Checkline MGT User Manual

Controls
All Series MG torque gauges have three keys for controlling all functions of the instrument.
POWER Turns the instrument on and off.
Also used in the calibration mode.
PEAK Used to select Clockwise Pea k,
Counterclockwise Peak or Nor­mal (real time) display mode. The actual peak readings are al­ways captured and can be dis­played at any time.
ZERO Zeros any tare value (up to the
full capacity of the gauge) and clears the peak readings stored in memory.
Display
The display consists of a 4 1/2-digit section and several indicators. Their functions are listed below.
LO BAT Low battery voltage indicator CW Clockwise torque indicator CCW Counterclockwise torque
indicator CW PEAK Peak clockwise indicator CCW PEAK Peak counterclockwise indicator OZIN, LBIN, Units of measurement (Model NCM, KGMM dependent)
- - - - (dashes) Overload (>110% of range)
Operation
The default mode of operation is the normal (real time) mode. If the peak readings are to be ob­served as they occur, then the mode of operation can be changed by pressing PEAK until the de­sired mode (CW PEAK or CCW PEAK) appears on the display. Please note that this action affects only the display. The peak readings are captured automatically and can be cleared from the memory by either pressing ZERO or shutting off the gauge.
The gauge is equipped with an automatic shut­down feature and it will shut off after approximately 30 minutes of inactivity (readings do not change by more than ±10 counts and no keys are pressed). The entire display will flash for 5-7 seconds as a warning of the imminent shutdown. To change the default setting of 30 minutes, hold PEAK while turning on the gauge. Press PEAK repeatedly un­til 'AOFF' appears. Press POWER to select this function. The current setting will flash on the dis­play. Use PEAK to scroll through the displayed choices and POWER to select. Press POWER again at the 'donE' prompt in order to save the setting.
The displayed units of measurement and the de­fault mode of operation (peak or normal) can be changed by entering the setup mode as described above, pressing POWER at the 'init' prompt and selecting the desired settings using the PEAK and the POWER keys.
Power
The instrument may be powered by the internal 9V battery, or by the included AC adapter. The need for battery replacement is indicated by a 3­step sequence: 1 - a steady LO BAT appears on the display indicating the last 10% of the battery life, 2 - LO BAT begins to flash indicating the need for an immediate battery replacement, 3 - the en­tire display except LO BAT flashes for several sec­onds and then the instrument shuts off.
Calibration
To properly calibrate this instrument, application of a precise torque value equal to the full capacity
of the gauge in lbin or ozin (regardless of the displayed units) will be required.
While holding PEAK, turn the gauge on. When ‘CAL’ appears on the display press POWER three times to select the calibration mode. At the ‘null’ prompt press ZERO. At the ‘SPAn’ prompt apply the cali­bration torque and press POWER. The display will show “uuuu” or “nnnn” if the calibration torque is insufficient or excessive accordingly. If this happens, the only way to terminate the calibration mode is by momentarily disconnecting the battery or connect­ing the AC adapter without plugging it into a wall outlet. This will stop the calibration procedure with­out making any changes to the previous calibration data. Successful calibration is indicated by ‘donE’ on the display. Press POWER to save the changes and resume normal operation.
2.53 [64.3]
3.84 [97.5]
10.0 [25.4] RETRACTED LENGTH
CHUCK CAPACITY:
0.38 [9.5] MAX DIA MGT50Z AND SMALLER
0.50 [12.7] MAX DIA MGT12 AND LARGER
3.50 [88.9]
4.82 [122.4] MGT50Z AND SMALLER
5.19 [131.8] MGT12 AND LARGER
1.69 [42.4]
IN[MM]
X2
1.75 [44.4] DIA
Model Capacity x graduation
MGT10Z 10.00 x 0.01 ozin, 7.000 x 0.005 kgmm, 7.000 x 0.005 Ncm
MGT20Z 20.00 x 0.02 ozin, 14.00 x 0.01kgmm, 14.00 x 0.01 Ncm
MGT50Z 50.00 x 0.05 ozin, 36.00 x 0.05 kgmm, 35.00 x 0.05 Ncm
MGT12 12.00 x 0.01 lbin, 140.0 x 0.1 kgmm, 135.0 x 0.1 Ncm
MGT50 50.00 x 0.05 lbin, 580.0 kgmm, 570.0 x 0.5 Ncm
MGT100 100.0 x 0.1 lbin, 1150 x 1 kgmm, 1150 x 1 Ncm
Specifications
Series MG
ACCURACY ±0.5% of full scale ±1digit
SAMPLING 30 readings/second RATE
DISPLAY 2.5/s in normal mode, 30/s in RATE peak mode
SAFE 150% of gauge capacity. Display OVERLOAD shows ---- (dashes) above 110%
POWER 9V battery or AC adapter
BATTERY 30 hours of continuous operation LIFE
SIZE See drawing
WEIGHT 1.8 lb [0.8 kg]
Mark-10 Corporation has been an innovator in the Force and Torque measurement field since 1979. We strive to achieve 100% customer satisfaction through excellence in product design, manufac­turing and customer support. In addition to our stan­dard line of products, we can provide modifica­tions for special applications. Please contact us for further information or with suggestions for im­provement.
DIGITAL TORQUE GAUGE
USER'S GUIDE
Warranty
Mark-10 Corporation expressly warrants to its buyer for one year from the date of delivery that the goods sold are free from defects in workmanship and materials. Mark-10 Corporation will, at its option, repair or replace or refund the purchased price of goods found to be defective. This remedy shall be the buyer's sole and exclusive remedy. Any modification, abuse, exposure to corrosive environment or use other than intended will void this warranty. This warranty is in lieu of all other warranties, including implied warranties of merchantability and fitness for an in­tended purpose. In no event shall Mark-10 Corporation be liable for any incidental and consequential damages in connection with goods sold or any part thereof.
We Make A Measurable Difference In Force Measurement