Cewe Prometer 100 User Manual

Prometer 100
High-precision meter
User Manual
BGX501-943-R01
Prometer 100 User Manual
BGX501-943-R01
Copyright © 2014, SIHPL
BGX501-943-R01 Prometer 100 User Manual Page 3 of 72
Table of Contents
1 Important Safety Information ..................................................................................................... 5
2 Disclaimer .................................................................................................................................... 5
3 Introduction ................................................................................................................................. 5
3.1 Prometer 100 meters ..................................................................................................................................... 5
3.1.1 The Prometer 100 meter in an energy management system ......................................................... 6
4 Physical Features ........................................................................................................................ 7
4.1 Power Supply, Auxiliary and Other Details ................................................................................................... 7
4.2 Front Panel of Prometer 100-R ..................................................................................................................... 9
4.3 Front Panel of Prometer 100-W .................................................................................................................. 10
4.4 Connections to the Prometer 100-R ............................................................................................................ 11
4.5 Connections to the Prometer 100-W ........................................................................................................... 13
4.6 Sealing Arrangement in Prometer 100 - R .................................................................................................. 14
4.6.1 Front Cover Sealing ...................................................................................................................... 14
4.6.2 Rear Sealing Arrangement ........................................................................................................... 14
4.7 Sealing Arrangement in Prometer 100 – W ................................................................................................ 15
4.8 Pulse Inputs and Outputs ............................................................................................................................ 15
5 Prometer 100 and M-Cubed BCS.............................................................................................. 16
5.1 M-Cubed BCS ............................................................................................................................................. 16
6
Using the Display ...................................................................................................................... 16
6.1 Auto Scroll ................................................................................................................................................... 16
6.2 Manual ......................................................................................................................................................... 18
6.3 Display Buttons ............................................................................................................................................ 19
6.4 Menu Example Screens .............................................................................................................................. 20
6.5 Events .......................................................................................................................................................... 34
7 Functions ................................................................................................................................... 35
7.1 Meter clock .................................................................................................................................................. 35
7.1.1 Time Set ........................................................................................................................................ 35
7.1.2 Time Advance and Time Retard (Sliding adjustment) .................................................................. 35
7.1.3 Daylight saving time ...................................................................................................................... 35
7.1.4 External synchronisation ............................................................................................................... 35
7.2 Support for Different Types of Energy ......................................................................................................... 36
7.3 Instant values .............................................................................................................................................. 37
7.3.1 Overview ....................................................................................................................................... 37
7.3.2 Prefix for units in the display ......................................................................................................... 38
7.3.3 Harmonics measurement .............................................................................................................. 38
7.3.4 THD ............................................................................................................................................... 38
7.4 Daily Energy Snapshot ................................................................................................................................ 38
7.5 Digital inputs and outputs ............................................................................................................................ 39
7.5.1 Inputs ............................................................................................................................................ 39
Prometer 100 User Manual
BGX501-943-R01
7.5.2 Outputs ......................................................................................................................................... 40
7.6 Communications and Security ..................................................................................................................... 41
7.6.1 Communication speed .................................................................................................................. 41
7.6.2 Security ......................................................................................................................................... 41
7.7 Loggers ........................................................................................................................................................ 42
7.7.1 Overview ....................................................................................................................................... 42
7.7.2 Logging interval and total channels .............................................................................................. 43
7.7.3 Storage of logged values .............................................................................................................. 44
7.8 Alarms ......................................................................................................................................................... 45
7.8.1 Overview ....................................................................................................................................... 45
7.8.2 Indication ....................................................................................................................................... 46
7.8.3 Display of events .......................................................................................................................... 46
7.9 Maximum demand ....................................................................................................................................... 47
7.10 Historical registers ....................................................................................................................................... 48
7.11 Time of use .................................................................................................................................................. 48
7.11.1 Day t ype ........................................................................................................................................ 49
7.11.2 Season .......................................................................................................................................... 49
7.11.3 Holiday dates ................................................................................................................................ 49
7.11.4 TOU registers................................................................................................................................ 49
7.12 Billing Cycle Support ................................................................................................................................... 49
7.12.1 Billing Cycle .................................................................................................................................. 49
7.12.2 History of Energy, Rate and MD Register .................................................................................... 49
7.12.3 History for the Cause of Billing Register ....................................................................................... 49
7.12.4 Cumulative Maximum Demand Registers .................................................................................... 50
7.13 Meter Reading ............................................................................................................................................. 50
7.14 Scaling Tariff ................................................................................................................................................ 50
7.15 Transformer compensation ......................................................................................................................... 50
7.15.1 Overview ....................................................................................................................................... 50
7.15.2 Instrument transformer compensations ........................................................................................ 51
7.15.3 Power transformer losses ............................................................................................................. 51
7.16 Quality of Supply ......................................................................................................................................... 51
7.16.1 Voltage monitoring ........................................................................................................................ 51
Appendix A: Abbreviations ............................................................................................................ 52
Appendix B: Material Declaration .................................................................................................. 52
Appendix C: Communication Ports ............................................................................................... 53
Appendix D: How to Read Meters through Ethernet Port ............................................................ 55
Appendix E: Calculation Principles ............................................................................................... 61
Appendix F: Connection and General Details ............................................................................... 66
Appendix G: List of DLMS Parameters .......................................................................................... 68
Frequently Asked Questions (FAQs) ............................................................................................. 70
BGX501-943-R01 Prometer 100 User Manual Page 5 of 72
1 Important Safety Information
Installation, wiring, testing and service must be performed in accordance with all local and national electrical codes.
Apply appropriate personal protective equipment (PPE) and follow safe electrical work practices.
This equipment must only be installed and serviced by qualified electrical personnel.
Turn off all power supplying this device and the equipment in which it is installed before working on the
device or equipment.
Always use a properly rated voltage sensing device to confirm that all power is off.
Do not perform Dielectric (Hi-Pot) or Megger testing on this device.
Connect protective ground (earth) before turning on any power supplying this device.
Failure to comply with the above safety measures could cause serious injuries. If the meter is used in a manner not specified by the manufacturer, the protection provided by
connections may be impaired. The manufacturer shall not be held responsible for failure to comply with the instructions in this manual.
2 Disclaimer
This user manual covers all types of the Prometer 100 energy meter. Depending on the product offering based on business proposal, some features or functionalities may or may not be available in the supplied version. It is therefore recommended to refer the features or functionalities according to the business offered.
The details of complete software’s features are out of the scope for this document, please contact concern sales representative for its details if required. Note that due to variations between computers and improvements in software, the screen shots shown in this manual may vary slightly from the appearance of the software on your system.
3 Introduction
This manual discusses the Prometer 100 meter features and provides the information needed to configure and use the meter. The manual covers all versions of Prometer 100-R (rack-mount) and Prometer 100-W (wall­mount) meters.
By the time you are ready to use this guide, your meter should be installed, most basic setup should have been performed, and communications/basic operation should have been verified. If the unit is not yet installed and operational, refer to the Installation Guide shipped with the meter.
This section provides an overview of Prometer 100 meters and summarizes many of their key features.
3.1 Pro met er 100 met ers
Prometer 100 meters provide revenue-accurate, true RMS measurements of voltage, current, power and energy, and are com plemented by extensive I/O capabilities, comprehensive logging, and advanced power quality measurement and compliance verification functions. The meters come with an extensive selection of pre­configured data screens and measurements, so you can use the meters as they are shipped from the factory or customize them to fit your unique requirements.
You can integrate the meters with software such as M-Cubed or wit h other energy management, SCADA, automation and billing systems, using multiple industry-standard communication channels and protocols.
Common meter applications
Transmission and distribution metering
Revenue and tariff metering
Total harmonic distortion monitoring
Prometer 100 User Manual
BGX501-943-R01
Load management
System stability monitoring
Energy pulsing and total i za tion
CT/VT error compensation
Transformer loss compensation
3.1.1 The Prometer 100 meter in an energy management system
You can use Prometer 100 meters as standalone devices, but their extensive capabilities are fully realized when used with software as part of an energy management system (EMS). EMS gives energy suppliers, service providers, and large industrial and commercial energy consumers the tools to meet all the challenges and opportunities of the new energy environment. EMS uses real-time information and control to directly address a broad range of requirements throughout the power delivery chain. This system offers an integrated solution to managing new billing structures, distributed generation, energy purchasing, energy cost control, operational efficiency, power quality and reliability.
Applications that include the meter typically require additional equipment. Display and analysis software tools are almost always used to manage, interpret and distribute the data measured or logged by a meter. There are usually a variety of tools used, and often these tools are connected using different communications standards and protocols. In many cases, a meter must also provide control capabilities and device-level data sharing.
The meter can adapt to many situations. Advanced communications allow data to be shared simultaneously across multiple networks, built-in I/O provides monitoring and control capabilities, and a variety of display and analysis tools to help you get the most from your power system.
Figure 1: An Overview
Prometer
100
Industry
Standard
Mounting
Graphical
Display
Multiple Pulse
Input and
Output
Wide range Voltage and
Current supply
Scalability and
Modularity
Multiple
Communication
Channels
BGX501-943-R01 Prometer 100 User Manual Page 7 of 72
4 Ph ysical Features
Prometer 100 is configurable for HV 3-phase 3-wire, HV 3-phase 4-wire or LV 3-phase 4-wire and is suitable for mounting in a panel or on a wall. Prometer 100 has self-powered and auxiliary-powered variants. The auxiliary­powered variant also comes with dual auxiliary support so that you can put AC or DC voltage for main and backup supply for powering up the meter. The auxiliary circuit is not intended to be connected to the secondary of measurement VT. For example, the VT secondary supply of 63.5 V AC (phase to neutral voltage) or 110 V AC (phase-to-phase voltage) needs to be supplied as a voltage input to the product. Similarly three CTs, namely R/L1, Y/L2 and B/L3, need (as applicable) to be given as a current input to the product of 1A/ 5A from secondary side.
Note:
Only Prometer 100-W variants can be configured as LV 3-phase 4-wire.
4.1 Power Supply, Auxiliary and Other Details
Details for the power supply and measurement options are shown below:
Figure 2: Power Supply Options
Variants Main supply Backup supply
No. of Aux power
supply inputs
supported
Range of Aux supply
Field Replaceable
Battery
(for meter reading
and viewing
display)
Self+Aux power supply
Aux. supply input
(Aux 2)
Measurement voltage terminals (VT supply)
One (Aux 2)
60-240 V AC/DC ± 20% or 24-48 V DC ± 20%
Optional
Self power supply
Measurement voltage terminals (VT supply)
Not available Not available Not applicable Optional
Auxiliary power supply
Aux. supply input
(Aux 2)
Aux 1 (opt iona l)
Two (dual) (Aux. 1 & Aux. 2)
Both as 60-240 V AC/DC ± 20% or
one as 24-48 V DC ± 20%
Optional
Table 1: Different Power Supply and other Variants available in Prometer 100
Prometer 100 User Manual
BGX501-943-R01
In case of auxiliary-powered and self-powered variants, meter will draw power from Aux 2 supply input. In its absence it will shift to Aux 1 supply (in case of auxiliary-powered variant) and VT supply (in case of self-powered variant).
Do not connect voltage transformer (VT) to any of the Aux supply input terminal as a general practice. If it is necessary to connect VT secondary to Aux input then make sure to connect it to the Aux1 supply only in presence of Aux 2 so that burden on measurement VT secondary can be minimal.
Connect your reliable auxiliary supply source like DC bank/AC lighting load/ Aux power transformer etc to Aux 2 terminal only so that meter burden will be handled by it; or take your best judgment to connect auxiliary supply source considering the rating and suitable operation of meter and best installation practices followed.
The field replaceable battery can be configured to support meter reading in absence of mains power supply depending on the requirement. It only supplies sufficient power to the meter reading and display circuitry and will not fully power-up the meter. Contact the concern sales representative or technical team regarding this useful feature. The battery can be replaced in the field. Take care while inserting the battery; make sure that the polarity and fitment are correct.
Ensure that the correct auxiliary voltage rating is used with the meter. The wrong voltage rating could cause damage to the meter. Therefore it is recommended to verify and crosscheck the rating-plate on the actual product in use at the site.
Figure 3: Rating Plate of Prometer 100-W (To be verified with realistic information and updated as per latest
changes)
Figure 4: Rating Plate of Prometer 100-R (To be verified with realistic information and can show one of the
rating plates with Aux 1 supply, ‘MADE in UK’ & S. No. here shown in bold and in Wall-mount not bold so
we should use same font style in order to maintain consistency, and updated as per latest changes)
BGX501-943-R01 Prometer 100 User Manual Page 9 of 72
4.2 Front Panel of Prometer 100-R
Figure 5: Front View
The front cover is made of translucent plastic with a transparent window to view the display. The cover has two top hinges which allow the front cover to swing-up, allowing access to the sealed button and field replaceable battery. The cover is secured in position by a retaining screw and also has provision to seal it.
Figure 6: Front Cover Opened
Prometer 100 User Manual
BGX501-943-R01
4.3 Front Panel of Prometer 100-W
Figure 7: Front View
The front cover is made of translucent plastic with a transparent window to view the display. The top cover is used to seal the MD (Maximum Demand) reset button and field replaceable battery. The extended terminal block cover is secured in position by retaining screws and also has provision to seal it.
Figure 8: Top Cover Removed
BGX501-943-R01 Prometer 100 User Manual Page 11 of 72
4.4 Connections to the Prometer 100-R
The diagram shows the rear connector with its pin diagrams for the meter. There are different options that may be provided as per the specification agreed with the customer at the time of order.
Connect
R/L1
Y/L2
B/L3
N
CT IN
A01
A02
A03
NC
CT OUT
A1
A2
A3
NC
VT
B1
B2
B3
B0
Dual AUX
Supply
P/+ Terminal N/- Terminal
AUX1
B8
B9
AUX2
B5
B6
Digital Input/Output
8 Outputs and 4 configurable Inputs/Output
O/P 1
C0, C1
I/O 1
D3, D2
O/P 2
C1, C2
I/O 2
D9, D4
O/P 3
C4, C3
I/O 3
D7, D8
O/P 4
C4, C9
I/O 4
D5, D6
O/P 5
C7, C8
O/P 6
C7, C6
O/P 7
D0, D5
O/P 8
D0, D1
Figure 9: Rear Connector with Pin Details
Prometer 100 User Manual
BGX501-943-R01
Note: For 3-phase 3-wire CT/VT operated refer to figure A
For 3-phase 4-wire CT/VT operated refer to figure B For 3-phase 3-wire and 3-phase 4-wire connections from B5 to D9 are the same. For self-powered variant Aux.1 (B8 and B9) supply is not available.
Figure 10: Connection Diagram
Note: One side of the CT secondary wiring should be earthed according to local practice.
The recommended size of the CT, VT and Aux power supply cable is 2.5 sq mm with lugs type as ring (M3 type) for CT and H type for Aux and VT terminals.
In case of Prometer 100-R the internal earthing cable between the meter and the rack should also be connected, and for this an M4 size screw is used in the meter and in the rack. The same ring type connector as used in the CT connection can also be used.
The internal earthing cable should be about 5-10 cm in length to allow for ease of fitting and access and not fouling with sharp edges of rack etc. Finally a proper earthing cable from rack to earth should be put up by installer. Other accessories or shipway kit is supplied based on the requirement like seals, communication cords etc.
BGX501-943-R01 Prometer 100 User Manual Page 13 of 72
4.5 Connections to the Prometer 100-W
Connections to the Prometer 100-W are made on the meter terminal under the terminal cover.
Connect
R/L1
Y/L2
B/L3
N
CT IN
1 4 7
NC
CT OUT
3 6 9
NC
VT 2 5 8 11
Dual AUX
Supply
P/+ Terminal N/- Terminal
AUX 1
13
14
AUX 2 15 16
Digital Input/Output
8 Outputs and 4 configurable Inputs/Output
O/P 1
18, 19
I/O 1
32, 33
O/P 2
20, 21
I/O 2
34, 35
O/P 3
22, 23
I/O 3
36, 37
O/P 4
24, 25
I/O 4
38, 39
O/P 5
26, 27
O/P 6
28, 29
O/P 7 30, 31
Figure 11: Rear Connector with Pin Details
Current, Voltage and Auxiliary supply connections
Pulse Input/ Output connections
Figure 12: Connection Diagram
Prometer 100 User Manual
BGX501-943-R01
4.6 Sealing Arrangement in Prometer 100 - R
4.6.1 Front Cover Sealing
Figure 13: Front Cover Sealing Arrangement
The front cover can be sealed in the closed position. This will stop the front cover from being opened and restrict unauthorised access to the MD Reset pushbutton and internal areas. There are also sealing points on either side of the meter for securing it to a rack or frame. The sealing bore diameter is 2.0 mm and is suitable for seals.
4.6.2 Rear Sealing Arrangement
When the meter is fitted into the rack, a cover can be fitted which conceals all the rear connectors. The figure below show the sealing points for the rear cover. The figure shows the single rack example with an enlarged detailed view of the sealing points.
Figure 14: Front and Rear Sealing Points – 11” Rack Installation
Front
Cover
Translucent
Transparent
Window
Locking Scr ew
& Sealing points
Right-hand side
Sealing & Locking point
Left-hand s ide
Sealing & Locking point
BGX501-943-R01 Prometer 100 User Manual Page 15 of 72
4.7 Sealing Arrangement in Prometer 100 – W
The meter incorporates sealing bars and screws with through-holes, through which traditional lead/wire seals can be inserted. When utilized, these lead/wire seals can help prevent unauthorized personnel from gaining access to meter internals or to button under the cover. The sealing provision in Prometer 100-W is as follows:
The front cover can be sealed through the two sealing points under the terminal cover.
The secondary terminal cover can be sealed through a sealing point.
The extended terminal cover can be sealed through the two sealing points.
The top cover can be sealed through the two sealing points.
The 1107 optical communication port can be sealed using rotational seal.
Figure 15: Sealing Points
4.8 Pulse Inputs and Outputs
The product support multiple pulse inputs and outputs. A maximum of eight pulse outputs (for Prometer 100-R) and seven pulse outputs (for Prometer 100-W); and four configurable pulse input/ output can be provided as per the specification agreed at the time of order.
Two pulse outputs (3 and 4 in case of Prometer 100-R / 2 and 3 in case of Prometer 100-W, refer to Figure 10 and Figure 12 for details) are linked to two pulse output LEDs indication as available on front side of meter so that user can have a visualisation sort of feature by physically seeing the LEDs. The pulse output 3 (in case of Prometer 100-R) / 2 (in case of Prometer 100-W) is linked to pulse output LED 1 and the pulse output 4 (in case of Prometer 100-R) / 3 (in case of Prometer 100-W) is linked to pulse output LED 2. The Pulse Output LEDs can be configured through M-Cubed 100 either at factory or in field. One pulse input can be used for time synchronization applicat ion .
Configurable pulse input/output rating: 24 to 24 0 V AC/DC. Isolation will be available for each individual input/output.
Pulse output rating: 24 to 40 V DC or 48 to 240 V AC/DC @ 100 mA (Pulse outputs will have volt free contact). The only one rating will be applicable for entire block (each containing four outputs) and so isolation available will be for entire block. The outputs are of solid-state type and when the meter is turned off, they are open.
Prometer 100 User Manual
BGX501-943-R01
5 Prometer 100 and M-Cubed BCS
This section describes the various operations that can be performed using the M-Cubed BCS with Prometer 100 meter.
5.1 M-Cubed BCS
Figure 16: Prometer100 and M-Cubed BCS
M-Cubed stands for Modular Meter Management and is the name of Secure’s software suite for programming meters, reading data and reporting from energy meters. M-Cubed has separate modules that can be configured to suit particular applications and access rights required by individual users.
M-Cubed can be used for:
Configuration
Commissioning
Meter Reading
Data Viewing
The M-Cubed helpfile contains a detailed description of all these functions.
6 Using t he Display
There are two types of display mode: Auto Scroll and Manual Mode. The two types of display modes and their associated settings and parameters are discussed in this section. All single parameters, e.g. voltage, will be displayed with their OBIS codes. A large selection of parameters (Auto + Manual) can be chosen for display.
6.1 Auto Scroll
Auto Scroll is the default mode. A large number of parameters can be chosen for this mode. The display time out time can be configured in the field using M-Cubed or by the display keys. Once the display button is pressed, the Auto Scroll mode will be interrupted and will switch to manual mode. If no button is pressed in manual mode, the display will time out and revert to auto mode. The display will resume from the last displayed parameter. The display LCD backlight is always on.
The Auto Display pages will look as shown below (details can be checked as agreed in purchase order):
BGX501-943-R01 Prometer 100 User Manual Page 17 of 72
Prometer 100 User Manual
BGX501-943-R01
Note:
‘T’ stands for Total (Fundamental with harmonics) ‘F’ stands for Fundamental ‘C’ stands for Current ‘MD’ stands for Maximum Demand
6.2 Manual
Display parameters are grouped in to a number of pages. Each page parameters can be individually selected. A considerable number of parameters can be assigned in manual mode.
Display Groups
The display is arranged into groups for easier navigation.
Fixed Display Page: Fixed default displays (not configured by tariff)
User Configurable Pages: User selected parameters. Page name can also be configured. Up to 7 pages
can be configured.
Favourite: These are selected from the user configurable pages and up to 20 parameters can be selected in the field.
Configuration: Configuration page for Metrological LED, Display time out, Delete Favourite page displays, Reset Battery time, Meter Cons tant, MODBUS and language configuration.
BGX501-943-R01 Prometer 100 User Manual Page 19 of 72
6.3 Display Buttons
The meter has four user buttons. One button can be used for performing MD reset or for navigating through the sealed button display sequence depending on the requirement and is behind a sealable door. The other three buttons are used for navigating through the display and configuration menu options.
Figure 17: User Buttons
A page can be selected by pressing the enter button. A navigation screen will appear showing all the available pages. The Up and Down buttons can then be used to choose a page. Press the Enter button to select your choice.
Figure 18: Selecting a Page (instead of the word ‘Tamper’, ‘event’ can be used in the display menu)
Once you have selected a target page, you can then use the Up/Down buttons to scroll through that selected page. Parameters are cyclically displayed in the selected page, i.e. after the last parameter in the list is displayed, the display will return to the first in the list, and so on.
Press the Enter button to return to the immediate parent page.
Prometer 100 User Manual
BGX501-943-R01
6.4 Menu Example Screens
Fixed Display Page
Select the Fixed Display Page from the top line menu. The following screen will be displayed.
Figure 19: Fixed Display Page
The Fixed Display pages are shown below:
BGX501-943-R01 Prometer 100 User Manual Page 21 of 72
Refer to tariff for Multiplier value
Note: ‘I’ stands for Import ‘E’ stands for Export ‘T’ stands for Total (Fundamental with harmonics) ‘F’ stands for Fundamental ‘R’ stands for Right ‘L’ stands for Left ‘Cfg. IO’ stands for Configurable Input/Outp ut
User Configurable Pages
Up to seven user configurable pages can be defined complete with page title e.g. Instant Parameters, see Figure 20 below. The page title can have a maximum of 20 alphanumeric characters. The illustrations shown below are indications of how the pages and their respective displays will be displayed. The final pages and their corresponding displays are dependent on the specification of your purchase or der.
Prometer 100 User Manual
BGX501-943-R01
Figure 20: User Configurable Pages
Instantaneous Parameter s
Use the Up/Down buttons to scroll to the Instant Parameters page. The selection will be highlighted.
Press the Enter button to open the page. Use the Up/Down buttons to scroll within the page and view the various
screens.
BGX501-943-R01 Prometer 100 User Manual Page 23 of 72
Prometer 100 User Manual
BGX501-943-R01
Note:
‘D’ stands for degree
‘M’ stands for minutes
Energy Registers
In this example the user defined page has been configured to view the energy registers and an appropriate page title has been chosen for easy identification. Always choose a user friendly and self-explanatory name for your titles. We have used “Energy Register” for our title example.
Use the Up/Down buttons to scroll to the Energy Registers page. The selection will be highlighted.
Press the Enter button to open the page. Use the Up/Down buttons to scroll within the page and view the various
screens.
BGX501-943-R01 Prometer 100 User Manual Page 25 of 72
Demand Data
This page has been set up to capture demand values.
Use the Up/Down buttons to scroll to the Demand Data page. The selection will be highlighted.
Press the Enter button to open the page. Use the Up/Down buttons to scroll within the page and vie w the various screens.
Prometer 100 User Manual
BGX501-943-R01
High Resolution Energy Register
This page has been set up to capture hi gh re solution energy registers .
Use the Up/Down buttons to scroll to the High Resolution Energy Registers page. The selection will be highlighted.
Press the Enter button to open the page. Use the Up/Down buttons to scroll within the page and view the various screens.
BGX501-943-R01 Prometer 100 User Manual Page 27 of 72
Previous IP data
This page has been set up to capture p revious IP data .
Use the Up/Down buttons to scroll to the Previous IP Data registers page. The selection will be highlighted.
Press the Enter button to open the page. Use the Up/Down buttons to scroll within the page and view the various screens.
Prometer 100 User Manual
BGX501-943-R01
Event Information
This page details tamper events and status. Events are discussed in section 6.5
Use the Up/Down buttons to scroll to the Tamper Information page. The selection will be highlighted.
Press the Enter button to open the page. Use the Up/Down buttons to scroll within the page and view the various screens.
Miscellaneous
This page details miscellaneous displays.
Use the Up/Down buttons to scroll to the Miscellaneous information page. The selection will be highlighted.
Press the Enter button to open the page. Use the Up/Down buttons to scroll within the page and view the various screens.
BGX501-943-R01 Prometer 100 User Manual Page 29 of 72
Scroll Lock
Locking of individual para meter page in user configurable pag es (manual pages) –
Figure 21: Locking the manual page parameter
The screen can be locked to show a desired parameter. To screen lock a parameter, select the parameter using the menu buttons (this is applicable under user configurable pages only). Press the Up key for 5 seconds. During this process the display will temporarily move to the next parameter, then after 5 seconds will display your selection.
T o unlock the screen (i.e. revert to auto scroll mode) press the Up key for 5 seconds.
Favourite Page
This page is used for your selection of display parameters.
Figure 22: Adding parameter to favourite page
To add a parameter, go to the page containing the parameter that you require. Once the required page is selected press the down key for 5 seconds, the selected parameter will now be added to your favourite page.
When the favourite page is full, you will need to delete complete list in this page to make space for the new one, so be sure before selecting parameters as maximum 20 parameters can be selected in favourite display page.
To delete the complete list of Favourite parameters, go to the configuration page and select ‘Del Fav Parameters’.
press and hold the Key for
5 seconds
Up
Up
press and hold
the Key
for 5 seconds
Down
Down
Prometer 100 User Manual
BGX501-943-R01
Configuration Page
Figure 23: Front view showing metrological LEDs
The configuration pages are used to set-up and enable features such as the metrological LEDs, display timeout, meter constant, reset battery time, MODBUS configuration and language on display. They are also used to manage your favourite selections.
The configuration page is password protected which can be accessed only by users authorized to change the above configurations.
At the password prompt enter the 6-digit password (default password is ‘000000’) using Up and Down buttons Use the Down key to enter a digit at the cursor position and use the Up key to move the cursor to the right while entering a digit. After completing the entry, press the Enter button to confirm. Correct password gives access to the first screen of configuration page. An incorrect password will display ‘Incorrect Password’ message.
The Prometer 100 meter display can be set to different languages. The following languages are available:
English
Swedish
German
French
Spanish
Russian
Arabic
The configuration page displays are shown below.
Use the Up/Down buttons to scroll to the Configuration page. The selection will be highlighted.
Press the Enter button and enter the correct password using Up and Down buttons to open the page.
Use the Up/Down buttons to scroll within the page and view the various screens.
BGX501-943-R01 Prometer 100 User Manual Page 31 of 72
Display to be added
Prometer 100 User Manual
BGX501-943-R01
(Russian and Arabic language options are missing in the above display) The following displays are self-explanatory and occur at the time of events such as configuration, file download
etc. and are included here for your reference only.
BGX501-943-R01 Prometer 100 User Manual Page 33 of 72
User Selectable Display Examples (displays to be updated)
Description
HV (3-phase 4-wire) Display
HV (3-phase 3-wire) Display
Phasor Diagram
Bar graph
Same for HV3 and HV4
MD Value
Same for HV3 and HV4
Battery Status
Greater than equal to 30%
Same for HV3 and HV4
Battery Status
Less than 30% & Greater than equal to 20%
Same for HV3 and HV4
Battery Status
Less than 20%
Same for HV3 and HV4
Prometer 100 User Manual
BGX501-943-R01
Battery Remaining Hours
Same for HV3 and HV4
6.5 Events
The Prometer 100 has a number of defined events that are stored in the meter’s event log when they occur and restore. The events are arranged into different logical compartments with defined numbers of events logging. An event is displayed with a description and a time stamp in M-Cubed. All the events will be displayed in following categories in meter display as– voltage related, current related and events other than voltage & current related. The events do not enforce any electrical value changes inside the meter such as running the meter on I
max
etc.
Events are logged if the condition for the detection of an event persists for a specified duration, known as the persistence time. Events can have a different persistence time for occurrence and restoration as applicable. The persistence time for an occurrence as well as for restoration is configurable at factory end for respective event as applicable. However some events like power on/off, magnet will not have any restoration time because of the nature and type of the event. Similarly some events can have the same condition of tamper detection as per the nature and type of the event in a given circumstances.
The general events supported are as follows –
1. Phase-wise missing potential
2. Voltage unbalance
3. Invalid Voltage
4. Over Voltage
5. Under Voltage
6. Phase wise current circuit reversal
7. CT Miss
8. Current unbalance
9. Feeder Supply Fail
10. Po wer on/ of f
11. Front cover and ETBC Open detection in power on/off condition (applicable for Prometer 100-W)
There are other customised events like Neutral Disturbance (ND), Magnet immunity, CT open/ Bypass and %THD for voltage and current are also supported in meter.
BGX501-943-R01 Prometer 100 User Manual Page 35 of 72
7 Functions
This section provides an overview of the functions available in Prometer 100. All functions in the meter can be both configured and read in M-Cubed. In many cases, M-Cubed can also export data to a file or print out data.
7.1 Meter clock
The Prometer 100 has an integrated real-time clock for time-dependent functions. Information about time adjustment and daylight saving time status is recorded with tags on the logged values (refer to the section Loggers on page 42). Time set is also noted as an event in the event log; see the section Display of events (pg.
46).
7.1.1 Time Set
The meter’s date and time can be set to an absolute point in time. Instantaneous adjustment of the meter’s time can influence logged values. For this reason, instantaneous adjustment of the meter clock is primarily intended for use at initial configuration of the meter.
7.1.2 Time Advance and Time Retard (Sliding adjustment)
When the Prometer 100 is not being used for on-line application, the meter’s time can be advanced or retarded using time advance or time retard commands with M-Cubed BCS. The meter time can be retarded or advanced by t
tc
seconds spanned over Ntc consecutive blocks of 15 minutes (ttc/ N
tc
seconds for each block). The meter automatically adjusts for any time correction during the load survey reading period. Once a meter gets and accepts a time advance/retard command, it is not possible to do time adjustment for the next 7 days.
7.1.3 Daylight saving time
Prometer 100 offers the alternative of letting the meter clock follow daylight savings time. At a specified date, the meter clock is adjusted forward, and at another, adjusted backward. Prometer 100 can store 15 years of DST configuration.
Example: On 28 March the clock is to be adjusted forward, from 02:00 to 03:00. The adjustment back to standard time is to occur on 31 October at 3:00 (daylight savings time) when the clock is to be set back to 02:00. The following is set in the meters: Begin March, 28, 02:00. End October, 31, 03:00 and the standard time is to be adjusted by 60 minutes.
7.1.4 External synchronisation
The meter time can be adjusted by a pulse on one of the meter’s digital inputs. When a pulse is registered, the clock is adjusted to the closest multiple of a specified synchronisation interval. If the synchronisation interval is, for example, one minute and the time is 13:00:29, a pulse will adjust the clock to 13:00:00. If the time had instead been 13:00:31, the clock would have been adjusted to 13:01:00.
Available synchronisation intervals are:
1, 2, 5, 10, 15, 20, 30, 45 and 60 minutes
12 and 24 hours
A digital input must be configured for clock synchronisation (refer to the section Digital inputs and outputs on page 39).
Note: Meter will not sync the time if the time difference is more than the time adjustment limit configured in the meter. By default, the time adjustment limit is set as 25 seconds. The time adjustment limit can be configured as any value between 0 to 30 seconds.
Prometer 100 User Manual
BGX501-943-R01
7.2 Support for Different Types of Energy
Prometer 100 meter is used for feeder based applications where energy may flow in both the directions. Prometer 100 is an Import Export type meter; its metering unit is capable of logging energy in both directions.
III
III IV
Active power import
(+)
Active power export
(-)
Reactive power
export (-)
Reactive power
import (+)
S
Q
P
Capacitive
Inductive
Quadrants
Phase angle
Current relative to voltage
I
0 to 90°
Lagging
II
90 to 180°
Leading
III
-180 to -90°
Lagging
IV
-90 to 0°
Leading
Prometer 100 supports different tariff structures and number of Energy types (refer to Table 2). Any combination of energy types can be provided as per the specification agreed at the time of order.
The Energy Channel Registers are shown below:
Sr. No.
Parameters Instantaneous Billing
TOD MD / UMD
TOD Rate
Daily Energy
Logger 1
Logger 2
1
Active Import Total (Q1+Q4)
Y
Y Y Y Y Y
Y
2
Active Export Total
(Q2+Q3)
Y
Y Y Y Y Y
Y
3
Active Import Fundamental
(Q1+Q4)
Y
Y Y Y Y Y
Y
4
Active Export
Fundamental
(Q2+Q3)
Y
Y Y Y Y Y
Y
5
Reactive Import while Active Import – Q1
Y
Y Y Y Y Y
Y
6
Reactive Import while Active Export – Q2
Y
Y Y Y Y Y
Y
7
Reactive Export while Active Export – Q3
Y
Y Y Y Y Y
Y
8
Reactive Export while
Active Import – Q4
Y
Y Y Y Y Y
Y
9
Apparent – While Active Import (See
Note 2) Y Y Y Y Y Y
Y
10
Apparent – While
Active Export (See
Note 2) Y Y Y Y Y Y
Y
11
Reactive Import (Q1+Q2)
Y
Y Y Y Y Y
Y
12
Reactive Export (Q3+Q4)
Y
Y Y Y Y Y
Y
BGX501-943-R01 Prometer 100 User Manual Page 37 of 72
13
Reactive Inductive (Q1+Q3)
Y
Y Y Y Y Y
Y
14
Reactive Capacitive (Q2+Q4)
Y
Y Y Y Y Y
Y
15
Net Active (Imp –
Exp) * Y Y
N N
Y Y Y
16
Net Reactive
(Q1+Q2-Q3-Q4) *
Y
Y
N N
Y Y Y
17
Active Import Total (Q1+Q4) - Phase 1
Y
Y
N N
Y Y Y
18
Active Import Total (Q1+Q4) - Phase 2
Y
Y
N N
Y Y Y
19
Active Import Total (Q1+Q4) - Phase 3
Y
Y
N N
Y Y Y
20
Active Export Total (Q2+Q3) - Phase 1
Y
Y
N N
Y Y Y
21
Active Export Total
(Q2+Q3) - Phase 2
Y
Y
N N
Y Y Y
22
Active Export Total (Q2+Q3) - Phase 3
Y
Y
N N
Y Y Y
Table 2: Energy Channel Registers
The pre-defined energy channels can be configured for display, billing, TOD, MD, Load Survey and Daily Energy in line with purchase order requirements at the time of order.
Notes:
1) All the bi-directional energy register s (* marked) will have sign indicat ion (‘-‘sign will be ava ilable for negative value and no sign for positive value).
2) In ‘Apparent’ and ‘Net Active’ energy calculat ion, ‘Active’ energy can be either ‘fundam ental’ or ‘total’. This can be configured through the tariff tool. Both energies need to be of same type.
3) Single phase measurements (17-22) are supported by the 4 wire configuration.
For energy types 1-16: These registers are continuously increased depending on the selected energy.
Example: Main Energy Register kWh (I) will show the total imported active energy logged till date. Recording of all supported energy types is pos sible but only those energy t ypes are logged into the m emory
which is specified by the ta riff file. An indi vidua l registe r is pr ovided f or al l selec ted ener g y types. T hese regi ster are called Main Energy Registers . W henever an individual energy type is generated / consum ed, its value is updated in the corresponding main energy register. These registers cannot be reset.
7.3 Instant values
Besides energy, the Prometer 100 can also measure instant values. Instant values are constantly changing values such as current, voltage, power and harmonics. The instant values are updated every second. The formulas and definitions used to calculate the values are presented in Appendix E: Calculation Principles on page 61.
7.3.1 Overview
This table provides an overview of the instant values that can be read on the meter. Readings can be viewed with M-Cubed, on the displa y and with other software that has implemented Prometer 100’s communication protocol. Most instant values can be logged; for more information, see section Loggers on page 42.
Instant value
Available on 3-element meter
Available on 2-element meter
Real Time Clock – Date and Time Yes Yes
Phase Voltage
Yes (L1, L2, L3, Average)
No
Line Voltage (L12, L23, L31, Average)
Yes
Yes
Line Current
Yes (L1, L2, L3, Average)
Yes (L1, L3, Average)
Active Current
Yes (L1, L2, L3, Average)
Yes (L1, L3, Average)
Reactive Current
Yes (L1, L2, L3, Average)
Yes (L1, L3, Average)
Voltage Phase angle (L12, L23, L31) Yes Yes
Prometer 100 User Manual
BGX501-943-R01
Current Phase angle
Yes (L1, L2, L3)
Yes (L1, L3)
Active power total
Yes
Yes
Active power per phase
Yes
No
Active power fundamental Yes Yes
Active power fundamental per phase
Yes
No
Reactive power total
Yes
Yes
Reactive power per phase
Yes
No
Apparent power total
Yes
Yes
Apparent power per phase
Yes
No
Power factor total Yes Yes
Power factor per phase
Yes
No
Frequency
Yes
Yes
Average THD voltage
Yes
Yes
THD voltage per phase
Yes
No
Average THD current
Yes
Yes
THD current per phase
Yes
No
Average THD power Yes Yes
THD power per phase
Yes
No
7.3.2 Prefix for units in the display
The presentation for units and the number of decimals depends on the magnitude of the value.
7.3.3 Harmonics measurement
Harmonics numbers 2 to 31 are measured for all currents and voltages. At a fundamental frequency of 50 Hz, the second harmonic is 100 Hz, the third harmonic is 150 Hz, etc. Both the harmonics’ amplitude and phase angle are measured and included in the calculation of power and energy, and their magnitude can be read via the meter’s communication protocols. In M-Cubed, harmonic magnitude is presented with a diagram.
Voltage harmonics profile
Parameter
Available on 3-element meter
Available on 2-element meter
Real Time Clock – Date and Time
Yes
Yes
Voltage harmonics (2 to 31) – L1
Yes
Yes
Voltage harmonics (2 to 31) – L2 Yes No Voltage harmonics (2 to 31) – L3 Yes Yes
Current harmonics profile
Parameter
Available on 3-element meter
Available on 2-element meter
Real Time Clock – Date and Time Yes Yes Current harmonics (2 to 31) – L1 Yes Yes
Current harmonics (2 to 31) – L2
Yes
No
Current harmonics (2 to 31) – L3
Yes
Yes
7.3.4 THD
THD stands for Total Harmonics Distortion and is a measurement of the amount of harmonics present in a signal. Voltages and currents’ THD can be read via M-Cubed and on the display.
7.4 Daily Energy Snapshot
Energy Snapshot feature saves the value of a particular energy register at a particular time. Prometer 100 stores a snapshot of different energy registers (can be up to 28 energies) on a daily basis at predefined time as selected from tariff configuration (generally it is set at midnight). Snap shots are generally stored for 45 days and can be stored for a maximum of 100 days as configured at factory. The updating of Energy Snap shot records is done in a rollover fashion, i.e. each day a new energy snapshot is stored in the memory and the earliest record is deleted. So at any time a meter will have energy snapshot records for the last 45 days.
BGX501-943-R01 Prometer 100 User Manual Page 39 of 72
7.5 Digital inputs and outputs
The Prometer 100 has several inputs and outputs that can be configured to perform various tasks. Both inputs and outputs are protected against overv oltages b y vari s tors . T he y also have an isolated int erf ac e between the electronics and the surroundings to ensure personal safety. For electrical data on the meter’s inputs and outputs, see Appendix F: Connection and General Details (page 66).
7.5.1 Inputs
The inputs can be configured as follows:
Not used The input is not used.
Finish historical period (To be verified that only ‘Finish Hist Period’ Input can be set to inverted as in ConfigView S/W it allows none of the inputs to b e set as inverted)
An incoming pulse will result in the present period ending and registers being copied to historical registers. The meter registers pulses on positive or negative flanks, depending on if the input is set to inverted or not. By setting limits for maximum and minimum pulse lengths, the meter can be limited as to what it detects as a valid pulse. Pulses with lengths beyond the established limits are ignored. For a pulse to finish historical period, it is also necessary that the Data communication be configured to allow this. For more information, see the section Historical registers (pg.
48).
Pulse
length
GND
VCC
Pulse length
(inverted input)
The figure shows pulse lengths when an input is inverted or non-inverted, respectively.
Pulse input To register pulses from pulse-producing units such as energy meters, pulse inputs are used. Incoming pulses are accumulated in registers called external registers. There is an external register connected to each input on the meter. For external registers, a factor is configured by which the number of incoming pulses is multiplied. Prefixes and the number of decimals can also be configured for the registers. Moreover, the registers can be configured with descriptive texts.
Time synchronisation When incoming pulses are received, the meter’s clock is synchronised at a specific interval For available synchronisation intervals and more detailed information on time synchronisation, see the section Meter clock (pg. 35).
Rate input Up to three digital inputs can be configured to control the active rate. Each input will correspond to a bit and the significance is also defined for the digital input. A high level on the input will signify that the bit is “1” while a low level will signify “0”. How the digital input levels are mapped to rates is defined in the Time of use panel (pg.48).
7.5.1.1 Registration of pulses A pulse must be at least 20 ms (for 50 Hz) and 16 ms (for 60 Hz) long to be guarantee d of being det ected by the
meter. The maximum pulse width that the meter can handle is 300 ms.
Prometer 100 User Manual
BGX501-943-R01
7.5.2 Outputs
The outputs can be configured as follows:
Not used The output is not used.
Energy Pulse output The output is used to pulse an energy type that the meter is measuring. A multiplier is specified for the output as pulses/unit and the pulse length is specified for all pulse outputs. The shortest possible pulse length is 20 ms.
Pulse
Gap
Maximum pulse frequency at outputs is limited so that the gap is at least as long as the pulse length.
Long Pulse output The output is used to generate long pulse whose length is between 2 to 15 seconds. By default, the pulse length is set as 10 seconds. The long pulse output is used for the following functions:
o MD Register Change
On switching the MD registers, the output generates a long pulse.
o Rate Register Change
On switching the Rate registers, the output generates a long pulse.
o Billing Act ion
On performing billing action, the output generates a long pulse.
o DIP Start
At the start of a set demand integration period, the output will go active for the configured pulse length before returning to the inactive state. See the section Maximum demand (pg. 47).
Remote control With this function, the output can be made active or inactive by sending commands to the meter via the DLMS/PACT protocol. This function can be used to control anything that can be controlled with a digital relay output. For this, it is also necessary that the Data communication be configured to enable remote control.
State Output
o MD Register
With this function, the output remains active or inactive for the time zones during which the selected MD registers are activated.
o Rate Register
With this function, the output remains active or inactive for the time zones during which the selected Rate registers are activated.
o Alarm output
When an output is set to function as an alarm output, one of the user-defined alarms can be chosen to indicate at the output. When an alarm occurs, the output switches to active, and when the alarm state ceases, the output returns to inactive. In the section Alarms (pg. 45), user­defined alarms are described and how they can be configured.
Note that the outputs are inverted via firmware. If the meter loses its auxiliary power, the relay will open, regardless of it is inverted or not.
BGX501-943-R01 Prometer 100 User Manual Page 41 of 72
7.5.2.1 Output states An active output means a closed relay when the output is not inverted. When the output is inverted, the active
relay is open. The Prometer 100 outputs are of the solid-state type and when the meter is turned off, they are
open.
7.5.2.2 Meter vari ants
Inputs/Outputs
Prometer 100-R
Prometer 100-W
4 configurable input/output
8 outputs and 4 configurable input/output
7 outputs and 4 configurable input/output
7.6 Communications and Security
Prometer 100 meters have optical port, RS232, RS485 and Ethernet ports for communication. Note:
The Prometer 100-W supports attachment of field replaceable communication modules for RS232 and RS485. These modules are optional and can be procured separately.
Communication
channel
Default / Max
supported Baud
rate
Supported protocol Usage
Optical port- IEC1107
300 / 19200 bps DLMS
Local Meter Reading
RS485 (RJ-45 in & out) 9600 / 57600 bps DLMS, Modbus RTU
Meter reading, online
monitoring, third party interface
RS232 9600 / 57600 bps DLMS
Remote meter reading through external modem
Ethernet port 10 / 100 Mbps
Modbus TCP (port no.: 502) DLMS TCP (port no.: 4059)
Meter reading, online
monitoring, third party
interface
Note: For more information on protocol support, see the document ‘Prometer 100 meter reading’.
7.6.1 Communication speed
The meter’s optical port always starts with a baud rate of 300 bps, regardless of what is configured, before shifting over to the specified communication speed. This means that software (for example, M-Cubed) that communicates with the meter via the optical port does not need to know the speed that the meter’s optical port is set to. RS232 and RS485 communication ports differ in this respect. They start at the specified baud rate from the beginning, which means that connected software must be aware of the speed to be able to communicate. RS232 communication port can be set at a speed of between 9600 bps and 57600 bps, and RS485 communication port can be set at a speed of between 9600 bps and 57600 bps. The optical port can be set at a speed of between 300 and 19200 bps.
7.6.2 Security
The meter has five authorisation levels that can limit access to the meter during communication via any of the meter’s communication ports. Authorisation levels are password-protected.
Authorisation levels
1
Provides access to reading.
2 Provides access to everything in level 1 plus access to set the clock and
finish historical periods (also resets maximum demand values).
Prometer 100 User Manual
BGX501-943-R01
3 Provides access to everything in level 2 plus access to configure the
meter.
4 Provides access to everything in level 3 plus access to transfer new
firmware to the meter.
5 Provides access to everything in level 4 plus access to calibrate the
meter, reset registers and change certain security setting.
7.6.2.1 Limitation of total access attempts The meter limits the total number of access attempts to six when incorrect passwords are entered. At the
seventh attempt, the meter blocks access whether the password is correct or not. The block is in effect until the next hour shift. After that, new password attempts may be made.
7.6.2.2 Passwords A password consists of up to 12 case insensitive alpha-numerical characters. The authorisation check may be
deactivated for a level by deleting the password. When connecting to the meter, access is granted to the highest level that is lacking password regardless of the password given by the user.
If the setting ‘Require COP password compliance (min length 6)’ is activated, a new password is required to be at least 6 characters. The setting can only be changed at access level 5.
7.6.2.3 Security settings The following security settings modify what can be configured at what access level. The settings are of the type
active/not active, and can only be changed at access level 5.
Permit measuring configuration only at level 5 If this setting is active, settings in the
measurement form can only be changed at access level 5.
Block configuration of transformer compensation If this setting is active, transformer compensation
cannot be changed at all.
Block configuration of display sequence 4 If this setting is active, it is not possible to change
the content of display sequence 4, or change its name or activation.
7.6.2.4 Access restriction for measurement configuration Generally the meter allows to be reconfigured at access level 3. This function requires level 5 for measuring
configuration.
7.7 Loggers
The Prometer 100 has two identical, parallel and individually configurable loggers. That which is described in this section applies both to logger 1 and logger 2.
7.7.1 Overview
A logger in a Prometer 100 can log values for instant quantities, energy registers and external registers. Some quantities can be logged both by phase and as total values for all three phases, others only as total values. The table provides an overview of quantities that can be logged. Certain instant values in the table are not available in 2-element meters and thus cannot be logged; see the section I ns tant va lu es (pg. 37). Logger Parameters can be read as profile data and SIP wise.
Quantity
Computation Method
Real Time Clock – Date and Time
Instant
Active Import Total (Q1+Q4)
Consumption
Active Export Total (Q2+Q3)
Consumption
Active Import Fundamental (Q1+Q4)
Consumption
Active Export Fundamental (Q2+Q3)
Consumption
Reactive Import while Active Import – Q1
Consumption
Reactive Import while Active Export – Q2
Consumption
BGX501-943-R01 Prometer 100 User Manual Page 43 of 72
Reactive Export while Active Export – Q3
Consumption
Reactive Export while Active Import – Q4
Consumption
Apparent – While Active Import
Consumption
Apparent – While Active Export
Consumption
Reactive Import (Q1+Q2)
Consumption
Reactive Export (Q3+Q4)
Consumption Reactive Inductive (Q1+Q3) Consumption Reactive Capacitive (Q2+Q4) Consumption Net Active (Imp – Exp)* Consumption Net Reactive (Q1+Q2-Q3-Q4)* Consumption Active Import Total (Q1+Q4) - Phase 1 Consumption Active Import Total (Q1+Q4) - Phase 2 Consumption Active Import Total (Q1+Q4) - Phase 3 Consumption Active Export Total (Q2+Q3) - Phase 1 Consumption Active Export Total (Q2+Q3) - Phase 2 Consumption Active Export Total (Q2+Q3) - Phase 3 Consumption Phase Voltage - Phase wise, Av erage 3 phase Min / Max / Avg / Instant Line Voltage - Phase wise, Average 3 phase Min / Max / Avg / Instant Line Current - Phase wise, Average 3 phase Min / Max / Avg / Instant Active Power - Phase wise, Average 3 phase Min / Max / Avg / Instant Reactive Power - Phase wise, Av er age 3 phase Min / Max / Avg / Instant Apparent Power - Phase wise, Average 3 phase Min / Max / Avg / Instant Power Factor - Phase wise, Average 3 phase Min / Max / Avg / Instant THD Voltage (%) - Phase wise, Average 3 phase Min / Max / Avg / Instant THD Current (%) - Phase wise, Average 3 phase Min / Max / Avg / Instant THD Power (%) - Phase wise, Average 3 phase Min / Max / Avg / Instant Frequency Min / Max / Avg / Instant Voltage Angles Min / Max / Avg / Instant Voltage Current Angles Min / Max / Avg / Instant
Voltage Harmonics - Phase wise, Average 3 phase (3rd,
5th, 7th, 9th, 11th, 13th and 15th)
Min / Max / Avg / Instant
Current Harmonics - Phase wise, Average 3 phase (3rd, 5th, 7th, 9th, 11th, 13th and 15th)
Min / Max / Avg / Instant
Pulse Input Counter (1 to 4) Instant Status Flag (Time adjusted, Time disturbed, Alarm,
Parameter Changed, DST, Low / Missing Voltage, Battery
and Power Loss)
Instant
Energy is logged as consumption. Instantaneous values can be logged as average, maximum, minimum and instant value during the logging interval or as the instantaneous va lue at the end of the logging interval. Maximums and minimums are detected based on 1 sec. interval snapshots, and average is calculated based on 1 sec. interval snapshots.
Notes:
Maximum 50 parameters can be selected for each logger.
Pulse input must be configured through ConfigView.
7.7.2 Logging interval and total channels
A logger can store data in 1 to 50 channels. The logging interval is common for all channels in a logger and it can be configured from one minute up to one hour. A logger’s capacity is dependent on number of channels and logging interval. For example, Prometer 100 meters can be configured to store 960 days of load profile data at 30 minutes SIP for 10 parameters. When the logger is full, the oldest values will be written over. The table shows the capacity in number of days before the oldest value is written over.
Prometer 100 User Manual
BGX501-943-R01
Capacity in days
Number of logging channels
Logging
interval (min)
1
5
10
20
30
40
50
1
320
64
32
16
10 8 6
2
640
128
64
32
21
16
12 3 960
192
96
48
32
24
19 4 1000
256
128
64
42
32
25 5 1000
320
160
80
53
40
32
10
1000
640
320
160
106
80
64
15
1000
960
480
240
160
120
96
20
1000
1000
640
320
213
160
128
30
1000
1000
960
480
320
240
192
60
1000
1000
1000
960
640
480
384
Notes:
Survey Integration Period (SIP) for Instantaneous parameters can be configured as 1, 2, 3, 4, 5, 10, 15, 20, 30 or 60 minutes.
Survey Integration Period (SIP) for Energy parameters can be configured as 5, 10, 15, 20, 30 or 60 minutes.
Maximum 1000 days can be configured
7.7.3 Storage of logged values
Logged values are saved with time stamps and flags that indicate events that have occurred during the logging interval.
The time stamp indicates the end-time. If the logging interval is configured to one hour, a value with the time stamp 15:00 refers to the period 14:00 to 15:00.
To indicate events or states during an interval, a logged value can be stored with one or more flags.
Event or state
Name of flag
Explanation
Time adjusted
T
During the past interval, the meter clock has been adjusted either instantaneous or a sliding adjustment is in progress.
Time Disturbed D
The past interval is incomplete. For example, an interval shortened by the meter being without auxiliary power or if the logging memory has been reset. The first value after the logging memory having been configured will thus always be indicated with "Faulty value” (the logging memory is reset in conjunction with reconfiguration).
Alarm
A
In conjunction with user-d efi ned alarms being configured, it may be specified that
an alarm will also be indicated with logged values. When a user-defined alarm has triggered during the past interval, this is indicated with the flag “Alarm”.
Parameter changed
P
The Prometer 100’s configuration, calibration or initiation has changed. Which of these three the flag refers to can be seen in the event log.
Daylight savings time
S Daylight sav ing time has been in effect during the past interval.
Voltage
loss/missing
V
During the past interval all measuring voltages have been lost or missing.
Field
Replaceable Battery (for meter reading/ RTC backup)
B
Estimated battery lifetime is up.
Power loss
O
During the past interval, the meter has been without auxiliary power.
BGX501-943-R01 Prometer 100 User Manual Page 45 of 72
For several of the flags, additional information can be viewed in the event log. A more exact time for events is specified in the log.
Note: When power is calculated from logged energy values, the resulting values will be somewhat more precise than when power is logged directly. This is because power is logged as an average value of instant values. The instant values are read twice per second, while energy is accumulated continuously.
7.8 Alarms
The Prometer 100 is equipped with alarms to be able to indicate when measured quantities are over or under a factory configured limit value. The meter enters the alarm state when the limit value is reached. An alarm is generated only after the alarm state has continued for a configurable time (persistence time). Alarms are configurable by the user and are therefore called user-defined alarms. The persistence time can be configured to a maximum of 60 minutes.
Note: There are alarms and events that cannot be configured, but instead, are always active. Examples of such alarms are indication that the clock has been changed or that an auxiliary power loss has occurred. For more information, see the section Display of events (pg. 46).
7.8.1 Overview
For most user-defined alarms, the limit value is specified as a percentage of the nominal value, which is the configured, nominal primary value (current, voltage or power). For 3-element meters, the limit value corresponds to phase voltage, and for 2-element meters, phase to phase voltage. The following table provides an overview of available alarms.
Alarm
Description
Low voltage
Alarm can be configured for low voltage event. Occurred condition is treated as alarm condition.
High voltage
Alarm can be configured for high voltage event. Occurred condition is treated as alarm condition.
Voltage unbalance
Alarm can be configured for voltage unbalance event. Occurred condition is treated as alarm condition.
Current unbalance
Alarm can be configured for current unbalance event. Occurred condition is treated as alarm condition.
High THD voltage
Alarm can be configured for voltage THD event. Occurred condition is treated as alarm condition.
High THD current
Alarm can be configured for current THD event. Occurred condition is treat ed as
alarm condition. High THD power Average value of THD for all power phases exceeds the limit value. Any phase voltage
missing
Alarm can be configured for PT miss event (any phase). Occurred condition (any
phase) is treated as alarm condition. Frequency healthy Frequency is (< 49 Hz or >= 51 Hz) or (<59 Hz or >= 61 Hz) Auxiliary power
supply fail
Any auxiliary supply fails
Low power factor
System power factor is below limit value.
Low active power System active power is below limit value. High active power System active power is above limit value.
Single harmonic high voltage
Any phase voltage individual harmonics value (2
nd
to 31st) is above limit value.
Every 5 second meter will scan 2
nd
to 31st harmonic values for one phase and in next 5 sec meter will scan for next phase. Hence resolution of checking for each phase is 15 second.
Prometer 100 User Manual
BGX501-943-R01
Single harmonic high current
Any phase current individual harmonics value (2
nd
to 31st) is above limit value.
Every 5 second meter will scan 2
nd
to 31st harmonic values for one phase and in next 5 sec meter will scan for next phase. Hence resolution of checking for each phase is 15 second.
Internal error / Health Status
In case of internal failure like RTC fail (as per event logged) , Memory fail (as per event logged), Battery fail or System status.
Internal RTC battery fail
In case of internal RTC battery fail (Simulated battery failure alarm shall be provided)
Phase wise low voltage
Phase voltages is beneath the nom inal voltag e - limit value
Phase wise high voltage
Phase voltages is above the nominal voltage - limit value
Phase wise low power factor
Phase wise power factor is below limit value.
Reverse energy
direction
Alarm can be configured for CT reverse event (any phase). Occurred condition
(any phase) is treated as alarm condition.
Note:
Two LEDs are available for alarm information.
Multiple alarms can be selected on single LED.
Events selected only for alarm, shall not log event. For logging purpose ev ent has to be selec t ed in
event log section.
7.8.2 Indication
Alarms can be configured to indicate in one or more of the following ways:
Alarm LED on meter front
Changed digital output level
Indication of a logged value with a flag
The alarm LED stops flashing and the digital output returns to inactive low after the alarm state passes.
7.8.3 Display of events
The events status can be viewed on the meter’s display if the tamper information displays are configured to be included in one of the meter’s display sequences. A more detailed description of the meter's events can be viewed in M-Cubed. Via M-Cubed, the event log can also be printed out or saved to a file.
Sr.No Fixed Compartment Events
Maximum
number of events per
compartment
Snapshot
1
1
Phase wise PT Miss
100
Y
2
Over Voltage
3
Under Voltage
4
Voltage Unbalance
5
2
Phase wise CT Reversal
100 Y
6
Phase wise CT Open (HT Meter Only)
7
CT Bypass (HT Open Only)
8 Current Unbalance 9
3
Power Fail / Power On-Off
100 N
BGX501-943-R01 Prometer 100 User Manual Page 47 of 72
10 4 Time Set Transactions
100 Y
11
Profile Capture Period Transaction for logger 1
12 Profile Capture Period
Transaction for logger2
13
5
Neutral Disturbance
100 Y
14
Magnet Interference
15 6 Cover Open (In case of cover
open during power fail duration, time stamp and snapshot will be logged at
power up time)
100 Y
16
7
Device ID Change
100 N
17
Password Change
18
AES Key Change
19
Immediate Tariff Download
20
Feeder Supply Fail
21
Tamper Reset
22
Scaling Tariff Download
22
Remote control IO switch transaction
23
Energy register reset transaction
23
MD reset transaction
24
8
CT Miss
100 Y
25
Invalid Phase Association
26
Invalid Voltage
27
9
RTC Fail
100 N
28
Memory Fail
29
10 Phase wise Voltage THD (%) 100 Y
30
11 Phase wise Current THD (%) 100 Y
For snapshot refer to Appendix G
7.9 Maximum demand
Maximum Demand (MD) plays a crucial role in current scenario of electricity conditions. According to maximum demand conditions it is easy to monitor variation in the load condition and trend of load according to the time zone.
The Prometer 100 meter has the capability of logging Maximum Demand for all the selected energy types (except Net Active and Net Reactive). The Maximum Demand is computed for a fixed block of time which is called Demand Integration Period (DIP). DIP can be set to 5, 10, 15, 20, 30 or 60 minutes.
Maximum Demand registers are provided for each individual energy type. A separate register is available to record the Maximum Demand during the entire day (i.e. 00-24 hours). This is known as the Universal Demand Register. This is not configurable through the tariff.
The rules for logging Maximum Demand in these individual register may be set on the following basis: As per Time Zone: In this case the individual MD registers are assigned to a specific time zone of a day. A
particular MD is active in the assigned Time Zone only. In such case the MD register are called TOD MD
register. MD reset button: User can trigger the Maximum Demand by use of MD reset button provided under the front
cover of the meter. Please note that the Maximum Demand in any MD register is for the current billing period and is always reset to
zero whenever a billing cycle is finished.
Prometer 100 User Manual
BGX501-943-R01
7.10 Historical registers
Historical registers are used by the Prometer 100 to store current register values at defined points so as to be able to read them later. Stored in historical registers are all maximum demand values, TOU registers and energy registers, with the exception of energy registers by phase. The historical registers are time stamped to indicate when storage occurred. The Prometer 100 can store up to 15 historical registers.
Date and time
Energy registers
Maximum demand values
Time of use registers
7.10.1.1 Finish historical period By finishing a historical period, the current registers values are stored in historical registers and the maximum
demand values are reset. When a period is finished, an event is stored in the meter’s event log. Periods can be finished in various ways:
Via meter push button
The period is finished when the meter’s MD Reset button is pressed. This requires that the historical registers are configured to permit finish via the meter button.
Via M-Cubed
The period is finished when a command is given from M-Cubed or third-party software.
At any configured billing date
The period is finished when the meter clock reaches the configured billing date
Via digital input
The period ends when a pulse is received at a digital input on the meter.
This requires both that the historical registers are configured to permit ending via a digital input and that an input is configured for this purpose.
7.10.1.2 Lock out time for finish historical period The Lock out time prevents the user to create a new historical period within a configured time. It can be
configured from 1 hour to 40 days.
7.11 Time of use
Time of use is a function that enables energy to be divided up into various registers depending on the rate that applied when the energy was measured. In the Prometer 100, tariff structure can be stored that switch rates at predetermined times according to a configurable pattern. A tariff structure consists of seasons, day types and holiday dates. The maximum number of rates is eight.
Day types specify how rates change during a 24-hour day.
Seasons specify the day types that apply during the days of the week, Monday to Sunday.
Holiday dates specify the day type that applies on a certain date.
Rate input can be used to control active rate depending on the state of digital input signals.
BGX501-943-R01 Prometer 100 User Manual Page 49 of 72
7.11.1 Day type
Prometer 100 supports up to 16 day types. A day type specifies which rate, from a maximum of eight, should apply when during the day. A day can have up to 16 changing points. A day type can be connected to a day of the week in a season or to a holiday date.
7.11.2 Season
A season refers to a period. During this period, the season defines the day types that will apply during the days of the week. Prometer 100 supports up to 16 separate seasons. The seasons are arranged in a sequence where one season replaces the previous at a predetermined date.
7.11.3 Holiday dates
Days that change rates according to a pattern that is not covered by linking day types to seasons are called holiday dates. A holiday date specifies the day type that applies on a certain date. Holiday dates can be configured to apply every year on the same date or for a single year. In the Prometer 100, up to 30 holiday dates can be configured.
7.11.4 TOU registers
An energy register that is chosen to be divided into the rates is a TOU register. There are 14 TOU registers and each TOU register has separate registers for eight rates.
7.12 Billing Cycle Support
The concept of energy metering essentially consists of billing cycles. Prometer 100 has an inbuilt support for billing cycle.
7.12.1 Billing Cycle
The Prometer 100 provides following ways to perform a billing action. Performing a billing action finish current billing cycle and starts a new billing cycle.
According to bi lling dat es specified in Tarif f file. A b illing acti on from a billing dat e is done at the start of the day on the nominated date file
At the time of Tariff activation i.e. whenever a new tariff is activated. (Downloading of new tar iff having change in energy channels, scaling and TOU)
By pres s ing MD Reset butt on
Authenticated command for MD Reset from M-Cubed.
Note: The lock out time prevents the user to perform a billing action within a configured time.
7.12.2 History of Energy, Rate and MD Register
Following values are stored in the meter memory each time a billing cycle completes i.e. a billing action is performed.
1. Values of all main Energy Registers at the time of a billing action. This is maintained in a rollover fashion which may have a maximum of 15 histories .i.e. all last 15 historical values will be available to you.
2. Values of all Rate Registers. This is maintained in a rollover fashion which may have a maximum 15 histories .i.e. all las t 15 his t or ical values will be available to you.
3. Values for all MD Registers along with Date & Time of MD occurrence. This is also maintained in a rollover fashion which may have a maximum 15 histories .i.e. all last 15 historical values will be available to you.
7.12.3 History for the Cause of Billing Register
The cause of billing is available on display in the meter for last 15 billing actions along with date & time of billing.
Prometer 100 User Manual
BGX501-943-R01
7.12.4 Cumulative Maximum Demand Registers
A special register is also provided for energy type selected for maximum demand which stores cumulative Maximum Demand for all MD register whenever a billing action takes place. Each time Billing Action takes place, the cumulative register shall increase by the value of their source register.
The purpose of this register is to limit the scope of tampering with Max imum Demand by performing a billing action repeatedly.
7.13 Meter Reading
The Prometer 100 uses DLMS for meter reading. In DLMS protocol the BCS is considered the client and the meter is the server. On request by the BCS, the meter will send all its supported OBIS codes and expected queries. All DLMS meters will communicate data only after ‘getting associated’ with the BCS client. Meter reading is divided in to the following sessions:
Single parameter sessions Profile sessions
1. Instantaneous profile reading
2. Midnight energy profile reading
3. Load survey profile reading
4. Log wise events profile reading
7.14 Scaling Tariff
The Prometer 100 Scaling Tariff supports CT / VT ratio adjustment, and CT-VT error compensation in the field. The M-Cubed is used for preparing the scaling tariff. The existing meter configuration must first be opened and
read. The configuration can then be modified using the M-Cubed. The Prometer 100 does not require to be switched off during the configuration mode and will only implement the new configuration when all the blocks have been received and authenticated.
The Scaling Tariff contains the following sections:
1. Commissioning
2. Error Compensation
In a single scaling tariff, commissioning change and error compensation is treated & logged as three separate transactions.
It is also possible to download a scaling tariff in meter using a suitable tool. Complete meter data will be washed out by applying scaling tariff in meter.
7.15 Transformer compensation
Transformer compensation is a function for compensating for measurement errors in instrument transformers and for losses in power transformers. The function enables the Prometer 100 to present measurement values for which errors and losses have been compensated. The formulas used in the meter are presented in Appendix E: Calculation Principles (pg. 61).
Changing the transformer compensation can be blocked by the security setting ‘Block configuration of transformer compensation’. This setting can only be changed at access level 5.
7.15.1 Overview
The tables present an overview of the transformer compensations in Prometer 100.
Instrument transformer compensations
Name
Value to be entered
Voltage error L1, L2, L3 Amplitude error as percent
Phase angle in minutes
Current error L1, L2, L3 Amplitude error as percent
Phase angle in minutes
BGX501-943-R01 Prometer 100 User Manual Page 51 of 72
Power transformer compensation s
Name
Value to be entered
Copper losses, Total values Active loss as percent of nominal power
Reactive loss as percent of nominal power
Iron losses, Total values Active loss as percent of nominal power
Reactive loss as percent of nominal power
7.15.2 Instrument transformer compensations
To compensate for errors in instrument transformers, their amplitude error in percent and phase angle error in
minutes are configured in the meter. One minute is equal to the angle 1˚/60. The errors can be specified
separately for all voltages and currents. When instrument transformer compensations are used, current and voltage are affected, as well as all quantities t hat aris e f rom these: power, ener gy, etc.
7.15.2.1 2-element meter When voltage errors are compensated on a 2-element meter, this is done on phase to phase voltages L12, L23
and L31 instead of on phase voltages. Only L12 and L23 are included in the calcu lat ion of power and energy. Compensation of L31 has no significance in this respect. In compensation of current errors for 2-element meters, only L1 and L3 can be compensated for because they are the only currents measured.
7.15.3 Power transformer losses
Power transformer losses consist of copper and iron losses. They are expressed as percentages of nominal power. One value is specified for active loss and one for reactive. When compensation of losses is configured, power, energy, power factor, etc. are affected but not current and voltage.
When copper losses are added per phase, the resulting copper loss is the average of the value.
Calculating loss values
Based on the nominal total power and the measured loss value in watts, a loss value can be calculated as a percentage of nominal power. It is the loss value that is configured in the meter. Nominal power is calculated with configured nominal current and voltage.
Nominal power:
3=
nomnomnom
voltageLineCurrentPower
Loss value:
100/ =
nom
PowerLossvalueLoss
7.16 Quality of Supply
Power quality encompasses voltage monitoring and harmonics measurement. Harmonics measurement is described in the section Instant values (pg. 37).
7.16.1 Voltage monitoring
Voltage monitoring monitors the following states: swells (overvoltage), sags (under voltage), unbalance and interrupts. Monitoring is enabled by checking the check box and configuring limit values for the events. The occurrence and restoration limits are expressed as percentages of configured nominal voltage.
Example: For an occurrence limit of 110% and a restoration limit of 90%, and the c onfig ured nom in al prim ar y voltage of 10 kV, the limits attained are 9 kV and 11 kV.
These events are monitored phase wise every second. For 3-element meters, the average value for phase voltage is monitored, and for 2-element meters, the average value for phase to phase voltage is monitored.
The states shorter than three seconds are registered by accumulating registers. If the states last longer than three seconds, they are instead registered in the event log with time stamp and duration. The accumulating counters and the event log can be read in M-Cubed.
Prometer 100 User Manual
BGX501-943-R01
State Duration Registered in Duration presented as
Sags / Swells / Interrupt / Unbalance
1 s – 3 s
Accumulating counter
Duration not specified
3 s – 60 s
Event log
Duration of state
> 60 s
Event log as
Low voltage (for Sag) High voltage ( for Swell) Feeder fail (for Interrupt) Voltage unbalance (for unbalance)
Duration of state
Note: If voltage interrupt condition persists, then voltage sag and voltage unbalance conditions are not monitored.
Appendix A: Abbreviations
The following are commonly occurring abbreviations used throughout this manual. APS: Auxiliary Power Supply
BCS: Base Computer Software
DLMS: Device Language Message Specification
SCADA: Supervisory Control and Data Acquisition MD: Maximum Demand UMD: Universal Maximum Demand (0 to 24 hours)
Appendix B: Material Declaration
The material declaration for the Prometer 100-R is shown below:
Enclosure Rack Mild steel with Aluminium Meter Enclosure Mild steel Meter back plate Mild steel Meter current and voltage terminals Brass with gold and silver plating Handle Mild steel Screws M3 for sealing meter with rack and M4 elsewhere Front Cover PC (Polycarbonate) Front hinged plate PC (Polycarbonate)
*Rack carries the female part of Essailec connector and meter carries the male part of Essailec connector with all the back side communication ports in the form of RJ45 connector.
BGX501-943-R01 Prometer 100 User Manual Page 53 of 72
The material declaration for Prometer 100-W is shown below:
Enclosure PC FR (Polycarbonate Fire retardant) Meter current and voltage terminals Brass with nickel plating Sealing screws Brass with nickel plating
Appendix C: Communication Port s
The Prometer 100 can come as fitted with the following ports (see the variant supplied as per order):
1. Optical 1107 port
2. RS232/RS485 – Left Module
3. RS232/RS485 – Right Mod ule
4. Ethernet
Optical Communication Port
In Prometer 100-R, the optical 1107 port is protected by a sliding cover. The 1107 optical port cover can be slid upwards in the arrow direction to the open position. The cover has a captive design and cannot be removed and lost. There is an optional sealing point. While in Prometer 100-W, the optical 1107 port is protected by a rotational seal.
Figure 24: Optical 1107 Port in Prometer 100-R
Figure 25: Optical 1107 Port in Prometer 100-W
Cover Closed
Cover Open
Sealing Points
Prometer 100 User Manual
BGX501-943-R01
Communication through the 1107 infrared optical communications port is performed through PC with suitable communication cable. Care has to be taken to align the opto head so the IR transmitters and receivers are in the best position to exchange signals. Usually this means attaching the opto head with the connecting wire hanging straight down. Ambient visible light (daylight, incandescent or in particular fluorescent light) may disturb the exchange of IR signals if strong enough and cause the communication to fail. To ensure reliable communications, avoid strong ambient light positions when installing the Prometer 100.
RS232 Serial Communication Port The RS232 communication port is used for connecting to communication equipment such as computers and
modems through a serial cable. The meter side connector is an RJ-45 type. The user should ensure that the other side is fitted with a suitable connector before connecting an external device with the meter. See Figure 26 for pin description.
RS485 Serial Communication Port The RS485 serial half duplex communication ports are intended to be used to connect the Prometer 100 to a
network for multi-drop communications. The RS485 network is looped through the input connector to the output connector. See Figure 4 for pin description. If for example three meters needed to be “daisy chained”, then simply connect the “RS485 Out” on the first meter to the “RS485 In” on the next meter and so on. The pin connection for both ports is identical, so the same type of cable can be used through-out and is not polarity conscious.
Ethernet Port The Ethernet serial communication port is used for connecting directly to an internal Ethernet network. Each
meter needs to be provided with an IP addressed static to internal LAN for communicating the data over TCP/IP network. See Figure 26 for pin description.
Figure 26: Pin Description for RJ-45 Ports
Note- Recommended cable to use with all the above mentioned ports (RJ-45) is CAT6 type and shall be crimped with standard tools used in LAN connection to PC/ Laptop. The other end of the cable should be as desired by customer for its intended application like connecting to Modem or PC or LAN switch etc. In general, all the communication ports in meter are optically isolated with each other and can be configured by M-Cubed for data settings, TCP/IP settings etc. as desired by user. The Ethernet cable is normally connected between the meter and the main switch, although local IP installations may differ.
BGX501-943-R01 Prometer 100 User Manual Page 55 of 72
The Ethernet port has a 1-minute inactivity time-out period. If no activity is detected the Ethernet port will disconnect after 1 minute.
Appendix D: How to Read Meters through Ethernet Port
This section discusses the reading a Prometer 100 meter through Ethernet port (TCP/IP connection). Prerequisites:
Take static IP address, Subnet Mask and Gateway from network M-Cubed (6.0.0.6) Install the M-cubed on PC Connect meter with PC using optical port or RS232 port
Configuring the static IP address, Subnet Mask and Gateway
Connect Prometer 100 meter with PC on optical port
Select connection media, protocol, comport and baud rate. Click OK.
Prometer 100 User Manual
BGX501-943-R01
Check successfully connected
Click Transaction and then click TCP/IP Configuration
Enter the IP address, Subnet Mask and Gate wa y information to be configured in meter. Click OK
BGX501-943-R01 Prometer 100 User Manual Page 57 of 72
IP configured successfully.
You can verify on the meter display. Go to the Fixed Display Page
Scroll down to view the Ethernet configuration inform ation
Reading the Prometer 100 Meter on the Ethernet port
Connect the meter with the LAN network using the LAN cable. Confirm that the LAN cable is firmly connected to both the meter and PC Ethernet ports .
Prometer 100 User Manual
BGX501-943-R01
Click Connect and then select connection media as TCP/IP, communication type as DLMS. Select the checkbox “Use TCP profile over Ether net port .”
Enter the meter static IP information in Con nec ti on Ad dres s Click OK.
Static IP is 172.16.13.5
Service port is 4059. It is fixed for DLMS
BGX501-943-R01 Prometer 100 User Manual Page 59 of 72
Meter connected successfully with the M-Cubed.
Prometer 100 User Manual
BGX501-943-R01
To read data, click Get All Data
Action completed successfully will be displayed.
BGX501-943-R01 Prometer 100 User Manual Page 61 of 72
Appendix E: Calculation Principles
(to be updated)
Current and Voltage
Current and voltage are first harmonic component (fundamental).
Calculated phase to phase voltage
Phase to phase voltage is calculated from the phase voltages’ fundamental vectors. This calculation is only made in the 3-element meter; in the 2-element meter, the phase to phase voltage is measured.
)21cos(21221(12
1111
2
1
2
11
pUpUUUUUU +=
Calculated I2
In the 2-element meter, I2 is not measured but is calculated for monitoring purposes. It is calculated from the current’s fundamental vectors.
1
2I
Current I2’s fundamental harmonic.
2
11
2
11111
))31sin(3())31cos(31(2 pIpIIpIpIIII ++=
Voltage unbalance
Class A The basic measurement of voltage harmonics, for class A, is defined in IEC 61000-4-7 class I. That standard
shall be used to determine a 10/12-cycle gapless harmonic subgroup measurement, denoted U
isg,h
in IEC
61000-4-7.
Power
Harmonic component power
The calculations below are for active power, the calculations for reactive are identical except for that cos­functions are replaced with sin-functions.
n
P1
Active power in L1 is calculated for harmonic component n.
n
P
Total active power is calculated for harmonic component n.
n
ϕ
Phase angle between harmonic component
n
U1
and
n
I1
3-element meter:
)cos(111
nnnn
IUP
ϕ
=
2-element meter: For 2-element meters, only the total power is calculated in each harmonic component.
n
1
ϕ
Phase angle between harmonic component
n
U12
and
n
I1
n
2
ϕ
Phase angle between harmonic component
n
U32
and
n
I3
)2cos(332)1cos(112
nnnnnnn
IUIUP
ϕϕ
+=
Prometer 100 User Manual
BGX501-943-R01
Active and reactive power
Active and reactive power is calculated as the sum of harmonic component power up to 31st harmonic. The calculation is made with plus and minus signs, where negative power represents export direction and positive represents import direction.
P
Total active power
1P
Active power in L1
Q
Total reactive power
1Q
Reactive power in L1
321 PPPP ++=
321 QQQQ ++=
For 2-element meters, two elements are added instead of three.
Apparent power
S
Total apparent power
1S
Apparent power in L1
22
QPS +=
22
111 QPS +=
Energy
Energy is calculated by integrating power (P, Q and S) over time.
Definition of quadrants
The term phase angle is described under its own heading below. Quadrant I: phase angle 1–90° Quadrant II: phase angle 90–180° Quadrant III: phase angle -180–(-90)° Quadrant IV: phase angle (-90)–0°
Active energy
Active energy is calculated for import and export. The direction is controlled by the sign for active power (+ import, – export).
Active energy import: quadrant I and IV Active energy export: quadrant II and III
Reactive energy
Reactive energy is calculated for four quadrants. The quadrant is controlled by the sign for active and reactive power (e.g., active power >= 0 and reactive power >= 0 corresponding to quadrant I).
Reactive energy import: quadrant I and II Active energy export: quadrant III and IV
Reactive energy inductive: quadrant I and III
BGX501-943-R01 Prometer 100 User Manual Page 63 of 72
Reactive energy capacitive: quadrant II and IV
Apparent energy
Apparent energy is calculated for import and export. The direction is controlled by the sign for active power; apparent energy is registered for the direction that the active energy has during the same period.
Apparent energy import: quadrant I and IV Apparent energy export: quadrant II and III
Power Factor
1/1)1( SPLpf =
SPTotpf /)( =
The power factor is calculated without signs and is thus always positive.
Phase angle
pIpUL 11)1( =
ϕ
Phase angle for an element is calculated from the fundamental phase angles.
)1/
1arctan()(
fundfund
QPTot =
ϕ
Total phase angle is calculated from fundamental power. Phase angle values specified between –180 ° and 180°.
THD
Total harmonic distortion
%100
...
...
2
1
2
2
2
1
22
3
2
2
++
++
=
n
n
eur
III
III
THD
Where
1
I
n
I
are the current’s harmonic components. The calculation is made in the same ways for current
and voltage.
Prometer 100 User Manual
BGX501-943-R01
Transformer Compensations
Instrument transformer compensations
Instrument transformer compensations are specified as error in percent for amplitude, and as error in minutes (one-sixtieth of a degree) for phase angle. These errors can be specified separately for all voltages and currents.
AmpErr
Amplitude error as percent
PhaseErr
Phase angle error in minutes
meas
u
Measured voltage
u
Compensated voltage
Amplitude compensation
)1/( AmpErruu
meas
+=
Phase angle compensation
PhaseErrupup
meas
=
In 3-element meters, phase voltages are compensated and in 2-element meters, phase to phase voltages. The same calculation is used for all currents and voltages.
Power transformer losses
Power transformer losses consist of copper and iron losses. They are expressed as percentages of nominal power. One value is specified for active loss and one for reactive.
Iron loss
FeLoss
Active iron loss as percent of nominal power.
meas
P1
Measured power.
1P
Compensated power.
Nom
Nominal power per phase
NomFeLossPP
meas
+= 11
Copper loss
CuLoss
Active copper loss as percent of nominal power.
i
Phase current.
nom
i
Nominal current.
NomCuLossi
iPP
nommeas
+=
2
)/(11
Corresponding calculations made for active and reactive power.
Definition of phase order
Correct phase order (123) corresponds to phase position:
pU
1
Phase position for U1
BGX501-943-R01 Prometer 100 User Manual Page 65 of 72
°= 01pU
°= 1202pU
°= 1203pU
The same system is represented graphically below. The vectors rotate in an anti-clockwise direction.
Prometer 100 User Manual
BGX501-943-R01
Appendix F: Connection and General De t ai ls
Electrical
Connection type
Common product for HV3/HV4/LV4 applications
Wiring configuration
3-phase 3-wire, 3-phase 4-wire
Voltage range (L-N/L-L)
57.7/100 V to 69.3/120 V (configurable) for HV3 & HV4
240/415 V for LV4 (applicable for Prometer 100-W only)
Current range
I
b
: 1-5 A (configurable)
I
max
: maximum 10 A (configurable)
Accuracy
Class 0.2s and 0.5s
Mains frequency
50/60 Hz ± 5%
Burden
If power drawn from Aux supply – <0.1 VA /phase (for voltage and current circuit) <10 VA (burden on Aux supply) If power drawn from VT (i.e. self powered) – <0.1 VA /phase (for voltage and current circuit)
<6 VA /phase (burden on VT supply)
STOC
10 times I
max
for 1 second
Compliance
Standards
IEC62052-11, IEC62053-22, IEC62053-23, IEC62056-52, IEC61010-1,
IEC61010-2-030, CE, MID (EN50470-1, EN50470-3)
Mechanical
Prometer 100-R
Dimensions (L X W X D)
428 x 133 x 260 mm approx. (meter with 19” rack) 299 x 133 x 260 mm approx. (meter with 11” rack)
Enclosure
Translucent polycarbonate cover (with clear transparent window for display) and overall mild steel body
Sealing
Sealable screws on the front and back fascia of meter, sealing provision for optical port and MD reset button
Weight
Meter – 3.8 kg approx. (± 1 kg) 11” rack – 2.1 kg (± 0.1 kg)
19” rack – 3.3 kg (± 0.1 kg)
Ingress Protection (IP) IP53 on front fascia and IP20 on back side
Prometer 100-W
Dimensions (L X W X D)
300 x 200 x 100 mm (± 5 mm)
Enclosure
Plastic material (type Polycarbonate)
Sealing
Sealable base and cover
Weight
2.5 kg approx. (± 0.5 kg)
Ingress Protection (IP) IP53 and IP54 with panel mounting kit over the front fascia
BGX501-943-R01 Prometer 100 User Manual Page 67 of 72
Environmental
Temperature
-10
0
C to + 60 0C (operating)
-25
0
C to + 70 0C (limit range of operation)
-40
0
C to + 80 0C (storage)
-20 0C to + 70 0C (display operating range)
Humidity
95% non-condensing
Pollution degree
Type 2
Over voltage category
III
Prometer 100 User Manual
BGX501-943-R01
Appendix G: List of DLMS Parameters
List of Profile Parameters (DLMS)
Section : Event Log 1 to 9 Snapshot Profile
Sr.No
Profile
Parameter
1 Instantaneous
Real Time Clock – Date and Time
2
Instantaneous
Cumulative Energy – Active Import
3
Instantaneous
Cumulative Energy – Active Export
4 Instantaneous Voltage – VRN for 3Ф4W / VRY for 3Ф3W
5
Instantaneous
Voltage – VYN for 3Ф4W
6
Instantaneous
Voltage – VBN for 3Ф4W / VBY for 3Ф3W
7
Instantaneous
Line Current – IR
8 Instantaneous Line Current – IY
9
Instantaneous
Line Current – IB
10
Instantaneous
Active Current – IR
11
Instantaneous
Active Current – IY
12 Instantaneous Active Current – IB
13
Instantaneous
Power Factor – R phase
14
Instantaneous
Power Factor – Y phase
15
Instantaneous
Power Factor – B phase
16 Instantaneous Voltage angle – Angle between R and Y phase
17
Instantaneous
Voltage angle – Angle between Y and B phase
18
Instantaneous
Voltage angle – Angle between B and R phase
List of Profile Parameters (DLMS)
Section : Event Log 10 Snapshot Profi le
Sr.No Profile Parameter
1 Instantaneous
Real Time Clock – Date and Time
2
Instantaneous
Cumulative Energy – Active Import
3 Instantaneous Cumulative Energy – Active Export
4
Instantaneous
Voltage THD % - Phase 1
5
Instantaneous
3rd Harmonic Voltage - Phase 1
6
Instantaneous
5th Harmonic Voltage - Phase 1
7 Instantaneous 7th Harmonic Voltage - Phase 1
8
Instantaneous
9th Harmonic Voltage - Phase 1
9
Instantaneous
Voltage THD % - Phase 2
10
Instantaneous
3rd Harmonic Voltage - Phase 2
11 Instantaneous 5th Harmonic Voltage - Phase 2
12
Instantaneous
7th Harmonic Voltage - Phase 2
13
Instantaneous
9th Harmonic Voltage - Phase 2
14
Instantaneous
Voltage THD % - Phase 3
15 Instantaneous 3rd Harmonic Voltage - Phase 3
16
Instantaneous
5th Harmonic Voltage - Phase 3
17
Instantaneous
7th Harmonic Voltage - Phase 3
18
Instantaneous
9th Harmonic Voltage - Phase 3
BGX501-943-R01 Prometer 100 User Manual Page 69 of 72
List of Profile Parameters (DLMS)
Section : Event Log 11 Snapshot Profi le
Sr.No Profile Parameter
1 Instantaneous
Real Time Clock – Date and Time
2
Instantaneous
Cumulative Energy – Active Import
3
Instantaneous
Cumulative Energy – Active Export
4
Instantaneous
Current THD % - Phase 1
5
Instantaneous
3rd Harmonic Current - Phase 1
6
Instantaneous
5th Harmonic Current - Phase 1
7
Instantaneous
7th Harmonic Current - Phase 1
8
Instantaneous
9th Harmonic Current - Phase 1
9
Instantaneous
Current THD % - Phase 2
10 Instantaneous 3rd Harmonic Current - Phase 2
11
Instantaneous
5th Harmonic Current - Phase 2
12
Instantaneous
7th Harmonic Current - Phase 2
13
Instantaneous
9th Harmonic Current - Phase 2
14
Instantaneous
Current THD % - Phase 3
15
Instantaneous
3rd Harmonic Current - Phase 3
16 Instantaneous 5th Harmonic Current - Phase 3
17
Instantaneous
7th Harmonic Current - Phase 3
18
Instantaneous
9th Harmonic Current - Phase 3
Prometer 100 User Manual
BGX501-943-R01
Frequently Asked Questions (FAQs)
The message ‘Not configured’ is shown on the Prometer 100 display.
The message ‘Not configured’ is displayed when the parameters are not configured in the meter. For e.g, Bar Graph of Energy parameter for the logger will be available on meter display if user configures
logger energy in meter. Similarly Rate Registers and Demand Registers will be available if user configures TOU energy in meter.
Is it possible to have different IDs for the two RS485 communication modules attached to the meter?
For DLMS, different IDs can be used for both the RS485 communication modules. While for Modbus, same IDs will be used for both the RS485 communication modules.
Are the harmonics data up to 31
st
order available on meter display?
The voltage and current harmonics data up to 31
st
order is not available on meter display and can be accessed
using M-Cubed 100 and via the communication protocol.
Which information is not available on the Prometer 100’s display for 3-phase 3-wire?
On configuring Phase 2 displays for 3-phas e 3-wire, the Phase 2 parameters such as energy, voltage, current, Power, P.F, harmonics, L1-L2 voltage phase angle, L2-L3 voltage phase angle and current symmetry data will be displayed as “……..”.
Which events are not applicable for 3-phase 3-wire?
Phase 2 current reversal, Phase 2 current miss, CT open, CT bypass, neutral disturbance and invalid phase association are not applicable for 3-phase 3-wire.
Will the existing meter data reset on changing configuration?
The existing meter data will be reset on configuring/changing existing energy channels, logger parameters or scaling information in meter.
What happens if a new successive adjustment is made when one is already underway?
The current adjustment will be stopped and the new begun.
When summer time starts, the clock jumps one hour. What happens if the user sets the meter to a DST time within that hour?
The meter will ignore the new time because that hour “does not exist”.
Is logger’s data affected by daylight saving time change? What data is affected by daylight saving time change?
No, the logger’s data will not be affected by the daylight saving time change as the loggers, events and daily energy snapshot are logged as per base time. The daylight saving time is applicable only for TOU rate registers, TOU demand data and billing data.
BGX501-943-R01 Prometer 100 User Manual Page 71 of 72
Can user set meter clock via Modbus communication protocol?
Time set transaction as per base time or DST time will be provided via DLMS communication protocol. Only time synchronization as per base time or DST time will be provided via Modbus com munication protocol.
Meter will reject the time sync request if it is above time adjustment limit.
What is the maximum time adjustment limit and how it works ?
Time synchronization as per base time or DST time will be provided via Modbus communication protocol only. Meter will not sync the time if the time difference is more than the time adjustment limit configured in the meter.
By default, the time adjustment limit is set as 25 seconds. The time adjustment limit can be configured as any value between 0 to 30 seconds using ConfigView.
So, if user wants to change the meter time out of time adjustment limit then time set transaction has to be performed via DLMS communication protocol.
How the logged values affected by the daylight saving time change?
Logged values during daylight saving time are labelled with the flag “Yes”. “Yes” means that the DST offset is applied in the meter.
What are the possible consequences of adjusting forward the meter clock?
If the clock is adjusted forward over one or more interval limits, there will be empty spaces filled by zero value in the logging memory.
What are the possible consequences of adjusting backward the meter clock?
It the clock is adjusted backward over one or more interval limits, one or more values will exist with the duplicate day and same time stamps. Empty spaces will be filled by zero value in duplicate day. In the worst event, this can result in data read from the meter not matching the expected period. Previous integration period and rising demand data will be reset.
What are the possible consequences of adjusting backward or forward the meter clock within time adjustment limit?
Integration period will be compressed for forward time sync and stretched for backward time sync. Data will be logged as per integration period.
Cewe Instrument AB
Box 11006,
SE 611 29 Nyköping
Sweden
t: +46 155 775 00
f: +46 155 775 97
UK contact details
t: +44 (0) 1962 840048
f: +44 (0) 1962 841046
Loading...