2.45GΩ
3uA(0.27~3uA)
0.1MΩ~49.9MΩ
0.1MΩ~49.9MΩ
5.0GΩ~49.9GΩ
300nA(27~300nA)
0.1MΩ~49.9MΩ
0.1MΩ~49.9MΩ
50GΩ~60GΩ
30nA(1~30nA)
----------
0.1MΩ~49.9MΩ
50GΩ~60GΩ
User’s Manual
Insulation Resistance Test Mode (IR) (for Model 19020/19020-4/19022/19022-4)
VOLTAGE : It sets the voltage required for insulation resistance test.
LOW LIMIT : It sets the low limit of insulation resistance.
HIGH LIMIT : It sets the high limit of insulation resistance. The range is larger than the
low limit of insulation resistance or OFF.
TEST TIME : It sets the time test required, 0 means continue test.
RAMP TIME : It sets the time required for ramping to the set voltage, 0 means OFF.
FALL TIME : It sets the time required for falling to low voltage from set, 0 means OFF.
RANGE : It sets the current test range for insulation resistance, AUTO means
switching the range automatically. The table below lists the relationship
between current range and resistance measurement range.
Range
50MΩ~245MΩ
50MΩ~499MΩ
0.50GΩ~2.45GΩ
50MΩ~499MΩ
0.50GΩ~2.45GΩ
50MΩ~499MΩ
50MΩ~499MΩ
0.50GΩ~4.99GΩ
50MΩ~499MΩ
0.50GΩ~4.99GΩ
5.0GΩ~49.9GΩ
50MΩ~499MΩ
0.50GΩ~4.99GΩ
5.0GΩ~49.9GΩ
Operation
4-19
To select an appropriate IR current range, please calculate the current by
current range. It will show UUUUU if the IR display exceeds 60GΩ.
Please follow the model to set the output channel, for instance, the 19020
“Module Fail” if set otherwise and the test is unable to start.
PROGRAM
01.
TEST STEP :
1
AC
02.
TEST MODE :
OSC
03.
OPEN :
50%
DC
04.
SHORT :
OFF
05.
CHANNELS :
DEFAULT
IR
06.
SCANNER :
NONE
OSC
PA
EXIT
SELECT MODE
REMOTE
LOCK
CORR
ERROR
1. Before conducting the test or testing the new capacitance UUT or
2. Before reading standard capacitance (GET Cs), press Function Key
test voltage and UUT’s insulation impedance, and then select the proper
CHANNELS : It sets the high voltage channel for output. Select Function Key
[DEFAULT] means to use the “DEF. CHANNELS” settings in “TEST
CONTROL.” Select Function Key [SETUP] means to reset the high
voltage channel for output.
SCANNER : It sets the output mode of A190201/A190202 SCANNER. It can select
[ PS ], [ PC ], [ SC ], [ P+SC ] and [ PS+C ]. It is only valid
when the LINK SETUP in the SYSTEM of 19020/19020-4 is set to
SCANNER.
can set the output channel to CH1~CH10 while the 19020-4 can only set
the channel to CH1~CH4. The message line on the test screen will show
Open/Short Check (OSC) Test Mode (for Model 19020/19020-4/19021/19021-4/19022/
19022-4)
OPEN : It sets the OPEN condition for test result judgment. (Compare it with the
test reading and the read standard capacitance [Cs].)
SHORT : It sets the SHORT condition for test result judgment. (Compare it with the
test reading and the read standard capacitance [Cs].)
CHANNELS : It sets the high voltage channel for output. Select Function Key
[DEFAULT] means to use the “DEF. CHANNELS” settings in “TEST
CONTROL.” Select Function Key [SETUP] means to reset the high
voltage channel for output.
SCANNER : It sets the output mode of A190201/A190202 SCANNER. It can select
[ PS ], [ PC ], [ SC ], [ P+SC ] and [ PS+C ]. It is only valid
when the LINK SETUP in the SYSTEM of 19020/19020-4 is set to
SCANNER.
replacing the capacitance UUT in OSC Mode, the action of reading
standard capacitance (GET Cs) has to be done first.
[OFFSET] first to conduct OFFSET. OFFSET needs to be done
to start.
PROGRAM
01.
TEST STEP :
1
AC
02.
TEST MODE :
PA
03.
MESSAGE :
PAUSE-MODE
DC
04.
TIME :
CONTINUE
IR
OSC
PA
EXIT
SELECT MODE
REMOTE
LOCK
CORR
ERROR
User’s Manual
every time the cable or fixture is changed to ensure the test
accuracy.
3. When conducting tests in OSC Mode, the test condition for judging
OPEN/SHORT is the reading of GET Cs.
4. Please follow the model to set the output channel, for instance, the
19020 can set the output channel to CH1~CH10 while the 19020-4
can only set the channel to CH1~CH4. The message line on the test
screen will show “Module Fail” if set otherwise and the test is unable
Pause Mode (PA) (for Model 19020/19020-4/19021/19021-4/19022/19022-4)
MESSAGE : It sets the message for pause screen, maximum 13 characters for input.
TIME : It sets the action of PAUSE MODE.
(1) CONTINUE : The pause mode only ends when START on the
panel is pressed or START signal on the HANDLER
card is triggered again.
(2) 0.1~999.9sec : The pause mode ends when it reaches the time set.
Operation
4-21
MAIN INDEX
01 . PROGRAM
02 . STORE/RECALL
REMOTE
LOCK
CORR
ERROR
STORE/RECALL
01 . (01) CHROMA
STORE
02 . (00)
03 . (00)
RECALL
04 . (00)
05 . (00)
DELETE
06 . (00)
07 . (00)
08 . (00)
09 . (00)
10 . (00)
EXIT
REMOTE
LOCK
CORR
ERROR
4.6 Managing Memory for Programs
4.6.1 Entering Memory Screen
1. Press MAIN INDEX in any screen will appear the following:
2. When “MAIN INDEX” shows on the title bar, press ▲, ▼ to move the highlight to
[STORE/RECALL] and press ENTER to go to the setting screen as shown below:
3. Use ▲, ▼ to move the highlight to the memory to be processed and follow the
instruction of Function Key to read, save or erase that memory.
4. The number in parentheses means the test steps contained in the memory.
4.6.2 Saving Memory
Follow the steps below to save the set program data to memory:
1. Use ▲, ▼ to move the highlight to the memory to be saved and press Function Key
[STORE].
If there is data in the memory, it will be overwritten when save. Make sure
it is ok to do so before save.
STORE/RECALL
01 . (01) CHROMA
YES
02 . (00)
03 . (00)
NO
04 . (00)
05 . (00)
06 . (00)
07 . (00)
08 . (00)
09 . (00)
10 . (00)
REMOTE
LOCK
CORR
ERROR
DELETE MEMORY 1
User’s Manual
2. Use 1 2 3 4 4 5 6 7 8 9 0 to select the character for entering the
memory name. Then use the numeric/character keys to enter the memory name. Press
one numeric/character key repeatedly can switch the number and English letter display
in cycle. Pressing Function Key [NEXT CHAR] to move the cursor to the next character
position can use the same number/character key to enter the name in sequence.
3. Press ENTER, a confirmation dialog box will appear for save.
4. Press Function Key [YES] to confirm it or press Function Key [NO] to cancel it.
4.6.3 Deleting Memory
Follow the steps below to delete the programs from memory:
1. Use ▲, ▼ to move the highlight to the memory to be deleted and press Function Key
[DELETE].
2. A delete confirmation dialog box is prompted. Press Function Key [YES] to confirm it or
press Function Key [NO] to cancel it.
Operation
4-23
TEST – 1/2
MODE
CH
SOURCE
LIMIT
RES.
RECALL
01
50V
0.0 nF
02
50V
0.0 nF
OFFSET
03
50V
0.0 nF
04
50V
0.0 nF
GET Cs
05
50V
0.0 nF
06
50V
0.0 nF
07
50V
0.0 nF
08
50V
0.0 nF
CLEAR
09
50V
0.0 nF
10
50V
0.0 nF
SLAVE
1 2 3 4 5 6 7 8 9
STANDBY
REMOTE
LOCK
CORR
ERROR
STORE/RECALL
01 . (01) CHROMA
YES
02 . (00)
03 . (00)
NO
04 . (00)
05 . (00)
06 . (00)
07 . (00)
08 . (00)
09 . (00)
10 . (00)
REMOTE
LOCK
CORR
ERROR
RECALL FROM MEMORY 1
4.6.4 Recalling Memory
Follow the steps below to recall the programs stored in memory:
1. Use ▲, ▼ to move the highlight to memory to be recalled and press Function Key
[RECALL].
2. A recall confirmation dialog box is prompted. Press Function Key [YES] to confirm it or
press Function Key [NO] to cancel it.
4.7 Using OFFSET or GET Cs
4.7.1 How to Get in OFFSET or GET Cs Screen
1. Select OFFSET or GET Cs operation in TEST screen as shown below:
OSC
2. Press Function Keys [OFFSET] or [GET Cs] to select the desired function. If these two
keys are grayed out, it means they are invalid here.
1. Press Function Key [OFFSET] in TEST screen to go to GET OFFSET screen as shown
below:
2. GET OFFSET is to offset the leakage current of AC / DC / IR MODE test leads and
fixture as well as the stray capacitance of OSC MODE.
3. Remove the UUT from fixture and press START to offset the leakage current or leakage
capacitance.
4. The message box [CORR] will be reserved when the test time ends.
5. Press Function Keys [OFF] can cancel OFFSET.
4.7.3 Using GET Cs
1. Press Function Key [GET Cs] in TEST screen to go to GET Cs screen as shown below:
2. Use the standard sample under capacitance test as the UUT and connect it to the 1st
channel opened by OSC (this CHANNEL has to be the Master CHANNEL). Press
START to get the standard capacitance (GET Cs).
Operation
4-25
TEST – 1/2
MODE
CH
SOURCE
LIMIT
RES.
RECALL
01
0.050kV
0.500 mA
02
0.050kV
0.500 mA
OFFSET
03
0.050kV
0.500 mA
04
0.050kV
0.500 mA
GET Cs
05
0.050kV
0.500 mA
06
0.050kV
0.500 mA
07
0.050kV
0.500 mA
08
0.050kV
0.500 mA
CLEAR
09
0.050kV
0.500 mA
10
0.050kV
0.500 mA
SLAVE
1 2 3 4 5 6 7 8 9
STANDBY
REMOTE
LOCK
CORR
ERROR
Position 1
Position 2
Position 3
4.8 Conducting the Test
4.8.1 Connecting the UUT
First ensure there is no voltage output and the DANGER LED is off. Connect the low potential
test cable (black) to the Tester RTN/LOW terminal. Short-circuit the test cable and high
voltage output terminal and ensure there is no high voltage output. Next, plug in the high
voltage test cable (red or white) to high voltage output terminal. Then connect the low
potential test cable to UUT and the high potential test cable to UUT.
4.8.2 Procedure for AC/DC Test
1. Connect the UUT properly following the connection method.
2. In the power on screen shown below:
AC
Illustration:
TEST 1/2 means there are 2 test steps and it is running the 1
st
test step at present. AC
indicates the test mode. “Position 1” indicates the set voltage, “Position 2” is the high limit
set for current, while “Position 3” is the test result.
3. Press STOP to prepare for test. The status line shows “STANDBY”.
4. Press START to activate the test
When this key is pressed it starts to output voltage and the DANGER LED is on. The
status line shows a counter to count down. “Position 1” will show the output voltage value,
“Position 2” will show the current readings and “Position 3” will show the test result.
5. GOOD Judgment
When all tests are done and the results show PASS, the Tester will see the UUT as a
GOOD product and cutoff the output. The HANDLER interface outputs PASS signal and
the beeper acts at the same time.
6. NO GOOD Judgment
If the test value is abnormal, the Tester judges it as FAIL and cutoff the output
The current measured exceeds the range or the set high/low
limit.
ARC
The ARC measured exceeds the set high limit.
TEST – 1/2
MODE
CH
SOURCE
LIMIT
RES.
RECALL
01
0.050kV
1.0 MΩ
02
0.050kV
1.0 MΩ
OFFSET
03
0.050kV
1.0 MΩ
04
0.050kV
1.0 MΩ
GET Cs
05
0.050kV
1.0 MΩ
06
0.050kV
1.0 MΩ
07
0.050kV
1.0 MΩ
08
0.050kV
1.0 MΩ
CLEAR
09
0.050kV
1.0 MΩ
10
0.050kV
1.0 MΩ
SLAVE
1 2 3 4 5 6 7 8 9
STANDBY
REMOTE
LOCK
CORR
ERROR
Position 1
Position 2
Position 3
User’s Manual
immediately. The HANDLER outputs FAIL signal and the beeper acts at the same time
until the STOP key on Tester is pressed. The test result will show FAIL state.
FAIL State:
To stop test output in any condition, just press STOP.
4.8.3 Procedure for IR Test
1. Connect the UUT properly following the connection method.
2. In the power on screen shown below:
IR
Illustration:
TEST 1/2 means there are 2 test steps and it is running the 1
st
test step at present. IR
indicates the test mode. “Position 1” indicates the set voltage, “Position 2” is the low limit
set for insulation impedance, while “Position 3” is the test result.
3. Press STOP to prepare for test. The status line shows “STANDBY”.
4. Press START to activate the test
When this key is pressed it starts to output voltage and the DANGER LED is on. The
status line shows a counter to count down. “Position 1” will show the output voltage value,
“Position 2” will show the current readings and “Position 3” will show the test result.
5. GOOD Judgment
When all tests are done and the results show PASS, the Tester will see the UUT as a
GOOD product and cutoff the output. The HANDLER interface outputs PASS signal and
the beeper acts at the same time.
6. NO GOOD Judgment
If the test value is abnormal, the Tester judges it as FAIL and cutoff the output
immediately. The HANDLER outputs FAIL signal and the beeper acts at the same time
4-27
until the STOP key on Tester is pressed. The test result will show FAIL state.
Test Result Display
Meaning
FAIL
The resistance measured exceeds the range or the set high/low
limit.
TEST – 1/2
MODE
CH
SOURCE
LIMIT
RES.
RECALL
01
50V
0.0 nF
02
50V
0.0 nF
OFFSET
03
50V
0.0 nF
04
50V
0.0 nF
GET Cs
05
50V
0.0 nF
06
50V
0.0 nF
07
50V
0.0 nF
08
50V
0.0 nF
CLEAR
09
50V
0.0 nF
10
50V
0.0 nF
SLAVE
1 2 3 4 5 6 7 8 9
STANDBY
REMOTE
LOCK
CORR
ERROR
Position 1
Position 2
Position 3
FAIL State:
To stop test output in any condition, just press STOP.
4.8.4 Procedure for OSC Test
1. Connect the UUT properly following the connection method.
2. In the power on screen shown below:
Operation
OSC
Illustration:
OSC means it is in Open Short Check mode. “Position 1” is the setting voltage and
“Position 2” is the standard capacitance (Cs) while “Position 3” shows the test result.
3. Press STOP to prepare for test. The status line shows “STANDBY”.
4. Press START to activate the test
When this key is pressed it starts to output voltage and the DANGER LED is on. The
status line shows a counter to count down. “Position 1” will show the output voltage value,
“Position 2” will show the current readings and “Position 3” will show the test result.
5. GOOD Judgment
When all tests are done and the results show PASS, the Tester will see the UUT as a
GOOD product and cutoff the output. The HANDLER interface outputs PASS signal and
the beeper acts at the same time.
6. NO GOOD Judgment
If the test value is abnormal, the Tester judges it as FAIL and cutoff the output
immediately. The HANDLER outputs FAIL signal and the beeper acts at the same time
until the STOP key on Tester is pressed. The test result will show FAIL state.
The Open/Short Capacitance reading exceeds the OPEN/
SHORT setting.
1. Every time the cable or fixture is changed for OSC, be sure to run
3. For using OSC GET CS, see section 4.7.3 for detail information.
TEST – 1/2
MODE
CH
SOURCE
LIMIT
RES.
RECALL
01
PAUSE-MODE
02
PAUSE-MODE
OFFSET
03
PAUSE-MODE
04
PAUSE-MODE
GET Cs
05
PAUSE-MODE
06
PAUSE-MODE
07
PAUSE-MODE
08
PAUSE-MODE
CLEAR
09
PAUSE-MODE
10
PAUSE-MODE
SLAVE
1 2 3 4 5 6 7 8 9
STANDBY
REMOTE
LOCK
CORR
ERROR
Position 1
Position 2
User’s Manual
FAIL State:
FAIL
To stop test output in any condition, just press STOP.
OFFFSET in advance to ensure the test accuracy.
2. It is necessary to run GET Cs when testing a new UUT or replacing a
UUT for OSC tests. Read the standard capacitance from the test
sample as the standard value.
4.8.5 Test Procedure for PA Test
1. Connect the UUT properly following the connection method.
2. In the power on screen shown below:
PA
Illustration:
PA means it is in pause mode. “Position 1” is the message set to display and “Position 2”
is the test result.
3. Press STOP to prepare for test. The status line shows “STANDBY”.
4. Press START to activate the test
When this key is pressed it starts PA MODE. If action time is set for PA MODE, the
status line will show a counter to count down. If the test time is set to CONTINUE, the
status line will show PAUSE and wait for the input of START signal to end PA MODE.
4-29
4.8.6 Auto Range
TEST – 1/2
MODE
CH
SOURCE
LIMIT
RES.
RECALL
01
0.500kV
10.00 mA
02
0.500kV
10.00 mA
OFFSET
03
0.500kV
10.00 mA
04
0.500kV
10.00 mA
GET Cs
05
0.500kV
10.00 mA
06
0.500kV
10.00 mA
07
0.500kV
10.00 mA
08
0.500kV
10.00 mA
CLEAR
09
0.500kV
10.00 mA
10
0.500kV
10.00 mA
SLAVE
1 2 3 4 5 6 7 8 9
STANDBY
REMOTE
LOCK
CORR
ERROR
TEST – 1/2
MODE
CH
SOURCE
LIMIT
RES.
RECALL
01
0.500kV
0.050 mA
02
0.500kV
0.050 mA
OFFSET
03
0.500kV
0.050 mA
04
0.500kV
0.050 mA
GET Cs
05
0.500kV
0.050 mA
06
0.500kV
0.050 mA
07
0.500kV
0.050 mA
08
0.500kV
0.050 mA
09
0.500kV
0.050 mA
10
0.500kV
0.050 mA
SLAVE
1 2 3 4 5 6 7 8 9
TEST: 0.0s
REMOTE
LOCK
CORR
ERROR
Position 1
Position 1
1. Set Auto Range to ON.
2. As Position 1 shown in the figure below setting it to high current range.
AC
Operation
3. If the tested current can be displayed by low current range 0.6 seconds before the test
ends, the current range will change the range to low current range automatically as
Position 1 shows in the figure below.
AC
4.8.7 Hardware/Software AGC
AGC function is used due to load effect (the output voltage changes when the Load
changes.)
ACV : H/W AGC always ON, S/W AGC default is ON but can be set to OFF.
DCV : H/W AGC always ON, S/W AGC default is ON but can be set to OFF.
IR : No H/W AGC, S/W AGC is ON but can be set to OFF.
OSC : No H/W AGC, S/W AGC is OFF.
When the ACV/DCV test time is less than 0.29 second, the H/W AGC is
OFF.
_
Rs
12
+
RLVo
User’s Manual
Vs Vo / Vs = RL / (Rs + RL)
1. H/W AGC: Since the load effect caused Vo<Vs, the hardware comparator circuit is used
to make Vo compensate voltage to be the same as within Vs within 0.1sec.
2. S/W AGC: This tester uses software AGC only in IR MODE. Since the software
compensation is slow, it would not cause transient voltage shock to DUT and the
common IR impedance (R
thus Vo≒Vs.
L) is much larger than the output impedance (Rs) of this tester,
HANDLER Interface
5-1
Pin No.
Signal
Input/Output
Description
1
CHN1
2
CHN2
3
CHN3
4
CHN4
5
CHN5
6
CHN6
7
CHN7
8
CHN8
9
CHN9
10
CHN10
The output signal of PASS_FAIL indicates the
Lo: TOTAL PASS, Hi: TOTAL FAIL.
The output signal of nEOT shows if the test
standby.
It is the external START signal input that starts
when LOW.
It is the external STOP signal input that stops
when LOW.
5. HANDLER Interface
5.1 Introduction
The HANDLER interface socket can be purchased for the Tester rear panel. When it is
desired to control the Tester output via external signal or to send the signal outside, the
HANDLER card can be inserted for external control.
5.2 Specification
5.2.1 Driving Capability
Internal Signal Output Specification: DC 5V, 40~60mA
External Signal Output Specification: DC 3V~26V (HIGH), 10mA± 4mA,current limit is
10mA± 4mA for every circuit.
5.2.2 Pin Assignment
The each pin assignment of HANDLER is listed in the table below:
The output signals of CHN1~10 indicate the
Output
11 PASS_FAIL Output
12 nEOT Output
test results of CH1~10.
Lo: PASS, Hi: FAIL.
test result of all channels when each Step
ends.
procedure is ended.
When the signal is HIGH, it means the test
procedure is under execution.
When the signal is LOW, it means the test
procedure has been ended or the tester is
It is the external DC voltage input. The input
voltage range is between +3V~+26V.
16
nRecall1
nRecall1~nRecall3 signals indicate the memory
signal, the memory data can be retrieved.
17
nRecall2
18
nRecall3
20
+5VHan
It is the internal DC voltage output.
The output signal of nEOS indicates if the test
ended or standby.
It is the external DC voltage input and the low
voltage terminal for input/output signal.
It is the low voltage terminal for internal voltage
output.
This signal will change once when running PA
Mode.
User’s Manual
15 +VEXT
-
Input
19 nRecall4
-
21 nEOS Output
22 GNDF
23 GD3
-
-
position to be read.
It uses 3 bits to present 7 test steps.
The input format is binary code (001~011)
(nRecall1 is the low bit, while nRecall3 is the
high bit.)
001 means to recall memory 1
111 means to recall memory 7
nRecall4 signal is the switch for reading
memory. When nRecall4 inputs a LOW level
is ended.
When the signal is HIGH, it means the test is
undergoing.
When the signal is LOW, it means the test is
It sets the time required for external trigger signal
(/EXT_START) to sustain.
It sets the time for clear from external trigger signal
(/EXT_START) to /EOT signal.
T3
-
It sets the time required for test.
T4
> 5mS
It sets the waiting time for /PASS_FAIL signal to be stable.
It sets the waiting time for /EOT and /EOS signals to be
stable.
T6
Note
It is the time spent to switch step.
T7
Note
It is the time spent for starting 2nd test.
Note
The value is varied by conditions. For instance, the T6 under the condition of
the exact values shall base on the actual measurement.)
/EXT_START
/EOT
/EOS
/PASS_FAIL
/CHN1~10
T1
T2
T3
T3
STEP
STEP
T4
T7
T1
STEP
STEP
T5
T6
T4
T5
T4
T5
Test Process Control
HANDLER Interface
T1 > 10mS
T2 < 20mS
T5 > 5mS
Screen On is about 320mS and under the condition of Screen Off is about 40mS,
while the T7 under the condition of Screen On is about 280mS and under the
condition of Screen Off is about 60mS. (The values mentioned here are estimated,
The function of GPIB interface is to input the ASCII code composed commands in order to do
remote control and setting. The command string is formed by [command+parameter].
Semicolon “;” can be used to connect any two commands with end code at last. The End
Code is in one of the following formats which can be identified by the Tester itself:
End Code
6.2.5 Panel Description
1. Setting Address
●When “SYSTEM” appears on the title bar, press ▲, ▼ to move the highlight to
[SYSTEM CONFIG] and press ENTER to go to SYSTEM CONFIG screen.
●Press ▲, ▼ again to move the highlight to [GPIB] and use the Function Keys [UP]
and [DOWN] to select the GPIB Address.
●When the setting is done, press Function Key [EXIT] to end the setting.
2. Remote & Panel Control
●When the message box “Remote” is reversed it indicates the Tester is in remote
control state.
●When in remote control state, it can use the ENTER key on panel to switch the
Tester to panel control state.
●When in remote control state, all keys are invalid except ENTER (switch to panel
control state) and STOP (reset the Tester) keys.
●The GPIB LLO [Local Lockout] command can be used to make ENTER key invalid.
6.3 RS232 Specification
6.3.1 Data Format
Baud Rate: 9600∕19200∕38400
Transmission Bit: 1 start bit + 8 data bits + 1 stop bit
6.3.2 Command Format
The function of RS232 interface is to input the ASCII code composed commands in order to
do remote control and setting. The command string is formed by [command+parameter].
Semicolon “;” can be used to connect any two commands with end code at last. The End
Code one of the following formats:
GPIB/RS232 Interface (IEEE-488.2)
6-3
LF
CR+LF
Pin No.
Description
1 * Unused
2
TxD
Sending data
3
RxD
Receiving data
4 * Unused
5
GND
Grounding signal
6 * Unused
7 * Unused
8 * Unused
9 * Unused
TxD
RxD
GND
RxD
TxD
GND
2
3
5
2
3
5
2
3
5
2
3
5
9
Pin (female
)
Instrument
9
Pin (male)
9 Pin (
female)9
Pin (male
)
PC
Connecting Wire
12
3
45
67
89
End Code
6.3.3 Connector
The RS232 connector of the Tester is a female 9-pin connector.
6.3.4 Connection
The RS232 connector of the Tester is a female 9-pin connector.
The parameter syntax of SCPI command includes:
(1) Use “< >” to indicate the defined parameter format of standard SCPI command.
(2) “< numeric value >” is a decimal data while “< boolean >” is Boolean program data with
value 0 or 1.
(3) Use vertical bar “|” to indicate parameter OR.
(4) “< channel list >” indicates the Channel status and expression is:
(@C1, C2…) where C1, C2… indicates Channel number. The format of this model’s
Channel number is 3 digits, the first digit is Frame index and the last two seconds are
Channel index, for example, the Channel number of Master (Frame 0) Channel 4 (04) is
It clears the data structure of status in the following actions:
Clear the standard event register.
Clear the byte register except MAV bit (bit 4).
*ESE < decimal data>
It sets the value for standard event enable register. The value is a <decimal data> within
0~255.
*ESE?
It queries the standard event enable register value of device. The output format is
<decimal data> within 0~255.
*ESR?
It queries the standard event register value of device. The register is cleared to 0 when
this command is executed. The output format is <decimal data> within 0~255.
*IDN?
It reads the basic data of device. The output format is divided by comma into 4 columns,
which are manufacturer, device model no., serial no. and firmware version.
*OPC
It completes the operation.
*OPC?
It queries the operation for completeness. An ASCII character “1” is output when done.
*PSC 0 | 1
It clears the power on state.
*PSC?
It queries the power on state for clearing. The output format is an ASCII character “1” or
“0”.
*RST
It resets the device by stopping the test.
*RCL <decimal data>
It is a read back command.
This command reads back the settings saved in the memory of the device. The range of
< decimal data > is between 1~30. (This command is invalid when the value is 0.)
*SAV <decimal data>
It is a save command.
This command is to save the settings at present of the device to memory. The range of <
decimal data > is between 1~30. (This command is invalid when the value is 0.)
*SRE <decimal data>
It sets the value for service request register. The value is a <decimal data> within 0~255.
It reads the value of service request enable register. The output format is <decimal
data> within 0~255.
*STB?
It reads the value of status bit register. The output format is <decimal data> within
0~255.
●SCPI Command
:MEMory:DELete[:NAME] <name>
It deletes the parameter data specified by <name> in main memory. <name> is a string
with maximum 13 characters. .
Example: Input the command “MEM:DEL 123”.
Description: It deletes the parameter data named 123 in main memory.
:MEMory:DELete:LOCation <register number>
It deletes the parameter data specified by <register number> in main memory. <register
number> is an integer with the range between 0~30 where 0 means to clear the memory.
Example: Input the command “MEM:DEL:LOC 1”.
Description: It deletes the parameter data number 1 in main memory.
:MEMory:STATe:DEFine <name>, <register number>
It sets a name for memory specified by <register number>. <register number> is an
integer with the range between 1~30.
Example: Input the command “MEM:STAT:DEF TEST,1”.
Description: It defines the parameter data named TEST of the 1
st
memory set in
main memory.
:MEMory:STATe:DEFine? <name>
It queries the memory’s <register number> specified by <name>.
Example: Input the command “MEM:STAT:DEF? TEST”.
The tester returns “1”.
Description: The returned value “1” indicates the parameter data named TEST is
located at the 1
st
set.
:MEMory:FREE:STATe?
It queries the unused capacity in main memory.
Example: Input the command “MEM:FREE:STAT?”.
The tester returns “27”.
Description: The returned value “27” indicates the remaining data parameters for
set.
:MEMory:NSTates?
It queries the capacity of main memory. The value returned is the parameter maximum
plus 1 for *SAV and *RCL commands.
Example: Input the command “MEM:NST?”.
The tester returns “31”.
GPIB/RS232 Interface (IEEE-488.2)
6-11
String
Returned Data
STEP
The present STEP No.
MODE
The present MODE
OMETerage
The present reading of output meter
MMETerage
The present reading of measure meter
RELapsed
The time executed for RAMP
RLEave
The time remained for RAMP
DELapsed
The time executed for DWELL
DLEave
The time remained for DWELL
TELapsed
The time executed for TEST
When the Test Time is set to CONT, it responds 9.9000001E+37.
TLEave
The time remained for TEST
When the Test Time is set to CONT, it responds 9.9000001E+37.
FELapsed
The time executed for FALL
FLEave
The time remained for FALL
Mode
AC
DC
IR
OSC
ALL
Code
HEX
DEC
HEX
DEC
HEX
DEC
HEX
DEC
HEX
DEC
STOP
70
112
TESTING
73
115
PASS
74
116
HIGH FAIL
21
33
31
49
41
65
LOW FAIL
22
34
32
50
42
66
ARC FAIL
23
35
33
51
OCP
24
36
34
52
44
68
64
100
Description: The returned value “31” indicates the storage capacity of main
It queries the host for the measured result when the Screen sets to On. The format of
variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the
range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master. <item> is string as listed below:
When the Test Time is limited it responds the time remained.
When the Test Time is limited it responds the time remained.
Example: Input the command SAF:CHAN003:FETH? STEP,MODE,OMET
The tester returns 1,AC,+5.000000E+02.
Description: The returned string 1,AC,+5.000000E+02 indicate the result of STEP,
MODE and output result for CH3 of Address 0 host is STEP1, AC
MODE, 0.500kV.
It queries all judgment results when the Screen sets to On. The format of variable <m>
at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the range of 0~9 and
the last two digits are Channel index in the range of 01~10 (19020/19021/19022) or
01~04(19020-4/19021-4/19022-4). When the variable is ignored it is 001 to indicate
Channel 1 of the Master. The return format is First Step Result, Second Step Result, ...,
Last Step Result. The meaning of Code is listed as below:
It queries the readings of MEASURE METER in all STEPs when the Screen sets to On.
The format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame
index in the range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN004:RES:ALL:MMET?
The tester returns 7.000000E-05, 5.000000E-05, 4.000000E-05,
It queries the MODE of all STEPs when the Screen sets to On. The format of variable
<m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the range of 0~9
and the last two digits are Channel index in the range of 01~10 (19020/19021/19022) or
01~04(19020-4/19021-4/19022-4). When the variable is ignored it is 001 to indicate
Channel 1 of the Master. The returned string is AC|DC|IR|OSC|PA.
Example: Input the command SAF:CHAN005:RES:ALL:MODE?
The tester returns OSC,AC,AC,PA,DC,DC,DC,PA,IR,IR.
Description: The returned string is the MODE setting of Address 0 host for CH5
STEP1~STEP10, that is OSC Mode, AC Mode, AC Mode, PA Mode,
DC Mode, DC Mode, DC Mode, PA Mode, IR Mode, IR Mode.
It queries the readings of OUTPUT METER in all STEPs when the Screen sets to On.
The format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame
index in the range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN101:RES:ALL:OMET?
The tester returns 5.100000E+01, 5.000000E+02, 1.000000E+02,
It queries the DWELL time of all STEPs for test when the Screen sets to On. The format
of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the
range of 0~9 and the last two digits are Channel index in the range of 01~10 (19020/
19021/19022) or 01~04 (19020-4/19021-4/19022-4). When the variable is ignored it is
001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:ALL:TIME:DWEL?
The tester returns 5.000000E+00, 5.000000E+00, 5.000000E+00,
It queries the time of voltage fall for all STEPs when the Screen sets to On. The format
of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the
range of 0~9 and the last two digits are Channel index in the range of 01~10 (19020/
19021/19022) or 01~04 (19020-4/19021-4/19022-4). When the variable is ignored it is
001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:ALL:TIME:FALL?
The tester returns 1.500000E+00, 1.500000E+00, 1.500000E+00,
It queries the time required for voltage to ramp up of all STEPs when the Screen sets to
On. The format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is
Frame index in the range of 0~9 and the last two digits are Channel index in the range of
01~10 (19020/19021/19022) or 01~04 (19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:ALL:TIME:RAMP?
The tester returns 5.000000E+00, 5.000000E+00, 5.000000E+00,
It queries the test time of all STEPs when the Screen sets to On. The format of variable
<m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the range of 0~9
and the last two digits are Channel index in the range of 01~10 (19020/19021/ 19022) or
01~04 (19020-4/19021-4/19022-4). When the variable is ignored it is 001 to indicate
Channel 1 of the Master.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:AREPort < boolean > | ON | OFF
It sets if reporting the test result automatically (only valid for RS232 interface.) This
command is only valid for this Tester, so the Address 0 host <m> can use 001, Address
1 host <m> can use 101 and so forth. The Address 9 host <m> can use 901. The
returned data is Total Pass/Fail state and its format is string in “PASS” or “FAIL”.
Example: Input the command SAF001:RES:AREP ON
Description: It means to set the Address 0 host to return the test result
automatically after the test is done. If the test result of all channels of
Address 0 host is PASS, it returns “PASS” or “FAIL” will be returned.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:AREPort?
It queries the device if reporting the test result automatically. The return is 1 or 0.
(Applicable for RS232 interface only.)
Example: Input the command SAF001:RES:AREP?
The tester returns 1.
Description: The returned 1 means Address 0 host will report the test result
automatically once the test is done.
[:SOURce]:SAFety[:CHANnel]<m>:RESult:COMPleted?
It queries the device if all settings are executed completely. The return is 1 or 0.
Example: Input the command SAF001:RES:COMP?
The tester returns 1.
Description: The returned 1 means Address 0 host has completed the execution of
It queries the judgment of specified STEP when the Screen sets to On. The format of
variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the
range of 0~9 and the last two digits are Channel index in the range of 01~10 (19020/
19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is ignored it is
001 to indicate Channel 1 of the Master. Return format: First Step Result, Second Step
Result, ..., Last Step Result. The table below lists the definition of Code:
Code of Test Result
GPIB/RS232 Interface (IEEE-488.2)
6-15
OCP
24
36
34
52
44
68
64
100
SHORT FAIL
61
97
OPEN FAIL
62
98
Example: Input the command SAF:CHAN003:RES:STEP3?
The tester returns 116.
Description: The returned result indicates the judgment of CH3 STEP 3 for
It queries the MEASURE METER reading of specified STEP when the Screen sets to
On. The format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is
Frame index in the range of 0~9 and the last two digits are Channel index in the range of
01~10 (19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN004:RES:STEP3:MMET?
The tester returns 4.000000E-05.
Description: The returned result indicates the MEASURE METER reading of CH4
It queries the OUTPUT METER reading of specified STEP when the Screen sets to On.
The format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame
index in the range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN001:RES:STEP3:OMET?
The tester returns 1.000000E+02.
Description: The returned result indicates the OUTPUT METER reading of CH1
It queries the DWELL time tested by specified STEP when the Screen sets to On. The
format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index
in the range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:STEP1:TIME:DWEL?
The tester returns 5.000000E-01
Description: The returned result indicates the DWELL tested by CH5 STEP 1 for
It queries the time of voltage fall in specified STEP when the Screen sets to On. The
format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index
in the range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:STEP4:TIME:FALL?
It queries the time of voltage ramp in specified STEP when the Screen sets to On. The
format of variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index
in the range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:STEP1:TIME:RAMP?
The tester returns 5.000000E-01.
Description: The returned result indicates the time of voltage ramp in CH5 STEP 1
It queries the test time of specified STEP when the Screen sets to On. The format of
variable <m> at the end of [:CHANnel] is 3 digits. The first digit is Frame index in the
range of 0~9 and the last two digits are Channel index in the range of 01~10
(19020/19021/19022) or 01~04(19020-4/19021-4/19022-4). When the variable is
ignored it is 001 to indicate Channel 1 of the Master.
Example: Input the command SAF:CHAN005:RES:STEP2:TIME?
The tester returns 3.000000E+00.
Description: The returned result indicates the test time required in CH5 STEP 2 for
It queries the judgment of specified STEP for all channels of a Frame when the Screen
sets to On. The range of variable <f> at the end of FRAMe<f> command is 0~9 that
indicates the Frame address. When <f> is ignored the default is 1 it indicates to query
the judgment results of all channels for Slave 1. Return format: Ch1 Result, Ch2
Result, ..., Ch10 Result. The table below lists the definition of Code:
Code of Test Result
Example: Input the command SAF:FRAM0:RES:STEP3?
The tester returns 116,116,116,116,116,116,116,116,116,116.
Description: The returned result indicates the judgment result of STEP 3 for all
It queries the MEASURE METER readings of specified STEP for all channels of a
Frame when the Screen sets to On. The range of variable <f> at the end of FRAMe<f>
command is 0~9 that indicates the Frame address. When <f> is ignored the default is 1
it indicates to query the MEASURE METER readings of all channels for Slave 1.
Example: Input the command SAF:FRAME0:RES:STEP3:MMET?
The tester returns 7.000000E-05, 5.000000E-05, 4.000000E-05,
It queries the OUTPUT METER readings of specified STEP for all channels of a Frame
when the Screen sets to On. The range of variable <f> at the end of FRAMe<f>
command is 0~9 that indicates the Frame address. When <f> is ignored the default is 1
it indicates to query the OUTPUT METER readings of all channels for Slave 1.
Example: Input the command SAF:FRAM1:RES:STEP3:OMET?
The tester returns 5.100000E+01, 5.000000E+01, 5.000000E+01,
It queries the DWELL time tested by specified STEP for all channels of a Frame when
the Screen sets to On. The range of variable <f> at the end of FRAMe<f> command is
0~9 that indicates the Frame address. When <f> is ignored the default is 1 it indicates to
query the DWELL time of all channels for Slave 1.
Example: Input the command SAF:FRAM0:RES:STEP1:TIME:DWEL?
The tester returns 3.000000E+00, 3.000000E+00, 3.000000E+00,
It queries the time voltage fall elapsed of a specified STEP for all channels of a Frame
when the Screen sets to On. The range of variable <f> at the end of FRAMe<f>
command is 0~9 that indicates the Frame address. When <f> is ignored the default is 1
it indicates to query the time elapsed for voltage fall of all channels for Slave 1.
Example: Input the command SAF:FRAM0:RES:STEP4:TIME:FALL?
The tester returns 3.000000E+00, 3.000000E+00, 3.000000E+00,
It queries the time voltage ramp elapsed of a specified STEP for all channels of a Frame
when the Screen sets to On. The range of variable <f> at the end of FRAMe<f>
command is 0~9 that indicates the Frame address. When <f> is ignored the default is 1
it indicates to query the time elapsed for voltage ramp of all channels for Slave 1.
Example: Input the command SAF:FRAM0:RES:STEP1:TIME:RAMP?
The tester returns 3.000000E+00, 3.000000E+00, 3.000000E+00,
It queries the test time of a specified STEP for all channels of a Frame when the Screen
sets to On. The range of variable <f> at the end of FRAMe<f> command is 0~9 that
indicates the Frame address. When <f> is ignored the default is 1 it indicates to query
the test time of all channels for Slave 1.
Example: Input the command SAF:FRAM0:RES:STEP2:TIME?
The tester returns 3.000000E+00, 3.000000E+00, 3.000000E+00,
Description: The returned result indicates the test time required for STEP 2 of all
channels for Address 0 host is 3sec, 3sec, 3sec, 3sec, 3sec, 3sec,
3sec, 3sec, 3sec, 3sec.
[:SOURce]:SAFety:STARt[:ONCE]
It starts the test.
Example: Input the command SAF:STAR
Description: It means to start the host test.
[:SOURce]:SAFety:STARt:CORRection:OPEN GET | OFF
When the parameter is set to GET, it gets the correction value, the host may output high
voltage. When the parameter is set to OFF, correction is disabled.
Example: Input the command SAF:STAR:CORR:OPEN GET
Description: It means to start the function of getting correction value.
[:SOURce]:SAFety:STARt:CORRection:OPEN?
It queries if correction has been done.
Example: Input the command SAF:STAR:CORR:OPEN?
The tester returns 0.
Description: The returned 0 means the correction hasn’t been done on the host.
[:SOURce]:SAFety:STARt:CORRection:SAMPle GET
It starts the GET Cs function in Open Short Check mode.
GPIB/RS232 Interface (IEEE-488.2)
6-19
Example: Input the command SAF:STAR:CORR:SAMP GET
Description: It means to start the GET Cs function in Open Short Check mode.
[:SOURce]:SAFety:STARt:CORRection:SAMPle?
It queries if the GET Cs function is started in Open Short Check mode.
Example: Input the command SAF:STAR:CORR:SAMP ?
The tester returns 1
Description: The returned 1 means the GET Cs function has been started in Open
Short Check mode.
[:SOURce]:SAFety:STATus?
It queries the execution status of present device. The string returned is RUNNING|
STOPPED.
Example: Input the command SAF:STAT?
The tester returns RUNNING.
Description: The returned RUNNING means the host is conduction tests at
[:SOURce]:SAFety:STEP<n>:AC:CHANnel:PSC P | S | PS, S | C | SC
It sets the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:AC:CHAN:PSC P,S
Description: It means to set the scan channel of STEP 1 to PS.
[:SOURce]:SAFety:STEP<n>:AC:CHANnel:PSC?
It queries the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:AC:CHAN:PSC?
The tester returns P,S.
Description: It means the scan channel of STEP 1 is set to PS.
[:SOURce]:SAFety:STEP<n>:AC[:LEVel] < numeric value >
It sets the voltage required for AC withstand voltage test for selected STEP. The unit is
Volt (V).
Range: The range of 19020 is 50~5000 and the range of 19021 is 50~6000.
Example: Input the command SAF:STEP1:AC 3000.
Description: It means to set the voltage required by AC withstand voltage test for
the host STEP 1 is 3kV.
[:SOURce]:SAFety:STEP<n>:AC[:LEVel]?
It queries the voltage required for AC withstand voltage test for selected STEP.
Example: Input the command SAF:STEP1:AC?
The tester returns 3.000000E+03.
Description: The returned 3.000000E+03 means the voltage required by AC
withstand voltage test for the host STEP 1 is 3kV.
[:SOURce]:SAFety:STEP<n>:AC:LIMit:ARC < numeric value >
It sets the ARC test value for selected STEP. The unit is Ampere (A).
Range: 0 or 0.001~0.020, 0 is OFF.
Example: Input the command SAF:STEP1:AC:LIM:ARC 0.004.
Description: It means the ARC test value in AC Mode for the host STEP 1 is 4mA.
[:SOURce]:SAFety:STEP<n>:AC:LIMit:ARC?
It queries the ARC test value for selected STEP.
Example: Input the command SAF:STEP1:AC:LIM:ARC?
The tester returns 4.000000E-03.
Description: The returned 4.000000E-03 means the ARC test value for the host
STEP 1 is 4mA.
[:SOURce]:SAFety:STEP<n>:AC:LIMit[:HIGH] < numeric value >
It sets the high limit of AC withstand leakage current for selected STEP. The unit is
Ampere (A).
Range: The range of 19020 is 0.000001~0.01 and the range of 19021 is
0.000001~0.008.
GPIB/RS232 Interface (IEEE-488.2)
6-21
Example: Input the command SAF:STEP1:AC:LIM 0.01.
Description: It means the high limit of AC withstand leakage current for the host
STEP 1 is 10mA.
[:SOURce]:SAFEty:STEP<n>:AC:LIMit[:HIGH]?
It queries the high limit of AC withstand leakage current for selected STEP.
Example: Input the command SAF:STEP1:AC:LIM?
The tester returns 1.000000E-02.
Description: The returned 1.000000E-02 means the high limit of AC withstand
leakage current for the host STEP 1 is 10mA.
[:SOURce]:SAFety:STEP<n>:AC:LIMit:LOW < numeric value >
It sets the low limit of AC withstand leakage current for selected STEP. The unit is
Ampere (A).
Range: The range of 19020 is 0.000001~0.01 and the 19021 range is
0.000001~0.008 (the low limit of leakage current ≤ the set high limit.)
Example: Input the command SAF:STEP1:AC:LIM:LOW 0.00001.
Description: It means the low limit of AC withstand leakage current for the host
STEP 1 is 0.01mA.
[:SOURce]:SAFety:STEP<n>:AC:LIMit:LOW?
It queries the low limit of AC withstand leakage current for selected STEP.
Example: Input the command SAF:STEP2:AC:LIM:LOW?
The tester returns 1.000000E-05.
Description: The returned 1.000000E-05 means the low limit of AC withstand
leakage current for the host STEP 2 is 0.01mA.
[:SOURce]:SAFety:STEP<n>:AC:TIME:FALL < numeric value >
It sets the time required for the voltage to fall to 0 in selected STEP. The unit is second
(s).
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP1:AC:TIME:FALL 3
Description: It means the time required for set voltage to fall to 0 in the host STEP
1 is 3.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:FALL?
It queries the time required for the voltage to fall to 0 in selected STEP.
Example: Input the command SAF:STEP1:AC:TIME:FALL?
The tester returns 3.000000E+00.
Description: The returned 3.000000E+00 means the time required for the set
voltage to fall to 0 in the host STEP 1 is 3.0sec.
[:SOURce]:SAFety:STEP<n>:AC:TIME:RAMP< numeric value >
It sets the time required for the voltage to ramp up to the set for selected STEP. The unit
second (s).
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP1:AC:TIME:RAMP 5.
Description: It means the time required for testing the voltage to ramp up to set for
It queries the time required for the voltage to ramp to the set for selected STEP.
Example: Input the command SAF:STEP1:AC:TIME:RAMP?
The tester returns 5.000000E+00.
Description: The returned 5.000000E+00 means the time required for the voltage
to ramp up to the set for the host STEP 1 is 5.0sec.
[:SOURce]:SAFety:STEP<n>:AC:TIME[:TEST] < numeric value >
It sets the time required for test for selected STEP. The unit is second (s).
Range: 0 or 0.03~999.9, 0 is CONTINUE.
Example: Input the command SAF:STEP1:AC:TIME 10.
Description: It means the time required for test in host STEP 1 is 10.0sec.
[:SOURce]:SAFety:STEP<n>:AC:TIME[:TEST]?
It queries the time required for test for selected STEP.
Example: Input the command SAF:STEP1:AC:TIME?
The tester returns 1.000000E+01.
Description: The returned 1.000000E+01 means the time required for test in the
It queries the high voltage channel settings set by DEF. CHANNELS options in TEST
CONTROL for selected STEP.
GPIB/RS232 Interface (IEEE-488.2)
6-23
Example: Input the command SAF:STEP2:DC:CHAN:DEF:STAT?
The tester returns 1.
Description: The returned 1 means STEP 2 is using the high voltage channel
settings in DEF. CHANNELS options under TEST CONTROL.
[:SOURce]:SAFety:STEP<n>:DC:CHANnel:PSC P | S | PS, S | C | SC
It sets the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:DC:CHAN:PSC P,S
Description: It means to set the scan channel of STEP 1 to PS.
[:SOURce]:SAFety:STEP<n>:DC:CHANnel:PSC?
It queries the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:DC:CHAN:PSC?
The tester returns P,S.
Description: It means the scan channel of STEP 1 is set to PS.
[:SOURce]:SAFety:STEP<n>:DC[:LEVel] < numeric value >
It sets the voltage required for DC withstand voltage test for selected STEP. The unit is
Volt (V).
Range: 50~6000 for 19020, 50~8000 for 19022.
Example: Input the command SAF:STEP2:DC 4000.
Description: It means the voltage set for the DC withstand voltage test for the host
STEP 2 is 4kV.
[:SOURce]:SAFety:STEP<n>:DC[:LEVel]?
It queries the voltage required for DC withstand voltage test for selected STEP
Example: Input the command SAF:STEP2:DC?
The tester returns 4.000000E+03.
Description: The returned 4.000000E+03 means the voltage required for DC
withstand voltage test for STEP 2 is 4kV.
[:SOURce]:SAFety:STEP<n>:DC:LIMit:ARC < numeric value >
It sets the ARC test value for selected STEP. The unit is Ampere (A).
Range: 0 or 0.001~0.010, 0 is OFF.
Example: Input the command SAF:STEP2:DC:LIM:ARC 0.0025.
Description: It means the ACR test value is set to 2.5mA for the host STEP 2.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit:ARC?
It queries the ARC test value for selected STEP.
Example: Input the command SAF:STEP2:DC:LIM:ARC?
The tester returns 2.500000E-03.
Description: The returned 2.500000E-03 means the ARC test value is set to
2.5mA for the host STEP 2.
[:SOURce:]SAFety:STEP<n>:DC:LIMit[:HIGH]< numeric value >
It sets the high limit of DC withstand voltage leakage current for selected STEP. The unit
is Ampere (A).
Range: 0.000001~0.005 for 19020, 0.000001~0.0035 for 19022.
Example: Input the command SAFE:STEP2:DC:LIM 0.002999.
Description: It means the high limit of DC withstand voltage leakage current is set
to 2.999mA for the host STEP 2.
[:SOURce:]SAFety:STEP<n>:DC:LIMit[:HIGH]?
It queries the high limit of DC withstand voltage leakage current for selected STEP.
Example: Input the command SAFE:STEP2:DC:LIM?
The tester returns 2.999000E-03.
Description: The returned 2.999000E-03 means the high limit of DC withstand
voltage leakage current is set to 2.999mA for the host STEP 2.
[:SOURce:]SAFety:STEP<n>:DC:LIMit:LOW < numeric value >
It sets the low limit of DC withstand voltage leakage current for select STEP. The unit is
Ampere (A).
Range: 0.000001~0.005 for 19020, 0.000001~0.0035 for 19022.
(Low limit of leakage current ≤ set high limit.)
Example: Input the command SAF:STEP2:DC:LIM:LOW 0.000001.
Description: It means the low limit of DC withstand voltage leakage current is set
to 0.001mA for the host STEP 2.
[:SOURce:]SAFety:STEP<n>:DC:LIMit:LOW?
It queries the low limit of DC withstand voltage leakage current for selected STEP.
Example: Input the command SAF:STEP2:DC:LIM:LOW?
The tester returns 1.000000E-06.
Description: The returned 1.000000E-06 means the low limit of DC withstand
voltage leakage current is set to 0.001mA for the host STEP 2.
[:SOURce]:SAFety:STEP<n>:DC:TIME:DWELl < numeric value >
It sets the time required for DWELL for selected STEP. The unit is second (s).
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP2:DC:TIME:DWEL 2.5.
Description: It means the time required for DWELL for the host STEP 2 is 2.5sec.
[:SOURce]:SAFety:STEP<n>:DC:TIME:DWELl?
It queries the time required for DWELL for selected STEP.
Example: Input the command SAFE:STEP2:DC:TIME:DWEL?
The tester returns 2.500000E+00.
Description: The returned 2.500000E+00 mean the DWELL time is set to 2.5sec
for the host STEP2.
[:SOURce]:SAFety:STEP<n>:DC:TIME:FALL < numeric value >
It sets the time required for set voltage to fall to 0 for selected STEP. The unit is second
(s).
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP2:DC:TIME:FALL 3.
Description: It means the time required for set voltage to fall to 0 for selected
GPIB/RS232 Interface (IEEE-488.2)
6-25
STEP is 3.0sec.
[:SOURce]:SAFety:STEP<n>:DC:TIME:FALL?
It queries the time required for set voltage to fall to 0 for selected STEP.
Example: Input the command SAF:STEP2:DC:TIME:FALL?
The tester returns 3.000000E+00.
Description: The returned 3.000000E+00 means the time required for set voltage
to fall to 0 for the host STEP 2 is 3.0sec.
[:SOURce]:SAFety:STEP<n>:DC:TIME:RAMP< numeric value >
It sets the time required for selected STEP to ramp up to the set voltage. The unit is
second (s).
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP2:DC:TIME:RAMP 2.
Description: It means the time required for the host STEP 2 to ramp up to the set
voltage is 2.0sec.
[:SOURce]:SAFety:STEP<n>:DC:TIME:RAMP?
It queries the time required for selected STEP to ramp up to the set voltage.
Example: Input the command SAF:STEP3:DC:TIME:RAMP?
The tester returns 2.000000E+00.
Description: The returned 2.000000E+00 means the time required for selected
STEP 2 to ramp up to the set voltage is 2.0sec.
[:SOURce]:SAFety:STEP<n>:DC:TIME[:TEST] < numeric value >
It sets the time required for test for selected STEP. The unit is second (s).
Example: Input the command SAF:STEP2:DC:TIME 1.
Range: 0 or 0.1~999.9, 0 is CONTINUE.
Description: It means the time required for test for the host STEP 2 is 1.0sec.
[:SOURce]:SAFety:STEP<n>:DC:TIME[:TEST]?
It queries the time required for test for selected STEP.
Example: Input the command SAFE:STEP2:DC:TIME?
The tester returns 1.000000E+00.
Description: The returned 1.000000E+00 means the time required for STEP 2 test
is 1.0sec.
[:SOURce]:SAFety:STEP<n>:DELete
It removes the STEP of <n> and the subsequent STEP moves forward.
Example: Input the command SAF:STEP1:DEL.
Description: It means to clear the settings of STEP in the working memory.
It queries if the high voltage output channel uses the options in DEF. CHANNELS under
TEST CONTROL for selected STEP.
Example: Input the command SAF:STEP3:IR:CHAN:DEF:STAT?
The tester returns 1.
Description: The returned 1 means the high voltage output channel are set using
the options in DEF. CHANNELS under TEST CONTROL for STEP 3.
[:SOURce]:SAFety:STEP<n>:IR:CHANnel:PSC P | S | PS, S | C | SC
It sets the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:IR:CHAN:PSC P,S
Description: It means to set the scan channel of STEP 1 to PS.
[:SOURce]:SAFety:STEP<n>:IR:CHANnel:PSC?
It queries the scan channel status for selected STEP.
Example: Input the command SAF:STEP1:IR:CHAN:PSC?
The tester returns P,S.
Description: It means the scan channel of STEP 1 is set to PS.
[:SOURce]:SAFety:STEP<n>:IR[:LEVel] < numeric value >
It sets the voltage required for insulation resistance test for selected STEP. The unit is
Volt (V).
Range: 50~1000
Example: Input the command SAF:STEP3:IR 1000.
Description: It means the voltage required for insulation resistance test for the host
STEP3 is 1kV.
[:SOURce]:SAFety:STEP<n>:IR[:LEVel]?
It queries the voltage required for insulation resistance test for selected STEP.
GPIB/RS232 Interface (IEEE-488.2)
6-27
Example: Input the command SAFE:STEP3:IR?
The tester returns 1.000000E+03.
Description: The returned 1.000000E+03 means the voltage required for
insulation resistance test for the host STEP 3 is set to 1kV.
[:SOURce]:SAFety:STEP<n>:IR:LIMit:HIGH < numeric value >
It sets the high limit of insulation resistance for selected STEP. The unit is ohm.
Range: 100000~50000000000
Example: Input the command SAF:STEP3:IR:LIM:HIGH 50000000000.
Description: It means the high limit of insulation resistance for the host STEP 3 is
50GΩ.
[:SOURce]:SAFety:STEP<n>:IR:LIMit:HIGH?
It queries the high limit of insulation resistance for selected STEP.
Example: Input the command SAF:STEP3:IR:LIM:HIGH?
The tester returns 5.000000E+10
Description: The returned 5.000000E+10 query the high limit of insulation
resistance for selected STEP. 50GΩ.
[:SOURce]:SAFety:STEP<n>:IR:LIMit[:LOW]< numeric value >
It sets the low limit of insulation resistance for selected STEP. The unit is ohm.
Range: 100000~50000000000 (low limit of insulation resistance ≤ set high
limit.)
Example: Input the command SAFE:STEP3:IR:LIM 100000.
Description: It means the low limit of insulation resistance for the host STEP 3 is
0.1MΩ.
[:SOURce]:SAFety:STEP<n>:IR:LIMit[:LOW]?
It queries the low limit of insulation resistance for selected STEP.
Example: Input the command SAFE:STEP3:IR:LIM?
The tester returns 1.000000E+05.
Description: The returned 1.000000E+05 query the high limit of insulation
resistance for selected STEP. 0.1MΩ.
[:SOURce]:SAFety:STEP<n>:IR:RANGe:UPPer < numeric value >
It selects the upper current range for measurement based on the inputted current. The
unit is Ampere (A).
Example: Input the command SAF:STEP3:IR:RANG 0.0003.
Range: 0.000000001~0.005 for 19020, 0.000000001~0.0035 for 19022.
Description: It means the current for IR measurement in the host STEP 3 is 300uA,
so the upper current range selected for IR measurement is 3mA.
[:SOURce]:SAFety:STEP<n>:IR:RANGe:UPPer?
It queries the upper range set.
Example: Input the command SAF:STEP3:IR:RANG?
The tester returns 3.000000E-03.
Description: The returned 3.000000E-03 means the range set for the host STEP 3
When AUTO is not set, giving the parameter OFF will remain the original set
range. When AUTO is set, giving the parameter OFF will set the range to 5mA.
User’s Manual
[:SOURce]:SAFety:STEP<n>:IR:RANGe[:LOWer] < numeric value >
It selects the lower current range for measurement based on the inputted current. The
unit is Ampere (A).
Range: 0.000000001~0.005 for 19020, 0.000000001~0.0035 for 19022.
Example: Input the command SAF:STEP3:IR:RANG:LOW 0.0003.
Description: It means the current for IR measurement in the host STEP 3 is 300uA,
so the lower current range selected for IR measurement is 300uA.
[:SOURce]:SAFety:STEP<n>:IR:RANGe[:LOWer]?
It queries the lower range set.
Example: Input the command SAF:STEP3:IR:RANG:LOW?
The tester returns 3.000000E-04.
Description: The returned 3.000000E-04 means the range set for the host STEP 3
is 300uA.
[:SOURce]:SAFety:STEP<n>:IR:RANGe: AUTO< boolean > | ON | OFF
It sets if IR range switches to AUTO. Parameter ON or 1 means AUTO while OFF or 0
means AUTO is disabled.
Example: Input the command SAF:STEP3:IR:RANG:AUTO ON.
Description: It means the current range for the host STEP 3 IR measurement is
AUTO.
[:SOURce]:SAFety:STEP<n>:IR:RANGe:AUTO?
It queries if IR range switches to AUTO. 1 means it is set to AUTO while 0 means AUTO
is disabled.
Example: Input the command SAF:STEP3:IR:AUTO?
The tester returns 1.
Description: The returned 1 means the range is set to AUTO for the host STEP 3.
[:SOURce]:SAFety:STEP<n>:IR:TIME:DWELl < numeric value >
It sets the dwell time required for the selected STEP. The unit is second (s).
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP2:IR:TIME:DWEL 2.5.
Description: It means the dewell time set for host STEP 2 is 2.5 sec.
[:SOURce]:SAFety:STEP<n>:IR:TIME:DWELl?
It queries the dwell time required for the selected STEP.
Example: Input the command SAFE:STEP2:IR:TIME:DWEL?
The tester returns 2.500000E+00.
Description: The returned 2.500000E+00 indicate the dwell time set for STEP 2 is
2.5 sec.
[:SOURce]:SAFety:STEP<n>:IR:TIME:FALL < numeric value >
It sets the time required for set voltage to fall to 0 for selected STEP. The unit is second
(s).
GPIB/RS232 Interface (IEEE-488.2)
6-29
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP3:IR:TIME:FALL 3.
Description: It means the time required for set voltage to fall to 0 for the host STEP
3 is 3.0sec.
[:SOURce]:SAFety:STEP<n>:IR:TIME:FALL?
It queries the time required for set voltage to fall to 0 for selected STEP. The unit is
second (s).
Example: Input the command SAF:STEP3:IR:TIME:FALL?
The tester returns 3.000000E+00.
Description: The returned 3.000000E+00 indicate the time required for set voltage
to fall to 0 is 3.0sec.
[:SOURce]:SAFety:STEP<n>:IR:TIME:RAMP< numeric value >
It sets the time required for selected STEP to ramp up to the set voltage. The unit is
second (s).
Range: 0 or 0.1~999.9, 0 is OFF.
Example: Input the command SAF:STEP3:IR:TIME:RAMP 0.5.
Description: It means the time required for the host STEP 3 to ramp up to the set
voltage is 0.5sec.
[:SOURce]:SAFety:STEP<n>:IR:TIME:RAMP?
It queries the time required for selected STEP to ramp up to the set voltage.
Example: Input the command SAF:STEP3:IR:TIME:RAMP?
The tester returns 5.000000E-01.
Description: The returned 5.000000E-01 means the time required for the host
STEP 3 to ramp up to the set voltage is 0.5sec.
[:SOURce]:SAFety:STEP<n>:IR:TIME[:TEST] < numeric value >
It sets the time required for test for selected STEP. The unit is second (s).
Range: 0 or 0.3~999.9, 0 is CONTINUE.
Example: Input the command SAFE:STEP3:IR:TIME 1.
Description: It means the time required for test for the host STEP 3 is 1.0sec.
[:SOURce]:SAFety:STEP<n>:IR:TIME[:TEST]?
It queries the time required for test for selected STEP.
Example: Input the command SAFE:STEP3:IR:TIME?
The tester returns 1.000000E+00.
Description: The returned 1.000000E+00 means the time required for test in STEP
It sets the measurement range and standard capacitance for selected STEP. This
command requires two parameters. The first one is to select measurement range from
1~3 where 1 is the smallest capacitance range. The second parameter is to set standard
capacitance in the unit farad. The valid standard capacitance range is 0.001E-09 ~
500E-09.
Example: Input the command SAF:STEP4:OSC:CORR:SAMP 2,5e-9.
Description: It means OSC is set to use the 2
It queries the measurement range and standard capacitance for selected STEP. It will
return two values separated by comma (,). The first value is measurement range while
GPIB/RS232 Interface (IEEE-488.2)
6-31
the second parameter is standard capacitance in the unit farad.
Example: Input the command SAF:STEP4:OSC:CORR:SAMP?
The tester returns 2,3.000000E-09.
Description: The returned 2 means host STEP 4 use the 2
It sets the percentage of open check when doing OSC for selected STEP. The unit is
100%.
Range: 0.1~1.0
Example: Input the command SAF:STEP4:OSC:LIM 0.3.
Description: It means the percentage of open check when doing OSC for the host
STEP 4 is 30%.
[:SOURce]:SAFety:STEP<n>:OSC:LIMit[:OPEN]?
It queries the percentage of open check when doing OSC for selected STEP.
Example: Input the command SAF:STEP4:OSC:LIM?
The tester returns 3.000000E-01.
Description: The returned 3.000000E-01 means the percentage of open check
when doing OSC for the host STEP 4 is 30%.
[:SOURce]:SAFety:STEP<n>:OSC:LIMit: SHORt <numeric value>
It sets the percentage of short check when doing OSC for selected STEP. The unit is
100%.
Range: 0 or 1~5, 0 is OFF.
Example: Input the command SAF:STEP4:OSC:LIM:SHOR 3.
Description: It means the percentage of short check when doing OSC for the host
STEP 4 is 300%.
[:SOURce]:SAFety:STEP<n>:OSC:LIMit: SHORt?
It queries the percentage of short check when doing OSC for selected STEP.
Example: Input the command SAF:STEP4:OSC:LIM:SHOR?
The tester returns 3.000000E+00.
Description: The returned 3.000000E+00 mean the percentage of short check
when doing OSC for the host STEP 4 is 300%.
[:SOURce]:SAFety:STEP<n>:PAuse[:MESSage] <string data >
It sets the message prompt string for PAUSE mode.
Example: Input the command SAF:STEP5:PA “WAIT”.
Description: It means the message string is WAIT for the host STEPS 5.
[:SOURce]:SAFety:STEP<n>:PAuse[:MESSage]?
It queries the message prompt string set.
Example: Input the command SAF:STEP5:PA.
The tester returns “WAIT”.
Description: The returned “WAIT” means the message is “WAIT” for the host
Description: The returned 6.000000E-01 means the ration set for MIN. VOLTAGE
is 60%.
:SYSTem:TCONtrol:WRANge[:AUTO] < boolean > | ON | OFF
It sets the WV mode to enable or disable auto range function.
Example: Input the command SYST:TCON:WRAN ON
Description: It means the WV auto range function is enabled.
:SYSTem:TCONtrol:WRANge[:AUTO]?
It queries the WV mode if auto range is enabled.
Example: Input the command SYST:TCON:WRAN?
Description: The returned 1 means auto range is enabled.
:SYSTem:TCONtrol:WVAC:FREQuency < number value >
It sets the ACV FREQUENCY. The unit is Hz.
Range: 50/60
Example: Input the command SYST:TCON:WVAC:FREQ 50.
Description: It means the ACV FREQUENCY is set to 50Hz.
:SYSTem:TCONtrol:WVAC:FREQuency?
It queries the setting of ACV FREQUENCY.
Example: Input the command SYST:TCON:WVAC:FREQ?
The tester returns 5.000000E+01.
Description: The returned 5.000000E+01 means the setting of ACV FREQUENCY
is 50Hz
:SYSTem:TCONtrol:WVAC:OFFSET < number value >
It sets AC OFFSET in ampere (A).
Range: 0.000001~0.002.
Example: Input the command SYST:TCON:WVAC:OFFS 0.001.
Description: It means AC OFFSET is set to 1mA.
:SYSTem:TCONtrol:WVAC:OFFSet?
It queries the setting of AC OFFSET.
Example: Input the command SYST:TCON:WVAC:OFFS?
The tester returns 5.000000E-04.
Description: The returned 5.000000E-04 means AC OFFSET is set to 0.5mA.
:SYSTem:VERSion?
It queries the supported SCPI version of this device.
Example: Input the command SYST:VERS?
GPIB/RS232 Interface (IEEE-488.2)
6-37
The tester returns 1990.0.
Description: The returned 1990.0 means the SCPI version supported by this
device is 1990.0.
6.6 Error Messages
●The error messages stored in error queue will be returned in the way of first in first out
(FIFO) which means the first error message returned is the first one being saved.
●When error messages exceed 30, the last one be stored in the error queue will be –350
“Queue overflow”. It means the error queue is unable to store another error message
until the error messages are extracted.
●If no error is generated, +0 “No error” will be stored in the first position of error queue.
-101 Invalid character
There is invalid character in the command.
-102 Syntax error
The syntax is error due to wrong character is used when piping the command.
-103 Invalid separator
There is invalid separator in the command string.
-104 Data type error
The parameter format is incorrect.
-108 Parameter not allowed
The device receives unallowable parameter.
-109 Missing parameter
The parameter is missing.
-111 Header separator error
The command header separator is incorrect.
-112 Program mnemonic too long
The simple command program header exceeds 12 characters.
-113 Undefined header
The device receives undefined command header.
-114 Header suffix out of range
The command header suffix variable is out of range.
-120 Numeric data error
The numeric parameter is incorrect.
-141 Invalid character data
The device receives invalid character data.
-151 Invalid string data
The device receives invalid string data usually the missing double quote.
-158 String data not allowed
The device receives unallowable string parameter.
-168 Block data not allowed
The device receives unallowable block parameter.
-171 Invalid expression error
The device receives invalid parameter of math expression.
-178 Expression data not allowed
The device receives unallowable parameter of math expression.