CHARDER MEDICAL MS 3910, DP3700 User Manual

- 0 -
MS 3910 (DP3700)
Medical Scale
USER MANUAL
Please keep the instruction manual at hand all the time for future reference.
1
TABLE OF CONTENTS
PREFACE.....................................................................................................2
GENERAL INFORMATION.......................................................................2
SAFETY INSTRUCTION ...........................................................................2
ENVIROMENTAL.......................................................................................3
CLEANING..................................................................................................3
MAINTENANCE.........................................................................................3
WEIGHING OPERATION ..........................................................................3
WARRANTY-LIABILITY...........................................................................4
DISPOSING OF THE SCALE.....................................................................5
EXPLANATION OF THE GRAPHIC SYMBOLS.....................................6
EMC guidance and manufacturer's declaration............................................8
SPECIFICATION.......................................................................................12
POWER ADAPTOR STANDARDS..........................................................13
PANEL........................................................................................................14
KEY FUNCTION.......................................................................................15
LCD SYMBOL DESCRIPTION................................................................16
TIME SETTING.........................................................................................17
USING SCALE..........................................................................................17
USING HOLD FUNCTION.......................................................................18
USING BMI FEATURE.............................................................................19
USING TARE FUNCTION........................................................................20
USING INTERNAL SETTING .................................................................21
STORING AND RE-CALLING PRE-TARE WEIGHT............................22
PRINTING FUNCTION............................................................................25
INSTRUCTION FOR CHARGING AND CONNECTING......................27
INSTRUCTION FOR REPLACING BATTERIES...................................28
MS 3910 ASSEMBLY................................................................................31
SSEMBLING HEIGHT ROD HM201M/201D/210D ON WIDE
COLUMN...................................................................................................32
ASSEMBLING TP21xx SERIES THERMO PRINTER ON SCALE.......34
ERROR MESSAGE...................................................................................35
TROUBLESHOOTING.............................................................................36
2
PREFACE Thank you for choosing CHARDER MEDICAL product. All features
of this product were designed to state of the art and are optimized for simple and straightforward use. If you have any queries or experience any problems not addressed in the operating instructions, please contact your CHARDER MEDICAL service partner, or visit us on the Internet at www.chardermedical.com
GENERAL INFORMATION We strongly recommend you use the scales on flat and hard surface.
Any soft surface, like carpet will cause inaccuracy. SAFETY INSTRUCTION
Before putting the device into use, please read with care the
information given in the Operating Instructions. They contain important instructions for installation, proper use and maintenance of the device.
The manufacturer shall not be liable for damages arising out of failure to heed the following instructions: These batteries should be kept away from small children. If
swallowed, promptly seek medical assistance.
Expected Service Life: 5 years  When using electrical components under increased safety
requirements, always comply with the appropriate regulations.
Improper installation will render the warranty null and void.  Ensure the voltage marked on the power supply unit matches
your main power supply.
This device is designed for use indoors.  Observe the permissible ambient temperatures for use  The device meets the requirements for electromagnetic
compatibility. Do not exceed the maximum values specified in
the applicable standards. If you have any question, contact your local CHARDER MEDICAL service partner.
3
ENVIROMENTAL All batteries contain toxic compounds; disposal of batteries
should be delegated to a competent organisation, complying with the deposit of Poisonous Waste Regulation 1972.
Please do not incinerate batteries. The optimum operating temperature for the scale is 5 to +35
; although it will operate at higher and lower temperatures the scales battery life will be adversely effected.
CLEANING We would recommend using alcohol based wipes or similar
when cleaning the scales. Please do not use large amounts of water when cleaning the scales as this will cause damage to the scales electronics, you should also refrain from using corrosive liquids or high pressure washers. Always disconnect the scales from the mains power supply before cleaning.
MAINTENANCE The scale does not require any routine maintenance. However,
we recommend checking the scale’s accuracy at regular intervals. The regularity of these checks is dependent on the level of use and the state of the scale. If any inaccuracies occur, please contact your local dealer or CHARDER MEDICAL service partner.
WEIGHING OPERATION Before reading detailed instructions on how to use all the weighing
functions that are built into your scale, please read the following
4
important guidelines: Always be sure that the display shows `Zero` before use, if it does
not then please press the ZERO key.
The device is designed to detect when a stable weight is achieved,
the indicator will `bleep` twice to indicate a stable weight value, your reading should be taken at this point.
WARRANTY-LIABILITY If a fault or defect is present on receipt of the unit which is within
CHARDER MEDICAL’s scope of responsibility, CHARDER shall have the right to either repair the fault or supply a replacement unit. Replaced parts shall be the property of CHARDER. Should the fault repairs or replacement delivery not be successful, the statutory provisions shall be valid. The period of warranty shall be two years, beginning on the date of purchase. Please retain your receipt as proof of purchase. Should your scale require servicing, please contact your dealer or CHARDER MEDICAL Customer Service. No responsibility shall be accepted for damage caused through any of the following reasons: Unsuitable or improper storage or use, incorrect installation or commissioning by the owner or third parties, natural wear, changes or modifications, incorrect or negligent handling, overuse, chemical, electrochemical or electrical interference or humidity, unless this is attributable to negligence on the part of CHARDER MEDICAL. If operating, climatic or any other influences lead to a major change in conditions or material quality, the treaty for perfect unit functioning shall be rendered null and void. If CHARDER provides and individual warranty, this means that the unit supplied will be free of faults for the length of the warranty period.
5
DISPOSING OF THE SCALE This product is not to be treated as regular household waste, but
should be handed in to an electrical/electronic equipment recycling centre. You can obtain further details from your local council, your municipal waste disposal company or the firm which you purchased the product.
6
EXPLANATION OF THE GRAPHIC SYMBOLS
SN-T13000001
Designation of the serial number of every device, applied at the device.
(Number as an example) “Please note the accompanying
documents“ or “Observe operating instructions”
Charder Electronic Co., Ltd.
No.103, Guozhong Rd., Dali Dist., Taichung City 412, Taiwan (R.O.C.)
Identification of manufacturer of medical product including address
Type B applied part
Dispose of old appliances separately from your household waste!l! Instead, take them to communal collection points.
Carefully read this operation manual
before setup and commissioning, even if you are already familiar with Charder scales.
7
Transport and storage temperature limit indicating the upper and the lower limit (Transport and storage temperature on packaging)
-20°C
+60°C
8
EMC guidance and manufacturer's declaration
Guidance and manufacturer’s declaration-electromagnetic emissions
The MEDICAL SCALE MS3910 is intended for use in the electromagnetic environment specified below. The customer or the user of the MEDICAL SCALE MS3910 should assure that it is used in such an environment.
Emission test Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The MEDICAL SCALE MS3910 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations /flicker emissions IEC 61000-3-3
Compliance
The MEDICAL SCALE MS3910 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Guidance and manufacturer’s declaration-electromagnetic immunity The MEDICAL SCALE MS3910 is intended for use in the electromagnetic environment specified below. The customer or the user of the MEDICAL SCALE MS3910 should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance
9
Electrostatic discharge(ESD) IEC 61000-4-2
± 6 kV contact ± 8 kV air
± 6 kV contact ± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%
Electrical fast transient/burst IEC 61000-4-4
± 2kV for power supply lines + 1kV for input/output lines
± 2kV for power supply lines Not applicable
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
± 1kV line(s) to line(s) ± 2kV line(s) to earth
± 1kV differential mode Not applicable
Mains power quality should be that of a typical commercial or hospital environment.
Voltage Dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5% UT(>95% dip in UT) for 0,5 cycle 40% UT(60% dip in UT) for 5 cycles 70% UT(30% dip in UT) for 25 cycles <5% UT(>95% dip in UT) for 5 s
<5% UT(>95% dip in UT) for 0,5 cycle 40% UT(60% dip in UT) for 5 cycles 70% UT(30% dip in UT) for 25 cycles <5% UT(>95% dip in UT) for 5 s
Mains power quality should be that of a typical commercial or hospital environment. If the user of the MEDICAL SCALE MS3910 requires continued operation during power mains interruptions, it is recommended that the MEDICAL SCALE MS3910 be powered from an uninterruptible power supply or a
battery. Power frequency(50/60 Hz) magnetic field IEC 61000­4-8
3 A/m 3 A/m
The MEDICAL SCALE
MS3910 power
frequency magnetic
fields should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment. NOTE UT is the a.c. mains voltage prior to application of the test level.
10
Guidance and manufacturer’s declaration-electromagnetic immunity The MEDICAL SCALE MS3910 is intended for use in the electromagnetic environment specified below. The customer or the user of the MEDICAL SCALE MS3910 should assure that is used in such and environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidance
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms 150 KHz to 80 MHz
3 V/m 80MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of the
MEDICAL
SCALE
MS3910 including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 P d = 1,2 P 80MHz to 800 MHz d = 2,3 P 800MHz to 2,5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya, should be less than the compliance level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked with the following symbol:
11
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be les
than 3 V/m.
Recommended separation distance between portable and mobile RF
communications equipment and the MEDICAL SCALE
The MEDICAL SCALE MS3910 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the MEDICAL SCALE MS3910 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the MEDICAL SCALE MS3910 as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter m
Rated maximum output
power of transmitter
W
150 kHz to 80 MHz
d =1,2P
80 MHz to 800 MHz
d =1,2P
800 MHz to 2,5 GHz
d =2,3P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
12
SPECIFICATION
Model MS 3910
Capacity 300kg x 0.1kg
Accuracy ±0.15kg
OIML Approval
Class III
Weight Unit kg
LCD Display 1.0 inch LCD display with 5 and 1/2 digits
Dimension
Pole Height: 800 mm
Platform size: 340 x 450 mm
Key Functions
ON/OFF, ZERO, PRINT, BMI, HOLD, PRE-
TARE, TARE , CLEAR, ENTER, 0~9 , M1-5
Power Supply
6-AA size batteries 12V AC Adaptor Rechargeable battery pack (optional).
Operation
Temperature
and Humidity
5 - 35
15% - 85% RH
Storage
Temperature
and Humidity
- 20°C … + 60°C 10% - 95% RH
HM201M manual height rod
HM200D digital height rod
HM201D digital height rod without display
Options
Thermo Printer TP2100
13
POWER ADAPTOR STANDARDS
CAUTION: The device is only compatible with restricted power
adaptors in dashed block below.
14
Standard Accessories
No.
Accessories Item Spec.
Qty.
1
RS232 cable WR-8159 1
2
12V Adaptor 1
3
MS3910 user
manual
IN-1178 1
4
Screw
(Optional: for
wheel version)
M4*20 4
PANEL
15
KEY FUNCTION
Key Description
ON/OFF
Turn ON/OFF the scale
PRINT
Data transmission via interface (Print-out the results)
BMI
To determine Body Mass Index
HOLD
To determine stable weighing value in case of unstable weight.
ZERO
To reset the display to 0.0kg display / Zero the scale (±2% of full capacity)
M1-5
To store 5 pre-tare values (Approval Model).
PRE-TARE
Pre-tare function for subtraction of a known container weight or any other object before weighing process.
TARE
Tare allows the user to zero the instrument to cancel the weight of a container/ clothes from the reading of the instrument, thus giving the true weight of the product/ person being tested, while weighing.
CLEAR
To clear the wrong entry while entering digits.
ENTER
To confirm the performing functions.
0-9
Entering digits
16
LCD SYMBOL DESCRIPTION
Definitions
Stable symbol: To indicate that the weight is stable. Minor weight: Weight under zero. Zero symbols: Weight is at zero point.
Low battery: Battery need to charge or replaced.
17
TIME SETTING Long press HOLD key for 3 seconds to enter the TIME SETTING
mode, beginning with the top row, with the flashing digit. Press HOLD key after successful change to move to next step. EX: To input Dec 25, 2008, 8:00am.
Enter year using keys from 0~9. Press HOLD key after successful entry to move to next.
Enter date. Ex: 12/25 enter “12.25” Press HOLD key after successful entry to move to next.
Enter time.
Display Format YYYYMM.DDTT:SS
USING SCALE
Switch on the scale using key. The diagnostic scale self-
check is performed and the software version is displayed.
The „0,00 kg” weight displays on the screen, now the scale is
ready for weighing.
Note: If „0,00 kg” won’t display on the screen, press
key to
zero the scale, can be used any time to zero the scale.
18
Place a person in the middle of the scale. Wait until the scale
stabilizes and stable sign (ο) shows on the screen
Direction:
If a person is heavier than the scale capacity, the display will show the „Err” prompt (= overload).
USING HOLD FUNCTION MS 3910 is provided with the integrated hold function to determine
the average weight. It enables people to be weighed accurately although they are not still on the scale platform.
Note: Determining average weight is not possible incase of big fluctuation between weight.
Switch ‘ON’ the scale using [ON/OFF] key. The diagnostic self-
check will be performed and after that „0.0 kg” will display on the screen with ‘stable’ and ‘zero sign’.
Move the object/ person on the scale platform..  Press the [HOLD] key. The ‘HOLD’ will display on the screen
with blinking triangle and after few seconds the average of fluctuating weight will display on the screen and will lock on the display.
To release the locked weight on display simply press the
[HOLD] key again and the display will return to normal mode automatically.
[HOLD] key function can be activated before or after putting the
weight on the tread platform. But in case of weighing unstable person it is recommended to press [HOLD] key after the person moves on the tread platform.
19
USING BMI FEATURE
1. In normal mode, press BMI key to enter into BMI mode.
2. The display will show last height and the extreme left digit of figure will blink.
3. Enter the height using (0~9) number keys. (ex:170cm)
4. Proceed to weigh as normal. The instrument will show the weight, height, and BMI value.
5. At this stage, the weight and height can be changed, and the MS 3910 will automatically calculate the BMI value according to the changed weight and height.
6. HOLD feature can also be used if the weight is unstable.
7. Press BMI button to return normal weighing mode.
Body Mass Index Categories Classification of weight for adults over 18 years on the basis of Body
Mass Index according to WHO, 2000 EK IV and WHO 2004 (WHO ­World Health Organization).
20
USING TARE FUNCTION Tare allows the user to zero the instrument to deduct the weight of a
container/ clothes from the reading of the instrument, thus giving the true weight of the product/ person being tested.
Place the object need to be tare on the tread platform. Press “TARE” key after the weight stabilizes and stable sign
displays. The display will become zero.
Place the object need to be weighed (without removing the
tare object) To delete the saved tare value, remove the tare object from the tread platform and press “TARE” key.
21
USING INTERNAL SETTING Switch ON the scale and long press [ZERO] key for 3 seconds,
first “SETUP” and then “A.OFF” will display successively.
AUTO-OFF TIME SETUP This enables operator to select the auto turn OFF time of the device.
Auto off time: 120 sec/180 sec/240 sec/300 sec/off **Please Press ‘Tare’ key to confirm the setting. If you want to select
as auto off time, press ‘Tare’ key when displays on
the screen.
BUZZER ON/OFF SETUP
This enables operator to select beep sound ON/OFF.
Auto off time: 120 sec/180 sec/240 sec/300 sec/off
Buzzer: On/Off
** To apply the changes proceed till END displays on the screen, and then press HOLD to confirm changes.
240 S
240 S
22
STORING AND RE-CALLING PRE-TARE WEIGHT MS 3910 can store 5 sets of pre-tare values, and there are two methods to store pre-tare value – Using Dead Weight or Enter
weight using 0~9 keys:
A. Using Dead Weight:
DESCRIPTION EXAMPLE
Press M1-5 key after loading the weight on the platform; the display shows blinking m sign.
Press numeral key 1 ~ 5 to assign the position to pre-tare weight.
Press ENTER key to store pre­tare weight; the instrument will make beep sound.
B. Enter Weight Using 0~9 Keys:
DESCRIPTION EXAMPLE
Press PRE-TARE key, and the extreme left digit will blink.
23
Enter pre-tare weight using 0~9 keys. For example- 5kg.
Press ENTER key to confirm the pre-tare weight; the display will show the minus sign on left of the pre-tare weight value.
Press M1-5 key; the blinking m sign will appear on the display.
Press numeral key 1 ~ 5 to assign the position to pre-tare weight.
Press ENTER key to store pre­tare weight; the instrument will make beep sound.
24
C. RECALL PRE-TARE VALUE
DESCRIPTION EXAMPLE
Long press PRE-TARE key for 3 seconds; the display will show pre-tare value- m1 firstly. The pre-tare value will blink.
Press numeral key 1 ~ 5 to choose pre-tare value
Press ENTER key to confirm the pre-tare weight; the instrument will automatically perform tare function.
Press PRE-TARE key to return to Normal Mode.
25
GROSS WEIGHT 60.00kg TARE WEIGHT 30.00kg NET WEIGHT 30.00kg PATIENT HEIGHT 100.0cm PATIENT B.M.I 37. 5
29/12/2008 17:00
PRINTING FUNCTION
Weight, BMI and Height results can be printed for records using RS232 interface cable (included in accessory kit). After weighing and calculating BMI simply press PRINT key to print out the results. The format presented below is the standard format of results print­out and cannot be changed.
Parameters of RS232 interface
Set parameters of the scale interface on the connected device. It is not possible to change the scale parameters.
Baud rate: 9600 bps  Parity check: None  Data length: 8 bits  Stop bit: 1 bit  Handshake: RTS/CTS  Data code: ASCII
26
RS232 Pin out
Connecting with PC
1. Start Hyper Terminal Start Hyper Terminal program from clicking Start Menu Programs Accessories Communication
Hyper Terminal.
2. New Connection Description Give new connection a name then click OK.
3. Select Your COM Port Click Connect to select your COM port. Usually there’s only one
option for select. Then click OK.
4. Port Settings Click Bits per second to set up rate at 9600, Data bits at 8, Parity at
None, Stop bits at 1 and Flow control at Hardware. Then click OK to complete your setting.
5. Output Data When the patient has been weighed and BMI calculated, simply
press the PRINT key to output data from scale to PC or an Optional Printer.
27
CAUTION:
Always connect the AC adaptor with the indicator before
connecting to the mains power supply.
Please disconnect the adaptor from main power supply before
taking out the AC adaptor pin from indicator.
INSTRUCTION FOR CHARGING AND CONNECTING
If
prompt displays on the LCD, please charge the scale
with MS 3910 exclusive adaptor or replace the batteries.
Locate adaptor plug on the right side of indicator.
RS232 Connector
AC Jack Connector
28
INSTRUCTION FOR REPLACING BATTERIES
1.
2.
Open the battery housing cover
Take out the battery housing
29
3.
4.
Rechargeable Battery and AA-size Battery Housing
AA-size Battery Housing
Rechargeable Battery Housing
Make sure that all the batteries are properly placed
30
5.
6.
7.
Place the battery housing back and make sure that the housing pin is connecting to the right point
Place the housing back
Close the battery housing cover
31
MS 3910 ASSEMBLY
1. Assemble the column with the platform.
2. Tighten the four(4) screws of column through base.
32
SSEMBLING HEIGHT ROD HM201M/201D/210D ON WIDE COLUMN
A
Item Name Quantity 1 M5x0.8x11 round head screw 4 2 Relief Bushing 2 3 Bracket for HM-200D 2
* Photo of display for reference only. Please refer to the actual product.
Step 1. Assemble two brackets on the column
33
HM201M is mechanical height rod which has same assembly procedure with HM200D/201D.
Item Name Quantity 1 HM201M/HM201D/HM200D 1 2 M5x10L flat head screw 2 3 Fixing block 2 4 M5x0.8x11 4
* Photo of display for reference only. Please refer to the actual product.
Step 2. Fix HM200D/201D/201M on the bracket and tighten the screw.
34
ASSEMBLING TP21xx SERIES THERMO PRINTER ON SCALE
Item Parts Qty 1 M5*15L head screw 1 2 Screws for printer bracket 2 3 Printer bracket 1
35
ERROR MESSAGE
36
TROUBLESHOOTING
Troubleshooting for defective modes:
Original purchaser can enjoy the benefits under the effective Warranty against functional defects in material and workmanship subject to the terms and conditions listed in the yearly Warranty Program & Return Policy. Our warranty service program includes the following:
1. Technician repair service under warranty or at a service maintenance charge depending on the workmanship for the defective functionality or cause of damage covered by the warranty.
2. Parts replacement from the manufacturing factory under the warranty or at a certain cost for the replaced parts plus the workmanship charge if not covered under the warranty.
Before you contact our Authorized Dealer in your country for technician repair service, please read through the following section carefully:
Self-checking Tips:
Some functional defects can be identified and maintained by users as listed below:
1. Power-on failure
Check if the main power adaptor has not plugged onto the scale
properly
Check if the battery power is running low - Replace with new
batteries
2. Indicator showing “0000” ZERO SPAN out of range
Incorrect weighing result - Avoid damages by external environment force such as free-drop to the ground, collision by external objects, etc.
Proper re-calibration procedure required to correct the setting of weighing accuracy.
Interference due to RF disturbance, ground vibrationetc.
37
Unstable platform feet adjustments according to bubble level
indication
Incorrect position or other external objects within weighing area
The weighing-scale is not put in a solid & firm ground area, such as carpet floor or lawn.
3. Connection failure for data transmission to PC or printer
Wrong connection wires or faulty wires for transmission between
the digital indicator & load cells.
Wrong indicator models Wrong internal wiring or wire broken
In case of the following defective mode occurs, it is suggested to contact your nearest Authorized Dealer for further technician service & repair:
1. POWER switch-on failure :
Push-button faulty Short circuit wires - Wire broken Safety fuse burnt out Wire connection problem Main power adaptor faulty – Parts Replacement
2. LCD display faulty
Possible hardware defects include: Uneven brightness in the LCD
display screen & texts color blurred, smeared rainbow screen, incorrect decimal display
LCD PIN broken or short circuit PCB cooper foil broken & loosed welding Unable to save or read data – IC or transistor faulty, internal parts
broken.
LCD showing ERRL after switch on - Load cell damaged Overload may cause the weigh to malfunction.
Software system crash
Resonator faulty
Load cells with faulty grinding standard. Key buttons failure - Front key panel damaged or disconnected
3. Buzzer malfunction
38
Wrong welding of PVC wire Key buttons & control panel damaged or disconnected
39
40
Manufacturer’s Declaration of Conformity
This product has been manufactured in accordance with the harmonized European standards, following the provisions of the below stated directives:
93/42/EEC as amended by
2007/47/EC Medical Device Directive
2014/31/EU Non-
automatic Weighing Instruments
Directive
Please see separate document showing on sticker of device for above CE marking.
Authorized EU Representative:
Manufactured by:
Charder Electronic Co., Ltd.
No.103, Guozhong Rd., Dali Dist., Taichung City 412,Taiwan (R.O.C.)
FDA no.: D051882
IN-1178 [10141E] 05/2017