Thank you for purchasing this C.A 10141 conductivity meter.
For best results with your instrument:
read these operating instructions carefully, comply with the precautions for use.
Information or useful tip.
Battery.
The product is declared recyclable following an analysis of the life cycle in accordance with standard ISO 14040.
Chauvin Arnoux has adopted an Eco-Design approach in order to design this appliance. Analysis of the complete life-
cycle has enabled us to control and optimize the eects of the product on the environment. In particular this appliance
exceeds regulation requirements with respect to recycling and reuse.
The CE marking indicates conformity with European directives, in particular LVD and EMC.
The rubbish bin with a line through it indicates that, in the European Union, the product must undergo selective disposal
in compliance with Directive WEEE 2012/19/EU. This equipment must not be treated as household waste.
PRECAUTIONS FOR USE
This instrument is compliant with safety standard IEC 61010-2-030, for voltages 5V with report to ground. Failure to observe the
safety instructions may result in electric shock, re, explosion, and destruction of the instrument and of the installations.
The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be taken in
use. Sound knowledge and a keen awareness of electrical hazards are essential when using this instrument.
Observe the conditions of use, namely the temperature, the relative humidity, the altitude, the degree of pollution, and the place
of use.
Do not use the instrument if it seems to be damaged, incomplete, or poorly close.
Before each use, check the condition of the housing. Any item of which the insulation is deteriorated (even partially) must be set
aside for repair or scrapping.
All troubleshooting and metrological checks must be done by competent, accredited personnel.
2
CONTENTS
1. FIRST USE ............................................................................................................................................................................... 4
1.3. Replacement parts ......................................................................................................................................................... 5
1.4. Inserting the batteries ..................................................................................................................................................... 6
2. PRESENTATION OF THE INSTRUMENT ............................................................................................................................. 7
2.2. Functions of the instrument ............................................................................................................................................ 8
2.5. Setting the time ............................................................................................................................................................. 10
3. USE IN STAND-ALONE MODE ............................................................................................................................................11
3.1. Connecting the cell ........................................................................................................................................................11
4. USE IN RECORDING MODE ...............................................................................................................................................21
4.2. Get Data Logger Transfer software .............................................................................................................................. 21
4.3. USB link ........................................................................................................................................................................ 21
4.4. Data Logger Transfer software ..................................................................................................................................... 23
5.4. USB .............................................................................................................................................................................. 28
5.5. Power supply ................................................................................................................................................................ 28
5.8. Compliance with international standards ...................................................................................................................... 29
6.2. Replacement of batteries .............................................................................................................................................. 31
6.3. Serial number ............................................................................................................................................................... 31
Rendez-vous sur notre site Internet pour télécharger la notice
de fonctionnement de votre appareil et de votre cellule :
Touche Fonction
INTERNATIONALChauvin Arnoux Group
Tél : +33 1 44 85 44 38Fax : +33 1 46 27 95 69
Our international contacts
www.chauvin-arnoux.com/contacts
www.chauvin-arnoux.com
Allez dans l’onglet ProduitsHWHႇHFWXH]XQHUHFKHUFKHDYHF
le nom de votre appareil. Une fois l’appareil trouvé, allez sur
sa page. La notice de fonctionnement se trouve sur la droite.
cor lin
Téléchargez-la.
Procédez de même pour la cellule de conductivité.
ATC
Ref 20°C
MISE EN PLACE DES PILES
cor lin
ATC
Ref 25°C
cor lin
ATC
LOGICIEL DATA LOGGER TRANSFER
Rendez-vous sur notre site Internet pour télécharger la dernière version du logiciel d’application :
www.chauvin-arnoux.com
Allez dans l’onglet Support, puis Télécharger nos logi-
Téléchargez le logiciel puis installez-le sur votre PC.
N
N
Touche Marche / Arrêt (appui long)
MEM : un appui court pour enregistrer la
MEM
mesure et la température.
REC
REC : un appui long pour démarrer ou
arrêter un enregistrement.
: un appui court pour choisir le type de
mesure : la conductivité, la résistivité, le
TDS ou la salinité.
SET : un appui long pour modifier le
paramètrage des mesures : l’unité de tem-
SET
pérature, la correction en température, la
température de référence, le facteur TDS.
Un deuxième appui long pour sortir du
paramètrage des mesures
Rétroéclairage.
CAL : un appui court pour consulter les
paramètres d’étalonnage et lancer un
CAL
étalonnage.
END
END : un appui long pour terminer l’éta-
lonnage.
Un appui court pour régler la température
Ÿ
en MTC ou les paramètres en mode SET.
Un appui long pour choisir la solution tam-
ź
pon pendant l’étalonnage.
ÉTAT DE LIVRAISON
Le conductimètre C.A 10141 est livré dans une mallette de
transport avec :
Une cellule de conductivité avec sonde de température
intégrée XCP4ST1.
Quatre piles alcaline AA ou LR6.
Une gaine de protection montée sur l’appareil.
Une solution étalon de conductivité 1408 µS/cm.
Un bécher.
Un cordon USB - micro USB.
Une dragonne.
Des guides de démarrage rapide (un par langue).
8QHDWWHVWDWLRQGHYpUL¿FDWLRQ
➉
One C.A 10141 conductivity meter.
One XCP4ST1 conductivity cell with integrated temperature probe.
Four AA or LR6 alkaline batteries.
ATTESTATION DE VERIFICATION
CHECKING ATTESTATION
190, rue Championnet
Numéro de l'appareil :
75876 PARIS Cedex 18
Equipment number
FRANCE
Type /
:
Model
Désignation de l'instrument :
Instrument designation
Signature :
Vérifié par :
Signature
Tested by
Établi en usine, ce document atteste que le produit ci-dessus a été vérifié et est conforme aux
conditions d'acceptation définies dans nos procédures de fabrication et de contrôle.
Tous les moyens de mesure et d'essai utilisés pour vérifier cet appareil sont raccordés aux
étalons nationaux et internationaux soit par l'intermédiaire d'un de nos laboratoires de métrologie
accrédités COFRAC soit par un autre laboratoire accrédité.
Après sa mise en service, cet instrument doit être vérifié à intervalle régulier
auprès d'un service de métrologie agréé.
Pour tout renseignement veuillez contacter notre service après vente et d'étalonnage.
At the time of manufacture, this document certifies that the above product have been verified and
complies with acceptance conditions defined in our manufacturing and testing procedures.
Every test or measuring equipment used to verify this instrument are related to national
and international standards through one of our laborator
ies of metrology certified by french COFRAC
equivalent to NAMAS in the UK or through another certified laboratory.
After being in use, this instrument must be recalibrated within regular intervals
by an approved metrology laboratory. Please contact our after sales and calibration department:
Service après vente et d'étalonnage TEL: +33 (2) 31 64 51 55 FAX: +33 (2) 31 64 51 72
After sales and calibration department e-mail: info@manumesure.fr
WEB : www.manumesure.com
www.chauvin-arnoux.com
ATTESTATION DE CONFORMITE
COMPLIANCE ATTESTATION
Nous certifions que ce produit a été fabriqué conformément aux spécifications
techniques de constuction applicables.
We certify that this product is manufactured in accordance with applicable
constructing specifications.
11
907 009 119 - 02/03
➂
➄
➇
➆
One protective sheath tted on the instrument.
4
One standard solution having a conductivity of 1408µS/cm.
5
One beaker.
6
One USB to micro USB cable.
7
One carrying case.
8
One wrist strap.
9
Quick start guides (one per language).
10
One certicate of verication.
11
4
1.2. ACCESSORIES
One adapter, DIN male to S7 female (for the cell) and female
jack (for a PT1000 temperature probe). The cable is 1m long.
One USB power adapter with USB to micro USB cable
One adapter DIN male to BNC female (for the cell) and
female jack (for a PT1000 temperature probe). The cable
is 10cm long.
One standard solution having a conductivity of 12,85mS/cm One standard solution having a conductivity of 147µS/cm.
1.3. REPLACEMENT PARTS
One XCP4ST1 conductivity/temperature cell One standard solution having a conductivity of 1408µS/cm One set of 3 plastic beakers One protecting sheath
For the accessories and spares, consult our web site:
www.chauvin-arnoux.com
5
1.4. INSERTING THE BATTERIES
Turn the instrument over. Press the locking tab and lift o the battery compartment cover.
Remove the rubber plug. Insert the 4 batteries provided, with the polarities as shown.
Put the rubber plug back in place. Push it in correctly.
Place the two front ends before pushing in the central part.
Put the battery compartment cover back in place; make sure
that it is completely and correctly closed.
6
2.1. C.A 10141
2. PRESENTATION OF THE INSTRUMENT
Cable.
Tight 8-point connector.
Clamping ring.
MEM
REC
SET
CAL
END
C.A 10141
Conductivity Meter
Conductivity cell.
Backlit LCD display unit.
Keypad with 6 keys.
Protecting sheath.
On/O key.
Wrist strap.
Type B micro USB connector protected by a cap.
7
2.2. FUNCTIONS OF THE INSTRUMENT
The C.A 10141 is a conductivity meter in a watertight housing. It is used to measure conductivity, resistivity, TDS (total dissolved
solids), salinity, and temperature.
This instrument is easy to use. It has extensive stand-alone capabilities and can be used:
to calibrate the cell using a standard solution, to display temperature measurements in °C or in °F, to display the time, to record the measurements, to communicate with a PC via a USB cable.
The Data Logger Transfer software can be installed on a PC and is used to congure the instruments, to program a recording
session and to recover the recorded measurements.
2.3. KEYPAD
The functions indicated above the line on the keys are accessed by a short press.
The functions indicated below the line are accessed by a long press.
To prevent inadvertently switching the instrument on, the On/O key requires a long press.
KeyFunction
A long press on the key switches the instrument on.
A second long press on the key switches the instrument o when it is on.
The instrument cannot be switched o while it is recording.
MEM
REC
SET
CAL
END
▲
▼
A short press on the MEM key records the measurement and the temperature. A long press on the REC key starts or stops a manual recording session.
A short press on the key is used to choose the quantity to be displayed: conductivity, resistivity, TDS, or
salinity.
A long press on the SET key lets you view and modify the calculation parameters:
the temperature unit (°C or °F), the temperature correction, the reference temperature for the conductivity display (20 or 25°C), the TDS (Total Dissolved Solids) factor.
A second long press is used to return to measurement mode.
A short press on the key toggles the backlighting on or o. Once on, it goes o at the end of 30 seconds.
A short press on the CAL key is used to look up the current calibration and start the calibration procedure for
the cell connected.
A long press on the END key is used to terminate the calibration procedure in progress.
A short press on the ▲ or ▼ key is used to set the temperature in manual mode or choose the parameter to
be modied in SET mode.
During the calibration, a long press on the ▲ or ▼ key is used to select the standard solution.
8
2.4. DISPLAY
MEM FULL
REC
MEM
SET COR. REF TEMP TDS
no cor
cor lin
cor f(T)
Display of the measurement.
Display of the unit.
CAL SET
MTC
Display of the temperature.
ATC
Ref 20°C Ref 25°C
AM
Display of the time.
PM
Indicates the battery voltage level. When the symbol is empty, the batteries must be replaced.
Indicates that the instrument is connected to a PC via the USB (cable) or that the power adapter is connected.
Indicates that auto-o is disabled and the instrument is in permanent mode.
This occurs when:
P
MEM FULL
REC
MEMIndicates that a measurement is recorded.
SET COR.Indicates adjustment of the temperature correction.
SET REFIndicates the reference temperature selected, 20 or 25°C (68 or 77°F)
SET TEMPIndicates the temperature measurement unit selected.
SET TDSIndicates adjustment of the TDS factor.
no cor
cor lin
cor f(T)
CAL SETIndicates that calibration is in progress.
MTCIndicates that temperature compensation is manual.
ATCIndicates that temperature compensation is automatic.
Ref 20°C
Ref 25°C
parameterizing of the instrument is in progress, a calibration is in progress, a recording is in progress, the instrument is connected via the USB cable, auto o is deactivated (see § 4.4.3).
When lit steadily, indicates that the instrument’s memory is full.
When blinking, indicates that the instrument’s memory is 90% full.
Indicates the progress of the measurement. When all segments are lit, the measurement is stable.
When blinking, indicates that the measurement is over or that the calibration has been applied.
When lit steadily, indicates that recording is in progress.
When blinking, indicates waiting for the start of a recording session.
Indicates the type of temperature correction:
no cor: no correction cor lin: linear correction from 1.00 to 7.40%/°C. cor f(T): nonlinear correction, exceeding 7.40%/°C
Indicates the reference temperature, 20 or 25°C (68 or 77°F)
When the measurement exceeds the limits (whether positive or negative), the instrument displays OL.
9
2.5. SETTING THE TIME
The time of your instrument is set using the Data Logger Transfer software. Refer to §4.4.2
2.6. PROP
To make reading easier, the instrument can be set on its prop.
10
3. USE IN STAND-ALONE MODE
The instrument can operate in two modes: the stand-alone mode described in this section, the record mode, in which it is controlled by a PC. This mode is described in the next section.
To ensure proper operation of the instrument, always leave the cell connected and the cap on the USB connector.
3.1. CONNECTING THE CELL
When you rst use the instrument, connect the conductivity cell provided to the measuring instrument. Connect the connector of
the cell to the instrument and screw the ring all the way on.
3.2. CALIBRATION
3.2.1. PARAMETERIZING THE MEASUREMENTS
Before calibrating the cell, you can set various parameters.
The unit of temperature (°C or °F)
The reference temperature, 20 or 25°C (68 or 77°F). This is the temperature to which the measurement will be referred.
The temperature correction used to correct and display the measured conductivity as of the chosen reference temperature.
This is useful because the conductivity of a solution increases with increasing temperature.
cor lin: linear temperature correction
The measured conductivity is corrected and displayed using the following formula:
Σ
= σT / (1 + α (T - T
displayed
where: σ
σ
) / 100%)
ref
= the conductivity displayed by the instrument.
displayed
= the conductivity measured at temperature T.
T
α = the linear temperature correction coecient.
T = the measured temperature (°C)
T
= the reference temperature (20 or 25°C)
ref
Acids: 1 to 1.6%/°C Bases: 1.8 to 2.2%/°C Salts: 2.2 to 3.0%/°C Drinking water: 2.0%/°C Ultra-pure water: 5.2%/°C
no cor: no correction, α = 0%/°C. cor f(T): nonlinear temperature correction. See appendix 1 in §8.
TDS factor (f): between 0.40 and 1.00. This is used to calculate TDS (Total Dissolved Solids). It is expressed in mg/l or in ppm.
TDS = σ.f
Where: σ = conductivity on μS/cm
f = TDS factor
11
Long-press the SET key.
Choose the temperature unit (°C or °F) using the ▲ and
▼ keys.
> 2s
SET
Press the SET key and use the ▲ and ▼ keys to set the
temperature correction.
SET COR. SET COR.
no cor
SET
SET
SET
A long press scrolls the values faster.
cor lin
TEMP
Press the SET key and choose the reference temperature
(20 or 25°C) using the ▲ and ▼ keys.
SET REF
SET
Ref 25°C
Ref 25°C
Press the SET key and set the TDS factor (f) to between
0.40 and 1.00 using the ▲ and ▼ keys.
A sustained press on ▲ or ▼ scrolls through the values
faster.
SET
SET
TDS
Long-press the SET key to exit from parameterizing the
measurements.
A short press on the SET key takes you back to the choice
of temperature unit.
> 2s
SET
12
3.2.2. CHOICE OF STANDARD SOLUTION
> 2s
The C.A 10141 is calibrated for conductivity.
The calibration is done at one point. Choose the solution that lies in the range of interest to you.
Conductivity at 25°C
Conductivity147 µS/cm1408 µS/cm12.85 mS/cm
Resistivity6.8 kΩ.cm710 Ω.cm77.8 Ω.cm
TDS73,5 mg/l
TDS factor = 0,5
Salinity--7,4 psu
Rinse the cell in deionized water, then dry it. Pour the standard solution into a beaker.
746 mg/l
TDS factor = 0,53
7,45 g/l
TDS factor = 0,58
Long-press the key to switch the instrument on.
HO
2
Immerse the end of the cell in the standard solution.
Take care to completely immerse the sensitive part of the cell in the solution.
Minimum level of immersion.
Sensitive part of the cell.
The cell provided with the instrument has an integrated temperature sensor. There is therefore no need to measure the temperature of the solution.
13
The instrument displays the conductivity measurement along with the measured temperature, the reference temperature, and the time.
ATC = Automatic Temperature Compensation.
cor lin
ATC
Ref 20°C
3.2.3. CALIBRATION PROCEDURE
Calibration serves to determine the constant of the conductivity cell.
In order to avoid the inuence of the temperature, do the calibration at the temperature at which you will be making measurements.
With the instrument set to conductivity measurement, press
the CAL key.
The instrument briey displays the current calibration (the
coecient of the cell in cm-1).
CAL
END
There are six sets available, corresponding to the six international standard conductivity solutions.
Conductivity at 25°C
Calibration set 1147.0 µS/cm
Calibration set 21.408 mS/cm
Calibration set 312.85 mS/cm
Calibration set 484.0 µS/cm
Calibration set 51.413 mS/cm
Calibration set 612.88 mS/cm
It then prompts you to choose the calibration set.
CAL SET
ATC
You can modify these values in the Set.csv le (see § 4.3).
14
Choose the calibration set by long presses on the ▲ and
▼ keys.
Conrm the set by pressing the CAL key.
The instrument displays the measured conductivity and
the temperature.
cor lin
CAL SET
CAL
ATC
It performs the conductivity measurement and indicates its progress.
Do not withdraw the cell from the solution until the measurement is over.
To abort the calibration of the cell, long-press the END key before the end of the measurement.
Otherwise, when the measurement is over, the new calibration is applied.
Press the CAL key. The instrument exits from the calibration procedure and briey displays the cell coecient before changing
back to measurement mode.
END
CAL SET
ATC
Ref 20°C
CAL
END
During the calibrations, the cell coecient for a given cell must vary only a little. If you observe a large variation, it may be
necessary to redo the calibration, check the condition of the standard solution (in particular the use-by date), or replace the
cell.
15
3.2.4. RESTORE THE INITIAL CALIBRATION
Press the CAL key, then the MEM key.
CAL
END
If you do not want to restore the initial calibration, choose no before pressing the CAL key.
Otherwise, choose YES and press the CAL key. The cell coecient returns to 1.000.
3.3. CONDUCTIVITY MEASUREMENT
When the calibration is over, the cell is ready to make measurements.
Use a cell suited to the medium to be measured.
Between measurements, the cell must be rinsed, then dried.
For each measurement, wait for the measurement to stabilize and for the temperature to be correctly established.
MEM
REC
CAL
END
Conductivity is the capacity of a solution to conduct electric current. It is ions that carry the charges. The more ions the solution
contains, the more current it will conduct.
Immerse the end of the cell in the solution, making sure that the sensitive part is completely immersed in the solution.
The instrument displays the measurement at the chosen
reference temperature (20 or 25°C). Wait for it to stabilize
before reading it on the instrument’s display unit.
16
3.3.1. USING ANOTHER CONDUCTIVITY CELL
The cell provided with the instrument has an integrated temperature probe. If you use another cell, one without an integrated
temperature sensor, you must measure the temperature of the solution.
Use a four-pole cell suited to the medium to be measured.
The instrument indicates that the temperature can be modied by displaying MTC next to the temperature
MTC = Manual Temperature Compensation.
cor lin
MTC
Ref 20°C
You must then correct the temperature displayed, using the ▲ and ▼ keys, so that it is equal to the measured temperature of the
solution.
The instrument corrects the response of the cell as a function of the temperature.
To calibrate the cell, always correct the temperature rst.
To connect your cell, use an optional connection kit (see §1.2). These adapters also let you connect a PT1000 probe.
17
3.4. RESISTIVITY MEASUREMENT
> 2s
Resistivity is the reciprocal of conductivity.
You must rst calibrate the cell for conductivity and parameterize the measurements (in particular the temperature correction
and the reference temperature) before making resistivity measurements.
Long-press the key to switch the instrument on.
Press the key to switch to resistivity measurement.
cor lin
SET
Immerse the cell, after it has been rinsed and dried, in the solution to be measured. The instrument displays the measurement at the chosen reference temperature (20 or 25°C). Wait for it to stabilize to read it
on the instrument’s display unit.
The resistivity ranges from 2Ω.cm (σ = 500mS/cm) for a conducting solution to 18mΩ.cm (σ = 55nS/cm) for pure water.
ATC
Ref 20°C
18
3.5. TDS MEASUREMENT
> 2s
> 2s
The TDS (Total Dissolved Solids) measurement is used to estimate the level of dissolved solids in a solution.
You must rst calibrate the cell for conductivity and parameterize the measurements (in particular the temperature correction,
the reference temperature, and the TDS factor) before making TDS measurements.
Long-press the key to switch the instrument on.
Press the key twice to switch to TDS measurement.
cor lin
x 2
SET
Immerse the cell, after it has been rinsed and dried, in the solution to be measured. The instrument displays the measurement at the chosen reference temperature (20 or 25°C). Wait for it to stabilize to read it
on the instrument’s display unit.
3.6. SALINITY MEASUREMENT
The salinity measurement serves to evaluate the salt content of seawater. It is expressed in PSU (Practical Salinity Units).
The conversion from conductivity to salinity is done using the UNESCO formula (PSS-78), for a temperature of the solution ranging
from -2 to +35°C. Outside of this temperature range, the instrument displays O.L. The conductivity displayed is referred to 15°C.
You must rst calibrate the cell for conductivity and parameterize the measurements (in particular the temperature correction
and the reference temperature) before making salinity measurements.
Long-press the key to switch the instrument on.
Press the key three times to switch to salinity measurement.
ATC
Ref 25°C
cor lin
x 3
SET
ATC
19
3.7. RECORDING THE MEASUREMENTS
A short press on the MEM key records the measurement with the date and time. The MEM symbol is displayed briey.
It is not possible to record a single/an isolated measurement when/while the instrument is (already?) recording.
A long press on the REC key starts or stops a recording session. The REC symbol remains displayed for the duration of a
recording session. Auto o is deactivated (this means that the instrument is in permanent mode) and the P symbol is dis-
played.
Before starting a recording session, make sure that the battery life is sucient, or else connect the instrument to an external
power supply on/to a wall outlet using a micro USB cord.
When the memory is 90% full, the MEM FULL symbol blinks. When the memory is full, the MEM FULL symbol stops blinking.
To see the records, you must use a PC and install the Data Logger Transfer software (see §4).
3.8. ERRORS
The instrument detects errors and displays them in the form Er.XX. The main errors are the following:
Er.01: Hardware malfunction detected. The instrument must be sent in for repair.
Er.02: Error in internal memory. Format it using Windows.
Er.03: The update of the internal software is not compatible with the instrument (the software is that of another instrument of the
line). Install the correct internal software in your instrument.
Er.10: The instrument has not been calibrated or the calibration is not in conformity. The instrument must be sent back to customer
service.
Er.12: The update of the internal software is not compatible with the electronic boards in the instrument. Reload the previous
internal software into your instrument.
Er.13: Recording scheduling error. Check that the instrument’s time and the time of the Data Logger Transfer software are the
same.
Er.14: Calibration error. The measured value is too far from the value of the standard solution of the selected calibration set. Check
that the solution used in fact belongs to the selected set. If necessary, return to the initial calibration (see § 3.2.4).
Er.15: Calibration error. The stabilization time is too long.
Er.19: Calibration error. The temperature (ATC or MTC) is outside of the specications of the standard solutions. Redo the cali-
bration in a room where the temperature lies within the specications of the standard solution (see the Set.csv le, §4.3).
Er.20: Calibration error. The le dening the set of calibration solutions is missing. Download it from our web site:
www.chauvin-arnoux.com
Er.21: Calibration error. The le dening the set of calibration solutions is not in conformity. Check that it is the right le. If you have
modied it, check the format; in particular, the decimal separators must be points, not commas.
Er.22: Recording error. Power was cut o while recording was in progress.
Er.50: Calibration error. Measurement error for the calibration.
To exit from calibration errors, press the CAL key or the END key.
20
4. USE IN RECORDING MODE
The instrument can operate in two modes:
the stand-alone mode described in the previous section, the record mode, in which it is controlled by a PC. This mode is described below.
4.1. CONNECTION
The instrument communicates by a USB link, using the USB to micro USB cord provided.
4.2. GET DATA LOGGER TRANSFER SOFTWARE
Visit our web site to download the latest version of the application software:
www.chauvin-arnoux.com
Go to the Support tab, then Download our software. Then search on the name of your instrument.
Download the software, then install it on your PC.
You must have administrator privileges on your PC to install the Data Logger Transfer software.
Minimum computer requirements:
Windows 7 (32/64 bits)
2 GB of RAM
200 MB of disc space
Windows® is a registered trade mark of Microsoft®.
Do not connect the instrument to the PC until you have installed the Data Logger Transfer software.
4.3. USB LINK
Long-press the key to switch the instrument on.
Once the Data Logger Transfer software has been installed,
connect the instrument to the PC.
21
The symbol blinks.
The instrument is treated as a USB key.
Clicking Open the folder and display the les gives you access to its content.
This content includes the Set.csv le. You can open this le in a spreadsheet and modify it:
add or remove a calibration set modify a calibration set by adding, removing, or modifying standard solutions.
Do not change the structure of the le.
Number of calibration sets.
Calibration set number (1, 2, 3, etc.).
Conductivity of the standard solutions in mS/cm.
Evolution of the conductivity of the standard solution
as a function of temperature.
22
4.4. DATA LOGGER TRANSFER SOFTWARE
Once the instrument is connected to the PC, open the Data Logger Transfer software.
For context-sensitive information about the use of the Data Logger Transfer software, refer to the Help menu.
4.4.1. CONNECTING THE INSTRUMENT
To connect an instrument, click Add an instrument, then choose the type of connection (USB).
A window opens with a list of all instruments connected to the PC.
The name of the instrument will be formed from the model of the instrument and the warranty number: CA10141 - 123456ABC.
You can personalize your instrument by adding a name and a location, by clicking on or .
Choose your instrument in the list. The instrument then displays complete information about the instrument and its measure-
ments in progress.
4.4.2. DATE AND HOUR
In the Instrument menu, the icon lets you set your instrument’s date and time.
These cannot be changed while recording or when a recording session has been scheduled.
By clicking , you can choose the date and time display formats.
4.4.3. AUTO OFF
As default, the instrument switches itself o automatically after 5 minutes of operation without the user’s presence being conrmed
by a key-press. By clicking , you can change this value to 3, 10, or 15 minutes.
It is also possible to disable this auto-o function; the instrument then displays the P symbol.
23
4.4.4. PROGRAMMED RECORDING SESSIONS
By clicking , you can program a recording session. Assign a name to the recording session. Then enter a starting date and time
and an ending date and time or a duration. The maximum duration of a recording session depends on the memory available.
Choose a sampling period. The possible values are: 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 1 min, 2 min, 5 min, 10 min, 15 min, 30 min and
1 hour. The shorter the sampling period, the larger the recorded le.
Before and after the recording session, if the instrument is switched on, the sampling period will be that of the stand-alone mode (1s).
If the instrument is o when recording starts, it switches itself on by itself. Then it displays the measurement, which it refreshes at
each sampling period.
Before starting a recording session, make sure that the battery life is sucient, or else connect the instrument to an external
power supply to a wall outlet using a USB cord.
4.4.5. DISPLAY
By clicking on , then going to the Conductivity meter tab, you can modify the quantities displayed on the instrument and the
various values contained in Set-up.
4.4.6. READING THE RECORDS
The Data Logger Transfer software lets you read the records made. Click Recorded Sessions under the name of your instrument
to obtain a list of the records.
4.4.7. EXPORTING RECORDS
Once the list of the records is displayed, choose the one you want to export and convert it into a word-processing document (docx)
or a spreadsheet (xlsx), in order to be able to use it in the form of reports or curves.
It is also possible to export the data to the DataView application software (see § 1.2).
4.4.8. REAL-TIME MODE
Click Real-time data under the name of your instrument to see the measurements being made on the instrument as they are made.
24
4.4.9. FORMATTING THE MEMORY OF THE INSTRUMENT
The internal memory of the instrument is already formatted. But if there is a problem (if it becomes impossible to read or to write),
it may be necessary to reformat it (in Windows).
In this case, all of the data will be lost.
25
5. TECHNICAL CHARACTERISTICS
5.1. REFERENCE CONDITIONS
Quantity of inuenceReference values
Temperature23 ± 3 °C
Relative humidity45% to 75%
Battery supply voltage4 to 6.4 V
USB supply voltage5 V ± 5%
Electric eld< 1 V/m
Magnetic eld< 40 A/m
The intrinsic uncertainty is the error specied for the reference conditions.
It is expressed in % of the reading (L).
5.2. ELECTRICAL CHARACTERISTICS
The intrinsic uncertainties on the measurements are given for the instrument alone. The uncertainty of the cell used must be added
to them.
5.2.1. CONDUCTIVITY MEASUREMENTS
Specied measurement range0.050 to 4.999 µS/cm5.00 to 49.99 µS/cm50.0 to 499.9 µS/cm
Resolution (R)1 nS/cm10 nS/cm100 nS/cm
Intrinsic uncertainty of the instrument alone
without the cell (E)
Specied measurement range0.500 to 4.999mS/cm5.00 to 49.99mS/cm50.0 to 500.0mS/cm
Resolution (R)1 µS/cm10 µS/cm100 µS/cm
Intrinsic uncertainty of the instrument alone
without the cell (E)
5.2.2. RESISTIVITY MEASUREMENTS
Specied measurement range2.000 to 4.999 Ω.cm 5.00 to 49.99 Ω.cm50.0 to 499.9 Ω.cm500 to 4999 Ω.cm
Resolution (R)1 mΩ.cm10 mΩ.cm100 mΩ.cm1 Ω.cm
Intrinsic uncertainty of the instrument
alone without the cell (E)
± 0.5% L ± R *
± 0.5% L ± R *
± 0.5% L ± R *
Specied measurement range5.00 to 49.99 kΩ.cm 50.0 to 499.9 kΩ.cm 500 to 4999 kΩ.cm 5.00 to 19.99mΩ.cm
Resolution (R)10 Ω.cm100 Ω.cm1 kΩ.cm10 kΩ.cm
Intrinsic uncertainty of the instrument
alone without the cell (E)
*: provided that the cell used has a coecient:
≥ 1 from 2µS/cm to 499.9 mS/cm
≤ 0.5 from 50nS/cm to 49.99 mS/cm
26
± 0.5% L ± R *
5.2.3. TDS MEASUREMENTS
Specied measurement range0.001 to 4.999 mg/l5.00 to 49.99 mg/l50.0 to 499.9 mg/l
Resolution (R)1 µg/l10 µg/l100 µg/l
Intrinsic uncertainty of the instrument
alone without the cell (E)
Specied measurement range500 to 4999 mg/l5.00 to 49.99 g/l50.0 to 499.9 g/l
Resolution (R)1 mg/l10 mg/l100 mg/l
Intrinsic uncertainty of the instrument
alone without the cell (E)
5.2.4. SALINITY MEASUREMENTS
Specied measurement range: 2.0 to 42.0 PSU
Intrinsic uncertainty of the instrument alone without the cell: ± 0.5% L ± R
5.2.5. TEMPERATURE MEASUREMENTS
Temperature measurements are made with a PT1000 resistive probe built into the cell.
Specied measurement range- 10,0 to + 120,0°C14,0 to 248,0°F
ResolutionDisplay in °C: 0.1°C Display in °F: 0.1°F
Intrinsic uncertainty of the instrument
alone without the cell
< 0.4°C< 0.7°F
± 0.5% L ± R *
± 0.5% L ± R *
5.2.6. INFLUENCE OF TEMPERATURE
Inuence of temperature (from -10°C to 55°C at 50% RH) on the C.A 10141.
Type of measurementTypical inuenceMaximum inuence
If the calibration is done at the temperature of use, then the inuence of temperature is zero.
5.2.7. INFLUENCE OF HUMIDITY
Inuence of humidity (from 25 to 90% RH at 25°C) on the C.A 10141.
Type of measurementTypical inuenceMaximum inuence
Conductivity measurement > 0.2 µS/cm0.25%0.5%
Conductivity measurement ≤ 0.2 µS/cm0.5%1.5%
Temperature measurement≤ 0.2°C
27
5.3. MEMORY
The size of the ash memory containing the records is 8 MB.
This capacity makes it possible to record 100,000 measurements. Each measurement is recorded with the date and time.
5.4. USB
Protocol: USB Mass Storage
Maximum transmission speed: 12 Mbit/s
Type B micro-USB connector
5.5. POWER SUPPLY
The instrument is supplied by three 1.5V LR6 or AA alkaline batteries. It is possible to replace the batteries by rechargeable NiMH
batteries of the same size. But the rechargeable batteries, even when correctly charged, will not reach the voltage of the alkaline
batteries and the life indicated will be or .
The voltage range ensuring correct operation is from 4.0 to 6.4 V for the alkaline batteries and 4.0 to 5.2 V for the rechargeable
batteries.
Below 4V, the instrument stops making measurements and displays BAt.
Battery life is 300 h.
The instrument can also be powered via a USB-micro USB cord, connected either to a PC or to a wall outlet via a mains adapter.
The symbol is displayed in this case.
> 110 VAC
< 240 VAC
50 / 60 Hz
1 W
Connecting the external USB supply does not recharge the storage batteries.
28
5.6. ENVIRONMENTAL CONDITIONS
-4050
-10
401030200608070
Instrument for indoor and outdoor use.
Altitude < 2000m, and 10,000m in storage.
Pollution degree 2
%RH
90
80
70
60
321
50
40
30
20
10
0
°C
-30 -20
1 = Range of reference.
2 = Range of use.
3 = Range for storage (with neither primary nor rechargeable batteries. Except conductivity cell and standard solutions).
5.7. MECHANICAL CHARACTERISTICS
Dimensions (L x W x D) 211 x 127 x 54mm with the sheath
Dimensions (L x W x D) 206 x 97 x 49mm without the sheath
Mass of the instrument approximately 600g
Mass with the cell approximately 720g
Inrush protection IP 67 per IEC 60 529, with the USB connector closed and the cell connected.
IP 20 otherwise.
Drop impact test 1.50m.
5.8. COMPLIANCE WITH INTERNATIONAL STANDARDS
The instrument is compliant with standard IEC 61010-2-30.
5.9. ELECTROMAGNETIC COMPATIBILITY (CEM)
The instrument is compliant with standard IEC 61326-1.
29
6. MAINTENANCE
Except for the batteries, the instrument contains no parts that can be replaced by personnel who have not been
specially trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent” may gravely impair safety.
6.1. CLEANING
6.1.1. INSTRUMENT
Switch the instrument o.
Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth or forced air. Do not use
alcohol, solvents, or hydrocarbons.
The two parts of the connector (instrument end and cell end) must remain perfectly clean and dry. This is why it is best to leave
the cell connected to the instrument at all times.
If immersed, dry the instrument at the USB connector and the cell connector.
6.1.2. CELL
For the cleaning and maintenance of the cell, refer to its user manual.
6.1.3. PROTECTIVE SHEATH
To withdraw the protective sheath, rst disconnect the cell, then remove the wrist strap. Free the bottom of the casing from the sheath Then withdraw the housing from the sheath.
30
6.2. REPLACEMENT OF BATTERIES
The symbol indicates the remaining battery life. When the symbol is empty, all of the batteries must be replaced.
Switch the instrument o. Refer to §1.4 for the replacement procedure.
Spent batteries must not be treated as ordinary household waste. Take them to the appropriate recycling collection
point.
When the batteries are removed, the time is preserved for approximately two minutes.
6.3. SERIAL NUMBER
If you ever need to send your instrument in for repair, it will be helpful to know its serial number. It is found in the guarantee.txt le.
This le is in your instrument’s memory. To access it, simply connect the USB cord (see § 4.3).
The serial number is also marked on a label under the batteries.
6.4. CALIBRATION LOG
At each calibration, information is entered in the calib_log.txt le:
the date and time, the coecient of the cell on which the calibration was done.
This le is in your instrument’s memory. To access it, simply connect the USB cord (see § 4.3).
6.5. EMBEDDED SOFTWARE VERSION
To look up the version number of the embedded software in your instrument, press the MEM and keys simultaneously. The
instrument displays the number for a few seconds, then returns to measurement mode.
6.6. UPDATING THE EMBEDDED SOFTWARE
With a view to providing, at all times, the best possible service in terms of performance and technical improvements, Chauvin
Arnoux oers you the possibility of updating the internal software of this instrument by downloading, free of charge, the new
version available on our web site.
See you on our site:
www.chauvin-arnoux.com
Then go to “Support”, then “Download our software”, then “C.A 10141”.
Updating the embedded software may reset the conguration and cause the loss of the recorded data. As a precaution,
save the data in memory to a PC before updating the embedded software.
31
Embedded software update procedure
> 2s
Download the .bin le from our web site, then press and hold the MEM key and switch the instrument on by pressing the key.
The instrument displays BOOT.
+
MEM
REC
Release the keys and the instrument displays COPY, indicating that it is ready to receive the new software.
Connect the instrument to your PC using the USB cord provided.
32
Copy the .bin le to the instrument as if were a USB key. When the copying is done, press the MEM key and the instrument displays LOAD, indicating that the software is being installed.
When installation is done, the instrument displays PASS or FAIL according to whether or not the operation succeeded. If installation
fails, download the software again and repeat the procedure.
Then the instrument restarts normally.
After the internal software is updated, it may be necessary to recongure the instrument; see § 4.4.
33
7. WARRANTY
Except as otherwise stated, our warranty is valid for 24 months starting from the date on which the equipment was sold. Extract
from our General Conditions of Sale provided on request.
The warranty does not apply in the following cases:
Inappropriate use of the equipment or use with incompatible equipment;
Modications made to the equipment without the explicit permission of the manufacturer’s technical sta;
Work done on the device by a person not approved by the manufacturer;
Adaptation to a particular application not anticipated in the denition of the equipment or not indicated in the user’s manual;
Damage caused by shocks, falls, or oods.
The conductivity cells are wear parts. The life of your cell depends on how you use and maintain it. The cells are warranted for
12 months.
34
8. APPENDIX 1: NONLINEAR TEMPERATURE CORRECTION OF
THE CONDUCTIVITY
The nonlinear correction concerns natural water: underground water, surface water, drinking water, and sewage. It is dened by
ISO/DIN standard 7888, between 0 and 35.9°C. It is especially useful for low conductivity values.
The table below indicates the nonlinear correction, f25, applied to refer the conductivity measured at a temperature T to the 25°C
reference temperature.