This material does not contain any ingredients having
known health hazards in concentrations greater than 1%.
This material does not contain any known or suspected
carcinogens.
SECTION 3
PHYSICAL DATA (nominal)
Boiling Point: >220°F pH: 7.35 – 7.9
Freezing Point: <20°F Acoustic Imp.: 1.726x106
Vapor Pressure: N/A Vapor Density: N/A
Evaporation Rate: N/A Specic Gravity: >1.02
Solubility in Water: complete
Appearance and Odor: water white, opaque gel;
bland odor
SECTION 4
FIRE AND EXPLOSION HAZARD DATA
Flash Point: None
Upper Exposure Limit: None
Lower Exposure Limit: None
Special Fire Fighting Procedures: N/A
Extinguishing media: N/A
Unusual Fire and Explosion Hazards: None
SECTION 5 – REACTIVITY DATA
Stability: Stable
Conditions to Avoid: None
Incompatibility (Materials to Avoid): None known
Hazardous Polymerization: Will not occur
Hazardous Decomposition
or Byproducts: None known
SECTION 6
HEALTH HAZARD AND FIRST AID DATA
Routes of Entry: 1
Skin: not likely
Ingestion: not normally
Eyes: not normally
Inhalation: no
Effects of Overexposure:
Acute: May cause temporary eye
irritation
Chronic: none expected
First Aid Procedures:
Skin: Remove with water if
desired.
Eyes: Flush with water for 15
minutes.
Ingestion: For large quantities, induce
vomiting, call a physician.
Inhalation: N/A
SECTION 7
STORAGE AND HANDLING INFORMATION
Precautions to be taken in handling and storage:
Store between 20°F and 120°F.
Spills are slippery and should be cleaned up
immediately.
Steps to be taken in case material is released or spilled:
Pick up excess for disposal. Clean with water.
Waste disposal method: Dispose of in accordance with
federal, state, and local regulations.
SECTION 8
CONTROL MEASURES
Respiratory Protection: Not required
Ventilation: Not required
Protective Gloves: On individuals
demonstrating
sensitivity to
SOUNDSAFE®
Eye Protection: As required by
working conditions
Other Protective Equipment: Not required
SOUNDSAFE® contains only food grade and cosmetic
grade ingredients.
16.0 Material Safety Data .............................................................................. 104
17.0 Warranty ........................................................................... Inside Back Cover
– 102 –
– 3 –
14.0 APPENDIX A – VELOCITY TABLE
1.0 Introduction
MaterialSound Velocity
in/µs
Aluminum0.25106375
Beryllium0.508012903
Brass0.17304394
Bronze0.13903531
Cadmium0.10902769
Columbium0.19404928
Cooper0.18304648
Glass (plate)0.22705766
Glycerine0.07601930
Gold0.12803251
Onconel0.22905817
Iron0.23205893
Cast iron0.1800approx.4572
Lead0.08502159
Magnesium0.23005842
Mercyry0.05701448
Molybdenum0.24606248
Monel0.21105359
Nickel0.22205639
Nylon0.1060approx2692
Platinum0.15603962
Plexiglas0.10602692
Polystyrene0.09202337
PVC0.09402388
Quartz glass0.22605740
Rubber vulcanized0.09102311
Silver0.14203607
Steel (1020)0.23205893
Steel (4340)0.23305918
Steel Stainless*0.22305664
Sound Velocity
m/s
The Check•Line model TI-CMXDLP is an ultrasonic thickness gauge that measures
with extreme versatility. It has the ability to measure simultaneously measure coatings
and material thicknesses while maintaining the ability to locate pits, aws and defects
in the material. Based on the same operating principles as SONAR, the TI-CMXDLP is
capable of measuring the thickness of various materials with accuracy as high as ± 0.001
inches, or ± 0.01 millimeters. The principle advantage of ultrasonic measurement over
traditional methods is that ultrasonic measurements can be performed with access to
only one side of the material being measured.
1.1 Disclaimer
Inherent in ultrasonic thickness measurement is the possibility that the instrument will
use the second rather than the rst echo from the back surface of the material being
measured. This may result in a thickness reading that is TWICE what it should be.
Responsibility for proper use of the instrument and recognition of this phenomenon
rest solely with the user of the instrument. Other errors may occur from measuring
coated materials where the coating is insufciently bonded to the material surface.
Irregular and inaccurate readings may result. Again, the user is responsible for
proper use and interpretation of the measurements acquired.
– 101 –
– 4 –
13.0 USING THE UTILITY SOFTWARE
2.0 QUICK STARTUP GUIDE
13.1 Computer System Requirements
Datacomm will run on many different operating systems: Windows 98 (1st or
2nd edition), Windows NT 4.0 with Service Pack 5, Windows ME, Windows XP,
Windows2000 Professional, Windows 2000 Server, or Windows 2000 Advanced Server
operating systems running on Intel or AMD hardware. A Pentium 166MHz or faster
processor with at least 32 megabytes of physical RAM is required. You should have
40 megabytes of free disk space before attempting to install Datacomm. Datacomm
requires an available communications port in order to transfer data to and from the
TI-CMXDLP. Datacomm supports COM1, COM2, COM3, and COM4. USB to Serial
Converter. Some newer laptop computers do not have standard serial ports. In this case
it is possible to use a USB to Serial converter. If a serial to USB cable is needed, contact
Electromatic.
13.2 Installing Datacomm
Datacomm comes on a CD-ROM with an automatic installer program. Place the CD
in your computer’s CD tray and close the door. Open the CD-ROM by double clicking
on the My Computer ICON, then double click on the CD. Finally, double click on the
SETUP icon to begin the installation. Refer to the help section in Datacomm software
for the complete operating manual,setup, and operation.
13.3 Using the XFER menu (TI-CMXDLP)
The XFER menu of the TI-CMXDLP will be used in
conjunction with the Datacomm PC software. The steps
below outline the procedure for accessing the XFER menu
and basic operation as follows:
1. Press the MENU key once to activate the menu items
tab. Press the MENU key multiple times to tab right,
and the ESC key multiple times to tab left,until
the XFER menu is highlighted and displaying the
submenu items.
2. Use the UP and DOWN arrow keys to scroll
through the sub menu items until the desire option is
highlighted.
3. Press the ENTER key to activate the option selected.
4. Once the Backup or Restore function has been
completed, press the MEAS key once to return to the
menu items, or twice to return to the measurement
mode.
Turn the TI-CMXDLP on and off using the switch located on the bottom right corner of
the keypad. When TI-CMXDLP is initially turned on, a ash logo and blinking lights
will be displayed, followed by the identity of the transducer (probe) currently plugged
into the gauge. The TI-CMXDLP is equipped with an “Auto Probe Recognition”
feature that attempts to identify special transducers. If the TI-CMXDLP doesn’t nd a
transducer equipped with this feature, the user will be see a list of transducers and will
need to select a specic transducer type. The following sections outline each scenario.
NOTE: This section is primarily written as a basic startup guide only.
2.1 TI-CMXDLP Overview
The TI-CMXDLP has great features
and tools that will be a benet for the
variety of applications you face on
a continual basis. Let’s have a brief
look at the screens you’ll be looking
at most often:
A. Repeatability/Stability Indicator: This is used in
conjunction with the digital
thickness values displayed.
When all the vertical bars are
fully illuminated and the last digit
on the digital thickness value is
stable, the TI-CMXDLP is
reliably measuring the same
value 3 to 200 times per second,
depending on which measurement mode and features
are enabled.
B. Battery Icon: Indicates amount
o f battery life remaining.
C. Velocity: The material velocity
value the TI-CMXDLP is currently using or calibrated for. It can be set to display in English or Metric units.
D. Feature Status Bar: Indicates the features currently enabled and in use in the following order:
1. Measurement Mode 3. High-Speed Scan Mode 5. Gain Setting
2. Differential Mode 4. Alarm Mode
– 100 –
– 5 –
E. Digital Material Thickness Value: Extra large font size for viewing ease.
F. Scan Bar: Another view of material thickness in a deection style horizontal bar.
This visual tool enables the user the ability to see thickness changes during high
speed scans from aws and pits.
G. Units: The current measurement units being used (English, Metric)
H. Digital Material Thickness Value: Smaller font size when the B-Scan display
view is enabled.
I. Coating Thickness Value: Displays the actual thickness of any coating adhered
to a metallic material surface (PECT Mode), or a coating adhered to anon-metallic
surface (CT Mode).
J. Minimum Material Thickness: Part of the Alarm feature. Displays the minimum
thickness value found during a scan.
K. Maximum Material Thickness: Part of the Alarm feature. Displays the maximum
thickness value found during a scan.
L. B-Scan Display: Cross section view of the material. Provides a graphical view
of the opposite/blind surface (i.e. inside pipe wall surface), to give some idea of the
condition, or integrity of the material being tested.
M. RF A-Scan Display: The actual sound wave reection returned from the detection of the opposite surface of the material being measured. In this view, the entire
sine wave is displayed, showing both the positive and negative half cycles from the
detection/reection. This view is commonly used to “see the big picture” and make
ne adjustments to the scope settings, prior to selecting another view option.
N. Hot Menu items: Items in this menu section are the most commonly adjusted
features. They can be displayed and scrolled by pressing the MEAS key at anytime.
The ESC key is used in conjunction with MEAS key to reverse the direction
scrolled.
O. Reference Note: When the “hot menu” items are displayed/activated, the base
material thickness value shown at reference point “O”. However, when the “hot
menu” items have been deactivated, the current base material velocity value is
displayed, as shown in reference point “P”. When the “hot menus” are deactivated,
the entire “hot menu” section becomes a multiple measurement display area,
allowing the user to simultaneously view: base material, coating,and base material
minimum/maximum thickness values dynamically, as shown in reference point “R”
P. Reference Note: Please refer to “O”, for a detailed explanation.
Q. RECT A-Scan Display: The actual sound wave reection that is returned from the
detection of the opposite surface of the material being measured. In this view, only
half of the sine wave is being displayed (positive or negative). This view is comonly
used as a “aw detection” mode, once the user made all the necessary adjustments
using the RF mode (refer to M).
12.5 Using the Default Setup
The default setup feature was added to the TI-CMXDLP to use, as a last resort, if
there are no setups stored in the gauge –factory or otherwise. The only time this might
possibly occur is if the setup le in the TI-CMXDLP was somehow corrupted, and
the user does not have access to a computer to re-load the factory setups back into the
TI-CMXDLP. This gives the user the ability to load and modify a basic setup as follows:
NOTE: The default le contains no probe zero data.
Therefore, a zero must be performed after loading. Also,
the default setup can be loaded when using the aw mode
feature, to eliminate the probe zero delay value, and start
at the initial pulse or zero.
1. Press the MENU key once to activate the menu items
tab. Press the MENU key multiple times to tab right
and the ESC key multiple times to tab left until the
SETUP menu is highlighted and displaying the
submenu items.
2. Use the UP and DOWN arrow keys to scroll through
the sub menu items until DEFAULT SETUP is
highlighted.
12.6 Selecting a Language
The TI-CMXDLP is equipped with a language option.
Currently, the only languages supported are English,
Spanish, and German.
1. Press the MENU key once to activate the menu items
tab. Press the MENU key multiple times to tab right
and the ESC key multiple times to tab left until the
SETUP menu is highlighted and displaying the
submenu items.
2. Use the UP and DOWN arrow keys to scroll through
the sub menu items until LANGUAGE is highlighted.
3. Press the LEFT and RIGHT arrow keys to toggle the
language options.
4. Once the desired language is displayed, press the
MEAS key to return to the measurement screen.
– 6 –
– 99 –
7. Press the OK key to return to the Save Setup
Parameters List Box.
8. If both parameters will be edited, repeat steps 4 – 7.
9. Use the UP and DOWN arrow keys to scroll to and
highlight SAVE SETUP.
10. Press the ENTER key to activate the Setup List
Box.11) Use the UP and DOWN arrow keys to scroll
through the setups until the target location to save the
Setup is highlighted.
12. Press the OK key to activate the conrmation screen.
13. Press the OK key to save the Setup, or ESC to cancel
saving the Setup.
14. Finally, press the MEAS key to return to the
measurement screen.
NOTE: The Name and Comments of the Setup can
be edited at any time by simply repeating the Save Setup routine described above.
Therefore, the Save Setup function can also be considered an Edit Function.
12.4 Deleting a Saved Setup
This option allows a user to delete setup les that were
previously saved and no longer needed. It’s a simple
feature to allow the user to do a bit of “house cleaning”.
1. Press the MENU key once to activate the menu items
tab. Press the MENU key multiple times to tab right
and the ESC key multiple times to tab left until the
SETUP menu is highlighted and displaying the
submenu items.
2. Use the UP and DOWN arrow keys to scroll through
the sub menu items until DELETE is highlighted.
3. Press the ENTER key to display the Setups List.
4. Press the UP and DOWN arrow keys to scroll to the
Setup Name.
5. When the Setup Name is highlighted, press the
ENTER key to display the conrmation screen.
6. Press the OK key to delete the Setup File.
7. Finally, press the MEAS key to return to the
measurement screen.
2.2 Auto Probe Recognition
When the TI-CMXDLP is initially powered up, the gauge will automatically check to
see if the transducer plugged into the gauge can be recognized. The steps that follow
assume the TI-CMXDLP recognized the probe type:
1. Press the OK key once to use the identied probe,
or ESC to display a list of optional transducers.
NOTE: If the TI-CMXDLP recognizes a specic transducer,
the user should always select OK to use the identied probe.
The only time an alternative probe should be selected from a
list is if the user switched probes following initial power up
and recognition
2. Assuming the TI-CMXDLP recognized the probe and
the OK key was pressed, the TI-CMXDLP will advance
to a Zero Probe menu. If the transducer was identied
as a special transducer capable of measuring coating
thickness, a menu will be displayed giving the user the
ability to toggle the coating thickness display on/off as
follows:
3. Press the UP and DOWN arrow keys to toggle the
coating option on/off.
4. Wipe all couplant from the transducer face and advance
to the Probe Zero & Calibration section.
2.3 Selecting the Transducer Type
If the TI-CMXDLP does not identify a specic transducer
type on initial power up, the user will be required to select
a type from a predened list of types by diameter and
frequency. By selecting a transducer type from a predened
list, the TI-CMXDLP can recall specic properties about the
transducer.
NOTE: Once the transducer has been selected, the
TI-CMXDLP will store and recall this transducer type every
time the TI-CMXDLP is powered on/off. The type will only
change if the user physically selects another transducer type
from the list, or selects a previously saved setup. However,
the TI-CMXDLP will continue to take you through these
steps each time the gauge is powered up. You’ll notice that
the probe type previously selected will be highlighted every
time the probe type screen is displayed. Use the following
steps to select your transducer type:
– 98 –
– 7 –
1. Press the OK or ESC keys to display the factory list of transducer types
(by diameter and frequency).
2. Press the UP and DOWN arrow keys to scroll through the transducer list until the
appropriate type is highlighted.
3. Press the ENTER key to select the transducer type and display overwrite existing
probe screen.
4. Press the OK key to overwrite the existing probe type with the newly selected probe
type. The zero probe screen will be displayed. Proceed to the zero probe section that
follows.
2.4 Probe Zero & Calibration
The next steps are to perform a probe zero and calibrate the TI-CMXDLP to the material
and transducer being used. If the sound velocity is unknown, the TI-CMXDLP can be
calibrated to a known thickness sample. This demo will briey explain both of the set
techniques. The TI-CMXDLP is equipped with two zero options:
1. Off Block Zero (Automatic Probe Zero): When this feature is enabled the
TI-CMXDLP will do an electronic zero automatically, eliminating the need for a
zero disk or block.
2. On Block Zero (Manual Probe Zero): When this feature is enabled the transducer
must be placed on the Probe Zero Disk (battery cover located on the top of the unit.
NOTE: Transducers of the same type will have very slight mechanical and electrical
variations. If it’s discovered that the linearity is off following an initial auto probe
zero and extreme accuracy is required, a manual zero should be performed followed
by an auto zero. This will adjust and eliminate any error. This is only required if
it’s discovered the transducer is non-linear following an initial auto probe zero. The
procedures are outlined as follows:
4. Use the UP and DOWN arrow keys to scroll through
the setups until the target setup is highlighted.
5. Press the ENTER key to activate the conrmation
screen.
6. Press the OK key to load the setup from memory.
7. Press the MEAS key to return to the measure screen.1
12.3 Saving a Setup
Once the TI-CMXDLP parameters and features have be adjusted for an application,
the user may elect to save these setting to a specic setup location for future use. This
can potentially save time and reduce error between users. It is sometimes necessary to
rename a previously saved setup, or add additional comments about a particular setup.
The setup name may have been entered incorrectly, or the user needs to use the setup for
a completely different project. An inspector’s name or other comments about the project
may also be required for additional documentation purposes. The following procedures
outline the necessary steps for saving and editing a setup:
1. Press the MENU key once to activate the menu items tab.
Press the MENU key multiple times to tab right and the
ESC key multiple times to tab left until the SETUP menu
is highlighted and displaying the submenu items.
2. Use the UP and DOWN arrow keys to scroll through the
sub menu items until SAVE is highlighted.
3. Press the ENTER key to display the
Save Setup Parameters List Box.
4. Press the UP and DOWN arrow keys to
scroll the Name and Note parameters.
5. When the parameter to edit is
highlighted, press the ENTER key to
activate the Alpha Edit Box.
6. Use the UP, DOWN, LEFT, and
RIGHT arrow keys to scroll through
the characters, the ENTER key to
select characters, and the CLR key to
backspace through the characters, until
the Name or Note elds have been
edited.
– 8 –
– 97 –
12.0 SETUPS – CREATE, STORE, EDIT, & RECALL
12.1 Introduction to Setups
Often times, users are faced with a variety of tasks and applications that are sometimes
similar, but often times very different. With a standard thickness gauge,the user would
have to recalibrate for each individual application respectively. With all the features of
the TI-CMXDLP, the number of potential applications also increases based on ability
alone. This is primarily in reference to the addition of those very difcult applications,
where a standard thickness gauge would not be feasible, or capable of accomplishing
the tasks. The increased number of features and parameters also adds to the overall
setup time, or amount of time to set up marginal applications with perfection and
understanding. Because of the additional time involved, the TI-CMXDLP has been
equipped with the ability to save these setups to memory and be recalled at any time.
The TI-CMXDLP can store up to 64 custom setups. These setups can be bidirectionally
transferred to and from a PC. Therefore, the user can save as many setups as necessary
for all their individual applications requirements. This saves a great deal of time and
knowledge for future inspections of the same job or project. This feature also eliminates
error between two or more users during the setup and calibration process.
12.2 Opening a Setup
The TI-CMXDLP is loaded with a number of setups from the factory. These setups
can be opened, edited, and saved to any one of 64 setup locations. If a factory setup
is written over, the user can simply reload the default factory setups at anytime using
the utility software included with the TI-CMXDLP. The
factory setups are general setups Only. However, they
may serve as an excellent starting point, requiring only
minor adjustments to work for custom applications. It
is recommended that user customized setups be saved
to an empty location, and save the factory setups for
the purpose of a general starting point. The following
procedures outline how to open factory and custom
setups:
1. Press the MENU key once to activate the menu
items tab. Press the MENU key multiple times to tab
right and the ESC key multiple times to tab left until
the SETUP menu is highlighted and displaying the
submenu items.
2. Use the UP and DOWN arrow keys to scroll through
the sub menu items until OPEN is highlighted.
3. Press the ENTER key to display the Setup List Box.
Performing An Auto Zero (Off Block)
1. Be sure all couplant has been removed
from the face of the transducer.
2. Press the OK key to perform the
automatic probe zero, or ESC key
to cancel the zero operation.
3. The screens illustrated will be
briey displayed followed by the
main measurement screen. The
TI-CMXDLP is ready to be calibrated.
Performing a Manual Probe Zero (On Block)
NOTE: When the zero probe option is set to manual, the Probe
Zero Disk (battery cap) located on the top of the gauge, will be
used as a zero standard and the warning screen illustrated above
will be displayed.
1. Press the OK or ESC keys to enter the main measurement
screen and begin the manual zero process.
2. Apply a drop of couplant on the transducer and place the
transducer in steady contact with the Probe Zero Disk, and
obtain a steady reading.
3. Press the MENU key once to activate the menu items tab. Press the MENU key
multiple times to tab right and the ESC key multiple times to tab left until the
PRB menu is highlighted and displaying the submenu items.
4. Press the UP and DOWN arrow keys to scroll through the sub menu items until
ZERO PROBE is highlighted.
–96 –
– 9 –
5. Press the ENTER key to display the
conrmation screen.
6. If a coating transducer was identied
use the UP and DOWN arrow keys to
toggle coating on/off.
7. Press the OK key to complete the
probe zero function, or ESC key to
cancel the Probe Zero function.
8. Remove the transducer from the Probe
Zero Disk, and proceed to the calibration section.
NOTE: The value that is displayed will change depending on the current velocity setting
in the TI-CMXDLP. Disregard the number that is displayed. It is not important. What is
important is accurately performing the steps outlined above to insure reliability of the
probe zero calculation.
One Point Material Calibration
For the purposes of this quick start section, we’ll only be covering the most common
one point calibration option to determine the sound velocity of the test material. It
would be very handy to carry a set of mechanical calipers to use in conjunction with the
TI-CMXDLP for calibration in the eld.
Using a Known Thickness
NOTE: Be sure that the probe zero procedure has been performed prior to performing
this calibration procedure.
1. Physically measure an exact sample of the material or a
location directly on the material to be measured using a set
of calipers or a digital micrometer.
2. Apply a drop of couplant on the transducer and place the
transducer in steady contact with the sample or actual
test material. Be sure that the reading is stable and the
repeatability indicator, in the top left corner of the display, is
fully lit and stable. Press the MENU key once to activate the
menu items tab. Press the MENU key multiple times to tab
right and the ESC key multiple times to tab left until the CAL menu is highlighted
and displaying the submenu items.
3. Use the UP and DOWN arrow keys to scroll through the sub menu items until
MATL 1PT is highlighted.
2. Use the UP and DOWN arrow keys to scroll through
the sub menu items until CLOSE is highlighted.
3. Press the ENTER key to close the active le.
NOTE: Following the keypress, the CLOSE text will
be grayed out indicating the le has been close and is no
longer active.
– 10 –
– 95 –
11.7 Changing the Active File - Open
The user may have transferred grid/
seq templates from a PC to the
TI-CMXDLP, or setup grids/seq using
the TI-CMXDLP at an earlier time.
The name of the currently active le is
always displayed at the top of the Grid/
Seq Box in measurement mode (refer to
photo below). It’s not only important to
recognize what le is currently active,
but also be able to change the active le
at any time. The following procedures
outline this process:
1. Press the MENU key once to activate the menu
items tab. Press the MENU key multiple times
to tab right and the ESC key multiple times to
tab left until the DATA menu is highlighted and
displaying the submenu items.
2. Use the UP and DOWN arrow keys to scroll through
the sub menu items until OPEN is highlighted.
3. Press the ENTER key to display the Grid/Seq List
Box.
4. Use the UP and DOWN arrow keys to scroll through
the grids until the target grid is highlighted.
5. Press the ENTER key to activate the conrmation
screen.
6. Press the OK key to load the le from memory.
7. Press the MEAS key to return to the measure screen.
11.8 Closing an active File - Close
A user might not have a current requirement to store
measurements, but a le is currently open or active and
needs to be closed. The following procedures outline how
to close an open or active le :
1. Press the MENU key once to activate the menu items
tab. Press the MENU key multiple times to tab right
and the ESC key multiple times to tab left until
the DATA menu is highlighted and displaying the
submenu items.
2.5 Zero Coating
In order to account for very slight electronic differences in transducers of the same type,
frequency, and diameter, the TI-CMXDLP has been equipped with a “zero coating”
feature. This enables the TI-CMXDLP to obtain very accurate readings on coatings,
eliminating potential errors incurred from slight differences in the manufacturing
processes. The procedure is outlined below:
Performing A Coating Zero
1. Press the MULTI MODE key once to activate the
measurement mode options.
2. Use the UP and DOWN arrow keys to scroll through
the sub menu items until Coating Only (CT) mode is
highlighted.
3. Press the ENTER key to select the measurement mode
and return to the measurement screen.
4. Apply a drop of couplant on the transducer and place the
transducer in steady contact with the Probe Zero Disk
(battery cover) and obtain a steady reading.
NOTE: The coating measurement displayed will potentially
be a value greater or less than 0.
5. Press the MENU key once to activate the menu items tab.
Press the MENU key multiple times to tab right and the
ESC key multiple times to tab left until the PRB menu is
highlighted and displaying the submenu items.
6. Use the UP and DOWN arrow keys to scroll through the
sub menu items until ZERO COATING is highlighted.
7. Press the ENTER key to display the conrmation screen.
8. Press the OK key to zero the coating and return to the
PRB menu, or ESC to cancel the coating zero process.
9. Press the MULTI MODE key once to activate the
measurement mode options.
10. Use the UP and DOWN arrow keys to scroll through the
sub menu items until Coating On (PECT) is highlighted
11. Press the ENTER key to select the measurement mode
and return to the measurement screen, and begin taking
readings.
–94 –
– 11 –
2.6 Coating Calibration
The TI-CMXDLP has been preset to a default coating velocity of 0.0850 in/sec
(2159m/sec). This will be very close to the most common coating material velocities
used in the eld. If the velocity of the coating is known, and different than the above
default setting, the user can simply enter the coating velocity into the TI-CMXDLP.
However, if the velocity is unknown, the TI-CMXDLP can also be calibrated to a
specic coating sample/type using the 1pt calibration option in PECT (pulse-echo
coating)mode, or a two point calibration is CT (coating only) mode. For the purpose
of this quick start section only the 1pt option PECT (pulse-echo coating) mode will
be covered. Refer to the calibration section of the manual for a complete explanation
on the coating calibration options. The following steps below outline the necessary
steps to either set the velocity of the coating, or perform a one point calibration to
calculate the coating velocity:
Known Coating Velocity
If the coating velocity is known, the user may wish to
simply enter the velocity number into the TI-CMXDLP,
rather than have the TI-CMXDLP calculate the velocity
value using a known thickness on a coating sample. The
steps for entering the velocity are outlined below:
1. Press the MENU key once to activate the menu items
tab. Press the MENU key multiple times to tab right and
the ESC key multiple times to tab left until the CAL
menu is highlighted and displaying the submenu items.
2. Use the UP and DOWN arrow keys to scroll through
the sub menu items until COATING VEL is highlighted.
3. Press the ENTER key to display the Digits Edit Box.
4. Press the UP and DOWN arrow keys to scroll the
highlighted value.
5. Press the LEFT and RIGHT arrow keys to scroll the
digit locations.
6. Repeat steps 4 & 5 until the velocity number is correctly
displayed.
7. Press the OK key to set the coating velocity and return
to the menu screen,or ESC to cancel entering the
coating velocity.
8. Finally, press the MEAS key to return to the
measurement screen and begin taking readings.
4. Press the OK key to delete All Files from memory, or the ESC key to abort.
5. Press the MEAS key to return to the measurement screen.
11.6 Editing a Grid (File)
Once a grid has been created and saved to memory, the user can edit the Comments or
Increment Direction at a later time. The following procedures outline this process
1. Press the MENU key once to activate
the menu items tab. Press the MENU
key multiple times to tab right and the
ESC key multiple times to tab left until
the DATA menu is highlighted and
displaying the submenu items.
2. Use the UP and DOWN arrow keys to
scroll through the sub menu items until
EDIT is highlighted.
3. Press the ENTER key to display the Edit List Box.
4. Use the UP and DOWN arrow keys to scroll through
the Edit options until NOTE or INCR. DIR is
highlighted.
NOTE: If editing the INCR. DIR, simply use the LEFT
or RIGHT arrow keys to scroll NONE, NORTH, EAST,
SOUTH, WEST for a Grid, or INC, DEC for a SeqLog.
Proceed to step 10.
5. Press the ENTER key to activate the Alpha Edit box –
Only used when editing the NOTE.
6. Use the UP, DOWN, LEFT, & RIGHT arrow keys to
highlight the appropriate alpha characters.
7. Press the ENTER key to select a character and
advance to the next eld of the Comments.
8. Use the CLR key to backspace if necessary.
9. Repeat steps 6 - 8 until the Comments are completed
10. Press the UP or DOWN arrow key to highlight
SAVE CHANGES, and the OK key to activate the
conrmation screen.
11. Press the OK key to save the changes or the ESC key
to cancel editing the le parameters.
12. Press the MEAS key to return to the measurement screen.
– 12 –
– 93 –
4. The user may opt to clear a specic reading
and save a new one at anytime. Press the CLR
key in the appropriate cell location to clear the
reading,take a new measurement, and press the
ENTER key to save the new reading.
5. Abort the Grid/Seq Log View Box by pressing
the MEAS key at any time.
11.5 Deleting Grids (Files)
Deleting One File
1. Press the MENU key once to activate the menu
items tab. Press the MENU key multiple times
to tab right and the ESC key multiple times to
tab left until the DATA menu is highlighted and
displaying the submenu items.
2. Use the UP and DOWN arrow keys to scroll
through the sub menu items until DELETE ONE FILE is highlighted.
3. Press the ENTER key to display the File List
Box.
4. Use the UP and DOWN arrow keys to scroll
through the stored Files until the target File to
delete is highlighted.
5. Press the OK key to delete the File.
6. Press the MEAS key to return to the
measurement screen.
Deleting All Data
1. Press the MENU key once to activate the menu
items tab. Press the MENU key multiple times
to tab right and the ESC key multiple times to
tab left until the DATA menu is highlighted and
displaying the submenu items.
2. Use the UP and DOWN arrow keys to scroll
through the sub menu items until DELETE ALL DATA is highlighted.
3. Press the ENTER key to activate the
conrmation screen.
Known Coating Thickness
When the exact velocity of a coating is unknown, the user has the option of performing
a one point calibration on a sample of the coating with a known thickness to determine
the sound velocity. It would be very handy to carry a set of mechanical calipers to use in
conjunction with the TI-CMXDLP for calibration in the eld:
1. Physically measure a location on a coating sample using a set of calipers or a digital
micrometer.
IMPORTANT NOTE: In PECT (pulse-echo coating) mode, the coating sample must be
coupled to metal in order to calibrate successfully. Simply place a drop of couplant on
a piece of metal, lay the coating sample over the couplant on the metal and proceed to
step 2.
2. Apply a drop of couplant on the transducer and place the
transducer in steady contact with the coating (on metal)
sample or actual test material. Be sure that the reading is
stable and the repeatability indicator, in the top left corner
of the display, is fully lit and stable. Press the MENU
key once to activate the menu items tab. Press the MENU
key multiple times to tab right and the ESC key multiple
times to tab left until the CAL menu is highlighted and
displaying the submenu items.
3. Use the UP and DOWN arrow keys to scroll through the
sub menu items until COATING 1PT is highlighted.
4. Press the ENTER key to display the Digits Edit Box.
5. Press the UP and DOWN arrow keys to scroll the
highlighted value.
6. Press the LEFT and RIGHT arrow keys to scroll the digit
locations.
7. Repeat steps 5 & 6 until the known thickness value is
correctly displayed.
8. Press the OK key to calculate the coating velocity and
return to the menu screen, or ESC to cancel the one point
calibration.
9. Finally, press the MEAS key to return to the measurement
screen and begin taking readings.
NOTE: CHECK YOUR CALIBRATION! Place the transducer back on the calibration
point. The coating thickness reading should now match the known coating thickness
sample. If the thickness is not correct, repeat the steps above.
– 92 –
– 13 –
2.7 Measure
The TI-CMXDLP is now ready to measure. There are four different measurement view
options, each with a specic purpose – Digits, RF, RECT, & B-Scan. The below outline
how to toggle between the different view mode options:
Selecting the Measurement View Option
1. Press the MENU key once to activate the menu items
tab. Press the MENU key multiple times to tab right and
the ESC key multiple times to tab left until the DISP
menu is highlighted and displaying the submenu items.
2. Use the UP and DOWN arrow keys to scroll through the
sub menu items until VIEW is highlighted.
3. Use the LEFT and RIGHT arrow keys to scroll the view
options.
4. Once the view is displayed, press the MEAS key to
return to measurement mode.
DIGITS: Displays the digital thickness value using a large font size. This view is useful
when the TI-CMXDLP is being used as a basic thickness gauge.
RF: Displays the actual waveform signal, much like an oscilloscope, from the reection
of the opposite surface, pit, aw, crack or void. This view shows both the positive and
negative peaks, and is often used to ne tune the scope settings, prior to inspection
RECT: Displays a half waveform signal, either positive or negative, from the reection
of the opposite surface, pit, aw, crack or void. The user can select the polarity or
“phase” displayed. This is typically determined by rst using RF view to select the most
optimal polarity “phase”, to ne tune the scopes settings. The RECT view is commonly
used as the primary “aw detection” view.
BSCAN: The Time Based B-Scan provides the user with a cross sectional view of the
material being tested. This mode is useful when there is concern regarding the prole of
the blind surface. This can also be a useful view when scanning for pits and aws.
Once the view has been selected according to the application requirements, the Delay
and Range of the screen will potentially need to be adjusted, if the view has been set to
RF or RECT. Alternatively, if BSCAN was selected, the B-Start and B-Depth settings
will need to be adjusted. These settings serve the same purpose,with only differences
in terminology. The Delay the same as B-Start, and the Range is the same as B-Depth.
Therefore, these items will be grouped together for the duration of this manual, as
follows: Delay (B-Start) and Range (B-Depth). Use the following steps to adjust these
settings directly from the measurement screen as follows:
NOTE: The Delay (B-Start) and Range (B-Depth) are also used to adjust the parameters
of Scan Bar.
1. Press the UP, DOWN, LEFT,
and RIGHT arrow keys to scroll
the target cell cursor to the desired
storage location.
2. Press the ENTER key to save the
current reading in the highlighted
cell location. It’s as simple as that!
3. Press the MEAS key to hide the
GRID/SEQ View Box, or ENTER
to display it as necessary.
Note: Once the le is open, it will remain open until it’s closed or another le is opened
by the user. If the gauge is powered off, the TI-CMXDLP will automatically open
the le when powered on. Simply press the ENTER key to display the le from the
measurement screen.
11.4 Viewing stored readings & A/B Scans
It is sometimes necessary to
go back and view the stored
readings and B-Scans using the
TI-CMXDLP without a PC. The
following procedures outline this
process:
1. Press the MEAS key once to
activate measure menu items. Press
the MEAS key multiple times
to move right and the ESC key
multiple times to move left until the
LOG cell is highlighted.
2. Press the ENTER key to display
the Grid Log Box.
3. Press the UP, DOWN, LEFT, and
RIGHT arrow keys to scroll the stored
readings and corresponding display
view. Notice as the cursor is moved to a
different cell, the display will be updated
with the display view saved with the
reading. Readings stored in memory
are indicated by displaying a MEM in
the top left corner of the measurement
screen.
– 14 –
– 91 –
3. Continue on to the next section “Saving a Grid or Seq File”.
Saving a Grid
Once all the parameters are set, the user
has the option of saving or canceling
the new grid.
4. Use the UP and DOWN arrow
keys to scroll through the new Grid
or Seq List Items until CREATE GRID or LOG? is highlighted.
5. Press the ENTER key to accept
the grid or seq log parameters and
activate the conrmation screen
6. Press the OK key to save the New
Grid or Seq Log, or the ESC key
to cancel the New Grid or Seq Log
setup and return to the DATA menu.
7. Press the MEAS key to return to
the measurement screen and begin
storing readings.
11.3 Storing a reading
Now that a grid or sequential log has been created, it’s time to make some measurements
and store the readings. The following procedures outline this process: Storing a Reading
NOTE: Once the gird or sequential log has been created it will automatically be
displayed following the create conrmation screen. It is located at the bottom of the
measurement screen, and can be left open for the duration of the inspection process.
This section assumes the grid or sequential log was just created. Refer to the section on
“opening a grid or seq log” if the le was created at another time.
Key Feature: When the TI-CMXDLP is displaying a “grid log”, pressing the OK key
will initiate an advance to row number option. Use the Left, Right, UP, and DOWN
arrow keys to enter the row number. Once the row number is correctly displayed, press
the OK key to advance directly to that row number in the grid log.
Key Feature: When the TI-CMXDLP is displaying a “grid log”, pressing the ESC key
multiple times, will toggle through the display options: Digits, RF, RECT,and B-SCAN
views.
Adjusting Delay (B-START & Range (B-DEPTH)
1. Press the MEAS key once to activate the measure menu
items. Press the MEAS key multiple times to move right
and the ESC key multiple times to move left, until the
either the DELAY (START) or RANGE (DEPTH) cell is
highlighted.
2. Use the UP, DOWN, LEFT, or RIGHT arrow keys
to scroll the DELAY(START) and RANGE (DEPTH)
values.
3. Repeat steps 1 & 2 until the range is correctly being
displayed. Alternatively, the DELAY (START) and RANGE (DEPTH) values can be changed using the Digit
Edit Box as follows:
4. Press the MEAS key once to activate measure menu
items. Press the MEAS key multiple times to move right
and the ESC key multiple times to move left, until the
either the DELAY (START) or RANGE (DEPTH) cell is
highlighted.
1. Press the ENTER key to display the digits edit box.
2. Press the UP and DOWN arrow keys to scroll the
highlighted value.
3. Press the LEFT and RIGHT arrow keys to scroll the digit
locations.
4. Repeat steps 3 & 4 until the DELAY (START) or RANGE (DEPTH) value is correctly displayed.
5. Press the OK key to set the DELAY (START) and WIDTH (DEPTH) value and return to the measure
screen, or ESC to cancel entering the DELAY(START)
or WIDTH (DEPTH) value.
6. Finally, press the MEAS key to return to the
measurement screen and begin taking readings.
NOTE: The DELAY (START) & WIDTH (DEPTH) can
also be adjusted from the tabbed menu item DISP. However,
using the hot menu keys is the easiest method.
– 90 –
– 15 –
In the upper left corner of each of the display photos above, is the repeatability indicator.
The repeatability indicator is represented by six vertical bars and represents how
repeatable the measurements are. In regular measurement mode,the TI-CMXDLP makes
8 measurements a second. In scan mode, the TI-CMXDLP makes 200 measurements a
second. If the coating mode option is activated, the TI-CMXDLP makes 3 measurements
a second in regular measurement mode and 65 measurements a second in scan mode.
When the TI-CMXDLP is idle, only the left vertical bar will be displayed. However,
when the TI-CMXDLP is making a measurement,four or ve of the bars should
be displayed on the repeatability indicator. If fewer than four bars are showing, the
TI-CMXDLP is having difculty achieving a stable measurement and the thickness
value displayed is potentially unstable.
4. Press the OK key to select the coordinate or end ID and return to the Grid or Seq
List Items screen, or ESC to cancel the selection and return to the Grid or Seq List
Items menu.
NOTE: If there is not enough memory available to create the grid or sequential log, an
error message box “NOT ENOUGH MEMORY“ will be displayed. Press the OK or
ESC key to return to the Grid or Seq List Items menu. It may be necessary to free some
memory in the TI-CMXDLP at this time. Refer to section 11.5 for more information on
Deleting a File.
Selecting the Auto Increment Direction
The Auto Increment feature gives the user
the ability to specify which direction to
advance the cursor after storing a reading.
5. Use the UP and DOWN arrow keys
to scroll through the new Grid or
Seq List Items until INCR. DIR or
DIRECTION is highlighted.
6. Use the LEFT & RIGHT arrow keys to
toggle the Increment direction NONE,
NORTH, EAST, SOUTH, or WEST for
a grid, or INC, DEC for a sequential
log.
7. When the correct Increment direction is displayed, continue on to the next section
“Graphics”.
Saving Graphics
The TI-CMXDLP provides the user with the ability to save a snapshot of the display
screen and all the current settings of the TI-CMXDLP with every reading, or just save
the reading only. Saving the graphics might be advantageous to the user when the A/BScan views will be used to graphically save a picture of the scanned areas for reporting
purposes. It might also come in handy when user requires backup of all the current
TI-CMXDLP parameters for each reading. However, if neither of these requirements are
necessary, this option should be disabled
in order make more efcient use to the
TI-CMXDLP storage capacity.
1. Use the UP and DOWN arrow keys
to scroll through the new Grid or Seq
List Items until SAVE GRAPHICS
is highlighted.
2. Use the LEFT & RIGHT arrow keys
to toggle the Increment direction to
YES or NO.
– 16 –
– 89 –
Setting the Top Left(Grid) or Start ID(Seq)
1. Use the UP and DOWN arrow keys to
scroll through the new Grid or Seq List
Items until TOP LEFT or START ID is
highlighted.
2. Press the ENTER key to activate the
Coordinate or Start ID Edit Box.
3. GRID LOG: Use the LEFT, & RIGHT
arrow keys to scroll the Columns, and
the UP, DOWN arrow keys to scroll the
Rows.
SEQ LOG: Use the UP, DOWN, LEFT,
& RIGHT arrow keys to highlight the
appropriate alpha characters. Press the
ENTER key to select a character and
advance to the next character eld, in
conjunction with using the CLR key to
backspace if necessary.
4. Press the OK key to select the coordinate
or start ID and return to the Grid or Seq
List Items screen, or ESC to cancel the
selection and return to the Grid or Seq
List Items menu.
Setting the Bottom Right (Grid) or END ID (Seq)
1. Use the UP and DOWN arrow keys to
scroll through the new Grid or Seq List
Items until LOWER RIGHT or END ID
is highlighted.
2. Press the ENTER key to activate the
Coordinate or End ID Edit Box.Grid Log
Sequential Log
3. GRID LOG: Use the LEFT, & RIGHT
arrow keys to scroll the Columns,and
the UP, DOWN arrow keys to scroll the
Rows. SEQ LOG: Use the UP, DOWN,
LEFT, RIGHT arrow keys to highlight
the appropriate alpha characters. Press
the ENTER key to select a character and
advance to the next character eld, in
conjunction with using the CLR key to backspace if necessary.
3.0 Keyboard, Menu & Connector Reference
3.1 Menu Key (Operation * Sub Menus)
The Menu key activates the primary menu structure
containing 8 menu tab groups. These tab groups then
contain sub menu items, or functions. The sub menu
items have been organized in tab groups according
to how closely they are related to the individual tab
group names. Let’s rst get familiar with how to move
around in these tabs before continuing on to the sub
menu functions. This procedure is outlined below:
Activating and Getting Around in the Menus Items
1. Press the MENU key once to activate the menu items
tab. Press the MENU key multiple times to tab right,
and the ESC key multiple times to tab left until the
desired tab group is highlighted and displaying the
submenu items. The tab groups are illustrated above
(A).
2. Use the UP and DOWN arrow keys to scroll
through the sub menu items until the desired
function is highlighted. The sub menu items are
illustrated in the diagram above (B).
3. Depending on which function is highlighted, use
the LEFT, RIGHT, and Enter keys to scroll the
options or activate the Digit Edit and List Box options.
3.2 Probe – Menu
ZERO PROBE: The TI-CMXDLP is zeroed in much the
same way that a mechanical micrometer is zeroed. If the
TI-CMXDLP is not zeroed correctly, all of the measurements
made using the TI-CMXDLP may be in error by some xed
value. The TI-CMXDLP is equipped with an optional automatic or manual zero feature.
Refer to section 6.2, for an explanation of this important procedure.
ZERO COATING: In order to account for very slight electronic differences in
transducers of the same type, frequency, and diameter, the TI-CMXDLP has been
equipped with a “zero coating” feature. This enables the TI-CMXDLP to obtain
very accurate readings on coatings, eliminating potential errors incurred from slight
differences in the manufacturing processes. Refer to section 6.2.
– 88 –
– 17 –
Loading...
+ 37 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.