Read, understand and follow the Safety Precautions and information contained in this manual.
This manual contains the necessary safety, and field service information for those Field Service Technicians, approved by
Chattanooga Group, to perform field service on the Fluido DHT Models 1480 and 1481 units.
At the time of publication, the information contained herein was current and up to date. However, due to continual
technological improvements and increased clinical knowledge in the field of fluidotherapy, as well as Chattanooga Group’s
policy of continual improvement, Chattanooga Group reserves the right to make periodic changes and improvements to their
equipment and documentation without any obligation on the part of Chattanooga Group.
It is the sole responsibility for field technicians to stay informed and trained in the latest technology utilized in the Fluido DHT
Models 1480 and 1481 units by Chattanooga Group. From time to time, as significant improvements are incorporated, Service
Bulletins will be produced and made available on our web site (www.chattgroup.com) in lieu of reprinting a complete manual
prematurely. These Service Bulletins will provide updated service information and technological improvements to the Fluido
DHT Models 1480 and 1481 for use by approved service technicians.
Due to the complex nature of the technology utilized by Chattanooga Group, the recommended troubleshooting techniques
for PC Boards are to determine “Bad Board” and PC Board replacements only. No board component level troubleshooting is
recommended nor will information or parts be supplied by Chattanooga Group. Any PC Board component level troubleshooting
performed will be at sole risk and liability of the Service Technician performing such troubleshooting techniques.
This equipment is to be sold and used only under the prescription and supervision of a licensed medical practitioner.
Fluido DHT™ Dry Heat Therapy Unit
This equipment is to be serviced only by an “Approved Service Technician”.
For Additional Service Contact:
Chattanooga Group
DHT Support Department
Toll Free: 1-866-864-0598
Outside USA: +1-423-870-7200
1
Page 6
SYMBOL DEFINITION
The following symbols are located on the Chattanooga Group Fluido DHT Dry Heat Therapy Unit Control Panel. Understand the
meaning of each symbol before attempting any operation or use of the unit.
This symbol indicates therapy session Time adjustment.
This symbol indicates therapy session Temperature adjustment.
This symbol indicates therapy session Air Speed adjustment.
This symbol indicates Upward adjustment in function parameters.
This symbol indicates Downward adjustment in function parameters.
This symbol with the illuminated Blue indicator light indicates the Intake Filter requires changing.
Fluido DHT™ Dry Heat Therapy Unit
This symbol with the illuminated Blue indicator light indicates the Cellex® media requires changing.
This symbol indicates Start therapy session.
This symbol indicates Stop therapy session.
This symbol indicates Pulse Mode for pulse therapy sessions.
This symbol indicates Preheat Standby function.
2
Page 7
SAFETY PRECAUTIONS
Precautionary Symbol Definitions
The precautionary instructions found throughout this
manual are indicated by specific symbols. Understand
these symbols and their definitions before operating or
servicing this equipment. The definitions of these
symbols are as follows:
CAUTION
CAUTION
Text with a “CAUTION” indicator will explain
possible safety infractions that could have the
potential to cause minor to moderate injury or
damage to equipment.
WARNING
WARNING
Text with a “WARNING” indicator will explain possible safety infractions that will potentially cause
serious injury and equipment damage.
DANGER
DANGER
Text with a “DANGER” indicator will explain possible
safety infractions that are imminently hazardous
situations that would result in death or serious injury.
NOTE:
Throughout this manual “NOTE” may be found.
The Notes are helpful information to aid in the
particular area or function being described.
Safety Precautions
Read, understand and follow all safety precautions
found in this manual. The following are general safety
precautions that must be read and understood before
attempting any service techniques on these units.
Throughout this manual, specific safety precautions will
be found. Read, understand and follow all safety
precautions.
Fluido DHT™ Dry Heat Therapy Unit
CAUTION
• Read, understand and practice the precautionary and
operating instructions found in this manual. Know the
limitations and hazards associated with using any electrical
device. Observe the precautionary and operational decals
placed on the unit.
• DO NOT operate the unit when connected to any unit other
than Chattanooga Group devices.
• Refill unit daily to proper fill level with Chattanooga Group
Cellex® Dry Heat Medium.
• Change Cellex Dry Heat Medium every six (6) months.
• Use only Cellex Dry Heat Medium in the Fluidotherapy units.
• Clean Inlet Filter(s) daily before unit startup.
• Use only fingers to operate button controls on the control
panel(s). Use of sharp objects such as pencils or pens will
result in damage to the unit.
• Turn unit to the “Standby” mode before positioning a patient
or removing a patient from the unit.
• After or between treatments, do not immediately unplug or
turn the power off from the unit. Allow the unit to process
through the “Cool Down” cycle. The unit cycles into standby
mode after treatment time has elapsed. Standby can be
disabled by pressing the preheat/standby button. The unit
goes to a 3 minute cool down mode after standby is turned
off. Turning the power off before cool down completes is
potentially hazardous to the equipment, and could lead to
failure of the unit. It is recommended that power be supplied
to the unit at all times. Keep in mind that the recommended
treatment air speed is 50%
• Secure all entry ports before turning the unit ON.
• Check unit temperature before treating patient to ensure
correct temperature.
• Place the patient in a comfortable position allowing for
correct placement of the limb being treated.
• Proper storage and transport temperatures for the Fluido DHT
units are 40°F - 158°F (4.5 °C - 70°C). Relative Humidity
85%.
• This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to
other devices in the vicinity. However, there is no guarantee
that interference will not occur in a particular installation.
Harmful interference to other devices can be determined by
turning this equipment on and off. Try to correct the
interference using one or more of the following: Reorient
or relocate the receiving device, increase the separation
between the equipment, connect the equipment to an outlet
on a different circuit from that to which the other device(s)
are connected and/or consult the factory field service
technician for help.
3
Page 8
SAFETY PRECAUTIONS
Fluido DHT™ Dry Heat Therapy Unit
WARNING
• Federal law restricts this device to sale by, or on the order
of, a physician or licensed practitioner. This device should
be used only under the continued supervision of a physician
or licensed practitioner.
• For continued protection against fire hazard, replace fuses
only with ones of the correct type and rating.
•
Make certain the unit is electrically grounded by connecting
only to a grounded electrical service receptacle conforming to
the applicable national and local electrical codes.
• This device should be kept away from children.
• Care must be taken when operating this equipment around
other equipment. Potential electromagnetic or other
interference could occur to this or to the other equipment.
Try to minimize this interference by not using other equipment
in conjunction with it.
•
Before administering any treatment to a patient you should
become acquainted
mode of treatment available, as well as the indications,
contraindications, warnings and precautions. Consult other
resources for additional information regarding the application
of Dry Heat Therapy.
• To prevent electrical shock, disconnect the unit from the
power source before attempting any maintenance
procedures.
• Use only Cellex® processed dry heat medium in the unit to
prevent excessive dusting.
• Adequate precautions should be taken when treating
individuals with suspected or diagnosed medical
conditions or diseases such as heart problems, epilepsy,
diabetes, etc.
• Prior to treatment, consult a medical professional
familiar with the precautionary measures to be taken for
patients that may experience allergic reactions to dust
and pollen.
• Properly dispose of used Cellex according to National and
local laws, rules and regulations.
with the operating procedures for each
DANGER
• Perform all Required Maintenance as described in this and
the User Manual. Strict adherence to the Required
Maintenance for the Fluido DHT units is mandatory. Failure
to perform the Required Maintenance could result in the
Cellex medium entering the heat chamber of the unit(s) and
cause severe injury to patients as well as smoke damage to
the facility and the Fluido DHT unit(s).
• Make certain the unit is unplugged from the power source
before attempting any removal and replacement procedures
on the unit.
4
Page 9
THEORY OF OPERATION
The Fluido DHT utilizes ambient air and pressurizes it via an internal blower. The pressurized air is directed across a heating
element bringing the media to the desired treatment temperature. The heated, pressurized air is then diffused across and
through a baffle to fluidize and heat the Cellex media in a patient treatment reservoir. All treatment parameters; air speed,
temperature, treatment time and the unit preheat settings are programmed by the operator with the touch panel user interface.
The Fluido DHT unit base incorporates four locking casters and a manually operated hydraulic lift to adjust height and rotation
of the treatment reservoir for patient comfort.
Fluido DHT™ Dry Heat Therapy Unit
5
Page 10
NOMENCLATURE
Fluido DHT
The nomenclature graphics below, Figure 3.1, indicate the
general locations of the major components of the Fluido
DHT unit.
Fluido DHT™ Dry Heat Therapy Unit
Know the components and their functions before performing
any operation of or service to the Fluido DHT Model 1480
and 1481 unit.
13
12
14
15
1
2
3
4
5
6
11
7
10
1. Reservoir Lid - Allows access to the treatment
cavity for adding Cellex® medium.
2. Latches - Secures Reservoir Lid.
3. Side Access Ports - Four available Side Access
Ports: two on each end of the unit.
4. Reservoir/Treatment Cavity - Patient treatment
cavity and Cellex Reservoir.
5. Treatment Limb Sleeves - Replaceable and
launderable patient limb treatment sleeves.
6. Heat Chamber - Houses Heating element.
7. Locking Casters - Four Locking Casters for
securing the unit in place for treatment.
8. Elevation Adjustment Cylinder - Adjusts and
maintains reservoir to desired height for patient
comfort.
8
9
9. Unit Base - Rigid unit base for ease in
transporting unit to different locations for
treatment.
10. Height Adjustment Pedal - Used to raise, rotate,
release, and lock the cylinder for height
adjustment of the unit.
11. Intake Filter - Air intake filter. Requires periodic
replacement.
12. Blower Housing - Houses Blower Motor.
13. Control Panel - Operator Controls. See Page 7
for detail description of each control.
14. Treatment Cavity Viewing Window - View
treatment area during treatment.
15. Top Access Port - Top Treatment access port
with sleeve.
6
Page 11
NOMENCLATURE
Fluido Control Panel
The Control Panel nomenclature graphics below, Figure 3.2,
indicate the location and functions of the Fluido DHT Model
1480 and 1481 control panel.
12
Fluido DHT™ Dry Heat Therapy Unit
Know the components and their functions before performing
any operation of or service to the Fluido DHT Model 1480
and 1481 unit.
1
2
3
11
10
9
8
1. Panel Display - Displays settings and operational
parameters.
2. Temperature - Use in conjunction with the Up
and Down arrows to set operation temperature.
Temperature can be adjusted in 1° increments.
The available temperature range is 88° F to
125° F (31° C to 52° C).
3. Air Speed - Use in conjunction with the Up and
Down arrows to adjust air speed for fluidization
of the Cellex® Medium. Available speeds range
from 5% to 100%.
4. Pulse Mode - Turn Pulse Mode on and off as
well as adjust pulse time. Available Pulse Time
is from 1 second on/1 Second off to 6 Seconds
on/6 Seconds off.
5. Up and Down Arrows - Use in conjunction with
other mode or function buttons to set desired
parameters.
4
5
6
7
6. Preheat Standby - Turn the Preheat Standby
Mode on and Off as well as set the parameters
desired for the Preheat Standby Mode to
automatically start.
7. Treatment Stop - Press to stop treatment.
8. Treatment Start - Press to start treatment.
9. Clear Button - Used to turn off maintenance
indicators after maintenance has been properly
performed.
10. Media - Indicator will light when it is time to
change the Cellex® Medium.
11. Filter - Indicator will light when it is time to
change the required filters.
12. Time - Use in conjunction with the Up and
Down arrow buttons to set treatment time.
Available time is 1 to 99 minutes in one minute
increments, or Continuous.
Attention, consult accompanying documentation.
Ordinary equipment: Not designed to prevent ingress of water.
*Dependent upon airspeed selected and ambient temperature.
8
Page 13
MAINTENANCE
Fluido DHT™ Dry Heat Therapy Unit
DAILY MAINTENANCE
WARNING
• Before any Maintenance is performed or attempted,
unplug the unit from the power source to prevent the
possibility of electrical shock.
CLEAN INLET FILTERS
At the end of each work day,
unplug the unit and clean the Inlet
Filters on the unit by brushing
them off.
If your filter screens become
stained with dust residue and will
no longer brush clean, carefully
remove the filter retainer and
wash the filter and screen with a
mild antibacterial soap and water.
Thoroughly dry the filter and
screen before placing back on the
unit.
Should the filter become damaged,
torn, or punctured, call your dealer
for replacement of the filter before
resuming operation.
REFILL WITH CELLEX® MEDIUM
Refill the unit with Cellex Dry Heat
Medium to the fill level indicator
in the reservoir.
NOTE:
Use of other than Cellex Medium
CELLEX
FILL LEVEL
INSPECT SLEEVE CONDITION
Inspect the port sleeves for tears, rips, and weak seams.
Replace all sleeves that show signs of tears, rips, weak
or loose seams, or excessive wear. Keeping the sleeves in
excellent condition prevents excessive spillage of the Cellex
medium.
may cause premature failure of the
Fluido DHT.
REMOVING TOP SLEEVES
Unplug the unit from the power
source.
Remove Treatment Reservoir Top
turn upside down and remove the
retaining ring from the top sleeve
with a 1/8” Allen Wrench. Remove
sleeve and launder.
Replace in reverse order making
certain the retaining ring has the
sleeve captured completely to
prevent escape of the Cellex®
around the sleeve.
REMOVING END SLEEVE
Remove Treatment Reservoir Top.
Drain the Cellex from the Reservoir.
See page 10.
Vacuum the reservoir making
certain to clean all Cellex from the
screw heads.
Remove seven screws retaining
the top piece of perforated metal.
Remove perforated metal from
reservoir.
MONTHLY MAINTENANCE
Each month, all sleeves of the Fluido DHT unit should be
laundered in a mild antibacterial detergent. Allow the sleeves
to air dry or dry on a low temperature setting. Drying the
sleeves in high temperatures could cause the sleeves to shrink
or become distorted resulting in the sleeve(s) not properly
fitting when placing them back onto the unit.
Pull the Reservoir Insert free from
the reservoir upper lip and hold
away from sleeve.
9
Page 14
MAINTENANCE
Fluido DHT™ Dry Heat Therapy Unit
Pull the sleeve off of the mounting
boss on the inside of the unit.
Launder and replace in reverse
order.
Reseat the Reservoir Insert into
the upper lip of the reservoir.
Refill the reservoir with Cellex
Medium to the proper level. See
image on the right.
BI-ANNUAL MAINTENANCE
In order to maintain the Chattanooga Group Fluido DHT in
optimum operating condition, the Cellex® Media should be
changed bi-annually or when the Media indicator on the control
panel is illuminated.
CHANGING CELLEX MEDIA
Fill the reservoir with new Cellex
to the Fill Line. The reservoir
should require approximately 30
lbs (13.5 kg) of Cellex Media.
Run the unit for one hour at 50%
CELLEX FILL
LEVEL INDICATOR
Air Speed after filling with new
Cellex and before any patient
treatment to ensure proper
operation of the unit.
WARNING
• Do not allow soap and water to penetrate the
distributor when cleaning the unit. Components
below the distributor can be damaged and the
unit may not operate properly should they become
wet. If moisture is suspected to have penetrated
the distributor, call a certified service technician
for service before the unit is filled with Cellex and
operated.
• Use only Cellex
unit to prevent excessive dusting.
• Properly dispose of used Cellex according to
national and local laws, rules, and regulations.
®
processed dry heat medium in the
Set the Air Speed to 5% or 10%.
Remove the Reservoir Drain
Plug located on the back of the
unit using a coin or flat blade
screwdriver.
Place an empty container in
position to catch the media and
press the Treatment Start button
on the control panel. The media
should freely flow out of the unit
into the catch container.
After the media has stopped
flowing out of the unit, press
Treatment Stop and disconnect
the unit from the power outlet.
Vacuum out any remaining Cellex
and clean the inside of the
Reservoir using a cloth dampened
with a mild antibacterial soap
and water. Allow the reservoir to
completely dry before filling with
new Cellex.
10
Page 15
MAINTENANCE
Fluido DHT™ Dry Heat Therapy Unit
VERIFYING TEMPERATURE CALIBRATION
After each change of Cellex media, perform a temperature
check to ensure that calibration of unit has not changed.
The temperature displayed on the display is an average of
the temperature throughout the media chamber. Because the
media is fluidized by the movement of forced air, there are
zones throughout the chamber. The temperatures measured in
each zone may vary by ± 8° F or ± -13°C.
Once the Cellex has been changed and operated at 50% for
1 hour, turn off the unit and immediately remove the lid to
the unit. With a calibrated thermometer, take a temperature
reading in the following zones – approximately half way down
from the top of the media.
ADDITIONAL FLUIDOTHERAPY
MAINTENANCE REQUIREMENTS
The following additional maintenance requirements must be
scheduled and performed as described to ensure that the unit
is operating efficiently, safely, and functioning optimally. A
blank Maintenance Record is provided on page 12 to aid in the
scheduling and record keeping of this prescribed maintenance
program. The following maintenance procedures must be
performed by a Chattanooga Group qualified service technician
trained in the maintenance requirements of the Chattanooga
Group Fluido DHT units.
QUARTERLY (Every 3 Months)
The following maintenance must be performed on all units
quarterly by a certified service technician.
• Internal Cavity Inspection and Cleaning
• Full Functional and Performance Tests
BI-ANNUAL (Every 6 Months)
Calculate the average temperature by adding all five readings
together and dividing by 5. The resultant number should reflect
the set temperature ± 2° F or ± -17°C.
The following maintenance must be performed on all units
every six months in addition to the quarterly maintenance
requirements by a certified service technician.
• Change Cellex® Medium
AS NEEDED
The following maintenance must be performed on all units only
if performance test results indicate replacement is necessary
in addition to the quarterly and bi-annual maintenance
requirements by a certified service technician.
• Intake Filter Replacement
• Distributor Replacement
11
Page 16
MAINTENANCE
Fluido DHT™ Dry Heat Therapy Unit
FLUIDOTHERAPY MAINTENANCE RECORD
UNIT SERIAL NUMBER UNIT MODEL NUMBER DATE PLACED IN SERVICE
DEALER: PHONE: CONTACT
Completing this form:
“DATE”- Date Service is performed
“MAINTENANCE PERFORMED”- Quarterly, Bi-Annual, or Annual
“TECH INITIALS”- Certified Tech’s Initials
12
Page 17
TROUBLESHOOTING
Fluido DHT Model 1480 and 1481 Software Error Messages
Fluido DHT™ Dry Heat Therapy Unit
Error Possible Cause Action to be Taken
Motor/Blower Error
The AIRSPEED shows
four dashes on the
Control Panel LCD.
• The diffuser is clogged. Refer to the Removing and Replacing of the
Diffuser Section on pages 20-21 of this manual
for proper procedure.
the heat cavity.
excess Cellex Medium.
Temperature Sensor Error
Section of the User Manual for proper procedure.
Diffuser Section on pages 20-21 of this manual
for proper procedure.
the heat cavity.
• The intake filter is clogged. Clean the intake filters. Refer to the Maintenance
Section of the User Manual for proper procedure.
• The motor has overheated. Allow the motor to cool before resuming
operation.
• Cellex Medium has entered
CAUTION
Make certain that you do not spill any Cellex into
the heater enclosure cavity.
Use a wet/dry vacuum to remove any exposed,
• The intake filter is clogged.
• The diffuser is clogged.
• Cellex Medium has entered
Clean the intake filters. Refer to the Maintenance
Refer to the Removing and Replacing of the
CAUTION
Make certain that you do not spill any Cellex into
the heater enclosure cavity.
excess Cellex Medium.
• The temperature sensor is Reset the temperature sensor by pressing and
faulty. holding the “Stop” button for 10-15 seconds.
Release the “Stop” button and turn the unit on.
It will be necessary to reset the clock (refer to the
Installation and Setup Section of the User Manual). If error persists, remove and replace the
temperature sensor (P/N 14215). Refer to the Wiring Diagrams on pages 31-32.
Control Panel Lock-Up
manual for proper procedure.
page 18 for proper procedure.
• The parameters are not
correct.
• The Control Panel PC Board
is faulty.
Troubleshooting Section on page 15 of this
Replacing the Control Panel PC Board Section on
Reset the Factory Default Settings. Refer to the
Use a wet/dry vacuum to remove any exposed,
Replace the Control Panel PC Board. Refer to the
13
Page 18
TROUBLESHOOTING
Fluido DHT System Testing
General
The following information is intended to aid in
troubleshooting the major components of the DHT Units
to “Board Level” only. These tests are OEM standard
testing procedures and methods used at the factory before
shipment of any Fluido unit.
Due to the complex nature of the technology utilized by
Chattanooga Group, the recommended troubleshooting
techniques are to determine “Bad Board” and
board replacement only. No board component level
troubleshooting is recommended nor will information
or parts be supplied by Chattanooga Group. Any board
component level troubleshooting performed will be at sole
risk and liability of the Service Technician performing such
troubleshooting techniques.
Once the PC Board has been determined as bad, replace the
board only with Chattanooga Group OEM replacement parts
and hardware.
Special Tools, Fixtures & Materials Required
Certain tests require the use of special tools and fixtures.
These will be listed at the particular test where they are
required. Testing with any other special tool or fixture
other than those stated could give erroneous readings or
test results. Always perform the tests exactly as stated to
ensure accurate results.
Standard test equipment settings will be listed for each test
performed to aid in performing the test to OEM standards
and ensure proper readings.
The troubleshooting and repair of the Fluido DHT units,
should be performed only by authorized technicians trained
and certified by Chattanooga Group.
Fluido DHT™ Dry Heat Therapy Unit
CAUTION
•
The following tool requirements are critical to the
component removal and replacement of the Fluido DHT
unit.
•
All hardware, bolts, nuts and screws used to assemble
the Fluido DHT are SAE Standard. Due to the size of
these components no metric equivalent is available.
Therefore, it will be necessary to obtain the proper size
tools for removal and replacement of certain components.
Required SAE Tools
#1 Phillips Screwdriver
#2 Phillips Screwdriver
5/16, 7/16, 9/16, 1/2, 5/32, and 11/32 Wrenches
1/8 and 5/32 Allen Wrenches (or drill bits)
Utility Knife
Wet/Dry Vacuum
NOTE:
The tool requirements will be listed at the respective
removal and replacement procedures throughout this
manual.
Equipment Required
Digital Multimeter
Dielectric Withstand (Hi-Pot) and ground resistance
tester
Milliohm Meter
Calibrated Thermometer
Calibrated Stop Watch
NOTE:
Adjust Dielectric Withstand tester to indicate fault with
120k Ohm Load across the output when at specified test
voltage.
14
Page 19
TROUBLESHOOTING
Fluido DHT™ Dry Heat Therapy Unit
Visual Inspection
General
Visually inspect the Fluido DHT unit. A visual inspection can,
to an experienced Technician, indicate possible abuse of the
unit and/or internal problems.
Conduct all necessary leakage tests as required per
“Chapter 7 Electrical Equipment” of the 1999, or later,
edition of the NFPA (National Fire Protection
Association) “Health Care Facilities” standards.
Input: 120 VAC~50/60 Hz, 40 Watts
1000V
Resetting Fluido DHT Factory Default Settings
To reset all factory default settings of the unit, turn unit Off.
Depress and hold the “Clear” button on the control panel
and turn unit on simultaneously.
Preheat Test
Tools & Equipment Required
Calibrated ThermometerPreheat Test Procedures
Power Requirements:
Plug the unit power cord
service receptacle conforming to
national and local electrical codes.
to a grounded electrical
the applicable
POWER
SWITCH
WARNING
• Unit failing Dielectric Withstand and/or Leakage Tests could
indicate serious internal system problems.
• Do not place unit back into service.
• Send unit to factory for repair.
• Do not attempt to repair in the field.
Turn unit power switch On.
Start the preheat function by depressing the
“PREHEAT STANDBY” button on the control panel.
After approximately 30 seconds, the Blue LED
should illuminate.
15
Page 20
TROUBLESHOOTING
Fluido DHT™ Dry Heat Therapy Unit
VERIFYING TEMPERATURE CALIBRATION
After each change of Cellex media, perform a temperature
check to ensure that calibration of unit has not changed.
The temperature displayed on the display is an average of
the temperature throughout the media chamber. Because the
media is fluidized by the movement of forced air, there are
zones throughout the chamber. The temperatures measured in
each zone may vary by ± 8° F or ± -13°C.
Once the Cellex has been changed and operated at 50% for
1 hour, turn off the unit and immediately remove the lid to
the unit. With a calibrated thermometer, take a temperature
reading in the following zones – approximately half way down
from the top of the media.
Calculate the average temperature by adding all five readings
together and dividing by 5. The resultant number should reflect
the set temperature ± 2° F or ± -17°C.
Pulse Mode Test
Tools & Equipment Required
Calibrated Stop WatchPulse Mode Test Procedures
Reset the factory defaults. Refer to the section
entitled “Resetting Fluido DHT Factory Default
Settings” on page 15.
Press the “PULSE MODE” button, the Blue LED should
illuminate. Press the “TREATMENT START” button.
Using the calibrated stop watch, time the pulses of
the unit while it is running. Record the readings taken
and verify the settings you entered for the pulse.
The unit should pulse at approximately four seconds
on, after reaching its maximum blower speed, and
four seconds off.
Treatment Time Test
Tools & Equipment Required
Calibrated Stop WatchTreatment Time Test
With the unit On, press and release the “Time”
button. Using the down arrow, adjust the time to
“1:00”.
Press “TREATMENT START” and time with the
Calibrated Stop Watch. Record the reading.
Treatment Time Spec . . . . . . . . . . 1 minute ± 1 second
Air Speed Test
Airspeed Test Procedures
Turn unit On. Press “Treatment Start”. While unit is
running, press the “Air Speed” button. Use the Up
and Down arrows to adjust the air speed to 100%
and back down to 5%. Listen for the increase and
decrease in the blower speed. Look for increased
and decreased fluidization of the medium in the
reservoir.
NOTE:
For proper operation and setting of the unit parameters
outside the Factory Defaults, refer to the User Manual
for the Fluido DHT.
16
Page 21
REMOVAL AND REPLACEMENT
Fluido DHT™ Dry Heat Therapy Unit
Sleeves and Cellex
®
WARNING
•Unplug the unit from the power source before
attempting any removal or replacement procedures to
prevent electrical shock.
•Follow electronic repair protocols for grounding to
prevent damage to the electronic components from
static electricity.
Refer to the Fluido DHT User Manual for the proper
removal and replacement of the following items:
• Unit End Sleeves
• Unit Top Sleeves
• Cellex Medium
• Intake Filters
Control Panel PC Board
Tools & Equipment Required
Small Flat Blade Screwdriver
5/16 Open End Wrench
2. Carefully remove the Bezel from the Control
Panel.
3. Remove the four retaining nuts securing the
Control Panel to its mounting base (two on each
side).
Removing the Control Panel PC Board
1. Remove the three retaining screws from the
Control Panel Bezel.
17
Page 22
REMOVAL AND REPLACEMENT
Fluido DHT™ Dry Heat Therapy Unit
4. Remove the Control Panel from the mounting
base.
5. Remove the five wiring harnesses from the PC
Board.
Replacing the Control Panel PC Board
To replace the control panel PC board, reverse steps 1
through 7 of “Removing the Control Panel PC Board”
starting on page 1 7.
NOTE: Keep in mind the following when replacing the
Control Panel PC Board:
• Attach the connector with the two brown wires to
the heater terminal.
• Attach the connector with one brown wire and one
blue wire to the line terminal.
• Attach the connector with one black wire and one
white wire to the motor terminal.
• Attach the small, three-wire connector wires to
the speed terminal.
• Attach the small, two-wire connector to the
temperature terminal.
18
Page 23
REMOVAL AND REPLACEMENT
Fluido DHT™ Dry Heat Therapy Unit
Fluido Heater
Tools Required
1/8 and 5/32 Allen Wrenches
3/8 Wrench
Wire Cutters
Removing the Fluido Heater
Before attempting to remove the Fluido Heater, you
must complete steps 1 through 3 of “Removing the
Muffler and Muffler Cover” on page 22. Next, complete
steps 1 through 2 of “Removing the Fluido Motor” on
page 17.
1. Remove the gasket from the motor and support
assembly.
NOTE: Once the gasket is removed, always replace with a new gasket.
2. Remove the two bolts that secure the motor
housing mount bracket with the 1/8 Allen Wrench.
Remove the motor housing mount bracket from
unit.
4. Slowly remove the heater enclosure assembly.
3. Remove the two pieces of insulation.
NOTE: Make sure that the insulation is placed back into the unit correctly or the insulation may burn.
5. Using a 3/8 wrench, remove the three (two blue
and one brown) harnesses from the heater
enclosure assembly.
6. Using wire cutters, cut the wire between the
brown and black wire from the heater enclosure
assembly.
19
Page 24
REMOVAL AND REPLACEMENT
Fluido DHT™ Dry Heat Therapy Unit
Replacing the Fluido Heater
To replace the Fluido heater, reverse steps 1 through 6
of “Removing the Fluido Heater” on page 19 (i.e.,
instead of cutting wires, simply solder or reattach them).
Ensure all connections are insulated.
NOTE: Make sure that the insulation is placed back
into the unit correctly or the insulation may burn.
Fluido Diffuser
Tools RequiredWet/Dry Vacuum #2 Phillips Screwdriver
1/8 Allen Wrench
Removing the Fluido Diffuser
1. Remove lid completely from unit.
2. Using the wet/dry vacuum, remove the Cellex from the tub.
3. Remove the seven screws from the perforated
metal at the bottom of the tub.
CAUTION
Make certain to vacuum the screw heads at the base of
the tub. This will ensure that Cellex will not fall in the
standoff’s holes, and you will not round out the heads of
the screws.
5. Remove the rubber tub insert closest to the control
panel. Then, remove the other rubber tub insert.
6. Using the wet/dry vacuum, remove the newly
exposed, excess Cellex.
CAUTION
Make sure to remove the excess Cellex from the heads
of the hex screws to avoid rounding out the heads of the
screws.
7. Remove the ten hex screws from the bottom of the
tub.
4. Using the wet/dry vacuum, remove the newly
exposed, excess Cellex.
NOTE: When vacuuming, be careful not to damage the temperature sensor.
20
Page 25
REMOVAL AND REPLACEMENT
Fluido DHT™ Dry Heat Therapy Unit
9. Remove the perforated metal.
10. Remove the diffuser baffle.
11. Remove the diffuser foam.
CAUTION
Make certain that you do not spill any Cellex into the
heater enclosure cavity.
Fluido Motor
Tools Required
1/8 and 5/32 Allen Wrenches
Removing the Fluido Motor
Before attempting to remove the Fluido motor, you
must complete steps 1 through 3 of “Removing the
Muffler and Muffler Cover” on page 22.
1. Remove the five screws outside of the motor
support plate, being sure to hold the plate while
removing the last screw.
CAUTION
Hold the motor support plate before removing the last
screw. Otherwise, the motor may fall from the housing.
12. Remove bottom (and final) piece of perforated
metal.
13. Using the screwdriver, remove the ground screw.
Replacing the Fluido Diffuser
To replace the Fluido diffuser, reverse steps 1 through
13 of “Removing the Fluido Diffuser” on pages 20-21.
After replacing the Fluido diffuser, trim excess foam
with razor knife while stretching away from the tub.
2. With one hand, hold the motor while using your
other hand to disconnect the motor wiring from the
harness. See the following 2 figures.
21
Page 26
REMOVAL AND REPLACEMENT
CAUTION
Be careful not to damage the metal shaft on the end of the
motor. See the following figure.
Replacing the Fluido Motor
To replace the Fluido Motor, reverse steps 1 and 2 of
“Removing the Fluido Motor” on page 21.
Fluido DHT™ Dry Heat Therapy Unit
2. Slide the muffler cover down on the shaft.
3. Remove the foam insert from inside the muffler
cover.
NOTE: Notice the hole in the foam insert. You will need to line up this hole with the Morris taper
on the shaft when you re-assemble the muffler.
Muffler and Muffler Cover
Tools Required
1/8 Allen Wrench
Removing the Muffler Cover and Muffler
1. Remove the 14 screws that secure the muffler
cover.
Replacing the Muffler Cover and Muffler
To replace the muffler and muffler cover, reverse
steps 1-3 of “Removing the Muffler Cover” on this
page.
NOTE: After re-installing the muffler, remove the
air filters to ensure that the muffler is not obstructing
the air intakes.
22
Page 27
REMOVAL AND REPLACEMENT
Casters
Materials and Tools Required
9/16 Wrench(4) Caster - 14135
Base Frame- 14104
(4) 3/8-16 Black Acorn Nuts - 14146
Removing the Casters
1. Make sure casters are locked by pressing the
metal flap until the wheel does not move. This will
prevent the shaft of the caster from moving while
you are attempting to remove it.
Fluido DHT™ Dry Heat Therapy Unit
2. Remove each of the four 3/8-16 black acorn nuts.
3. Lift the base to remove each of the shafts of the
four casters from the unit.
Replacing the Casters
1. Make sure caster is locked before attempting to
install.
2. Place one caster on each corner of the base, and
attach with four 3/8-16 black acorn nuts.
23
Page 28
REPLACEMENT PARTS
Fluido DHT™ Dry Heat Therapy Unit
Part No. Description
14105 BASE ASSEMBLY FLUIDOTHERAPY
14111 PATIENT TUB LID FLUIDO DHT
14113 LID LATCH FLUIDO DHT
14116 SCREW #2 X 3/8 PAN HD. PHIL.
14117 DISTRIBUTOR PLATE FLUIDO DHT
14118 PLEXIGLASS TRIM FLUIDOTHERAPY
14119 FOAM DISTRIBUTOR PLATE FLUIDOTHERAPY
14120 TUB DRAIN PLUG FLUIDOTHERAPY
14122 6-32 X 1/2 OD X1 LG CERAMIC STD. OFF
14123 6-32 X 1/2 OD X 1-1/2 LG CERAMIC STD.
14124 GASKET TUB LID FLUIDOTHERAPY
14125 MANUAL USER FLUIDOTHERAPY
14128 MANUAL SERVICE FLUIDOTHERAPY
14130 SLEEVE ASSEMBLY FLUIDO DHT
14135 CASTER TENTE 006874 FLUIDO DHT
14136 FOAM MUFFLER FLUIDO DHT
14137 MUFFLER COVER FLUIDO DHT
14139 BOTTOM W/ MOTOR HOUSING FLUIDO DHT
14146 ACORN NUT FLUIDO DHT
14148 SCREW 10-32 x 3/4 BUTT. HD. SOC
14150 SCREW 10-32 X 1 BUTT HD SOC
14151 SWITCH PLATE GASKET FLUIDO DHT
14160 SCREW 10-32 X 1/2 BUTT HD SOC
14161 HEX STANDOFF 6-32 X .875 ALUM.
14162 MULTI STACK ADAPTER
14166 USER INTERFACE COVER FLUIDO DHT
14168 PLASTIC COVER FLAT FLUIDO DHT
14169 SLEEVE RETAINING CLIP FLUIDO DHT
14170 POWER CORD FLUIDO DHT
14171 SWITCH PLATE FLUIDO DHT
14174 GASKET MOTOR MOUNTING PLATE
14175 ELBOW SLEEVE FLANGE FLUIDO DHT
14188 SPONGE GASKET 3/4" X1/4" ADHES.
14189 HARNESS CONTROL POWER SWITCH
14202 HARNESS CONTROL TO HEATER SUB
14203 HARNESS HEATER TO INDICATOR
14204 HARNESS FUSE TO RECEPTACLE
14206 HARNESS RECEPTACLE TO POWER SWITCH
Part No. Description
14207 HARNESS O.T. TO STAT H.E.
14208 HARNESS MOTOR SPEED CONTROL
14209 LINE FILTER BRACKET
14211 HARNESS CONTROL INTERCONNECT
14213 HARNESS GROUND 7"
14214 HARNESS GROUND 24"
14215 HARNESS TEMP SENSOR
14219 1/4-20 X 5/8 SANDWICH MOUNT
14220 10 x 24 INSULATION MOTOR HOUSING
14221 5' X 4" INSULATION MOTOR HOUSING
14222 5" X 5" INSULATION MOTOR HOUSING
14223 9" X 9" INSULATION MOTOR HOUSING
14225 HARNESS FUSE TO POWER SWITCH 230V
14226 HARNESS POWER SWITCH TO LINE FILTER
14227 HARNESS O.T. STAT TO H.E.
14228 HARNESS LINE FILTER GROUND TO CHASSIS
14229 ELBOW SLEEVE ASSEMBLY
14230 INLET FILTER 19155K22
14231 FILTER NYLON MESH 3-5/8" SQ.
14232 GLIDE ETS 32 X 50 D80153
14233 USER INTERFACE BRACKET PAINTED
14234 MOTOR HOUSING SUPPORT PAINTED
14235 FOAM INSERT
14236 MOTOR SUPPORT PLATE PAINTED
14237 SET SCREW 6-32 X 3/4 91375A151
14238 SPACER NYLON 1/2 ODX1/4 ID x 5/8 94639A143
14239 SCREW #1 X 3/8" 90253A033
14240 POP RIVET 1/8 X 5/8 ALUMINUM 97447A135
14241 CAPTIVE STUD 6/32 X 3/8 93580A016
14244 RIVNUT 10-32 S10P175
14245 STRAIN RELIEF PLATE FLUIDO DHT
14247 230V. MOTOR HOUSING ASSY.
14251 HYDRAULIC LIFT (CHROME) 4500
14253 FUSE HOLDER SCHURTER
14254 INLET AC FAST CONNECT
14255 10 AMP. 5MM X 20MM (FUSE)220V 3 AG SLOW BLOW
14259 2-56 X 1/4" PAN HD.PHIL PLT.
14260 MOTOR ASSY. 120V
24
Page 29
REPLACEMENT PARTS
Part No. Description
14261 MOTOR ASSY. 230V
14262 HARNESS RECP. TO SWITCH PLATE
14263 HARNESS RECP. GROUND TO GROUND
14268 TOP DIFFUSER FLUIDO DHT
14269 DIFFUSER AIR BAFFLE FLUIDO DHT/24 GA. CARBON
14270 DIFFUSER AIR BAFFLE FLUIDO DHT/ADHESIVE BACKED FOIL
14272 HOUSING BRACE SPACER
14274 HOUSING BRACE (PAINTED #)
14275 FOAM INSERT
14276 POWER SWITCH
14277 ADAPTER 1/8 NPT 15090-1
14279 HEATER INDICATOR LIGHT BLUE 230V
14285 POP RIVET 1/8"
14286 FICHE PAPER 10 MIL THK.
14287 FITTING ADAPTER 1/8 TO 10-32 15090-1
14288 HEATER INDICATOR LIGHT BLUE 120V
14291 THERMAL OVER TEMP SWITCH
14299 CONTROL PNL. ASSY 120V
14300 MOTOR ASSY 230V
14301 MOTOR ASSY 120V
14302 MOTOR HOUSING WARNING LABEL
14304 1" ALMN. STANDOFF
31498 900 WATT HEATER FINNED STRIP 120V
31506 900 WATT HEATER FINNED STRIP 230V
14309 230V HEATER ASSEMBLY
14308 120V HEATER ASSEMBLY
Fluido DHT™ Dry Heat Therapy Unit
25
Page 30
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Base Assembly-
120V and 230V
26
Apply small bead of
clear silicone around
perimeter of base
Page 31
DIAGRAMS
Control Panel Assembly120V and 230V
Fluido DHT™ Dry Heat Therapy Unit
Note:
(1) Set selector switch for
desired voltage- 115V or 230 V
(2) For 115V- use Fuse #71552
For 230V- use Fuse #20656
27
Page 32
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Apply small bead of
clear silicone around
perimeter of window
Lid Assembly120V and 230V
secure with screws
Slip outer edge of sleeve
assembly under flange and
28
Page 33
DIAGRAMS
Switch Plate Assembly 120V
Fluido DHT™ Dry Heat Therapy Unit
Apply silicone sealant
around components
after assembly
29
Page 34
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Switch Plate Assembly-
230V
Apply silicone sealant
around components
after assembly
30
Page 35
DIAGRAMS
Wiring Diagram120V
Fluido DHT™ Dry Heat Therapy Unit
screw interface
Connect to user
innerface mounting
31
metal plate
To perforated
Page 36
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Wiring Diagram-
230V
screw interface
Connect to user
innerface mounting
32
metal plate
To perforated
Page 37
DIAGRAMS
Heater Enclosure Assembly120V
Fluido DHT™ Dry Heat Therapy Unit
33
Page 38
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Heater Enclosure Assembly-
230V
34
Page 39
DIAGRAMS
Motor & Support Assembly120V
Fluido DHT™ Dry Heat Therapy Unit
Apply light grease to
outer side of gasket to
ease assembly
35
Page 40
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Motor & Support Assembly-
230V
Apply light grease to
outer side of gasket to
ease assembly
36
Page 41
DIAGRAMS
Motor Housing Assembly120V
Apply silicone sealant
around components
after assembly
Fluido DHT™ Dry Heat Therapy Unit
37
Page 42
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Motor Housing Assembly-
230V
Apply silicone sealant
around components
after assembly
38
Page 43
DIAGRAMS
Motor Housing Assembly120V
Fluido DHT™ Dry Heat Therapy Unit
39
Page 44
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Motor Housing Assembly-
230V
40
Page 45
DIAGRAMS
Motor Housing Assembly120V
Fluido DHT™ Dry Heat Therapy Unit
Break tab off
ground lug side
Fit P/N 14221
snug against
inside wall and
tuck under
flange
Apply thin film
of clear silicone
between P/N 14139
and P/N 14245
Fit P/N 14221 snug against
inside wall and tuck
under flange
41
Page 46
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Motor Housing Assembly-
230V
Break tab off
ground lug side
Fit P/N 14221
snug against
inside wall and
tuck under
flange
Apply thin film
of clear silicone
between P/N 14139
and P/N 14245
Fit P/N 14221 snug against
inside wall and tuck
under flange
42
Page 47
DIAGRAMS
Final Assembly120V
Fluido DHT™ Dry Heat Therapy Unit
Trim all three
edges 1” x 60”
before assembly
43
Page 48
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Final Assembly-
230V
Trim all three
edges 1” x 60”
before assembly
44
Page 49
DIAGRAMS
Final Assembly120V
Fluido DHT™ Dry Heat Therapy Unit
45
Trim foam along outside
edge of housing after
assembly
Page 50
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Final Assembly-
230V
46
Trim foam along outside
edge of housing after
assembly
Page 51
DIAGRAMS
Final Assembly120V
Fluido DHT™ Dry Heat Therapy Unit
47
Install as shown
4 places
Install sleeve retainer
clip as shown 4 places
Page 52
DIAGRAMS
Fluido DHT™ Dry Heat Therapy Unit
Final Assembly-
230V
48
Install as shown
4 places
Install sleeve retainer
clip as shown 4 places
Page 53
WARRANTY
Chattanooga Group ("Company") warrants that the Fluido DHT units ("Product") are free of defects in material and workmanship. This warranty shall
remain in effect for two years (24 months) from the date of original consumer purchase. If this Product fails to function during the two year warranty
period due to a defect in material or workmanship, Company or the selling dealer will repair or replace this Product without charge within a period of
thirty (30) days from the date on which the Product is returned to the Company or the dealer.
All repairs to the Product must be performed by a service center authorized by the Company. Any modifications or repairs performed by unauthorized
centers or groups will void this warranty.
The warranty period for replaceable intake filter(s) is 90 days.
The warranty period for sleeves is one year (12 months).
To participate in warranty coverage, this Product's warranty registration card (included with Product) must be filled out and returned to the Company by
the original owner within ten (10) business days of purchase.
This Warranty Does Not Cover:
Replacement parts or labor furnished by anyone other than the Company, the selling dealer or a certified Company service technician.
Defects or damage caused by labor furnished by someone other than Company, the selling dealer or a certified Company service technician.
Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonable and required
maintenance or any use that is inconsistent with the Product User's Manual.
COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.
To Obtain Service From Company or the selling dealer under this warranty:
1. A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to the Company
should be sent to:
4717 Adams Road
P.O. Box 489
Hixson, TN 37343 US
Telephone: (423) 870-2281
Tel (International): +1 (423) 870-7200
Facsimile: (423) 870-7200
Fax (International): +1 (423) 870-2046
and
2. The Product must be returned to the Company or the selling dealer by the owner.
Fluido DHT™ Dry Heat Therapy Unit
This warranty gives you specific legal rights and you may also have other rights which vary from state to state.
The Company does not authorize any person or representative to create for it any other obligation or liability in connection with the sale of the Product.
Any representation or agreement not contained in the warranty shall be void and of no effect.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
When the Fluido DHT unit requires service, or preventive maintenance, contact the selling dealer or Chattanooga Group Service
Department.
All units returned to the factory for service must include the following;
WARRANTY REPAIR/OUT OF WARRANTY REPAIR
1. Written statement containing the following information;
·
RGA Number- Obtain from Factory
·
Unit Model Number
·
Unit Serial Number
·
Contact person with Phone and Fax Numbers
·
Billing Address (for Out of Warranty Repair)
·
Shipping Address (Where to Ship Unit after Repair)
·
Detailed Description of Problem or Symptoms
2. Copy of original invoice issued at purchase of the unit.
3. Ship unit to Factory in the original container with all
accessories and information as required in item 1 above to:
Chattanooga Group
4717 Adams Road
Hixson, TN 37343