Checkline TMT-425 User Manual

TMT-425 / TMT-425-USB
TEXTILE MOISTURE METER
OPERATING INSTRUCTIONS
13.0 WARRANTY
Electromatic Equipment Co., Inc. (Electromatic) warrants to the original purchaser that this product is of merchantable quality and confirms in kind and quality with the descriptions and specifications thereof. Product failure or malfunction arising out of any defect in workmanship or material in the product existing at the time of delivery thereof which manifests itself within one year from the sale of such product, shall be remedied by repair or replacement of such product, at Electromatic’s option, except where unauthorized repair, disassembly, tampering, abuse or misapplication has taken place, as determined by Electromatic. All returns for warranty or non-warranty repairs and/or replacement must be authorized by Electromatic, in advance, with all repacking and shipping expenses to the address below to be borne by the purchaser.
THE FOREGOING WARRANTY’S IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY AND FITNESS FOR ANY PARTICULAR PURPOSE OR APPLICATION. ELECTROMATIC SHALL NOT BE RESPONSIBLE NOR LIABLE FOR ANY CONSEQUENTIAL DAMAGE, OF ANY KIND OR NATURE, RESULTING FROM THE USE OF SUPPLIED EQUIPMENT, WHETHER SUCH DAMAGE OCCURS OR IS DISCOVERED BEFORE, UPON OR AFTER REPLACEMENT OR REPAIR, AND WHETHER OR NOT SUCH DAMAGE IS CAUSED BY MANUFACTURER’S OR SUPPLIER’S NEGLIGENCE WITHIN ONE YEAR FROM INVOICE DATE.
Some State jurisdictions or States do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you. The duration of any implied warranty, including, without limitation, fitness for any particular purpose and merchantability with respect to this product, is limited to the duration of the foregoing warranty. Some states do not allow limitations on how long an implied warranty lasts but, not withstanding, this warranty, in the absence of such limitations, shall extend for one year from the date of invoice.
Electromatic Equipment Co., Inc. 600 Oakland Ave. Cedarhurst, NY 11516 - USA Tel: 1-800-645-7330 / Tel: 516-295-4300 / Fax: 516-295-4399
Every precaution has been taken in the preparation of this manual. Electromatic assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of information contained herein. Any brand or product names mentioned herein are used for identification purposes only, and are trademarks or registered trademarks of their respective holders.
TABLE OF CONTENTS
1.0 Powering on …………………………………………………………… 2
2.0 Calibration curves …………………………………………………..… 2
3.0 Operation of the instrument ………………………………………..… 3
4.0 View of the instrument ………………………………………………… 4
5.0 Menu level overview ………………………………………………..… 5
6.0 Measuring procedure ……………………………………..………….… 6
7.0 PC connection and data print ………………………………...… 7
8.0 Device maintenance instructions ……………………………….… 9
9.0 Changing the batteries …………………………………………..… 9
10.0 Important information ………………………………………………… 10
11.0 Exemption of liability ………………………………………………… 10
12.0 Technical information …………………………………………………... 11
13.0 Warranty ……………………………………………………………… 12
Electromatic Equipment Co., Inc.
-12-
Electromatic Equipment Co., Inc.
-1-
1.0 POWERING ON
Textil
e Measuring range
Turn the gauge on: keep the key pressed for three seconds. If a menu item will be shown, press exit or the rolling menu until the measuring window is
displayed.
2.0 CALIBRATION CURVES
Wool 13% - 26% Cotton 1,5% - 13% Rayon 4% - 26%
Flax yarn 5,5% - 15% Jute yarn 8% - 24%
Hemp Garn 5% - 17%
Acetate 3,5% - 13%
Nylon 1,5% - 6,5%
Polyacrylic 0,2% - 3%
Polyester 0,5% - 4%
Digit 0 - 100
The list of calibration curves is displayed if one of the two arrow buttons is pressed for about three seconds in the measuring mode. Select the desired calibration curve and confirm with
.
Note: The humimeter TMT-425 shows the water content directly on the display. A conversion to the moisture value is not necessary anymore. A factory-set conversion to the moisture content is possible. Please contact the manufacturer about this.
Definition water content: The water content is declared as the amount of water in a material, which stands in relationship to the whole weight of the material. For example: You have 1kg material with 40% water content; the material consists of 0,4kg water and 0,6kg material.
Definition material moisture: The material moisture is declared as the amount of water in a material, which stands in relationship to the weight of the dried material. The same example: You have 0,6kg material with 0,4kg water; the 0,6kg material represents 100%, so the 0,4kg water will be 66,7% material moisture.
12.0 TECHNICAL DATA
Resolution 0,1% material moisture 0,5°C temperature Measuring range 0,3 to 26% water content Operating temperature range 0°C to 40°C Temperature measuring range -20°C to 60°C Storage temperature range -20°C to 60°C Temperature compensation Automatic Power supply 4 pieces 1,5Volt AA Alkaline Batteries
(ca. 1000 measuring cycles)
Power off time after approx. 6 minutes Power consumption 55 mA (with light) Display 128 x 64 Matrixdisp., with LED Backlight Dimensions 740 x 65 x 40 mm Weight approx. 450g (with batteries) Protection class IP 40 Scope of delivery Gauge TMT-425
4x1,5Volt AA Alkaline Batteries
Plastic case, electrodes
Optional Humimeter USB data plug for PC
connection
Option 2 (only with opt.1) Thermal printer
Electromatic Equipment Co., Inc.
-2-
Electromatic Equipment Co., Inc.
-11-
10.0 IMPORTANT INFORMATION
Common reasons for incorrect measurements:
Product temperature out of application range Material below 0°C resp. above +40°C (32 to 104 °F ) may cause faulty
measurements. The storage of cold material in a warm storage area usually creates condensed water which may lead to major measuring errors.
Not adjusted temperature of the device Make sure that your measuring gauge and the test material are stored under the same temperature conditions. Protect your gauge from direct sunlight. A too high temperature difference can affect the accuracy of your gauge negatively
Wrong calibration curve
Before you measure the sample, check the correct selection of the calibration curve.
Wet or moldy test material
Moisture/Water on the measuring head
After a measurement of wet material, a water film can be left on the measuring head. This can result in higher moisture values at the next measurement. Therefore, clean all plastic parts of the measurement head with a dry towel.
Warning: Tip of electrodes can lead to injury!
Keep away from children under 16 years.
11.0 EXEMPTION OF LIABILITY
The manufacturer is not liable for any incorrect measurements and resulting damages. As this rapid measurement procedure is influenced by product specific conditions, we recommend to carry out a plausibility check on the measurement results. Each instrument has a serial number and warranty seal. If this is broken, we cannot provide warranty. If the instrument is defective, contact us via www.checkline.com.
3.0 OPERATION OF THE INSTRUMENT Switch on the device: Press the
Power off: Press the
key 5 Seconds in the measuring window; the device will be switched of when you leave the key. The device also switched off automatically, when no key is pressed for 4 minutes.
Set the clock: press 3 times the date and the time by pressing (YY.MM.DD). When you entered the Year push the
key to get to the month and to the day. Press the key again to get to the
time. If you entered date and time confirm it by pressing
Hold the measure value: Press the activated in the Datalog Time menu)
Display background lighting: Press the after 20 seconds. Backlight will be activated by pressing any key. Furthermore the Power off time is reset to 4 minutes.
Activation of superuser function: 2 times
- Options – Unlock Enter the 4-digit code using the number). Confirm by pressing
Changing the user level:: From advanced user to simple user:
2 times Options – O userlevel. From simple user to advanced user:
Keep both the buttons and pressed immediately after switching on the device as long as the logo is shown on the display. The device automatically starts the main menu. Navigate to options – o userlevel and confirm by pressing
key for 3 seconds
key - Options – Date / Time Here you set the
.
key in the save mode (Function must be
key; Backlight will turn off automatically
button (standard are the last 4 digits of the serial
.
.
Electromatic Equipment Co., Inc.
-10-
Electromatic Equipment Co., Inc.
-3-
4.0 VIEW OF THE INSTRUMENT
Electrodes
Measuring head
Power button
Temperature value
Water content
Changing
calibration curves
Batteries (rearside)
8.0 DEVICE MAINTENANCE INSTRUCTION
Do not drop the instrument or expose it to excessive temperatures. Clean it only with fuzz-free, dry cloth. The device is not waterproof.
For more information concerning the treatment of the device, see www.checkline.com
The frequency for technical check of the instrument depends on the use and required level of accuracy.
For a fee, we can also carry out a calibration at the factory. In this case you will receive a calibration certificate.
9.0 CHANGING THE BATTERIES
Please follow following steps to change the batteries:
1. Use a finger to press the arrow on the battery cover and pull it back.
2. Remove the empty batteries and replace them with four new 1,5 Volt AA Alkaline Batteries. Take care of the correct polarization when you put in the batteries.
3. Press down the batteries so you can close the cover.
If the battery symbol appears in the measuring window resp. if a critical charge of battery is shown in the status, the batteries have to be changed IMMEDIATELY. If you do not
use your humimeter device for a longer period, remove the batteries. For eventual resulting damages we cannot provide any warranty.
Electromatic Equipment Co., Inc.
-4-
Electromatic Equipment Co., Inc.
-9-
Print data (logs):
Switch off
Switch off
Keypad symbols
OK
To print the stored logs, connect the supplied printer cable with the printer. Open the protective cover at the TMT-425 carefully. First insert the plug, which is closer to the plastic housing, into the gauge. Then turn the gauge on. Only then insert the other plug of the cable into the
printer and turn him on by the
button. The green light should blink. If this is not the case, change the batteries and try again. Press the button until the main menu is opened. Select the item “Print logs” and confirm with
. Now you can choose whether all stored values or only the last values (logs) should be printed.
5.0 MENU LEVEL OVERVIEW
Overview main menu
Data Memory
Manual logs Delete logs
Print logs
Last row All logs Delete logs
Send logs
Manual logs Delete logs
Type selection menu
Next calibration curve Previous calibration curve
Store menu
Previous log New log
Main menu
Switch lower Switch upper Open this menu / Enter
Options
Date/Time Datealog Time Language Unlock
° C / °F
Userlevel Lighting time Auto off time Setting Materialcalib. Password Reset
Status
Measuring window:
Rolling Menu Power ON / OFF Switch upper Switch lower Save
Hold
Watch saved
data
Add supplier’s
data
Menu:
Enter Switch upper Switch lower Back
Enter numbers
Enter letters Next or right Left Yes No Shift
Electromatic Equipment Co., Inc.
-8-
Electromatic Equipment Co., Inc.
-5-
6.0 MEASURING PROCEDURE
Important: For a correct measurement please ensure that the device has the same
temperature (+/-3°) as the material under test. For that reason, let your humimeter adjust to the surrounding temperature of the material for at least half an hour before measuring (protect from direct sunlight!).
1. Switch on the device by pressing the power button (
2. If desired, the calibration curves can be changed by using the arrow keys.
3. Insert the electrodes of the TMT-425 in the material. The measuring values will be displayed directly.
4. Save the displayed value in the store menu (1x
value succeeded if the number in front of the
5. To get into the store menu, press the left button ( ) until the symbol appears.
6. Name the saved value by selecting
) for 3 sec..
) by pressing ( button). Saving the
symbol increased.
.
7.0 PC CONNECTION AND DATA PRINT The TMT-425 provides the possibility to transfer your
saved data to a computer. Use the supplied connection cable to connect your gauge. Open the protective cover of the TMT-425 carefully and insert the USB mini port B. The bigger port has to be inserted in the USB output of your PC. Start the LogMemorizer Software and turn on the TMT-
425. The data transfer can be started either by the software or by measuring gauge.
Data transfer by the humimeter TMT-425: Press the button until the main menu is opened (see right picture). Select the menu item “Send logs” and confirm with again with be transferred to your pc then.
Data transfer by your computer: Open the LogMemorizer Software. Click on the button “remote access” and a drop down menu will open. For the data transfer, select the option “Import last manual log” (last stored log will be transferred) or “Import all manual logs” (all stored values will be transferred). Click on one option and the data will be imported on the pc. It is recommended to read the manual on the LogMemorizer CD about further settings.
. Then select “Manual logs” and confirm
. All stored values on your humimeter will
Electromatic Equipment Co., Inc.
-6-
Electromatic Equipment Co., Inc.
-7-
Loading...