Chemtrol PC6000, PC3000 Instruction Manual

MANUALPC-2012
© SBCS 2012
i
INSTRUCTION MANUAL
CHEMTROL® PC PROGRAMMABLE
WATER TREATMENT CONTROLLERS
WARRANTY
This CHEMTROL® Controller S/N _________________ is warranted by SANTA BARBARA CONTROL SYSTEMS (SBCS) to be free from defects in manufacturing and workmanship for a period of FIVE (5) YEARS from the date of purchase for the printed circuit boards and ONE (1) YEAR for all other components. SBCS will repair or replace at its option any defective part during the warranty period. Labor, shipping or incidental expenses are specifically excluded from this warranty. For warranty coverage, defective parts should be returned immediately to your CHEMTROL® Dealer or to our factory postpaid with a copy of your purchase receipt and a description of the malfunction.
TECHNICAL SUPPORT US/CANADA OTHER COUNTRIES
AUSTRALIA
EMAIL
support@chemtrol.com.au
800-621-2279 805-683-8833
1-300-585-820
MANUALPC-2012
ORP AUTO Calibrate mV 750 Setpoint mV 700 Alarm Low mV 650 Alarm High mV 850 Time Limit min 30 Run Time 10 125 Last Shock 05/01/98
SANITIZER OFF Calibrate ppm 1.5 Setpoint ppm 1.5 Alarm Low ppm 0.3 Alarm High ppm 3.0 Time Limit min 30 Run Time 10 125 Last Shock 00/00/00
CONTROL TYPE
OFF Manual Automatic Timer
CALIBRATION
1 Point (Zero) 2 Point (Slope) 3 Point (Curve)
ALARM OPTIONS Feed Lockout YES
Buzzer YES
SHOCK & SAVER
Shock Treatment Chemical Saver
ON/OFF Proportional 4-20 mA
Cycle ON 0.1 Cycle OFF 1.0
Deadband
Deadband
Progressive Zone
Top Limit Bottom Limit
pH AUTO Calibrate 7.5 Setpoint 7.5 Alarm Low 7.0 Alarm High 8.0 Time Limit min 30 Run Time 10 25 Probe Clean AUTO
CLEAN ON OFF Mo 19:00 19:05 Tu 00:00 00:00 We 19:00 19:05 Th 00:00 00:00 Fr 00:00 00:00 Sa 19:00 19:05 Su 00:00 00:00
TDS AUTO Calibrate ppm 1650 Setpoint ppm 2000 Alarm Low ppm 50 Alarm High ppm 2500 Time Limit min 30 Run Time 10 25 Select Scale
Cell Constant 1 TDS Factor 0.5
HEATER AUTO Calibrate F 82 Setpoint F 82 Alarm Low F 70 Alarm High F 85 Time Limit min 30 Run Time 15 60 Energy Saver AUTO
Heating Cooling
Acid Feed Base Feed Both
Oxidizer Reducer Both
ENERGY SAVER Mo 19:00 06:00 Tu 19:00 06:00 We 19:00 06:00 Th 19:00 06:00 Fr 19:00 06:00 Sa 19:00 06:00 Su 19:00 06:00
PUMP AUTO P influent psi 25 P effluent psi 20 Flow Rate gp m 800 Alarm low gp m 500 Flow Total Mga 2.20 Heater Delaymi n 20 Water Level AUTO
PUMP ON OFF Mo 05:00 20:00 Tu 05:00 20:00 We 05:00 20:00 Th 05:00 20:00 Fr 05:00 20:00 Sa 05:00 20:00 Su 05:00 20:00
PRESSURE ALARMS Low psi + 10 High psi + 50
FLOWMETER FACTOR Pulses/ga: 6.705
FILTER AUTO Backwash Date 8/10 Start Time 15:30 Filter Time (min)10.0 Advance (min) 1.0 Number of Filters 5 Pump shutdownYES Pump Override YES
CONFIGURATION Initial Setup Operations Reports Communications
Time Interval P Differential Time or Pressure Time & Pressure Flow rate Volume of water Volume or Pressure Volume & Pressure
Number of Filters5 Priority Valve YES
PUMP Override YES Stop Time between valves (sec) 10
COMMUNICATIONS Phone Unit ID Alarm calling YES
INITIAL SETUP Language Units Password Clock Set Readings Data Logging Model Options
MODEL OPTIONS ORP Yes SANITIZER Yes p H Yes CONDUCTIVITY Yes HEATER Yes PUMP Yes FILTER Yes
PHONE NUMBERS 1: ______________ 2: ______________ 3: ______________ 4: ______________ 5: ______________ 6: ______________
SATURATION Alkalinity 150
Hardness 300
pH 7.5 Temperature 80 Langelier Limits Langelier Index Condition OK
OPERATIONS Audio Alarms YES Bypass Line YES Saturation OK Print Reports Reset Battery Probe Monitor YES
Complete Reset Partial Reset
ORP SANITIZER pH CONDUCTIVITY HEATER PUMP FILTER
LANGUAGE English French Spanish
UNITS U.S. Metric
MENU TREE CHEMTROL
(R)
PC CONTROLLERS
Decrease Increase
SHOCK AUTO Date 05/08/98 Cycle (days) 7 Time ON pH 18:00 Level pH 7.8 Time ON shock 21:00 Level ORP 820 Time OFF 21:30
View data log Print Reports
ii
MENU TREE
MANUALPC-2012
1. READ AND FOLLOW ALL INSTRUCTIONS
2. WARNING - To reduce the risk of injury, do not permit children to use this product unless they are closely supervised at all times.
3. A wire connector is provided on this unit to connect a minimum No. 8 AWG solid copper conductor between this unit and any metal equipment, metal enclosures or electrical equipment, metal water pipe or conduit within 5 feet of this unit.
4. DANGER - Risk of injury. a) Replace damaged cord immediately. b) Do not bury cord. c) Connect to a grounded, grounding type receptacle only.
5. WARNING - This product must be connected to a power source equipped with a ground-fault circuit interrupter
(GFCI). The GFCI must be tested before each use. With the product operating, open the service door. If the product stops operating, this merely indicates that the door is equipped with an electrical interlock. Next, push the test button on the GFCI and close the service door. The product should not operate. Now open the service door, push the reset button on the GFCI and close the service door. The product should now operate normally. If the product fails to operate in this manner, there is a ground current flowing indicating the possibility of an electric shock. Disconnect the power until the fault has been identified and corrected.
8. DANGER - Risk of electric shock. Install at least 5 feet (1.5 m) from inside wall of tub or spa using nonmetallic
plumbing.
7. DANGER - Risk of electric shock. Do not permit any electric appliance, such as a light, telephone, radio, or
television, within 5 feet (1.5 m) of a spa or hot tub.
8. WARNING - To reduce the risk of injury:
a) The water in a spa should never exceed 40 ºC (104 ºF). Water temperatures between 38 ºC (100 ºF) and 40 ºC (104 ºF) are considered safe for a healthy adult. Lower water temperatures are recommended for young children and when spa use exceeds 10 minutes.
b) Since excessive water temperatures have a high potential for causing fetal damage during early months of pregnancy, pregnant or possibly pregnant women should limit spa water temperatures to 38 ºC (100 ºF).
c) Before entering a spa or hot tub, the user should measure the water temperature with an accurate thermometer since the tolerance of water temperature-regulating devices varies.
d) The use of alcohol, drugs or medication before or during spa or hot tub use may lead to unconsciousness with the possibility of drowning.
e) Persons suffering from obesity or with a medical history of heart disease, low or high blood pressure, circulatory system problems or diabetes should consult a physician before using a spa.
f) Persons using medication should consult a physician before using a spa or hot tub since some medication may induce drowsiness while other medications may affect heart rate, blood pressures and circulation.
iii
IMPORTANT SAFETY INSTRUCTIONS
Specified by ITS Testing Services for Swimming Pools and Spas
9. SAVE THESE INSTRUCTIONS
MANUALPC-2012
TABLE OF CONTENTS
MENU TREE ...................................................................II
IMPORTANT SAFETY INSTRUCTIONS .......................II
IMPORTANT SAFETY INSTRUCTIONS ......................III
TABLE OF CONTENTS ............................................... IV
LIST OF FIGURES .........................................................3
CHAPTER I - INTRODUCTION ......................................1
Water Maintenance ...........................................................1
Reference Manuals........................................................1
CONTROLLER FEATURES.................................................2
New Features....................................................................2
Models...............................................................................2
Automated Water Treatment .............................................2
Control Functions ..............................................................2
Free Chlorine.................................................................3
Free Chlorine control .....................................................3
ORP control ...................................................................3
Probe Failure Analysis ......................................................3
CONTROL PANEL ...............................................................4
LCD Display Screen ..........................................................4
Display Readings ..............................................................4
Data Entry Keyboard ......... E
Remote Communications ..................................................4
PROGRAM MENUS .............................................................5
Initial Screens....................................................................5
Display Screens ................................................................5
Main Menus and Submenus..............................................5
Alarm Displays .................. Error! Bookmark not defined.
CHAPTER II - INSTALLATION ......................................6
SAFETY NOTICE .................................................................6
UNPACKING........................................................................6
INSTALLATION REPORT....................................................6
TECHNICAL SUPPORT.......................................................6
PRINCIPLE OF INSTALLATION..........................................6
CONTROLLER CABINET ....................................................6
LOCATION...........................................................................6
ELECTRICAL .......................................................................8
Electrical Codes ................................................................8
AC Power Input .................................................................8
Grounding (GFI) ................................................................8
Main Power Interlock.........................................................8
Panel Interlock ..................................................................8
PC Boards.........................................................................8
Power Board .....................................................................8
115V/230V Power Transformer ........E
defined.
Relay Jumpers (PC7000) ..................................................8
rror! Bookmark not defined.
rror! Bookmark not
iv
Relays and Fuses ............. E
Remote Alarm ................................................................... 9
Mother Board ....................................................................9
Sensor Connections.......................................................... 9
Buzzer............................................................................... 9
Display Brightness ............................................................9
Backup Battery................................................................ 11
Remote Communications................................................ 11
Multiple Serial Connections (Option)............................... 11
4-20 mA Converter Boards .............................................12
Field Installation .............................................................. 12
4-20 mA Connections ..................................................... 12
PPM Sensor Board .........................................................12
PLUMBING ........................................................................13
Installation of Sensors..................................................... 13
In-line Installation (2" Pipe) ............................................. 13
Flow Cell Assembly......................................................... 13
Sensor Cell Cabinet (Option) .......................................... 13
Water Flow...................................................................... 14
Paddle Wheel Rotary Flow Switch .................................. 14
PPM SENSORS ................................................................. 15
PPM Sensor Flow Cell .................................................... 15
PPM Sensor Installation.................................................. 15
ORP AND PH SENSORS....................................................16
Properties........................................................................ 16
Packaging .......................................................................16
Sensor cables .................................................................16
Electrical Interference .....................................................16
Storage and Winterizing.................................................. 17
Sensor Warranties ..........................................................17
TEMPERATURE SENSOR ................................................ 17
CONDUCTIVITY SENSOR.................................................17
FLOW SENSOR.................................................................18
Flow Sensor Location ..................................................... 18
Flow Sensor Saddles ...................................................... 18
Flow Sensor Connections ............................................... 18
PRESSURE TRANSDUCERS ........................................... 19
Transducer Connections ................................................. 19
WATER LEVEL SENSOR .................................................. 19
Installation....................................................................... 19
Maintenance ................................................................... 19
Figure 28 - Optical Level Sensor..................................... 19
CHEMICAL FEEDERS....................................................... 20
WARNING: Tank Sizing .................................................. 20
Chemical Feed Pumps.................................................... 20
Carbonic Acid (CO2) Valve ............................................. 20
Gas Chlorinator............................................................... 20
Erosion Feeders.............................................................. 20
Back Diffusion ................................................................. 20
FILTER BACKWASH......................................................... 21
Main Pump Shutoff ......................................................... 21
Single Filter Backwash.................................................... 21
Multiple Filter Backwash .................................................21
Backwash Stager ............................................................ 21
Electrical Valves.............................................................. 21
Hydraulic Valves .............................................................21
Pneumatic Valves ...........................................................21
rror! Bookmark not defined.
CHAPTER III - STARTUP ............................................ 22
MANUALPC-2012 Page 1
CONTROLLER STARTUP .................................................22
Initial Reset .....................................................................22
Configuration Menu .........................................................22
Alarm Buzzer...................................................................22
Bypass Line.....................................................................22
Battery Check..................................................................22
PPM Board Initialization ..................................................22
CHEMICAL CONTROL ......................................................22
Initial Activation of Sensors .............................................22
Bypass Line Test.............................................................22
Water Chemistry Adjustment...........................................22
Water Sampling...............................................................22
pH Calibration .................................................................22
pH Feed (Acid or Base)...................................................23
pH Setpoint .....................................................................23
ORP Calibration ..............................................................23
ORP Setpoint ..................................................................23
Sanitizer Calibration ........................................................23
Time Limits......................................................................23
Shock Treatment .............................................................23
Chemical Saver...............................................................23
WATER SATURATION ......................................................23
CHAPTER IV - CONTROLLER OPERATION..............24
Access.............................................................................24
Default Setup ..................................................................24
8 - CONFIGURATION MAIN MENU ...................................24
8.1 - INITIAL SETUP SUBMENU ....................................25
8.1.1 - Language..........................................................25
8.1.2 - Units .................................................................25
8.1.3 - Code Number ...................................................25
8.1.4 - Clock ................................................................26
8.1.5 - Readings ..........................................................26
8.1.6 - Data Logging ....................................................26
8.1.7 - Model Options ..................................................26
8.2 - OPERATIONS SUBMENU......................................27
8.2.1 - Audio Alarms ....................................................27
8.2.2 - Bypass Line ......................................................27
8.2.3 - Langelier Saturation Index................................27
8.2.4 - Reset ................................................................28
8.2.5 - Backup Battery .................................................28
8.2.6 - Probe Monitor ...................................................28
8.3 - REPORT SUBMENU ..............................................28
8.3.1 - View Data Log ..................................................28
8.3.2 - Print Reports.....................................................29
On-Site Printing ...........................................................29
On-Site Download........................................................29
8.4 - COMMUNICATIONS SUBMENU............................30
8.4.1 - Phone Numbers................................................30
8.4.2 - Unit Identification ..............................................30
8.4.3 - Alarm Calling ....................................................30
8.4.4 - 4-20 mA Output ................................................30
1 - ORP MENU....................................................................31
Operation.....................................................................31
1.1 - Control Mode...........................................................31
1.2 - Display and Calibration ...........................................31
1.3 - Setpoint...................................................................31
1.4 - Low Alarm ...............................................................31
1.5 – High Alarm .............................................................32
1.6 - Time Limit ...............................................................32
1.7 - Run Time ................................................................32
1.8 - Last Shock ..............................................................32
1.8.1 - Shock Treatment ..............................................32
1.8.2 - Deshock............................................................32
1.8.3 - Chemical Saver ................................................32
2 - SANITIZER MENU ........................................................ 33
Input Selection ................................................................33
Free Chlorine Control...................................................... 33
Calculated PPM Readings .............................................. 33
2.1 - Display and Calibration ........................................... 33
Free Chlorine Calibration............................................. 33
Calculated PPM Calibration......................................... 33
2.3 - Setpoint...................................................................34
2.4 - Low Alarm............................................................... 34
2.5 - High Alarm .............................................................. 34
2.6 - Time Limit ...............................................................34
2.7 - Run Time ................................................................34
2.8 - Last Shock .............................................................. 34
2.8.1 - Superchlorination .............................................34
2.8.2 - Deshock ...........................................................34
2.8.3 - Chemical Saver ................................................ 34
3 - PH MENU.......................................................................35
Operation ........................................................................35
3.1- Control Mode ...........................................................35
3.2 - Display and Calibration ........................................... 35
3.3 - Setpoint...................................................................35
3.4 - Low Alarm............................................................... 36
3.5 - High Alarm .............................................................. 36
3.6 - Time Limit ...............................................................36
3.7 - Run Time ................................................................36
3.8 - Probe Clean............................................................ 36
4 - CONDUCTIVITY MENU ................................................ 37
Conductivity and TDS .....................................................37
Conductivity or TDS Displays.......................................... 37
4.1- Control Mode ...........................................................37
4.2 - Display and Calibration ........................................... 37
4.3 - Setpoint...................................................................37
4.4 - Low Alarm............................................................... 37
4.5 - High Alarm .............................................................. 37
4.6 - Time Limit ...............................................................38
4.7 - Run Time ................................................................38
4.8 - Select Scale............................................................ 38
4.8.1 - Scale Selection................................................. 38
4.8.2 - Cell Constant....................................................38
4.8.3 - TDS Factor....................................................... 38
5 - TEMPERATURE MENU ................................................ 39
Operation ........................................................................39
5.1 - Control Mode ..........................................................39
5.2 - Display and Calibration ........................................... 39
5.3 - Setpoint...................................................................39
5.4 - Low Alarm............................................................... 39
5.5 - High Alarm .............................................................. 39
5.6 - Time Limit ...............................................................39
5.7 - Run Time ................................................................39
5.8 - Energy Saver .......................................................... 39
6- PUMP MENU.................................................................. 40
Operation ........................................................................40
6.1- Control Mode ...........................................................40
6.2 - Influent Pressure..................................................... 41
Pressure Alarms.......................................................... 41
Calibration ...................................................................41
6.3 - Effluent Pressure ....................................................41
Differential Pressure ....................................................... 41
6.4 - Flow Rate.................. E
6.5 - Flow Total ................. Error! Bookmark not defined.
6.6 - Heater Delay (Cooldown Safety) ...........................43
6.7 - Water Level Control ................................................ 43
7 - FILTER MAIN MENU..................................................... 44
Filter Backwash............................................................... 44
7.1 - Automatic Backwash............................................... 44
7.1.2 - Pressure Differential.........................................45
rror! Bookmark not defined.
MANUALPC-2012 Page 2
7.1.3 - Time or Pressure Differential ............................45
7.1.4 - Time and Pressure Differential .........................45
7.1.5 – Flow Rate.........................................................45
7.1.6 – Volume of Water ..............................................46
7.1.7 – Volume or Pressure Differential .......................46
7.1.8 - Volume and Pressure Differential .....................46
Safety Shutoff ..............................................................46
7.6 - Priority Valve ...........................................................46
7.7 - Pump Shutdown......................................................46
7.8 - Pump Override ........................................................46
CONTROL SUBMENUS.....................................................47
Features ..........................................................................47
X.1 - Control Type ...........................................................47
X.0.1 - Deadband.........................................................47
X.0.2 - Progressive Zone .............................................47
X.1.3 - Automatic Control.............................................48
X.1.4 - Timer Control ...................................................48
CALIBRATION SUBMENUS..............................................49
X.2 - Calibration Options .................................................49
X.2.1 - One-Point Calibration.......................................49
X.2.2 - Two-Point Calibration.......................................49
X.2.3 - Three-Point Calibration ....................................49
SHOCK AND SAVINGS SUBMENUS ................................50
X.8 - Treatment Selection................................................50
X.8.1 - Shock Treatment Program ...............................50
X.8.2 - Deshock Program ............................................50
X.8.3 - Chemical Saver Program .................................50
Limit Timers (Overfeed Safety) ....................................... 57
Timer Settings .............................................................57
PERIODIC MAINTENANCE............................................... 57
Water Testing.................................................................. 57
Shock Treatment............................................................. 57
Precautions ................................................................. 57
PORTABLE TESTER......................................................... 58
ORP SENSOR TESTING................................................ 58
pH SENSOR TESTING................................................... 58
ORP AND pH SIMULATION ........................................... 58
CONDUCTIVITY AND TEMPERATURE.........................58
TROUBLESHOOTING ................................................. 59
PARTS, ACCESSORIES AND UPGRADES ............... 60
INDEX ........................................................................... 62
CHAPTER V - COMMUNICATIONS ............................51
CHEMCOM
MODEM CONNECTIONS ..........................................................52
COMPUTER SOFTWARE INSTALLATION.....................................52
REMOTE OPERATION .............................................................52
Secondary Units..............................................................52
Duplex Operation ............................................................52
DATA COLLECTION ................................................................53
AUTOMATIC SCANNING ..........................................................53
Direct Scanning...............................................................53
Data Log Download.........................................................54
DATA DISPLAY ......................................................................54
Text Data Display ............................................................54
Graphic Data Display ......................................................54
PROGRAM .....................................................51
CHAPTER VI - MAINTENANCE...................................55
CONTROLLER MAINTENANCE........................................55
Regular Maintenance ......................................................55
The Acid Test ..................................................................55
Sensor Cleaning..............................................................55
PPM Sensor: ...............................................................55
ORP and pH Sensors: .................................................55
Conductivity Sensor.....................................................55
PPM Sensor Storage.......................................................55
ORP and pH Sensor Storage ..........................................55
Sensor Winterizing ..........................................................55
Battery Replacement.......................................................55
Software Upgrade ...........................................................55
CHEMICAL MAINTENANCE..............................................56
Overview .........................................................................56
pH Control .......................................................................56
ORP and Sanitizer Control ..............................................56
MANUALPC-2012 Page 3
LIST OF FIGURES
Figure 1 - Models Comparision Chart.......................................2
Figure 2 - Equilibrium of Free Chlorine.....................................3
Figure 3 - Control Panel and Main Screen ...............................4
Figure 4 - Remote Computer....................................................4
Figure 5 – Installation Schematics............................................7
Figure 6 - Mother Board ........... Error! Bookmark not defined.
Figure 9 - Power Board ............ Error! Bookmark not defined.
Figure 10 - Multiplex Connections ..........................................11
Figure 11 - Installation of Optional Boards .............................12
Figure 12 - 4-20 mA Converter Board ....................................12
Figure 13 - Sensor Installation................................................13
Figure 14 - Flow cell Assembly .
Figure 15 - Sensor Cell Cabinet (Option)............................13
Figure 16 - PPM Sensor Flow Cell .........................................15
Figure 17 - PPM Sensor Package ..........................................15
Figure 18 - ORP and pH Sensors...........................................16
Figure 19 - Temperature Sensor ...........................................17
Figure 20 - Conductivity Sensor .............................................17
Figure 21 - Flow Sensor Installation .......................................18
Figure 22 - Model 2536 Flow Sensor .....................................18
Figure 23 - Model 2540 Flow Sensor .....................................18
Figure 24 - Saddle for FS2536 ...............................................18
Figure 25 - Saddle for FS 2540 ..............................................18
Figure 26 - Pressure Transducer............................................19
..........................................13
Figure 27 - Optical Level Sensor............................................ 19
Figure 28 - Optical Beam Path ............................................... 19
Figure 29 - Water Level Assembly ......................................... 19
Figure 30 - Erosion Feeder Control........................................ 20
Figure 31 - Single Filter Backwash......................................... 21
Figure 32 - Multiple Filter Backwash ...................................... 21
Figure 33 - On-screen Data Log............................................. 28
Figure 34 - Proportional Feed Rate........................................ 48
Figure 35 - CHEMCOM™ Program Menu.............................. 51
Figure 36 - CHEMCOM™ System Setup ............................... 51
Figure 37 - CHEMCOM™ Facility Menu ................................ 51
Figure 38 - CHEMCOM™ Facility Selection ..........................52
Figure 39 - CHEMCOM™ Secondary Units ........................... 52
Figure 40 - CHEMCOM™ Remote Operation Screen............ 52
Figure 41 - CHEMCOM™ Automatic Scanning .....................53
Figure 42 - CHEMCOM™ Scan Facilities ............................. 53
Figure 43 - CHEMCOM™ Manual Scanning.......................... 53
Figure 44 - CHEMCOM™ Recent Scans ............................... 53
Figure 45 - CHEMCOM™ Data Display ................................. 54
Figure 46 - CHEMCOM™ Text Data Display ......................... 54
Figure 47 - CHEMCOM™ Graphic Data Display....................54
Figure 48 - PORTA-PROBETM Portable Tester......................58
Figure 49 - pH / mV Scale...................................................... 58
MANUALPC-2012 Page 1
PLATE COUNT
CHAPTER I - INTRODUCTION
Congratulations on your selection of a CHEMTROL® PC Programmable Controller for your water treatment.
CHEMTROL® Automation uses advanced electronic sensing technology to monitor and control the most control critical parameters in water, such as sanitizer activity, pH, temperature, conductivity or total dissolved solids (TDS). These are all essential for maintaining safe and enjoyable water.
Water Maintenance
The primary purpose of water maintenance is to maintain clean water that meets the bacteriological and physiological requirements of state and local Health Departments, as shown, for example, in the table for the State of California.
Equally important is the protection of the equipment from the aggressiveness of water and its constituents.
The guidelines adopted by the National Spa and Pool Institute (NSPI), as shown in the second table.
For chemical automation, we recommend:
- Adequate filtration with a maximum turnover rate of six hours for a pool or 30 minutes for a spa,
- Proper water balance with a pH between 7.4 and 7.6, alkalinity between 80 and 120 ppm (mg/l) and Langelier Saturation Index between 0 and 0.3,
- Minimum Free Chlorine level of 1.0 ppm (mg/l) ,
- Oxidation-Reduction Potential (ORP) above 650 mV,
- Total Dissolved Solids (TDS) less than 1,500 ppm (mg/l).
- If required, stabilization not exceeding a cyanuric acid
level above 40 ppm (mg/l),
CHEMTROL® PC controllers are designed to automate and facilitate the attainment of these goals.
Reference Manuals
For more information on water treatment, see the Maintenance section in this Instruction Manual or refer to one of the following reference manuals:
- “Basic Pool and Spa Technology
National Spa and Pool Institute (NSPI)
- “Pool/Spa Operators Handbook
National Swimming Pool Foundation(NSPF)
- “Aquatic Facility Operator Manual
National Recreation and Park Association (NRPA).
1986 CALIFORNIA HEALTH CODE
for Swimming Pools
CLARITY
STANDARD at 35o, per ml
TOTAL COLIFORM COUNT per 100 ml
Main drain to be visible from deck.
Less than 200 Bacteria
Less than 2.2 organisms
No eye or skin
PHYSIOLOGICAL QUALITY
irritation. No objectionable physiological effect.
TEST
MIN IDEAL MAX
pH 7.2 7.5 7.6
FREE CL, ppm
Non-stabilized
Stabilized
Spa
1
1.5
2
1.5
2
3
COMBINED CL, ppm 0 0 0.2
BROMINE, ppm 2 3 4
OZONE, ppm 0.01 - 1
ORP, mV 650 750 -
CYANURIC ACID, ppm Outdoor Pool Indoor Pool Spa
30 0
0
40 0
0
TOTAL ALKALINITY, ppm 100 125 150
TDS, ppm 300 1000 3000
Based on Standards for Public Pools and Spas.
Published in 1988 by the
NATIONAL SPA AND POOL INSTITUTE
2
3
5
100 0
100
MANUALPC-2012 CHAPTER I – INTRODUCTION Page 2
CONTROLLER FEATURES
CHEMTROL® PC Controllers are microprocessor-based
programmable controllers designed for automatic control of water chemistry, water balance and filtration in swimming pools, spas, cooling towers and industrial applications.
New Features
This manual introduces two new features:
- On-screen data display, available on all models,
- New solid state PPM sensor for Free Chlorine, available
with PC5000 and PC7000 controllers.
Models
Five different, yet very similar, CHEMTROL® controller models are covered in this manual: PC2100, PC3000/PC5000, and PC6000/PC7000. The latter 4 models use similar hardware
and thus are referred to as PC3000 Product Line.
Please refer to Figure 1 showing the standard features and available options for each model. Note that these models have few differences and can do mostly all the same functions based on how they’re equipped. While PC2100 lacks level and filtration control capability for example, PC5000/PC7000 only have Free Chlorine control included when compared to their otherwise identical counterparts PC3000/PC6000, respectively.
Automated Water Treatment
CHEMTROL® PC controllers are advanced automation
systems for water maintenance. Designed around a sophisticated microprocessor, the controllers display user­friendly menus and submenus on an easy-to-read LCD (Liquid Crystal Display) screen. All the displays and adjustments are accessible from menu screens that are laid out in a logical and intuitive order. They can in fact be used without reference to the instruction manual.
All sensing devices are connected to the central control module. That makes it possible to monitor the status of all operational parameters at a glance. Also, in case of malfunction or alarm, the operator is immediately alerted.
A 24-hour clock/calendar shows the date and time on the main screen.
Control Functions
The following display and control functions are:
- Oxidation-Reduction Potential (ORP or Redox) in
millivolts with programmable oxidizer feed, superoxidation and chemical savings program,
- Sanitizer Concentration in parts per million (ppm) or
milligrams per liter (mg/l) with programmable sanitizer feed, superchlorination and chemical savings program,
- pH from 0 to 14 with capability for acid and base feed,
- Conductivity in microsiemens/cm or Total Dissolved
Solids (TDS) in ppm or mg/l with capability for programmed dumping and replacement of water and for addition of three
different chemical additives, such as inhibitor, different biocides or probe cleaner.
- Heater Control with temperature display in Fahrenheit or Celsius, programmable heater control, and energy savings program.
- Water Recirculation with automatic programming or manual control of main pump, plus display of flowrate in gallons per minute (gpm) or liters per minute (l/m) and cumulative flow in Millions of Gallons (Mga) or cubic meters (m3),
- Water Balance and Saturation Condition derived from the Langelier Saturation Index and showing water balance
conditions as either OK, corrosive or scaling.
- Filtration with display of inlet and outlet pressures and programmable filter backwashing (PC6000/PC7000 only) based on choice of time, inlet pressure, pressure differential or combinations of the above.
OPTIONS PC2100 PC3000 PC5000 PC6000 PC7000
Free Chlorine Control / PPM sensor ORP Control/ Sensor pH Control/ Sensor PPM Display TDS/ Conductivity Control Water Level Control Heater Control Shock Program Chemical Saver Program Langlier Saturation Index Display Safteytimers Energy Saver Program Adjustable Setpoints Proportional Feed Automatic Data Logging Influent/ Effluent Pressures Main Pump Control and Flow Rate Automated Filter Backwash Remote Computer Operation Telephone Voice Communications Recirculation Pump Control Remote Alarm Autodialer for Alarm Callout Sensor Cell Cabinet Flow Cell Assembly Safety Flowswitch 4-20mA Data Output 4-20mA Control Ethernet/ Internet w/ Chemcon Software RS-485 Direct Computer Operation RS-232 for direct download USB Data Port – SD Card Reader for Firmware Updates – Modbus Priority Valve Control - Interlock Waste Pit and Surge Pit Alarms – Multiple Backwash Programs Full Screen Menus Compatible w/ all common Sanitizers and A.C.T. Program 5 Year Electronics Warranty
Standard Feature Optional Add-On – Not Available
Figure 1 - Models Comparision Chart
MANUALPC-2012 CHAPTER I – INTRODUCTION Page 3
Sanitizer Control
CHEMTROL® PC controllers offer two choices for sanitizer control:
- Direct chlorine control using a membrane solid state sensor for Free Chlorine (standard on PC5000/PC7000 and available on PC2100/PC3000/PC6000)
- Oxidation-Reduction Potential control for chlorine or bromine using an ORP sensor with a platinum ring (available on all PC controllers).
Free Chlorine
When chlorine in any form is introduced in water it forms Hypochlorous Acid HOCl, which is called Free Chlorine and is an excellent bactericide.
As shown in Figure 2, HOCl is a weak acid. It dissociates to produce an hypochlorite ion OCl- and a hydrogen ion H+:
HOCl ========= OCl- + H+
Free chlorine in water is therefore always present in two forms:
- Hypochlorous acid HOCl, which is Fast Acting Free Chlorine, and
- Hypochlorous ion OCl-, which is a much slower acting sanitizer.
Note in that same figure that the concentration of HOCl decreases very rapidly with increasing pH in the range of 7 to
8. At a pH of 7.5, it is equally divided between HOCl and OCL-
.
Free Chlorine control
CHEMTROL® PC5000/PC7000 include true Free Chlorine
control using a solid state sensor with a selective membrane for Free Chlorine.
The controller displays the concentration of Total Free Chlorine (HOCl + OCl-) in PPM (parts per million or milligrams/liter). It does not respond to Bromine and it is not affected by oxidizers.
The Free Chlorine display does not vary with pH.
The PPM setpoint default value on the controller is 1.0 ppm, which is recommended to kill germs and bacteria and maintain good water quality. It can be adjusted to meet local conditions and Health Department requirements.
ORP control
The oxidizing activity of the sanitizer (chlorine or bromine) can be monitored with a special electrode for Oxidation-Reduction Potential, called ORP or Redox.
ORP does not measure sanitizer concentration but only the activity of the more active forms: HOCl or HOBr. As shown in Figure 2, their concentration varies with pH. Therefore, for ORP control, it is important to maintain a constant pH in the water.
Figure 2 - Equilibrium of Free Chlorine
ORP readings are measured in millivolts (mV). They are valuable because they provide an excellent index of bacteriological water quality.
In 1971, the third edition of the "International Standards for Drinking Water Quality" of the WORLD HEALTH ORGANIZATION (WHO) stated:
"A redox potential of 650 mV (measured between platinum and calomel electrodes) will cause almost instantaneous inactivation of even high concentrations of virus. Such a potential can be obtained with even a low concentration of free chlorine but only with an extremely high concentration of combined chlorine."
In 1985, a study of thirty commercial spas by the OREGON DEPARTMENT OF HEALTH also confirmed 650 mV to be the
minimum value required to prevent growth of germs and bacteria. Public pools and spas in many European countries are required by law to maintain an ORP level of 750 mV or more.
In the US, the 1988 revision of the Standards for Public Spas published by the NATIONAL SPA AND POOL INSTITUTE (NSPI) states:
"When chlorine or bromine is used as a primary disinfectant, ORP can be used as a supplemental measurement of proper sanitizer activity. The recommended minimum is 650 mV with no ideal and no maximum. The use of ORP testing does not eliminate the need for testing the sanitizer level with standard test kits".
Probe Failure Analysis
All CHEMTROL® PC controllers include the exclusive safety feature called Probe Failure Analysis.
With this proprietary technology (US Patent No. 5,895,565), the controller continuously monitors and tests the dynamic response of the ORP and pH sensors (see page 28). This important feature allows early detection of probe failure, thereby avoiding the dangerous out-of-range conditions that can develop with other controllers that alert the operator only after a failure has actually occured. This could result in serious damage and liability.
MANUALPC-2012 CHAPTER I – INTRODUCTION Page 4
CONTROL PANEL
LCD Display Screen
The operator interacts with the controller using the LCD display screen and the 16-key data entry keypad on the front panel of the controller, both shown in Error! Reference source not found.. Newer controllers come equipped with touch screens.
The display shows different screens each consisting of up to eight lines of text. Each data line can be highlighted using the UP and DOWN arrow keys to navigate to its submenu using the LEFT and RIGHT arrow keys.
A line flashes to alert the operator if an alarm condition for that specific data is met.
For models that do not come equipped with all functions available, only the applicable screens and menus are shown.
The LCD display screen features backlight illumination for better viewing at night and in dark areas. To protect the display, the backlight turns off automatically after a short time and on again as soon as a key is pressed.
The brightness of the LCD display screen is adjusted at the factory before shipping. If required, it can be re-adjusted at any time with a potentiometer on the Mother Board (R16 for PC2100 or R39 for PC3000 product line).
Display Readings
As shown on the Display Screen (Error! Reference source not found.), the operator can check all operating parameters
at a glance.
Line 1 shows an ORP reading of 750 mV with the feed pump
in Automatic feed mode (A) and running (>)
Line 2 shows a Sanitizer level of 1.5 ppm with the feed mode
OFF (X)
Line 3 shows a pH reading of 7.5 with the Acid feed mode in
Automatic (A) and the pump not running (no >)
Line 4 shows the Total Dissolved Solids (TDS) at 801 ppm
and bleed valve control OFF (X)
Line 5 shows a Temperature reading of 82 F with the heater
control on Automatic (A)
Line 6 shows Flow Rates of 1.5 and 800 gpm in the Bypass
line and Main line and the Pump in Manual mode (M)
Line 7 shows an Influent Pressure of 25 psi, an Effluent
Pressure of 20 psi and the backwash program in Automatic mode (A)
Line 8 shows the Date and Time and an indication that the
water saturation condition is OK. The line is highlighted to give access to the Configuration Menu
Figure 3 - Control Panel and Main Screen
emote Communications
R
CHEMTROL® controller features remote operation by computer (Figure 4).
Unlike less advanced controllers, the CHEMTROL® PC operates in true duplex mode with full access to all menus and submenus.
This means that any change on the controller screen is immediately reproduced on the remote computer screen. Similarly, any operation that is performed on the remote computer screen is reproduced immediately on the controller.
Both the controller as well as your computer have the same available commands and interface. This facilitates operator training and allows instant verification and adjustment of all control parameters anywhere from any remote computer. To prevent unauthorized access, the controller could be programmed with different authorization level password protection.
Please see Chapter V – Communications on the CHEMTROL® communication software CHEMCOM
Figure 4 - Remote Computer
TM and how it works.
MANUALPC-2012 CHAPTER I – INTRODUCTION Page 5
WELCOME TO
>ORP
650
mV A
>ORP
710
mV A
PROGRAM MENUS
Initial Screens
When power is applied to the controller, CHEMTROL® PC displays the Welcome Screen shown on the right. It includes the firmware version of the operating software installed on the controller and the factory’s Technical Support phone numbers. Canada has the same toll-free support numbers as the USA.
These screens can also be accessed at any time from the Main Display Screen by pressing the LEFT ARROW key.
Display Screens
The Welcome Screen is followed by the main menu showing the main functions of the controller and their conditions.
The adjustable operational information are shown from left to right as follows:
1. Operational status arrow indicating outlet activation,
2. Function identification: ORP, SANITIZER, pH,
CONDUCTIVITY, TEMPERATURE, PUMP and FILTER,
3. Function/Sensor readings,
4. Units of measurement (US or metric),
5. Function operational mode:
A: Automatic control, M: Manual operation, T: Timer control, X: OFF.
Main Menus and Submenus
As shown on the Menu Tree included in the manual, there are several submenus that are accessed directly from the Main Screen; one for each operating function, and one for system configuration.
Navigation through the menus is done with the four arrow keys located on the front panel keypad. To access any menu or submenu, use the UP and DOWN ARROW keys to highlight the desired function and press the RIGHT ARROW key. To exit from any submenu or screen, press the LEFT ARROW key.
Alarm Displays
Probe failure, out-of-range and overfeed conditions are indicated on the main screen and submenues with flashing lines. For probe failure, the display flashes “Probe” on the corresponding line.
If the display shows an out-of-range or overfeed alarm condition, press the DOWN ARROW key to the flashing line and enter that function’s submenu with the RIGHT ARROW key to determine the cause of alarm.
CHEMTROL® 6000
Version 2.16
TECHNICAL SUPPORT
USA 800-621-2279
PHONE 805-683-8833
FAX 805-683-1893
Welcome Screen
SANITIZER 1.0 ppm X >ACID 7.5 pH A COND 2000 uS X TEMP 82 F A >PUMP 200 gpm A FILTER 25 / 20 psi X 08/10/11 5:30 LSI OK
PC5000/7000 Main Display
SANITIZER 1.4 ppm X >pH 6.7 pH A COND 1730 uS X TEMP 79 F A FLOW 2 4 gpm X
06/03/12 12:47 OK
PC2100 Main Display
MANUALPC-2010 CHAPTER II – INSTALLATION Page 6
CHAPTER II - INSTALLATION
SAFETY NOTICE
See important safety information on the first page of the manual.
UNPACKING
Immediately upon receipt of your shipment, check the shipping carton carefully for damage and report any damage directly to the shipping company. Please report any shortage immediately to the factory.
Before opening the carton, check the outside label and verify the model number, voltage, tee size and feed mode. Unpack the carton carefully and accounting for any of small parts such as PVC fittings.
The controller carton should include the following:
- Controller Cabinet,
- Sensors as required for selected model,
- Instruction Manual and Warranty Card,
- Installation Report to be mailed back to factory upon completion of installation,
- Bypass Line Assembly (Optional).
INSTALLATION REPORT
he Installation Report is a triplicate form designed to assure
T warranty coverage, technical updates and factory support.
1. White copy: to mail back to factory.
2. Pink copy: to Facilities Manager.
3. Yellow copy: to Qualified Dealer.
Upon completion of installation, it must be filled out and signed by the Qualified Dealer and by the facilities manager.
TECHNICAL SUPPORT
lease take your time to read this detailed Instruction Manual
P to insure proper installation and operation. If you need further technical assistance, you can contact your Qualified CHEMTROL® Representative, call our Technical Department toll free at 800-621-2279 or e-mail us at chemtrol@sbcontrol.com.
PRINCIPLE OF INSTALLATION
C
HEMTROL® PC constitutes an integrated command center
for complete monitoring and control of all water treatment operations, including chemistry, filtration, heating recirculation pump and water level.
All information provided by the sensors is processed by the microprocessor on the Mother Board and displayed on the Main display screen. Command signals are then sent to the different control outputs on the Power Board (see ELECTRICAL below).
The schematic of installation in Figure 5 shows the principle of installation with all options installed. Figure 11 show the details of installation for the bypass line and sensor cell assembly.
CONTROLLER CABINET
HEMTROL® PC controllers are contained in a rain proof and
C
splash proof NEMA Type 3 cabinets.
For ease of installation, there are two external mounting ears on the top of the cabinet.
All the electronic and electrical components are mounted inside the cabinet on two separate PC Boards, the Mother Board and the Power Board. Outlets are provided on the bottom of the cabinet for ½" conduits connectors.
LOCATION
ount the cabinet on a wall in a secure location:
M
- More than 10' (3 m) away from the water edge to comply with electrical code requirements,
- Within 10' (3 m) of the main recirculation line or of the bypass line - unless special extension cables are used for the sensors (see Sensor Cables),
- Not exposed to direct sunlight as the LCD display screen will darken at high temperature,
- Easily accessible to maintenance personnel,
- If possible, in a separate or well-ventilated room as far away
as possible from corrosive chemicals and storage tanks,
- At a safe distance from power transformers, pump motors or high voltage power lines,
- Safe from unauthorized access or vandalism
MANUALPC-2010 CHAPTER II – INSTALLATION Page 7
Figure 5 – Installation Schematics
MANUALPC-2010 CHAPTER II – INSTALLATION Page 8
ELECTRICAL
Electrical Codes
INSTALLATION MUST FOLLOW
ALL APPLICABLE ELECTRICAL CODES
The controller is available in either hard-wiring or plug-in configurations. Make sure to use the proper type of wiring according to the local electrical code, usually the same as for the chemical feeders.
The internal wiring and labeling for the Power Board of the controller (Error! Reference source not found.) are as follows:
RELAYS POWER IN
GREEN GROUND GND GND BLACK HOT NO1 L1 WHITE NEUTRAL NO2 L2
AC Power Input
The CHEMTROL® PC is a dual-voltage controller with a voltage selector switch located inside the cabinet on the Power Board. Before connecting the unit to an external power supply, make sure that the voltage selector switch is set to the proper AC power input: 115 V or 230 V.
CAUTION: Damage resulting from improper voltage
selection is not covered by manufacturer warranty.
Grounding (GFI)
grounding lug is provided on the right side of the cabinet. It
A is important to connect it to a proper earth ground to prevent dangerous current leakage and electrical shock. Ground Fault Interruption (GFI) protection is also strongly recommended for all installations.
Main Power Interlock
To prevent accidental chemical feeding, the controller and the chemical feeders should always be interlocked (i.e. wired in parallel) with the manual switch for the main pump. This prevents feeding chemicals when there is no water flow in the recirculation line.
Panel Interlock
For safety of operation, a panel interlock switch is mounted inside the cabinet to shut off all internal power when the control panel is open.
DO NOT ATTEMPT TO DEFEAT ITS PURPOSE !!!
PC Boards
CHEMTROL® PC controllers use two PC boards connected together with a ribbon connector.:
1) A Mother Board behind the front panel containing the
software, memories, main processor and operating electronics,
.
2) A Power Board in the back of the cabinet containing the
power input and all output relays.
The components on the two boards are labeled with their respective identifiers and/or descriptions.
The PC boards are protected with a 1A fuse located on the Power board. If the fuse has to be replaced, make sure to use a 1A fuse only. The use of a larger fuse may cause irreparable damage to the electronic boards.
Power Board
CHEMTROL® PCs are equipped with a switchable dual voltage power transformer for 115v or 230v, and is mounted on the Power Board inside the cabinet.
Always verify that the switch is set to the correct voltage. Connecting the controller to higher voltage than its settings may cause damage to the electronics that is not covered by the manufacturer’s warranty.
Jumpers, Relays and Fuses
Jumpers on the Power Board determine the connections for the output relays and should only be changed by a Qualified CHEMTROL® dealer.
All output relays are fused on the Normally Open (NO) and Normally Closed (NC) sides. Both the Hot and the Neutral sides are fused with 5A Slow Blow fuses.
Make sure not to overload the relays. Chemical feed pumps normally draw less than 5A. If a pump draws more than 5A, it will need a motor starter or a magnetic switch.
Relays on your Power Board are found only for the specific options your PC is configured for.
Fuses for the Power Supply are AGC-1 Fast Blow. The two fuses for the dialup modem if present are 250mA Fast Blow. All other fuses for relay outputs are 5A Slow Blow.
The Mother Board is mounted directly behind the face panel of the controller and contains all the low voltage circuitry including the microprocessor and program chips, the LCD display and the keyboard pad. It is also used to connect all the sensor inputs.
MANUALPC-2010 CHAPTER II – INSTALLATION Page 9
Remote Alarm
The remote alarm is a 5A DPDT relay located on the upper right corner of the Power Board. The remote alarm relay can be set for dry or hot contacts, or for any external signal.
To avoid damaging the Power Board, make sure to use the right type of contacts. Call your dealer or the factory if you are not sure.
With hot contacts, the controller powers the alarm with 110 or 230V, depending on the setting of the input voltage of the controller (see preceding page). Connect the leads to the alarm to the Normally Open contacts (NO1 and NO2) on the terminal strip located next to the alarm relay.
For an external power source, wire the input power to the terminals marked NC1 and NC2. Wire the remote alarm to the normally open contact (NO1 and NO2). The alarm voltage will be the same as the external power source.
Mother Board
he Mother Board is mounted directly behind the face panel of
T the controller and contains all the low voltage circuitry including the microprocessor and program chips, the LCD display and the keyboard pad. It is also used to connect all the sensor inputs.
The key electronic components are the microprocessor and the programmable chips for Program, Display and Memory. The program chips are located in the center at the very top of the board. They can be replaced for upgrading of the software program, which should be done only by an experienced technician.
Sensor Connections
ll sensor connections are on the Terminal Barrier strips on the
A Mother Board, as shown on Error! Reference source not found.. The pH and ORP sensors are connected externally to the bulkhead BNC connectors on the left side of the cabinet.
The temperature sensor has two leads, black and red. The combined conductivity/temperature sensor has four leads: white, green, red and black. Connect all leads as shown on the Mother Board schematic (Error! Reference source not found.).
The color coding for the connections are as follows.
TB1 - Level control sensor and blade-type flow switch (PC3) or high alarm sensor (PC6) 5 = Low White wire from level sensor 4 = Ground 3 = Flow switch NO f/s wire or white wire high alarm 2 = Ground Black wires 1 = +5vdc Red wires from level sensors
TB2 - Pressure transducer 4 = In White wire from influent transducer 3 = Out White wire from effluent transducer 2 = Ground Black wire 1 = 24vdc Red wire from transducers
TB3 - Bypass line flow pulser 3 = +5VDC Red wire 2 = Signal White wire 1 = Ground Black wire
TB4 - Main flow pulser (Signet) 3 = + Black 2 = Signal Red 1 = Ground Shield
TB5 - Temp & TDS, pH, ORP 8 = TDS Black
7 = TDS Red 6 = Temperature White Red 5 = Temperature Green Black 4 = pH Shield 3 = pH Signal 2 = ORP Shield 1 = ORP Signal
TB6 - RS-485 Communications
Connect to A and B only, for multiple serial installations, using Host/Slave configurations (see Remote Communications below).
TB7 - RS-232 Communications
5 = Green 4 = Brown 3 = Black 2 = White 1 = Red
Buzzer
The buzzer is located near the bottom of the Mother Board, as shown on Error! Reference source not found.. It can be turned on for specific alarm conditions through the software program or for all alarms using the Audio Alarm Submenu
8.2.1 (see page 27).
Display Brightness
MANUALPC-2010 CHAPTER II – INSTALLATION Page 10
The brightness of the display can be adjusted with the potentiometer marked R-39 that is located in the center of the Mother Board (Error! Reference source not found.).
MANUALPC-2010 CHAPTER II – INSTALLATION Page 11
Backup Battery
The 3V Backup Battery is located on the left side of the Mother Board (Error! Reference source not found.). It is used to maintain the memory settings in case of loss of AC power. This battery is designed to last for several years in normal operation.
Under normal conditions, the controller will operate without battery power. However, the clock and other memory settings will have to be restored in case of complete power shutdown. Replace the battery if the voltage falls below 2.6 V. The voltage is displayed in Configuration Menu / Battery Submenu.
Remote Communications
he Remote Communications option connects the modem with
T a standard US-type, 6-position RJ12 phone jack located on the center right hand side of the Mother Board (Error! Reference source not found.).
If possible, connect the controller to a direct outside telephone line dedicated for remote operation. If the line is also used for voice communications, users should wait for at least three rings to allow the modem to answer a call.
The modem is a combination Data/Fax/Voice multimedia device registered by the Federal Communications Commission (FCC), Number B46USA-22429-MN-E.
MODEM SPECIFICATIONS
FCC Registration Number
B46USA-22429-MN-E
The Ring Equivalency Number (REN) is 0.2 A. Most telephone companies require that the sum of all devices connected to a telephone line do not exceed 5. If a problem arises as a result of operating this equipment, you may have to provide information about this modem to the Phone Company or to the FCC. If the equipment causes disruption to the telephone network, the Phone Company may disconnect your service.
Multiple Serial Connections (Option)
Up to thirty (30) CHEMTROL® PC controllers can be multiplexed to a single phone line or a single computer line by using a network consisting of a host and several slaves. The connections are made through RS485 Serial Ports (Option RS485).
Different models of CHEMTROL® PC controllers can be mixed and matched in a network. Each controller has its own ID number. This number is set up through software in the Communications Submenu 8.3.1 (see Page 30)
Ring Equivalency Number
(REN) 0.2 A
As shown in Figure 6, the host controller includes a modem and an RS485 connection (Option REM). It is connected to the telephone line by modem or to a computer equipped with an RS485 communication card. The slave units need only an RS485 connection (no modem needed) and an internal ID number.
Any controller can be set up at any time as either host or slave by orienting the jumper JP9 as marked on the motherboard (Error! Reference source not found.).
The slave units must be located within 3,000 feet of the host controller and connected with two Category 5 wires (one pair). The RS485 terminal has four terminals, only two of which, marked A and B, are used. Each unit must be wired A to A and B to B for proper communications. The slave units can be connected directly to the host unit or through any other slave units.
CAUTION: Do not wire the connections marked 24 VDC or
GND on the RS485 terminal. This could cause serious damage to the terminal.
Figure 6 - Multiplex Connections
MANUALPC-2010 CHAPTER II – INSTALLATION Page 12
4-20 mA Converter Boards
The 4-20 mA Converter Boards are two optional piggyback boards. They convert the digital outputs of the controller (sensor or control outputs) into analog signals that can be used by analog monitoring and control equipment.
The two boards are identical and are normally installed at the factory on the motherboard before shipping, as shown in (Figure 7). The location on the motherboard determines the function, i.e. sensor signal or control outputs. Either one of the boards or both can be installed, depending on requirements.
If the boards are installed properly, the controller software automatically shows the 4-20 mA menu line in the Submenu
8.3 - Communications.
Field Installation
or field installation, turn off all power to the controller.
F Position the converter board on top of the motherboard as shown on Figure 7. Press the electrical connector J21 into the socket of the motherboard marked JP1 (or JP2 depending on option) and the three plastic standoffs fittings into the three corresponding holes on the motherboard.
4-20 mA Connections
Figure 8 shows the connections for each Converter Board. Determine the type of signal required by the host system. There are two types of signals used in 4-20mA communications, Internal source or external sink. Each communication channel has four sets of jumpers that must be shunted to the proper setting. (Please note that the factory settings are generally for internal source).
The 4-20mA output signals from the board are located on the terminal boards marked TB1 and TB2. Each channel has a + and - indication corresponding to the markings on the board.
The 4-20mA Signal Board has five communication channels for pH, ORP, TDS, temperature and PPM readouts. Each analog signal requires two wires for connection to the central monitoring system.
The 4-20mA Control Board also has five pump control channels for acid, sanitizer, oxidizer, deoxidizer and base feed. Each analog control signal requires two wires for connection to the corresponding pump.
PPM Sensor Board
For new controllers, the PPM sensor board is already installed on the lower left quadrant of the motherboard, as shown in Error! Reference source not found..
For upgrades and retrofits, the PC board can be easily installed on the motherboard.
At the bottom of the keypad between S15 / S16 locate the jumper marked JP3. Insert the black pin receptacle on the back of the PPM board into the JP3 pin connector. The two plastic standoffs located on opposite sides of the pin receptacle will snap into the two holes on the motherboard to secure it.
Figure 7 - Installation of Optional Boards
Figure 8 - 4-20 mA Converter Board
MANUALPC-2010 CHAPTER II – INSTALLATION Page 13
PLUMBING
This section covers the installation of the sensors and the connection of the chemical feeders or control valves.
All chemical injection should be done on the return line, i.e. downstream of the sensors and pool equipment, as explained in the Chemical Feeders section.
Installation of Sensors
Figure 9 - Sensor Installation
The CHEMTROL® PC controllers use up to nine different sensors for measurement of water chemistry, temperature, flow rate, pressure and water level:
- amperometric sensor for Free Chlorine,
- potentiometric sensors for pH and ORP,
- thermistor for temperature,
- conductivity sensor for Total Dissolved Solids,
- Hall effect pulse generator for flow rate,
- piezoelectric sensors for influent and effluent pressures,
- electro-optical water level sensor.
The first five sensors measure water chemistry (Free Chlorine, ORP, pH, conductivity and temperature). These can be mounted directly on a 2-inch main recirculation line using PVC reducing tees Figure 9). On larger diameter lines, they must be mounted on a ½-inch bypass line - using a Sensor Cell Cabinet (Figure 11).
In-line Installation (2" Pipe)
n smaller installations (2 “ pipe diameter), the sensors can be
O mounted directly on the main recirculation line between the strainer and the pump (Figure 9).
Use only 2x2x½ in. SST reducing tees without reducers. Do not install the sensors near an elbow or a constriction where there might be excessive turbulence.
Install the tees on the suction side of the pump and make sure that the tip of the sensor is oriented downward - as shown in Figure 9 - to avoid formation of air pockets near the tip. The sensors should be readily accessible for servicing but not exposed to physical damage.
After inserting the sensor, be careful not to overtighten the compression fitting as it can crush the small glass tube inside the sensor. Make it finger tight (no wrench).
Flow Cell Assembly
Figure 10 - Flow cell Assembly
For ease of installation and maintenance, the sensors should be mounted on the Flow Cell Assembly, as show above. It includes a sensor cell with an air vent and a clear cover, two compression fittings for the ORP and pH sensors, a water sampling tap and two ball valves for controlling the water flow in and out.
Sensor Cell Cabinet (Option)
Figure 11 - Sensor Cell Cabinet (Option)
MANUALPC-2010 CHAPTER II – INSTALLATION Page 14
For even greater ease of installation and maintenance, the components of the bypass line assembly can be supplied in a pre-plumbed Sensor Cell Cabinet (also called Wet Box). The Flow Cell Assembly is mounted in a fiberglass cabinet containing the sensor cell assembly. Also included is a Y-filter, a flowmeter and a paddle wheel safety flow switch. Install on a ½" bypass line, as shown in Figure 11,
Make sure that the Sensor Cell Cabinet is located within 2' (60 cm) of the controller cabinet or order sensor extension cables (see Sensor Cables). See wiring instructions in the ELECTRICAL section and operational instructions under WATER FLOW below.
Water Flow
Proper flow of water past the sensors is essential to obtaining good readings. To check the water flow in the bypass line, start the main recirculation pump. Open both the intake and the return valves on the bypass line and read the flow rate on the flowmeter. It should be in the middle of the range, i.e. about 2-3 gpm (about 8 to 12 l/min). If the water flow is too high, reduce it by closing down the valve on the RETURN SIDE of the bypass line. If there is no water flow, replumb the bypass line as shown on the schematic.
NOTE: The most common installation problems with bypass
line or wet box installations are caused by faulty hydraulics.
To ensure proper water flow, make sure that the bypass line is properly connected. The intake side should be off the pressure side of the recirculation system, i.e. after the filter. The return side should be to a low-pressure area - such as the vacuum side before the recirculation pump, or downstream after the heater, or atmospheric pressure in the pit of a vacuum sand filter or balancing tank.
Paddle Wheel Rotary Flow Switch
The Rotary Flow Switch is a safety switch for the bypass line.
The switch contains a transducer that generates an electrical signal proportional to the water flow. The relay trip point is factory adjusted for a minimum flow rate of 1 gpm (about 4 l/min). (Make sure to use the flow restrictor included in package).
CAUTION: Improper wiring will result in switch burnout
(not covered under warranty).
The black, red and white leads should be connected to the contacts marked “BYPASS FLOW PULSER” on the terminal board marked TB3 on the right side of the Mother Board (Error! Reference source not found.), as indicated:
BLACK Ground WHITE Signal RED + 5VDC or +24 VDC
MANUALPC-2010 CHAPTER II – INSTALLATION Page 15
PPM SENSORS
The following new, improved PPM Sensors (Model
2010) are available with PC controllers:
- PPM002 for 0-2 PPM (mg/l) of Free Chlorine
- PPM010 for 0-10 PPM (mg/l) of Free Chlorine
Both use the same membrane Part Number PPMM01
- PPM200 for 0-200 PPM (mg/l) of Free Chlorine
ith membrane part number PPMM02
w
- PPMSLT 0-10 PPM Membraneless Sensor for electrolytic chlorine generators.
NOTES
1. These new sensors are not affected by cyanuric acid - therefore the new CH255/265 (Version 2010) have no GAIN FACTOR adjustment.
2. All sensors use the same flow cell PPMCLL and do not require the plastic spacer ring.
PPM Sensor Flow Cell
Figure 12 - PPM Sensor Flow Cell
The PPM Sensor must be installed in the specially designed flow cell for better water flow control. Install the flow cell on a bypass line with 3/8” tubing as shown above.
Make sure that the bypass line is located after the filter and that there is a sufficient but not excessive flow of water. The recommended flow rate is between 20 to 100 l/h (0.1 to 0.6 gal/min.) with a best value of about 30l/h (0.2 gal/min).
SPAS
For a spa, it is recommended to increase the flow rate sufficiently to avoid the formation of air bubbles and to install the cell horizontally with the inflow coming in from the bottom.
PPM Sensor Installation
The new PPM Sensors are shipped in a molded foam package containing:
- One (1) chlorine sensor with warranty serial number,
- One (1) compression fitting with O-ring,
- One (1) electrolyte cap with membrane,
- One (1) electrolyte bottle and screw-on filling tip,
- One (1) screwdriver (not needed if pre-wired).
In order to prevent serious damage to the sensor, be careful not to touch the membrane or the electrodes. To facilitate shipment and storage of the sensor, it is not fitted with the electrolyte cap until ready for installation. The sensor cap must be filled with the electrolyte solution immediately before usage and emptied during storage.
Fill the sensor cap with the electrolyte taking care to prevent air bubbles. Fill the cap to the bottom of the threaded section. Thread the sensor into the cap as far as it will go while making sure not to touch the membrane film at the bottom with your fingers. Excess air and electrolyte may escape through the hole below the rubber seal. Wipe any excess electrolyte with a soft tissue. Contact your CHEMTROL additional electrolyte.
Slide the O-ring over the sensor tip up to the compression ring. Finally, insert the sensor with O-ring in the flow cell and make it finger tight.
For best performance, it is recommended to change the membrane cap and electrolyte at least once a year.
PPMSLT SENSOR
The PPMSLT sensor does not use a membrane and has a different open-through cap that holds the electrolyte solution around the electrodes glass shaft. Because this cap is opened from both sides, the procedure to fill it with the necessary electrolyte is slightly different in that you must do so with the accompanied lid on before fully screwing that cap onto the electrodes then finally removing that lid.
The PPMSLT sensor is the only sensor that requires cleaning of its gold electrodes tip and you do so with the small special abrasive cleaner the sensor is shipped with.
Figure 13 - PPM Sensor Package
(R)
dealer if you need
Loading...
+ 50 hidden pages