SD Acquisition, Inc., DBA CETAC Technologies (“CETAC”) warrants
any CETAC unit manufactured or supplied by CETAC for a period of
twenty five (25) months from the date of shipment. Units found in the
reasonable judgment of CETAC to be defective in material or
workmanship will be repaired or replaced by CETAC without charge for
parts and labor. CETAC reserves the right to change or improve the
design of any unit without assuming any obligation to modify any unit
previously manufactured.
This warranty does not cover any unit that has been subject to misuse,
neglect, negligence, or accident. The warranty does not apply to any
damage to the unit that is the result of improper installation or
maintenance, or to any unit that has been operated or maintained in
any way contrary to the instructions specified in the CETAC instruction
and operation manual. Operation of the CETAC unit inside a
laboratory fume hood is contra-indicated and will void the warranty.
Any attempt to repair or alter any CETAC unit by anyone other than by
CETAC authorized personnel or agents will void this warranty. If any
non-CETAC component is installed in the CETAC manufactured unit
without the approval of CETAC, the warranty will be voided. In
addition, this warranty does not extend to repairs made necessary by
the use of parts, accessories or fluids which are either incompatible with
the unit or adversely affect its operation, performance or durability.
CETAC’s obligation under this warranty is strictly and exclusively
limited to repair or replacement of defective CETAC parts and no claim
of breach of warranty shall be cause for cancellation or rescission of the
contract of sale of any unit.
The foregoing express warranty is in lieu of all other warranties,
expressed or implied, including warranties of merchantability and
fitness for a particular purpose. CETAC shall not be bound by any
representations or statements on the part of its employees or agents
whether oral or in writing and including any made in catalogues and
other promotional material including technical details and
specifications except where such representations and statements are
expressly made part of this contract. CETAC assumes no responsibility
for incidental, consequential or other damages, even if advised of such a
possibility, including but not limited to loss or damage of property, loss
of revenue, loss of use of the unit, loss of time, or inconvenience.
CETAC’s liability on any claim for loss or damage arising out of the
sale, resale or use of any of its products shall in no event exceed the
selling price of the unit.
Purch aser shall indemnify CETAC against any c laim or liability wh ich
may be asserted as relates to the following: (i) the use to which any
product supplied hereunder is put infringes the patent, copyright or
other intellec tual proper ty rig hts of any third party ; or (ii) an y liability
resulting from the failure by Purchaser to observe the terms of this
Warranty.
Retu rned Product Procedures
Claims for shipmen t damage (eviden t or conc ealed) must be filed with
the carr ier by the buyer . CETAC must be notified within nin ety (90)
days of shipment of incorrect materials. No product may be returned,
whether in warranty or out of warranty, without first obtaining
approval from CETAC. No replacements will be provided nor repairs
made for products returned without such approval. Any returned
product mu st be accompanied by a return auth orization number . The
expense of r eturning the unit to CET AC for servic e will be paid by the
buyer. The status of an y produc t return ed later than th irty (30) days
after issuance of a return authorization number will be subject to
review. Shipment of repaired products will generally be made forty
eight (48) hours after the receipt.
Products may not be returned which are contaminated by radioactive
materials, infectious agents, or other materials constituting health
hazards to CETAC employees.
Returned Product Warranty Determi n ation
After CETAC’s examination, warranty or out of warranty status will be
determined. If a w arranted defect ex ists, the product w ill be repaire d
at no charge and shipped prepaid back to the buyer. If the buyer
desires an air freight return, the product will be shipped collect.
Warranty repairs do not extend the original warranty period.
If an out of warranty defect exists, the buyer shall be notified of the
repair cost. At such time the buyer must issue a valid purchase order
to cover the cost of repair and freight, or authorize the products to be
shipped back as is, at the buye r’s e xpense . Failure to obtain a purchas e
order number approval within fifteen (15) days of notification will
result in the products being returned as is, at the buyers expense.
COPYRIGHT
Copyright 204-2012 CETAC Technologies
480123 Version 1.1, March, 2012
REPRODUCTION
All rights reserved. Reproduction or
transmission of this document in whole or
in part, and by any means without the
express written consent of the copyright
owner or authorized agent is prohibited.
Requests for additional copies of this, or
any other CETAC publication, can be
filled by contacting an authorized
distributor or
CETAC Technologies
Customer Service & Support
14306 Industrial Road
Omaha, Nebraska 68144, USA
Phone (800) 369-2822 (USA only)
Phone (402) 733-2829
Fax (402) 733-1932
E-mail custserv@cetac.com
DISCLOSURE
This document contains CETAC
proprietary data and is provided solely to
its customers for their express benefit of
safe, efficient operation and maintenance
of the product described herein. Use or
disclosure of CETAC proprietary data for
the purpose of manufacture or
reproduction of the item described herein,
or any similar item, is prohibited, and
delivery of this document shall not
constitute any license or implied
authorization to do so.
SAFETY
Instruments, accessories, components or
other associated materials may not be
returned to CETAC Technologies if
contaminated with biohazard or
radioactive materials, infectious agents, or
any other materials and/or conditions that
could constitute a health or injury hazard
to CETAC employees. Call Customer
Service and Support if there is any
question or doubt relative to
decontamination requirements.
CAUTION and WARNING statements, as
applied in this document, shall be
interpreted consistent with the following
context: CAUTION applies only to
potential property damage conditions;
WARNING applies to potential personal
injury conditions, in combination with or
exclusive of potential property damage.
WARNING
The handling of organomercurial
concentrates which may be used in the
preparation of process standards presents
a substantial (potentially lethal) safety
hazard. Only an experienced,
professionally trained organo-metallic
chemist, knowledgeable and skilled
specifically in the safe handling of
organomercurials (using approved
apparatus and approved protection
measures in an approved facility) should
attempt to prepare diluted
organomercurial process standards from
concentrates.
REVISIONS
CETAC Technologies strives to provide
the scientific community with an
unparalleled combination of effective
technology and continuing value.
Modular upgrades for existing
instruments will continue to be a prime
consideration as designs progress.
CETAC Technologies reserves the right to
revise this document and/or improve
products described herein at any time
without notice or obligation. Warranty
registration entitles the named owner
exclusively to manual change pages/new
editions as they are published.
NOTE
SD Acquisition, Inc., DBA CETAC
Technologies assumes no liability for the
handling of organomercurial concentrates
or the preparation, handling, or use of
diluted organomercurial process
standards.
recommends use of appropriate standard
reference materials to validate sample
preparation (dissolution/digestion) and
use of inorganic mercury standards for
instrument calibration.
All user-serviceable components are
specifically identified in this document as
such; the balance shall be assumed to
Instead
, CETAC Technologies
require the expertise of a factory service
technician/engineer for adjustment,
repair, replacement, modification, etc.
Others not so qualified and performing
these actions
Furthermore, never operate the
instrument without first reading and
understanding the ASX-260 Operator’s Manual and ensuring that it is operated
safely and properly.
shall do so at their own risk.
WARNING
ORIGINAL PACKAGING
Retain original factory packaging for
moves and factory return shipments.
Shipping in anything other than the
original fitted foam and container can
result in incidental damage from which
the purchaser will not be protected under
warranty.
Under all conditions the user must observe safe laboratory
procedures during the operation of this product.
Operator’s Manual Addendum
Notices and Compliance Declarations
FEDERAL COMMUNICATIONS
COMMISS ION (FCC) NOT ICE
This equipment has been tested and f ound
to comply with the limits for a Class A
digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to
provide reasonable protection against
harmful interference in a commercial
installation.
This equipment generates, uses, and can
radiate radio frequency energy and, if not
installed and u sed in accord ance with th e
instructions, may cause harmful
interference to radio communications.
Operation of this equipment in a
residential environment is likely to cause
harmful interference, in which case the
user will be required to correct the
interference at his own expense.
MODIFICATIONS
The FCC requires the user to be notified
that any change s or mo difications mad e to
this device that are not expressly approved
by CETAC may void the user's authority to
operate the equipment.
CABLES
CANADIAN NOTICE
This digital apparatus does not exceed the
Class A limits for radio noise emissions
from digital apparatus as set out in the
interference-causing equipment standard
entitled "Digital Apparatus." ICES-003 of
the Depar t me nt of Communications.
AVIS CANADIEN
Cet appareil numerique respecte les
limites de bruits radioelectriques
applicables aux appareils numeriques de
Classe A prescrites dans la norme sur le
materiel brouilleur: "Appareils
Numeriques," NMB-003 edictee par le
ministre des Communic atio ns.
Connections to this device must be made
with shielded cables with metallic RFI/EMI
connector hoods to maintain compliance
with FCC Rules and Regulations.
AD-1
Operator’s Manual Addendum
Notices and Compliance Declarations
POWER CORD SET REQUIREMENTS
The power cord set supplied with your
instrument meets the requirements of the
country where you purchased the
instrument.
If you use the instrument in another
country, you must use a power cord set
that meets the requirements of that
country.
WARNING
CAUTION
WARNING
This equipment is designed for connection to a grounded (earthed) outlet. The
grounding type plug is an important safe ty feature. To reduce the risk of electrical
shock or damage to the instrument, do not disable this feature.
To reduce the risk of fire hazard and electrical shock, do not expose the unit to rain or
humidity. To reduce the risk of electrical shock, do not open the cabinet. All maintenance
is to be performed by an Authorized CETAC Service Provider.
Protection provided by the equipment may be impaired if the equipment is used in a
manner not spec ified by the manufacturer.
CLEANING INSTRUCTIONS
To clean the exterior surfaces of the instrument, complete the following steps:
1
Shut down and unplug the instrument.
2
Wipe the instrument exterior surfaces
only using a towel dampened wit h a
lab-grade c leaning agent.
Do not a l low any li qu i d to enter th e i n st ru men t cabin et, or c ome i n to con tact with
any electrical compon ents. The instrument must be thoroughly dry before you
reco nne ct po wer, or turn t he instrume nt o n.
COOLING FAN OBSTRUCTION
The instru ment co oling fan( s) shall remain un obstruc ted at all t imes. Do not oper ate the
instrument if the cooling fan(s) are blocked or obstructed in any manner.
3
Repeat step 2, using a towel dampened
with clear water.
4
Dry the instrument exterior using a dry
towel.
ENVIRONMENTAL
Operating Temperature:10° to 30°C
Relative Humidity0% to 95%
Altitude0 to 10000 Ft
AD-2
AVERTISSEMENT
POUR UNE PROTECTION CON T INUÉ
CONTRE LES RISQUES D’INCENDIE,
REMPLACER UNIQUEMENT PAR DES
FUSIBLES DE MÊME TYPE ET
AMPÈRAGE.
Operator’s Manual Addendum
Notices and Compliance Declarations
PANNEAU NE DOIT ÊTRE ENLEVE QUE
PAR UN RÉPARATEUR QUALIFIÉ.
AVERTISSEMENT
TOUT CONTACT AVEC LES HAUTES
TENSIONS PEUT ENTRAINER LA MORT
OU DES BLESSURES SÉVÈRES. CE
PANNEAU NE DOIT ÊTRE ENLEVE QUE
PAR UN RÉPARATEUR QUALIFIÉ.
AVERTISSEMENT
TOUT CONTACT AVEC LES HAUTES
TENSIONS PEUT ENTRAINER LA MORT
OU DES BLESSURES SÉVÈRES. CE
AVERTISSEMENT
TOUT CONTACT AVEC LES HAUTES
TENSIONS PEUT ENTRAINER LA MORT
OU DES BLESSURES SÉVÈRES. CE
PANNEAU NE DOIT ÊTRE ENLEVE QUE
PAR UN RÉPARATEUR QUALIFIÉ.
AVERTISSEMENT
TOUT CONTACT AVEC LES HAUTES
TENSIONS PEUT ENTRAINER LA MORT
OU DES BLESSURES SÉVÈRES. CE
AD-3
Operator’s Manual Addendum
Notices and Compliance Declarations
PANNEAU NE DOIT ÊTRE ENLEVE QUE
PAR UN RÉPARATEUR QUALIFIÉ.
WARNING
HIGH LEAKAGE CURRENT ENSURE PROPER GROUNDING
AVERTISSEMENT
COURANT DE FUITE ÉLEVÉ — FORNIR
UNE MISE À LA TERRE E FFICACE.
Pinch point on X, Y, Z axis movement.
Attention – refer to the manual. T his symbol
indicates that information about usage of a
feature is contained in the manual.
WARNING
If the auto sampler is used in a manner not specified by CETAC
Technologies, the protection provided the equipment may be
WARNING
The power switch on the rear panel is not the mains disconnect.
Mains disconnect is a cco mplished by disconnecting the deta chable
power suppl y cord at the appliance coupler or at the mains plug.
Ensure the power cord is easily accessible and removable, in the
event of an emergency, which requires immediate disconnection.
AD-4
Operator’s Manual Addendum
Notices and Compliance Declarations
CAUTION
The CAUTION notice denotes a hazard. It calls attention to a
procedure, practice, or the like, that, if not correctly per for med or
adhered to, could result in personal injury. Do not proceed beyond a
CAUTION notice until the indicated conditions are fully understood
and met.
WARNING
This is a Safety Class 1 Product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall only be
inserted in a socket-outlet provided with a protective earth contact.
Intentional interruption is prohibited.
Safety Mainte nance
The operator should check the detachable power supply cord
condition. The equipment should not be operated if the mains inlet is
cracked or broken. Any obvious damage to the case (from a drop or
fall) should be checked by service personnel for loose or damaged
parts. See individual parts lists for approved replac ement parts.
Safety and Reg u lato r y Information
Review this product and related documentation to familiarize with
safety markings and instructions before you operate the instrument.
WARNING
The WARNING no tice denotes a hazard. It calls attention to a
procedure, practice, or the like, that, if not correctly per for med or
adhered to, could result in personal injury. Do not proceed beyond a
WARNING notice until the indicated conditions are fully understood
AD-5
Operator’s Manual Addendum
A
quip
p
A
p
Notices and Compliance Declarations
WARNING
Ensure that power is diconnected before removal of any covers.
DC Voltage and Current
24V
AC Voltage and Current
100-240V ~
47-63Hz 1.9A
3.33A
TM
of the European community
Other Fuse(s) – The equipment
uses one 2A 250V AC fuse which
Is not operator replaceable.
WARNING
Equipment is not intended for wet locations. Miscellaneous liquids in the
ment could cause hazardous conditions.
e
WARNING
Pollution Degree - 1
WARNING
Do not o
erate in explosive atmosphere.
WARNING
ll terminals allow a maximum current of 500MA @ 24VDC.
D-6
NOTE: For instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, or
auxiliary items for proper disposal please contact the
supplier or importer. In the event a supplier cannot be
reached, contact CETAC Technologies customer service
artment at 1 (800) 369 2822.
de
Conte nts
Conte nts
Prefacexii
Who Should Read This Bookxii
How to Use This Bookxii
Conventions Used in This Bookxiii
Instructionsxiii
Menu Itemsxiv
Terminologyxiv
Notesxv
Cautionsxv
Warningsxv
Where to Go for More Informationxvi
1Introduction1–2
Auto Sampler Standard Components1–2
Optional Accessories1–6
2 Preparing for Installation2–2
Choosing a Location2–2
Space Requirements2–2
Water Requirements2–2
Power Requirements2–3
Unpacking the ASX-2602–4
ASX-260 Auto Sampler Operator’s Manual
Contents
3 Installing the Auto Sampler3–2
Mounting the Z-Drive Assembly3–3
Attaching the Z-drive Assembly to the
Auto Sampler Arm3–4
Attaching the Z-drive Mounting Blocks3–4
Installing the Sample Probe3–6
Setting the Z-Axis Travel3–7
Connecting the Rinse Station3–9
Assembling a n d Pla cing the Sample Vial Racks3–14
Establishing External Connections3–15
Connecting the Auto Sampler to the Power Source3–16
Connecting the Auto Sampler to an Analytical Instrument3–16
Connecting the ASX-260 Auto Sampler
to the Host Computer3–16
Establishing a Serial Communications Interface3–18
Establishing an IEEE Communications Interface3–19
4 Verifying Installation4–2
Testing the Interface4–2
vii
ASX-260 Auto Sampler Operator’s Manual
Contents
Checking the Auto Sampler Components4–6
Testing the Sample Probe4–6
5 Using the Auto Sampler5–2
Estab lish in g Optimal Operating Conditions5–2
Creating the Lab Environment5–3
Replacing Auto Sampler Components5–4
Purchasin g Su pplies5–4
Arranging the Sample Vial Racks5–5
Starting the Auto Sampler5–5
Shutting Down the Auto Sampler5–7
Flushing the Rinse Station and Flow Path5–7
6 Maintaining the Auto Sam pler6–2
Cleaning t h e Auto Sampler6–2
Daily Exter n al Cleani n g6–3
Weekly Cleaning6–3
Checking for Leaks6–5
Replacing Peristaltic Pump Tubing6–6
Replacing the Sample Probe6–7
Replacing the Rinse Station Tubing6–7
Replacing the Sample Tray6–8
The ASX-260 Au t o Sampler Operator’s Manual explains the procedur es
for installing, using, and maintaining the CETAC ASX-260. It also
provides information about troubleshooting ASX-260 problems and
describes the design of the Auto Sampler.
Who Should Read This Book
The primary audience for the ASX-260 Operator’s Manual consists of
analytical chemis ts and lab technicians. To use this manual effectively,
you should have a strong knowledge of chemistry, a basic knowledge of
electronic sampling equipment, at least a beginning level of computer
experience, and working knowledge of the analytical instrument being
used with the ASX-260.
How to Use This Book
The ASX-260 Auto Sampler Operator’s Manual contains seven
chapters. You should read the chapters sequentially the first time.
Thereafter, refer to the chapters separately as needed. The first
chapter provides an introduction to the Auto Sampler. Subsequent
chapters d e t ail the primary tasks ass ociated with the ASX-260.
This manual contains the foll o wi ng chap ters:
Chapter 1, “Introduction,”
ASX-260 Auto Sampler’s function and de sign.
Cha pter 2, “Preparing for Installation,”
requireme nts that must be met before the ASX- 260 is installed. It also
provides i nstructions for unpa cking the Auto Sampler.
provides you with an overview of the
discusses spac e and pow er
ASX-260 Auto Sampler Operator’s Manual
Preface
Chapter 3, “Installing the Auto Sampler,”
procedures for installing the ASX-260 and connecting it to the
analytical instrument.
Chapter 4, “Verifying Installation,”
communic ation s inter fac e betw een the ASX- 260 and th e host c ompu ter.
It also explains how to check Auto Sampler components and test the
sample probe.
Chapter 5, “Using the Auto Sampler,”
perform during daily operation of the ASX-260.
Chapter 6, “Maintaining the Auto Sampler,”
weekly, and peri o dic maintenance tasks.
Cha pter 7, “Tr oubl eshooti ng t he Aut o Sa mpl er, ”
diagnose and correct ASX-260 problems.
These chapters are followed by a glossary of related terms.
provides step-by-step
explains how to test the
describes the tasks you
explains daily,
describes how to
Conventions Used in This Book
This book uses certain conventions to distinguish different types of
information easily. This section describes these conventions.
Instructions
All step-by-step instructions are numbered and in bold, as in the
following example.
1 Replace the sample vial racks.
Many numbered instructions are followed by more detailed
explanations.
xiii
ASX-260 Auto Sampler Operator’s Manual
Preface
Menu Items
This book uses the following format for referring to menu items:
Settings»Communication
The text before the arrow symbol is the name of the menu; the text
after the ar row symbol is the menu c hoice. Th is example re fers to the
Communications menu choice in the Settings menu.
Terminology
This book frequently uses the following terms:
ASX-260
Host co m puter
Hz
ICP-AES
ICP-MS
ID
LED
PEEK
VAC
VDC
The ASX-260 Auto Sampler.
The computer that controls operation of the ICP-AES
or ICP-MS in strument to wh ich the Au to Sampler is
attached.
Hertz.
An inductively coupled plasma atomic emission
spectrometer.
An inductively coupled plasma mass spectrometer.
Inside diameter.
Light-emitting diode.
Polyetheretherketone.
Volts alternating current .
Volts d irect current.
xiv
ASX-260 Auto Sampler Operator’s Manual
Preface
CAUTION
X-axis
Y-axis
Z-axis
The left-to-right axis of the Auto Sampler.
The front-to-back axis of the Auto Sampler.
The up-and-down axis of the Auto Sampler.
Notes
Notes contain a reminder about the e ffect of particular actions. Th ey
are indicated as follows:
Note:
This example shows how a note is displayed.
Cautions
Cautions indicate situations that require immediate attention to
prevent harm to the Auto Sampler. Cautions are indicated as follows:
This example shows how a caution is displayed.
WARNING
Warnings
Warnings indicate situations that could cause bodily harm. Warnings
are indicated as follows:
This example shows how a warning is displaye d.
xv
ASX-260 Auto Sampler Operator’s Manual
Preface
Where to Go for More Information
In addition to th e ASX-260 Auto Sampler Operator’s Manual, you can
refer to the following resources:
•The software manual for the analytical instrument you are using
The ASX-260 Auto Sampler is designe d to be sturdy , reliable, an d easy
to use. It pr ovides automated sample intr oduction that en ables you to
perform other tasks while the Auto Sampler runs. The ASX-260
automatically introduces up to 180 samples when fully loaded. It
contains a mic roprocessor that allo ws sequential or random sampling ,
providing f lexibility.
The ASX-260 is typically in terfaced to and c ontrolled by the analytical
instrument host computer using a serial or IEEE communications
protocol.
Auto Sampler Standard Components
Auto sampler components are made of corrosion-resistant stainless steel
alloys or anodized aluminum. The enclosure and base are made from a
high-strength aluminum alloy that is chromated and finished with an
epoxy powder coating.
The ASX-260 operates reliably under a wide variety of conditions.
Components in the sample flow path are made of polyetherimide (PEI)
and polytetrafluoroethylene (PTFE). When these inert, non-metallic
materials are used at temperatures less than 135°C, they can withstand
repeated exposure to the following substances:
•predominantly aqueous solutions of strong acids (less than 40%)
•predominantly aqueous solutions of strong bases (less than 10%)
•common organic solvents such as acetone, alcohols, ethyl acetate,
methylethylketone (MEK), petroleum oils and derived fuels,
tetrachloroethylene, toluene, and xylene
ASX-260 Auto Sampler Operator’s Manual
Introduction
CAUTION
Prolonged or repeated exposure to temperatures greater than 135°C
and to the following substances can cause failure of the PEI flow path
components:
•solutions of concentrated acids (greater than 40%)
•solutions of concentrated bases (greater than 10% potassium,
ammonium, or sodium hydroxides)
•partially halogenated hydrocarbons or extremely aggressive organic
The following standard components are located on the front of the ASX260 and are shipped with the Auto Sampler. Each lettered item
corresponds with a callout in Figure 1–1.
ASample tray.
The sample tray has 10 re ar standar d positions and
accommodates up to 2 sample vial racks. Ribs located on the bottom
of the sample tray hold the sample vial racks in place.
B Sample vial racks.
ASX-260 includes four sets of sample vial
racks. You can choose from five different rack sizes (with either 21,
24, 40, 60 or 90 vials per rack). You can combine racks of different
sizes as long as the host computer’s software accepts the
combination.
C Standards vials.
Ten standar ds vials are in cluded w ith the Auto
Sampler. The standards vials, which fit into the standards
positions at the back of the sample tray, are 50-milliliter conical
centrifuge vials with caps.
D Flowing rinse station.
The rinse station is located in the extreme
left position at th e back of the sample tray. It comes with tubing
used to connect the rinse station to the rinse sourc e and the w aste
container.
EZ -drive assembl y.
The z-drive assembly includes a y-axis slider
block and guide plate as well as the sample probe. The z-drive
assembly fits onto the Auto Sampler arm.
1–4
ASX-260 Auto Sampler Operator’s Manual
Introduction
Figure 1---- 2. ASX-260 Auto Sampler------ Back View.
The following standard components are located on the back of the ASX260 and are shipped with the Auto Sampler. Each lettered item
corresponds with a callout in Figure 1- -- 2.
AOn-board two-channel peristaltic pump. The on-board
peristaltic pump is located in the lower right-hand corner on the
back of the Auto Sampler. The pump moves the rinse solution from
the rinse source through the flowing rinse station.
1- -- 5
ASX-260 Auto Sampler Operator’s Manual
Introduction
B Two RS-232 serial I/O ports (COM1 and COM2).
ports are located on the left side at the back of the Auto Sampler.
The COM1 port is the communications interface between the ASX260 and the analytical instrument’s host computer.
The serial
C Optional 5-channel input/output (I/O) auxiliary port.
DUSB Option.
port can be used to interface the ASX-260 with the host computer.
The following standard components are also shipped with the ASX-260:
•Externa l d esk-top power suppl y
240V, 1 A, with an output of DC 24V, minimum 3.3 A.
•S ampl e probe kit.
tubing. The sample probe fits into the z-drive assembly.
•Serial interface kit.
computers with normal AT-style DTE serial ports, and a
1.828-meter mod ular cable.
TheASX-260 c omes standar d with a USB p ort. Th is
. The inpu t rating is AC 100V-
The kit includes the sample probe with sample
The kit includes DB9F port adapters for host
Optional Accessories
If you are doing a specialized type of analysis or are connecting the
ASX-260 to a host computer that uses a non-RS-232 communications
protocol, you may need optional accessories in addition to the standard
componen ts inc lud ed w ith the Au to Sampler . Th e follow ing acc essor ies
are available for the ASX-260:
•
IEEE-488 interface kit.
converter box, an IEEE-488 cable, a power cord, and instructions.
It is used to convert an IEEE communications protocol to a serial
protocol.
1–6
The kit includes an RS-232/IEEE-488
ASX-260 Auto Sampler Operator’s Manual
Introduction
Serial interface null adapter.
•
The null adapter replaces one
DB9F port adapter at the host computer. It is used for computers
with DCE-AT style serial ports.
•
Serial interface special adapter kit.
The adapter kit replaces
one or both standard serial port adapters with unwired DB9M,
DB25M, and DB25F adapters for special applications or host
computers with serial po rt s not conforming to the RS-232 standard.
•
Sample probes.
Available Probe Sizes
.3 mm IDBlack band
.5 mm IDBlue band
.8 mm IDRed band
.9 mm IDYellow band
1.0 mm ID2 Blue bands
Note:
Contact CETAC Technologies if you need additional accessories not
listed, need added features to int egrate the A SX-260 Auto Sam pler i nto
your analytical system, or have unique requirements. Research and
development of new features and accessories for the ASX-260 Auto
Sampler , of ten i nspir ed by custom er r equests, is a conti nui ng acti vity of
CETAC Technologies.
1–7
2
Preparing for Installation
Preparing for Installation
Installing the ASX-260 requires preparation. Before you install the
Auto Sampler, you should evaluate the physical arrangement of the
laboratory to choose a suitable location. Once you choose a location,
you must carefully unpack the Auto Sampler prior to beginning the
installation.
This chapter discusses what requirements must be met when you
choose a location for the Auto Sampler. It also describes how to unpack
the ASX-260 before installation.
Choosing a Location
Choosing a location for the ASX-260 involves evaluating the lab
environment for the availability of space, water, and power. For the
Auto Sampler to fu nction optimally, the location you selec t must meet
specific requirements associated with each of these items. The
following se ctions discus s space, wat er, and power requirem ents.
Space Requirements
Most analytical applications ben efit fr om th e sho rtest samp le flow path.
Therefore, you should place the Auto Sampler close to the analytical
instrument. The recommended minimum footprint for countertop
installation of the ASX-260 is 50 x 71 x 76 centimeters. The normal
operating position is horizontal, on a flat surface. The ASX-260
dimensions are 33 x 51 x 25 centimeters and weighs 8.4 kg.
Water Requirements
For most applications, deion ized water is used as a rinse agen t in the
ASX-260. If a different rinse agent is routinely used, place the rinse
agent source within 2 meters of the ASX-260.
ASX-260 Auto Sampler Operator’s Manual
Preparing for Installation
Ensure that there is a liquid waste receptacle within 2 meters of the
ASX-260. The waste receptacle inlet should be at least 30 to 60
centimeters lower than the Auto Sampler rinse station outlet.
Power Requirements
Place the ASX-260 within 1.2 meters of a power outlet. The Auto
Sampler’s input requirements are 110–240 VAC ±10%, 50/60 Hz, and
40 W. Voltage selection is automatic.
The ASX-260 uses an ex ternal d esktop pow er supp ly. Th e inpu t rating
is AC 110V-240V 1 A with an output of DC 24V, maximum 3.33 A.
Ensure that you position the ASX-260 so that the location where the
power supply cord plugs into it is easily accessible (is not blocked) and it
can be quickly disconnected if needed.
WARNING
The power supply socket is on the back of the Auto Sampler right below
the power switch. Conne ct the power supply to the Auto Sampler first
and then connect a line c ord to the power supply. Do not apply power
to the power supply until ready to operate the Auto Sampler.
The ASX-260 is intended to operate from an AC power source
that will not apply more than 253V ac between the supply
cond uct ors and ground. A pro tective gr ou nd c onne c tion by way
of the grounding connector in the power cord is required for safe
operation.
Unpacking the ASX-260 Auto Sampler
Inspect external packaging upon receipt for holes, tears, smashed
corners, or any other outward signs of damage from rough handling or
abuse during shipment. Inspect all items during unpacking and notify
the carrie r immediately of any con cealed damage.
2–3
ASX-260 Auto Sampler Operator’s Manual
Preparing for Installation
If the ASX-260 is shipped or removed f rom storage during cold weather,
allow the packaged equipment to attain room temperature before
opening and exposing to warm, humid air. It is usually sufficient to
provide 4 to 8 hours for this purpose.
CAUTION
If condensation forms on or inside the Auto Sampler, allow it to dry
thoroughly before connecting it to a power source and operating it.
Failure to do so may cause equipment damage.
Remove the packing checklist from the shipping container, and check
off items against it. Leave accessories in the packing until you are
ready to install them on the Auto Sampler.
Note:
Do not throw away the factory packaging. Keep it for possible future
use. This is one of the warran t y c ondi tions.
2–4
3
Installing the Auto Sampler
Installing the Auto Sampler
The ASX-260 Auto Sampler is designed for easy installation.
Installation consists of two parts: assembling the Auto Sampler and
connecting it to the host analytical instrument.
For the most p art, yo u can in stall the ASX- 260 with out using tools. In
fact, using tools such as screwdrivers or pliers to perform most
installation tasks may result in a damaged or unusable instrument.
You can remove thumbscrews with tools if necessary, but do not tighten
them with anything other than your fingers.
To install the Auto Sampler, you must complete the following tasks.
Each of these tasks will be discussed in detail later in this chapter.
1 Mount the z-drive assembly.
2 Connect th e rinse station.
3 Assemble the sample vial racks.
4 Establish external connections.
5 Conne ct the Auto Sampler to the h o s t c o mputer.
CAUTION
Ensure that AC power is off before proceeding with installatio n.
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Mounting the Z-Drive Assembly
Mountin g the z-drive asse mbly on the Auto Sampler is th e first major
task. The z-d rive asse mbly must be attached to the Au to Sample r ar m
to allow movement and function of the sample probe. Figure 3–1
illustrates the z-drive assembly components.
Figure 3–1.
Z-drive Assembly.
3–3
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Attaching the Z-drive Assembly to the
Auto Sampler Arm
To attach the z-drive assembly to the Auto Sampler arm, complete the
following steps, with reference to Figure 3-1.
1 Position the z-drive assembly at the free end of the Auto
Sampler arm with the z-drive assembly pointing up.
2 Match the 6-x 3-millimeter grooves in the Y-axis slider block
with the guide rails on the Auto Sampler arm, and slide the
block along the arm tube until the holes in the block align
with the matching holes in the Y-axis lead screw nut.
3 Secure the Y-axis slider block to the Y-axis lead screw nut
using the 12-millimeter nylon thumbscrews installed from the
top (through the bushings).
Take care to only finger tighten the nylon thumbscrews.
Attaching the Z-drive Mounting Blocks
The mounting block is already attached to the Auto Sampler.
3- -- 5
ASX-260 Auto Sampler Operator’s Manual
CAUTION
Installing the Auto Sampler
The PEEK cable goes to the right side of the
z-axis rotor (as viewed from the back of the Auto Sampler. See Figures
3-1 and 3-2. Failure to connect the cables correctly will result in Auto
Sampler malfunction. See also the Z-Drive Assembly Quick Installation Guide.
Figure 3---- 2. Back View of Auto Sampler Showing Z-Axis Rotor and
3- -- 6
Mounting Block.
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Installing the Sample Probe
1 Install a clamp in the slot on the z-axis slider (Figure 3---- 1).
2 Install the sample probe through the slider block and push
3 Move the z-axis slider plus attached sample probe to the top of
4 Leave approximately 105 millimeters of the sample probe’s
To install the sample probe, complete the following steps:
through the clamp (Figure 3----1).
the z-axis drive.
yellow-colored support tube extending above the top of the
z-axis slider (with the slider at the top of the z-axis drive).
Tighten the clamp. Clamp must be positioned as shown in figure
3-1.
5 Verify that the probe tip clears the top of the rinse station when
the Auto Sampler is in the home position (Figures 1-1, 3-1) above
the rinse station.
You can manually move the Auto Sampler arm with attached z-axis
drive to the rinse station without damage to the Auto Sampler.
6 Retain the sample transfer tubing with the spiral-wrap at tie
points approximately 15 and 40 centimeters above the top of the
z-axis drive, leaving an untangled service loop of approximately
13 to 15 centimeters above the probe.
3- -- 7
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
The sample transfe r tubing should still hav e slack re maining w hen th e
probe is a t the maximu m d ownw ard limit.
Setting the Z-Axis Travel
To set the z-axis travel of the z-dr ive assembly, co mplete the following
steps:
1 Adj u st the z-axis sl id er ( with atta ch ed sa m pl e probe) so that the
slider is approximately 3-millimeters below the top of z-axis drive
(see Figure 3-1).
2 Rotate the z-a xis rotor (Fi gur e 3–2) cl ockwise so the rotor stop
pin is against the rotor stop.
3 Finger tighten the rotor cla mp.
Ensure that th e PEEK pull-pu ll tube is fully located in the rotor clamp
groove . Oth erwise the PEEK tube will slip, resulting in no movement
of the z-axis slider.
4 Manually rotate the z-axis rotor back and forth several times
and check for full unhinder ed movement of the z-axis slider.
5 With the z-axis in the full-up position, hold the z-axis slider and
move the sample probe tube up and down so that 3 to 6
millimeters extends below the sample probe guide plate.
3–7
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
CAUTION
Figure 3–3.
Z-Drive Assembly with Z-Axis Slider.
Do not maneuver the sam p le probe directly as damage may result.
Note:
To make fine adjustments to the sample pr obe X/Y targeting, loosen the
nylon screw on the probe guide plate 1/8 to 1/4 turn and move the guide
plate up to 0.2 millimeters from the original location. Be sure to tighten
the nylon screw before operating the ASX-260.
3–8
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Connecting the Rinse Station
The cabinet-mounted rinse station is located at the extreme left position
in the standards rack. Typically, deionized water is used as the rinse
solution and is pumped into the rinse station by the on-board peristaltic
pump. An acid rinse such as 2% HNO
Since the p eristaltic pump in let is at the top of th e pu mp an d th e ou tlet
is at the bottom, the r inse water flows fr om th e bottom to th e top of th e
rinse station. Up-flow rinsing is the most effective method for
decontaminating the sample probe tube between samples. Reversing
the connections and the rinse water flow reduces the effectiveness of
the rinse station and can cause cross-contamination and unsatisfactory
performance.
The waste rinse solution drains from the top of the rinse station by
means of a pumped drain, which is the standard arrangement for
draining the rinse station. If it is not desirable to use a pumped drain,
you can create a gravity drain arrangement. This section contains
instructions for both the pumped drain arrangement and the gravity
drain arrangement.
solution may also be used.
3
Pumped Drain Arrangement
In a pumped dr ain arran gement, th e rinse solutio n moves thr ough the
on-board pe ristaltic pump to th e inle t at the bo ttom of th e r inse statio n,
as shown in Figure 3–4. It then drains out the top of the rinse station
and into the rinse solution waste container through the second channel
of the on-board peristaltic pump.
3–9
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Figure 3–4.
To connect the rinse station using the pumped drain, complete the
following steps:
1 Connect the rin se water sour ce to the on-boar d peri sta lti c pum p
by inserting the 3-mill imeter i.d. Tygon
the top of the pump.
1
Tygon is a registered trademark of Norton Co.
3–10
Pumped Drain Arrangement.
1®
tubin g on to the inlet at
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Use the tubing provided for the rinse solution uptake.
2 Connect th e peri stal tic pump to the ri nse stati on by completi ng
the following steps.
Use approximately 30 centimeters of the 3-millimeter i.d. Tygon
tubing provided for the rinse solution uptake.
a I nser t on e end of th e 3- mil l im eter i .d . T yg on® tubin g onto the
outlet at the bottom of the pump.
Insert the tubing car efully because th e peristaltic pump fitting gr ips
the tubing tightly . If you apply too much force, the fitting can break
off.
b Ins ert the oth er end of the 3-mill imeter i. d. Tyg o n® tubin g on to
the rinse tube inlet at the bottom of the rinse station.
Again, insert the tubing carefully to avoid breaking the fitting.
3 Connect the rinse station to the second channel of the on-board
peristaltic pump by completing the following steps.
Use approximately 30 centimeters of the 3-millimeter i.d. Tygon
tubing provided.
®
®
a Insert th e 3- millimeter i . d . Tygon® tubing onto th e top outl et of
the rinse station.
Insert the tubing c arefu lly becau se the rin se station fitting gr ips the
tubing tig htly. If you apply too mu ch force, the fitting can break off.
b Pl ace the other end of th e tubi ng onto th e pu mp inl et (top of
channel 2).
Again, insert the tubing carefully to avoid breaking the fitting.
4 Connect the second channel of the on-boa rd peristaltic pump to
the rinse solution waste container by completing the following
steps.
Use up to 1.8 meters of the tubing provided for the pumped drain.
3–11
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
a Insert th e 3- mill imeter i.d . Tygon® tubin g onto the peri stal tic
pump outlet.
Insert the tubing c arefu lly becau se the rin se station fitting gr ips the
tubing tig htly. If you apply too mu ch force, the fitting can break off.
b Place the other end of the tubing into the waste container.
Ensure that the tubing outlet is placed in the waste container so that
it will not be immersed in the waste solution. Immersion of the drain
tube outlet may cause the waste solution to back up and overflow.
Gravity Drain Arrangement
Figure 3–5.
In a gravity drain ar rangement, the rinse solution move s through the
peristaltic pump to the inlet at the bottom of the rinse station, as sho w n
in Figure 3-5. It then drains out the top of the rinse station by means
of a gravity drain.
To connect the rinse station using a gravity drain, complete the
following steps:
3–12
Gravity Drain Arrangement.
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
1 Connect the rinse solution source to the on-board peristaltic
pump by inserting the 3-millimeter i.d. Tygon
®
tubing onto the
inlet at the top of the pump.
Use the tubing provided for the rinse solution uptake.
2 Connect th e peri stal tic pump to the ri nse stati on by completi ng
the following steps.
Use approximately 30 centimeters of the 3-millimeter i.d. Tygon
tubing provided for the rinse solution uptake.
a I nser t on e end of th e 3- mil l im eter i .d . T yg on® tubin g onto the
outlet at the bottom of the pump.
Insert the tubing car efully because th e peristaltic pump fitting gr ips
the tubing tightly . If you apply too much force, the fitting can break
off.
b Ins ert the oth er end of the 3-mill imeter i. d. Tyg o n® tubin g on to
the rinse station inlet at the bottom of th e rinse stat ion.
Again, insert the bottom tubing carefully to avoid breaking the
fitting.
3 Connect the rinse station to the wa ste container by completing
the following steps.
®
Use up to 1.8 meter s of th e 5-millimeter i.d. Tyg on® tubing provided for
the gravity drain. Ensur e that the waste c ontainer is at le ast 30 to 60
centimeters lower than the rinse station outlet.
a Insert the 5-millimeter i.d. Tygon® tubing onto the rinse
station outlet (on top).
Insert the tubing carefully because the rinse station fitting grips the
tubing tig htly. If you apply too mu ch force, the fittings can break off.
b Place the other end of the tubing into the rinse solution waste
container.
Ensure that the tubing outlet is placed in the waste container so that
it will not be immersed in the waste solution. Immersion of the drain
tube outlet may cause the waste solution to back up and overflow.
3–13
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Assembling and Placing the Sample Vial Racks
Sample vial racks for the ASX-260 Auto Sampler are shipped
unassembled. However, you can easily assemble them without using
tools. Once you assemble the sample vial racks, place them in the
sample tray before proceeding with the installation.
WARNING
Before loading or unloading any sample vial racks on the
sample tray, park the sampling arm and probe in the home
posi tion by cycl ing the power off an d on. The home positi on is
the ini tia l positi on at power-up. Never a ttempt to loa d, unl oad ,
or reposition a sa mple via l rack or sa mple vial whil e the Auto
Sampler is operating.
To assemble and place the sample vial racks, complete the following
steps:
1 Snap the ra cks together as shown in the instructions included
with each ra ck.
You can also easily disassemble the r acks if you ne ed to ship or store
them.
Note:
Keep at least one copy of the assembly i nstructions prov ided for each
rack. Keep the copy with this manual for reference.
2 P lac e th e fi rs t sa m pl e via l r a ck a t th e extr eme l eft- han d si de of
the sa mple tra y so tha t the feet on th e ra ck’s undersi de eng age
the locating ribs on the sample tray's surface.
Correctly placed sample vial racks will not move more than ±2
millimeters in either a left/right or forward/backward direction unless
3–14
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
you lift them. Tilted sample vials in dicate an improperly placed r ack,
which must be corrected before you operate the Auto Sampler.
3 Pl a ce the next sampl e via l r a ck to the ri gh t of the previous ra ck
so th a t th e feet on t he r a ck ’s u nd er sid e en ga g e th e l oca t in g ri bs
on the sample tray’s surfac e.
Ensure that the rack is not tilted and is properly placed.
4 Repeat step 3 until you place all the sample vial racks.
When viewed from the front of the ASX-260, the sample vial racks
should now be arran ged on the sample tray as follows: rack #1 at th e
extreme left position, rack #2 at the left center position, rack #3 at the
right center po sition, and rack #4 at the extreme right position.
Establishing External Connections
WARNING
The next step in the installation process involves connecting the
ASX-260 Auto Sampler to the power source and to an analytical
instrument. The following sections explain how to establish these
connections.
Connecting the Auto Sampler to the Po wer Source
A voltage-specific external desktop power supply is supplied with each
ASX-260 Auto Sampler.
Use only this external desktop power supply or an exact
replacement.
To connect the Auto Sampler to a power source, plug the external
desktop power supply cord into the power connector located on the back
panel of the Auto Sampler. Then, plug the power supply’s power cord
into a 100–240-VAC ±10%, 50/60-Hz utility power outlet.
3–15
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Connecting the ASX-260 to an Analytical Instrument
You can connect the Auto Sampler directly to a sample introduction
peristaltic pump and then to any sample introduction device, such as
the CETAC U-5000AT
complete the following steps:
+
or U-6000AT+ Ultrasonic Nebulizer . To do so,
1 Determine the length of the sample transfer tubing you need,
and cut it to siz e.
2 Connect the free end of the sa mpl e transfer tu bin g to the inlet of
the analytical instrument’s pe ristaltic pump tubing.
Connecting the ASX-260 Auto Sampler to the Host
Compute r
You cannot operate the ASX-260 Auto Sampler until you establish a
communications interface between the Auto Sampler and the host
computer. It is through this interface that the host computer directs
the operation of the ASX-260 Auto Sampler. The ASX-260 Auto
Sampler supports the following two communications protocols:
•The serial ( RS-232C) protoco l is the standard c onfigu ration. There
are two RS-232C serial ports on the ASX-260, an d a serial in terface
kit is shipped with the A uto Sampler.
•The parallel (IEEE-488) protocol is less common than the serial
configu r ation. An IEEE-488 interface kit is av ailable as an option al
accessor y to the ASX- 260. See Chapter 1, “Introduction,” for more
information about this kit.
3–16
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Note:
Although the ASX-260 supports both serial and IEEE communications
protocols, the host computer governs which protocol is used. To
determine which protocol is required f or the analytical instrument you
are using, ref er to the hardware or software manual provided wit h the
instrument. The following sections explain how to establish a
communications interface using a serial protocol and using an IEEE
protocol.
CAUTION
When interconnecting any computing devices, keep the
communications cables away from sources of electromagnetic or radio
frequency (RF) interference, such as electric motors, transformers,
fluorescent light ballasts, or RF energy sources. Limit cable runs for
RS-232C to less than 16 meters. If these conditions cannot be satisfied,
use low-impedance, fully shielded cables to provide satisfactory
operation. The cables are available from many sources, but you will
need to specify the correct mating connectors and “straight-through”
(DTE-DCE) wiring.
Establishing a Serial Communications Interface
The serial interface kit provided with the ASX-260 Auto Sampler
includes an interface cable equipped with two modular port adapters.
Use the interface kit to establish a serial communications interface with
the host computer. To do so, complete the following steps:
1 Plug one end of the cable into the host computer’s serial (COM)
port selected for Auto Sampler communications.
3–17
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Make sure that the COM port you s elect m a tches the port selected in
the host computer’s software.
2 Finger tighten both screws of the cable adapter.
Note:
If a host computer serial port with a DB9F, a DB25M, or a DB25F
connector (9 pin D-submini receptacle or 25 pin D-submini plug or
receptacle) must be used, use the mating connec tor from the CETAC
Technologi es univ ersal port adapter ki t. You can order the adapter kit
from CETAC Technologies or purchase an adapter locally to convert
the serial port to a DB9M. Do not use a “null modem” ad apter.
3 Connect the other end of the cable to the ASX-260 COM1 port.
CAUTION
Ensure that you are connecting the adapter to the COM1 port.
Connecting the adapter to the COM 2 port on the Auto Sampler will
cause a malfunction. The Auto Sampler COM 2 port is used for
communications to other devices other than the host computer.
4 Finger tighten both sc rews of the cable adapter.
3–18
ASX-260 Auto Sampler Operator’s Manual
Installing the Auto Sampler
Establishing an IEEE Communications Interface
Before you can establish an IEEE communications interface with the
host computer, you must have the IEEE-488 interface kit, which
includes a conv erter box and an IEEE-488 cable. Use the IEEE-488 kit
to establish a parallel interface with the host computer. To do so,
complete the following steps:
1 Connect the converter box according to the manufacturer’s
instructions.
Manufacturer’s instructions are provided with the kit.
2 Plug one end of the cable into the serial port on the 232/488
conver ter box.
3 Finger tighten both screws of the cable adapter.
CAUTION
4 Connect the other end of the cable to the ASX-260 COM1 port.
Ensure that you are connecting the adapter to the COM1 port.
Connecting the adapter to the COM 2 port on the Auto Sampler will
cause a malfunction. The Auto Sampler COM 2 port is used for
communications to other devices other than the host computer.
5 Finger tighten both screws of the cable adapter.
3–19
4
Verifying Insta llation
ASX-260 Auto Sampler Operator’s Manual
Verifying Installation
Verifying Installation
Once installation of the ASX-260 Auto Sampler is complete, it is
important to verify that you have in stalled it correctly . Attempting to
use it b e f o re ensur ing that it is installed correctly may result in damage
to the Auto Sampler.
Verifying installation of the Auto Sampler consists of two parts:
•ensuring that the communications interface between it and the host
computer is working
•ensuring that the sample probe functions properly.
This chapter explain s how to test the abo ve ite ms befor e u sing the Auto
Sampler.
Note:
The procedures given in thi s chapter are for use in a W indows® 95 or
Windows® NT environment.
Testing the Interface
If the communications interface between the Auto Sampler and the host
computer is not established correctly, the Auto Sampler will not
function . Befor e you test the in terface , ensu re that th e commun ication
port connectors are properly attached between the host computer and
the Auto Sampler.
4–2
ASX-260 Auto Sampler Operator’s Manual
Verifying Installation
Note:
The fol lowing procedures assume that you have opened W indows® 95
or Windows® NT and the Program Manager w indow is showing.
To test the communications interface, complete the following steps:
1 S ta rt the host computer and go to the m ain Windows® screen.
2 Turn on power to the Auto Sampler.
3 Click the start button in the lower left corner of the Program
Mana ger window.
A selection list will appear.
4 Select Programs>>>Accessorie s>>>Hyperterminal and double-
click.
The Hyperterminal®1 window appears (Figure 4–1).
Figure 4–1.
1
Hyperterminal i s a registered tr ademark of Hilgraeve, Inc.
Example of Hyperterminal® Window.
4–3
ASX-260 Auto Sampler Operator’s Manual
Verifying Installation
5 Double click on the Hyperterminal® icon .
The Connection Description box appears (Figure 4–2).
Figure 4–2.
Connection Description Box.
6 T ype the name C O M 1 _ tes t f or t h e c on n ec t i on a n d ch o ose an icon
from the list given, and click OK .
The phone number box appears (Figure 4-3).
4–4
ASX-260 Auto Sampler Operator’s Manual
Verifying Installation
Figure 4-3.
Phone Number Box.
7 Select Direct to COM1 in the Connect Using box. Click OK.
The COM1 Properties box appears (Figure 4-4).
Figure 4-4.
COM1 Properties Box.
4–5
ASX-260 Auto Sampler Operator’s Manual
Verifying Installation
8 Change the bits per second to 9600, set the data bits to 8, the
parity to none, the stop bits to 1, the flow control to none, and
click OK.
9 Select File>>>Prope rtie s.
The COM1_Test Properties box appears.
10 Se lect Settings on the COM_Te st Prope rties box and click on the
ASCII setup button on the lower right of the box.
The ASCII setup box appears (Figure 4-5).
Figure 4-5.
11 In the ASCII Setup box, select the following items:
a. Echo typed characters lo cally.
b. Append line feeds to incoming line ends.
c. Wrap lines that exceed terminal width.
4–6
ASCII Setup Box.
Click OK.
ASX-260 Auto Sampler Operator’s Manual
Verifying Installation
12 Click OK on the COM1_Test Properties box.
13 Type HOME at the cursor in the upper lef t of the main
Hyperterminal® scre en and press Enter.
The Auto Sampler resets, with the sample probe moving out and back
into the home position. If the Au to Sampler does no t reset, see Ch apter
7, “Troubleshooting the Auto Sampler”.
Checkin g the Aut o S ampler Components
The following Auto Sampler components may be damaged during
shipping or installation: the sample probe, the peristaltic pump tubin g,
and the rin se station and tubin g. It is impor tant that you chec k these
components for damage before you operate the Auto Sampler. To do so,
complete the following steps:
1 Shut down and unplug the Auto Sampler.
2 Visu al l y in spect the sampl e probe, perista l tic pump tu bing , a nd
rinse station and tubing for leaks or sig ns of damage.
If you detect a leak or other damage to an Auto Sampler component,
you must replace it. For more information, see the appropriate section
in Chapter 6, “Maintaining the Auto Sampler”.
Testing the S ample Probe
The sample probe must descend into the center of each sample vial to
ensure satisfactory sample uptake. Shipping or rough handling can
disturb the Auto Sampler’s cabinet-to-base alignment. If it is
incorrec tly aligned, the sample probe will n ot function pr operly. It is
therefore important to test the sample probe before you actually run
sampl e s with the Auto Sampler.
4–7
ASX-260 Auto Sampler Operator’s Manual
Verifying Installation
Note:
Before testing t he sampl e probe, ensure that you hav e i nstal led al l Auto
Sampler components correctly. Also, ensure that you have securely
tightened all thumbscrews and connected the communications cable
from the host computer to the COM1 port on the Auto Sampler.
Testing the sample probe involves observing the operation of the
sample probe. To do so, complete the following steps:
1 Load the Auto Sampler sample tray with empty sample vial
racks.
For information about placing the sample vial racks, see Chapter 3,
“Installing the Auto Sampler”.
2 Turn the Auto Sampler power switch on and veri fy that the LED
power indicator is on.
The LED power indicator is green in color. The indicator is located
behind the z-drive assembly when it is in the home position.
3 Using the host computer control software, designate sample
positions at the left r ea r, l eft fr ont, ri g ht r ea r, and r ig h t fr ont of
the sample tray.
4 Place sample vial racks and sample vials at the designated
positions.
5 Command the Auto Sampler to move the sample probe to the
designated sample positions. Check that the sample probe
correctly accesses each position a n d that the probe descends into
the center of each sample vial.
4–8
ASX-260 Auto Sampler Operator’s Manual
Verifying Installation
Note:
If the Auto Sampler alignment is not correct, contact CETAC
Technologies Customer Service and Support or an authorized
representative.
4–9
5
Using the Auto Sampler
ASX-260 Auto Sampler Operator’s Manual
Using the Auto Sampler
Usin g the Auto Sampler
The ASX-260 Auto Sampler is both reliable and easy to use. Before
using it, however, ensure that your lab environment provides operating
conditions that will prolong the life of the ASX-260 Auto Sampler. Once
the proper operating conditions are met, you can arrange the sample
vial racks and start the Auto Sampler sequence run. When you finish
using the Auto Sampler, you may need to flush the rinse station and
flow path before shutting the Auto Sampler down.
This chapter explains how to create the proper operating conditions for
using the ASX-260 Auto Sampler . It also explains h ow to arrange the
sample vial racks, start and shu t dow n the Au to Sampler , and f lush the
rinse station and flow path.
Establishing Optimal Operating Co nditions
The ASX-260 Auto Sampler operates reliably even under less than ideal
conditions. It is not, however , indestructible. Malfu nction or damage
can occur if specific operating conditions are not met. Meeting these
conditions requires that you create the proper lab en viron ment, r eplace
Auto Sampler components that wear out under normal use, and
purchase the appropriate supplies for use with the Auto Sampler. The
following sections explain how to meet these conditions.
Note:
Damage or malfunction that results from unsatisfactory operating
conditions may constitute misuse and abuse and be excluded from
warranty coverage.
5–2
ASX-260 Auto Sampler Operator’s Manual
Using the Auto Sampler
Creating the Lab Environment
To create satisfactory operating conditions in your lab environment,
follow these guidelines:
•Operate the ASX-260 Auto Sampler in a conventional lab
environment where the temperature is 50–95 °F (10–35 °C), the
humidity is 20–70% non-condensing, and the unit is not exposed to
excessive flammable or corrosive materials.
•Avoid rough handlin g of the ASX-260 Auto Sample r. If p ossible, do
not expose the Auto Sampler to vibration or shock.
•Protect the Auto Sampler from long-term exposure to condensation,
corrosive materials, solvent vapor, continual standing liquids, or
large spills into the Auto Sampler cabinet or arm. Exposures of this
type can damage the drive mechanisms as well as the electronics.
•Observe the same general electrostatic discharge precautions as
with any other integrated circuit electronic dev ices. Low humidity
environments, especially when combined with static-generating
materials, require maximum care.
WARNING
Disc har ge stati c buil dup a nd gr ound to the Auto S ampl er ba se
or ca binet before performing any ma intena nce. Do not touch or
short-circuit bare conta cts, COM1, COM2, or auxiliary ports.
•Avoid using the ASX-260 Auto Sampler if strong electromagnetic
interference, radio frequency interference, or radioactivity is
present. Interference fields can cause erratic operation of the Au to
Sampler. The Auto Sampler will not function properly if the level of
radioactivity is above back ground.
5–3
ASX-260 Auto Sampler Operator’s Manual
Using the Auto Sampler
Replacin g Auto Sampler Components
The following ASX-260 Auto Sampler components wear out under
normal use and must be replaced periodically.
•peristaltic pu mp tu bing
•sample probe
If you fail to replace these components when they deteriorate, the Auto
Sampler will not function properly. For information about replacing
Auto Sampler components, see Chapter 6, “Maintaining the Auto
Sampler.”
Purchasing Suppli es
Because the life-span of the sample and standards vials varies, you
should maintain an adequate supply of spare vials. When you need to
purchase additional supplies, it is extremely important that you choose
the appropriate sizes and materials.
When you purchase sample and/or standards vials, make sure they
meet the following requirements:
•The diameter of the sample or standards vial matches the rack size
you are using. The diameter of the standar ds vial is 28 millimete rs,
plus or minus 1 millimeter.
•The height does not exceed 125 millimeters.
•Vials used for standards are conical-bottom centrifuge vials.
Anything other than tape red-bottom vials will not remain pro perly
positioned in the rear standards rack.
•The material is c ompatible with the samples y ou are analy zing or
the reference standards you are using. This requirement also
applies to the peristaltic pump tubing.
5–4
ASX-260 Auto Sampler Operator’s Manual
Using the Auto Sampler
WARNING
CAUTION
Use of mismatched sample vials and sample vial racks may
resul t in ma l fun ction s or sa mpl e spil ls. Be sur e you r via l s m eet
the given requirements.
To order additional supplies, refer to the CETAC Accessories and
Supplies Catalog for the ASX-260 Auto Sampler.
Arra n ging the Sample Vial Racks
You can change the arrang ement of the sample vial racks to meet your
needs. The ASX-260 Auto Sampler accommodates up to two sample vial
racks of 21, 24, 40, 60, or 90 positions each. The Auto Sampler
physically accepts any and all size combinations and arrangements of
sample vial racks. However, you can use racks of mixed sizes only if
the host computer’s software enables you to do so.
For more information about placing sample vial racks in the sample
tray, see Chapter 3, “Installing the Auto Sampler.”
Incorrec tly defining the position c ount can result in sample spills and
invalid analysis results .
Starting the Auto Sampler
Once you arrange the sample vial racks and ensure that the
arrangement is correctly defined in the software, you can start the Auto
Sampler and let it run until the sampling sequen ce is finished. T o do
so, complete the following steps:
5–5
ASX-260 Auto Sampler Operator’s Manual
Using the Auto Sampler
1 Ensure that the rinse station is properly connected.
For more information about proper connections, see Chapter 3,
“Installing the Auto Sampler.”
2 Turn the ASX-260 Auto Sampler power switch on.
The green LED indicator along the Auto Sampler x-axis lights up when
the power is on.
3 Adj ust the perista l tic pump sh oe until the desir ed rin se soluti on
flow rate is achieved.
4 Purge air from the rinse system by placing the rinse solution
uptake tubing in the rinse solution source and running the rinse
solution through the rinse station.
Ensure there are no air bubbles visible in the rinse uptake tubing
before you run samples with the ASX-260 Auto Sampler.
Note:
If you are flushing the rinse system dur ing init ial start up, fi rst use a 2%
nitri c acid sol ution as the ri nse agent. Fl ush the rinse system a second
time using deionized water as the rinse agent.
5 Access the host computer’s software and activate the Auto
Sampler program.
The ASX-260 Auto Sampler runs until it reaches the end of the
sampling sequence.
5–6
ASX-260 Auto Sampler Operator’s Manual
Using the Auto Sampler
Sh utting Down the Auto Sampler
To shut down the Auto Sampler, complete the following steps:
1 Drain the rinse system by removing the rinse solution uptake
tubin g from the rin se solu tion sourc e. L et the peri stal tic pump
run unti l a l l solu tion dr a in s from th e tube atta ch ed to the rin se
station outlet.
If you use a rinse solution other than deionized water, flush the rinse
system with deionized water before shutting down the ASX-260 Auto
Sampler. For more information, see the following section, “Flushing
the Rinse Station and Flow Path.”
2 Release the pressure shoe on the peristaltic pump.
Releasing the pressure shoe decreases wear on the pump tubing.
3 Turn off the ASX-260 Auto Sampler power switch.
Flushing the Rinse S tation and Flow Path
Generally, you can operate the ASX-260 Auto Sampler without flushing
the rinse system. Under normal circumstances, you can simply drain
the rinse sy stem prior to shutting down the Auto Sampler . However,
you need to flush the rinse station and flow path under two
circumstances:
•during initial startup of the ASX-260 Auto Sampler after
installation
•following the use of strong bases, acids, or organic solvents as rinse
agents
5–7
ASX-260 Auto Sampler Operator’s Manual
Using the Auto Sampler
Flushing the rinse system during initial startup of the ASX-260 Auto
Sampler removes any contaminants that could cause interference
during sample analysis. Flushing the rinse system after using strong
rinse agents prevents degradation and failure of the flow path
components.
To flush the rinse station and flow path, complete the following steps:
1 Insert the rinse uptake tubing into a deionized water source.
Note:
If you are f lushing the rinse system during initial startup, first use a 2%
nitric acid solution as the rinse agent, followed by deionized water.
2 Run the rinse solution through the rinse station and flow path
for 5 to 10 minutes.
Once you flush the rinse system, you can proceed with the sampling
sequence or drain the rinse system as part of the shutdown procedure.
For information about running the sampling sequence, see “Starting
the Auto Sampler ” earlier in this chapter. For more information about
draining the rinse system, see the previous section, “Shutting Down the
Auto Sampler.”
5–8
6
Maintaining the Auto Sampler
Maintaining the Auto Sampler
Routine maintenance of the ASX-260 Auto Sampler consists of daily
and weekly cleaning of specific Auto Sampler components. Routine
maintenance also includes checking ASX-260 components for leaks or
other damage. Additional periodic maintenance tasks may be req uired,
including replacement of the following Auto Sampler components:
peristaltic pump tubin g , sample pro be, rin se station tubing, and sample
tray.
This chapter exp lains how to c lean the ASX-260 Au to Sampler, in spect
it for leaks, and replace damaged components.
WARNING
Disc har ge stati c buil dup a nd gr ound to the Auto S ampl er ba se
or ca binet before performing any ma intena nce. Do not touch or
short-circuit bare contacts, COM1, COM2, or auxiliary port s.
Cleaning the Auto Sampler
Cleaning the ASX-260 Au to Sampler is the primary maintenance task
you perform. Failure to do so regularly causes increased wear and
reduces the Auto Sampler’s life.
You must clean the ASX-260 Auto Sampler both daily and weekly to
prevent damage and extend its life. It is especially important to clean
up spills and remove contaminants, such as abrasives, from the Auto
Sampler’s moving parts. It may also be necessary to chemically
neutralize spills. The following sections explain daily and weekly
cleaning procedures.
ASX-260 Auto Sampler Operator’s Manual
Maintaining the Auto Sampler
Daily External Cleaning
Use of the ASX-260 Auto Sampler often results in spills on Auto
Sampler components such as the sample tray. Good maintenance
requires that you clean the Auto Sample r daily. To do so, c omplete th e
following steps:
1 Shut down and unplug the Auto Sampler.
For infor mation abou t shutting down the Auto Samp ler, see Chapter 5,
“Using the Auto Sampler.”
2 Wipe the sa mple tra y, Auto Sa mpl er ca binet, a nd Auto S am pler
arm using a towel dampened with a lab-grade cleaning agent.
CAUTION
Do not allow the cleaning agent to come into contact with the lead
screws. Also, never lubricate either of the two lead screws.
3 Repeat step 2, using a towel dampened with clear water.
This process removes any remaining contaminants.
4 Dry the sample tray, Auto Sampler ca binet, and Auto Sampler
arm using a dry towel.
The ASX-260 Auto Sampler must be thoroughly dry before you turn the
Auto Sampler pow e r on.
Weekly Cleaning
Although cleaning it daily r emoves spills and con taminants from most
of the Auto Sampler componen ts, it is necessary to c lean the ASX-260
Auto Sampler more thoroughly once a week. To do so, complete the
following steps:
6–3
ASX-260 Auto Sampler Operator’s Manual
Maintaining the Auto Sampler
1 Shut down and unplug the Auto Sampler.
2 Remove the sample tray.
For information about removing the sample tray, see “Replacing the
Sample Tray” later in th is ch apte r.
3 W ipe l o os e pa rticles o ff the Y-axis lea d scr ew with a d ry, lin t-free
cloth.
The Y-axis lead screw is a large metal screw located inside the Auto
Sampler arm tubing , as s h own in Figure 6–1.
Figure 6–1.
6–4
Z-Drive Assembly on Auto Sampler Arm.
ASX-260 Auto Sampler Operator’s Manual
Maintaining the Auto Sampler
WARNING
Never lubricate the lead screws. The lead screw nuts are
compounded with a dry film lubricant. Oiling the lead screws
will cause gumming, galling, and binding of the sample probe
asse mbly.
4 Wipe th e Auto Sa mpl er exterior and base until they a re clea n,
using a towel dampened with a lab-grade cleaning agent,
followed by a towel dampened with clear water.
Pay special attentio n to the slider bloc k and gu ide rails alo ng the tube
of the Auto Sampler arm.
5 Wash the sample tray in a warm detergent solution.
Make sure you remove all spills and stains.
6 Rinse the sample tray with water and then dry it.
Ensure tha t the sampl e tray is thoroughly dry.
7 Replace the sample tray on the Auto Sampler base.
For information about replacing the tray, see “Replacing the Sample
Tray” later in this chapter.
Checking for Le aks
Several of the Auto Sampler components have a limited life and will
wear out under normal use: the sample probe, the peristaltic pump
tubing, and the rinse station and tubing. Standard maintenance
procedures require that you periodically check these components for
leaks. To do so, complete the following steps:
6–5
ASX-260 Auto Sampler Operator’s Manual
Maintaining the Auto Sampler
1 Shut down and unplug the Auto Sampler.
2 Visu al l y in spect the sampl e probe, perista l tic pump tu bing , a nd
rinse station and tubing for leaks or signs of deterioration.
If you detect a leak or other damage to an Auto Sampler component,
you must replace it. For more information, see the appropriate section
in this chapter.
Replacing Peristaltic Pump Tubing
Routine maintenance of the ASX-260 Auto Sampler includes
replacement of the peristaltic pump tubing. Because of the operating
nature of peristaltic pumps, the tubing will probably be the most
frequently replaced item on the Auto Sampler. If you use strong bases,
acids, or solvents as rinsing agents, the tubing may break down rapidly.
To replace the peristaltic pump tubing, complete the following steps:
1 Shut down and unplug the Auto Sampler.
2 Release the pressure shoe and remove the old tubing.
Carefully pull or cut the old tubing to remove it.
3 Replace the pump tubing by pushing the new tubing onto the
mounting block fittings.
Replace the new tubing carefully. Damage can result if you apply too
much force.
4 Reconnect the pressure shoe.
6–6
ASX-260 Auto Sampler Operator’s Manual
Maintaining the Auto Sampler
Replacing the Sample Probe
You must replace the sample probe if it is leaking or shows other signs
of deterioration. To do so, complete the following steps:
1 Shut down and unplug the Auto Sampler.
2 Remove the old sample probe and tubing.
Be careful not to use excessive force when removing the sample probe.
Applying too much force can result in damage to the z-drive assembly.
3 Ins tall the new sample probe.
For information about installing the sample probe, see Chapter 3,
“Installing the Auto Sampler.”
4 With the z-drive in the full -up position, hold the z-axis slider and
move the sample probe tube up and down so that 3 to 6
millimeters extends below the sample probe guide plate.
Replacing the Rinse Station Tubing
If the rinse station tubing is typically exposed to deionized water as a
rinsing agent, you do not need to replace it often. However, if you use
other rinsing agents, such as acids or solvents, the tubing is likely to
deteriorate more rapidly. To replace the rinse station tubing, complete
the following steps:
1 Shut down and unplug the Auto Sampler.
2 M ove the A ut o S a mpl er a rm 20 to 3 0 c ent im eter s a wa y f rom th e
home position by gently pushing it.
6–7
ASX-260 Auto Sampler Operator’s Manual
Maintaining the Auto Sampler
Moving the Auto Sampler arm ensures th at the sample probe will not
be damaged while you replace the rinse station tubing.
3 Disconnect the rinse s oluti on uptake and drain tubing.
Apply only a linear force when removing the tubing to prevent the
fittings from breaking.
4 Remove the rinse station tube by completing the following steps:
a Rotate the rinse station tube counterclockwise 1/4 turn.
b Remove the rinse station tube from the mounting block by
lifting the tube straight up.
Repla c e t h e r ins e st ati o n tube by pu shin g th e n ew r ins e s t ati on
5
tube into the mounting block and rotating it clockwise 1/4 turn.
6 Reconnect the rinse solution uptake and drain tubing.
Apply only a linear force when replacing the tubing to prevent the
fittings from breaking.
7 Move the Auto Sampler arm back to the home position.
Replacing the Sample Tray
Cleaning the ASX -260 Au to Sampler sample tray each w eek ex ten ds its
life and makes frequent replacement unnecessary. However, if the
sample tray needs to be replaced, complete the following steps:
1 Shut down and unplug the Auto Sampler.
2 Remove all sample vial racks.
3 M ove the A ut o S a mpl er a rm 20 to 3 0 c ent im eter s a wa y f rom th e
home position by gently pushing it.
6–8
ASX-260 Auto Sampler Operator’s Manual
Maintaining the Auto Sampler
Moving the Auto Sampler ar m ensur es that the sample probe assembly
will not be damaged while you replace the sample tray.
4 Raise the rinse station tube approximately 2 centimeters.
5 Ra i se the front edge of the da m a g ed tray at l ea st 2. 5 centi meters
and slide it forward.
If you have difficulty removing the sample tray, raise the front edge
higher be f ore sliding it forward.
6 Install the new tray.
7 Lower the rinse station tube.
Ensure the rinse station tube is positioned securely.
8 Move the Auto Sampler arm back to the home position.
9 Replace the sample vial racks.
6–9
7
T rou bleshooting the Auto Sampler
Troubleshooting the Auto Sampler
The ASX-260 Auto Sampler is both easy to operate and reliable.
However, problems with it may occur. When the Auto Sampler does
not function properly, isolate the problem to determine if it originates
in the host computer, the analytical instrument, the RS-232 cable, or
the Auto Sampler. If you determine the problem is in the ASX-260 ,
check the power system, the communications interface, or the sample
probe assembly to find the cause of the problem and resolve it.
This chapter explains how to troubleshoot ASX-260 Auto Sampler
problems. If you cannot solve a problem using the steps given in this
chapter, contact CETAC Technologies Customer Service and Support.
Power System Problems
A possible cause of ASX -260 Auto Sampler malfu nction is a pr oblem in
the power system. If the Auto Sampler is not func tional, ther e may be
no power getting to it. If this is the case, the green LED power
indicator will be off. To troubleshoot this problem, complete the
following steps:
1 Check the wall outlet and see if the external power supply is
plugged in.
2 Check that the power switch of the external power supply is
turned on.
ASX-260 Auto Sampler Operator’s Manual
Troubleshooting the Auto Sampler
Interface Problems
Operation of the ASX-260 Auto Sampler is directed by the host
computer. A malfunction of the Auto Sampler can in dicate a problem
with the RS-232 cable or with the c onfiguration of the software o n the
host computer. The following sections explain how to troubleshoot
these problems.
RS-232 Cable Problems
The first step in troubleshooting interface problems is to check the
RS-232 cable. To do so, complete the following steps:
1 Check the Auto Sampler power switch to ensure it is on.
2 Check the RS-232 cable to ensure it is plugged in to the COM1
port on the Auto Sampler.
If the cable is plugged in, ensure that it is tightened properly.
3 Check the host computer to ensure that the RS-232 cable is
connected to the appropriate COM port.
If the cable is plugged in, ensure that it is tightened properly. For more
information about connecting the RS-232 cable, see Chapter 3,
“Installing the Auto Sampler.”
Software Configuration P roblems
If the RS-232 cable is connec ted properly and the Auto Sampler is still
not communicating with the host computer, ensure that the host
software is configured correctly. To do so, complete the following steps:
1 Run the host software and ensure that the instrument is
functioni ng properly.
7–3
ASX-260 Auto Sampler Operator’s Manual
Troubleshooting the Auto Sampler
2 Check the software configuration for the correct COM port
sel ection and bau d rate (9600, N, 8, 1).
If the wrong port or baud rate is selected, change the configuration.
For information about changing the software configuration, see
Chapter 4, “Verifyin g Installation.”
Z-Drive Assembly Problems
A ASX-260 Auto Sampler malfunction may be c aused by a problem in
the z-drive assembly . You can easily deter mine that a malfunction is
related to the z-drive assembly if you hear a loud chattering noise when
the ASX-260 pow er switch is on or if the sample probe is not mov ing.
To troubleshoot z-drive assembly problems, complete the following
steps:
1 Ensure that the Y-axis slider block and z-drive assembly are
installed.
If the z-drive assembly is not installed, fo llow the instr uctions pro vided
in Chapter 3, “Installing the Auto Sampler ,” to install it. If the z-driv e
assembly is already installed, co ntinue with step 2.
2 Check the Y-axis block home position flag for damage.
The home position flag is shown in Figure 7–1. If the flag is damaged,
you must r eplace the en tire y-axis slider bloc k. For inf ormation abo ut
mounting the y-axis slider block on the Auto Sampler arm, see
Chapter 3, “Installing th e Auto Sampler.”
If the home position flag is undamaged, continue with step 3.
7–4
ASX-260 Auto Sampler Operator’s Manual
Troubleshooting the Auto Sampler
Figure 7–1.
Z-Drive Assembly with Y-Axis Block Home Position Flag.
7–5
ASX-260 Auto Sampler Operator’s Manual
Troubleshooting the Auto Sampler
3 Check t hat th e sample probe i s moving.
If the sample probe is binding, free the sample probe assembly.
Note:
If you cannot free the z-drive assembly, you will need to replace it. See
Chapter 3, “Installing the Auto Sampler,” for inf or m ati on about m ounti ng
a new z-driv e assembly. You can order a new z-dri ve assembl y from
CETAC Technologies.
7–6
Glossary
Glossary
This glossary defines the terms used in the ASX-260 Auto Sampler
Operator’s Manual.
analytical instrument
instrument, typically an ICP-AES
or ICP-MS, to which the Auto
Sampler is connected.
Auto Sampler arm
that extends from the front of the
ASX-260 cabinet. It governs the
left / right and forward /
backward travel of the z-drive
assembly.
I/O ports
for establishing communication
between the ASX-260 and the
host computer or other external
devices.
perista lti c pump
pump controlling the movement
of the rinse solution to the rinse
station.
rinse solution
typically deionized water, used to
clean the sample probe.
The connections used
The on-board
The solution,
The
The arm
Y-axis slider block
componen t that fits over the Auto
Sampler arm and executes the
forward/backward movement of
the z-drive assembly along the
Auto Sampler arm.
Z-axis rotor
that contr ols the mov emen t of th e
Z-axis slider.
Z-axis slider
assembly component th at gov erns
the up-and-down travel of the
sample probe.
The component
The z-drive
The
rinse station
component used to clean the
sample probe with a rinse
solution.
sample probe
moves the analyte from the
sample vial to the sample transfer
tubing.
The Au to Sampler
The tube that
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.