CGS Instruments SAM-1950, SAM-3000 Operating Instructions Manual

Tel.: +49 8466 / 94 15 - 0 Fax: +49 8466 / 94 15 - 49
sales@cgs-company.de www.cgs-company.de
CGS ANALYSENTECHNIK CGS PROZESSANALYTIK
Keltenstraße 3 D-85095 Denkendorf
SAM-3000 / SAM 1950
Operating instructions
Language: English Revision: Version 1.0 Issue: September 2015
NOTE
Read these operating instructions carefully before assembly, installation and implementation.
These operating instructions are also available as a digital document on:
www.cgs-company.de/downloads/MDZ_M_B_SAM-3000.pdf
MDZ_M_B_SAM-3000_1.0
42/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
Table of contents
Table of contents ....................................................................33
1 General information ................................................................35
1.1 Device version and validity .....................................................35
1.2 Information for our customers ..................................................35
1.3 General ....................................................................35
1.4 Warnings and explanatory texts .................................................35
1.5 Intended use ................................................................36
1.6 Qualified personnel ...........................................................36
1.7 Warranty information .........................................................36
1.8 Delivery information ..........................................................37
1.9 Norms and regulations ........................................................37
1.10 Disclaimer .................................................................38
2 Warnings and safety instructions ......................................................39
3 Packaging and transport ............................................................40
3.1 Packaging material ...........................................................40
3.2 Shipping labels ..............................................................40
3.3 Transport inspection ..........................................................41
3.4 Storage ....................................................................41
4 Installation, assembly information ....................................................42
5 Setup and function .................................................................43
5.1 Functional description .........................................................43
5.2 Design and operating mode ....................................................44
5.3 Dimensional drawing .........................................................45
5.4 Gas flow schematic ...........................................................46
5.5 Control PCB .................................................................47
5.5.1 Design and operating mode of the control PCB ...............................47
5.5.2 PROFIBUS DPV2 (Optional) ................................................51
5.6. Design and functions of solenoid valve manifold ...................................51
6 Technical data .....................................................................53
7 Maintenance & repair ...............................................................54
7.1 Replacing the solenoid valves / replacing the filter ..................................54
8 Device configuration (order code) .....................................................55
9 Spare parts .......................................................................56
10 Declaration of conformity ...........................................................57
11 Labelling-CD .....................................................................58
12 Cleaning ........................................................................59
12.1 Cleaning the surface .........................................................59
12.2 Cleaning the interior .........................................................59
12.3 Cleaning the control panel ....................................................59
43/79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
44/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
1 General information
1.1 Device version and validity
This documentation is restricted to the specified product and is only valid for this respective product. The documentation describes the installation, implementation and maintenance of the device and provides information about its function.
1.2 Information for our customers
Read this manual before commencing work on this device! It contains important information and data that, when observed, will ensure the device functions properly and save service costs. Implementation of this device is thereby considerably facilitated, leads to reliable measurement results and prevents damage caused by improper use of the product.
1.3 General
This device has left the factory in a technically and operationally safe condition. To maintain this condition and to ensure safe operation of this product, it must be implemented only in the manner specified by the manufacturer. In addition, correct and safe operation of this product depends upon proper transport, proper storage and installation as well as careful operation and maintenance. This manual contains the information necessary for the proper use of the product described herein. It is written for technically qualified personnel who are specially trained or have relevant knowledge in the field of automation technology (measurement, control and feedback control systems). Knowledge and technically correct implementation of the safety instructions and warnings contained in this manual are required for the safe installation and implementation of the product described and for its safety during operation and maintenance. Only qualified personnel have the necessary expertise to interpret the generally applicable safety instructions and warnings in specific cases and to apply them in practice. This manual is an integral part of the scope of delivery even if a separate order is possible due to logistical reasons. Because of the large number of technical details, it is not possible to consider all the details for all versions of the described product and for every conceivable installation, operation, maintenance and implementation in systems. If you need additional information, or should particular problems arise which are not sufficiently covered in this document, please request the necessary information from your local or responsible CGS representative.
1.4 Warnings and explanatory texts
This manual describes how you can properly run, commission, operate and maintain this device. You must pay particular attention to all warning and explanatory texts. These are separated from the text and specifically identified by the relevant icons. They provide you with valuable tips for avoiding operating errors and accidents.
45/79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
1.5 Intended use
The intended use for the purposes of this manual means that this product must only be used for the applications described in the catalogue and in the technical description and only in conjunction with external devices and components approved or recommended by CGS. The product described in this manual has been developed, manufactured, tested and documented in compliance with the applicable safety standards. No danger to personal health or property damage therefore normally ensues from compliance with the handling guidelines and safety instructions described for its project planning, assembly, intended use and maintenance.
1.6 Qualified personnel
Only those persons who are familiar with the installation, assembly, implementation and operation of the product are qualified.
These persons have the following qualifications:
•They are authorised, trained or instructed to operate and maintain equipment and systems in accordance
with established safety procedures for electrical circuits, high pressures and aggressive and hazardous media.
•For devices with explosion protection: They are authorised, trained or instructed in carrying out work on
electrical circuits for hazardous systems.
•They are trained or instructed in the maintenance and use of protective equipment in accordance with
established safety procedures.
1.7 Warranty information
The contents of this manual is neither part of any previous or existing agreement, commitment or a previous or existing legal relationship, nor should anyone modify these. All obligations by CGS arise from the respective sales contract, which also contains the complete and solely applicable warranty terms and conditions. These contractual warranty regulations are neither extended nor limited by the contents of the manual. The manufacturer accepts no liability for damage or malfunctions resulting from failure to observe the operating instructions. Apart from the operating instructions and regulations regarding accident prevention applicable in the country and place of use, you must also observe the recognised technical regulations for safe and proper work.
The warranty claim expires if the device is not installed, implemented, operated and maintained in accordance with regulations.
46/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
Proper maintenance means:
•Adherence to the maintenance plan (see Chapter: Maintenance)
•Maintenance by trained maintenance personnel
•Use of original CGS spare parts
Maintenance should be performed by the CGS service technician or, on the operator side, by CGS-trained service technicians.
The warranty is void in case of:
•improper use
•use of unauthorised equipment
•faulty connection and existing components that are not part of the scope of delivery and service
•failure to use original spare parts and accessories
•retrofitting that was not performed by CGS itself
•non-compliance of prescribed maintenance work
The contents reflect the state of the art at the time of publication. We reserve the right to make any technical changes due to further development.
1.8 Delivery information
The respective scope of delivery is listed in the shipping documents accompanying the delivery based on the valid contract. Please observe the corresponding information on the packaging material when opening the packaging. Check the delivery for completeness and intactness. In particular, you should, where applicable, compare the order number on the nameplates with the ordering data. If possible, please keep the packaging materials so that you can use it for future return deliveries (see Chapter: Packaging and transport).
1.9 Norms and regulations
The specification and production of this device were based on the harmonised European Norms whenever possible. If no harmonized European Norms have been applied, then the norms and regulations of the Federal Republic of Germany shall apply. When using this product outside the scope of these norms and regulations, you must comply with the norms and regulations valid in the operator's country.
47/ 79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
1.10 Disclaimer
All technical information, data and operating instructions contained in this manual correspond to the latest state of development at the creation of this manual and are based on our previous experience and knowledge to the best of our knowledge.
We reserve the right to make any technical changes in the course of further development of the device described in these operating instructions. No claims can be derived from the data, illustrations and descriptions in this manual.
We assume no liability for any damage and interruptions caused by operator error, non-observance of these operating instructions or improper repairs.
We expressly state that only original spare parts and accessories approved by us must be used. This also applies to the subassemblies of other manufacturers.
The installation or use of non-approved replacement parts and accessories and any unauthorised modifications and changes are not permitted for safety reasons and exclude liability by CGS for any resulting damages.
Claims for damages are excluded regardless of the legal grounds of such claims.
Translations are made in good faith. We cannot accept any liability for any translation errors not even when the translation was made by us or on our behalf. Only the original text is binding.
The text and illustrations do not necessarily reflect the scope of delivery or a possible spare parts order. The drawings and diagrams are not to scale.
48/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
2 Warnings and safety instructions
This manual contains important Warnings and safety instructions which you must observe to ensure your own personal safety and to avoid property damage. The instructions referring to your personal safe­ty are indicated by a warning symbol. Instructions referring only to property damage have no warning symbol. Depending on the hazard level, warnings are displayed in descending order as follows.
If more than one hazard level occurs then the warning representing the highest degree of danger will be used. If a warning triangle is used to warn against personal injury then the same warning can be added to warn against property damage.
DANGER
indicates that death or severe personal injury will result if no appropriate precautions are taken.
WARNING
indicates that death or severe personal injury can result if no appropriate precautions are taken.
CAUTION
with warning triangle indicates that minor personal injury can result if no appropriate precautions are taken.
NOTE
means that an undesirable event or situation can occur if the corresponding information is not observed.
CAUTION
without warning triangle indicates that property damage can result if no appropriate precautions are taken.
49/ 79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
3 Packaging and transport
3.1 Packaging material
The individual system components and modules are shipped in different packaging. Primary packaging materials: wood, cardboard and plastics (films, foams), but also including hoop iron (for bundles), for example. Packing materials include materials which are added to the packages for moisture protection (e.g. silica gel bags). If no agreement was reached on returning the packaging material then the packaging material remains with the customer. Waste disposal that is environmentally friendly and in accordance with the applicable disposal regulations must be ensured. If necessary, assign waste disposal companies with the disposal of the packaging material. The system components are to be transported as far as possible in the package to the installation site and kept in the package until assembly.
3.2 Shipping labels
Symbols are attached on the packages which must be observed during all transport phases including storage.
1. TOP! The package must be stored, handled and transported in such a way that the arrows always point upwards. Rolling, swinging, severe tipping or walking of goods marked in this way must be avoided.
2. Do not use forklift truck here
3. Do not destroy barrier!
4. Fragile, handle with care! Goods marked with this symbol should be handled carefully and should
never be rolled over or tied tightly.
5. Keep dry. Since goods of this type should also be protected from high levels of humidity, they must be transported in closed vehicles or containers.
6. Center of gravity. This symbol is intended to provide a clear indication of the position of the center of gravity.
50/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
3.3 Transport inspection
Important:
Check the entire shipment for completeness and possible transport damage immediately upon arrival. Failure to observe the following instructions in the event of damage shall cease the obligation of the insurer to render performance.
•Even in case of suspected damage: Only acknowledge receipt with reservation (e.g. including shipping
documents) by specifying the alleged damage.
•Assert claims for compensation against third parties:
Carriers, freight forwarders, warehouse keepers:
- demand joint damage inspection
- request a certificate of damage
- hold these liable for damages in writing and describe the damage in detail, namely:
- in case of visible damage prior to accepting delivery of the goods
- in case of no apparent outward damage immediately upon discovery
•Set deadlines for complaints and observe these!
•Ensure mitigation of incurred damage and avoidance of further damage.
•Immediately consult an insurance adjuster to record the damage and provide advice for securing claims
against third parties and for measures to mitigate the damage incurred.
•Never change the status of the consignment and its packaging until the arrival of the adjuster insofar
this is not required to mitigate and prevent further damage.
•Immediately notify the insurer about the insurance case and quickly submit the complete claim
documents to accelerate claims settlement (latest time before any preclusion and/or limitation periods for claims against third parties).
3.4 Storage
After unloading, the packages must be stored until installation in accordance with the shipping markings. Packed machine parts and accessories may not be unpacked.
Observe the applicable regulations for storage:
•Store in a cool dry place. Maximum humidity: 60 %.
Make sure the packages are not stored outdoors. Also ensure that the storage room floor is dry during storage.
•Protect against direct sunlight. Storage temperature 5° C to 35 °C.
•Store in a dust-free location!
•Avoid mechanical shock and damage.
•For longer storage over 3 months, check the conservation measures.
In case of aggressive weather conditions, conservation must be renewed if necessary.
51/79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
4 Installation, assembly information
Safety measures:
Installation and assembly must only be performed by specially trained and qualified personnel. The general accident prevention regulations must thereby be observed. A formal hand over to the operator occurs in c ase of implementation by CGS o r its authorised rep resentatives. The device can subsequently be properly operated according to the operating instructions and applicable safety regulations. Before opening the device, disconnect it from the power supply in any case.
Requirements for the installation site:
•No exposure to direct sunlight
•Low exposure to dust
•Sufficient space for operation and maintenance
•Available and easily accessible escape routes
•No tripping hazards caused by cables that run close to the ground, etc.
•Maximum height: 2000 m above sea level
Technical prerequisites (supply voltage, gas pressures, etc.) can be found in the Technical Data section.
52/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
5 Setup and function
5.1 Functional description
The cabinet sample module series SAM-1000_DP is used to connect different sample points to analysers. Because of its modular design, it can be used for various applications in the chemical and petrochemical industries for gaseous high-purity gases.
Special features: The SAM-3000 especially distinguishes itself through the following features:
• Compactconstruction
• Minordeadvolume
• Modularconguration
• Highleakthightness
• Multi-purposeapplicationspossible
• PROFIBUSconnectivitypossible(optional)
Depending on the expansion stage of SAM-3000 can be switched up to 4 sample gases and 2 calibration gases to an analyzer. For monitoring the gas ways at a glance LED’s shows the selected way. The LED’s can be optionally arranged as Pushbuttons. Filter frits at the inlet of the switching module protects the following analysers. So that impurities are filtered out of the sample gases and avoids pollutions and damages at the analyzer. Filter frits are available in 20 μm and 50 μm fineness of filtration. With the solenoid valve manifold concerning purging or no purging the dead volumes are minimized. For high pu­rity applications the solenoid valves on the manifold are optimally arranged. By activation of a solenoid valve a potential free contact can be used to send a feedback signal to the DCS. Up to 4 flow indicators can be used to monitor the flow rate of sample gases. The SAM-3000 optional includes an integrated pressure transmitter. The estimation of additional attended connectors and tubing connections can be much reduced and supports a fast installation. A high variety of connectors can be chosen for different tubing diameters. Depending on application different materials for the tubing inside of SAM-3000 can be selected. Free configurable bridge circuits on control PCB compliance individual requirements. So for e.g. the LED’s displays the individual task of function. Optionally the SAM-3000 can be connected to a Profibus.
53/79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
5.2 Design and operating mode
The SAM-3000 is designed as a plug-in module for 19“ racks or 19“ cabinets.
• Aluminiumhousingwithmountedfrontplate
• ControlPCB
• Solenoidvalvemanifold
• Outletmanifold
• Dependingontheexpansionstagebulkheadunions,owmetersandowregulators
Bypass-Outlet
Outlet to analyzer
Pressure transmitter
Solenoid valve manifold
Housing with front plate
Control PCB
Flow controller
Quick connector
Flow low indicator
Flow meters with needle valve
In the main case the inner of the module includes a solenoid valve manifold with selectable tubings in different materials (metal, plastic or mixture of both) available. To control the solenoid valves and to evaluate the signals is a control PCB integrated. The top of the LED’s the Pushbuttons are implemented in the front plate of the SAM-3000. Depending on expansion stage there are bulkhead unions, flow meters and flow regulators monted at the front plate.
54/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
5.3 Dimensional drawing
The following drawing shows the necessary space required:
25-250l/h
bypass AP4711
bypass AP4712
25-250l/h
25-250l/h
bypass AP4713
bypass AP4714
25-250l/h
6-60l/h
sample
gas
SAM-3000
Manual
Sampling
Sample
Flow
Power/LT
Flow 1
Flow 2
Flow 4
Flow 3
Service
Zero Gas
Sample 1
Span Gas
Sample 2
Sample 4
Sample 3
AP2615
AP4711
AP4712
AP4714
AP4713
Zero Gas
Span Gas
Service
Flow 3
Flow 4
Flow 1
Flow 2
55/7 9
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
5.4 Gas flow schematic
A exemplary gas diagram is shown below:
25-250 l/h
25-250 l/h
6-60 l/h
25-250 l/h
SAM-3000
sample gas 1
25-250 l/h
to analyser
ATM
example
sample gas 2
sample gas 3
sample gas 4
zero gas
span gas
56/7 9
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
5.5 Control PCB
5.5.1 Design and operating mode of the control PCB with LED v.1.0
The dimensions of control PCB is developed for using in the 19“plug-in device. Via guiding rails the con­trol PCB gets through the module. On control PCB mounted diodes and Pushbuttons for inspection and service are implemented in the front plate. The connection plugs are located at the rear side of control PCB. Power supply plug, the connection to analyzer and the plug for connection to the DCS. The control PCB is supplied by 24 V DC and protected by a fuse. For a failure-free operation the voltage is stabilized onboard.
1
2
3
4
1
3
2
4
Pushbuttons
Green LEDs to display sample gases
Yellow LEDs to display service, zero and span gas
Red LEDs for FLOW- Sample gas/ or. costomized display variants
Used to test the lights on the front panel
57/7 9
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
5.5.1 Design and operating mode of the control PCB with button v.2.0
1
2
3
4
5
6
7
Pushbuttons
Used to test the lights on the front panel
Pushbuttons for different sample gases
Switches the sample gases
LED for the service mode
Pushbutton for zero gas
Pushbutton for span gas
LED for FLOW- Sample gas/ or. costomized display variants
Pushbutton for Reset- Sample gas/ span gas
1
2
1
1
1
3
1
4
1
5
1
6
1
7
58/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
Micro-fuse
Power supply 24 VDC
Connection to analyzer SUB-D 9-pin
Connection to DCS SUB-D 37-pin
1
3
2
4
1
3
2
4
5.5.3 Design of both Control PCBs (back view)
59/ 79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
Description of display- and operating controls:
Power/LT (green): LED is shining permanently, if the module is connected to the power supply. By pressing the button the function of LEDs can be checked (all LEDs are shining).
Sample 1 (green): LED is shining permanently, if sample gas 1 is switched to the analyzer.
Sample 2 (green): LED is shining permanently, if sample gas 2 is switched to the analyzer.
Sample 3 (green): LED is shining permanently, if sample gas 3 is switched to the analyzer.
Sample 4 (green): LED is shining permanently, if sample gas 4 is switched to the analyzer.
Service (yellow): LED is shining permanently, if DCS releases manual service handling.
Zero Gas (yellow): LED is shining permanently, if zero gas is switched to the analyzer.
Span Gas (yellow): LED is shining permanently, if span gas is switched to the analyzer.
Flow 1 (red): LED is shining by individual customer requirements. The LED can shine for instance if the gas flow of sample gas 1 is in tolerance range or outside of tolerance range.
Flow 2 (red): See above „Flow 1“.
Flow 3 (red): See above „Flow 1“.
Flow 4 (red): See above „Flow 1“.
60/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
Plug connections:
Power supply (2-pin plug):
Pin Function Description
1 24 V DC voltage + 24 VDC 2 GND-connection GND
SUB-D 9-pin plug for connection to analyzer:
The left side „functions of device“ of diagram shows the schematic design of the control PCB. The right side of diagram shows the connections to peripheral devices.
5
9
4
8
6
2
7
3
1
binary input sample gas 1
binary input zero gas
analog output pressure transmitter +
analog output pressure transmitter -
binary input span gas
binary output service release
output GND
output 24VDC
binary input flow
functions of device
PIN-
assign
P
I
61/7 9
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
Connection to DCS via a SUB-D 37-pin plug:
The left side „functions of device“ of diagram shows the schematic design of the control PCB. The right side of diagram shows the connections to peripheral devices.
5
23
4
22
20
2
21
3
1
binary input sample gas 1
binary input zero gas
contact sample gas 2.1
contact sample gas 2.2
binary input span gas
binary input service release
binary input flow 1
functions of device
PIN-
assign
7
8
9
6
10
11
12
13
14
15
16
17
19
18
24
25
26
27
28
29
30
31
32
37
36
35
34
33
GND 24VDC
binary input sample gas 2
binary input sample gas 4
binary input sample gas 3
binary input flow 2
binary input flow 4
binary input flow 3
nc
nc
contact sample gas 1.2
contact sample gas 1.1
nc
nc
nc
nc
contact sample gas 3.1
contact sample gas 3.2
contact sample gas 4.2
contact sample gas 4.1
contact span gas 1.1
contact span gas 1.2
contact zero gas 1.2
contact zero gas 1.1
contact flow 1.1
contact flow 1.2
contact flow 2.2
contact flow 2.1
contact flow 4.1
contact flow 4.2
contact flow 3.2
contact flow 3.1
62/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
5.6. Design and functions of solenoid valve manifold
5.6.1 MVB-S
Solenoid valves screwed in manifold
Solenoid valve manifold (to analyzer)
Sample gas inlets
The standard solenoid valve block of 1.4305 is used for stream switching up to 10 different streams. It has rapid switching solenoid valves, which have a very long life time – tested over 500 million switch­ing operations. Due to the small design the solenoid valve block can be installed in an 19“-rack directly behind the analyzer. The connections for the individual sample gases and the calibration gases are made by flange adapt­ers, which have integrated sinter metal filters. In addition the sample gas inlet can be equipped with a continuously bypass to minimize the total response time. The valve block has a common tubule which is connected to the analyzer. The other side of the tubule has a blind flange. Alternatively at low con­centrations this flange will be changed by a capillary for a very low split flow.
63/79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
5.6.1 MVB-K
Solenoid valves screwed in manifold
Solenoid valve manifold (to analyzer)
Sample gas inlets
The standard solenoid valve block compact is a further development of the standard block. Due to the new design the solenoid valves are mounted on opposite sides of the block. This reduces the length of the block by half. The solenoid valve block can now be equipped with up to 20 gas inlets. New is also the continuously purging of the solenoid valves. These rapid switching solenoid valves have a very long life time – tested over 500 million switching operations. The connections for the individual sample gases and the calibration gases are made by flange adapters, which have integrated sinter metal fil­ters. In addition the sample gas inlet can be equipped with a continuously bypass to minimize the total response time. The valve block has a common tubule which is connected to the analyzer. The other side of the tubule has a blind flange. Alternatively at low concentrations this flange will be changed by a capillary for a very low split flow.
64/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
1. Analyser without control function
PCS
SAM-3000
4-20 mA
Measuring/Calibrating
Test gas valves
Test gas valves
Maintenance
Measuring gas valves
Measuring gas valves
Fault
Measuring gas fl ow alarm
Service release
Measuring range identifi cation
Bypass fl ow alarm
Analyser
analogue
2. Analyser with Autocal function
Measuring/Calibrating
PCS
SAM-3000
Measuring gas valves
Test gas valve
Zero gas valve
Test gas valves
Measuring gas valves
Measuring gas fl ow alarm
Service release
Bypass fl ow alarm
Analyser
analogue
Fault
Calibration
Measuring range identifi cation
4-20 mA
Maintenance
3. Analyser with Profi bus interface
Selection measuring point
PCS
SAM-3000
Analyser Profi bus
Maintenance messages
Measuring/Calibrating
Calibration
Test gas valve
Measuring point valve
Zero gas valve
Measuring gas fl ow alarm
Bypass fl ow alarm
Measuring point
Measured value
Fault messages
Measured value
Feedback
Controls
6 Connection options
65/79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
7 Technical data
Inlets Quantity
Inlet filter fineness Diameter (outer)
1 ... 6 without, 20 μm or 50 μm 2, 3, 6 mm or 1/8“, 1/4“
Solenoid valves Standard
Option 1 Option 2
2/2-ways normally closed 2/2- ways normally open
3/2-ways Bypass flow meters
Quantity Flow rate
1 ... 4
variabel 5-1200 l/h Flow alarm Quantity
Flow rate
1 ... 4
variabel 5 - 1200 l/h Flow limitation
Standard Option
Needle valve
Flow Controller 0 .. 96 l/h Pressure transmitter
Measuring range (Standard) Output signal (Standard)
0 ... 1 bar
4 ... 20 mA Materials All-wetted parts
Hounsing
Stainless steel 1.4571 PTFE
Aluminium, anodised Interfaces DCS
Optional Analyzer
bit parallel SUB-D37-pin
PROFIBUS DPV2
SUB-D9-pin Dimensions SAM-3000
SAM-1950
483 x 4HU x 270 mm
241,5 x 5HU x 270 mm Weight By complete expansion stage circa. 7 kg Electrical connections
Power supply Fuse
24 V DC
M 1,6 A
66/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
8 Maintenance & repair
8.1 Replacing the solenoid valves / replacing the filter
• To replace the solenoid valve, unscrew it using a hook spanner (Item No.: 1017105)
• O-ring (Item No.: 1026311) and the filter (Item No.: 1026309 (50μ)) are located on the bottom of the
connection
• To change the filter frits, loosen the Allen screws and remove the connection
• Remove the O-ring and take out the filter
• Before inserting the new filter and O-ring, clean the contact surfaces of the O-ring so that it can
properly seal.
Maintenance or repair must only be performed by CGS service technicians. This can also be done by CGS-instructed / trained technicians. The warranty and liability expires in case of unauthorised maintenance or repair work. The applicable safety and accident prevention regulations must be observed in any case.
Ill. similar
67/ 79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
9 Cleaning
9.1 Cleaning the surface
The front panels are washable. We recommend using a sponge or cloth soaked in a mild detergent.
9.2 Cleaning the interior
After opening the device, you can, if necessary, carefully blow out the interior with a compressed air gun.
9.3 Cleaning the control panel
Clean with warm or hot water (if necessary, add a slightly acidic, neutral or slightly alkaline cleaning agent). Do not use abrasive cleaners or a high-pressure cleaner. Only limited resistance to organic solvents (e.g. alcohol, benzine, etc.).
CAUTION
Make absolutely sure that no water enters the device while cleaning. Before cleaning the device, disconnect it from the power mains!
68/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
10 Device configuration (order code)
- - - -
Cabinet switching module type
SAM-3000
SAM-1950
Gas inlets
Quanty of inlets
1
Quanty of inlets
2
Quanty of inlets
3
Quanty of inlets
4
Quanty of inlets
5
Quanty of inlets
6
Calibraon gas
Quanty
0
Quanty
1
Quanty
2
Inlet filter
Without inlet filter
0
Input filter 20µm
2
Input filter 50µm
5
Solenoid valve design
Solenoid valve STD
S
Solenoid valve for moisture
analysis
H
Solenoid valve (NC) for oxygen
applicaon
O
Solenoid valve (NC) for moisture
analysis and oxygen applicaon
Y
Manual measuring input
Without manual measuring inlet
0
manual measuring inlet
1
Bypass ow meter
Quanty of bypass ow meter
0
Quanty of bypass ow meter
1
Quanty of bypass ow meter
2
Quanty of bypass ow meter
3
Quanty of bypass ow meter
4
measuring range bypass ow
meter
15 - 150 l/h
A
25 - 250 l/h
B
Others
C
Flow and pressure control
without
A
Flow controller 18l/h
B
Flow controller 96l/h
C
Flow controller made of SS 96l/h
D
Needle valve
E
Pressure reducer
F
Back pressure regulator with ow
meter
H
69/ 79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
Flow meter for sample gas
Without ow meter
0
Flow meter before analyzer
1
Flow meter with ring iniator
before analyzer
2
Flow meter aer analyzer
3
Flow meter with ring iniator
aer analyzer
4
Measuring range sampl gas ow
meter
0,5 - 5 l/h
A
1,6 - 16 l/h
B
6 - 60 l/h
C
10 - 100 l/h
D
Others
X
Pressure sensor
Without pressure sensor
0
pressure sensor incl. threaded
connecon
1
Capillary
Without capillary
0
Kapillare
1
Capillary
Pipe design sample gas input
Diameter for process connecon
2mm
A
Diameter for process connecon
3mm
B
Diameter for process connecon
6mm
C
Diameter for process connecon
1/8"
E
Diameter for process connecon
1/4"
F
pipe calibraon gas input
Diameter for process connecon
2 mm
A
Diameter for process connecons
3 mm
B
Diameter for process connecon
6 mm
C
Diameter for process
connecon1/8"
E
Diameter for process connecon
1/4"
F
Without calibraon gas pipe
W
Pipe design bypass
Pipe design with plasc 6mm
1
Without bypass
3
Valve block
MVB-S
A
MVB-SP
B
MVB-GC
C
MVB-K
D
MVB-KSP
E
Pilot valves for membran valves
F
Special version
X
LED push buon
LED indicator
1
LED push buon
2
- - - -
70/7 9
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
11 Delivery contents
Immediately after receipt, check whether the shipment has arrived complete and without damage.
•SAM-1000_ONE / SAM-3000 [1x]
•Documentation [1x]
•Sub-D plug (9-pole) [1x]
•Sub-D plug (37-pole) [1x]
•Power plug (2-pole) [1x]
•PVC-hose (for suction) 6mm [1 meter]
•Stainless steel nut + ferrule set [according to configuration]
•Phillips-head screw (for rack mounting) M6x16 [4x]
•Captive nuts (for rack mounting) M6 [4x]
71/79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
12 Spare parts
Pos
*PU Article
Order No.
1 1 pcs. Control and display board 1017998 (V1.0) 2 1 pcs. Control and display board with button 1018004 (V2.0) 3 1 pcs. Flow controller; 18 l/h; 1/8“; Alu-body; SS-diaphragm 1011221 4 1 pcs. Flow controller; 96 l/h; 1/8“; Alu-body; SS-diaphragm 1011223 5 1 pcs. Flow controller; 96l/h; 1/8“; SS-body; SS-diaphragm 1011224
6 1 pcs.
Pressure Regulator with pressure gauge; Inlet: 20 bar; Outlet: 0.5 - 3 bar; NPT 1/4“ female; Chrome-plated brass; with Panel-Mounting­Ring for SL 20-P
1021835
7 1 pcs.
Flowmeter without valve; 10...100 Nl/h; without limit switch; nickel plated brass; 1/4“ NPT; Air
1028362
8 1 pcs. Ring sensor; inductive; NAMUR bistable; 10 mm 1014583
9 1 pcs.
Pressure transmitter S-10; 0-25 bar; for general applications; stan­dard version
1019509
10 1 pcs.
Solenoid valve; 2-ways; 24 VDC; D= 1.25mm; FKM; with connection cable 700mm
1027695
11 1 pcs.
Solenoid valve; 3/2-ways; 24 VDC; D=1.25mm; with connection cable 300mm
1019762
12 1 pcs.
Solenoid valve; O2; 2-way; 24 VDC; FKM; with connection
cable 700 mm
1030682
13 1 pcs.
Solenoid valve; O2; 2-way; 24 VDC; FKM; with connection
cable 700 mm
1030686
14 1 pcs.
Solenoid valve; O2; 3/2-ways; 24 VDC; with connection
cable 300 mm
1030684
15 1 pcs.
Solenoid valve; H2O; O2; 3/2-way; 24 VDC; with connection
cable 300 mm
1030688
16 1 pcs.
Solenoid valve; H2O; 2-ways; 24 VDC; D=1.25mm; with connection cable 700mm
1024306
17 1 pcs.
Solenoid valve; 3/2-ways; 24 VDC; D=1.25mm; with connection cable 300mm
1024308
18 1 pcs.
SS Integral Bonnet Needle Valve;Cv = 0.09; 3 mm Swagelok Tube Fitting; PCTFE Stem Tip
1021802
19 1 pcs.
Flowmeter with needle valve; 25...250 Nl/h; without limit switch; nickel plated brass; 1/4“ NPT; Air
1027287
20 1 pcs.
Pressure regulator GO; back pressure regulator; outlet: 3.5 bar; 50 PSI; 3 x 1/4“ NPT
1019170
21 1 pcs.
Flowmeter with needle valve; 15...150 Nl/h; without limit switch; nickel plated brass; 1/4“ NPT; Luft
1021111
22 1 pcs. O-ring set (10 pcs); FKM 6.00x2.00mm 1026311 23 1 pcs. Sintered metal frit set (10 pcs); D=6.35mm; pore size 50μ 1026309 24 1 pcs. Quick coupling plug; RBE 03; internal thread M8x1 1015026 25 1 pcs. Quick coupling plug; RBE 03; internal thread G 1/8 1015027 26 1 pcs. Tube fitting set 1/16“ SS 1027699
72/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
Pos
*PU Article
Order No.
27 1 pcs. Tube fitting set 2 mm SS 1027700 28 1 pcs. Tube fitting set 1/8“ SS 1027701 29 1 pcs. Tube fitting set 3 mm SS 1027702 30 1 pcs. Tube fitting set 1/4“ SS 1027703 31 1 pcs. Tube fitting set 6 mm SS 1027704 32 1 pcs. Cap nut set 4.3 mm hose 1027705 33 1 pcs. Cap nut set 6 mm hose 1027706 34 1 pcs. Phillips-head screw set M6x16 with captive nuts 1027707 35 1 pcs. PTFE-hose; do=12.0 mm di=10.0 mm; nature 1015713 36 1 pcs. Hook spanner for screw in solenoid valve; DIN 1810 1017105
37 1 pcs.
Pressure Regulator GO; Back Pressure Regulator; SBPR Series; Control Range 0 - 2.07 bar; 1/4“ NPT; stainless steel / Teflon
1023812
*PU = Packing Unit
- ferrule set = union nut , front clamping ring, back clamping ring
73/ 79
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
13 Appendix A (EG Declaration of conformity)
Tel.: +49 8466 / 9415 - 0 Fax: +49 8466 / 9415 - 49
sales@cgs-company.de www.cgs-company.de
CGS ANALYSENTECHNIK CGS PROZESSANALYTIK
Keltenstraße 3 D-85095 Denkendorf
EG-Konformitätserklärung
EC Declaration of Confi rmity
Hersteller:
Manufacturer: CGS Analysen-, Mess- und Regeltechnik GmbH
Keltenstraße 3
D-85095 Denkendorf
Hiermit erklären wir, dass unser Produkt, Typ:
Hereby, we declare that the product, type:
SAM-3000
folgenden einschlägigen Bestimmungen entspricht:
is in confi rmity with the following applicable standards:
20 12/19/E U WEEE-Richtlinie
WEEE-Directive
2014/30/EU EMV-Richtlinie
EMC-Directive
2014/68/ EU Druckgeräterichtlinie
Pressure Equipment Directive
Angewendete harmonisierte Normen, insbesondere:
Applied harmonised standards, especially:
DI N EN 61326 -1:2013- 07 (VDE 0843-20-1:2013-07)
Elektrische Mess-, Steuer-, Regel- und Laborgeräte EMV-Anforderungen - Teil 1: Allgemeine Anforderungen
Electrical equipment for measurement, control and laboratory use EMC requirements - Part 1: General requirements
DIN EN 61010-1 ( VD E 0 411-1:20 11- 07 )
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel­und Laborgeräte - Teil 1: Allgemeine Anforderungen“
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements
DIN EN 61010-2-201:2 014-01 (V DE 0 411-2-201:2014 -01)
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel­und Laborgeräte - Teil 2-201: Besondere Anforderungen für Steuer- und Regelgeräte
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-201: Particular requirements for control equipment
Jahreszahl der CE-Kennzeichenvergabe: 15
Denkendorf, den 11.09.2015
Alfons Geyer
Geschäftsführer / Managing director
74/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
Tel.: +49 8466 / 9415 - 0 Fax: +49 8466 / 9415 - 49
sales@cgs-company.de www.cgs-company.de
CGS ANALYSENTECHNIK CGS PROZESSANALYTIK
Keltenstraße 3 D-85095 Denkendorf
EG-Konformitätserklärung
EC Declaration of Confi rmity
Hersteller:
Manufacturer: CGS Analysen-, Mess- und Regeltechnik GmbH
Keltenstraße 3
D-85095 Denkendorf
Hiermit erklären wir, dass unser Produkt, Typ:
Hereby, we declare that the product, type:
SAM-1950
folgenden einschlägigen Bestimmungen entspricht:
is in confi rmity with the following applicable standards:
20 12/19/E U WEEE-Richtlinie
WEEE-Directive
2014/30/EU EMV-Richtlinie
EMC-Directive
2014/68/ EU Druckgeräterichtlinie
Pressure Equipment Directive
Angewendete harmonisierte Normen, insbesondere:
Applied harmonised standards, especially:
DI N EN 61326 -1:2013- 07 (VDE 0843-20-1:2013-07)
Elektrische Mess-, Steuer-, Regel- und Laborgeräte EMV-Anforderungen - Teil 1: Allgemeine Anforderungen
Electrical equipment for measurement, control and laboratory use EMC requirements - Part 1: General requirements
DIN EN 61010-1 ( VD E 0 411-1:20 11- 07 )
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel­und Laborgeräte - Teil 1: Allgemeine Anforderungen“
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements
DIN EN 61010-2-201:2 014-01 (V DE 0 411-2-201:2014 -01)
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel­und Laborgeräte - Teil 2-201: Besondere Anforderungen für Steuer- und Regelgeräte
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-201: Particular requirements for control equipment
Jahreszahl der CE-Kennzeichenvergabe: 15
Denkendorf, den 11.09.2015
Alfons Geyer
Geschäftsführer / Managing director
75/7 9
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
Qualitätszeugnis
Hiermit bestägen wir, dass unsere Produkte nach aktuellem Stand der Technik produziert und
geprü wurden.
Die verwendeten Materialien sind nach den Einsatzspezifikaonen ausgesucht und nach
handwerklichen Richtlinien verarbeitet.
Hierbei kommen maßgebliche naonale und internaonale Normen und Standards zur Anwendung.
Das ferge Produkt wurde nach technischen Gesichtspunkten in einem Qualitäts-
sicherungssystem geprü und kontro
lliert.
Quality cerficate
Hereby we conrm that our products were produced and examined aer
current state of the art.
The used materials were selected aer
guidelines relang to cras and processed according to the employment
specificaons.
Here relevant naonal and internaonal norms and
standard are used.
The nished product was examined and controlled in quality insurance system on
the basis of technical criteria.
Denke
ndorf, 29.09.2015
Alfons Geyer
14 Appendix B (Inspection protocols)
76/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
Prüescheinigung für Gaskomponenten in Sauerstoff
Inspecon document for gas components in oxygen
Hersteller:
manufacturer:
CGS Analysen-, Mess- und
Regeltechnik GmbH Keltenstraße 3 D-85095 Denkendorf
Denkendorf, 29.09.2015
Alfons Geyer
Verfahren:
method:
Nr.
no.
Reinigungsmethode:
cleaning method:
Ultraschallbad
ultrasonic bath
1
Aceton
acetone
2
Spülen der Gaswege:
purge of gas pathes:
Scksto
nitrogen
3
Temperatur:
temperature:
± 20°C
4
Trocknung:
drying:
Sckstospülung
nitrogen purging
5
Drucktest:
pressure test:
3 bar (a) / 5 min.
6
1 bar (a) / 10 min.
7
Produktreihe:
product series:
SAM-3000
Gaskomponente:
gas compontent:
Sauersto
oxygen
Anwendung:
applicaon:
Gasanalyse / Sauersto
Öl- und Ferei
gas analysis / oxygen
free of oil and grease
Sauerstoffführende Komponenten:
oxygen transmied components:
Bauteile:
components:
Verfahren:
method:
Edelstahlrohr
stainless steel tube
2, 3, 4, 6
Magnetvenlblock
solenoid valve block
1, 3, 4, 5, 6
Druckregler
pressure regulator
3, 4, 5, 6/7
Manometer
gauge
3, 4, 5, 6/7
Flow Controller
ow controller
3, 4, 5, 6
Durchussmesser
ow meter
3, 4, 5, 6
Nadelvenl
needle valve
3, 4, 5, 6
Absperrvenl
ball valve
3, 4, 5, 7
Umschalthahn
manifold valve
3, 4, 5, 7
Reguliervenl
control valve
3, 4, 5, 6
Rückschlagvenl
check valve
1, 3, 4, 5, 6/7
Überströmvenl
relief valve
3, 4, 5, 6/7
Verschraubungen
fings
1, 3, 4, 5, 6
Filterfrien für Magnetvenlblock
filter frit for solenoid valve block
3, 4, 5
7 7/7 9
E
Product: SAM-3000 / SAM-1950
Revision 1.0– Issue September 2015
Generelle Konformitätserklärung für Sauerstoanwendungen
Für Komponenten und Systeme, die für Produkon, Lagerung und Transport von gasförmigem Sauersto verwendet werden und mit ihm in Berührung kommen, ist eine extrem saubere Oberäche erforderlich.
CGS Analysen-, Mess- und Regeltechnik GmbH bestägt, dass die gasführenden Bauteile unserer Analysenanlagen für Sauerstoffanwendungen gemäß den aktuellen Vorschrien gebaut werden. Somit sind die We
rkstoe für Druckregler, Rohr, Manometer… speziell gereinigt und mit Inertgas
gespült. Materialien sind frei von
- Fe, Öl und anderem organischen Material (Kohlenwasserstoverbindungen)
- Wasser
- Lacken, Anstrichsystemen und anderen Beschichtungen
- losem Rost, Walzhaut, Füllmaterial, Schweißverfärbungen und Schweißspritzern.
General conrmaon cerficaon for oxygen applicaons
Components and systems used for producon, storage and transportaon of gaseous oxygen and come into contact with, an extremely clean surface is required.
CGS Analysen-, Mess- und Regeltechnik GmbH conrms that gas-bearing components of your analyzer systems for oxygen applicaons are build according to the current standard regulaons. Therefore the materials of pressure regulators, tubes, gauges… are special cleaned and inert gas­purged. Materials are clear of
- grease, oil and othe
r organic materials (hydrocarbon)
- water
- laquer, painng systems and other coangs
- loose rust, mill scale, filling material, sweat stains and spaers.
Denkendorf, 29.09.2015
Alfons Geyer
78/79
E
Product: SAM-3000 /SAM-1950
Revision 1.0 – Issue September 2015
Tel.: +49 8466 / 94 15 - 0 Fax: +49 8466 / 94 15 - 49
sales@cgs-company.de www.cgs-company.de
CGS ANALYSENTECHNIK CGS PROZESSANALYTIK
Keltenstraße 3 D-85095 Denkendorf