Chattanooga DHT 1480, DHT 1481 User manual

ISO 13485 CERTIFIED
SERVICE MANUAL
Models: 1480
(Serial Numbers- 1000 and above)
1481
(Serial Numbers- 1000 and above)
Fluido DHT™Dry Heat Therapy Unit
TABLE OF CONTENTS
©2003 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial or layout composition of this publication without express written consent from Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated and prepared for print by Chattanooga Group of Encore Medical, L.P.
Foreword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Precautionary Symbol Definitions . . . . . . . . . . . .2
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . .3-4
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . .5
Nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . .9-12
Fluido Software Error Messages . . . . . . . . . . . .9
Motor/Blower Error . . . . . . . . . . . . . . . . . . . . . . .9
Temperature Sensor Error . . . . . . . . . . . . . . . . . .9
Fluido System Testing . . . . . . . . . . . . . . . . . .10-12
Visual Inspection . . . . . . . . . . . . . . . . . . . . .11
Ground Resistance Test . . . . . . . . . . . . . . .11
Leakage Tests . . . . . . . . . . . . . . . . . . . . . . . .11
Resetting Fluido Factory Defaults . . . . . . .11
Preheat Test . . . . . . . . . . . . . . . . . . . . . . . . .11
Pulse Mode Test . . . . . . . . . . . . . . . . . . . . . .12
Treatment Time Test . . . . . . . . . . . . . . . . . . .12
Air Speed Test . . . . . . . . . . . . . . . . . . . . . . . .12
Control Panel Lock-Up . . . . . . . . . . . . . . . . .12
Removal & Replacement Procedures . . . . . .13-20
Sleeves and Cellex . . . . . . . . . . . . . . . . . . . . . . .13
Control Panel PC Board . . . . . . . . . . . . . . . .13-14
Fluido Heater . . . . . . . . . . . . . . . . . . . . . . . . .15-18
Fluido Diffuser . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Fluido Motor . . . . . . . . . . . . . . . . . . . . . . . . . .18-19
Muffler and Muffler Cover . . . . . . . . . . . . . . . . .19
Casters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Replacement Parts . . . . . . . . . . . . . . . . . . . . . .21-22
Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23-45
Base Assembly - 120v and 230v . . . . . . . . . . . .23
Control Panel Assembly - 120v and 230 v . . . .24
Lid Assembly - 120v and 230v . . . . . . . . . . . . . .25
Switch Plate Assembly - 120v . . . . . . . . . . . . . .26
Switch Plate Assembly - 230v . . . . . . . . . . . . . .27
Wiring Diagram - 120v . . . . . . . . . . . . . . . . . . . .28
Wiring Diagram - 230v . . . . . . . . . . . . . . . . . . . .29
Heater Enclosure Assembly - 120v . . . . . . . . .30
Heater Enclosure Assembly - 230v . . . . . . . . .31
Motor & Support Assembly - 120v . . . . . . . . . .32
Motor & Support Assembly - 230v . . . . . . . . . .33
Motor Housing Assembly (1) - 120v . . . . . . . . .34
Motor Housing Assembly (1) - 230v . . . . . . . . .35
Motor Housing Assembly (2) - 120v . . . . . . . . .36
Motor Housing Assembly (2) - 230v . . . . . . . . .37
Motor Housing Assembly (3) - 120v . . . . . . . . .38
Motor Housing Assembly (3) - 230v . . . . . . . . .39
Final Assembly (1) - 120v . . . . . . . . . . . . . . . . . .40
Final Assembly (1) - 230v . . . . . . . . . . . . . . . . . .41
Final Assembly (2) - 120v . . . . . . . . . . . . . . . . . .42
Final Assembly (2) - 230v . . . . . . . . . . . . . . . . . .43
Final Assembly (3) - 120v . . . . . . . . . . . . . . . . . .44
Final Assembly (3) - 230v . . . . . . . . . . . . . . . . . .45
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Warranty Repair/Out of Warranty Repair
. . . . . .46
Fluido DHT™Dry Heat Therapy Unit
FOREWORD
1
Read, understand and follow the Safety Precautions and information contained in this manual.
This manual contains the necessary safety, and field service information for those Field Service Technicians, approved by Chattanooga Group, to perform field service on the Fluido DHT Models 1480 and 1481 units.
At the time of publication, the information contained herein was current and up to date. However, due to continual technological improvements and increased clinical knowledge in the field of fluidotherapy, as well as Chattanooga Group’s policy of continual improvement, Chattanooga Group reserves the right to make periodic changes and improvements to their equipment and documentation without any obligation on the part of Chattanooga Group.
It is the sole responsibility for field technicians to stay informed and trained in the latest technology utilized in the Fluido DHT Models 1480 and 1481 units by Chattanooga Group. From time to time, as significant improvements are incorporated, Service Bulletins will be produced and made available on our web site (www.chattgroup.com) in lieu of reprinting a complete manual prematurely. These Service Bulletins will provide updated service information and technological improvements to the Fluido DHT Models 1480 and 1481 for use by approved service technicians.
“Approved Service Technician” Definitions;
1. Level I- Those Field Service Technicians that have successfully completed the minimal training required by Chattanooga
Group, in basic service techniques.
2. Level II- Those Field Service Technicians that have successfully completed Level I Training as well as Level II Training as required to perform specific troubleshooting and repair techniques and procedures.
3. Level III- Those Field Service Technicians that have successfully completed Levels I & II Training as well as Level III Advanced Training as required to perform all necessary Troubleshooting and Repair techniques. The Technician having successfully completed the three levels of training and coupled with experience should have the ability to train other technicians in Level I and Level II Training with the necessary Training Materials from Chattanooga Group.
4. Temporary- Chattanooga Group, at its discretion and based on known experience of the technician, may grant a
“Temporary Approval” to a field technician for particular troubleshooting and repair of a specific unit requiring immediate attention. This “Temporary Approval” in no fashion acknowledges the training level of a technician as defined above. This “Temporary Approval” is utilized only in unique situations for a specific unit for a specific service technique only and is documented as such.
Due to the complex nature of the technology utilized by Chattanooga Group, the recommended troubleshooting techniques for PC Boards are to determine “Bad Board” and PC Board replacement only. No board component level troubleshooting is recommended nor will information or parts be supplied by Chattanooga Group. Any PC Board component level troubleshooting performed will be at sole risk and liability of the Service Technician performing such troubleshooting techniques.
This equipment is to be sold and used only under the prescription and supervision of a licensed medical practitioner.
This equipment is to be serviced only by an “Approved Service Technician”.
For Additional Service Contact:
Chattanooga Group
DHT Support Department
Toll Free: 1-866-864-0598
Outside USA: +1-423-870-7200
Fluido DHT™Dry Heat Therapy Unit
SYMBOL DEFINITION
2
The following symbols are located on the Chattanooga Group Fluido DHT Dry Heat Therapy Unit Control Panel. Understand the meaning of each symbol before attempting any operation or use of the unit.
This symbol indicates therapy session Time adjustment.
This symbol indicates therapy session Temperature adjustment.
This symbol indicates therapy session Air Speed adjustment.
This symbol indicates Upward adjustment in function parameters.
This symbol indicates Downward adjustment in function parameters.
This symbol with the illuminated Blue indicator light indicates the Intake Filter requires changing.
This symbol indicates Start therapy session.
This symbol indicates Stop therapy session.
This symbol indicates Pulse Mode for pulse therapy sessions.
This symbol indicates Preheat Standby function.
This symbol with the illuminated Blue indicator light indicates the Cellex®media requires changing.
Fluido DHT™Dry Heat Therapy Unit
3
SAFETY PRECAUTIONS
Precautionary Symbol Definitions
The precautionary instructions found throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating or servicing this equipment. The definitions of these symbols are as follows:
CAUTION
Text with a “CAUTION” indicator will explain possible safety infractions that could have the potential to cause minor to moderate injury or damage to equipment.
WARNING
Text with a “WARNING” indicator will explain possible safety infractions that will potentially cause serious injury and equipment damage.
DANGER
Text with a “DANGER” indicator will explain possible safety infractions that are imminently hazardous situations that would result in death or serious injury.
EXPLOSION HAZARD
Do not use this equipment in the presence of flammable anesthetics. This symbol is also prominently displayed on the serial number plate of the unit.
NOTE:
Throughout this manual “NOTE” may be found. The Notes are helpful information to aid in the particular area or function being described.
Safety Precautions
Read, understand and follow all safety precautions found in this manual. The following are general safety
precautions that must be read and understood before attempting any service techniques on these units. Throughout this manual, specific safety precautions will be found. Read, understand and follow all safety precautions.
!
CAUTION
WARNING
!
!
DANGER
Read, understand and practice the precautionary and operating instructions found in this manual. Know the limitations and hazards associated with using any electrical device. Observe the precautionary and operational decals placed on the unit.
DO NOT operate the unit when connected to any unit other than Chattanooga Group devices.
Refill unit daily to proper fill level with Chattanooga Group Cellex®Dry Heat Medium.
Change Cellex Dry Heat Medium every six (6) months.
Use only Cellex Dry Heat Medium in the Fluidotherapy units.
Clean Inlet Filter(s) daily before unit startup.
Use only fingers to operate button controls on the control panel(s).
Use of sharp objects such as pencils or pens will result in damage to the unit.
Turn unit to the “Standby” mode before positioning a patient or removing a patient from the unit.
After or between treatments, do not immediately unplug or turn the power off from the unit. Allow the unit to process through the “Cool Down” cycle. The unit cycles into standby mode after treatment time has elapsed. Standby can be disabled by pressing the preheat/standby button. The unit goes to a 3 minute cool down mode after standby is turned off. Turning the power off before cool down completes is potentially hazardous to the equipment, and could lead to failure of the unit. It is recommended that power be supplied to the unit at all times. Keep in mind that the recommended treatment air speed is 50%
Secure all entry ports before turning the unit ON.
Check unit temperature before treating patient to ensure correct
temperature.
Place the patient in a comfortable position allowing for correct placement of the limb being treated.
Proper storage and transport temperatures for the Fluido DHT units are 40°F - 158°F (4.5 °C - 70°C). Relative Humidity 85%.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. Harmful interference to other devices can be determined by turning this equipment on and off. Try to correct the interference using one or more of the following: Reorient or relocate the receiving device, increase the separation between the equipment, connect the equipment to an outlet on a different circuit from that to which the other device(s) are connected and/or consult the factory field service technician for help.
!
CAUTION
Fluido DHT™Dry Heat Therapy Unit
4
SAFETY PRECAUTIONS
Safety Precautions Continued
Federal law restricts this device to sale by, or on the order of, a physician or licensed practitioner. This device should be used only under the continued supervision of a physician or licensed practitioner.
For continued protection against fire hazard, replace fuses only with ones of the correct type and rating.
Make certain the unit is electrically grounded by connecting
only to a grounded electrical service
receptacle conforming to
the applicable national and
local electrical codes.
This device should be kept away from children.
Care must be taken when operating this equipment
around other equipment. Potential electromagnetic or other interference could occur to this or to the other equipment. Try to minimize this interference by not using other equipment in conjunction with it.
Before administering any treatment to a patient you should become acquainted
with the operating procedures for each mode of treatment available, as well as the indications, contraindications, warnings and precautions. Consult other
resources for additional information regarding the application of Dry Heat Therapy.
To prevent electrical shock, disconnect the unit from the power source before attempting any maintenance procedures.
Use only Cellex
®
processed dry heat medium in the unit
to prevent excessive dusting.
Adequate precautions should be taken when treating individuals with suspected or diagnosed medical conditions or diseases such as heart problems, epilepsy, diabetes, etc.
Prior to treatment, consult a medical professional familiar with the precautionary measures to be taken for patients that may experience allergic reactions to dust and pollen.
Properly dispose of used Cellex according to National and local laws, rules and regulations.
WARNING
!
Explosion hazard if used in the presence of flammable anesthetics. The warning symbol for this hazard is prominently displayed on the serial number plate.
Perform all Required Maintenance as described in this and the User Manual. Strict adherence to the Required Maintenance for the Fluido DHT units is mandatory. Failure to perform the Required Maintenance could result in the Cellex medium entering the heat chamber of the unit(s) and cause severe injury to patients as well as smoke damage to the facility and the Fluido DHT unit(s).
Make certain the unit is unplugged from the power source before attempting any removal and replacement procedures on the unit.
DANGER
Fluido DHT™Dry Heat Therapy Unit
5
THEORY of OPERATION
The Fluido DHT utilizes ambient air and pressurizes it via an internal blower. The pressurized air is directed across a heating element bringing the air to the desired treatment temperature. The heated, pressurized air is then diffused across and through a baffle to fluidize and heat the Cellex media in a patient treatment reservoir. All treatment parameters; air speed, temperature, treatment time and the unit preheat settings are programmed by the operator with the touch panel user interface. The Fluido DHT unit base incorporates four locking casters and a manually operated hydraulic lift to adjust height and rotation of the treatment reservoir for patient comfort.
Fluido DHT™Dry Heat Therapy Unit
6
NOMENCLATURE
Fluido DHT
The nomenclature graphics below, Figure 3.1, indicate the general locations of the major components of the Fluido DHT unit.
Know the components and their functions before performing any operation of or service to the Fluido DHT Model 1480 and 1481 unit.
11
22
33
44
55
66
99
1100
1111
88
77
1122
1133
1144
1155
1. RESERVOIR LID - Allows access to the treatment cavity for adding Cellex
®
medium.
2. LATCHES - Secures Reservoir Lid
3. SIDE ACCESS PORTS - Four available Side Access
Ports. Two on each end of the unit.
4. RESERVOIR/TREATMENT CAVITY - Patient treatment cavity and Cellex Reservoir.
5. TREATMENT LIMB SLEEVES - Replaceable and launderable patient limb treatment sleeves.
6. HEAT CHAMBER - Houses Heating element.
7. INTAKE FILTER - Air intake filter. Requires periodic
replacement.
8. LOCKING CASTERS - Four Locking Casters for securing the unit in place for treatment.
9. ELEVATION ADJUSTMENT CYLINDER - Adjusts and maintains reservoir to desired height for patient comfort.
10. UNIT BASE - Rigid unit base for ease in transporting unit to different locations for treatment.
11. HEIGHT ADJUSTMENT PEDAL - Used to raise, rotate, release and lock the cylinder for height adjustment of the unit.
12. BLOWER HOUSING - Houses Blower Motor.
13. CONTROL PANEL - Operator Controls. See Page 7 for
detail description of each control.
14. TREATMENT CAVITY VIEWING WINDOW - View treatment area during treatment.
15. TOP ACCESS PORT - Top Treatment access port with sleeve.
Fluido DHT™Dry Heat Therapy Unit
7
NOMENCLATURE
Fluido Control Panel
The Control Panel nomenclature graphics below, Figure
3.2, indicate the location and functions of the Fluido DHT Model 1480 and 1481 control panel.
Know the components and their functions before performing any operation of or service to the Fluido DHT Model 1480 and 1481 unit.
11
22
33
44
55
66
77
88
99
1100
1111
1122
1. PANEL DISPLAY- Displays settings and operational parameters.
2. TEMPERATURE- Use in conjunction with the Up and Down arrows to set operation Temperature. Temperature can be adjusted in 1° increments. The available temperature range is 88° F to 130° F (31° C to 54° C).
3. AIR SPEED- Use in conjunction with the Up and Down arrows to adjust air speed for fluidization of the Cellex
®
Medium. Available speeds range from 5% to 100% (in 5% increments).
4. PULSE MODE- Turn Pulse Mode On and Off as well as adjust pulse time. Available Pulse Time is from 1 second On/1 Second Off to 6 Seconds On/6 Seconds Off.
5. UP and DOWN ARROWS- Use in conjunction with other mode or function buttons to set desired parameters.
6. PREHEAT STANDBY- Turn the Preheat Standby Mode On and Off as well as set the parameters desired for the Preheat Standby Mode to automatically start.
7. TREATMENT STOP- Press to stop treatment.
8. TREATMENT START- Press to start treatment.
9. CLEAR BUTTON- Used to turn off maintenance
indicators after maintenance has been properly performed.
10. MEDIA- Indicator will light when it is time to change the Cellex
®
Medium.
11. FILTER- Indicator will light when it is time to change the required filters.
12. TIME- Use in conjunction with the Up and Down arrow buttons to set treatment time. Available time is 1 to 99 minutes, in one minute increments, or Continuous.
Fluido DHT™Dry Heat Therapy Unit
8
SPECIFICATIONS
DDeepptthh
OOvveerraallll HHeeiigghhtt
WWiiddtthh
PPaattiieenntt HHeeiigghhtt
MMOODDEE OOFF OOPPEERRAATTIIOONN
. . . . . . . . . . . . . . . . . . . . . . Continuous
OOPPEERRAATTIIOONNAALL FFUUNNCCTTIIOONNSS
VVaarriiaabbllee AAddjjuussttmmeennttss
. . . . . . . . Time, Temp and Air Speed
PPuullssee MMooddee
. . . . . . . . . . . . . . . . . . . . OFF to 6 Sec ON/OFF
AAvvaaiillaabbllee LLaanngguuaaggeess
. . . . . . . . English, Spanish and French
TTRREEAATTMMEENNTT TTIIMMEE
. . . . . . . . 1 to 99 minutes and Continuous
**OOPPEERRAATTIINNGG TTEEMMPPEERRAATTUURREE
. 88 °F (31 °C) to 130 °F (54 °C)
AAIIRR SSPPEEEEDD
. . . . . . . . . . . . . . . . 5% to 100% (5% increments)
PPRREEHHEEAATT TTIIMMEERR
. . . . . Per default setting with 50% Air Flow
during 30 min. preheat then 5% airflow for standby.
CCeelllleexx
®®
MMEEDDIIUUMM CCAAPPAACCIITTYY
. . . . . . . . . . . 20-25 lbs (9-11 kg)
IINNPPUUTT PPOOWWEERR
. . . . . . . . . . . . . . . . . . . . . 120V, 50/60 Hz, 12A
FFUUSSEE RRAATTIINNGG
. . . . . . . . . . . . . . . . 10 A Time Delay (Slo-Blow)
PPHHYYSSIICCAALL DDIIMMEENNSSIIOONNSS
UUnniitt DDeepptthh
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31” (79 cm)
UUnniitt WWiiddtthh
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 29.5” (75 cm)
OOvveerraallll HHeeiigghhtt
. . . . . . . . . . . . . 41” - 49” (104 cm - 124 cm)
PPaattiieenntt HHeeiigghhtt
. . . . . . . . . . . . . . . 31” - 39” (79 cm - 99 cm)
RReesseerrvvooiirr SSwwiivveell
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 360°
WWeeiigghhtt
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145 lbs (66 kg)
SShhiippppiinngg WWeeiigghhtt
(Including Stool) . . . . . . . . 210 lbs (95 kg)
EElleeccttrriiccaall::
Class I
TTyyppee BB EEqquuiippmmeenntt
AAtttteennttiioonn,, ccoonnssuulltt aaccccoommppaannyyiinngg ddooccuummeennttaattiioonn.. OOrrddiinnaarryy eeqquuiippmmeenntt:: NNoott ddeessiiggnneedd ttoo pprreevveenntt iinnggrreessss ooff wwaatteerr.. *Dependent upon airspeed selected
Model 1480
MMOODDEE OOFF OOPPEERRAATTIIOONN
. . . . . . . . . . . . . . . . . . . . . . Continuous
OOPPEERRAATTIIOONNAALL FFUUNNCCTTIIOONNSS
VVaarriiaabbllee AAddjjuussttmmeennttss
. . . . . . . . . Time, Temp and Air Speed
PPuullssee MMooddee
. . . . . . . . . . . . . . . . . . . . . OFF to 6 Sec ON/OFF
AAvvaaiillaabbllee LLaanngguuaaggeess
. 9 Languages on Board (See Page 10)
TTRREEAATTMMEENNTT TTIIMMEE
. . . . . . . . 1 to 99 minutes and Continuous
**OOPPEERRAATTIINNGG TTEEMMPPEERRAATTUURREE
. 31 °C (88 °F) to 54 °C (130 °F)
AAIIRR SSPPEEEEDD
. . . . . . . . . . . . . . . . . 5% to 100% (5% increments)
PPRREEHHEEAATT TTIIMMEERR
. . . . . Per default setting with 50% Air Flow
during 30 min. preheat then 5% airflow for standby.
CCeelllleexx
®®
MMEEDDIIUUMM CCAAPPAACCIITTYY
. . . . . . . . . . . 9-11 kg (20-25 lbs)
IINNPPUUTT PPOOWWEERR
. . . . . . . . . . . . . . . . . . . 230 V~, 50/60 Hz, 10A
FFUUSSEE RRAATTIINNGG
. . . . . . . . . . . . . . . . 10 A Time Delay (Slo-Blow)
PPHHYYSSIICCAALL DDIIMMEENNSSIIOONNSS
UUnniitt DDeepptthh
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 cm (31”)
UUnniitt WWiiddtthh
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 cm (29.5”)
OOvveerraallll HHeeiigghhtt
. . . . . . . . . . . . . 104 cm - 124 cm (41” - 49”)
PPaattiieenntt HHeeiigghhtt
. . . . . . . . . . . . . . . 79 cm - 99 cm (31” - 39”)
RReesseerrvvooiirr SSwwiivveell
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 360°
WWeeiigghhtt
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 kg (145 lbs)
SShhiippppiinngg WWeeiigghhtt
(Including Stool) . . . . . . . . 95 kg (210 lbs)
EElleeccttrriiccaall::
Class I
TTyyppee BB EEqquuiippmmeenntt
AAtttteennttiioonn,, ccoonnssuulltt aaccccoommppaannyyiinngg ddooccuummeennttaattiioonn.. OOrrddiinnaarryy eeqquuiippmmeenntt:: NNoott ddeessiiggnneedd ttoo pprreevveenntt iinnggrreessss ooff wwaatteerr.. *Dependent upon airspeed selected
Model 1481
Fluido DHT™Dry Heat Therapy Unit
9
TROUBLESHOOTING
Fluido DHT Model 1480 and 1481 Software Error Messages
A. The information provided in this section is
intended to aid in troubleshooting the Fluido DHT Model 1480 and 1481 units. There are only two
software error messages that could indicate a “Bad Board”. Therefore, should tests contained in this section indicate the pc board is bad the board must be replaced as an assembly. No component level troubleshooting information is or will be provided by Chattanooga Group for field troubleshooting of pc board components.
B. Once the PCB has been determined as bad,
replace the suspected board as described in the Removal & Replacement section of this manual.
Motor/Blower Error
A. Should the software detect a problem within the
operation of the Blower/Motor, the display will show four dashes beneath “AIRSPEED” on the control panel LCD.
B. If this indicator is visible, there could be several
possible reasons:
The intake filters are clogged and require cleaning.
The motor has overheated and requires a cooling period before operation is resumed.
The Distributor is clogged and requires changing.
The Cellex Medium has entered the heat cavity of the unit and is restricting the air flow. Should this indicator be visible, remove the unit from service immediately. Perform any
and all maintenance and repair necessary before the unit is placed back into service. Refer to the appropriate sections of this manual for proper removal and replacement procedures.
Temperature Sensor Error
A. Should the Temperature Sensor experience
problems the software will display four dashes beneath “Temp” on the LCD.
Should this occur the system may “Lock Up,” preventing any further operation of the unit. This error could be caused by several different problems including:
The intake filters are clogged and require cleaning.
The Temperature Sensor is faulty and requires replacement.
The Distributor is clogged and requires changing.
The Cellex Medium has entered the heat cavity of the unit and is restricting the air flow.
B. Reset the Temperature Sensor by turning the
unit Off. Press and hold the “Stop” button for ten to fifteen seconds. Release the “Stop” button and turn the unit On. It will be necessary to reset the Clock (refer to Users Manual). If the Error persists, remove and replace the Temperature Sensor. If the problem continues to persist, replace the pc board in the control panel. Refer to the appropriate section of this manual for proper removal and replacement procedures.
BBLLOOWWEERR//MMOOTTOORR PPRROOBBLLEEMM
IINNDDIICCAATTOORR..
TTEEMMPPEERRAATTUURREE SSEENNSSOORR
PPRROOBBLLEEMM IINNDDIICCAATTOORR..
Fluido DHT™Dry Heat Therapy Unit
10
TROUBLESHOOTING
FFlluuiiddoo DDHHTT SSyysstteemm TTeessttiinngg
AA.. GGeenneerraall
11..
The following information is intended to aid in troubleshooting the major components of the DHT Units to “Board Level” only. These tests are OEM standard testing procedures and methods used at the factory before shipment of any Fluido unit.
22..
Due to the complex nature of the technology utilized by Chattanooga Group, the recommended troubleshooting techniques are to determine “Bad Board” and board replacement only. No board component level troubleshooting is recommended nor will information or parts be supplied by Chattanooga Group. Any board component level troubleshooting performed will be at sole risk and liability of the Service Technician performing such troubleshooting techniques.
33..
Once the PC Board has been determined as bad, replace the board only with Chattanooga Group OEM replacement parts and hardware.
BB.. SSppeecciiaall TToooollss,, FFiixxttuurreess && MMaatteerriiaallss RReeqquuiirreedd
11..
Certain tests require the use of special tools and fixtures. These will be listed at the particular test where they are required. Testing with any other special tool or fixture other than those stated could give erroneous readings or test results. Always perform the tests exactly as stated to ensure accurate results.
22..
Standard test equipment settings will be listed for each test performed to aid in performing the test to OEM standards and ensure proper readings.
33..
The troubleshooting and repair of the Fluido DHT units, should be performed only by authorized technicians trained and certified by Chattanooga Group.
CC.. EEqquuiippmmeenntt RReeqquuiirreedd
11..
Digital Multimeter
22..
Dielectric Withstand (Hi-Pot) and ground resistance tester.
33..
Milliohm Meter.
44..
Calibrated Thermometer
55..
Calibrated Stop Watch
NNOOTTEE:: Adjust Dielectric Withstand tester to indicate fault with
120k Ohm Load across the output when at specified test voltage.
66.. RReeqquuiirreedd SSAAEE TToooollss #1 Phillips Screwdriver #2 Phillips Screwdriver 5/16, 7/16, 9/16, 1/2, 5/32, and 11/32 Wrenches 1/8 and 5/32 Allen Wrenches (or drill bits) Utility Knife Wet/Dry Vacuum
77.. RReeqquuiirreedd LLuubbrriiccaannttss Silicon based multipurpose grease.
88.. RReeqquuiirreedd SSeeaalliinngg CCoommppoouunndd 100% pure silicon sealant
NNOOTTEE:: The tool, lubricant and sealing compound requirements will
be listed at the respective removal and replacement procedures throughout this manual.
!
CAUTION
The following tool, lubrication, and sealing compound requirements are critical to the component removal and replacement of the Fluido DHT unit.
All hardware,bolts, nuts and screws used to assemble the Fluido DHT are SAE Standard. Due to the size of these components no metric equivalent is available. Therefore, it will be necessary to obtain the proper size tools for removal and replacement of certain components.
The lubricants and sealing compounds listed below are crucial in the assembly of certain components to ensure patient safety and efficient operation of the unit. Use only the recommended products listed or an approved equivalent possessing the same properties and qualities.
VViissuuaall IInnssppeeccttiioonn
AA.. GGeenneerraall
Visually inspect the Fluido DHT unit. A visual inspection can, to an experienced Technician, indicate possible abuse of the unit and/or internal problems.
GGrroouunndd RReessiissttaannccee TTeesstt
AA.. VVoollttaaggee SSppeecciiffiiccaattiioonnss
MMooddeell 11448800
. . . . .
Input: 120 VAC~50/60 Hz, 40 Watts
MMooddeell 11448811
. . . . Input: 230 VAC~50/60 Hz, 40 Watts
BB.. SSppeecciiffiiccaattiioonn
Maximum Acceptable Resistance: 500 milliohms
CC.. EEqquuiippmmeenntt RReeqquuiirreedd
Milliohm Meter
DD.. TTeesstt
Place unit on level work surface. Place one meter probe on the ground prong of power
cord and the other to any exposed metal or screw on the unit.
LLeeaakkaaggee TTeessttss
TTeesstt VVoollttaaggee SSppeecc
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
1000V
Conduct all necessary leakage tests as required per “Chapter 7 Electrical Equipment” of the 1999, or later, edition of the NFPA (National Fire Protection Association) “Health Care Facilities” standards.
RReesseettttiinngg FFlluuiiddoo DDHHTT FFaaccttoorryy DDeeffaauulltt SSeettttiinnggss
To reset all factory default settings of the unit, turn unit Off. Depress and hold the “Clear” button on the control panel and turn unit on simultaneously.
PPrreehheeaatt TTeesstt
AA.. TToooollss && EEqquuiippmmeenntt RReeqquuiirreedd
Calibrated Thermometer
BB.. PPrreehheeaatt TTeesstt PPrroocceedduurreess
Power Requirements:
Model 1480. . . . . . . . . . . . . . . . . . . . . . 120 VAC
Model 1481. . . . . . . . . . . . . . . . . . . . . . 230 VAC
Plug the unit power cord
to a grounded electrical
service receptacle conforming to
the applicable
national and local electrical codes. Turn unit power switch On.
Start the preheat function by depressing the “PREHEAT STANDBY” button on the control panel. After approximately 30 seconds, the Blue LED should illuminate.
After approximately 30 minutes or less, depending on the ambient temperature of the medium, the blower should reduce from 50% airspeed to 5% airspeed and the Temp indicator should read the temperature of the Cellex
®
medium. Insert the calibrated thermometer into medium and record the reading. The Temperature reading should be ± 5 °F or 3 °C of Spec below.
Preheat Default Temp. Spec . . . . 115 °F (46.11 °C)
A temperature reading outside the specified range may be caused by clogged intake filters, clogged distributor, faulty temperature sensor, or faulty heating element. Refer to the appropriate section for the proper removal and replacement procedures.
Fluido DHT™Dry Heat Therapy Unit
TROUBLESHOOTING
11
!
UNIT FAILING DIELECTRIC WITHSTAND AND/OR LEAKAGE TESTS COULD INDICATE SERIOUS INTERNAL SYSTEM PROBLEMS. DO NOT PLACE UNIT BACK INTO SERVICE! SEND UNIT TO FACTORY FOR REPAIR! DO NOT ATTEMPT TO REPAIR IN THE FIELD!
POWER
SWITCH
WARNING
Fluido DHT™Dry Heat Therapy Unit
12
PPuullssee MMooddee TTeesstt
AA.. TToooollss && EEqquuiippmmeenntt RReeqquuiirreedd
Calibrated Stop Watch
BB.. PPuullssee MMooddee TTeesstt PPrroocceedduurreess
Reset the factory defaults. Refer to the section entitled “Resetting Fluido DHT Factory Default Settings”on page 11.
Press the “PULSE MODE” button, the Blue LED should illuminate. Press the “TREATMENT START” button.
Using the calibrated stop watch, time the pulses of the unit while it is running. Record the readings taken and verify the settings you entered for the pulse. The unit should pulse at approximately four seconds on, after reaching its maximum blower speed, and four seconds off.
TTrreeaattmmeenntt TTiimmee TTeesstt
AA.. TToooollss && EEqquuiippmmeenntt RReeqquuiirreedd
Calibrated Stop Watch
BB.. TTrreeaattmmeenntt TTiimmee TTeesstt
With the unit On, press and release the “Time” button. Using the down arrow, adjust the time to “1:00”.
Press “TREATMENT START” and time with the Calibrated Stop Watch. Record the reading.
TTrreeaattmmeenntt TTiimmee SSppeecc
. . . . . . . . . . . . 1 minute ± 1 second
AAiirr SSppeeeedd TTeesstt
AAiirrssppeeeedd TTeesstt PPrroocceedduurreess
Turn unit On. Press “Treatment Start”. While unit is running, press the “Air Speed” button. Use the Up and Down arrows to adjust the air speed to 100% and back down to 5%. Listen for the increase and decrease in the blower speed. Look for increased and decreased fluidization of the medium in the reservoir.
CCoonnttrrooll PPaanneell LLoocckk--UUpp
Should the control panel Lock Up and not allow any operation of the unit or parameter changes. Reset the unit Factory Defaults (Refer to the section entitled “Resetting Fluido DHT Factory Default Settings”on page 11). If the condition persists, replace the Control Panel PC Board. Refer to “Control Panel PC Board” on page 13 for removal and replacement procedures.
NNOOTTEE::
For proper operation and setting of the unit parameters outside the Factory Defaults, refer to the User Manual for the Fluido DHT.
TROUBLESHOOTING
Fluido DHT™Dry Heat Therapy Unit
13
SSlleeeevveess aanndd CCeelllleexx
®®
Refer to the Fluido DHT User Manual for the proper removal and replacement of the following items:
Unit End Sleeves Unit Top Sleeves Cellex Medium Intake Filters
CCoonnttrrooll PPaanneell PPCC BBooaarrdd
AA.. TToooollss && EEqquuiippmmeenntt RReeqquuiirreedd
Small Flat Blade Screwdriver 5/16 Open End Wrench #2 Phillips Screwdriver
BB.. RReemmoovviinngg tthhee CCoonnttrrooll PPaanneell PPCC BBooaarrdd
1. Remove the three retaining screws from the Control Panel Bezel.
2. Carefully remove the Bezel from the Control Panel.
3. Remove the four retaining nuts securing the Control Panel to its mounting base (two on each side).
REMOVAL & REPLACEMENT
!
UNPLUG THE UNIT FROM THE POWER SOURCE BEFORE ATTEMPTING ANY REMOVAL OR REPLACEMENT PROCEDURES TO PREVENT ELECTRICAL SHOCK. FOLLOW ELECTRONIC REPAIR PROTOCOLS FOR GROUNDING TO PREVENT DAMAGE TO THE ELECTRONIC COMPONENTS FROM STATIC ELECTRICITY.
WARNING
Loading...
+ 34 hidden pages