6
Sometimes Gage readings can be influenced by changes in substrate shape, composition, surface roughness or by measuring in
a different location on the part. This is why Calibration
Adjustments are made possible.
1- or 2-point Calibration Adjustments may be performed if
readings are not falling within the expected range of thickness for
the application being measured.
If a calibration adjustment method has not been specified, use a
1-point method first. If measuring the included shims reveals
inaccuracies, use the 2-point method. Factory Calibration
settings can be restored at any time by performing a Reset
(pg.16), creating a NEW calibration setting (pg.9), or by
DELETING the adjustments made to the Cal 1 calibration setting
(pg.9). The symbol appears on the display whenever factory
calibration settings are in use.
With “FN” probes, calibration adjustments for “F” and “N” prin-
ciples are stored independent of one another.
Once adjusted, you may “lock” the current calibration
adjustment to prevent further modification. (See “Cal Lock” on pg.8)
1-point Calibration Adjustment
Also known as an offset or correction value, there are 4 ways to
perform this adjustment:
(1) Simple
Zero Calibration Adjustment
Measure your uncoated part. If the Gage does not read "0" within the
tolerance of the probe being used, lift the probe from the surface and
adjust the display down (-) or up (+) until it reads "0". Measure and
adjust until the average of a series of readings on the uncoated surface
is within tolerance of "0" .
(2) Average
Zero Calibration Adjustment
To establish “0” on a rough or curved surface a preferred method
is to take several readings on the uncoated part and average the
result.
NOTE:
NOTE: