QPS400-3 Technical Data
INSTRUMENTS
++
/
Quadrupole Power Supply Unit 3×400 V
QPS400-3
Version 1.00
Page 3
QPS400-3 Technical Data
INSTRUMENTS
++
/
Technical Data
Characteristics
Precise high-voltage amplifier for electrodynamical quadrupole traps
for micro- and nanoparticles
DC accuracy
Extremely low temperature coefficient of amplification
High bandwidth
Low noise
Temporally short-circuit proof
19" case
Input
Input voltage: ±10 V max.
Input impedance: 47 k
Signal inputs: 50 BNC plug sockets
Output
Amplification:
AC input: 40 DC inputs: 4
Function:
V
Top
= –40VAC – 4V
DC
V
Mid
= +40VAC – 4V
M
V
Bot
= –40VAC + 4V
DC
Output voltage: ±400 V min. (> 800 V
pp
)
Output current: ±50 mA max.
Output power: 10 W / channel max.
Bandwidth:
AC input (±10%): 3.5 Hz ... > 1 MHz DC inputs (-3dB): DC ... 100 kHz
Power bandwidth: > 100 kHz (200 kHz typ.)
Output offset voltage: < 100 µV
Drift of the output voltage: < 1 µV/K
Noise (DC-10 MHz): 3 mV
eff
(typ.)
Precision of the amplification at low frequencies:
< 200 ppm (100 ppm typ.)
Page 4
QPS400-3 Technical Data
INSTRUMENTS
++
/
Temperature coefficient of the amplification at low frequencies:
< 10 ppm/K (5 ppm/K typ.)
Signal outputs: 93 BNC plug sockets
Output short circuit duration: 1 s max.
Monitor Output
Attenuation: 1:40 of the output voltage
Output voltage: ±12 V min.
Output current: ±20 mA max.
Output offset voltage: < 100 µV
Drift of the output voltage: < 25 µV/K (5 µV/K typ.)
Precision of the attenuation at low frequencies:
< 0.2% (0.1% typ.)
Temperature coefficient of the attenuation at low frequencies:
< 50 ppm/K (25 ppm/K typ.)
Bandwidth (±10%): DC ... > 1 MHz
Signal outputs: 50 BNC plug sockets
Output short circuit duration: indefinite
Power Supply
Rated voltage: 230 V or 115 V ±10% (voltage selector switch)
Rated frequency: 50/60 Hz
Fuse: T 0.63 A at 230 V, T 1.25 A at 115 V
Power consumption: about 70 VA without load, 130 VA max.
Main connection: IEC inlet
General
Dimensions:
2 HU 19" plug-in box for mounting in a 19" cabinet
(height: 88 mm, width: 483 mm)
depth with heat sink without handles: 265 mm
Weight: 6.8 kg
Page 5
QPS400-3 Technical Data
INSTRUMENTS
++
/
Shipment Content
Quadrupole power supply unit QPS400-3 in 19" case
Line cord (universal IEC mains lead, length 1.8 m)
Output cables
(3 pcs. 93 coaxial cable with 93 BNC jacks, length 1.2 m each)
Ground wire
(flexible wire 1 mm
2
with 4 mm cable eyes, length 2 m)
Material for mounting in a 19" cabinet
(4 pcs. M6 screws with washers and cage nuts)
Instruction manual
Page 6
QPS400-3 Terminals
INSTRUMENTS
++
/
Terminals
The QPS400-3 device is equipped with 3×3 input and output BNC plug sockets and a ground socket (4 mm banana socket) on the front panel and an IEC inlet for the mains connection on the rear panel. Also on the rear panel, a voltage selector switch allows changing the rated mains voltage between 230 V and 115 V. The device fuse is integrated in the IEC inlet. When changing the rated mains voltage, the fuse must be exchanged for one with a value corresponding to the new mains voltage (see Technical Data: Power Supply).
Power
Output
Bottom Middle Top
Monitor
Bottom Middle Top
V
ACVDCVM
Input
Quadrupole Power Supply 3x400V
p
QPS 400-3
Fig. 1. The terminals and control elements on the front panel of the quadrupole power sup­ply unit QPS400-3. Power: mains switch with signal lamp
Input: signal inputs (V
AC
, VDC, V
M
) Monitor: monitor outputs (Bottom, Middle, Top) Output: amplifier outputs (Bottom, Middle, Top)
: device ground
Page 7
QPS400-3 Getting Started
INSTRUMENTS
++
/
Getting Started
Before turning on the device, the rated value of the mains voltage must be set and a corresponding fuse must be installed.
The device must be properly cooled. The heat sink at the rear panel must not be covered and the fan must be able to supply sufficiently cool air. The temperature of the cases of the amplifier ICs must never exceed 80°C.
Page 8
QPS400-3 Adjustment Elements
INSTRUMENTS
++
/
Adjustment Elements
This device is shipped to you ready to use. Adjustments should not be necessary. They should be performed only for maintenance purposes or to optimize the frequency response in the case of a special load.
Attention: Adjustments can be made by trained personnel only:
A wrong setting can damage the device.
Manipulation of the device when the case is open is perilous. Even sev-
eral minutes after turning off the device the capacitors of the power supply unit can be charged to high voltages.
!
Fig. 2. The inner assembly of the quadrupole power supply unit QPS400-3. High-voltage power-supply units (Fig. 3.): left – positive, right – negative supply voltage, Amplifier (Fig. 4): center.
Page 9
QPS400-3 Adjustment Elements
INSTRUMENTS
++
/
Fig. 3. Adjustment elements of the high-voltage power-supply units. POT1: adjusting the value of the supply voltage
Page 10
QPS400-3 Adjustment Elements
10/10
INSTRUMENTS
++
/
Fig. 4. Adjustment elements of the amplifier. CT1: frequency compensation of the input stage (balancing the channels) CT2-4: frequency compensation of the monitor outputs POT1-3: compensation of the offset voltage of the monitor outputs CT5-7, POT4-6: frequency compensation of the main amplifier