TECHNICAL BULLETIN PPE-5034.E
Calibration Unit for Testers Installation, Operation and Maintenance
Operation
COMBO TESTER
Testing the Wrist Strap Circuit
1. Plug one of the included test leads into the black banana jack labeled “E” on the Calibration Unit. Connect the opposite end of the test lead into the yellow banana jack labeled “WRIST STRAP” on the Combo Tester. Toggle the switch to the “WRIST STRAP” position.
Figure 1. Charleswater 99090 Calibration Unit
The Charleswater 99090 Calibration Unit is designed to
verify whether a tester is operating within specifications.
The 99090 Calibration Unit is a passive device and requires no power source. The Calibration Unit is manufactured with industry accepted test ranges for both wrist straps and foot grounders. Both the wrist strap and footwear pass range is set at 750K - 35M. The Charleswater Calibration Unit is calibrated to NIST traceable standards.
The Charleswater 99090 Calibration Unit is to be used with the following items:
Item Description
99037 Combo Tester
99038 Combo Tester with Stand and Foot Plate
Packaging
1 Calibration Unit
2 Test Leads with Banana Plug Terminals
1 Certificate of Calibration
Figure 2. Connecting the test lead from banana jack “E” on the Calibration Unit to the WRIST STRAP banana jack on the Combo Tester
2. Connect the second test lead to the red 675K banana jack labeled “A” on the Calibration Unit. Touch the opposite end of the test lead to test plate on the Combo Tester and press down to activate
the test. Be sure not to touch the banana plug or test plate with your skin.
3. Observe the response from the Combo Tester. The alarm should sound, and the LOW LED should illuminate red.
4. Repeat the procedure for test points B (825K), C (35M) and D (40M). The expected test results can be found in the following table.
Test
Point
Resistance
Value
A 675K FAIL LOW
B 825K PASS
C 35M PASS
D 40M FAIL HIGH
Made in the
United States of America
Test Output
PPE-5034.E Page 1 of 2
Unit 17. Millbrook Business Park, Sybron Way • Crowborough, East Sussex TN6 3JZ United Kingdom
Phone: 00 44 (0) 1892-665313 • Internet: Charleswater.co.uk
August 2015
© 2015 Charleswater
Page 2
Figure 3. Verifying the calibration of the wrist strap circuit in the Combo Tester
Testing the Footwear Circuit
1. Plug one of the included test leads into the black banana jack labeled “E” on the Calibration Unit. Connect the opposite end of the test lead into the yellow banana jack labeled “FOOT PLATE” on the Combo Tester. Toggle the switch to the “FOOT PLATE” position.
Figure 4. Connecting the test lead from banana jack “E” on the Calibration Unit to the FOOT PLATE banana jack on the Combo Tester
2. Connect the second test lead to the red 675K banana jack labeled “A” on the Calibration Unit. Touch the opposite end of the test lead to test plate on the Combo Tester and press down to activate
the test. Be sure not to touch the banana plug or test plate with your skin.
Figure 5. Verifying the calibration of the footwear circuit in the Combo Tester
Specifications
Accuracy:
±5%
Dimensions:
8.1 cm x 11.2 cm x 2.3 cm
Weight:
.1 kg
Limited Warranty, Warranty Exclusions, Limit of Liability and RMA Request Instructions
See Charleswater’s Warranty -
http://www.charleswater.co.uk/LimitedWarranty.aspx
3. Observe the response from the Combo Tester. The alarm should sound, and the LOW LED should illuminate red.
4. Repeat the procedure for test points B (825K), C (35M) and D (40M). The expected test results can be found in the following table.
Test
Point
Resistance
Value
Test Output
A 675K FAIL LOW
B 825K PASS
C 35M PASS
D 40M FAIL HIGH