Thank you for purchasing a Fluidotherapy Dry Heat Therapy Unit.
This manual contains general safety, operating, maintenance, and care instructions for the owners and operators of the
Fluidotherapy Dry Heat Therapy Units.
The specifications put forth in this manual were in effect at the time of the publication. However, owing to DJO, LLC's
policy of continuous improvement, changes to these specifications may be made at any time without obligation on the
part of DJO, LLC.
Read, understand, and follow the information contained in this manual.
Stay current with the latest clinical developments in the field of Dry Heat Therapy and observe all applicable
precautionary measures for treatment.
Keep informed on appropriate indications and contraindications for the use of Dry Heat Therapy.
Fluidotherapy® - Dry Heat Therapy
This equipment is to be used only under the prescription and supervision of a licensed practitioner.
1
1
ABOUT DRY HEAT THERAPY
Fluidotherapy® - Dry Heat Therapy
PRECAUTIONARY INSTRUCTIONS
The precautionary instructions found in this section and throughout
this manual are indicated by specific symbols. Understand these
symbols and their definitions before operating this equipment. The
definitions of these symbols are as follows:
= CAUTION - Text with a “CAUTION” indicator will explain
possible safety infractions that could have the potential to
cause minor to moderate injury or damage to equipment.
= WARNING - Text with a “WARNING” indicator will explain
possible safety infractions that will potentially cause serious
injury and equipment damage.
= DANGER - Text with a “DANGER” indicator will explain
possible safety infractions that are imminently hazardous
situations that would result in death or serious injury.
= EXPLOSION HAZARD - Text with an “Explosion Hazard”
indicator will explain possible safety infractions if this
equipment is used in the presence of flammable anesthetics.
= NON-IONIZING ELECTROMAGNETIC RADIATION - Text with a
Non-Ionizing Electromagnetic Radiation" indicator informs
the user of possible hazards resulting from elevated,
potentially dangerous, levels of non-ionizing radiation.
= Protective Earth (ground)
NOTE: Helpful information marked as "NOTE" may be found throughout this manual. Each "NOTE" is beneficial in the aid of
a particular area or function being described for this product.
DESCRIPTION OF DEVICE MARKINGS
Degree of Protection Against Electrical Shock ...........................
Complies with UL 2601-1 CSA C22.2 No. 601.1
Medical Electrical Equipment ............ ..............................................
Refer to Instruction Manual/Booklet ......... ....................................
2
ABOUT DRY HEAT THERAPY
Fluidotherapy® - Dry Heat Therapy
Read, understand, and practice the precautionary and operating •
instructions found in this manual. Know the limitations and hazards
associated with using any electrical device. Observe the precautionary
and operational decals placed on the unit.
•
DO NOT operate this unit in an environment where other devices
are being used that intentionally radiate electromagnetic energy in
an unshielded manner. Medical Electrical Equipment needs special
precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in this manual.
•
Portable and Mobile RF Communications equipment can affect Medical
Electrical Equipment.
•
This unit generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instructions, may cause
harmful interference to other devices in the vicinity. However, there is
no guarantee that interference will not occur in a particular installation.
Harmful interference to other devices can be determined by turning
this unit on and off. Try to correct the interference using one or more
of the following: reorient or relocate the receiving device, increase the
separation between the equipment, connect the equipment to an outlet
on a different circuit from that which the other device(s) are connected
and consult the DJO, LLC Service Department for help.
•
Clean Inlet Filter(s) daily before unit startup.
Turn unit OFF before positioning a patient or removing a patient from •
the unit.
•
Place the patient in a comfortable position allowing for correct
placement of the limb being treated.
•
Secure all entry ports before turning the unit ON.
Use only Cellex® Dry Heat Medium in the Fluidotherapy units.•
Refill unit daily to proper fill level with Chattanooga Cellex Dry Heat •
Medium.
Use only fingers to operate button controls on the control panel(s). Use of sharp •
objects such as pencils or pens will result in damage to the unit.
Check unit temperature before treating patient to ensure correct temperature.•
This unit should be transported and stored in temperatures between -40 °F and •
158 °F (-40 °C and -70 °C) with Relative Humidity between 10% - 100% to prevent
damage to the unit or its components.
This unit should be operated in temperatures between 110 °F and 125 °F (43 °C and •
52 °C) to prevent damage to the unit or its components.
DO NOT operate the unit when connected to any unit other than Chattanooga •
devices.
Change Cellex Dry Heat Medium every six (6) months.•
Do not use accessories other than those supplied with the unit, or recommended •
by DJO, LLC. The safety of other products has not been established, and their use
could result in injury to the patient.
3
ABOUT DRY HEAT THERAPY
Fluidotherapy® - Dry Heat Therapy
Federal law restricts this device to sale by, or on the order of, a physician or licensed •
practitioner. This device should be used only under the continued supervision of a
physician or licensed practitioner.
Make certain the unit is electrically grounded by connecting only to a grounded •
electrical service receptacle conforming to the applicable national and local
electrical codes.
Before administering any treatment to a patient, you should become acquainted •
with the operating procedures for each mode of treatment available, as well as the
indications, contraindications, warnings, and precautions. Consult other resources
for additional information regarding the application of Dry Heat Therapy.
The Fluidotherapy Dry Heat Therapy Unit should not be used adjacent to or stacked •
with other equipment. If adjacent or stacked use is necessary, the Fluidotherapy Dry
Heat Therapy Unit should be observed to verify normal operation in the configuration
in which it will be used.
For continued protection against fire hazard, replace fuses only with ones of the •
same type and rating.
This device should be kept away from children.•
To prevent electrical shock, disconnect the unit from the power source before •
attempting any maintenance procedures.
Use only processed dry heat medium in the unit such as Cellex to prevent excessive •
dusting.
Use only accessories that are specially designed for this •unit. Do not use accessories
manufactured by other companies on this unit. DJO, LLC is not responsible for any
consequence resulting from using products manufactured by other companies. The use
of other accessories or cables may result in increased emissions or decreased immunity
of this unit.
Dispose of all products in accordance with local and national regulations and codes.•
Fluidotherapy equipment not in use should be protected against unqualified use.•
The Fluidotherapy Dry Heat Therapy Unit is not suitable in the •
presence of a flammable anesthetic mixture with air, oxygen, or nitrous
oxide.
Perform all Preventive Maintenance as described in this manual. Failure to perform •
the Preventive Maintenance could result in the Cellex medium entering the heat
chamber of the unit and cause severe injury to patients as well as smoke damage
to the facility and the Fluidotherapy unit.
The solvents of adhesives and flammable solutions used for cleaning and •
disinfecting the unit should be allowed to evaporate before the unit is used.
4
ABOUT DRY HEAT THERAPY
Indications & Contraindications
Fluidotherapy® - Dry Heat Therapy
Indications
• Relief of local pain
• Treatment of local blood flow insufficiency
• Treatment in range of motion when combined with exercise
• Treatment for symptoms of non-rheumatoid arthritis
Contraindications
• This device should not be used for symptomatic pain relief unless
etiology is established or unless a pain syndrome has been
diagnosed.
• This device should not be used when cancerous lesions or open
wounds are present in the treatment area.
• Other counterindications are patients suspected of carrying
serious infectious disease or disease where it is advisable, for
general medical purposes, to suppress heat or fevers.
Adequate precautions should be taken when treating individuals •
with suspected or diagnosed medical conditions or diseases such as
heart problems, epilepsy, diabetes, etc.
Prior to treatment, consult a medical professional familiar with the •
precautionary measures to be taken for patients that may experience
allergic reactions to dust and pollen.
55
NOMENCLATURE
FLU110D & 110DE Unit Familiarization
1
8
2
Fluidotherapy® - Dry Heat Therapy
1. RESERVOIR LID & TREATMENT VIEWING WINDOW
Access to add/change medium, view patient treatment, and
clinician access to patient treatment area
2. MEDIUM/TREATMENT RESERVOIR
Treatment access and medium reservoir
3
3. HAND/ARM ACCESS PORT
Treatment port for Hand/Arm Treatment (end of unit)
7
4. REPLACEABLE INLET FILTERS
Filters room air entering the Fluidotherapy unit (1 each side)
5. FOUR POINT BASE
Casters included to convert to a Mobile Base
6. BLOWER HOUSING
6
4
Houses Blower & Heater
7. CONTROL PANEL
See page 8 for detailed description of controls
8. ELBOW/FOOT TREATMENT ACCESS PORT
Treatment port for Elbow and Foot
5
6
NOMENCLATURE
FLU115D & FLU115DE Unit Familiarization
Fluidotherapy® - Dry Heat Therapy
1
2
8
3
1. ELBOW/FOOT ACCESS PORT
Treatment port for Elbow and Foot Treatment, access to add/
change medium, clinician access to patient treatment area
2. HAND/ARM ACCESS PORTS
Treatment ports for Hand/Arm Treatment (end of unit)
3. INLET FILTERS
Filters room air entering the Fluidotherapy unit (1 each side)
4. MEDIUM/TREATMENT TUB
Treatment access and medium reservoir
7
5. MOBILE BASE
4
6. LOCKING CASTERS
7. BLOWER HOUSING
Houses Blower & Heater
8. CONTROL PANEL
See page 8 for detailed description of controls
5
6
77
NOMENCLATURE
Operating Controls (All Units)
12
1
11
78
10
9
Fluidotherapy® - Dry Heat Therapy
1.DISPLAYDisplays Treatment Time, Temperature, Airspeed, Pulse Time,
Preheat Timer, and Clock when the respective Mode is indicated
2. PREHEAT TIMER ON
2
3
4
5
6
Indicator light for Preheat Timer
3. INCREASE/HOUR
Use to increase Mode parameters and use to set the Hour when
setting Clock
4. DECREASE/MINUTE
Use to decrease Mode parameters and use to set the Minutes
when setting Clock
5. STOP
Use to stop treatment and Preheat Timer
6. MOTOR FUSE
7. START
Use to start treatment
8. INDICATOR BAR
Indicates Mode as they are chosen
9. HEATER FUSE
10. PREHEAT TIMER BUTTON
Press once to turn Preheat Timer On. Press twice to start
Preheat function.
11. MODE SELECT
Use to select desired Mode
12. PM
PM indicator for Clock
8
SPECIFICATIONS
FLU110D & FLU110DE
Length
Height
Fluidotherapy® - Dry Heat Therapy
MODES OF OPERATION
Continuous (Default)
Variable Adjustments for Time, Temp, and Air Speed
Pulse Mode.....................................................................OFF to 6 Sec ON/OFF
TREATMENT TIME........................................................................1 to 99 minutes
OPERATING TEMPERATURE...............................110 °F (43.3 °C) to 125 °F (51.6 °C)
AIR SPEED.............................................................0% to 100% (5% increments)
PREHEAT TIMER........................................115 °F (46.1 °C) with 50% Air Flow
MEDIUM CAPACITY.........................................Approximately 30 lb (13.6 kg)
INPUT POWER (FLU110D)..............................................120V, 50/60 Hz, 12A
INPUT POWER (FLU110DE).....................................230-240V, 50/60 Hz, 8A
Weight...............................................................70 lbs (31.7 kg) Less Medium
Shipping Weight....................................................................100 lbs (45.4 kg)
DEGREE OF PROTECTION AGAINST INGRESS OF WATER...............IPX0
ELECTRICAL CLASS.....................................................................................CLASS I
DEGREE OF PROTECTION AGAINST ELECTRICAL SHOCK......TYPE B
POWER CORD......................................2 meter cable length, 14 AWG, shielded
Width
Make certain the unit is electrically grounded by connecting only to a
grounded electrical service receptacle conforming to the applicable
national and local electrical codes.
99
SPECIFICATIONS
FLU115D & FLU115DE
Width
Height
Length
Fluidotherapy® - Dry Heat Therapy
MODES OF OPERATION
Continuous (Default)
Variable Adjustments for Time, Temp, and Air Speed
Pulse Mode.....................................................................OFF to 6 Sec ON/OFF
TREATMENT TIME........................................................................1 to 99 minutes
OPERATING TEMPERATURE..............................110 °F (43.3 °C) to 125 °F (51.6 °C)
AIR SPEED.............................................................0% to 100% (5% increments)
PREHEAT TIMER........................................115 °F (46.1 °C) with 50% Air Flow
MEDIUM CAPACITY.......................................Approximately 40 lbs (18.1 kg)
INPUT POWER (FLU115D)...............................................120V, 50/60 Hz, 12A
INPUT POWER (FLU115DE)...................................230-240V, 50/60 Hz, 10A
Weight................................................................60 lbs (27.2kg) Less Medium
Shipping Weight....................................................................100 lbs (45.4 kg)
DEGREE OF PROTECTION AGAINST INGRESS OF WATER. ..............IPX0
ELECTRICAL CLASS.....................................................................................CLASS I
DEGREE OF PROCTECTION AGAINST ELECTRICAL SHOCK....TYPE B
POWER CORD......................................2 meter cable length, 14 AWG, shielded
Make certain the unit is electrically grounded by connecting only to a
grounded electrical service receptacle conforming to the applicable
national and local electrical codes.
10
SPECIFICATIONS
Fluidotherapy® - Dry Heat Therapy
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES
TABLE 1: GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC EMISSIONS
The Fluidotherapy - Dry Heat Therapy Unit is intended for use in the electromagnetic environment specified in the table below. The user
of the Fluidotherapy - Dry Heat Therapy Unit should assure that it is used in such an environment.
The Fluidotherapy - Dry Heat Therapy Unit is intended for use in the electromagnetic environment specified in the table below. The user of the
Fluidotherapy - Dry Heat Therapy Unit should assure that it is used in such an environment.
Immunity
Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Fluidotherapy - Dry Heat Therapy Unit is used exceeds the applicable RF compliance level above, the Fluidotherapy - Dry Heat Therapy Unit should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Fluidotherapy - Dry Heat Therapy Unit.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Portable and mobile RF communications equipment should be used no closer to any part of the Fluidotherapy - Dry Heat Therapy Unit,
including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
3 V
d = [3.5]√P
V
1
d = [3.5]√P 80 MHz to 800 MHz
E
3 V/m
1
d = [7]√P 800 MHz to 2.5 GHz
E
1
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya, should be less than the compliance level in each
frequency rangeb.
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE: UT is the a.c. mains voltage prior to application of the test level.
±6 kV contact
±8 kV air
(60% dip in UT) for 5 cycles
(>95% dip in UT) for 5 sec
T
13
Floors should be wood, concrete, or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the Fluidotherapy - Dry Heat Therapy Unit requires
continued operation during power mains interruptions, it is recommended that
the Fluidotherapy - Dry Heat Therapy Unit be powered from an uninterruptible
power supply or a battery.
location in a typical commercial or hospital environment.
TABLE 4: RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT
AND THE FLUIDOTHERAPY - DRY HEAT THERAPY UNIT (CONTINUED)
The Fluidotherapy - Dry Heat Therapy Unit is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Fluidotherapy - Dry Heat Therapy Unit can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Fluidotherapy - Dry
Heat Therapy Unit as recommended below.
Separation Distance According to Frequency of Transmitter
Rated Maximum Output Power of
Transmitter
150 kHz to 80 MHz
W
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable
to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
d = [3.5]√P
V
1
14
m
80 MHz to 800 MHz
d = [3.5]√P
E
1
800 MHz to 2.5 GHz
d = [7]√P
E
1
SETUP
TREATMENT MODE PARAMETERS
ATTENTION!!!
ATTRITION OF PARTICLES DURING SHIPMENT
MAY RESULT IN INITIAL DUSTING AND SMALL
PARTICLE RELEASE. DUST AND ODOR WILL
DISSIPATE WITH USE.
Use the following instructions to set the various Mode Parameters to
the desired settings.
TREATMENT TIME
Press the “PRESS TO CHANGE MODE”
button until the indicator light is beside
“TREATMENT TIME”.
NOTE:
The default time of “20:00” will be
displayed.
Fluidotherapy® - Dry Heat Therapy
Press the “INCREASE” or “DECREASE”
buttons until desired Treatment Time is
displayed.
NOTE:
The “TREATMENT TIME” can be adjusted in
one minute increments from 1 to 99.
TREATMENT TEMPERATURE
Press the “PRESS TO CHANGE MODE”
button until the indicator light is beside
“TEMPERATURE”.
NOTE:
The default temperature of “115 °F (46.1 °C)
will be displayed when unit is first plugged
into wall outlet.
The “F” or “C” and “TEMPERATURE” indicator
will flash while the programmed Treatment
Temperature is being displayed.
When the Bed Temperature is displayed, the
“F” or “C” and the “TEMPERATURE” indicator
will illuminate steadily.
15
15
SETUP
TREATMENT MODE PARAMETERS (CONTINUED)
Fluidotherapy® - Dry Heat Therapy
TREATMENT TEMPERATURE (CONTINUED)
Press the “INCREASE” or “DECREASE”
buttons until desired Treatment Temperature
is displayed.
NOTE:
°F can be changed to °C and vice versa as
desired. See page 20 for instructions.
Treatment Temperature can be adjusted in 1°
increments from 88 °F to 130 °F
°C).
AIR SPEED
Press the “PRESS TO CHANGE MODE”
button until the indicator light is beside
“AIR SPEED”.
NOTE:
The default speed of “50” will be displayed.
(31.1 °C to 54
Press the “INCREASE” or “DECREASE”
buttons until desired Air Speed is
displayed.
NOTE:
The “AIR SPEED” is adjusted in increments
of 5 from 0 to 100.
PULSE TIME
Press the “PRESS TO CHANGE MODE”
button until the indicator light is beside
“PULSE TIME”.
NOTE:
The “PULSE TIME” allows the unit to
operate by pulsing the medium during
treatment. If selected and set, the medium
will pulse on and off in equal increments.
The Factory Default is “OFF”.
EXAMPLE: “PULSE TIME” is set at “5”.
The unit will pulse the medium during
treatment, five seconds on and five
seconds off.
16
SETUP
TREATMENT MODE PARAMETERS (CONTINUED)
Fluidotherapy® - Dry Heat Therapy
PULSE TIME (CONTINUED)
Press the “INCREASE” or “DECREASE”
buttons until desired Pulse Time is
displayed.
NOTE:
“PULSE TIME” is adjusted in one second
increments from “OFF” to 6 seconds.
START TREATMENT
Press the “START” button.
17
17
STOP TREATMENT
Press the “STOP” button.
SETUP
TIME CONTROLLED PARAMETERS
Fluidotherapy® - Dry Heat Therapy
PREHEAT TIMER
Press the “PRESS TO CHANGE MODE”
button until the indicator light is beside
“PREHEAT TIMER”.
NOTE:
The “PREHEAT TIMER” allows the unit to
preheat the medium. The unit will run at 50%
Air Speed until the unit reaches the default
preheat temperature or 90 minutes, whichever
comes first. The “PREHEAT TIMER” will operate
Monday through Friday only and can be set
to automatically start at a predetermined time
and heat the medium to a predetermined
temperature. See page 21 for setting the
default parameters of the “PREHEAT TIMER.
Press the “PRESS ONCE TO TURN ON PREHEAT
TIMER” button to illuminate the “PREHEAT
TIMER ON” indicator light. Press again to start
the “PREHEAT TIMER”.
NOTE:
The “PREHEAT TIMER” indicator light must be
set at the end of each day in order for it to
automatically come on the next day.
The “STOP” button is used to turn off the
Preheat if desired.
CLOCK
Press the “PRESS TO CHANGE MODE” button
until the indicator light is beside “CLOCK”.
NOTE:
The “PREHEAT TIMER” will not function until
the Clock is set.
Press the “INCREASE” button to set hours.
Press the “DECREASE” button to set minutes.
NOTE:
The “PM” indicator light will illuminate when
PM hours are reached.
18
SETUP
PREFERENCE MODE DEFAULT PARAMETERS
Fluidotherapy® - Dry Heat Therapy
ENTERING “PrEF” MODE (PREFERENCE MODE)
Simultaneously press the “PRESS TO CHANGE MODE”, “INCREASE”, and
“DECREASE” buttons. “PrEF” will be displayed.
NOTE:
The “PrEF” mode allows the user to change the default settings.
TREATMENT TIME DEFAULT
Press the “PRESS TO CHANGE MODE” button
until the indicator light is flashing beside
“TREATMENT TIME”.
Press the “INCREASE” and “DECREASE”
buttons to set unit to the desired default
treatment time.
Press the “PRESS TO CHANGE MODE” button
to save the new setting.
NOTE:
The default setting is 20:00 minutes.
1919
SETUP
PREFERENCE MODE DEFAULT PARAMETERS (CONTINUED)
TEMPERATURE READING DEFAULT
Fluidotherapy® - Dry Heat Therapy
TREATMENT TEMPERATURE DEFAULT
Press the “PRESS TO CHANGE MODE”
button until the indicator light is flashing
beside “TEMPERATURE”.
Press the “INCREASE” button to set unit
to display °F or °C as desired for the
default setting.
Press the “PRESS TO CHANGE MODE”
button to save the new setting.
NOTE:
The default setting is °F.
Press the “PRESS TO CHANGE MODE”
button until the indicator light is flashing
beside “TEMPERATURE” and the existing
default temperature is displayed.
Press the “INCREASE” and “DECREASE”
buttons to set unit to the desired default
treatment temperature.
Operating Temperature Range is:
110 °F - 125 °F (43.3 °C - 51.6 °C)
Press the “PRESS TO CHANGE MODE”
button to save the new setting.
20
SETUP
PREFERENCE MODE DEFAULT PARAMETERS (CONTINUED)
PULSE MODE (Enable/Disable)
Fluidotherapy® - Dry Heat Therapy
PREHEAT TIMER MODE (Enable/Disable)
Press the “PRESS TO CHANGE MODE”
button until the indicator light is flashing
beside “PULSE TIME”.
“EnAb” (enable) or “dISA” (disable)
will display.
Press the “INCREASE” button to set unit to
display “EnAb” or “dISA” as desired.
Press the “PRESS TO CHANGE MODE”
button to save the new setting.
Press the “PRESS TO CHANGE MODE”
button until the indicator light is flashing
beside “PREHEAT TIMER”.
“EnAb” (enable) or “dISA” (disable)
will display.
Press the “INCREASE” button to set unit to
display “EnAb” or “dISA” as desired.
Press the “PRESS TO CHANGE MODE”
button to save the new setting.
NOTE:
If Clock is disabled, the “PREHEAT TIMER”
will automatically be disabled.
2121
SETUP
PREFERENCE MODE DEFAULT PARAMETERS (CONTINUED)
CLOCK (Enable/Disable)
Fluidotherapy® - Dry Heat Therapy
SETTING DAY OF WEEK
Press the “PRESS TO CHANGE MODE”
button until the indicator light is flashing
beside “CLOCK”.
“EnAb” (enable) or “dISA” (disable)
will display.
Press the “INCREASE” button to set unit to
display “EnAb” or “dISA” as desired.
Press the “PRESS TO CHANGE MODE”
button to save the new setting.
NOTE:
If Clock is disabled, the “PREHEAT TIMER”
will automatically be disabled.
Press the “PRESS TO CHANGE MODE”
button until “dAy__” (__= day code) is
displayed.
Press the “INCREASE” or “DECREASE” button
to set unit to display the desired day code.
Press the “PRESS TO CHANGE MODE” button
to save the new setting.
22
SETUP
PREFERENCE MODE DEFAULT PARAMETERS (CONTINUED)
EXITING “PrEF” MODE
Press the “PRESS TO CHANGE MODE” buttons until “PrEF” is displayed.
Simultaneously press the “INCREASE”, “DECREASE”, and “STOP” buttons. The
default treatment time will display and the indicator will be beside “TREATMENT
TIME”.
Fluidotherapy® - Dry Heat Therapy
2323
OPERATION
Fluidotherapy® - Dry Heat Therapy
PATIENT PREPARATION
Clean the patient treatment area thoroughly with an antimicrobial •
soap and clean water as per industry, facility, regulatory standards,
and universal skin washing procedures.
Following the skin washing procedure, apply a hospital grade •
antiseptic skin cleanser, according to the cleanser manufacturer's
recommended instructions for use.
Reference the maintenance schedule and make certain the •
preventive maintenance has been performed prior to starting the
unit.
STARTING TREATMENT
STARTING
Make certain all control panel settings have
been made, unit is preheated, and patient
is in proper position with sleeve(s) secure
prior to starting the treatment. Refer to
pages 15 - 17 for setup of the unit.
Press “START” button to begin treatment.
STOPPING TREATMENT
STOPPING
Treatment will automatically stop when the
treatment time reaches zero.
Should it be desired to stop treatment
before the timer reaches zero, push the
“STOP” button.
NOTE:
Should the treatment be stopped before
the timer reaches zero, it will be necessary
to re-set the “TREATMENT TIME” in order to
complete the prescribed treatment.
24
PREVENTIVE MAINTENANCE
Fluidotherapy® - Dry Heat Therapy
DAILY MAINTENANCE
Before any maintenance is performed or attempted, unplug the unit
from the power source to prevent the possibility of electrical shock.
CLEAN INLET FILTERS
At the end of each work day, unplug the
unit and clean the Inlet Filters on the unit.
Carefully remove the filter retainer and
wash the filter and screen with a mild
antibacterial soap and water. Thoroughly
dry the filter and screen before placing
back on the unit.
NOTE:
Should your unit have the earlier style inlet
filters, gently clean the filter using a soft
bristle brush. Be careful not to puncture or
damage the filter.
Should the filter become damaged, torn, or
punctured, call your dealer for replacement
of the filter before resuming operation.
REFILL WITH CELLEX MEDIUM
Refill the unit with Cellex Dry Heat Medium to
approximately one (1) inch above the bottom of
the arm treatment ports of the unit.
NOTE:
Use of other than Cellex Medium may cause
premature failure of the Fluidotherapy unit(s).
INSPECT SLEEVE CONDITION
Inspect the port sleeves for tears, rips, and weak seams. Replace all
sleeves that show signs of tears, rips, weak, or loose seams, or excessive
wear. Keeping the sleeves in excellent condition prevents excessive
spillage of the Cellex medium and prevents the medium from entering
the heat chamber of the unit.
ARM PREHEAT TIMER
Plug the unit back into an approved power
source. If the “Preheat Timer” is used, press
the “PREHEAT TIMER” button once and the
“PREHEAT TIMER ON” indicator will illuminate.
NOTE:
The "PREHEAT TIMER ON" indicator must be
illuminated in order for the Preheat Timer to
automatically come on at the default time the
next morning.
The Fluidotherapy Preheat Timer does not
operate on Saturday or Sunday. However,
for Monday morning preheat, set the unit as
25
25
described above on Friday evening.
PREVENTIVE MAINTENANCE
Fluidotherapy® - Dry Heat Therapy
WEEKLY MAINTENANCE
Each week all sleeves of the Fluidotherapy unit should be laundered in a
mild antibacterial detergent. Allow the sleeves to air dry or dry on a low
temperature setting. Drying the sleeves in high temperatures could cause
the sleeves to shrink or become distorted, resulting in the sleeve(s) not
properly fitting when placing them back onto the unit.
REMOVING TOP SLEEVES (All Models)
Unplug the unit from the power source.
Carefully remove the rectangular bezel(s)
from the top of the unit.
Carefully pull the hook and loop fasteners
apart to remove the sleeve from the unit.
Replace in reverse order making certain
the corners are well seated to form the seal
required to prevent excessive escape of
medium.
REMOVING END SLEEVE
(Model FLU110D & FLU110DE)
Carefully pull the hook and loop fastener on
the sleeve to remove sleeve from unit.
Replace in reverse order making certain
the corners are well seated to form the seal
required to prevent excessive escape of
medium.
26
Loading...
+ 84 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.