This manual is intended for users of Vectra® Neo Clinical
Therapy System. It contains general information on
operation, precautionary practices, and maintenance.
In order to maximize use, e ciency, and the life of the
system, please read this manual thoroughly and become
familiar with the controls, as well as the accessories before
operating the system.
In addition to the above information, this manual
contains care and installation instructions for the optional
Cart, Channel 1/2 Electrotherapy module, Channel 1/2
Electrotherapy module + sEMGmodule, Channel 3/4
Electrotherapy module, Laser module, and Ultrasound
module for the users of the Vectra® Neo Clinical Therapy
System.
Speci cations put forth in this manual were in e ect at
the time of publication. However, owing to DJO’s policy of
continual improvement, changes to these speci cations
may be made at any time without noti cation on the part
of DJO.
Before administering any treatment to a patient, the
users of this equipment should read, understand, and
follow the information contained in this manual for each
mode of treatment available, as well as the indications,
contraindications, warnings, and precautions. Consult
other resources for additional information regarding the
application of electrotherapy, ultrasound, and laser.
PRECAUTIONARY INSTRUCTIONS
The precautionary instructions found in this section and
throughout this manual are indicated by speci c symbols.
Understand these symbols and their de nitions before
operating this equipment. The de nition of these symbols
are as follows:
CAUTION
Text with a “CAUTION” indicator explains possible
safety infractions that have potential to cause minor or
moderate injury or damage to the equipment.
WARNING
Text with a “WARNING” indicator explains possible safety
infractions that will potentially cause serious injury and
equipment damage.
DANGER
Text with a “DANGER” indicator will explain possible safety
infractions that are imminently hazardous situations that
would result in death or serious injury.
Text with a “DANGEROUS VOLTAGE” indicator
serves to inform the user of possible hazards
resulting in the electrical charge delivered to
the patient in certain treatment con gurations
of TENS waveforms.
Warning; Laser beam
Explosion Hazard - Text with an “Explosion
Hazard” indicator will explain possible safety
infractions if this equipment is used in the
presence of ammable anesthetics, mixture
with air, oxygen, or nitrous oxide.
Wear eye protection
NOTE: Throughout this manual, “NOTE”
indicators provide helpful information
regarding the particular area of function being
described.
4
Page 7
Vectra® Neo Clinical Therapy System
INTRODUCTION
GENERAL TERMINOLOGY
The following are de nitions for the terminology used throughout this manual. Study these terms to become familiar
with them for ease of system operation and control functionality of the Vectra® Neo Clinical Therapy System.
SYSTEM SOFTWARE SYMBOLS
Back Arrow
Home
Electrode Placement
Increase/Decrease Parameter
Stim
Ultrasound
Combo
Scroll Up or Down in a text box
sEMG
Select
Page Up
Laser
CPS
Page Down
Customize
Save Data
When pressed, will print the screen
contents or Patient Treatment Results
Report to the USB ash drive
Indicates a USB ash drive is inserted
Patient Remote/Laser Shut O Icon
Indicates the Remote is plugged in
Custom Protocols
Patient Data
Anatomical Library
5
Page 8
INTRODUCTION
DESCRIPTION OF DEVICE MARKINGS
The markings on the unit are assurance of its conformity to the highest
applicable standards of medical equipment safety and electromagnetic
compatibility. One or more of the following markings may appear on the
device:
This unit is considered to be a Class 3B laser product and thus emits visible
and invisible laser radiation (IR). Avoid direct eye exposure to the Laser beam.
The symbol to the right is located on the back of the applicator and indicates
the active radiant surface (the area on the applicator that emits infrared laser
energy and the direction of the beam of light) . . . . . . . . . . . . . . . . . . . . . . . . . . .
For VMS (Pulsed Mode, Burst Mode, or FR Mode),
Russian, Monophasic Hi-Volt (NMES) & Interferential, and
Premodulated (IFS):
• Relaxation of muscle spasms
• Prevention or retardation of disuse atrophy
• Increasing local blood circulation
• Muscle re-education
• Maintaining or increasing range of motion
• Immediate postsurgical stimulation of calf muscles to
prevent venous thrombosis
Additionally for Microcurrent, Interferential, Premodulated
(IFS), VMS (Pulsed Mode, Burst Mode, or FR Mode),
Asymmetrical Biphasic (TENS), and HANS:
• Symptomatic relief or management of chronic,
intractable pain
• Post-traumatic acute pain
• Post-surgical acute pain
For DC (Direct Current) Continuous Mode:
• Relaxation of muscle spasm
For FES:
• Stimulation of the muscles of the leg and ankle of
partially paralyzed patients to provide flexion of the
foot and thus improve the patient’s gait.
Contraindications
The Vectra® Neo Clinical Therapy System should NOT be
used under the following conditions:
• Do not use for symptomatic local pain relief unless
etiology is established or unless a pain syndrome has
been diagnosed.
• Do not use when cancerous lesions are present in the
treatment area.
• Do not apply stimulation over swollen, infected,
inflamed areas or skin eruptions (e.g., phlebitis,
thrombophlebitis, varicose veins, etc.).
• Do not use when patient is suspected or known to
have infectious disease and/or disease where it is
advisable, for general medical purposes, to suppress
heat or fevers.
• Do not place electrode placements to the carotid
sinus region (anterior neck) or transcerebrally (through
the head).
• Do not use on pregnant women. Safety has not
been established for the use of therapeutic electrical
stimulation during pregnancy.
• Do not use powered muscle stimulators or TENS
waveforms on patients with cardiac demand
pacemakers.
• Do not use Vectra® Neo Clinical Therapy System
on patients who have or have had implantable
neurostimulating cardiac demand pacemakers, ICD, or
other implantable electronic devices.
• Do not use Vectra® Neo Clinical Therapy System on
patients with body worn electro mechanical medical
devices, i.e. insulin pump.
• Do not use this system in an MRI or CT environment.
The Vectra® Neo Clinical Therapy System, its
components, and accessories are not to be present in
an MRI or CT environment.
Additional Precautions
• Use caution for patients with suspected or diagnosed
heart problems.
• Use caution for patients with suspected or diagnosed
epilepsy.
• Use caution in the presence of the following:
- When there is a tendency to hemorrhage
following acute trauma or fracture
- Following recent surgical procedures when
muscle contraction may disrupt the healing
process
- Over a menstruating or pregnant uterus
- Over areas of the skin that lack normal sensation
• Some patients may experience skin irritation or
hypersensitivity due to the electrical stimulation
or electrical conductive medium. The irritation can
usually be reduced by using an alternative conductive
medium or an alternative electrode placement.
• Electrode placement and stimulation settings
should be based on the guidance of the prescribing
practitioner.
• Powered muscle stimulators should be used only with
the lead wires and electrodes recommended for use
by the manufacturer.
• With TENS waveforms, isolated cases of skin irritation
may occur at the site of electrode placement following
long-term application.
• The effective management of pain by TENS waveforms
is highly dependent upon patient selection by a
person qualified in pain management.
Adverse E ects
• Skin irritation and burns beneath the electrodes
have been reported with the use of powered muscle
stimulators.
• Potential adverse effects with TENS are skin irritation
and electrode burns.
7
Page 10
INDICATIONS FOR USE
SEMG & STIM INDICATIONS
Vectra® Neo Clinical Therapy System
Indications
For EMG triggered Stim:
• Stroke rehab by muscle re-education
• Relaxation of muscle spasms
• Prevention or retardation of disuse atrophy
• Increase local blood circulation
• Muscle re-education
• Maintaining or increasing range of motion
For EMG:
• To determine the activation timing of muscles for:
- Retraining of muscle activation
- Coordinating of muscle activation
• Any indication of the force produced by muscle for
control and maintenance of muscle contractions
- Relaxation muscle training
- Muscle re-education
Contraindications
The Vectra® Neo Clinical Therapy System should not be
used under the following conditions:
• Do not use for symptomatic local pain relief unless
etiology is established or unless a pain syndrome has
been diagnosed.
• Do not use when cancerous lesions are present in the
treatment area.
• Do not apply stimulation over swollen, infected,
inflamed areas or skin eruptions (e.g., phlebitis,
thrombophlebitis, varicose veins, etc.).
• Other contraindications are patients suspected of
carrying serious infectious disease and/or disease
where it is advisable, for general medical purposes, to
suppress heat or fevers.
• Do not place electrode placements to the carotid
sinus region (anterior neck) or transcerebrally (through
the head).
• Safety has not been established for the use of
therapeutic electrical stimulation during pregnancy.
• Do not use powered muscle stimulators or TENS
waveforms on patients with cardiac demand
pacemakers.
• Do not use Vectra® Neo Clinical Therapy System
on patients who have or have had implantable
neurostimulating cardiac demand pacemakers, ICD, or
other implantable electronic devices.
• Do not use Vectra® Neo Clinical Therapy System on
patients with body worn electro mechanical medical
devices, i.e. insulin pump.
• Do not use this system in an MRI or CT environment.
The Vectra® Neo Clinical Therapy System, its
components, and accessories are not to be present in
an MRI or CT environment.
Additional Precautions
• Use caution for patients with suspected or diagnosed
heart problems.
• Use caution for patients with suspected or diagnosed
epilepsy.
• Use caution in the presence of the following:
- When there is a tendency to hemorrhage
following acute trauma or fracture
- Following recent surgical procedures when
muscle contraction may disrupt the healing
process
- Over a menstruating or pregnant uterus
- Over areas of the skin that lack normal sensation
• Some patients may experience skin irritation or
hypersensitivity due to the electrical stimulation
or electrical conductive medium. The irritation can
usually be reduced by using an alternative conductive
medium or an alternative electrode placement.
• Electrode placement and stimulation settings
should be based on the guidance of the prescribing
practitioner.
• Powered muscle stimulators should be used only with
the lead wires and electrodes recommended for use
by the manufacturer.
• With TENS waveforms, isolated cases of skin irritation
may occur at the site of electrode placement following
long term application.
• The effective management of pain by TENS waveforms
is highly dependent upon patient selection by a
person qualified in the management of pain patients.
Adverse E ects
• Skin irritation and burns beneath the electrodes
have been reported with the use of powered muscle
stimulators.
• Potential adverse effects with TENS are skin irritation
and electrode burns.
8
Page 11
Vectra® Neo Clinical Therapy System
ULTRASOUND INDICATIONS
INDICATIONS FOR USE
Indications
Application of therapeutic deep heat for the treatment of
selected sub-chronic and chronic medical conditions such
as:
• Relief of pain, muscle spasms, and joint contractures
• Relief of pain, muscle spasms, and joint contractures
that may be associated with:
- Adhesive capsulitis
- Bursitis with slight calcification
- Myositis
- Soft tissue injuries
- Shortened tendons due to past injuries and scar
tissues
• Relief of sub-chronic and chronic pain and joint
contractures resulting from:
- Capsular tightness
- Capsular scarring
Contraindications
• This device should not be used for symptomatic local
pain relief unless etiology is established or unless a
pain syndrome has been diagnosed.
• This device should not be used when cancerous
lesions are present in the treatment area.
• Other contraindications are patients suspected
of carrying serious infectious disease and disease
where it is advisable for general medical purposes to
suppress heat or fevers.
• This device should not be used over or near bone
growth centers until bone growth is complete.
• This device should not be used over the thoracic area
if the patient is using a cardiac pacemaker.
• This device should not be used over a healing fracture.
• This device should not be used over or applied to the
eye.
• This device should not be used over a pregnant uterus.
• Tissue necrosis might result if the device is used on
ischemic tissues in individuals with vascular disease,
where the blood supply would not keep up with the
metabolic demand.
• Do not use Vectra® Neo Clinical Therapy System
on patients who have or have had implantable
neurostimulating cardiac demand pacemakers, ICD, or
other implantable electronic devices.
• Do not use Vectra® Neo Clinical Therapy System on
patients with body worn electro mechanical medical
devices, i.e. insulin pump.
• Do not use this system in an MRI or CT environment.
The Vectra® Neo Clinical Therapy System, its
components, and accessories are not to be present in
an MRI or CT environment.
Additional Precautions
Additional precautions should be used when ultrasound
is used on patients with the following conditions:
• Over an area of the spinal cord following a
laminectomy, i.e., when major covering tissues have
been removed
• Over anesthetic areas
• On patients with hemorrhagic diatheses
9
Page 12
INDICATIONS FOR USE
LASER INDICATIONS
Vectra® Neo Clinical Therapy System
Indications
To provide topical heating for the following:
• Temporary increase in local blood circulation
• Temporary relief of minor muscle and joint aches,
pains, and stiffness
• Relaxation of muscles
• Temporary relief of muscle spasms
• Temporarty relief of minor pain and stiffness
associated with arthritis
Contraindications
The Vectra® Neo Clinical Therapy System Laser should NOT
be used:
• Where analgesia may mask progressive pathology,
and where the practitioner would normally avoid
the use of any other analgesia in order to retain the
beneficial aspects of pain
• For direct aim into the eyes of humans over areas
injected with steroids in the past 2-3 weeks
• Over areas that are suspicious or contain potentially
cancerous tissue
• Over areas of active hemorrhage
• Over a pregnant uterus
• Over the neck (thyroid or carotid sinus region) or chest
(vagus nerve or cardiac region of the thorax)
• Directly over areas with open wounds, unless covered
with a clear protective barrier
• Treatment over sympathetic ganglia
• For symptomatic local pain relief unless etiology
is established or unless a pain syndrome has been
diagnosed
• On patients suspected of carrying serious infectious
disease and/or disease where it is advisable, for
general medical purposes, to suppress heat or fevers
• Over or near bone growth centers until bone growth is
complete
• Over the thoracic area if the patient is using a cardiac
pacemaker
• Over or applied to the eye
• On ischemic tissues in individuals with vascular
disease where the blood supply would be unable to
follow the increase in metabolic demand and tissue
necrosis might result
• Do not use Vectra® Neo Clinical Therapy System
on patients who have or have had implantable
neurostimulating cardiac demand pacemakers, ICD, or
other implantable electronic devices.
• Do not use Vectra® Neo Clinical Therapy System on
patients with body worn electro mechanical medical
devices, i.e. insulin pump.
• Do not use this system in an MRI or CT environment.
The Vectra® Neo Clinical Therapy System, its
components and accessories, are not to be present in
an MRI or CT environment.
Additional Precautions
Additional precaution should be used when the Laser is
used on patients with the following conditions:
• Over an area of the spinal cord following a
laminectomy, i.e., when major covering tissues have
been removed
• Over anesthetic areas
• On patients with hemorrhagic diatheses
Preventing Adverse E ects
Perform the following procedures to avoid the negative
e ects of Laser therapy:
• Inspect the treatment area during and following
treatment, and discontinue if an adverse reaction does
occur.
• Higher output levels have a greater potential for
patient discomfort. Choose a lower dosage to reduce
output or select a pulsed duty cycle to decrease
patient discomfort.
Factors that A ect Treatment
The following factors may a ect Laser treatment:
• Color of skin (light or dark)
• Age of lesion
• Depth of lesion
• Sensitivity of patient
• Type of tissue
• Medications that increase sensitivity to light
10
Page 13
Vectra® Neo Clinical Therapy System
PRODUCT DESCRIPTION
The Vectra® Neo Clinical Therapy System is a modular
system used with or without an optional Cart, allowing for
the inclusion of Channel 1/2 Electrotherapy module with
or without sEMG, Channel 3/4 Electrotherapy module,
Laser module, and Ultrasound module.
To maximize functionality and life of Vectra® Neo, be sure
to:
• Stay current with the latest clinical developments in
the field of electrotherapy, ultrasound, laser therapy,
sEMG, and sEMG + electrotherapy.
• Observe all applicable precautionary measures for
treatment.
• Keep informed of appropriate indications and
contraindications for the use of the Vectra® Neo
Clinical Therapy System.
NOTE: This equipment is to be used only under the
prescription and supervision of a licensed medical
practitioner.
DEVICE DESCRIPTION
11
Page 14
DEVICE DESCRIPTION
OPERATOR INTERFACE
Vectra® Neo Clinical Therapy System
The Vectra® Neo Clinical Therapy System Operator
Interface contains all the functions and controls necessary
for operator access to all operator utilities, modalities, and
parameters for modi cation and system set up.
1. Color Display
2. Intensity Dial (Gray outer ring)
3. Start/Pause button
4. Stop button
5. ON/OFF switch
6. Ultrasound Applicator holder, left and right sides
7. Laser Applicator holder, left and right sides
8. Patient Remote/Laser Interrupt Switch port
9. Mains Power Cord
10. Rear Access Panel
11. Serial Label
12. USB Flash Drive Port (USB flash drive not included)
13. Tilt Screen
14. Swivel function
15. Laser Interlock Port and Icon
Front Controls
1
2
4
3
Rear Access Panel
12
5
16. Lead wire holders
Side Holders
16
45
11
6,7
10
8
6,7
º
90
º
13
14
15
9
12
Page 15
Vectra® Neo Clinical Therapy System
GENERAL WARNINGS AND PRECAUTIONS
CAUTION
• Read, understand, and practice the precautionary and operating
instructions. Know the limitations and hazards associated with using
any electrical stimulation, Laser device, or ultrasound device. Observe
the precautionary and operational decals placed on the unit.
• All modalities should be routinely checked before each use to
determine that all controls function normally, especially that the
intensity control does properly adjust the intensity of the ultrasonic
power output in a stable manner. Also, determine that the treatment
time control does actually terminate ultrasonic power output when
the timer reaches zero.
• DO NOT use sharp objects such as a pencil point or ballpoint pen to
operate the buttons on the control panel.
• This unit should be operated at 10°C to 45°C and 0% to 90% relative
humidity. The unit should be transported and stored at 0°C to 60°C and
0% to 95% relative humidity.
• Handle Ultrasound Applicator and Laser Applicator with care.
Inappropriate handling may adversely affect its characteristics.
• Before each use, inspect Ultrasound Applicator for cracks, which may
allow the ingress of conductive fluid.
CAUTION
• Failure to use and maintain the Vectra® Neo Clinical Therapy System,
its modules, and its accessories in accordance with the instructions
outlined in this manual will invalidate the warranty.
• DO NOT permit foreign materials, liquids, or cleaning agents (including,
but not limited to, inflammables, water, and metallic objects) to enter
the unit to prevent unit damage, malfunction, electrical shock, fire, or
personal injury.
• If you have difficulty operating the unit after carefully reviewing this
user manual, contact your DJO dealer for assistance.
• DO NOT remove the cover. Doing so may cause unit damage,
malfunction, electrical shock, fire, or personal injury. There are no
user-serviceable parts inside the unit. If a malfunction occurs,
discontinue use immediately and consult dealer for repair service.
• Use of parts or materials other than DJO’s can degrade minimum
safety.
• The Vectra® Neo Clinical Therapy System is not designed to prevent the
ingress of water or liquids. Ingress of water or liquids could cause
malfunction of internal components of the system and therefore
create a risk of injury to the patient.
• Inspect Applicator cables and associated connectors before each use.
• Device is designed to comply with electromagnetic safety standards.
This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with instructions,
may cause harmful interference to other devices in the vicinity.
However, there is no guarantee that interference will not occur in a
particular installation. Harmful interference to other devices can be
determined by turning this equipment on and off. Try to correct the
interference using one or more of the following:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment to an outlet on a different circuit from that
to which the other device(s) are connected and consult the factory
field service technician for help.
- Consult your authorized DJO dealer for help.
• Do not operate this unit when connected to any unit other than DJO
devices or accessories specifically described in user or service manuals.
• Use of controls, adjustments or performance of procedures other than
those specified herein may result in hazardous exposure to Laser
energy.
• DO NOT operate the Vectra® Neo Clinical Therapy System within the
vicinity or environment as any microware and RF shortwave diathermy
system.
• DO NOT operate the Vectra® Neo Clinical Therapy System within the
vicinity or environment as an ultrasonic diathermy system. The
Ultrasound (diathermy) Module of the Vectra® Neo Clinical Therapy
System does not require separation distance.
• DO NOT use electrodes with an active area less than 7.92 cm
2
, as there
will be a risk of suffering a burn injury. Always exercise caution with
current densities more than 2mA/cm
2
. Refer to the CURRENT DENSIT
CALCULATIONS FOR DJO ELECTRODES table in Appendix 3.
• DO NOT disassemble, modify, or remodel the unit or accessories. This
may cause unit damage, malfunction, electrical shock, fire, or personal
injury.
13
Page 16
GENERAL WARNINGS AND PRECAUTIONS
Vectra® Neo Clinical Therapy System
WARNING
• U.S.A. Federal Law restricts these devices to sale by, or on the order of,
a physician or licensed practitioner. This device should be used only
under the continued supervision of a physician or licensed
practitioner.
• Be sure to read all instructions for operation before treating patient.
• Make certain the unit is electrically grounded by connecting only to a
grounded electrical service receptacle conforming to the applicable
national and local electrical codes.
• Care must be taken when operating this equipment around other
equipment. Potential electromagnetic or other interference could
occur to this or to the other equipment. Try to minimize this
interference by not using other equipment in conjunction with it.
• The safety of TENS waveforms for use during pregnancy or birth has
not been established.
• TENS is not effective for pain of central origin. (This includes
headache.)
• TENS waveforms have no curative value.
• Electronic monitoring equipment (such as ECG monitors and ECG
alarms) may not operate properly when electrical stimulation is in use.
• TENS is a symptomatic treatment, and as such, suppresses the
sensation of pain which would otherwise serve as a protective
mechanism.
• Inspect the plastic lens of the laser head for blemishes, deformation,
pitting, scratches, discoloration, and cleanliness before each use.
• Do not drop the applicator or unit on hard surfaces or submerge in
water. These actions will damage the applicator and unit. Damage
resulting from these conditions is not covered under the warranty.
• Use of controls or adjustments or performance of procedures other
than those specified herein may result in hazardous exposure to Laser
energy.
• This device should be kept out of the reach of children.
WARNING
• Do not treat through clothing.
• Stop treatment immediately if patient experiences discomfort or pain.
• Do not apply laser on an area of skin that has lotion or ointments
applied as burns may occur.
• Do not use laser on or over a tattoo.
• The laser head must be cleaned with a disinfectant cleaner
(i.e. Virex® II 256) or germicidal cloth (i.e. PDI Sani-Cloth® Plus/Hb)
between each therapy session. Ensure no liquids enter into the laser
head while cleaning. Do not use any chlorine-based cleaners on the
laser head.
• The color of skin, age of lesion, depth of lesion, sensitivity of the
patient, tissue type, and medications that increase sensitivity to light
may affect therapy.
• Powered muscle stimulators should be used only with the leads and
electrodes recommended for use by the manufacturer.
• In the event of all 300-Level or a 200-Level error message that cannot
be resolved, immediately stop all use of the system, and contact the
dealer or DJO for service. Errors and Warnings in these categories
indicate an internal problem with the system that must be tested by
DJO or a Trained Technician before any further operation or use of the
system.
- Use of a system that indicates an Error or Warning in these
categories may pose a risk of injury to the patient, user, or
extensive internal damage to the system.
• Use of controls or adjustments or performance of procedures other
than those specified herein may result in hazardous exposure to
ultrasonic energy.
• Before administering any treatment to a patient you should become
acquainted with the operating procedures for each mode of treatment
available, as well as the indications, contraindications, warnings, and
precautions. Consult other resources for additional information
regarding the application of each mode of treatment.
• Use of other accessories other than those specified in this User Manual
may increase electrical emissions and decrease electrical immunity of
the device.
• Contaminated sponges, electrodes, lead wires, and gel can lead to
infection.
• Use of electrode with degraded hydrogel can result in burn to the skin.
• DO NOT operate this unit in an environment where other devices are
being used that intentionally radiate electromagnetic energy in an
unshielded manner.
• Use of electrode on multiple patients can lead to infection.
• Clean applicators after each use, otherwise it can lead to cross
contamination and infection.
• When the Laser Module is not in use, it should be protected against
unqualified use.
• Disconnect the system from the power source before attempting any
maintenance, installation, removal, or replacement procedures to
prevent electrical shock and possible damage to system.
• Keep electrodes separated during treatment. Electrodes in contact
with each other could result in improper stimulation or skin burns.
14
Page 17
Vectra® Neo Clinical Therapy System
GENERAL WARNINGS AND PRECAUTIONS
WARNING
• The Vectra® Neo Clinical Therapy System may be susceptible to ElectroStatic Discharge (ESD) at greater than ±4 kV when first grasping either
the Ultrasound or Laser Applicator. In the event of such a discharge,
the Vectra® Neo Clinical Therapy System may experience
communication loss with the installed modules. The Vectra® Neo
Clinical Therapy System will terminate all active outputs (stim,
ultrasound, laser), automatically place the unit in a safe state, and
issue an error message 301 or 307.
- To recover from an error message 301 or 307, turn the unit off and
on using the ON/OFF switch located at the top of the display. Once
the system restarts, re-initiate all treatments that were
interrupted.
• To prevent Electro-Static Discharge (ESD) at greater than ±4 kV:
- Grasp and hold the Ultrasound or Laser Applicator prior to starting
treatment. If the applicator must be put down prior to completion
of treatment, stop the current treatment first and then place the
applicator in the holder.
- Maintain humidity in the use environment to at least 50% relative
humidity.
- Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, DJO recommends implementing
additional controls to maintain relative humidity to at least 50%.
- Communicate these ESD-precautionary procedures to healthcare
staff, contractors, visitors, and patients.
• Long term effects of chronic electrical stimulation are unknown.
WARNING
• Output current density is related to electrode size. Improper
application may result in patient injury. If any question arises as to the
proper electrode size, consult a licensed practitioner prior to therapy
session.
• Do not apply the Ultrasound Applicator to the patient during the Head
Warming period. Applicator must remain in Applicator Hook during the
Head Warming period.
• Some patients are more sensitive to Laser output (i.e., patients taking
medications that increase sensitivity to light) and may experience a
reaction similar to a heat rash.
• Before each Laser use, clean the plastic lens with a clean cloth. Make
certain to apply with a clean cloth. Failure to clean the lens between
patient therapy sessions could cause beam fragmentation, which may
reduce the effectiveness of the treatment.
• Medical electrical equipment needs special precautions regarding EMC.
Portable and mobile RF communication equipment can be affected by
other medical electrical devices. If you believe interference is
occurring, please consult the ELECTROMAGNETIC COMPATIBILITY (EMC)
section on page 73, to assist in removing the interference.
• Common RF emitting devices (e.g., RFID) and electromagnetic security
systems (e.g., metal detectors) may interfere with the operation of the
Vectra® Neo Clinical Therapy System. The Vectra® Neo Clinical Therapy
System has been tested in the presence of these types of devices and
while no adverse event occurred, the device should not be operated
within the vicinity or environment as another RF emitting device.
• Stimulation should not be applied over the anterior neck or mouth.
Severe spasm of the laryngeal and pharyngeal muscles may occur and
the contractions may be strong enough to close the airway or cause
difficulty in breathing.
• Stimulation should not be applied transthoracically in that the
introduction of electrical current into the heart may cause cardiac
arrhythmia.
• Stimulation should not be applied over swollen, infected, and
inflamed areas or skin eruptions, e.g., phlebitis, thrombophlebitis,
varicose veins, etc.
• Stimulation should not be applied over, or in proximity
to, cancerous lesions.
• Electrotherapy output current density is related to electrode size.
Improper application may result in patient injury. If any question
arises as to the proper electrode size, consult a licensed practitioner
prior to therapy session.
• The Vectra® Neo Clinical Therapy System optional modules and
associated accessories are designed for use only with the Vectra® Neo
Clinical Therapy System.
• Remove the Ultrasound or Laser Applicator by pulling the cable
connector only. DO NOT remove by pulling the cable.
15
Page 18
GENERAL WARNINGS AND PRECAUTIONS
DANGER
• Stimulus delivered by the TENS waveforms of this device,
in certain configurations, will deliver a charge of
25 microcoulombs (μC) or greater per pulse and may be
sufficient to cause electrocution. Electrical current of this
magnitude must not flow through the thorax because it
may cause a cardiac arrhythmia.
• Patients with an implanted neurostimulation device must
not be treated with or be in close proximity to any
shortwave diathermy, therapeutic ultrasound diathermy,
or laser diathermy anywhere on their body. Energy from
diathermy (shortwave, microwave, ultrasound, and laser)
can be transferred through the implanted
neurostimulation system, can cause tissue damage, and
can result in severe injury or death. Injury, damage, or
death can occur during diathermy therapy even if the
implanted neurostimulation system is turned off.
• Handle, clean, and dispose of components and accessories
that have come in contact with bodily fluids according to
National, Local, and Facility rules, regulations, and
procedures.
Vectra® Neo Clinical Therapy System
• This unit is considered to be a Class 3B Laser product and
thus emits visible and invisible Laser radiation (IR). Avoid
direct eye exposure to the Laser beam. The symbol to the
left is located on the back of the applicator and indicates
the active radiant surface (the area on the applicator that
emits infrared Laser energy and the direction of the beam
of light). When the unit is on, not all wavelengths are
visible to the naked eye. Therefore, when performing any
operational or functional check, always wear Chattanooga
laser protective eyewear.
• The solvents of adhesives and flammable solutions used
for cleaning and disinfecting should be allowed to
evaporate before the unit is used.
• DO NOT connect the unit to an electrical supply without
first verifying that the power supply is the correct voltage.
Incorrect voltage may cause unit damage, malfunction,
electrical shock, fire, or personal injury. Your unit was
constructed to operate only on the electrical voltage
specified on the Voltage Rating and Serial Number Plate.
Contact your DJO dealer if the unit is not properly rated.
• Laser protective eyewear should be worn during laser
treatment by the operator and patient to block infrared
light energy from the eyes during treatment.
• DO NOT point the Laser beam directly into human or
animal eyes. The lens of the eye does not detect the
invisible, coherent Laser beams, potentially resulting in
permanent retinal damage.
• Device is not designed to be used in oxygen rich
environment. Explosion hazard if the device is used in the
presence of flammable anesthetic mixture with air,
oxygen, or nitrous oxide.
16
Page 19
Vectra® Neo Clinical Therapy System
COMPONENTS
DETAIL DEVICE DESCRIPTION
Throughout these instructions the terms “left” and “right”
referring to the machine sides are from the perspective of
a user standing in front of the unit.
The Vectra® Neo Clinical Therapy System allows
installation of optional modality modules by the user.
Speci cally designed for use with the Vectra® Neo Clinical
Therapy System, these modules con gure the system to
meet virtually every therapeutic need that a clinician may
have. The components of the Vectra® Neo Clinical Therapy
System are shown below.
NOTE: The Vectra® Neo Clinical Therapy System, when
ordered as a Tabletop System, without cart, is assembled
with Base (shown below). The only assembly required is
the installation of the desired Modules described in the
MODULE INSTALLATION section on page 23.
Head
Modules
• Stimulation Channel 1/2
• Stimulation Channel 1/2 + sEMG
• Stimulation Channel 3/4
• Laser
• Ultrasound
Lead Wires
The available lead wires are shown below. If the user
orders Stimulation Channel 1/2 module, the box will
include the blue and green lead wires. Stimulation
Channel 3/4 is the cranberry and orange lead wires.
If both modules are ordered, the box contains all four
colored lead wires. Stimulation modules channel 1/2 with
sEMG includes blue and green sEMG lead wires.
• Stimulation module (2 channel Stimulation with
sEMG)
• sEMG Lead Wires
• DURA-STICK® 2 in (5 cm) Round Disposable Electrodes
(2 packs of 4)
• Faceplates (to cover module after inserted into main
unit)
Laser Module – PN 70005
• Laser module
• Protective Eyewear, 2 Pair
• Interlock
• Patient Remote/Laser Interrupt Switch
• Faceplates (to cover module after inserted into main
unit)
18
Page 21
Vectra® Neo Clinical Therapy System
ULTRASOUND APPLICATOR
1. Applicator Head
The component of the applicator that makes
contact with the patient during Ultrasound or
1
Combination therapy.
2. Applicator
The assembly that connects to the system and
incorporates the Applicator.
3. LED
The component of the applicator that indicates if the
Applicator is coupled or uncoupled on the treatment area.
2
Coupling is not available on the 1 cm
applicator.
LASER APPLICATOR
DETAIL DEVICE DESCRIPTION
3
2
Aperture
Laser Head
Pause/Resume Button
Laser Diode
LED’s/SLD’s
Laser Head
LED Indicator (Output Power)
LED Indicator (Output Power) This orange light illuminates
when Laser energy is being distributed by the applicator.
19
Page 22
DETAIL DEVICE DESCRIPTION
PATIENT REMOTE/LASER INTERRUPT SWITCH
The Vectra® Neo Patient Remote/Laser Interrupt Switch
buttons are described in the table below. By default, the
remote is not assigned to any treatment. When assigned,
the buttons function as follows:
Only) - Increases the intensity of the assigned stim
treatment; button is not active if stim treatment is
unassigned. Button is not active and has no function for
Ultrasound or Laser treatments.
Vectra® Neo Clinical Therapy System
Decrease Intensity (2) (Electrical Stimulation Treatments
Only) - Decreases the intensity of the assigned stim
treatment; button is not active if stim treatment is
unassigned. Button is not active and has no function for
Ultrasound or Laser treatments.
M (Manual Stimulation) (4) (Electrical Stimulation
Treatments Only) - Provides one cycle of stimulation.
Can only be operated when the clinician enables
manual mode on the base unit (head). This mode is
clinician monitored and is not for use when the patient is
unattended. Button is not active and has no function for
Ultrasound or Laser treatments.
Stop (3)
Increase (1)
Decrease (2)
Manual (4)
20
Page 23
Vectra® Neo Clinical Therapy System
HEAD TO CART ASSEMBLY
SETUP INSTRUCTIONS
The optional Therapy System Cart, PN 70001, is designed
for use with the Vectra® Neo Clinical Therapy System only
and allows the user to easily transport the System from
patient to patient within the clinic as well as store all
necessary accessories, supplies, and applicators used for
the various modalities of the System.
Tools required (not included): #2 Phillips screwdriver and
standard slotted screwdriver.
Remove the Vectra® Neo Clinical Therapy System from the
shipping carton. Visually inspect for damage. Report any
damage to the carrier immediately.
To assemble the Neo Head to the Cart, follow these steps:
1. Remove the top drawer from the Cart. Pull the drawer
open. Press the plastic tabs on both drawer slides
simultaneously in opposite directions, as shown. Pull
the drawer completely out.
3. Place the Neo Head on the cart facing toward the
drawers.
4. Fasten the Neo Head to the cart using four screws to
connect the base to the Neo Head.
2. Remove the base from the head unit prior to placing
it on the cart. Do this by removing the four screws
from the underside of the base where they secure to
the Neo Head. Retain for use when attaching the Neo
Head to the Cart.
5. If desired, replace the closed Handles with the open
Handles. Each Handle is attached with four screws as
shown.
6. Reinstall the drawer.
21
Page 24
SETUP INSTRUCTIONS
NEO LEG TO CART ASSEMBLY/ADJUSTMENT
Vectra® Neo Clinical Therapy System
The Neo Cart is shipped without the legs attached.
• Socket Head Cap Screw ¼-20 x 1-1/4”, quantity 6
(provided)
To install or adjust the leg assemblies onto the Neo Cart,
follow these steps:
1. Remove the bottom drawer from the Cart. Pull the
drawer open. Press the plastic tabs on both drawer
slides simultaneously in opposite directions, as shown.
Completely pull the drawer out.
2. There are two Cart height adjustments. Standard
shown on the left and lowered, shown on the right.
For initial installation, determine the desired height.
Locate three Allen-style bolts for each leg, left and
right and insert, by hand, in their respective slots. Use
the Allen wrench to secure the legs.
NOTE: To adjust height at a later time, simply remove
the Allen-style bolts, re-position the legs, and re-insert
the bolts.
3. Reinstall the Bottom Drawer.
22
Page 25
Vectra® Neo Clinical Therapy System
MODULE INSTALLATION
SETUP INSTRUCTIONS
All modules are installed from the left side (when facing
the screen) of the Neo Head unit and are each installed
in the same manner. Each has color-coded lead wires
that correspond to the appropriate colored labeling on
the modules. Module-speci c installation instructions
are shown after the generic instructions. To install the
modules in the Vectra® Neo Clinical Therapy System,
follow the steps shown.
Tools required (not included): #2 Phillips screwdriver and
standard slotted screwdriver.
The System is programmed to automatically recognize
the new Module(s), therefore, no software installation is
required.
UltrasoundLaser
3. Insert a standard slotted screwdriver (not provided)
into the top slot, pressing down with slight pressure.
Pull the faceplate off (in this example showing
Electrotherapy Channel 3/4).
4. The module is inserted on the left side of the Neo
Head in the slot as shown in this example (with the
Ultrasound module).
Stimulation3/4Stimulation 1/2
1. Ensure that the power cord is removed from the
device.
2. Remove the blank faceplate over the slot from the
left and right sides of the Neo head. (The example
displays the Ultrasound module.)
5. Carefully insert the module into the slot, with 32 pins
(2x16) in first. Secure the module in place with gentle
pressure until you feel the module is seated.
6. Secure the module with a screw provided at the
bottom as shown (using channel 3/4 as example).
7. In this example showing the Laser faceplate, insert
the faceplate at the bottom and snap into place at the
top, as shown on the left and right sides (Vectra® Neo
Clinical Therapy System allows Laser access on the left
and right sides.)
Ultrasound Cable Insertion
Shown below is the Ultrasound Cable Insertion location.
INSERTING PLUGS
When inserting the plugs for the Ultrasound and Laser
modules, be sure to align the at side of the plug with
the at side of the slot and push in gently. This is to avoid
bending the pins in the plug
8. Plug in the unit and press the power button. Allow
the unit to initialize, and then verify that the newly
installed module is shown as available on the Home
screen.
To operate the Patient Remote/Laser Interrupt Switch,
plug the remote into the device on the Rear Access Panel
receptacle, as shown below:
CAUTION
SETUP INSTRUCTIONS
2. Press the Remote ON/OFF toggle icon to assign or
unassign the remote to the selected treatment. The
remote can be assigned to only one treatment at
a time, however the remote can be reassigned as
needed.
When not in use, the Patient Remote/Laser Interrupt
Switch can be stored by hooking it onto the lead wire
holder clips in the same manner as lead wires and
cables (shown below).
• Patient Remote/Laser Interrupt Switch is to be used under
supervision of a physician or licensed practitioner only.
Complete the following steps to assign the remote to a
treatment:
1. When the remote is plugged into the unit, a Remote
ON/OFF toggle icon is displayed on the Treatment
review screen in the upper right corner (shown
below).
25
Page 28
SETUP INSTRUCTIONS
INSTALLING THE LASER INTERLOCK
DOOR INTERRUPT SWITCH
The Laser Interlock is an optional safety device designed
to interrupt Laser therapy if the door to the therapy room
is opened. The Laser Interlock Kit consists of a switch
resistor and a jack. Customers must supply the necessary
cable that complies with local and international codes.
Use only quali ed electricians to install the Laser Interlock
Kit.
The diagrams to the right provide installation guidelines
for therapy room with single and multiple doors.
Operation of the Laser Interlock
Laser Interlock works as an interrupt switch once it is
installed and connected to the Vectra® Neo Clinical
Therapy System with Laser module.
Laser Interlock monitors the state of the door(s) of the
therapy room and only allows start of Laser treatment if all
of the doors are closed.
Vectra® Neo Clinical Therapy System
Diagram for Therapy Room with One Door
Diagram for Therapy Room with Multiple Doors
If any door is open, it will not allow user to start the
Laser treatment; and if treatment is already started and
someone opens the door, it interrupts the system to stop
the Laser treatment.
WARNING
• Disconnect the system from the power source before attempting
any maintenance, installation, removal, or replacement
procedures to prevent electrical shock and possible damage to
system.
• The Laser Interlock must be installed by a professional or qualified
electrician. Serious eye injury can result if the device is not
properly installed. Also, when installing the device for multiple
doors, the resistance total may not exceed 4800 ohm.
26
Page 29
Vectra® Neo Clinical Therapy System
THERAPY SYSTEM STARTUP
SETUP INSTRUCTIONS
Complete the following steps for initial setup of the
Vectra® Neo Clinical Therapy System:
1. Plug the Power cord into the back of device. Plug the
other end of the cord into an electrical outlet.
NOTE: The Power Cord may be unplugged from the
back of the cart in an emergency situation.
2. Press the Power button located on the top left portion
of the LCD casing (shown below):
3. Select desired function on the Home screen (shown
below).
27
Page 30
SYSTEM SPECIFICATIONS
Vectra® Neo Clinical Therapy System
SYSTEM SPECIFICATIONS AND DIMENSIONS
WidthDepthHeightWeight
Module 11.12” (28.2448 cm)6.34” (16.1036 cm)1.43” (3.6322 cm)1lb (0.453592 kg)
Head @ 45 degree with Base (Tabletop)15.89” (40.3606 cm)15.89” (40.3606 cm)22.05” (56.007 cm)20.7lb (9.389362 kg)
NOTE: All waveforms except High Voltage Pulsed Current (HVPC) have been
designed with a 200 mA current limit. VMS™, VMS™ Burst ,and all TENS
waveform output intensities are measured, speci ed, and listed to peak, not
peak to peak.
GENERAL SYSTEM OPERATING AND STORAGE
TEMPERATURE
Operating Conditions
The device will meet its requirement under the following conditions:
The following charts represent the distribution of the
ultrasonic radiation eld and the orientation of the eld
with respect to each applicator (Y-plane represents
voltage in Vrms and X-plane represents applicator head
surface in 1mm resolution).
2
Crystal (model 27333)
1 cm
2 cm2 Crystal (model 27334)
E ective Radiating Areas (ERA)
ERA HighERA Low
2
2
DescriptionERA (cm
2
Crystal
10 cm
2
Crystal 45+25%3-25%
5 cm
2
Crystal1.82+11%1.4-22%
2 cm
2
Crystal0.91+11%0.4-55%
1 cm
)cm
8.510+18%7-18%
%cm2%
Head Warming Feature
The Head Warming feature of an Vectra® Neo Clinical Therapy System utilizes
Ultrasound output, resulting in warming of the Applicator to increase patient
comfort.
With Head Warming enabled, ultrasound is emitted without pressing the Start
button while an ultrasound treatment is being set up. The Applicator LED will
not illuminate during the Head Warming period. US Channel will indicate
“Head Warming”.
For all single diode and cluster laser and LED applicators, the expected increase in the measured quantities after
manufacture added to the values measured at the time of manufacture is
The software incorporates a cooling function that forces the user to cool the laser cluster prior to the next treatment.
The software will calculate the cooling time needed when treatment times exceed 3 minutes per application. For a
3 minute treatment, it will force a 15 second cool down period before the next treatment can begin. For a 4 minute
treatment, it will force a 2 minute cool down period before the next treatment can begin. The software extrapolates for
times between 3 and 4 minutes.
A message will display on the screen informing the user that the probe is cooling down and the time period required.
After 5 seconds, this message will disappear. If the user attempts to use the probe before the cool down period is
completed, the message will re-display to signify that the applicator is still in cool down mode. After the cool down period
is complete, a message displays that informs the user that the unit is ready for use.
670 nm (10mW) LED x6
880 nm (25mW) LED x7
950 nm (15mW) LED x6
670 nm (10mW) LED x7
950 nm (15mW) LED x3
850 nm (200mW) Laser x3
3250.0437.550.3860.698N/A
7150.0947.5512.4430.0970.543
32
Page 35
Vectra® Neo Clinical Therapy System
SYSTEM SPECIFICATIONS
LASER PROTECTIVE EYEWEAR SPECIFICATIONS
The graph below illustrates optical density in relation to wavelength. Each unit is shipped with laser protective eyewear
that is L3 rated and approved, as well as EN207 compliant.
The device is labeled with the date of
manufacture.
The Laser Applicator handle label
includes important safety information.
The LED Applicator handle label includes
important safety information.
DJO, LLC
1430 Decision Street
Vista, CA 92081
USA
Rx Only
13-7687-5_C
34
Page 37
Vectra® Neo Clinical Therapy System
WAVEFORMS
For additional information on Waveforms and output
energy, refer to APPENDIX 3 on page 78.
CC: Constant Current
CV: Constant Voltage
IFC (Interferential) Traditional (4 Pole)
Interferential Current is a medium frequency waveform. Current is distributed
through two channels (four electrodes). The currents cross each other in the
body at the area requiring treatment. The two currents interfere with each
other at this crossing point, resulting in a modulation of the intensity (the
current intensity increases and decreases at a regular frequency).
The Symmetrical Biphasic waveform has a short pulse duration and is capable
of strong stimulation of nerve bers in the skin and in muscle. This waveform is
often used in portable muscle stimulation units, and some TENS devices.
SYSTEM SPECIFICATIONS
TENS- Asymmetrical Biphasic
The Asymmetrical Biphasic waveform has a short pulse duration. It is capable
of strong stimulation of the nerve bers in the skin as well as of muscle tissue.
This waveform is often used in TENS devices.Because of its short pulse, the
patient typically tolerates the current well, even at relatively high intensities.
The HAN Waveform provides optimal parameters with a precisely controlled
sequence of Dense-and-Disperse (DD) modes of stimulation where 2 Hz is
alternating with 15 or 70 Hz, each lasting for 3 seconds..
• Stimulus delivered by the TENS waveforms of this device, in
certain configurations, will deliver a charge of
25 microcoulombs (μC) or greater per pulse and may be
sufficient to cause electrocution. Electrical current of this
magnitude must not flow through the thorax because it
may cause a cardiac arrhythmia.
35
Page 38
SYSTEM SPECIFICATIONS
WAVEFORMS CONTINUED
Vectra® Neo Clinical Therapy System
VMS™
VMS™ is a symmetrical biphasic waveform with a 100 µsec interphase interval.
Because the pulse is relatively short, the waveform has a low skin load,
making it suitable for applications requiring high intensities, such as in muscle
strengthening protocols.
The VMS™ FR version of the VMS ™ waveform is a physiologically based channel
interaction in which one channel stimulates the agonist and the other the
antagonist of the muscle group that is being exercised. VMS is a symmetrical
biphasic waveform with a 100 sec interphase interval. Because the pulse
is relatively short, the waveform has a low skin load, making it suitable for
applications requiring high intensities, such as in muscle strengthening
protocols.
VMS ™ Burst is a symmetrical biphasic waveform delivered in a burst format.
Because the pulse is relatively short, the waveform has a low skin load,
making it suitable for applications requiring high intensities, such as muscle
strengthening protocols.
Premodulated Current is a medium frequency waveform. Current comes out of
one channel (two electrodes). The current intensity is modulated: it increases
and decreases at a regular frequency (the Amplitude Modulation Frequency).
Russian Current is a sinusoidal waveform, delivered in bursts or series of pulses.
This method was claimed by its author (Kots) to produce maximal muscle
strengthening e ects without signi cant discomfort to the patient.
The High Voltage Pulsed Current (HVPC) has a very brief pulse duration
characterized by two distinct peaks delivered at high voltage. The waveform is
monophasic (current ows in one direction only).
Microcurrent is a monophasic waveform of very low intensity. The physiological
working mechanism of this e ect is as yet not clearly understood. It is thought
to stimulate tissue healing by stimulating the ‘current of injury’, a current which
naturally occurs in healing tissue.
• Examine the skin for any wounds and clean the skin.
• Apply the electrodes to the treatment area.
• Ensure the electrodes are applied securely to the skin.
• Ensure good contact between each electrode and the
skin.
• Check the electrode contact regularly during the
treatment.
• Examine the skin again after the treatment.
• Choose electrodes that fit the anatomy.
• View the Electrode Placement recommendations
in the Treatment Review screen for the particular
modality being used for treatment as a reference point
only prior to administering treatment.
• Follow electrode manufacturer instructions.
• DO NOT use electrodes with an active area less than
7.92 cm
injury. Always exercise caution with current densities
more than 2mA/cm2. Refer to CURRENT DENSITY
CALCULATIONS FOR DJO ELECTRODES table in
Appendix 3.
2
, as there will be a risk of suffering a burn
ELECTROTHERAPY PATIENT PREPARATION
DURA-STICK® Electrode Instructions
Connecting Lead Wires
1. Insert the lead with the Red (+) electrode connector
into one DURA-STICK® Electrode.
2. Insert the lead with the Black (-) electrode connector
into the other electrode.
3. Make certain the lead wires are seated completely into
the electrodes.
NOTE: Use of conductive medium or sponges
is not required or recommended. DURA-STICK®
Electrodes are manufactured to ensure the optimum
conductivity during therapy when properly applied.
DURASTICK® ELECTRODES
DURA-STICK® Electrodes are a self adhesive, disposable
product designed speci cally for use with Vectra® Neo
Clinical Therapy System.
It is recommended that DURA-STICK® Electrodes be used
whenever possible to ensure the highest level of contact
with the treatment area and most uniform delivery of the
prescribed electrotherapy treatment.
Securing Electrodes
1. Remove the DURA-STICK® Electrodes from the
protective backing.
2. Apply to the treatment area as prescribed.
3. Ensure the entire electrode surface is in contact with
patient skin by pressing into place.
For ELECTROTHERAPY OPERATION, refer to page 48.
38
Page 41
Vectra® Neo Clinical Therapy System
SEMG & STIM PATIENT PREPARATION
PATIENT PREPARATION
Install DURA-STICK® Electrodes
1. Connect a DURA-STICK® 2 in (5 cm) disposable
electrode to each lead. These electrodes are designed
for use with DJO equipment and will provide an
accurate reading of sEMG activity. Active (Red) Lead,
Active (Black) Lead, and Reference (Green) Lead are
shown below.
2. Leave protective backing on electrodes until treatment
area has been prepared.
2. Select the Customize icon.
3. Select the Electrode Placement icon to view electrode
placement according to body area.
4. Touch the body part you wish to treat.
5. View the electrode placement(s) for desired body area
and press the Up and Down icons to scroll through
text relating to specific electrode placement and
typical conditions of the area.
Electrode Placement by Body Area
1. From the Home screen, select sEMG.
2. Choose one of the following modalities:
• sEMG (Ch 1)
• sEMG (Ch 2)
• sEMG (Ch 1 + 2)
• sEMG + Stim VMS
• sEMG + Stim Sym Biph
Connect the sEMG Lead Wires
1. Connect the sEMG Lead wires to the channel(s) desired
for use with the selected modality.
connect
SEMG lead
wires
6. Press the Back Arrow icon to return to desired screen.
7. Examine the skin for any wounds.
39
Page 42
PATIENT PREPARATION
sEMG & STIM PATIENT PREPARATION CONTINUED
Vectra® Neo Clinical Therapy System
8. Thoroughly cleanse the skin treating area.
NOTE: Thorough and proper cleaning of the
treatment area to remove any topical medication
and cream film as well as loose skin particles from
the treatment area is critical to the skin contact and
reception of the Electrodes during sEMG and sEMG +
Stim therapy.
Electrode Placement
3. Review the specific Electrode Placement graphic for
positioning of the Reference (green lead) electrode.
NOTE: The electrodes may be placed for specific,
general, and quasi-specific biofeedback muscle or
muscle group activity.
For sEMG and sEMG + Electrical Stimulation operation,
refer to page 55.
1. Using DURA-STICK® 2 in (5 cm) electrodes, place the
Active (red and black lead) electrodes in the center of
the muscle belly and parallel with the muscle fibers.
2. Position the Reference electrode (green lead) in close
proximity to the treatment area.
• Using small electrodes and placing them closer
together will render a more specific reading of
muscle activity during sEMG and sEMG + Stim
therapy.
• The Active electrodes may be placed farther apart
to obtain a general reading of a muscle or muscle
group activity during the session.
• DJO recommends using only DURA-STICK®
electrodes to obtain the most accurate sEMG
feedback.
• Follow the electrode manufacturer instructions.
• Trimming or cutting electrodes may interfere with
the reception of sEMG data and may affect the
delivery of electrical stimulation in the sEMG + Stim
Modality.
40
Page 43
Vectra® Neo Clinical Therapy System
PATIENT PREPARATION
LASER PATIENT PREPARATION
Preparing the Patient’s Skin for Laser Therapy
Before applying Laser therapy to the patient, you must
rst prepare the patient’s skin. By properly preparing the
patient’s skin for Laser therapy, you will allow more Laser
energy to reach the targeted areas and reduce the risk of
skin irritation.
To prepare the patient’s skin for Laser therapy, do the
following:
1. Thoroughly wash the skin on which you intend to
place the laser head with mild soap and water or
alcohol wipe.
2. Dry the skin thoroughly.
For LASER OPERATION, refer to page 56.
ULTRASOUND PATIENT PREPARATION
1. Examine the skin for any wounds and clean the skin.
2. View the Applicator recommendation in the
treatment.
3. Review screen for Ultrasound (as a reference point
only) prior to administering treatment.
NOTE: Applicators are available in the sizes shown
below:
1 cm
2
2 cm25 cm
2
Applicator Preparation and Use
1. Clean applicator before each therapy session with
warm soapy water.
10 cm
2
2. Liberally apply Conductor™ Transmission Gel to the
treatment area on the patient.
• Move the Applicator during therapy session in a
circular motion. The area treated should be twice
the diameter of the Applicator.
3. If US Coupling is On, the Applicator is properly
coupled to the patient and administering ultrasound
when the LED is constantly illuminated. Coupling is
2
not available on the 1 cm
applicator.
For ULTRASOUND OPERATION, refer to page 51.
41
Page 44
DEVICE USER INTERFACE
SCREEN DESCRIPTION
Vectra® Neo Clinical Therapy System
Each screen contains the following areas:
Title Bar
Located at the top of each screen and lists the current
screen and previous screens back to the Home screen. It
also contains a Print Screen icon at the top right, Patient
Remote/Laser Interrupt Switch, when installed, and a USB
connectivity icon when a USB ash drive is inserted.
Main Area
• Located under the Title bar, this area displays icons
unique to the current screen.
• All screens (except the Home screen) will contain the
Back Arrow icon to scroll to the previous screen and
the Home icon to return to the Home screen.
Channel Area
Located at the bottom of each screen, this screen displays
the following status information about each channel:
• Not Installed: Indicates the associated module is not
installed in the unit
• Available: Indicates the channel is available for use
• Setup: Indicates a treatment for the channel is
currently being set up but treatment has not yet
begun
• Running: Indicates a treatment for the channel is
currently running
• Paused: Indicates a treatment for the channel is
currently paused
• Completed: Indicates a treatment for the channel has
completed
• No applicator: Indicates there is not a valid applicator
plugged into the channel’s module (only valid for
Ultrasound and Laser)
Electrode Contact Quality Indicator
(Electrotherapy Channels Only)
Active Channels Indicator
The image below shows the Home screen with modality
and resource icons.
Title Bar
Main Area
Channels
Electrode
Contact
Quality
Indicator
Active Channels Indicator
42
Page 45
Vectra® Neo Clinical Therapy System
HOME SCREEN
The Vectra® Neo Clinical Therapy System Home
screen provides access to all of the system modalities
and functions. The Home screen has the following
information:
1. Utilities
Modality Icons:
2. Electrotherapy
3. Ultrasound
4. Combo
5. sEMG
6. Laser
7. Clinical Protocol Set-up (CPS)
8. Custom Protocols
9. Assign Shortcut Options
10. Patient Data Storage
11. Anatomical Library
DEVICE USER INTERFACE
1
23456
7
8
1110
9
43
Page 46
DEVICE USER INTERFACE
Vectra® Neo Clinical Therapy System
UTILITIES AND OPTIONS
The Utilities icon on the Home screen o ers users the opportunity to set the following preferences::
1
2
3
4
1. Print Screen
Select the Printer icon in the top right corner to
capture screen shots of current screen. Screen shots
are saved to the USB flash drive plugged into the unit
and are saved in Windows bitmap (.bmp) format. Note
that a screen cannot be printed while a treatment is in
process.
2. Volume
Select the Volume icon to set desired audio volume.
The volume range is 0% (off) to 100% (loudest) and is
measured in 20% increments. The default setting is
60%.
3. LCD Brightness
Select the Brightness icon to set the brightness of
the LCD screen. The brightness ranges from 50%
(dimmest) to 100% (brightest) in 10% increments. The
default setting is 100%.
5
6
7
8
12
10
11
13
9
4. US Coupling
Select the US Coupling icon to set the functionality of
an ultrasound treatment when unit detects that the
ultrasound applicator is uncoupled. Coupling is not
available on the 1cm2 applicator. The default is set to
Off. The following preferences are available:
• Off: Nothing happens when applicator becomes
uncoupled.
• Pause and Beep: When applicator becomes
uncoupled, treatment timer will pause, unit will
beep, and channel area indicates “uncoupled” state.
• Pause and No Beep: When applicator becomes
uncoupled, treatment timer will pause, unit will not
beep, and channel area indicates “uncoupled” state.
• Beep: When applicator becomes uncoupled,
treatment timer will continue to run, unit will beep,
and channel area indicates “uncoupled” state.
• No Beep: When applicator becomes uncoupled,
treatment timer will continue to run, unit will not
beep, and channel area indicates “uncoupled” state.
NOTE: Ultrasound output will continue to be emitted
in all US Coupling modes even if the applicator is
uncoupled.
44
Page 47
Vectra® Neo Clinical Therapy System
UTILITIES AND OPTIONS CONTINUED
DEVICE USER INTERFACE
5. US Head Warming
Select the US Head Warming icon to turn on or off
the ultrasound warming feature during ultrasound
treatment setup. The available choices are either On
or Off. The default setting is Off.
6. E-Stim Contact Quality
Select the E-Stim Contact Quality icon to turn on or
off the electrotherapy contact quality monitoring
function. The available choices are either On or Off.
The default setting is On, and in the On position, the
unit will monitor the contact quality of the electrodes
on all waveforms, excluding High Voltage Pulsed
Current (HVPC) and Microcurrent.
NOTE: If the impedance feedback exceeds 5000 ohms
(tolerance ±20%) and the set intensity is ≥ 10 mA CC
or 10 V CV, then the unit will display an information
message and set the intensity to zero for all channels
associated with this treatment.
7. Restore Default Unit Settings
Select the Restore Default Unit Settings icon to reset
all of the following settings back to their factory
defaults:
• Volume
• LCD Brightness
• US Coupling
• US Head Warming
8. Restore Default Protocols
Select the Restore Default Protocols icon to reset all
protocols (factory, custom, and the 5 custom protocol
shortcuts) to their factory defaults.
9. Display Unit Version Information
Select the Display Unit Version Information icon to
view the install status of each module. For all installed
modules, the PCB information and the Software
version for that module are displayed.
11. Import Protocols from USB Flash Drive
Select the Import Protocols from USB Flash Drive icon
to import all protocols from a valid USB flash drive.
Any protocols with the same name as an existing unit
protocol will be overwritten by the imported protocol,
and any protocols of a different name on the unit will
remain intact.
12. Date and Time
Select the Date and Time icon to set the date and time
on the unit.
13. Clinic Name
Select the Clinic Name icon to enter the name of your
clinic. The clinic displays on the Home screen and on
the patient treatment reports saved to the USB flash
drive. The following icons will perform the functions
listed:
• Left: Moves the cursor to the left one space per
instance pressed.
• Right: Moves the cursor to the right one space per
instance pressed.
• Space: Inserts one space per instance pressed.
• Clear: Clears the entire entry.
• Delete: Once the cursor has been moved using the
Left or Right icons, press the Delete icon to delete
the character.
• Caps: Press the Caps icon to capitalize the letters or
to make letters lowercase.
• Cancel: Press the Cancel icon to cancel the clinic
name entry function and return to the Utilities
screen.
• Accept: Press the Accept icon once entry of the
clinic name is complete. The clinic name should
now appear in the title bar area of the screen.
If a valid software upgrade USB flash drive is inserted
into the unit, the version of software for each module
that is on the USB flash drive is displayed.
NOTE: USB flash drive version is displayed in yellow
if it is newer than the current software version for the
applicable module.
10. Export Protocols to USB flash drive
Select the Export Protocols to USB flash drive icon to
export all protocols (both default and custom) to a
valid USB flash drive.
45
Page 48
DEVICE USER INTERFACE
Vectra® Neo Clinical Therapy System
TREATMENT SCREENS
The Vectra® Neo Clinical Therapy System Treatment screens for Electrotherapy and Ultrasound include the following
information:
43
1
2
8
5
6
101011121010
1. Modality Description Icon
Press the Modality Description Icon to view the text
explaining the rationale for the modality associated
with the specific Clinical Protocol selected.
2. Electrode Placement Icon
Press the Electrode Placement Icon to view the specific
electrode placement for the Clinical Protocol selected.
3. Home Icon
Press the Home Icon to return to the Home screen at
any time.
4. Back Arrow Icon
Press the Back Arrow Icon to return to the previously
viewed screen.
5. Remote Icon
Press either Off/On to deactivate or activate the
Remote Control feature.
6. Time Icon
Press the Time icon to adjust therapy time/duration.
7. Intensity Icon
Set intensity by rotating the Intensity Control Knob to
the prescribed level:
• Clockwise – Increases Intensity
• Counterclockwise – Decreases Intensity
13
9
7
8. Therapy Information Window
View selected Therapy information such as Waveform,
Electrode Size, CC/CV, Carrier Frequency, Cycle Time,
Frequency, and Vector Scan on the Therapy Information
Window.
9. Customize Icon
Press the Customize icon to edit the Therapy
information.
10. 4 Channel Icons
This icon shows the modalities in use. Channels are
automatically assigned to the next available channel.
Manual selection is done by touching the desired
channel.
11. Ultrasound Icon
This icon shows the Ultrasound information when in
use.
12. Laser Icon
This icon shows the Laser module information when in
use.
13. Remote Icon
This is how the Remote Icon is viewed from the main
screen.
46
Page 49
Vectra® Neo Clinical Therapy System
CPS CLINICAL PROTOCOL SETUP
DEVICE USER INTERFACE
The Vectra® Neo Clinical Therapy System has a Clinical
Protocol Setup (CPS) icon that is a series of protocol
presets where the Body Area, Clinical Indication,
Pathological Condition, and Pathological Severity are
selected by the user to assist in selecting the parameter
setting. These Clinical Protocols are to be used only as
guidelines. Each patient should be individually assessed to
determine the appropriateness of the protocol parameters
prior to use.
The following information gives general instructions to
access, select, and set up Clinical Protocols. Each Clinical
Protocol is set up and edited in the same basic manner.
1. Select CPS from the Home screen.
2. Touch the body part you wish to treat.
9. Overwrite a previously saved protocol by using the Up
and Down arrows or Previous Page/Next Page icons to
select the protocol to overwrite. Press the Save icon to
save the overwritten icon.
10. Return to Home screen by pressing the Home icon.
Press the Back arrow to scroll back one screen.
3. Using the Up and Down Arrows, scroll to the
appropriate indication and select by pressing the “√”
icon.
4. Using the Up and Down Arrows, scroll to the
appropriate condition you wish to treat and select by
pressing the “√” icon.
5. Select the modality desired.
6. Begin therapy by pressing the Start button or make
setting changes pressing the Customize icon.
7. To save the customized protocol, press the Save icon.
8. Save a new protocol by pressing the Save as a New
Name icon and typing in a name for the protocol.
Press the Save icon when finished.
47
Page 50
DEVICE USER INTERFACE
ELECTROTHERAPY OPERATION
Vectra® Neo Clinical Therapy System
All waveforms in the Vectra® Neo Clinical Therapy System
are set up and edited in the same basic fashion. The
Vectra® Neo Clinical Therapy System has an Electrotherapy
Icon with the following waveforms: IFC (Interferential)
Traditional (4 Pole), TENS- Asymmetrical Biphasic,
TENS- Symmetrical Biphasic, TENS HAN, High Voltage
Pulsed Current (HVPC), VMS™, VMS™Burst, VMS™ FR,
IFC Premodulated (Traditional 2 Pole), Russian, and
Microcurrent.
Complete the following steps to begin Electrotherapy
treatment:
1. Prepare patient and therapy system for Electrotherapy.
Refer to the ELECTROTHERAPY PATIENT PREPARATION
section on page 38 for electrode selection,
preparing the patient, and securing electrodes.
2. Touch the screen to activate the system. Read and
carefully follow the instructions on the screen.
3. From the Home screen, press the Electrotherapy icon.
4. Select desired waveform from the listing on the
screen by pressing the appropriate icon. Refer to the
SYSTEM SPECIFICATIONS section on page 28 for all
waveform specifications for the Vectra® Neo Clinical
Therapy System. The screen shown below will appear.
• To view information explaining the waveform, press
the Modality Description icon. Press the Up and
Down icons to view additional text. Press the Back
Arrow icon to return to the previous screen or the
Home icon to return to the Home screen.
• To view the most commonly used electrode
placement for the selected waveform, press the
Electrode Placement icon. A picture will appear
prompting you to select the body part you wish to
treat. Press the Down icon to scroll through text.
Press the Back icon to return to the previous screen
or the Home icon to return to the Home screen.
• To customize waveform settings shown in the
list box, press the Customize icon located in the
list box, and the screen shown below will appear.
Make the desired changes and press the Back icon
to return to the previous screen, the Home icon
to return to the home screen, or the Save icon to
save the customized settings. Refer to the CUSTOM
PROTOCOLS section on page 59 for detailed
instructions for saving customized settings.
48
• To set pain information for the treatment, press the
Pain Information icon. At this screen, you can do the
following:
- Map electrode placement by dragging the
electrodes to their appropriate locations on the
body and selecting either front, back, left, or
right side.
- Map pain information by touching the body area
in the list box.
- Edit pain scale information by pressing the Edit
Pain Scale icon. Press the Up and Down arrows
until pain is properly gauged. Press the Back
Arrow icon to return to the previous screen or the
Home icon to return to the Home screen.
Page 51
Vectra® Neo Clinical Therapy System
ELECTROTHERAPY OPERATION CONTINUED
5. If desired, connect optional Remote Control to device.
6. Use the Intensity dial to set therapy intensity:
• Clockwise - Increases Intensity
• Counterclockwise - Decreases Intensity
7. Press the Start button to begin therapy, the Pause
button to pause treatment, or the Stop button to
terminate the treatment.
NOTE: To make setting adjustments during therapy,
press the Customize icon or adjust intensity using the
intensity knob.
8. When treatment has completed, the Treatment
Summary screen will appear with the following
options:
• Save therapy information to a USB flash drive by
inserting a USB flash drive into the device and
pressing the Save to USB Flash Drive icon.
• Repeat the treatment by pressing the Run this
Treatment icon.
• Save the treatment protocol by pressing the Save
Protocol icon. Refer to the CUSTOM PROTOCOLS
section on page 59 for detailed instructions on
customized settings.
DEVICE USER INTERFACE
49
Page 52
DEVICE USER INTERFACE
SEQUENCING OPERATION
Vectra® Neo Clinical Therapy System
The Vectra® Neo Clinical Therapy System o ers sequencing
for special electrotherapy waveform treatment purposes
and stores these protocols in the system memory for
recall and use. Sequencing allows you to link up to 3
electrotherapy protocols together to encompass one
treatment session. Each sequence can have up to 3
protocols. The Vectra® Neo Clinical Therapy System memory
will accommodate up to 200 user-de ned protocols,
including all user protocols, sequences, and system default
protocols. The Vectra® Neo Clinical Therapy System denotes
default protocols with an asterisk (*) before the named
sequence.
Complete the following steps to create and access
customized sequences:
1. Press the Electrotherapy icon from the Home screen.
2. Press the Sequencing icon.
3. Select a sequence from the list box by pressing the Up
and Down arrows to find the desired sequence and
press the “√” icon. The screen shown below will appear.
4. Make any desired changes to the existing sequences
by using the Up and Down arrows to find the desired
sequence and press the Edit icon. This screen also
allows you to:
• Delete a sequence from the protocol by pressing
the Delete icon.
• Add a new sequence to the protocol by pressing
the New icon and using the Up and Down arrows to
find the desired protocol in the list box. Press the “√”
icon to add the new sequence.
• Save the newly created sequence by pressing the
Save icon and following the saving a treatment
protocol instructions listed in the ELECTROTHERAPY
OPERATION section on page 48.
5. Set the intensity for each protocol by using the Up and
Down arrows for each sequence and press the Start
button to begin treatment.
6. When treatment is completed, the Treatment
Summary screen will appear with the following
options:
• Save therapy information to a USB flash drive by
inserting a USB flash drive into the device and
pressing the Save to USB Flash Drive icon.
• Repeat the treatment by pressing the Run this
Treatment icon.
• Save the treatment protocol by pressing the Save
Protocol icon. Refer to the CUSTOM PROTOCOLS
section on page 59 for detailed instructions for
saving customized settings.
50
Page 53
Vectra® Neo Clinical Therapy System
ULTRASOUND OPERATION
DEVICE USER INTERFACE
The Vectra® Neo Clinical Therapy System Ultrasound
modality allows the user to select speci c Applicator
recommendations and edit treatment parameters for
various syndromes requiring the use of ultrasound
therapy. The following information gives general
instructions for the setup of Ultrasound Therapy when
selecting Ultrasound from the Home screen.
WARNING
• The Vectra® Neo Clinical Therapy System may be susceptible to
Electrostatic Discharge (ESD) at greater than ±4 kV when first grasping
either the Ultrasound or Laser Applicator. In the event of such a
discharge, the Vectra® Neo Clinical Therapy System may experience
communication loss with the installed modules. The Vectra® Neo
Clinical Therapy System will terminate all active outputs (stim,
ultrasound, laser), automatically place the unit in a safe state, and
issue an error message 301 or 307.
- To recover from an error message 301 or 307, turn the unit off and
on using the ON/OFF switch located at the top of the display. Once
the system restarts, re-initiate all treatments that were
interrupted.
Complete the following steps to begin Ultrasound
treatment:
1. Touch the screen to activate the system. Read and
carefully follow the instructions on the screen.
• To prepare the patient’s skin for Ultrasound Therapy,
prepare patient as described in the ULTRASOUND
PATIENT PREPARATION section on page 41.
2. From the Home screen, select the Ultrasound icon. The
screen shown below will appear:
• To prevent Electrostatic Discharge (ESD) at greater than ±4 kV:
- Grasp and hold the Ultrasound or Laser Applicator prior to starting
treatment. If the applicator must be put down prior to completion
of treatment, stop the current treatment first and then place the
applicator in the holder.
- Maintain humidity in the use environment to at least 50% relative
humidity.
- Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, DJO recommends implementing
additional controls to maintain relative humidity to at least 50%.
- Communicate these ESD-precautionary procedures to healthcare
staff, contractors, visitors, and patients.
NOTE: Prior to starting treatment, replace the ultrasound
faceplate on the side of the module not being used with a
blank faceplate.
• To view information explaining the modality, select
the Modality Description icon. Press the Up and
Down icons to view additional text. Press the Back
Arrow icon to return to the previous screen or the
Home icon to return to the Home screen.
• To view applicator information, press the Applicator
Information icon:
- Touch the body part you wish to treat.
- Information about applicator sizes, frequency,
and treatment directions will appear as seen on
the screen above. Press the Down icon to scroll
through the information.
- Press the Back Arrow icon to return to the
previous screen or the Home icon to return to the
Home screen.
• Use the Intensity dial to set therapy intensity:
- Clockwise – Increases Intensity
- Counterclockwise – Decreases Intensity
51
Page 54
ULTRASOUND OPERATION CONTINUED
3. Press the Start button to begin therapy, the Pause
button to pause treatment, or the Stop button to
terminate the treatment.
4. When treatment has completed, the Treatment
Summary screen will appear with the following
options:
• Save therapy information to a USB flash drive by
inserting a USB flash drive into the device and
pressing the Save to USB Flash Drive icon.
• Repeat the treatment by pressing the Run this
Treatment icon.
• Save the treatment protocol by pressing the Save
Protocol icon. Refer to the CUSTOM PROTOCOLS
section on page 59 for detailed instructions for
saving customized settings.
Vectra® Neo Clinical Therapy SystemDEVICE USER INTERFACE
52
Page 55
Vectra® Neo Clinical Therapy System
COMBINATION OPERATION
The Vectra® Neo Clinical Therapy System Combination
modality allows the user to select and use Ultrasound
therapy in combination with electrical muscle stimulation.
Combination therapy utilizes the Ultrasound modality
in conjunction with High Voltage Pulsed Current (HVPC),
IFC (4 pole), IFC Premodulated (2 pole), Asymmetrical
Biphasic, Symmetrical Biphasic, or VMS™ to generate a
therapeutic e ect. In this mode of therapy, the Applicator
of the Ultrasound Applicator becomes one-half of the
electrical circuit. An electrode attached to the Black (-)
lead wire completes the circuit.
Complete the following steps to begin Combination
treatment:
1. Touch the screen to activate the system. Read and
carefully follow the instructions on the screen.
2. Prepare patient and therapy system. Refer to the
PATIENT PREPARATION section on page 38 for
electrode selection, preparing the patient, and
securing electrodes. For ULTRASOUND PATIENT
PREPARATION, refer to page 41.
3. Connect the Black (-) lead wire from Channel 2 to the
electrode. Make certain the lead wire is completely
seated in the electrode. The Red (+) lead wire is not
used. The Ultra sound Applicator completes the circuit
for Combination Therapy.
4. From the Home screen, select the Combo icon.
5. Select the ultrasound combination therapy desired,
and press the corresponding icon. The Treatment
Review screen shown below will appear:
• To view information explaining the waveform,
select the Modality Description icon. Press the Up
and Down icons to view additional text. Press the
Back Arrow icon to return to the previous screen or
the Home icon to return to the Home screen.
• To view the most commonly used electrode
placement for the selected waveform, select the
Electrode Placement icon. A picture will appear
prompting you to select the body part you wish to
treat. Press the Down icon to scroll through text.
Press the Back Arrow icon to return to the previous
screen or the Home icon to return to the Home
screen.
53
Page 56
DEVICE USER INTERFACE
COMBINATION OPERATION CONTINUED
Vectra® Neo Clinical Therapy System
• To customize settings shown in the list boxes, press
the Customize icon located in the list box, and the
screen shown below will appear.
Electrotherapy Customization Screen
Ultrasound Customization Screen
6. Press the Start button to begin therapy, the Pause
button to pause treatment, or the Stop button to
terminate the treatment.
7. When treatment has completed, the Treatment
Summary screen will appear with the following
options:
• Save therapy information to a USB flash drive by
inserting a USB flash drive into the device and
pressing the Save to USB Flash Drive icon.
• Repeat the treatment by pressing the Run This
Treatment icon.
• Save the treatment protocol by pressing the Save
Protocol icon. Refer to the CUSTOM PROTOCOLS
section on page 59 for detailed instructions for
saving customized setting.
• Make the desired changes and press the Back Arrow
icon to return to the previous screen, the Home icon
to return to the home screen, or the Save icon to
save the customized settings. Refer to the CUSTOM
PROTOCOLS section on page 59 for detailed
instructions for saving customized settings.
• To adjust treatment time, press the left or right
arrows to reach desired treatment time.
• Use the Intensity dial to set therapy intensity:
- Clockwise – Increases Intensity
- Counterclockwise – Decreases Intensity
54
Page 57
Vectra® Neo Clinical Therapy System
sEMG OPERATION
DEVICE USER INTERFACE
The Vectra® Neo Clinical Therapy System sEMG + Stim
modality utilizes sEMG biofeedback activity coupled with
triggered electrical muscle stimulation using selected
electrotherapy waveforms for the maximum bene t in
muscle retraining. The following options are available:
sEMG (Channel 1), sEMG (Channel 2),
sEMG (Channels 1 & 2), sEMG + Stim VMS, and sEMG +
Stim Sym Biphasic.
The Electrical Muscle Stimulation is triggered when the
muscle contraction (sEMG portion of the therapy) reaches
the target, sEMG stops, and the muscle is then electrically
stimulated for the pre-set period. After stimulation, the
patient is given a short rest period and then repeats the
muscle contraction, attempting to reach the target to
again trigger the electrical stimulation. This is repeated
throughout the therapy session.
Session parameters can be stored on a USB storage drive
and accessed on the Home screen under the Patient Data
icon. The sEMG portion of sEMG + Stim modality is used
to force the patient to contract the muscle to a prescribed
target. The data cannot be recorded or stored on the
Patient Data Card or the sEMG Data Card.
NOTE: This section applies when Stimulation Channels
and sEMG module is installed.
Complete the following steps to begin sEMG treatment:
1. Touch the screen to activate the system.
2. Prepare patient and therapy system. Refer to the
PATIENT PREPARATION section on page 38 for
electrode selection, preparing the patient, and
securing electrodes.
3. Select sEMG icon from the Home screen.
4. Select the prescribed channel icon (see previous
for the available choices). The treatment screen will
appear (the screen shown below illustrates the sEMG
Channel 1 selection).
• To set the Target Threshold either Automatically or
Manually, touch the CUSTOMIZE Icon.
• Use the Intensity dial to set therapy intensity:
- Clockwise – Increases Intensity
- Counterclockwise – Decreases Intensity
5. Press Start/Pause button.to begin therapy. Screen
prompts willl appear:
• Contract: Instructs the patient to attempt reaching
the Target Threshold. “Contract” and a green vertical
line appear on the screen, indicating the patient
should attempt to contract the selected muscle(s).
“Contract” remains on the screen until the patient’s
sEMG output reaches the Target Threshold, at which
time Electrical Stimulation is delivered.
• Hold: Instructs the patient to contract the selected
muscle(s). When the Target Threshold is reached,
the “Hold” prompt appears, instructing the patient
to continue to contract the selected muscle(s) until
the pre-set time for the Stimulation ends. A yellow
vertical line appears at the end of the Hold period.
• Relax: Instructs the patient to relax. “Relax” appears,
indicating the patient should relax, stopping the
contraction. Relax continues for the pre-set time.
The cycle repeats when “Contract” re-appears, again
indicating the patient should attempt to contract
the selected muscle(s).
55
Page 58
DEVICE USER INTERFACE
sEMG OPERATION CONTINUED LASER OPERATION
Vectra® Neo Clinical Therapy System
6. The following options are available under the
Customize Treatment screen and accessed by pressing
the appropriate icon:
• To view information explaining the waveform, press
the Modality Description icon. Press the Up and
Down icons to view additional text. Press the Back
Arrow icon to return to the previous screen or the
Home icon to return to the Home screen.
• To view the most commonly used electrode
placement for the selected waveform, press the
Electrode Placement icon. A picture will appear
prompting you to select the body part you wish to
treat. Press the Down icon to scroll through text.
Press the Back Arrow icon to return to the previous
screen or the Home icon to return to the Home
screen.
NOTE:When using sEMG + Stim, the Customization
screen adds two icons to edit Stimulation settings
and to set the stimulation intensity.
• Use the Intensity dial to set therapy intensity:
- Clockwise - Increases Intensity
- Counterclockwise - Decreases Intensity
7. Make the desired changes and press the Back Arrow
icon to return to the previous screen and begin
treatment, the Home icon to return to the home
screen, or the Save icon to save the customized
settings. Refer to the CUSTOM PROTOCOLS section
on page 59 for detailed instructions for saving
customized settings.
Before applying Laser therapy to the patient, you must
rst prepare the patient’s skin as described in the
PATIENT PREPARATION section on page 38. By properly
preparing the patient’s skin for Laser therapy, you will
allow more Laser energy to reach the targeted areas and
reduce the risk of skin irritation.
DANGER
• Laser protective eyewear should be worn during laser
treatment by the operator and patient to block infrared
light energy from the eyes during treatment.
WARNING
• The Vectra® Neo Clinical Therapy System may be susceptible to
Electrostatic Discharge (ESD) at greater than ±4 kV when first grasping
either the Ultrasound or Laser Applicator. In the event of such a
discharge, the Vectra® Neo Clinical Therapy System may experience
communication loss with the installed modules. The Vectra® Neo
Clinical Therapy System will terminate all active outputs (stim,
ultrasound, laser), automatically place the unit in a safe state, and
issue an error message 301 or 307.
• To recover from an error message 301 or 307, turn the unit off
and on using the ON/OFF switch located at the top of the display.
Once the system restarts, re-initiate all treatments that were
interrupted.
• To prevent Electrostatic Discharge (ESD) at greater than ±4 kV:
8. Press the Start button to begin therapy, the Pause
button to pause treatment, or the Stop button to
terminate the treatment.
9. When treatment has completed, the Treatment
Summary screen will appear with the following
options:
• Save therapy information to a USB flash drive by
inserting a USB flash drive into the device and
pressing the Save to USB Flash Drive icon.
• Repeat the treatment by pressing the Run This
Treatment icon.
• Save the treatment protocol by pressing the Save
Protocol icon. Refer to the CUSTOM PROTOCOLS
section on page 59 for detailed instructions for
saving customized settings.
• Grasp and hold the Ultrasound or Laser Applicator prior to
starting treatment. If the applicator must be put down prior to
completion of treatment, stop the current treatment first and
then place the applicator in the holder.
• Maintain humidity in the use environment to at least 50%
relative humidity.
• Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, DJO recommends implementing
additional controls to maintain relative humidity to at least
50%.
• Communicate these ESD-precautionary procedures to healthcare
staff, contractors, visitors, and patients.
**Refer to the TREATMENT TIPS section on page 71.
NOTE: Prior to starting treatment, replace the laser
faceplate on the side of the module not being used with a
blank faceplate.
56
Page 59
Vectra® Neo Clinical Therapy System
LASER OPERATION CONTINUED
DEVICE USER INTERFACE
Complete the following steps to begin Laser therapy
treatment:
1. Touch the screen to activate the system. Read and
carefully follow the instructions on the screen.
2. Select the Laser icon from the Home screen.
3. Type in a four-digit PIN and press the Accept icon to
accept the entry or choose from the following options:
• The Cancel icon to return to the Home screen
• The Clear icon to clear the numbers entered
• The Left and Right arrows to scroll through the
numbers entered
• The Delete icon to delete numbers entered one at a
time
4. After successfully entering your PIN, the screen shown
below will appear:
• To customize settings shown in the list boxes, press
the Customize icon located in the list box, and the
screen shown below will appear.
- Make the desired changes and press the Back
Arrow icon to return to the previous screen, the
Home icon to return to the Home screen, or
the Save icon to save the customized settings.
Refer to the CUSTOM PROTOCOLS section on
page 59 for detailed instructions for saving
customized settings.
6. Use the Intensity dial to set therapy duration:
• To view information explaining the modality, press
the Modality Description icon. Press the Up and
Down icons to view additional text. Press the Back
Arrow icon to return to the previous screen or the
Home icon to return to the Home screen.
5. Verify that the Patient Remote/Laser Interrupt Switch
is connected and provided to the patient.
• To view applicator information, press the Applicator
Information icon:
- Touch the body part you wish to treat.
- Information about applicator sizes, frequency,
and treatment directions will display. Press the
Down icon to scroll through the information.
- Press the Back Arrow icon to return to the
previous screen or the Home icon to return to the
Home screen.
• Clockwise – Increases duration
• Counterclockwise – Decreases duration
7. Press the Start button to begin therapy, the Pause
button to pause treatment, or the Stop button to
terminate the treatment.
8. When treatment has completed, the Treatment
Summary screen will appear with the following
options:
• Save therapy information to a USB flash drive by
inserting a USB flash drive into the device and
pressing the Save to USB Flash Drive icon.
• Repeat the treatment by pressing the Run This
Treatment icon.
• Save the treatment protocol by pressing the Save
Protocol icon. Refer to the CUSTOM PROTOCOLS
section on page 59 for detailed instructions for
saving customized settings.
57
Page 60
DEVICE USER INTERFACE
SAVING TO USB FLASH DRIVE/PATIENT DATA
Vectra® Neo Clinical Therapy System
Patient treatment data can be saved to the USB ash drive
for viewing/printing on a PC, as well as for retrieving for
later use on the unit or multiple units.
Complete the following steps to view and access patient
data:
1. Press the Patient Data icon on the Home screen. The
screen will display a list box of all previously saved
patient data sessions found on the USB flash drive
connected to the unit.
2. Select the patient ID from the list box by using the Up
and Down arrows or Previous Page/Next Page icons to
locate the patient ID information. Select the patient ID
you wish to view and access by pressing the “√” icon.
3. Select the treatment date desired from the list box by
using the Up and Down arrows or Previous Page/Next
Page icons to locate the treatment date information.
Select the treatment date you wish to view and access
by pressing the “√” icon.
4. The Treatment Summary list box will appear with
detailed information about the specific treatment.
5. Choose one of the following options from the
Treatment Summary screen:
• Run the treatment by pressing the Run this
Treatment icon.
• Save the treatment protocol by pressing the Save
Protocol icon. Refer to the CUSTOM PROTOCOLS
section on page 59 for detailed instructions for
saving customized settings.
• View the pain information by pressing the View Pain
Information icon.
• Save therapy information to USB flash drive by
inserting a USB flash drive and pressing the Save
Protocol icon.
58
Page 61
Vectra® Neo Clinical Therapy System
CUSTOM PROTOCOLS
DEVICE USER INTERFACE
A new custom protocol may be saved at either the
Treatment Edit or Treatment Complete screen. The Vectra®
Neo Clinical Therapy System will allow a maximum of 200
custom protocols to be de ned, and the Home screen will
allow for 5 custom protocol shortcut assignments.
Complete the following steps to save a customized
protocol created from a Treatment Summary screen:
1. Press the Save Protocol icon from either the Treatment
Edit or the Treatment Complete screen.
3. After saving the protocol, you will return to the
Treatment Summary screen. Complete one of the
following actions listed on the Treatment Summary
screen:
• Return to Home screen by pressing the Home icon.
• Scroll back one screen at a time by pressing the
Back Arrow icon.
• Run the treatment again by pressing the Run this
Treatment icon.
• Save therapy information to a USB flash drive by
inserting a USB flash drive into the device and
pressing the Save to USB Flash Drive icon.
• Clear the treatment from “Completed” to “Available”
on the involved channel by pressing the Clear this
Treatment icon.
NOTE: The treatment protocol cannot be saved or run
once Clear the Treatment is selected.
2. Save the new protocol by pressing the Save as a New
Name icon and typing in a name for the protocol, or
overwrite a currently saved protocol by scrolling to the
protocol using the Up and Down arrow icons. Press
the Save icon when finished typing in the protocol
name. A yellow text box will appear confirming the
newly saved protocol and name.
59
Page 62
DEVICE USER INTERFACE
CUSTOM PROTOCOLS CONTINUED
Vectra® Neo Clinical Therapy System
Complete the following steps to delete a previously saved
protocol:
1. From the Home screen, press the Custom Protocols
icon.
2. Find the previously saved protocol in the list box by
using the Up and Down arrows or Previous Page/Next
Page icons to locate the protocol. Select the protocol
you wish to delete and press the Delete icon at the top
of the screen (shown below).
Complete the following steps to assign a Home screen
shortcut for a customized protocol:
1. Press one of the 5 Assign Shortcut icons on the Home
screen.
2. Find the previously saved protocol in the list box by
using the Up and Down arrows or Previous Page/Next
Page icons to locate the protocol. Select the protocol
for which you wish to create a shortcut by pressing the
“√” icon.
3. A confirmation prompt will appear asking, “Are you
sure you want to delete protocol _____”. Press the Yes
icon to delete the protocol.
4. Return to Home screen by pressing the Home icon, or
press the Back Arrow icon to scroll back one screen at
a time.
3. A yellow text box will appear confirming the newly
saved protocol and name, and you automatically
return to the Home screen. The newly saved protocol
will now appear on the Home screen as a unique icon.
60
Page 63
Vectra® Neo Clinical Therapy System
CUSTOM PROTOCOLS CONTINUED
DEVICE USER INTERFACE
Complete the following steps to deassign a Home screen
shortcut for a customized protocol:
1. From the Home screen, press and hold the shortcut
icon you wish to deassign.
NOTE: You can only deassign a shortcut if that
shortcut currently has a protocol assigned to it.
2. The unit will display a text box asking, “Are you sure
you want to de-assign protocol ______ shortcut?”
3. Press the No icon to quit the deassignment process
and return to the Home screen or press the Yes icon to
continue with the deassignment process.
4. After pressing the Yes icon to continue the
deassignment process, a confirmation message will
display for 5 seconds and return to the Home screen.
The previously assigned shortcut will no longer
appear on the Home screen.
Complete the following steps to view and access patient
data:
1. Press the Patient Data icon on the Home screen.
2. Select the patient ID from the list box by using the Up
and Down arrows or Previous Page/Next Page icons to
locate the patient ID information. Select the patient ID
you wish to view and access by pressing the “√” icon.
3. Select the treatment date desired from the list box by
pressing the Up and Down arrows or Previous Page/
Next Page icons to locate the patient ID information.
Select the treatment date you wish to view and access
by pressing the “√” icon.
4. The Treatment Summary list box will appear with
detailed information about the specific treatment.
5. Choose one of the following options from the
Treatment Summary screen:
• Run the treatment by pressing the Run the
Treatment icon.
• View the pain information by pressing the View Pain
Information icon.
61
Page 64
DEVICE USER INTERFACE
ANATOMICAL LIBRARY
Vectra® Neo Clinical Therapy System
The Vectra® Neo Clinical Therapy System contains a
unique Anatomical Library designed to aid the operator
in visually understanding and locating speci c muscle
groups and commonly found problems associated
with pathological conditions, as well as providing an
educational tool for the clinician to use with the patient.
Complete the following steps to view the Anatomical
Library:
1. Press the Anatomical Library icon on the Home screen.
2. Touch the body part for which you wish to view
information. Choose either anterior (on left of screen)
or posterior (on right of screen).
3. Select an image from either the Anatomical Image list
box or the Pathological Image list box by using the Up
and Down arrows or Previous Page/Next Page icons to
locate the desired image. Select the image you wish to
view and access by pressing the “√” icon.
4. View the selected image and, if desired, print the
selected image by pressing the printer icon:
5. Press the Back arrow icon to scroll to the previous
screen or Home icon to return to the Home screen.
62
Page 65
Vectra® Neo Clinical Therapy System
TROUBLESHOOTING
TROUBLESHOOTING CODES
The Vectra® Neo Clinical Therapy System incorporates information, warning, and error messages to inform the user
of problems or potential problems with the system, modality, or accessories. The numbering sequence is: 100-level
messages are general use information messages, 200-level messages are warnings, and 300-level messages are
errors. Use the following Troubleshooting Chart to de ne the code and locate the probable cause and possible remedies
before contacting your dealer or the DJO o ce.
Code
Type Message Probable CausePossible Remedies
Number
100Information
101InformationAttempting to save a Custom protocol with a blank protocol nameEnsure protocol name is not blank prior to saving the protocol.
102InformationAttempting to save a Sequence with a blank sequence nameEnsure sequence name is not blank prior to saving the sequence.
103InformationAttempting to save the Clinic Name with a blank nameEnsure Clinic name is not blank prior to saving the Clinic name.
104InformationAttempting to delete a factory sequenceFactory sequences (names begin with a “*”) cannot be deleted.
105InformationAttempting to delete a factory protocolFactory protocols (names begin with a “*”) cannot be deleted.
106Information
107
108
109
110InformationUltrasound applicator became unplugged during a treatment
111InformationNo ultrasound applicator is plugged into the unit
112Information
113Information
114Information
115Information
116InformationNo laser applicator is plugged into the unit
117InformationLaser Applicator became unplugged during a laser treatment
118InformationIncorrect Laser PIN was enteredEnter the correct laser PIN when prompted prior to starting a laser treatment.
119Information
120Information
Information
Information
Information
Attempting to save a treatment to a USB ash drive with a blank Patient
ID
Attempting to save a Custom protocol or Sequence after system memory
has reached the maximum allowed (200)
Attempting to access protocols or sequences and none are found in the
system
Attempting to select a treatment but available channels for desired
modality are currently in use
Attempting to select an sEMG treatment but sEMG channels are currently
in use
Attempting to select an sEMG treatment but no sEMG is installed on the
Stim 1/2 module.
Attempting to select a Laser treatment while another treatment is
running
Attempting to select a laser treatment but no laser module is installed in
the unit
Attempting to start a Laser treatment but the remote control is not
plugged into the unit
Type in or select a Patient ID prior to saving treatment to a USB ash drive.
Delete unused User protocols or sequences.
1. Custom protocols: No Custom protocols have been saved in the system. Refer to
the CUSTOM PROTOCOLS section on page 59.
2. Sequences: No User sequences have been saved in the system. Refer to the
SEQUENCING OPERATION section on page 50.
Ensure ultrasound applicator is securely connected to the unit prior to starting an
ultrasound treatment.
Ensure desired ultrasound applicator is securely connected to the unit prior to
selecting an ultrasound treatment.
Complete existing treatment before attempting to start another treatment on the
same channel.
Complete existing treatment before attempting to start another treatment on the
same channel.
1. If sEMG is installed on the Stim 1/2 module, reset the Therapy System by turning
the Therapy System O then back On via the Power button.
2. If sEMG is installed on the Stim 1/2 module and the problem persists, discontinue
use of the device and contact the dealer or factory for technical service.
Laser treatments cannot be run concurrently with another treatment. Complete
existing treatment(s) before attempting to select a laser treatment.
Ensure desired laser applicator is securely connected to the unit prior to selecting a
laser treatment.
Ensure desired laser applicator is securely connected to the unit prior to selecting a
laser treatment.
1. If laser module is installed in the unit, ensure module is securely inserted.
2. If laser module is not installed, contact the dealer or factory representative to
purchase one.
Prior to starting a laser treatment, the remote control must be plugged into the unit
and given to the patient to allow use as an emergency stop.
63
Page 66
TROUBLESHOOTING
TROUBLESHOOTING CODES CONTINUED
Vectra® Neo Clinical Therapy System
Code
Type MessageProbable CausePossible Remedies
Number
121Information
122Information
123Information
124
125
126
127
128
129
130
132Information
133Information
Information
Information
Information
Information
Information
Information
Information
Attempting to select an electrotherapy treatment but no Stim
module is installed in the unit
Attempting to select a Combination treatment but no Stim module
#1 (channels 1 and 2) is installed in the unit
Attempting to select an Ultrasound or Combination treatment but
no Ultrasound module is installed in the unit
Attempting to perform an action requiring the USB ash drive
but no USB ash drive is plugged into the unit or USB ash drive
cannot be accessed correctly
Attempting to perform an action requiring a particular le on the
USB ash drive but the le does not exist on the ash drive
Attempting to perform a Print Screen to the USB ash drive while a
treatment is running
Attempting to import protocols from the USB ash drive while a
treatment is running
1. If a Stim module is installed in the unit, ensure module is securely inserted into the
unit.
2. If a Stim module is not installed, contact the dealer or factory to purchase one.
1. If Stim module #1 is installed in the unit, ensure module is securely inserted into
the unit.
2. If Stim module #1 is not is installed, contact the dealer or factory to purchase one.
1. If Ultrasound module is installed in the unit, ensure module is securely inserted into
the unit.
2. If Ultrasound module is not installed, contact the dealer or factory to purchase one.
1. Ensure valid USB ash drive is rmly inserted into the unit or
2. Tr y a di erent USB ash drive.
1. Ensure USB ash drive contains the correct le for the desired function.
2. Ensure valid USB ash drive is rmly inserted into the unit or
3. Tr y a di erent USB ash drive.
The Print Screen function is not allowed while a treatment is running. Wait for
treatment to complete and then try again.
The Protocol Import function is not allowed while a treatment is running. Wait for
treatment to complete and then try again.
1. Reset the Therapy System by turning the power switch O and On.
203WarningError reading a protocol from internal storage
210
211
212
213
216WarningError reading a protocol from internal storage
233
234
235
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Laser application is out of calibration
Electrotherapy - bad electrode contact, overcurrent, or shorted
leads condition has been detected
2. If problem persists after resetting the unit, select the Restore Default Protocols in
the Utilities screen. Refer to the UTILITIES AND OPTIONS section on page 44.
3. If problem persists, discontinue use of the device and contact the dealer or factory
for technical service.
1. Discontinue use of this laser applicator and contact the dealer or factory for
technical service to have the laser applicator recalibrated.
2. If problem exists on multiple laser applicators then discontinue use of laser and
contact the dealer or factory for technical service.
1. Reset the Therapy System by turning the power switch O and On.
2. Restore the Default Protocols in the Utilities screen. Refer to the UTILITIES AND
OPTIONS section on page 44.
3. If problem persists, discontinue use of the device contact the dealer or factory for
technical service.
1. Ensure proper electrode placement use guidelines are followed.
2. Check electrode lead wire connections both at the unit and at the electrode.
3. Replace electrodes.
4. Replace lead wires.
64
Page 67
Vectra® Neo Clinical Therapy System
TROUBLESHOOTING CODES CONTINUED
TROUBLESHOOTING
Code
Number
Any 200-level
code not listed
above
Any 300-level
code
Type MessageProbable CausePossible Remedies
Warning
Error
Immediately stop all use of the system and contact the dealer or DJO for service. Warnings in these categories indicate an internal problem
with the system that must be tested and corrected by DJO or a Trained Technician before any further operation or use of the system.
Use of a system that indicates a Warning in these categories may pose a risk of injury to the patient, user, or extensive internal damage to the
system.
Immediately stop all use of the system and contact the dealer or DJO for service. Errors in these categories indicate an internal problem with
the system that must be tested and corrected by DJO or a Trained Technician before any further operation or use of the system.
Use of a system that indicates an Error in these categories may pose a risk of injury to the patient, user, or extensive internal damage to the
system.
65
Page 68
ACCESSORIES
Vectra® Neo Clinical Therapy System
REPLACEMENT ACCESSORIES
The following provides users of the Vectra® Neo Clinical Therapy System the necessary information to order replacement
accessories used with the system. This list of replacement accessories is designed for use with the Vectra® Neo Clinical
Therapy System. When ordering, provide the respective part number, description, and quantity desired.
ELECTRODES
Model NumberDescription
42045ELECTRODE DURA-STICK 2.75”X 5”
42159DURA-STICK+ 1.5”X2.5” OV 40/CASE
42160DURA-STICK+ 2”X4”OV 40/CASE
42170DURA-STICK PREM 1.25” RND 40/CASE
42171DURA-STICK PREM 2” RND 40/CASE
42172DURA-STICK P 1.5”X2.5” OV 40/CASE
42173DURA-STICK PREM 3”X5” OV 20/CASE
42174DURA-STICK PREM 2” SQ 40/CASE
42175DURA-STICK P 2”X3.5” REC 40/CASE
42176DURA-STICK BG 2” SQ 40/CA
42177DURA-STICK BG 1.5”X3.5” REC 40/CASE
42178DURA-STICK 2” SQ 40/CASE
42179DURA-STICK 2”X3.5”REC 40/CASE
42181DURA-STICK+ 2”X3.5” REC 40/CASE
42182DURA-STICK+ 2”RND 40/CASE
42183DURA-STICK+ 2” SQ 40/CASE
GENERAL ACCESSORIES
Model NumberDescription
70000Electrotherapy Module Channels 1/2
70001Neo Cart
70002Ultrasound Module
70003Electrotherapy Module Channels 3/4
70004Electrotherapy Module Channels 1/2 + sEMG
70005Laser Module
70008PATIENT REMOTE/LASER INTERRUPT SWITCH
79977 HIGH VOLT PROBE KIT- Includes Probe and Sponge Applicator Tips (15 and 8 mm)
With the system disconnected from the power source,
clean the system with a clean, lint-free cloth moistened
with water and mild antibacterial soap. If a more sterile
cleaning is needed, use a cloth moistened with an
antimicrobial cleaner.
Do not submerse the system in liquids. Should the unit
accidentally become submersed, contact the dealer or
DJO Service Department immediately. Do not attempt to
use a system that has been wet inside until inspected and
tested by a Trained Technician.
Do not allow liquids to enter the ventilation holes in the
optional modules. This could permanently damage the
modules.
Cleaning the LCD Screen
Clean the Therapy System LCD with a clean, dry cloth, in
the same way as cleaning the computer monitor screen.
Do not use abrasive materials or chemicals or liquids.
FUSE INFORMATION
3.15A, 250V AC
INSTRUCTION FOR SOFTWARE UPGRADE
1. Obtain a USB flash drive with the upgrade file in root
directory.
2. Power on Unit with the USB flash drive inserted in the
USB port. Allow initialization to complete.
3. Enter the Utilities screen by pressing the Utilities icon
on the Home screen.
4. On the Utilities screen, press the Display Unit Version
Information icon.
5. Press the Upgrade Unit Software from USB icon to
apply any upgraded software packages to the unit.
COPY OF MANUAL
CALIBRATION REQUIREMENTS
Annual factory calibration is required for all Ultrasound
and Laser Applicators. Only the Applicators should be sent
to the factory or a Trained Technician for this procedure.
NOTE: The unit was calibrated during the manufacturing
process and is ready to be placed into service upon
delivery.
DEVICE DISPOSAL
Council Directive 2012/19/EU concerning Waste
Electrical and Electronic Equipment (WEEE)
requires not to dispose of WEEE as municipal
waste. Contact your local distributor for
information regarding disposal of the unit and
accessories.
To obtain a copy of the Vectra® Neo Clinical Therapy System
User Manual, Item #13-7646 (CD Version, Item #13-7647),
contact your local representative or DJO Global Customer
Care.
68
Page 71
Vectra® Neo Clinical Therapy System
WARRANTY REPAIR/OUT OF WARRANTY REPAIR
Service
SERVICE AND WARRANTY
When the
accessory modules require service, contact the selling dealer
or DJO Service Department.
All Therapy System and accessory modules returned to
the factory for service must include the following:
1. Written statement containing the following
2. Copy of original invoice issued at purchase of the
3. Ship the unit to address specified by an authorized
Service to these units should be performed only by a
service technician certi ed by the Company.
Vectra® Neo Clinical Therapy System
information:
• RA Number – Obtain from DJO
• Therapy System or Module Model Number
• Therapy System or Module Serial Number
• Contact Person with Phone and Fax Numbers
• Billing Address (for Out of Warranty Repair)
• Shipping Address (Where to Ship Unit after Repair)
• Detailed Description of Problem or Symptoms
Therapy System or Module
Service Technician
or any of the
Ultrasound Applicators require annual calibration, from
the date placed in service, by the Factory or a Trained
Technician.
Through the purchase of a Service Manual, DJO, LLC
has made available circuit diagrams, component part
lists, descriptors, or other information which will assist
authorized technical personnel to repair those parts
of the equipment which are designated by DJO, LLC as
repairable.
Expected Life
Expected life of the applicator is 5 years and the control
unit is 10 years. Yearly calibration of the device would
extend the life of the device for as long as servicing is
available by DJO or other factory certi ed personnel.
69
Page 72
SERVICE AND WARRANTY
WARRANTY
Vectra® Neo Clinical Therapy System
DJO, LLC (“Company”) warrants that the Vectra® Neo
Clinical Therapy System, Channel 1/2 Electrotherapy
Module, Channel 1/2 Electrotherapy+EMG Module,
Channel 3/4 Electrotherapy Module, Laser Module, and
Ultrasound Module (“Products”) are free of defects in
material and workmanship. This warranty shall remain in
e ect for two years (24 months) from the date of original
consumer purchase. If these Products fail to function
during the two year warranty period due to a defect
in material or workmanship, at the Company’s option,
Company or the selling dealer will repair or replace the
respective Product without charge within a period of
thirty (30) days from the date on which the Product is
returned to the Company or the dealer.
All repairs to the Product must be performed by a service
center authorized by the Company. Any modi cations or
repairs performed by unauthorized centers or groups will
void this warranty.
The warranty period for certain accessories is 90 days.
Accessories consist of Lead Wires, Patient Remote/Laser
Interrupt Switch, and Electrodes.
The warranty period for the Therapy System Cart, Laser
Applicators and Ultrasound Applicators is one year
(12 months).
To participate in warranty coverage, this Product’s
warranty registration card (included with Product) must
be lled out and returned to the Company by the original
owner within ten (10) business days of purchase.
COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR
INCIDENTAL OR CONSEQUENTIAL DAMAGES.
Some states do not allow the exclusion or limitation
of incidental or consequential damages, so the above
limitation or exclusion may not apply to you.
To Obtain Service From Company or the selling dealer
under this warranty:
1. A written claim must be made within the warranty
period to the Company or the selling dealer. Written
claims made to the Company should be sent to:
DJO, LLC
1430 Decision Street
Vista, CA 92081-8553 USA
T: 1-800-592-7329 USA
F: 1-760-734-5608
and
2. The Product must be returned to the Company or the
selling dealer by the owner.
This warranty gives you speci c legal rights and you
may also have other rights which vary from location to
location.
The Company does not authorize any person or
representative to create for it any other obligation or
liability in connection with the sale of the Product.
Any representative or agreement not contained in the
warranty shall be void and of no e ect.
This Warranty Does Not Cover:
• Replacement parts or labor furnished by anyone other
than the Company, the selling dealer, or a Company
service technician
• Defects or damage caused by labor furnished by
someone other than Company, the selling dealer, or a
Company service technician
• Any malfunction or failure in the Product caused
by product misuse, including, but not limited to,
the failure to provide reasonable and required
maintenance or any use that is inconsistent with the
Product User’s Manual
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE.
70
Page 73
Vectra® Neo Clinical Therapy System
APPENDIX 1
OVERVIEW OF LASER THERAPY
Laser therapy results in energy absorbed into the patient’s
tissue from light, triggering biological changes at a
cellular level, resulting in the following:
• Topical heating for the temporary increase in local
blood circulation
• Temporary relief of minor muscle and joint aches,
pains, and stiffness
• Relaxation of muscles and relief of muscle spasm
• Temporary relief of minor pain and stiffness associated
with arthritis. The dose and frequency of treatment
can be adjusted to produce the desired effect.
There is no exact limit with respect to the penetration
of the light. The light gets weaker the further from the
surface it penetrates. There is, however, a limit at which
the light intensity is so low that no biological e ect of
the light can be registered. This limit, where the e ect
ceases, is called “the greatest active depth.” This depth is
also contingent on tissue type, pigmentation, and dirt
on the skin. For example, fat tissue is more transparent
than muscle tissue and more easily penetrable by Laser
than muscle tissue as blood hemoglobin is an important
absorber of light, and highly vascularized muscle tissue
makes Laser penetration more di cult.
NOTE: Some laser applicators may cause a noticeable
heat sensation, particularly in hairy areas and on sensitive
tissues such as lips.
TREATMENT TIPS
Contact
To obtain the most e ective results, the applicator should
be in contact with the patient’s skin.
Applicator Position
Due to the characteristics of Laser, the angle at which the
light enters the patient’s skin is very important. Therefore,
the applicator lens should always be parallel to the
treatment area.
Treating Joints
If you are applying Laser therapy to a patient’s joint, it is
more e ective to apply the Laser energy into the joint
by positioning the joint in an open position (e.g., knee
in exion). However, do not attempt this method if it is
uncomfortable to the patient.
Applicator Selection
If the injury you intend to treat is very small (pinpoint),
you should only need to treat the area with a single diode
applicator. If the area surrounding the treatment area
is sensitive, it is recommended that you apply therapy
with a single diode applicator rst, then use the cluster
applicator for the surrounding area.
Cold and Heat
If you intend to apply cold or heat as an adjunct to Laser
therapy, use the following guidelines:
• Use cold before Laser therapy. This slows the flow of
red blood cells and reduces the amount of energy
removed from the area.
• Use heat after the treatment. This speeds the flow of
red blood cells so that more energy can be removed
from the area.
71
Page 74
APPENDIX 1
COMMON TERMS
Vectra® Neo Clinical Therapy System
Applicator – Hand held assembly that delivers Laser
energy and includes laser head, diode, and related
electronics
Collimating – The shape of the Laser beam. It is neither
focused or dispersed and resembles a column when
applied
Continuous Mode – The output of the Laser is not
interrupted during the treatment time
Dosage – Intensity measurement of the Laser energy over
the treatment area. It is measured in Joules or Joules/cm
2
Energy – Measured in Joules, energy equals the treatment
time multiplied by the power. Energy density equals the
power output multiplied by the treatment time, and
divided by the spot size (cm2). This gives a more speci c
measurement of energy delivered
Frequency – Pulsed frequencies are selectable from 8 to
10,000 Hz
Laser Head – The clear lens applicator face contacts the
patient’s skin. It consists of laser diodes with or without
LED’s or SLD’s (depending on the applicator)
Treatment Time – Measured in seconds, it is the
suggested time per laser point that therapy is given
Wavelength – Wavelength is measured in nanometers
(nm) and is the key component in obtaining e ective
therapy as di erent wavelengths cause di erent
physiological e ects. Super cial skin disorders are most
e ectively treated at wavelengths 600-700 nm, while
deeper muscular or ligament lesions and joint conditions
are best treated at higher wavelengths of 700-1000 nm.
Power – Measured in Watts (W), power wattage is directly
proportional to treatment time and penetration of the
Laser energy. High-powered diodes will reduce patients’
treatment times and give a higher amount of energy at a
deeper depth. Power output can be either continuous or
pulsed
Power Density - Ratio of power divided by treatment
time
Pulsed Mode - This is the ratio of the “On” time to the
“Total” time of the cycle, expressed as a percentage.
The lower the percentage, the lower temporal average
intensity. 100% is continuous Laser. Pulsed Mode is 90%
on and 10% o
NOTE: Pulsed Mode is also equivalent to Duty Cycle.
Spot Size - Area of the LED, SLD, or laser beam when it
leaves the face of the lens
Treatment Area - Area of tissue a ected by LED, SLD, or
laser when wavelength, divergence angles, and depth of
penetration are factored. This is the area used to calculate
dosage
72
Page 75
Vectra® Neo Clinical Therapy System
ELECTROMAGNETIC COMPATIBILITY EMC
APPENDIX 2
The Vectra® Neo Clinical Therapy System has been
tested and found to comply with the electromagnetic
compatibility (EMC) limits for medical devices to
IEC 60601-1-2. These limits are designed to provide
reasonable protection against harmful interference in a
typical medical installation.
The guidelines below are intended to help promote
electromagnetic compatibility (EMC) in the identi ed use
environment for the Vectra® Neo Clinical Therapy System.
• Make use of available resources such as EMC
professionals and publications and Internet web pages
on the subject of medical device EMC;
• Assess the electromagnetic environment of the facility
(e.g., identify radio transmitters in around the facility)
and identify areas where critical medical devices are
used;
• Manage the electromagnetic environment, RF
transmitters and all electrical and electronic
equipment, including medical devices, to reduce the
risk of medical device electromagnetic interference
(EMI) and achieve EMC;
• Coordinate the purchase, installation, service,
and management of all electrical and electronic
equipment used in the facility to achieve EMC;
• Educate healthcare facility staff, contractors, visitors,
and patients about EMC and EMI and how they can
recognize medical device EMI and help minimize
associated risks;
• Establish and implement written policies and
procedures that document the intentions and
methods of the healthcare institution for reducing the
risk of medical device EMI and achieving EMC;
• Report EMI problems to the US FDA MedWatch
program and communicate EMI/EMC experiences to
colleagues in open forums such as medical/technical
publications and conferences.
More information is contained within a comprehensive
guidance document for EMC in healthcare facilities,
developed, with FDA participation, by the Association
for the Advancement of Medical Instrumentation (AAMI):
Technical Information Report (TIR) 18, Guidance on
Electromagnetic Compatibility of Medical Devices
for Clinical/Biomedical Engineers. AAMI TIR 18-1997.
Arlington, Virginia: Association for the Advancement
of Medical Instrumentation; 1997.
Caution:
Medical electrical equipment requires special precautions
regarding EMC and must be installed and operated
according to these instructions. It is possible that
high levels of radiated or conducted radio-frequency
electromagnetic interference (EMI) from portable
and mobile RF communications equipment or other
strong or nearby radio-frequency sources, could result
in performance disruption of the ultrasound system.
Evidence of disruption may include image degradation or
distortion, erratic readings, equipment ceasing to operate,
or other incorrect functioning. If this occurs, survey
the site of disruption, and take the following actions to
eliminate the source(s).
• Turn equipment in the vicinity off and on to isolate
disruptive equipment.
• Relocate or re-orient interfering equipment.
• Increase distance between interfering equipment and
your ultrasound system.
• Manage use of frequencies close to ultrasound system
frequencies.
• Remove devices that are highly susceptible to EMI.
• Lower power from internal sources within the facility
control (such as paging systems).
• Label devices susceptible to EMI.
• Educate clinical staff to recognize potential EMIrelated problems.
• Eliminate or reduce EMI with technical solutions (such
as shielding).
• Restrict use of personal communicators (cell phones,
computers) in areas with devices susceptible to EMI.
• Share relevant EMI information with others,
particularly when evaluating new equipment
purchases which may generate EMI.
• Purchase medical devices that comply with IEC 606011-2 EMC Standards (3V/meter EMI immunity, limit
interference level to 0.0014 V/meter).
73
Page 76
APPENDIX 2
Vectra® Neo Clinical Therapy System
ELECTROMAGNETIC COMPATIBILITY EMC TABLES
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Vectra® Neo Clinical Therapy System is intended for use in the electromagnetic environment speci ed below. The
customer or the user of the Vectra® Neo Clinical Therapy System should assure that it is used in such an environment.
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
The Vectra® Neo Clinical Therapy System is
suitable for use in all establishments other than
domestic and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Vectra® Neo Clinical Therapy System is intended for use in the electromagnetic environment speci ed below. The
customer or the user of the Vectra® Neo Clinical Therapy System should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD) IEC
61000-4-2
Electrical fast
transient/burst IEC
61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
eld
IEC 61000-4-8
NOTE: U
is the AC mains voltage prior to application of the test level.
Risk assessment on the Vectra Neo Clinical Therapy System
indicates the compliance levels claimed are acceptable when
ESD-precautionary measures are taken.
The VectraNeo Clinical Therapy System may be susceptible to
Electro-Static Discharge (ESD) at greater than ±4 kV when rst
grasping either the Ultrasound or Laser applicator. In the event
of such a discharge, the Vectra Neo Clinical Therapy System
may experience communication loss with the installed modules. The Vectra Neo Clinical Therapy System will terminate all
active outputs (stim, ultrasound, laser), automatically place the
unit in a safe state, and issue an error message 301 or 307.
To recover from an error message 301 or 307, turn the unit
o and on using the ON/OFF switch located at the top of the
display. Once the system restarts, re-initiate all treatments that
were interrupted.
To prevent Electro-Static Discharge (ESD) at greater than ±4 kV:
- Grasp and hold the Ultrasound or Laser Applicator prior
to starting treatment. If the applicator must be put
down prior to completion of treatment, stop the current
treatment first and then place the applicator in the
holder.
- Maintain humidity in the use environment to at least
50% relative humidity.
- Floors should be wood, concrete, or ceramic tile. If floors
are covered with synthetic material, DJO recommends
implementing additional controls to maintain relative
humidity to at least 50%.
- Communicate these ESD-precautionary procedures to
healthcare staff, contractors, visitors and patients.
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply
lines
Mains power quality should be that of a typical commercial or
hospital environment.
±1 kV for input/output
lines
+1 kV di erential mode (line
to line)
±1 kV di erential mode
±2 kV common mode
Mains power quality should be that of a typical commercial or
hospital environment.
+2 kV common mode (line
to ground)
<5% U
(>95% dip in UT) for
T
0,5 cycle
40% U
(60% dip in UT) for
T
5 cycles
70% U
(30% dip in UT) for
T
25 cycles
<5% U
(>95% dip in UT) for
T
5 sec
<5% U
(>95% dip in UT) for
T
0,5 cycle
40% U
(60% dip in UT) for
T
5 cycles
70% U
(30% dip in UT) for
T
25 cycles
<5% U
(>95% dip in UT) for
T
5 sec
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the Vectra® Neo Clinical
Therapy System requires continued operation during power
mains interruptions, it is recommended that the Vectra® Neo
Clinical Therapy System be powered from an uninterrupted
power supply.
3 A/m3 A/mPower frequency magnetic elds should be at levels
characteristic of a typical location in a typical commercial or
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Vectra® Neo Clinical Therapy System is intended for use in the electromagnetic environment speci ed below. The
customer or the user of the Vectra® Neo Clinical Therapy System should assure that it is used in such an environment.
Portable and mobile RF communications
equipment should be used no closer to any part of the
Vectra® Neo Clinical Therapy System, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
3,5
[V
] V, where V1 = 3V
1
] V/m, where E1 = 3V/m
[E
1
d =
V
d =
E
d =
E
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer
and d is the recommended separation distance in metres (m).
Field strengths from xed RF transmitters, as determined
by an electromagnetic site survey,a should be less than the
compliance level in each frequency range.
√P
[
]
1
3,5
√P 80 MHz to 800 MHz
[
]
1
7
√P 800 MHz to 2,5 GHz
[
]
1
b
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is a ected by absorption and re ection from structures,
objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the
Vectra® Neo Clinical Therapy System is used exceeds the applicable RF compliance level above, the Vectra® Neo Clinical Therapy System should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the Vectra® Neo Clinical Therapy System .
b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than [V
portable and mobile RF communications equipment and the Vectra® Neo Clinical Therapy System
The Vectra® Neo Clinical Therapy System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Vectra® Neo Clinical Therapy System can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Vectra® Neo Clinical Therapy System as recommended below, according to the
maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
d (m)
Rated maximum output
power of transmitter
P (W)
0,010,120,120,23
0,10,380,380,73
150 kHz to 80 MHz
3,5
d =
(where V1 = 3V)
√P
[
]
V
1
80 MHz to 800 MHz
3,5
d =
(where E1 = 3V/m)
√P
[
]
E
1
800 MHz to 2,5 GHz
7
d =
(where E1 = 3V/m)
√P
[
]
E
1
11,21,22,3
103,83,87,3
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is a ected by absorption and re ection from structures,
objects and people.
77
Page 80
APPENDIX 3
ξ
ξ
ξ
ξ
ξ
ξ
ξ
ξ
ξ
ξ
ξ
ξ
ξ
ξ
Vectra® Neo Clinical Therapy System
CALCULATIONS FOR OUTPUT CURRENT
The Vectra® Neo Clinical Therapy System controls total output current and output energy to comply with IEC 60601-2-
10. The System also controls total output current density for the selected electrode to comply with FDA requirements.
The Vectra® Neo Clinical Therapy System controls output by dynamically changing the range of controllable/selectable
parameters, based on the user selection of Intensity, Phase Duration, Frequency, and/or Electrode size.
The following tables provide the calculation and guidance used by the system to dynamically control the output and
comply with IEC 60601-2-10 and the 250 mW/cm2 power density-limit required by FDA. All calculations are based on
Resistive Load = 500 ohms and Accuracy of 10% (1.1).
CC: Constant Current
CV: Constant Voltage
Waveform
IEC current
Interferential (F <= 5 kHz) (CC) 100
limit
(in mA)
Output current equation
[I
in mA]
RMS
ܫ כ ܽܿܿݑݎܽܿݕ
ʹ
Intensity [I]
(in mA (CC), in
V (CV))
0 – 100 50 – 250 2000 – 10000 100% N/A
Phase Duration
[T]
(in uSec)
Frequency [F]
(in Hz)
Duty Cycle [D]
Number of
Pulse [N]
Interferential (F > 5 kHz) (CC) 100
Interferential (F <= 5 kHz) (CV) 100
Interferential (F > 5 kHz) (CV) 100
Premodulated (F <= 5 kHz) (CC) 100
Premodulated (F > 5 kHz) (CC) 100
Premodulated (F <= 5 kHz) (CV) 100
Premodulated (F > 5 kHz) (CV) 100
Russian (CC) 100
Russian (CV) 100
VMS™ (CC) 50
VMS™ (CV) 50
ܫ כܽܿܿݑݎܽܿݕ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ
ܫ כ ܽܿܿݑݎܽܿݕ
ʹ
ܫ כ ܽܿܿݑݎܽܿݕ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ
ʹ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ
ܫ כܽܿܿݑݎܽܿݕ כ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ כ
ܫ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ʹ
ʹ
ʹ
ܦ
ʹ
ʹ
ܰ כξܨ כξܶ
ܰ כξܨ כξܶ
0 – 90 50 – 250 2000 – 10000 100% N/A
ܦ
0 – 64 50 – 250 2000 – 10000 100% N/A
0 – 45 50 – 250 2000 – 10000 100% N/A
0 – 100 50 – 250 2000 – 10000 66% N/A
0 – 100 50 – 250 2000 – 10000 66% N/A
0 – 96 50 – 250 2000 – 10000 66% N/A
0 – 68 50 – 250 2000 – 10000 66% N/A
0 – 100 200 2500 50% N/A
0 – 90 200 2500 50% N/A
0 – 200 20 - 1000 1 – 200 100% 2
0 – 200 20 - 1000 1 – 200 100% 2
VMS™ Burst (CC) 50
VMS™ Burst (CV) 50
VMS™ FR (CC) 50
VMS™ FR (CV) 50
ܫ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ܫ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ܰ כξܨ כξܶ
ܰ כξܨ כξܶ
ܰ כξܨ כξܶ
ܰ כξܨ כξܶ
0 – 200 20 – 700 1 – 200 100% 6
0 – 200 20 – 700 1 – 200 100% 6
0 – 150 20 – 400 20 – 80 100% 2
0 – 150 20 – 400 20 – 80 100% 2
78
Page 81
Vectra® Neo Clinical Therapy System
ξ
ඥ
ඥ
ξ
ξ
ξ
CALCULATIONS FOR OUTPUT CURRENT CONTINUED
APPENDIX 3
Waveform
High Voltage Pulsed Current (CV) 80
Microcurrent (CC) 80
TENS Asymmetrical Biphasic (CC) 50
TENS Asymmetrical Biphasic (CV) 50
TENS Symmetrical Biphasic (CC) 50
TENS Symmetrical Biphasic (CV) 50
DC (Direct Current) (CC) 80
HAN-Stimulation 50
IEC current
limit
(in mA)
Output current equation
[I
RMS
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ כ
ܫ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ܫ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ሺܫȀܴሻ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
ܫ כ ܽܿܿݑݎܽܿݕ
ܫ כ ܽܿܿݑݎܽܿݕ כ
ͳͲͲͲ
in mA]
ܰ כξܨ כξܶ
ܦ
N/A
ܨ כ ሺܶ ʹͷͲሻ
ܨ כ ሺܶ ʹͷͲሻ
ܰ כξܨ כξܶ
ܰ כξܨ כξܶ
ܰ כξܨ כξܶ
Intensity [I]
(in mA (CC), in
V (CV))
0 – 500 20 10 – 120 33.33% 2
0 – 1 500 – 500000 0.1 – 1000 50% N/A
0 – 110 20 – 1000 1 – 250 100% N/A
0 – 110 20 – 1000 1 – 250 100% N/A
0 – 80 20 – 1000 1 – 250 100% 2
0 – 80 20 – 1000 1 – 250 100% 2
0 – 72 N/A N/A N/A N/A
0 – 100 180 80 100% 2
Phase Duration
[T]
(in uSec)
Frequency [F]
(in Hz)
Duty Cycle [D]
Number of
Pulse [N]
79
Page 82
APPENDIX 3
Vectra® Neo Clinical Therapy System
MAXIMUM CURRENT OUTPUT CALCULATIONS BASED ON 250 MW/CM2 POWER DENSITY LIMIT
, as there will be a risk of suffering a burn injury. Caution should always be exercised with current densities more than 2 mA/cm
2
Waveform IRMS [mA]
Interferential (F <= 5 kHz) (CC)
Interferential (F > 5 kHz) (CC)
Interferential (F <= 5 kHz) (CV)
Interferential (F > 5 kHz) (CV)
Premodulated (F <= 5 kHz) (CC)
Premodulated (F > 5 kHz) (CC)
Premodulated (F <= 5 kHz) (CV)
Premodulated (F > 5 kHz) (CV)
Russian (CC)
Russian (CV)
VMS™ (CC)
VMS™ (CV)
VMS™ Burst (CC)
VMS™ Burst (CV)
VMS™ FR (CC)
VMS™ FR (CV)
High Voltage Pulsed Current (CV)
Microcurrent (CC)
TENS Asymmetrical Biphasic (CC)
TENS Asymmetrical Biphasic (CV)
TENS Symmetrical Biphasic (CC)
TENS Symmetrical Biphasic (CV)
DC (Direct Current) (CC)
HAN-Stimulation
* Due to the area of electrode 42170, the IRMS value is automatically reduced by the Vectra® Neo Clinical Therapy System
Caution: DO NOT use electrodes with an active area less than 7.92 cm
81
Page 84
APPENDIX 3
Vectra® Neo Clinical Therapy System
ELECTRODE SIZE SELECTION
The Vectra® Neo Clinical Therapy System allows the user to select an electrode size from three di erent options, which
allows the System to more accurately control the output current. Please refer to following table for the selection guide
related to di erent DJO electrode part numbers.
Device Electrode Selection DJO Electrode Part Number
* The maximum output current is limited by the System to a value lower than what the electrode can deliver and
therefore for the purposes of electrode selection in modality setup, all elliptical electrodes are considered as “round” and