Checkline TEM-1 User Manual

– 1 –
TABLE OF CONTENTS
PAGE
1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
02
1.1 Complete Kit
1.2 Overview
04
3.0 Battery Test And Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . .
05
3.1 Battery Test
3.2 Battery Replacement
4.0 Taking A Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
06
6.0 Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
07
7.0 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
09
7.1 Temperature Error Compensation
8.0 Using the Cup Electrode (optional) . . . . . . . . . . . . . . . . . . . . . . . . 12
9.0 Conversion Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
10.0 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
– 24 –
10.0 LIMITED WARRANTY
ELECTROMATIC Equipment Co., Inc. (ELECTROMATIC) warrants to the original purchaser that this product is of merchantable quality and confirms in kind and quality with the descriptions and specifications thereof. Product failure or malfunction arising out of any defect in workmanship or material in the product existing at the time of delivery thereof which manifests itself within one year from the sale of such product, shall be remedied by repair or replacement of such product, at ELECTROMATIC’s option, except where unauthorized repair, disassembly, tampering, abuse or misapplica­tion has taken place, as determined by ELECTROMATIC. All returns for warranty or non-warranty repairs and/or replacement must be authorized by ELECTROMATIC, in advance, with all repacking and shipping expenses to the address below to be borne by the purchaser.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY AND FITNESS FOR ANY PARTICULAR PURPOSE OR APPLICATION. ELECTROMATIC SHALL NOT BE RESPONSIBLE NOR LIABLE FOR ANY CONSEQUENTIAL DAMAGE, OF ANY KIND OR NATURE, RESULTING FROM THE USE OF SUPPLIED EQUIPMENT, WHETHER SUCH DAMAGE OCCURS OR IS DISCOVERED BEFORE, UPON OR AFTER REPLACEMENT OR REPAIR, AND WHETHER OR NOT SUCH DAMAGE IS CAUSED BY MANUFACTURER’S OR SUPPLIER’S NEGLIGENCE WITHIN ONE YEAR FROM INVOICE DATE.
Some State jurisdictions or States do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you. The duration of any implied warranty, including, without limitation, fitness for any particular purpose and merchantability with respect to this product, is limited to the duration of the foregoing warranty. Some states do not allow limitations on how long an implied warranty lasts but, not withstanding, this warranty, in the absence of such limitations, shall extend for one year from the date of invoice.
ELECTROMATIC Equipment Co., Inc.
600 Oakland Ave. Cedarhurst, NY 11516—USA
Tel: 1-800-645-4330/ Tel: 516-295-4300/ Fax: 516-295-4399
Every precaution has been taken in the preparation of this manual. Electromatic Equipment Co., Inc., assumes no responsibility for errors or omissions. Neither is any liability assumed for damage result­ing from the use of information contained herein. Any brand or product names mentioned herein are used for identification purposes only, and are trademarks or registered trademarks of their respective holders
– 2 –
1.0 INTRODUCTION
The Check-Line TEM-1 textile moisture meter determines the residual moisture in yarn bobbins, fabric folls, filaments, fiber flocks, etc.
The measuring principle of the TEM-1 is based on the electrical conductivity of the material which always bears a fixed relation to the moisture. All changes in resistance inside the measuring ranges of interest are sufficiently pronounced to ensure a high degree of accuracy for the readings obtained with the TEM-1.
NOTE: The distribution of moisture is never uniform, but varies both
over the entire surface and also inside the material. Therefore, different probes types, referred to here as electrodes, are used to measure the moisture content and detect differences in moisture content.
NOTE: The TEM-1 is a high-quality instrument. However, over time,
even the best product is subject to a certain amount of wear. We recommend that you return the meter and accessories every twelve months for a routine check-up. The meter and its calibration will be thoroughly examined, so that you can be sure that the meter is operating correctly.
1.1 Complete Kit
The TEM-1 is supplied as a complete with leather carrying case, electrode holder (p/n # 204) electrodes (p/n #'s 205, 207) and connecting cable (p/n # 200
– 23 –
– 3 –
1.2 Overview
The TEM-1 Display provides direct readings for wool, rayon and cotton.
For other materials, the 0–100 scale is used in conjunction with the tables found in Section 9.0.
Connection Cable
Electrode Holder
Sample under test
Carrying Case
Retaining Ring
#207 Electrode
Red Battery Check Button
White Measure Button
– 22 –
– 4 –
2.0 TEM-1 SETUP
1. Open the carrying case and carefully remove the TEM-1 moisture meter.
2. Locate the cable receptacle on the top of the TEM-1 and insert either end of the connection cable. You will hear a click when the plug is seated. See A
3. Insert the opposite end of the cable into the receptacle in the base of the Electrode Holder. Press until you hear a click indicating the the plug is seated. See B.
4. Unscrew and remove the dark grey retaining ring. See C.
5. Select the proper electrode for your application. See Section 5.0 for a list and descriptions of the electrodes available for the TEM-1.
6. Once the proper electrode has been selected, insert the two connecting pins on the bottom of the electrode into the two receptacles on the top of the Electrode Holder. See D.
IMPORTANT NOTE: Inserting electrodes
The pins on the electrodes are very sharp. When inserting an electrode, grasp it by the white plastic base, keeping your hand away from the ends of the pins and press it into position.
IMPORTANT NOTE: Removing electrodes
The pins on the electrodes are very sharp. To remove the Electrode, first Unscrew and remove the retaining ring. Grasp the white plastic base of the electrode and carefully push up while wiggling the electrode from side to side until it is free. DO NOT place your hand above the pins and attempt to pull the Electrode out as an injury may result.
7. Replace the grey retaining ring and hand tighten.
8. Perform a Battery Check and described in Section 3.0
A.
B.
C.
D.
– 21 –
– 5 –
3.0 BATTERY TEST AND REPLACEMENT
To ensure an accurate reading, it is necessary to test the strength of the battery before each measuring session.
3.1 Battery Test
1. Press and hold the red test button on the front of the meter. If the measuring needle swings up and stops in the red area of the dial (see arrow), there is sufficient battery power for accurate measurements.
2. If the meter needle stops below the red area, the battery needs to be replaced before measurements can be taken.
3.2 Battery Replacement
1. Remove the gauge from the carrying case. Turn over locate the battery compartment.
2. Insert the tip of a small flat-head screwdriver (or similar instrument into the slot shown at left and gently twist open the lid.
NOTE: The battery compartment lid is attached to the case.
3. Remove and disconnect the old 9V battery and replace with a new one (alkaline or rechargeable).
4. Carefully close the lid by pressing on the two corners until the lid snaps back into position.
5. Repeat the Battery Test. If successful the TEM-1 is ready for use.
– 20 –
– 6 –
4.0 TAKING A MEASUREMENT
1. After completing the setup and battery check, insert the ends of the selected electrode into the test material.
(Do we need more explanation of how to do this?)
2. Press the white Test Button. The measuring needle will swing up and stop. Reading are made on the upper or lower scale depending upon the material being tested as follows:
Wool, rayon, cotton: The three lower scales indicate the moisture
content of the sample directly in percent.
Other materials: Note the reading on the upper scale then refer
to the conversion charts in Section 9.0 to cover the meter reading into an accurate measurement of percent moisture.
The three lower scales provide direct readings for wool, rayon and cotton
Upper scale. Conversion tables for 36 fiber types are included in Section 9.0.
– 19 –
– 7 –
5.0 ELECTRODES
A variety of measuring electrodes are available for the TEM-1 to accommodate different applications.
Ordering information?
Needle Electrodes
Needle Electrodes are used for making measurements of cops, bobbins or hanks or yarn
# 205 = 25mm length # 206 = 45mm length # 207 = 100mm length
Knife Electrode #210
The Knife Electrode is for measuring on webs, rolls and bales of textile fabrics and bulk fibers by applying the blades to or inserting them between the plies of fabric.
Roller Electrode #211
The Roller Electrode is used for measuring moving webs of material.
Flow Electrode #212
The Flow Electrode is used for measuring threads and yarns.
Surface Electrode #213
The Surface Electrode is used for measuring threads and yarns.
– 18 –
– 8 –
Compression Screw Electrode #201
The Compression Screw Electrode with automatic friction clutch applies constant pressure to assure accurate measurements of individual samples.
Cup Electrode #202
The Cup Electrode is used for measuring compressible material in the form of flocks or tufts. The constant pressure exerted upon the material thanks to the friction clutch assures accurate reproducible readings. the electrodes connects to the moisture meter directly, without the use of a universal connecting cable. See Section 8.0.
– 17 –
– 9 –
7.0 SPECIFICATIONS
7.1 Temperature Error Compensation
Indicating errors due to differences in ambient temperature as nor­mally encountered are negligible and withing the permissible toler­ance. For extremely accurate measurements, the temperature error may be corrected as follows:
• TEM-10 moisture meters are calibrated at 68 °F (20 °C). When making measurements at 68 °F, the temperature error is zero.
• When measuring below 68 °F, add 0.1% moisture to the dial reading for every 1.8 °F (1 °C) degree below 68 °F.
• When measuring above 68 °F, deduct 0.1% moisture from the dial reading for every 1.8 °F (1 °C) degree above 68 °F.
Measuring ranges Wool: 8 to 24.5% moisture content
Rayon: 3 to 23% moisture content Cotton: 2 to 12% moisture content
Comparison Measurement 0 - 100 scale
Power 9V battery (alkaline or rechargeable)
Weight 0.88 lbs. (400 g)
Dimensions 7.87"x 9.84" x 2.76" (250 x 200 x 70mm)
(L x W x H)
– 16 –
– 10 – – 15 –
– 11 –– 14 –
10.0 COMPARISON TABLES
The procedure for using the conversion tables in this section is as follows:
1. Follow the direction in section 4.0 Taking a measurement.
2. Note the reading on the upper scale of the TEM-1
3. Find that number on the upper scale of the appropriate conversion table. The lower scale provides a conversion to percent moisture for that reading.
– 12 –
9.0 USING THE CUP ELECTRODE (P/N 202)
Handwheel
Upper part of cup
electrode
Product to be measured
Lower part of cup
electrode
Lower part of cup
electrode
Plug
Retaining pin
Press disc
Locking ball
Built-in friction clutch
NOTE: Before beginning, perform a battery check, see section 3.0
1. Unlock the upper part of th electrode and pull up and away.
2. Fill the lower part or cup with the product to be measured up to the level mark.
3. Turn the handwheel of the upper part counterclockwise to move the press disc up as far as it will go.
– 13 –
4. Introduce the locking ball of the upper part in the retaining pins of the lower part and lock in the first notch.
5. Turn the handwheel. If the upper part clockwise to compress the material to be measured, The pressure required for a reliable measurement is reached as soon as the slipping clutch is felt to slip. The measurement can now be made. If the slipping clutch is not felt to slip, turn the pressure screw all the way up again and hook the locking ball into the second notch. If the clutch still fails to slip, the cup has not been filled with enough material. Add more material and proceed once more as described above until the clutch is felt to slip.
6. Now, plug the cup electrode into the measuring jack in the upper side of the TEM-1.
7. Depress the white measuring button and read the moisture content directly using the 3 lower scales; or note the upper scale reading and refer to the conversion tables in Section.
8. The material measured must be fully removed from the cup electrode after each measurement. in case of a pulpy or pasty substance, the cup electrode must be removed from the meter for a more thorough cleaning.
TEXTILE MOISTURE METER
TEM-1
Operating Instructions
CHECK•LINE
®
BY ELECTROMATIC
ELECTROMATIC
E Q U I P M E N T C O., I N C.
600 Oakland Ave., Cedarhurst, NY 11516–U.S.A. TEL: 516-295-4300 FAX: 516-295-4399
CHECK•LINE
®
INSTRUMENTS
Loading...