Chattanooga 2C, 4C User Manual

User Manual
COMBO
I
NTELECT
®
LEGEND
SERIES
Models 2C & 4C
Table of Contents
Foreword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Precautionary Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Package Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Indications, Contraindications and Adverse Effects for Electrical Stim . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
User Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Technical Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Unit Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-14
Pain Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Interferential . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Premodulated . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Muscle Contraction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
High Volt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-19
Russian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Ultrasound Indications and Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Ultrasound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Combination Therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24-25
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-28
INTLECT LEGEND COMBO
3
Foreword
This manual has been written for the owners and operators of the Intelect®Legend Combo, models 2C and 4C. It contains general instructions on operation, precautionary practices, maintenance and parts information. In order to maximize use, efficiency and the life of your unit, please read this manual thoroughly and become familiar with the controls as well as the accessories before operating the unit.
Specifications put forth in this manual were in effect at the time of publication.
However, owing to Chattanooga Group's policy of continual improvement, changes to these specifications may be made at any time without obligation on the part of Chattanooga Group.
Product Description
With the same legendary performance, quality and value that has made the Intelect®name respected world-wide, the Intelect®Legend Combo offers the convenience of a full-featured stimulator and ultrasound in one device. Easy to use 1-2-Go software makes this unit a pleasure to operate.
Features include Interferential, Premodulated, High Volt and Russian waveforms. Dual frequency ultrasound features a 5 cm
2
soundhead, which delivers 1 or 3.3 MHz frequencies, 10%, 20%, 50% and continuous duty cycle selections, and head warming. Unique Electronic Signature™ control allows all available ultrasound applicators to be interchangeable. Ultrasound operation is independent or in combination with Interferential, Premodulated and High Volt waveforms.
Models 2C and 4C of the Intelect Legend Combo are prescription devices used under the supervision or by the order of a physician or other licensed healthcare provider.
©2004 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial or layout composition of this publication without expressed written consent from the Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated and prepared for print by the Chattanooga Group of Encore Medical, L.P.
4
INTELECT LEGEND COMBO
Precautionary Instructions
1. CAUTION:
Read, understand and practice the precautionary and operating
instructions. Know the limitations and hazards associated with using any electrical stimulation or ultrasound device. Observe the precautionary and operational decals placed on the unit.
2. CAUTION: Do not operate the Intelect®Legend Combo when connected to any unit other than Chattanooga Group devices. Do
not operate the unit in an environment of short-wave diathermy use.
3. WARNING: Federal law restricts this device to sale by, or on the order of, a physician or licensed practitioner. This device should
be used only under the continued supervision of a physician or licensed practitioner.
4. CAUTION: This unit generates, uses and can radiate radio frequency energy, and if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. Harmful interference to other devices can be determined by turning this unit on and off or trying to correct the interference using one or more of the following: Reorient or relocate the receiving device, increase the separation between the equipment, connect the unit to an outlet on a different circuit from that to which the other device(s) are connected and/or consult the factory field service technician for help.
5. CAUTION:
The Ultrasound generator should be routinely checked before
each use to determine that all controls function normally; especially that the intensity control does properly adjust the intensity of the ultrasonic power output in a stable manner.
Also, determine that the treatment time control does actually
terminate ultrasonic power
output when the timer reaches zero.
6. CAUTION: Use of controls or adjustments or performance of proce
dures other than those specified herein may result in
hazardous
exposure to ultrasonic energy.
7. CAUTION: This unit is not designed to prevent the ingress of water or liquids. Ingress of water or liquids could cause
malfunction of internal components of the system and therefore create a risk of injury to the patient.
8. CAUTION: DO NOT use sharp objects such as a pencil point or ballpoint pen to operate the buttons on the control panel as
damage may result.
9. WARNING: Equipment not suitable for use in the presence of flammable anesthetics mixture with air, oxygen, or nitrous oxide.
The warning symbol for this hazard is prominently displayed on the cabinet.
5
INTELECT LEGEND COMBO
10. WARNING: Keep electrodes separated during treatment. Electrodes in contact with each other could result in improper
stimulation or skin burns.
11. WARNING: For continued protection against fire hazard, replace fuses only with ones of the same type and rating.
12. WARNING:
Make certain that the unit is electrically grounded by connecting
only to a grounded electrical service receptacle
conforming to the applicable national and local electrical codes.
13. WARNING: This device should be kept out of the reach of children.
14. WARNING: Care must be taken when operating this equipment around other equipment. Potential electromagnetic or other
interference could occur to this or to the other equipment. Try to minimize this interference by not using other equipment in conjunction with it.
15. CAUTION: This unit should be operated, transported and stored in temperatures between 59° F and 104° F (15° C and 40° C),
with Relative Humidity ranging from 30%- 60%.
16. CAUTION:
Handle ultrasound applicator with care. Inappropriate handling
of the ultrasound applicator may adversely affect its
characteristics.
17. CAUTION:
Inspect treatment head for cracks, which may allow the ingress
of conductive fluid before each use.
18. CAUTION: Inspect treatment head cables and associated connectors before each use.
19. WARNING: Use caution when operating in “Interferential” (IFC) mode. It is recommended to always use new electrode pads
when IFC modality is administered. Unseen residues on repeatedly used electrodes could possibly create an electrical shock or burn sensation to the patient.
20. CAUTION: DO NOT raise intensity after treatment has been started.
21. DANGER: Patients with an implanted neurostimulation device must not be treated with or be in close proximity to any shortwave
diathermy, microwave diathermy, therapeutic ultrasound diathermy or laser diathermy anywhere on their body. Energy from diathermy (shortwave, microwave, ultrasound and laser) can be transferred through the implanted neurostimulation system, can cause tissue damage and can result in severe injury or death. Injury, damage or death can occur during diathermy therapy even if the implanted neurostimulation system is turned "off."
Before administering any treatment to a patient you should become acquainted
with the operating procedures for each mode of treatment available, as well as the indications, contraindications, warnings and precautions. Consult other resources for additional information regarding the application of electrotherapy.
6
INTELECT LEGEND COMBO
Installation
Initial Setup Instructions
Remove the Intelect®Legend Combo unit and any additional items ordered from the carton and inspect for damage that may have
occurred during shipment. Check the voltage rating on the serial decal located
on the bottom of the unit. Plug the
system power supply in to a 100 Volt to 220-240 Volt AC outlet, as required.
CAUTION
• DO NOT attempt to use Direct Current (DC).
• DO NOT place unit in a location where the power cord could be tripped over or pulled out during treatment.
• DO NOT attempt to use the unit if it is not properly grounded.
Package Contents
Standard Accessories
The following accessories are included with your Intelect®Legend Combo:
78047 Applicator, Ultrasound 5 cm
2
4248 Conductor™Gel 12213 Lead, 120", Red/Black Channels 1 and 2 12214 Lead, 120", Channels 3 and 4 (Model 4C only) 72853 Electrodes, Carbonflex, 3" Round, Red 72852 Electrodes, Carbonflex, 3" Round, Black 10648 Nylatex, 2-
1
/2" x 24", Sewn
79079 Operator's Manual
7
INTELECT LEGEND COMBO
Optional Accessories
The following is a list of optional accessories available for the Intelect®Legend Combo:
78046 Applicator, Ultrasound 10 cm
2
78048 Applicator, Ultrasound 2 cm
2
79977 High Volt Probe Kit 10832 Strap, Nylatex, Long 2-1/2"x 48" 10648 Strap, Nylatex, Medium, 2-
1
/2"x 24" 10828 Strap, Nylatex, Short, 2-1/2"x 18" 78022 Patient Control Center
Indications/Contraindications Adverse Effects for Electrical Stimulation
Interferential and Premodulated
Indications
• Symptomatic relief of chronic, intractable pain.
• Management of pain associated with post-traumatic or postoperative conditions.
Contraindications
• This device should not be used for symptomatic pain relief unless etiology is established or unless a pain syndrome has
been diagnosed. This device should not be used on patients with demand type cardiac pacemakers. This device should not be used over cancerous lesions.
• Electrode placements must be avoided that apply current to the carotid sinus region (anterior neck) or transcereberally
(through the head).
8
INTELECT LEGEND COMBO
Warnings
• The long term effects of chronic electrical stimulation are unknown. Safety has not been established for the use of therapeutic electrical stimulation during pregnancy.
• Adequate precautions should be taken when treating individuals with suspected or diagnosed heart problems, or epilepsy.
• Benefits of Interferential stimulation have not been established for pain of central origin.
• This device is to be used as a symptomatic treatment for pain and has no curative value. Patients should be cautioned and their activities regulated if pain is suppressed that would otherwise serve as a protective mechanism.
• Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when electrical stimulation is being utilized.
Precautions
• Isolated cases of skin rash may occur at the site of electrode placement following long term applications. The irritation may be reduced by use of an alternate conductive medium or an alternative electrode placement.
• Effectiveness of this treatment is dependent upon patient selection.
Adverse Effects
• Skin irritation and burns beneath the electrodes have been reported with the use of therapeutic electrical stimulation.
Russian and High Volt
Indications
• Relaxation of muscle spasms.
• Prevention or retardation of disuse atrophy.
• Increasing local blood circulation.
• Muscle re-education.