Chattanooga Intelect Shortwave 100, 1602, Intelect Shortwave 100 Canada, 1604 User Manual

Supplied by EME Services Ltd
User Manual
Operation & Installation
Instructions for:
Intelect Shortwave 100
REF 1602
Intelect Shortwave 100 (Canada)
REF 1604
TABLE OF CONTENTS
W
Supplied by EME Services Ltd
FOREWORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Product Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
SAFETY INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Precautionary Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Dangers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Personal Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Protection of the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
ABOUT SHORTWAVE DIATHERMY . . . . . . . . . . . . . . . . . . . . . 16
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Pulsed Shortwave Diathermy - Introduction . . . . . . . . . . 16
INDICATIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
CONTRAINDICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
ADDITIONAL PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . 24
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Unpacking the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Requirements for Installation. . . . . . . . . . . . . . . . . . . . . . . . 27
Requirements for the Installation Location. . . . . . . . . . . . 27
NOMENCLATURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Front View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5
User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Software Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Capacitive Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Flexible Rubber Electrodes. . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Monode (Drum) Electrode . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Diplode Electrode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Indicator Discharge Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
SPECIFICATIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Unit Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Operating Data and Ratings. . . . . . . . . . . . . . . . . . . . . . . . . . 34
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Transport and Storage Conditions. . . . . . . . . . . . . . . . . . . . 35
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Applicator Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Load Resistance and Applicator Spacing. . . . . . . . . . . . . . 36
Description of Device Markings . . . . . . . . . . . . . . . . . . . . . . 37
OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Description of Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
i
TABLE OF CONTENTS
Supplied by EME Services Ltd
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Applications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Shortwave Diathermy in the
Capacitive (Dielectric) Field . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Shortwave Diathermy in the Inductive Field . . . . . . . . . . 40
Installing the Applicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Installing the Capacitive Electrodes . . . . . . . . . . . . . . . . . . 41
Installing the Monode (Drum) Electrode . . . . . . . . . . . . . . 43
Installing the Flexible Rubber Electrodes . . . . . . . . . . . . . 45
Preparing the Unit for Therapy. . . . . . . . . . . . . . . . . . . . . . . 46
Preparing the Patient for Therapy . . . . . . . . . . . . . . . . . . . . 47
Basic Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Starting Therapy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Home Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Edit Treatment Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Electrode Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Thermal Dosimetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Quick Link Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Starting Therapy from the Home Screen. . . . . . . . . . . . . . 52
Using the Indicator Discharge Tube . . . . . . . . . . . . . . . . . . 53
Stopping, Pausing, and Completing Therapy. . . . . . . . . . 54
Stopping Therapy Using the STOP Button . . . . . . . . . . . . 54
Stopping Therapy Using the Patient
Interrupt Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Pausing Therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Completing the Therapy Session . . . . . . . . . . . . . . . . . . . . . 55
Using Clinical Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Retrieving, Deleting, and Saving User Protocols. . . . . . . 58
Retrieving User Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Deleting User Protocols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Saving New User Protocols. . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Editing Existing User Protocols . . . . . . . . . . . . . . . . . . . . . . . 60
Using Clinical Resources - Educational Library . . . . . . . . 61
Anatomical Graphics Library . . . . . . . . . . . . . . . . . . . . . . . . . 61
Electrode Placement Library . . . . . . . . . . . . . . . . . . . . . . . . . 62
Diathermy Rationale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Contraindications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Using Clinical Resources - Patient Card . . . . . . . . . . . . . . . 65
Patient Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Edit Pain Profile. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Saving Treatments to the Patient Card. . . . . . . . . . . . . . . . 68
Erasing Patient Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Using Clinical Resources - Unit Settings. . . . . . . . . . . . . . . 70
Adding and Changing the Clinic Name . . . . . . . . . . . . . . . 70
Increasing and Decreasing the Unit Volume . . . . . . . . . . 70
Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Resetting Unit Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Resetting Default Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Changing Languages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Using the Applicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Using the Capacitive Electrodes . . . . . . . . . . . . . . . . . . . . . . 74
Using the Monode (Drum) Electrode . . . . . . . . . . . . . . . . . 75
ii
TABLE OF CONTENTS
T
T
W
T
T
T
Supplied by EME Services Ltd
Using the Flexible Rubber Electrode. . . . . . . . . . . . . . . . . . 76
Treatment Tips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Positioning the Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Positioning of Electrodes - Edge Effect. . . . . . . . . . . . . . . . 78
Positioning of Electrodes - Edge Effect
and Metal Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Positioning of Electrodes - Electrode
Constriction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Positioning of Electrodes - Normal Cases . . . . . . . . . . . . . 80
Setting the Electrode - Skin Distance . . . . . . . . . . . . . . . . . 81
Dosage Levels According to Schliephake . . . . . . . . . . . . . 82
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Error Codes and Descriptions . . . . . . . . . . . . . . . . . . . . . . . . 83
Using the Indicator Discharge Tube . . . . . . . . . . . . . . . . . . 84
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Maintaining the Intelect Shortwave 100 . . . . . . . . . . . . . . 85
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Cleaning and Disinfection. . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Safety Inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Warranty Repair/Out of Service Repair. . . . . . . . . . . . . . . . 87
ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Standard Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
APPENDIX A  EMC TABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Table 1: Guidance and Manufacturer’s Declaration–
Electromagnetic Emissions . . . . . . . . . . . . . . . . . . 90
Table 2: Guidance and Manufacturer’s Declaration–
Electromagnetic Immunity . . . . . . . . . . . . . . . . . . 91
Table 3: Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the
Intelect Shortwave 100. . . . . . . . . . . . . . . . . . . . . . 93
iii
FOREWORD
Supplied by EME Services Ltd
This manual has been written for the owners and operators of the Intelect® Shortwave 100 system. It contains general information on the operation, precautionary practices, and maintenance information. In order to maximize use, efficiency, and the life of your unit, please read this manual thoroughly and become familiar with the controls, as well as the accessories, before operating the unit. This manual contains general safety, operating, maintenance, and care instructions for the owners and operators of the Intelect Shortwave 100 system.
Specifications put forth in this manual were in effect at the time of publication. However, owing to Chattanooga Group's policy of continual improvement, changes to these specifications may be made at any time without obligation on the part of Chattanooga Group.
Before administering any treatment to a patient, the user of this equipment should read, understand, and follow the information contained in this manual for each mode of treatment available, as well as the indications, contraindications, warnings, and precautions. Consult other resources for additional information regarding the application of shortwave diathermy.
Product Description
The Intelect Shortwave 100 utilizes both inductive and capacitive electrodes to administer the clinical application of oscillating electromagnetic energy in the radio frequency of 27.12 megahertz to treat a wide variety of orthopedic and physiotherapeutic conditions. Diathermy generates deep heat within body tissues to increase circulation, decrease pain, decrease inflammation, increase ROM, influence muscle tone, and facilitate the sub acute healing phase.
Stay current with the latest clinical developments in the field of shortwave diathermy. Observe all applicable precautionary measures for treatment.
Keep informed on appropriate indications and contraindications for the use of shortwave diathermy.
This equipment is to be used by, and sold to, a trained clinician only under the prescription and supervision of a licensed practitioner.
©2009 Encore Medical, L.P. and its affiliates, Austin, Texas, USA. Any use of editorial, pictorial, or layout composition of this publication without express written consent from Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated, and prepared for print by Chattanooga Group of Encore Medical, L.P. Certain portions of this manual have been reproduced with the consent of gbo Medizintechnik AG.
1
FOREWORD
Supplied by EME Services Ltd
GENERAL INFORMATION
The Intelect Shortwave 100 unit and the accompanying components and individual elements fulfill, as a unit, the currently valid safety standards and comply with the stipulations of IEC 60601-1 and the medical products regulations.
The unit and its external components (accessory elements) are safe if used properly and in compliance with the explanations and instructions provided in this documentation. Nevertheless, the unit or its external components can pose dangers. Therefore, we urgently recommend that anyone operating the shortwave diathermy unit become aware of the potential dangers of the unit and its external components before beginning work.
Please read and observe all safety instructions in this operating manual. The following features are available on the Intelect Shortwave 100 unit:
Clear Touch Screen LCD
Allows the operator to select an option on the screen by simply touching it. Guides the operator through the setup process providing continuous feedback about treatment settings. Gives the operator optimal visibility during attended procedures. Allows the operator to turn on the unit simply by pressing the LCD.
Clinical Protocols™
An efficient approach for setting up a treatment using preset parameters.
Thermal Dosimetry
Allows the operator to set up a treatment using the four treatment dosages according to Schliephake (see page 82).*
*Low and Reed (1990)
2
FOREWORD
Supplied by EME Services Ltd
GENERAL INFORMATION CONTINUED
Quick Link Indications
The Intelect Shortwave 100 incorporates a unique Quick Link Indications section which allows the user to select specific clinical indications and apply the most common therapy for the Indication selected. All modalities are editable, in their normal editing fashion, in order to customize the treatment for each patient’s prescribed therapy.
User Protocols
User protocols allow you to set, save, and change the parameters of each program (protocol) in order to tailor it to meet your patients’ specific needs. You may create up to 200 user protocols.
Automatic Tuning
The Intelect Shortwave 100 has been programmed to automatically regulate the power between the electrodes and the patient.
Screen Saver
The screen saver function acts like any other screen saver on a normal computer to reduce LCD screen burn-in. The unit will enter screen saver mode after 10 minutes of inactivity. If the user presses the touch screen or any button on the User Interface, it will return to the Home screen.
Realtime Clock
The Realtime Clock keeps track of date and time even when the unit is turned off or unplugged.
3
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
PRECAUTIONARY DEFINITIONS
The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definition of these symbols are as follows:
Explosion Hazard
Text with an “Explosion Hazard” indicator will explain possible safety infractions if this equipment is used in the
Text with a “CAUTION” indicator will explain possible safety infractions that could have the potential to cause minor to moderate injury or damage to equipment.
Text with a “WARNING” indicator will explain possible safety infractions that will potentially cause serious injury and equipment damage.
presence of flammable anesthetics.
Dangerous Voltage
Text with a “Dangerous Voltage” indicator serves to inform the user of possible hazards resulting in the electrical charge delivered to the patient or operator in certain treatment configurations.
Non-ionizing Electromagnetic Radiation
Text with a “Non-ionizing Electromagnetic Radiation" indicator informs the user of possible hazards resulting from elevated, potentially dangerous, levels of non-ionizing radiation.
Text with a “DANGER” indicator will explain possible safety infractions that are imminently hazardous situations that would result in death or serious injury.
Refer to Instruction Manual/Booklet
NOTE: Throughout this manual, “NOTE” may be found. These
Notes are helpful information to aid in the particular area or function being described.
4
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
Read , understand, and practice the precautionary operating instructions. Know the limitations and hazards associated with using any shortwave diathermy device. Observe the precautionary and operational decals placed on the unit.
DO NOT operate the Intelect Shortwave 100 in conjunction with any
other devices. DO NOT operate this unit in an environment where other devices
are being used that intentionally radiate electromagnetic energy in an unshielded manner. Portable and mobile RF communications equipment can affect medical electrical equipment.
The unit should be routinely checked before each use to determine
that all controls function normally, especially that the output control does properly adjust the intensity of the shortwave diathermy power output in a stable manner. Also, determine that the treatment time control does actually terminate shortwave diathermy power output when the timer reaches zero.
DO NOT use sharp objects such as a pencil point or ballpoint pen to
operate the buttons on the control panel as damage may result. This unit should be transported and stored in temperatures
between -40 °C and 70 °C (-40 °F and 158 °F) to prevent damage to the unit or its components.
Handle shortwave diathermy accessories with care. Inappropriate
handling of the accessories may adversely affect their characteristics.
Inspect cables, electrodes, associated connectors, and accessories before each use. To test the Patient Interrupt Cord, follow the procedures listed in "Stopping Therapy Using the Patient Interrupt Cord" on page 54.
Ex
nal conductive material should be removed from the
ter
immediate treatment area. Do not use accessories other than those supplied with the unit, or
recommended by Chattanooga Group. The safety of other products has not been established, and their use could result in injury to the patient and degrade minimum safety.
Disconnect the power supply cord before removing covers on
this equipment. Refer the servicing of this unit to qualified service personnel.
This equipment has an output that is capable of producing a
physiological effect. Grounding reliability can only be achieved when the equipment is
connected to an equivalent receptacle marked "Hospital Only" or "Hospital Grade."
Medical electrical equipment needs special precautions regarding
EMC and needs to be installed and put into service according to the EMC information provided in this manual.
5
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
This unit generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. Harmful interference to other devices can be determined by turning this equipment on and off. Try to correct the interference using one or more of the following: reorient or relocate the receiving device, increase the separation between the equipment, connect the equipment to an outlet on a different circuit from that to which the other device(s) are connected, and consult the factory field service technician for help.
Medical devices that are to be used within
12 meters of the
Intelect Shortwave 100 should have an electromagnetic site survey completed. See NOTE 2 on page 92 for more information.
T
mine the distance of separation for all equipment operating
o deter
near the Intelect Shortwave 100, consult the EMC tables applicable to the other equipment. If the distance cannot be determined, make certain that the other equipment is operated from a distance of no less than 12 meters.
This equipment is to be used by, and sold to, a trained clinician only
under the prescription and supervision of a licensed practitioner.
Since relatively high powers are used, there is the possibility of producing shock, localized burns, and cataracts if the patient is unaware of the heat due to reduced thermal sensation, or if the patient does not know what to expect during treatment.
Improper installation, operation or maintenance of the shortwave
diathermy system may result in malfunctions of this unit or other devices.
In case of display failure or other obvious defects, switch the unit off
immediately by means of the power switch, disconnect the power cord from the power outlet, and notify a certified service technician.
Be aware that some synthetics and plastics, though assumed to be
non-conductive, may be heated by shortwave diathermy. Adjustments or replacement of components may result in the
equipment failing to meet the requirements for interference suppression.
If the unit cannot be installed immediately after delivery, the unit and
its external components or accessory elements must be stored in their original packaging in a dry place.
Do not store or operate the unit in a dusty environment.
Do not cross cables. Keep all electrodes, accessories, and their cords separated during
treatment by using the cable clips located on the side of the arm extenders. Electrodes or their cords in contact with each other during treatment could result in improper stimulation, skin burns, or damage to the cord or electrode.
6
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
Do not lean on or hold the cables during treatment. Keep all line cords away from the diathermy unit cables. Do not
store or coil line cords where they can come close to the cables on an operating shortwave diathermy unit.
This equipment is not designed to prevent the ingress of water
or liquids. Ingress of water or liquids could cause malfunction of internal components of the system and therefore create a risk of injury to the patient.
Care must be taken when operating this unit adjacent to or stacked
with other equipment. If adjacent or stacked use is necessary, the Intelect Shortwave 100 should be observed to verify normal operation in the configuration in which it will be used. Potential electromagnetic or other interference could occur to this or other equipment. Try to minimize this interference by not using other equipment in conjunction with it.
Use only accessories that are specially designed for this unit. Do
not use accessories manufactured by other companies on this unit. Chattanooga Group is not responsible for any consequence resulting from using products manufactured by other companies. The use of other accessories or cables (other than those specified) may result in increased emissions or decreased immunity of this unit.
Metal in treatment area will provide low impedance paths to the induced radio frequency current, producing local heating and the possibility of burning. In particular, treatment should never be given in the area of metal implants. Also, metal jewelry, buckles, cell phones, etc must be removed.
Use of controls or adjustments or performance of procedures other
than those specified herein may result in hazardous exposure to shortwave diathermy energy.
Make certain that the unit is electrically earthed by connecting
only to a earthed electrical service receptacle, conforming to the applicable national and local electrical codes.
This device should be kept out of the reach of children.
Induction field electrodes that are operated without a patient could be destroyed due to overheating.
Make certain that the electrode arms and arm extenders are locked
firmly into place during shortwave diathermy therapy using the front and rear handwheels to prevent unintentional movement. See page
28 for the nomenclature of the unit.
D
o not leav
therapy.
e patient unattended during shortwave diathermy
7
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
Remove the electrode applicator by pulling the cable connector only. DO NOT remove by pulling the cable.
To remove the cable from the applicator, make certain the power
is off. While the electrode applicator is being supported by the electrode arm, hold the electrode while removing the cable to prevent the electrode from dropping to the floor.
Tighten arms so that they do not move during therapy.
Observe the patient and the position of the arms at all times during therapy.
Inform the patient that the arms are not supposed to move during
therapy. Before using the unit, verify the patient is not in contact with the unit,
the electrode connection cable, the electrodes, or other devices or metal objects.
In the event that an Error message or Warning appears beginning with a 2 or 3, immediately stop all use of the unit and contact the dealer or Chattanooga Group for service. Errors and Warnings in these categories indicate an internal problem with the unit that must be tested by Chattanooga Group or a Field Service Technician certified by Chattanooga Group before any further operation or use of the system.Use of a unit that indicates an Error or Warning in these categories may pose a risk of injury to the patient, user, or extensive internal damage to the system.
8
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
Do not administer shortwave diathermy to a patient who has had an implant in the past unless you are absolutely certain that the implant and all leads in their entirety have been removed. Note that the leads are often left implanted after the implant is removed.
This unit generates non-ionizing radiation. Patients with
implanted electronic devices, such as cardiac pacemakers and defibrillators, cochlear implants, bone growth stimulators, deep brain stimulators, spinal cord stimulators, and other nerve stimulators, must not be treated, even if the device has been turned off.
The function of certain implanted devices (e.g.,
pacemakers) may be adversely affected during treatment with shortwave diathermy. In case of doubt, the advice of a licensed practitioner in charge of the patient should be sought.
Shortwave diathermy should not be used on patients who
have any implanted metallic lead or any implanted system that may contain a lead. Both the heating and non-heating modes of operation pose a risk of tissue destruction. If you are a licensed practitioner who implants or monitors patients with leads or implanted systems with leads, explain to the patient what diathermy is and stress that they should not receive shortwave diathermy treatment. If you are a licensed practitioner who uses diathermy in your practice, be sure to ask patients about possible implants before deciding to administer shortwave diathermy.
Other equipment, including patient connected devices, may be adversely affected when in close proximity to shortwave diathermy equipment.
Patients should not be treated with shortwave diathermy when
they have reduced thermal sensitivity over the proposed area of treatment, unless the physician in charge of the patient is notified.
Treatment should not be given through clothing, although it is
permissible to administer treatment through a dressing or plaster in pulsed modes.
Remove hearing aids prior to treatment.
At average power levels above 5 W, patients should not be allowed to come into contact with conductive parts which are earthed or which have an appreciable capacitance to earth and which may provide unwanted pathways for the radio frequency current. In particular, treatment must never be given with the patient on metal framed couches, chairs, or beds. Do not use conductive mattresses or mattress covers.
Before increasing the output in response to a report of inadequate
patient heating, verify that the cables are properly routed, spaced correctly, and away from metal or grounded objects. The heating effect may be misdirected and heating may be occurring in an unwanted area.
Before each use, check the condition of the housing and the
insulation of the electrodes, electrode connection cable, and the power supply cable. Also make sure that the cables have been routed correctly.
9
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
If the unit is not safe for operation, then it must be repaired by certified service personnel and the operators must be informed of the dangers posed by the unit.
In order to prevent electrical shock, unplug the power plug
from the socket before cleaning or disinfecting the unit. Under no circumstances may liquid penetrate the openings
on the unit, e.g. the connecting sockets of the electrode cables. Therefore, do not use cleaning or disinfectant sprays.
The unit, electrodes, and cables may not be sterilized using
steam or gas. Never clean the unit with abrasives, disinfectants or solvents
that could scratch the housing or damage the unit. Internal burns can occur with the incorrect application of
shortwave diathermy due to excessive intensity. Internal burns can occur with the incorrect application of
shortwave diathermy due to excessive exposure time. Do not perform unauthorized repairs under any
circumstances. The unit and the electrodes must be positioned so that
there is no danger of personal injury. Therefore, you must read and observe the safety instructions and the list of contraindications before putting the unit into operation.
10
The Electrode-Skin Distance (ESD) must be small for surface warming and large for depth warming. A larger Electrode-Skin Distance (ESD) is necessary for patients with a thick layer of subcutaneous fat in order to achieve the necessary warming of deep-lying tissue.
Explosion hazard if Intelect Shortwave 100 is used in
the presence of flammable anesthetics mixture with air, oxygen, or nitrous oxide.
The operator should not use diathermy over the heart in
order to prevent theoretical cardiac signal interference. Never, under any circumstances, attempt to hold any of
the electrodes in your hands during therapy. The unit must be installed so that there is no danger to
the patient, the operator or other persons. Therefore, you must read the safety instruction and contraindications.
Keep all unnecessary persons out of the treatment
location. No other person should be located within 3 meters of the unit.
The Intelect Shortwave 100 unit may contain Di (2-
ethylhexyl) phthalate (DEHP) which is the plasticizer for most PVC medical devices. Everyone is exposed to small levels of DEHP in everyday life. However, some
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
individuals can be exposed to high levels of DEHP through certain medical procedures. DEHP can leach out of plastic medical devices into solutions that come into contact with the plastic. The amount of DEHP that will leach out depends on the temperature, the lipid content of the liquid, and the duration of contact with the plastic. As a preventative measure, use PVC devices that do not contain DEHP, or devices made of other materials such as ethylene (EVA), silicone, polyethylene, or polyurethane for all clinical procedures. Avoid using DEHP-leaching PVC when performing all clinical procedures on male neonates, pregnant women who are carrying male fetuses, or peripubertal males.
The Intelect Shortwave 100 unit may contain bisphenol-A (BPA)
which is a building block of polycarbonate plastic used to make numerous consumer products (baby bottles, 5 gallon water bottles, etc). This chemical is considered an Unclassifiable Carcinogen and a suspected Endocrine disruptor which may interfere with or block hormones. BPA is highly toxic if swallowed, absorbed through the skin, or inhaled.
In case of damage from transport that could endanger personal safety, the unit must not be connected to the Mains Power Supply before inspection is complete.
Adhere to rules, regulations, and ordinances that may vary from
location to location concerning the appropriate use of high powered radio frequency fields.
Since the effects of high-frequency fields on unborn life have not yet
been sufficiently researched, we recommend that operators who are pregnant are not within 15 meters of the applicator when the unit is activated.
Any persons with pacemakers or implants must remain outside of
the treatment area during shortwave diathermy. No one wearing a cardiac pacemaker should be within 15 meters of an operating unit.
11
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
PERSONAL SAFETY
In case of improper or unauthorized use of the unit, the operator, the patient or other persons may be subjected to the danger of electric shock due to high voltage produced by the unit, the danger of influence on active implantations by magnetic fields produced by the unit and the danger of being burned due to erroneously positioned electrodes or false parameters such as the duration of treatment, power output or operating mode.
Before operating the unit, please read this instruction manual carefully and observe the information contained therein.
Pay special attention to the list of contraindications. Refer to "Safety Instructions" on pages 5-11, "Contraindications" on pages 20-23, and "Additional Precautions" on pages 24-25.
Before operating the unit each time, verify that:
the unit has been correc Supply.
the unit has been set up so that it is free-standing and the patient is not in direct contact with metal objects such as heating radiators, metal beds, or other equipment.
the insulation of the RF output jack and electrode connection cables is not damaged.
the electrode connection cable is connected properly and is not cross-routed (which may cause capacitive short circuits).
only accessories (cables, electrodes) approved by the manufacturer are connected.
tly connected to the Mains Power
the patient to be treated (and the personnel) have removed all electric devices (e. g. hearing aids, electrotherapy electrodes, mobile telephones) and all conductive objects (e.g. rings, chains, watches, earrings or other jewelry, eyeglasses) and that they are not in the immediate vicinity of the unit, the patient is in a composed state and the bodily areas to be treated are dry on the exterior.
the electrodes are positioned according to the doctor’s
instructions (to be checked by the doctor or physiotherapist if applied by assisting personnel).
there are no unneeded personnel in the room other than the operator and the patient.
there is no danger of unwanted local warming due to electrode constrictions.
Before using the unit, speak with the patient to verify:
the patient is in a comfortable position during the entire treatment.
the patient is not in contact with the unit, the electrode connection cable, the electrodes, or other devices or metal objects.
that the patient is comfortable before and during treatment.
12
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
PERSONAL SAFETY CONTINUED
Before using the unit, determine the maximum nominal output power of the respective accessory in order to avoid overheating the tissue.
At regular intervals during the treatment, verify:
that the unit is functioning properly
that there is no moisture development* (perspiration) in the area of the electrodes
if the patient feels well.**
After the treatment, ask the patient about the tolerance of the treatment. The treatment environment should be inspected by a licensed practitioner.
The affected parts of the body should be unclothed during * treatment, since accumulation of moisture on the skin or in folds can cause local overheating of the skin. This is especially important in the event that the patient is wearing clothing made of moisture-resistant fabric such as silk or synthetic fibers.
*The output power must always be set according to the * subjective response of the patient. Particular care is to be taken with patients who have a reduced capacity for heat perception (see the dosage levels according to Schliephake on page 82).
13
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
PROTECTION OF THE UNIT
Improper installation, operation or maintenance of the shortwave diathermy unit may result in malfunctions of this unit or other devices.
Observe the following instructions in order to prevent malfunctions:
In order to prevent electromagnetic disturbances, place the unit at least 12 meters (see the Caution on page 6 for more information) from any other devices. Also make sure that there is sufficient distance between the unit and Mains Power Supply or data cables in walls, ceilings and floors, since the electromagnetic radiation from the unit can pass these essentially without hindrance.
In selecting the location for the unit, make sure that the patient has contact during the treatment to the non­earthed application element and, due to equalizing currents in case of differing potentials, that the patient is never in contact with metal elements (especially if they are earthed), such as heating radiators, metal beds or other earthed devices.
Before connecting the unit, make sure that:
the voltage rating on the safety label corresponds to the available system voltage.
the frequency rating on the rating plate corresponds to the system frequency.
an earthed socket outlet with earthing contact is available for connecting the unit.
the routing of the power cable from the unit to the socket outlet with earthing contact does not pose a danger for personnel or the patient.
the Mains Power Supply is designed for the comparatively high (possibly additional) power input of the unit (~ 1500 VA) and the line is sufficiently protected in accordance with regulations.
Make certain that the unit is electrically earthed by connecting only to a earthed electrical service receptacle, conforming to the applicable national and local electrical codes.
14
SAFETY INSTRUCTIONS
Supplied by EME Services Ltd
PROTECTION OF THE UNIT CONTINUED
Do not connect the unit to the Mains Power Supply until the following requirements have been met:
Before putting the unit into operation, check to make sure
that the electrode connection cable and the electrodes are undamaged and have been connected correctly to the unit.
Never operate the unit with open outputs, (i.e. without electrodes).
Do not operate the unit for an extended period with no load (without a patient), especially in coil (induction field) mode. When operating the unit without power output, induction field electrodes could be destroyed due to overheating.
Pay attention to the routing of the electrode connection cables. These must always be in the air and must never lie on surfaces.
Other than Chattanooga Group equipment, keep chip cards, magnetic cards, audio and video cassettes, and other data media susceptible to interference away from the unit.
Clean and disinfect the unit only when the Mains Power Supply is deactivated (power switch off, power plug disconnected).
Clean and disinfect the unit only by means of disinfection by wiping. Disinfecting by spraying can damage the unit due to penetrating moisture.
Never clean the unit with abrasives, disinfectants, or solvents that could scratch the housing or damage the unit.
Never perform unauthorized service work. All service work
must be performed only by service technicians who have been authorized by the manufacturer.
15
ABOUT SHORTWAVE DIATHERMY
Supplied by EME Services Ltd
OVERVIEW
Shortwave refers to electromagnetic radiation in the frequency range 2 to 100 MHz. Shortwave diathermy is the application of electromagnetic energy to the body at shortwave frequencies. At these frequencies, the electromagnetic energy is converted to thermal energy by the induction of circulating currents in the tissue and dielectric absorption in insulating tissue. Shortwave diathermy units may produce varying output power levels providing significant heat to the area of the body being treated. To avoid equipment such as shortwave diathermy units interfering with radio communications, certain frequency ranges are designated by international agreement as ISM (Industrial, Scientific, and Medical) bands.
Shortwave diathermy equipment normally uses the band centered on 27.12 MHz. This corresponds to a wavelength, in a vacuum, of approximately 11 meters.
Shortwave diathermy is normally applied at a level which produces detectable heating and the benefits are those associated with the heating effect (encouragement of healing, pain relief, reduction of muscle spasm, increase in mobility, etc.).
The difference between shortwave diathermy and other methods of heating is that it provides “deep heat.” Other heating techniques such as infrared therapy, hot packs, etc., provide the heat externally whereas shortwave diathermy generates heat within the tissue.
Pulsed Shortwave Diathermy - Introduction
Pulsed shortwave diathermy equipment delivers the energy in pulses or bursts of shortwave energy. The pulses are typically 20 to 400 sec in duration (pulse width) and are repeated with a frequency of 10 to 800 Hz (pulse frequency). As with other modalities such as ultrasound, it is found that delivering the energy in pulses is often therapeutically more beneficial than providing the same amount of energy in continuous waveform. Pulsed shortwave diathermy appears to be effective for many conditions especially in the early stages of recovery.
Because the output is pulsed, the average output power levels can be very low (less than 1 W) and still produce the effective treatment. The Intelect Shortwave 100 in pulsed mode provides a peak power of 200 W and average powers from a few mW to 64 W.
As the power levels are lower than with conventional shortwave diathermy equipment, some of the potential dangers associated with the modality no longer apply.
16
INDICATIONS
Supplied by EME Services Ltd
GENERAL
Shortwave therapy is the application of electromagnetic energy to the body at shortwave frequencies. Shortwave therapy equipment normally uses 27.12 MHz to produce the desired deep heating effect on the tissues of the body. At these frequencies, electromagnetic energy is converted to thermal energy by the induction of circulating currents in the insulating tissue.
The heating effect produced by Shortwave diathermy aids the healing process by generating heat deep within the tissue resulting in numerous beneficial effects.
Shortwave therapy is indicated for the following:
Bursitis
Bursal synovitis
Distortions, Dislocations, and Contusions
Twisted joints, overextended joints, sprains, dislocations (luxation), crushing, injury due to brute force (contusion)
Epicondylitis
Tennis elbow, inflammation of tendon attachments on cubital or radial part of elbow joint (humeral)
Osteoarthritis
Chronic and acute joint disorders (knee, hip, shoulder, elbow, hand, foot and mandibular joints)
Chronic Polyarthritis of the Hip and Shoulder Joints
Inflammation of more than one joint
Tendonitis/Tendinosis
Chronic and acute inflammation or injury to the tendons
Shin Splints
Painful condition of the shins
Sinusitis
Inflammation of the paranasal passages
Fractures
Broken bones
Intercostal Neuralgia
Nerve pain. Acute, painful irritation starting from the thoracic spine; Possible causes of this are nerve root compressions and acute blockages in the area of the kinetic elements or the joint faces of the vertebral bodies
17
INDICATIONS
Supplied by EME Services Ltd
GENERAL CONTINUED
Ischialgia
Pain in ending of nervus ischiadicus, always radicular, usually caused by damaged intervertebral disc
Contracture
Loss of motion in a joint due to the shortening of soft tissue
Lumbago
Muscle pain in the lumbar region, lumbar rheumatism
Myalgia
Muscle pain
Neuralgia/Neuritis
Nerve pain, nerve inflammation
Frozen Shoulder
Shoulder pain accompanied by limitation of movement
Periostitis
Cortical osteitis
Fibrositis/Fibromyalgia
Condition classified by the presence of chronic widespread pain
Spondylosis / Osteochondrosis
Arthrosis of the vertebral bodies or degeneration of the intervertebral discs
Carpal Tunnel Syndrome
Medical condition in which median nerve is compressed at the wrist causing pain and muscle weakness
Tendovaginitis
Inflammation of tendon and sheath; Painful grating or chafing of the affected tendon after overstraining or dull trauma
Cervical syndrome
Post-traumatic neck syndrome; Refers to afflictions beginning in the cervical spine that can emanate into the shoulder muscles or arms
Piriformis Syndrome
Neuromuscular disorder when sciatic nerve is compressed or otherwise irritated by the piriformis muscle
18
INDICATIONS
Supplied by EME Services Ltd
GENERAL CONTINUED
Morton’s Neuroma
Benign neuroma of the interdigital plantar nerve
Plantar Fasciitis
Inflammatory condition of the foot caused by excessive wear to the plantar facia that supports the arch
Rotator Cuff Tear/Repair
Tears of one or more of the four tendons of the rotator cuff muscles
Torticollis
Condition in which head is tilted to one side, chin elevated and turned to opposite side
Plica Syndrome
Occurs when plica in the knee becomes irritated or inflamed
Retropatellar Pain Syndrome
Inflammation of the patellar tendon in the knee
19
CONTRAINDICATIONS
Supplied by EME Services Ltd
GENERAL
The Intelect Shortwave 100 unit is contraindicated for the following:
Any patient with an implanted electronic device such as a cardiac pacemaker, bladder stimulator, spinal cord stimulator or electrodes for a myoelectric prosthesis, or implanted metallic leads, must not be treated with shortwave diathermy and should not be subjected to shortwave diathermy. Do not use on patients who have had an implant in the past unless you are absolutely certain that the implant and all leads in their entirety have been removed. Note that leads are often left implanted after the implant is removed. The effects of the applied high frequency on the pacemaker could cause ventricular fibrillation. Any other persons with pacemakers must also remain outside of the treatment area during shortwave diathermy. No one wearing a cardiac pacemaker should be within 15 meters of an operating unit.
Patients whose condition could be negatively affected by heat.
Patients with hemorrhages or risk of hemorrhage. Patients with septic conditions and empyemas. Patients with malignant tumors and undiagnosed tumors.¹ Implants, areas where implants have been removed,
damaged implants, and metal inclusions.² Implants that could be impaired by shortwave diathermy
irradiation. Swellings that still feel warm.
Thermohypesthesia (diminished perception of temperature differences).
Thermohyperesthesia (very acute thermoesthesia or temperature sense; exaggerated perception of hot and cold).
Acute inflammations. Severe arterial obstructions (stage III and IV). Gynecological disorders involving acute inflammation³. Wetness, perspiration, or damp bandages. Permeating irradiation of the thorax in cases of severe heart
diseases (heart valve diseases, myocardial insufficiency, myocardiac infarct, severe coronary sclerosis).
Pregnancy, since irradiation of the abdomen could cause teratogenous damage due to alterations of blood circulation and diffusion.
During the menstrual cycle. Over the pregnant or potentially pregnant uterus. Therefore,
shortwave diathermy should not be applied over the uterus unless specific assurance can be attained from the patient that she is not pregnant.
Sudeck’s syndrome, stage I and II. Basedow’s disease (irradiation could cause serious states of
agitation). Varicose veins (irradiation could cause congestive pain). Particular care must be taken if the patient’s clothing is wet
20
CONTRAINDICATIONS
Supplied by EME Services Ltd
GENERAL CONTINUED
or damp, since the garments may heat up faster and more intensely than the patient’s body.
Synthetic fibers (perlon, nylon, etc.) are characterized by low
absorbency, which can cause the skin beneath such fabrics to quickly become moist. Therefore, it is recommended that the body areas to be treated be completely unclothed and the patient’s skin dried, particularly where perspiration accumulates in folds of the skin. This applies especially when a higher dosage is being applied. There is no danger, however, when applying shortwave diathermy irradiation to bandaged areas as long as the bandages are completely dry.
When treating small children, particular care is obviously required due to the low body weight. Very careful dosing and constant observation (manual checks of the skin temperature while the unit is switched off) are necessary.
Since the effects of high-frequency fields on unborn life have not yet been sufficiently researched, we recommend that operators who are pregnant are not within 15 meters of the applicator when the unit is activated.
The output power must always be set according to the subjective response of the patient. Therefore, special care must be taken in case of patients with a diminished capacity for perception of heat (refer to "Dosage Levels According to Schliephake on page 82).
It is advisable to post warnings for wearers of pacemakers
in the rooms where high-frequency therapy (e.g. shortwave diathermy) is applied.
A distance of at least 12 meters (see the Caution on page 6 for more information) must be maintained between the unit and any low-frequenc
Cardiac conditions. Deep vein thrombosis, phlebitis, varices. Arterial disease, circulatory insufficiency. Over eyes. Over reproductive organs. Over cardiac pacemakers and defibrillators, cochlear implants,
bone growth stimulators, deep brain stimulators, spinal cord stimulators, and other nerve stimulators.
Over open lamina (after laminectomy; spina bifida). Over superficial endoprosthesis or metal implants. Directly over the carotid sinuses, ceruical stellate ganglion, or
Vagus nerve located in the anterior neck triangle. Direct application over cancerous tumors or lesions due to its
potential to increase blood flow to the area of malignancy. Neoplastic tissues or space occupying lesions. Occlusive vascular disease, such as arteriosclerosis obliterans
and thromboangitis obliterans, in which organic occlusion and ischemia are evident.
Directly over the epiphysis of growing bones in children
y therapy that is being used.
21
CONTRAINDICATIONS
Supplied by EME Services Ltd
GENERAL CONTINUED
and adolescents because shortwave diathermy therapy may enhance or inhibit bone growth.
NOTE: The mean age for skeletal maturity is 15 ½ years in
females and 17 ½ years in males.
In the presence of systemic or local infection (sepsis, Osteomyelitis, tuberculosis) or if the patient has an elevated temperature.
In areas where metal is present due to eddy current generation of excessive and uneven heat distribution. Metal objects within the treatment area should be removed and placed outside the electromagnetic field. These include, but are not limited to metal:
Metal in the environment:
Within 0.3 m of beds, treatment tables, standard chairs, wheelchairs, swivel stools, step stools, splints, braces, scissors, forceps, and scalpels.
Within 12 meters (see the Caution on page 6 for more information) of electronically controlled medical devices such as, CPM devices, electric wheelchairs, electrotherapy devices or other electrical systems, computers, etc.
Metal near the patient:
Jewelry, body piercing earrings, watches, keys, coins, belt buckles, underwire bra, hearing aids, zipper in
clothing or pillow cases.
Metal objects within the treatment area that cannot be removed should be avoided. These include, but are not limited to:
External metal: orthodontic braces, dental fillings, staples, external fixation devices.
Internal metal: valves, joint replacements, metal IUDs, shrapnel, metal implants, internal fixation devices-rods, plates, screws, wires, etc.
NOTE: lf there is a scar in or near the treatment area,
check with the patient and/or the patient's chart to determine if there is metal under the scar.
¹According to Schneider (in Elektromedizin 7/62): Tissue and organ sections with inflammations, necroses, pus formation and abscesses. In such cases, the therapist must choose between the application of cold or heat in accordance with general pathological considerations, depending on the degree of inflammation. Inflammatory conditions that are still in statu nascendi are treated with cold. Inflammatory conditions with necroses and a cavitary tendency are treated with therapeutic means that generate heat and hyperemia. Chronic and unspecific inflammations are treated in the same way (heat and hyperemia), as this supports resorption, reparation and regeneration. Specific chronic inflammations (such as tuberculosis), however, are activated by heat. Accordingly, they represent a contraindication. The same applies in the case of malignant tumorous conditions. Heat application in the case of a tumorous disease can only be regarded as malpractice. Moreover, cardiac congestions must be removed prior to any heat application.
22
CONTRAINDICATIONS
Supplied by EME Services Ltd
GENERAL CONTINUED
²The higher conductivity of metals causes concentration of the fi eld, producing a high temperature in the border area of the tissue. This, in turn, can cause excessive local heat, leading to (irreparable) third-degree burns. Therefore, caution is also necessary in case of long-existing metal inclusions, such as shell fragments.
³Further contraindications relating to gynecological disorders include (see Möbius, Gynecological University Clinic, Jena): genital tuberculosis, endometriosis, pyosalpinx or pyo­ovarium, tubal carcinoma.
23
ADDITIONAL PRECAUTIONS
Supplied by EME Services Ltd
GENERAL
When administering shortwave diathermy, keep in mind the following:
Caution is advised in patients who cannot perceive or report pain or heat sensation accurately. Absent or diminished sensation should be avoided or, if unavoidable, treated with caution. Establishment of acceptable intensity levels for desensitized areas may be related to the intensity levels tolerated on normal skin in opposite or related body parts. Frequent monitoring of intensity level and skin response should occur during all treatments.
Caution is suggested when using shortwave diathermy immediately after the application of superficial and deep heat or cold modalities. The application of thermal modalities prior to shortwave diathermy can alter the patient's perception of warmth and pain.
Caution is advised when applying thermal shortwave diathermy directly to an area with impaired arterial blood supply because the compromised blood flow may not meet the increased metabolic demand placed on the tissues by the thermal energy dose. Always start with a low dose and observe patient response. Increase in dose may be made in subsequent treatments if the patient can tolerate the dosage given.
Caution should be exercised when treating over adipose tissue. Excessive superficial heating can occur due to the
high electrical resistance of subcutaneous fat to shortwave diathermy.
Shortwave diathermy should be applied with caution over bone where minimal (bony prominence) or no (Stage IV wounds) soft tissue is present.
Hearing aids should be removed. The function of other patient connected equipment may be
adversely affected by the operation of the pulsed shortwave diathermy equipment. Maintain maximum distance between units in order to reduce any tendency to interaction.
Any bleeding tendency is increased by heating because of the increase in blood flow and vascularity of the heated tissues. Care, therefore, should be used in treating patients with therapeutic shortwave diathermy who have bleeding disorders.
Heating of the joint capsule in acute or subacute arthritis should be avoided.
Use a single layer of toweling to absorb moisture during treatment with the inductive drum applicators.
Shortwave diathermy may interfere with other electronic therapeutic devices such as neuromuscular stimulators and therapeutic ultrasound units. Never use another electronic device on the same patient when shortwave diathermy is being applied.
24
ADDITIONAL PRECAUTIONS
Supplied by EME Services Ltd
GENERAL CONTINUED
Use caution when treating obese patients with capacitive electrodes since this method of application may heat fat excessively.
Frequent monitoring of intensity level and skin response should occur during all treatments.
25
SETUP
Supplied by EME Services Ltd
UNPACKING THE UNIT
The unit is generally delivered with the packaging material supplied by the manufacturer. Since the unit weighs approximately 60 lbs (27.22 kg), it must be unpacked by at least 2 persons.
Proceed as follows:
Position the transport packaging so that the arrows are pointing upward.
Remove the safety bands from the transport packaging. Remove the transport packaging upward. Remove the remaining foam material. Using at least 2 people, lift the unit from the lower packaging
element.
Inspection
Immediately upon unpacking the unit, perform the following steps:
Verify the delivery documents to make sure that the delivery is complete.
Check the external components and accessories for possible damage due to transport.
Verify that the packaging contains the unit, applicators, power cord, User Manual, Quick Start Guide, and indicator discharge tube.
In case of damage from transport that could endanger personal safety, the unit must not be connected to the Mains Power Supply before inspection is complete.
26
Loading...
+ 68 hidden pages