Masimo rainbow SET Pronto-7 Spot Check CO-Oximeter, also referred to as
Pronto-7, operating instructions provide the necessary information for proper
operation of Pronto-7. There may be information provided in this manual that
is not relevant for your system. Do not operate Pronto-7 without completely
reading and understanding the instructions in this manual.
Notice:
Purchase or possession of this instrument does not carry any express or
implied license to use this instrument with replacement parts which would,
alone or in combination with this instrument, fall within the scope of one of the
relating patents.
Caution: Federal law (U.S.) restricts this device to sale by or on the
order of a physician.
For professional use. See instructions for use for full prescribing
information, including indications, contraindications, warnings,
precautions and adverse events.
Masimo Corporation
40 Parker
Irvine, CA 92618
USA
Tel.: 949-297-7000
Fax.: 949-297-7001
www.masimo.com
EU Authorized Representative for Masimo Corporation:
About this Manual -------------------------------------------------------------------------- 55
Product Description ----------------------------------------------------------------------- 77
Indications for Use --------------------------------------------------------------------- 77
Contraindications ----------------------------------------------------------------------- 77
Safety Information, Warnings, and Cautions---------------------------------------- 99
Safety Information, Warnings, and Cautions --------------------------------- 1010
Compliance Information, Warnings and Cautions --------------------------- 1213
Sensor Information, Warnings and Cautions --------------------------------- 1414
Unpacking and Inspection the System ----------------------------------------- 2525
Preparation for Use ----------------------------------------------------------------- 2525
Common Screens ------------------------------------------------------------------- 2828
Commonly Used Icons ------------------------------------------------------------- 2828
Recommended Initial Testing Procedure -------------------------------------- 3030
www.masimo.com 3 Masimo
Pronto-7Pronto-7Contents
Patient Testing ----------------------------------------------------------------------- 3332
Completed Test ---------------------------------------------------------------------- 3534
Incomplete Test ---------------------------------------------------------------------- 3634
Key Features ------------------------------------------------------------------------- 3635
Main Menu Options ----------------------------------------------------------------- 5147
Battery ---------------------------------------------------------------------------------- 5854
Chapter 5: Alarms and Messages -------------------------------------------------- 6157
Chapter 8: Service and Repair ------------------------------------------------------ 7167
Service and Repair ------------------------------------------------------------------ 7167
Index --------------------------------------------------------------------------------------- 7973
www.masimo.com 4 Masimo
About this Manual
This manual explains how to set up and use Pronto-7. Important safety
information relating to general use of the product appears in this manual.
Read and follow any warnings, cautions, and notes presented throughout this
manual. The following are explanations of warnings, cautions and notes.
A warning is given when actions may result in a serious outcome (for example,
injury, serious adverse effect and, death) to the patient or user. The following
is an example of a warning:
WARNING: This is an example of a warning statement.
A caution is given when any special care is to be exercised by the patient or
user to avoid injury to the patient, damage to Pronto-7 or damage to other
property. The following is an example of a caution:
CAUTION: This is an example of caution statement.
A note is given when additional general information is applicable. The
following is an example of a note:
Note: This is an example of a note.
Read the entire safety information section before you operate the
instrument.
www.masimo.com 5 Masimo
Product Description
Indications for Use
The Masimo rainbow SET Pronto-7 Spot Check CO-Oximeter and
Accessories are indicated for noninvasive spot check testing of functional
saturation of arterial oxygen hemoglobin (SpO2), pulse rate (PR), and total
hemoglobin concentration (SpHb). The Masimo rainbow SET Pronto-7 Spot
Check CO-Oximeter and Accessories are indicated for use by trained
personnel, with adult and pediatric individuals, in clinical and non-clinical
settings (e.g., hospitals, hospital-type facilities, home, clinics, physician
offices, and ambulatory surgery centers).
The Masimo rainbow SET Pronto-7 Spot Check CO-Oximeter is designed to
simultaneously and noninvasively measure functional arterial oxygen
saturation (SpO2), pulse rate (PR), total hemoglobin (SpHb), perfusion index
(PI), and hematocrit (SpHct).
Advanced features available on the Pronto-7 include SpHct access, multi test
mode, low signal I.Q., measurement through motion of SpHb and SpO2,
wireless download of spot check tests, SpO2 only mode and EMR connectivity
for parameter and measurement download.
The instrument is voice automated, provides step by step instructions when
administering tests and is equipped with an interactive touchscreen for user
selectable options.
Contraindications
The Pronto-7 is contraindicated for use as an apnea monitor. The Pronto-7 is
also contraindicated for use as a continuous monitor.
www.masimo.com 7 Masimo
Safety Information, Warnings, and
Cautions
The following section lists warnings, cautions, notes, and safety information
for Pronto-7.
Pronto-7 is to be operated by qualified personnel only. The manual,
accessories, directions for use, all precautionary information, and
specifications should be read before use.
www.masimo.com 9 Masimo
Pronto-7 Safety Information, Warnings, and Cautions
Safety Information, Warnings, and Cautions
Always use the Pronto-7 precisely in accordance with the directions in
this manual, including finger selection, finger alignment in the sensor,
and subject behavior during testing. Failure to follow all of the
directions in this manual could lead to inaccurate measurements.
The Pronto-7 should be considered an early warning device. Blood
samples should be analyzed by laboratory instruments prior to clinical
decision making to completely understand the patient’s condition.
Variation in hemoglobin measurements may be profound and may be
affected by sample type, body positioning, as well as other
physiological conditions. As with most hemoglobin tests, Pronto-7 test
results should be scrutinized in light of a specific patient’s condition.
Any results exhibiting inconsistency with the patient’s clinical status
should be repeated and/or supplemented with additional test data.
Pulse rate measurement is based on the optical detection of a peripheral flow
pulse and therefore may not detect certain arrhythmias. The Pronto-7 should
not be used as a replacement or substitute for ECG based arrhythmia
analysis.
If SpO2 values indicate hypoxemia, a laboratory blood sample should be taken
to confirm the patient’s condition.
SpO2 is empirically calibrated in healthy adult volunteers with normal levels of
carboxyhemoglobin (COHb) and methemoglobin (MetHb).
Inaccurate SpO2 readings may be caused by:
• Elevated levels of COHb or MetHb: High levels of COHb or MetHb
may occur with a seemingly normal SpO2. When elevated levels of
COHb or MetHb are suspected, laboratory analysis (CO-Oximetry)
of a blood sample should be performed.
• Intravascular dyes such as indocyanine green or methylene blue
• Externally applied coloring and texture such as nail polish, acrylic
nails, glitter, etc.
• Elevated levels of bilirubin
• Severe anemia
• Low arterial perfusion
• Motion artifact
Inaccurate SpHb readings may be caused by:
• Improper sensor application
• Intravascular dyes such as indocyanine green or methylene blue
www.masimo.com 10 Masimo
Pronto-7 Safety Information, Warnings, and Cautions
• Externally applied coloring and texture such as nail polish, acrylic
nails, glitter, etc.
• Elevated PaO2 levels
• Elevated levels of bilirubin
• Low arterial perfusion
• Motion artifact
• Low arterial oxygen saturation levels
• Elevated carboxyhemoglobin levels
• Elevated methemoglobin levels
• Hemoglobinopathies and synthesis disorders such as thalassemias,
Hb s, Hb c, sickle cell, etc.
• Vasospastic disease such as Raynaud's
• Peripheral vascular disease
• Liver disease
• EMI radiation interference
SpHct is NOT reliable in cases of abnormal blood composition and is not
indicative of disease states. Abnormal values of SpHct that are displayed
should warrant repeat testing by conventional laboratory methods.
High intensity extreme lights (including pulsating strobe lights and direct
sunlight) directed on the sensor, may not allow the Pronto-7 to obtain
readings.
As with all medical equipment, carefully route patient cabling to reduce the
possibility of patient entanglement or strangulation.
Do not place the Pronto-7 or accessories in any position that might cause it to
fall on the patient. Do not lift the Pronto-7 by the cable or sensor.
Ensure the sensor is physically intact, with no broken or frayed wires or
damaged parts. Visually inspect the sensor and discard if cracks or
discoloration are found.
Interfering Substances: Dyes, or any substance containing dyes, that change
usual blood pigmentation may cause erroneous readings.
Do not use the Pronto-7 or sensor during magnetic resonance imaging (MRI)
scanning. Induced current could potentially cause burns. The Pronto-7 may
affect the MRI image and the MRI device may affect the accuracy of the Pulse
CO-Oximetry parameters and measurements.
Do not use the Pronto-7 during electrocautery.
Do not use the Pronto-7 or sensor during defibrillation.
Do not place the Pronto-7 where the controls can be changed by the patient.
www.masimo.com 11 Masimo
Pronto-7 Safety Information, Warnings, and Cautions
Do not expose the Pronto-7 to excessive moisture such as direct exposure to
rain. Excessive moisture can cause the instrument to perform inaccurately or
fail.
Do not place containers with liquids on or near the Pronto-7. Liquids spilled on
the instrument may cause it to perform inaccurately or fail.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment.
Explosion hazard. Do not use the Pronto-7 in the presence of flammable
anesthetics or other flammable substances in combination with air,
oxygen-enriched environments, or nitrous oxide.
EMI radiation interference such as computer displays and / or LCD / plasma
TVs can cause errors or incorrect measurements on the Pronto-7.
To protect against injury from electric shock, follow the directions below:
• Avoid placing the device on surfaces with visible liquid spills.
• Do not soak or immerse the device in liquids.
• Use cleaning solutions sparingly.
Always remove the sensor from the patient and completely disconnect the
patient from the Pronto-7 before bathing the patient.
A functional tester cannot be utilized to assess the accuracy of the Pronto-7 or
its rainbow 4D DC reusable sensors.
Do not place the Pronto-7 on electrical equipment; it may prevent it from
working properly. If Pronto-7 is used adjacent to other electrical equipment,
the device should be observed to verify normal operation.
Compliance Information, Warnings and Cautions
Use the Pronto-7 in accordance with the Environmental Specifications in this
manual.
Do not incinerate device and/or battery.
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
www.masimo.com 12 Masimo
Pronto-7 Safety Information, Warnings, and Cautions
instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Cleared Use Only: The device and related accessories are cleared by the
Food and Drug Administration (FDA) for noninvasive patient monitoring and
may not be used for any processes, procedures, experiments or any other use
for which the device is not intended or cleared by the FDA, or in any manner
inconsistent with the instructions for use or labeling. The device and related
accessories are not intended for use in combination with other medical
devices or in high-risk applications.
Disposal of product - Comply with local laws in the disposal of the instrument
and/or its accessories.
This equipment has been tested and found to comply with Class B limits for
medical devices according to the EN 60601-1-2. These limits are designed to
provide reasonable protection against harmful interference in all
establishments, including domestic establishments.
This Class B digital apparatus complies with Canadian ICES-003.
Do not open the Pronto-7 instrument. Only a qualified operator may perform
maintenance procedures specifically described in this manual. Refer servicing
to Masimo for repair of this equipment.
In accordance with international telecommunication requirements, the
frequency band of 5,150 MHz to 5,250 MHz is only for indoor usage to reduce
potential for harmful interference to co-channel mobile satellite systems.
Pronto-7 is provided with RF wireless capabilities. When used outside the US,
consideration should be taken to account for local/ national regulations or
restrictions for RF wireless technologies prior to using the wireless feature.
RF Exposure: The Pronto-7 was tested for SAR compliance with a 0mm
separation distance. In order to maintain FR exposure compliance, the
device should not be used for any other body worn accessories.
www.masimo.com 13 Masimo
Pronto-7 Safety Information, Warnings, and Cautions
Sensor Information, Warnings and Cautions
If using the Pronto-7 during full body irradiation, keep the sensor out of the
radiation field. If the sensor is exposed to the radiation, the reading might be
inaccurate or the instrument might read zero for the duration of the active
irradiation period.
Failure to apply the sensor properly may lead to incorrect measurements.
Do not loop the sensor cable into a tight coil or wrap around the device, as this
can damage the sensor cable.
Additional information specific to Masimo's rainbow 4D DC reusable sensor,
including information about parameter/measurement performance during
motion and low perfusion, may be found in the sensor's Directions For Use
(DFU).
Do not expose the rainbow 4D DC reusable sensor to moisture, liquids or a
humid environment, as this may make the sensor perform inaccurately or fail.
Do not use damaged sensors. Do not use a sensor with exposed optical or
electrical components. Do not immerse the sensor in water, solvents, or
cleaning solutions (the sensors and connectors are not waterproof). Do not
sterilize by irradiation, steam, autoclave or ethylene oxide unless otherwise
indicated in the sensor's directions for use. See the cleaning instructions in the
sensors' directions for use.
Do not attempt to reprocess, recondition or recycle any Masimo sensors or
patient cables as these processes may damage the electrical components,
potentially leading to patient harm.
www.masimo.com 14 Masimo
Chapter 1: Technology Overview
Signal Extraction Technology (SET)
Masimo Signal Extraction Technology's signal processing differs from that of
conventional pulse oximeters. Conventional pulse oximeters assume that
arterial blood is the only blood moving (pulsating) in the measurement site.
During patient motion, however, the venous blood also moves, causing
conventional pulse oximeters to read low values, because they cannot
distinguish between the arterial and venous blood movement (sometimes
referred to as noise).
Masimo SET pulse oximetry utilizes parallel engines and adaptive digital
filtering. Adaptive filters are powerful because they are able to adapt to the
varying physiologic signals and/or noise and separate them by looking at the
whole signal and breaking it down to its fundamental components. The
Masimo SET signal processing algorithm, Discrete Saturation Transform®
(DST®), in parallel with Fast Saturation Transform (FST®), reliably identifies
the noise, isolates it and, using adaptive filters, cancels it. It then reports the
true arterial oxygen saturation for display on the monitor.
rainbow Pulse CO-Oximetry Technology
Pulse CO-Oximetry is governed by the following principles:
(non-oxygenated blood), carboxyhemoglobin (blood with carbon
monoxide content), methemoglobin (blood with oxidized
hemoglobin) and blood plasma constituents differ in their absorption
of visible and infrared light (using spectrophotometry, see figure
below).
www.masimo.com 15 Masimo
Pronto-7 Chapter 1: Technology Overview
Emitting Diodes (LEDs)
2. The amount of arterial blood in tissue changes with a person’s pulse
(photoplethysmography). Therefore, the amount of light absorbed by
the varying quantities of blood changes as well.
The Pronto-7 uses a multi-wavelength sensor to distinguish between
oxygenated blood, deoxygenated blood and blood plasma. The Pronto-7
utilizes a sensor with various light-emitting diodes (LEDs) that pass light
through the site to multiple photodiodes (detectors). See the figure below.
Light (7+ Wavelengths)
Detectors
Signal data is obtained by passing various visible and infrared lights (ranging
from 500nm up to 1300nm) through a capillary bed (for example, a fingertip)
and measuring changes in light absorption during the blood pulsatile cycle.
This information may be useful to clinicians. The maximum radiant power of
the strongest light is rated at ≤ 25mW. The detectors receive the light, convert
it into an electronic signal and send it to the Pronto-7 for calculation.
Once the Pronto-7 receives the signal from the sensor, it utilizes proprietary
algorithms to calculate the patient's functional oxygen saturation (%SpO2),
total hemoglobin concentration (SpHb [g/dL]) and pulse rate (BPM). The
SpHb measurement relies on a multiwavelength calibration equation to
estimate the percentage of total hemoglobin in blood. In an ambient
temperature of 95º F (35º C) the maximum skin surface temperature has been
measured at less than 106º F (41º C), verified by Masimo sensor skin
temperature test procedure.
Functional Oxygen Saturation
The Pronto-7 is calibrated to measure and display functional oxygen
saturation (SpO2): the amount of oxyhemoglobin expressed as a percentage
of the hemoglobin that is available to transport oxygen. Refer to the Safety
information, Warnings and Cautions section in front of this manual for details.
Pronto-7 vs. Drawn Whole Blood Measurements
When SpO2 and SpHb measurements obtained from the instrument
(noninvasive) are compared to drawn whole blood measurements (invasive)
www.masimo.com 16 Masimo
Pronto-7 Chapter 1: Technology Overview
by blood gas and/or laboratory hematology, caution should be taken when
evaluating and interpreting the results. The blood gas and/or laboratory
hematology measurements may differ from the SpO2 and SpHb
measurements of the Pronto-7.
In the case of SpO2, different results are usually obtained from the arterial
blood gas sample if the calculated measurement is not appropriately
corrected for the effects of variables that shift the relationship between the
partial pressure of oxygen (PO2) and saturation, such as: pH, temperature, the
partial pressure of carbon dioxide (PCO2), 2,3-DPG, and fetal hemoglobin.
High levels of bilirubin may cause erroneous SpO2 and SpHb readings. As
blood samples are usually taken over a period of 20 seconds (the time it takes
to draw the blood) a meaningful comparison can only be achieved if the
oxygen saturation, carboxyhemoglobin and mehemoglobin concentration of
the patient are stable and not changing over the period of time that the blood
gas sample is taken. Subsequently, blood gas and laboratory hematology
measurements of SpO2 and SpHb may vary with the rapid administration of
fluids and in procedures such as dialysis. Additionally, drawn whole blood
testing can be affected by sample handling methods and time elapsed
between blood draw and sample testing.
Non-Invasive Total Hemoglobin (SpHb) Accuracy
Compared to Invasive Laboratory Methods*
The following is data of the sensor compatible with the Pronto-7: Rainbow 4D
DC Reusable Sensor.
rainbow 4D DC Reusable Sensor
For Normal Sensitivity Mode in 10,253 comparisons of SpHb® and invasive
hemoglobin (tHb) measurements from a laboratory reference device in the
range of 8-17 g/dL, SpHb accuracy was as follows:
• 0.90 correlation
• 1.05 g/dL A
For Maximum (MAX) Sensitivity Mode in 13,205 comparisons of SpHb and
invasive hemoglobin (tHb) measurements from a laboratory reference device,
SpHb accuracy was as follows:
accuracy
RMS
• 0.86 correlation
• 1.00 g/dL A
accuracy
RMS
www.masimo.com 17 Masimo
Chapter 2: Product Description
Pronto-7 Description
The Pronto-7 is designed to simultaneously and noninvasively measure
functional arterial oxygen saturation (SpO2), pulse rate (PR), total hemoglobin
(SpHb), perfusion index (PI) and hematocrit (SpHct ). The instrument is voice
automated, provides step by step instructions when administering tests, and is
equipped with an interactive touchscreen for user selectable options.
The Test Ready screen, as seen in the following image, indicates a
compatible sensor has been properly inserted into the Pronto-7 and the
instrument is ready for testing.
The Main Menu icon, also available on the Test Ready screen, provides
access to additional icons and user configurable options.
Parameters and measurements are displayed on the Test Results screen and
may include functional arterial oxygen saturation (SpO2), pulse rate (PR), total
hemoglobin (SpHb), perfusion index (PI), and hematocrit (SpHct).
Measurements are collected through the rainbow 4D DC sensor, connected
via the connector port.
www.masimo.com 19 Masimo
Pronto-7 Chapter 2: Product Description
Pronto-7 Front Panel and Touchscreen
www.masimo.com 20 Masimo
Pronto-7 Chapter 2: Product Description
Displays instrument's time.
Ref. Feature Description
1
Power On/Off
2
LCD Touchscreen
3
Test Ready Screen
4
Connector Port
Places instrument in Power
On and Power Off modes.
Provides an interface for
user interactions.
Indicates Pronto-7 is ready
for testing.
Provides connectivity for
rainbow 4D DC sensors.
5 Time
6
Battery Status
7
Action Icons
8
Test Icon
9
Main Menu Icon
10
Radical Sign
11
Sensor Size
12
Connectivity
Indicates the estimated
percentage of battery
power remaining.
Provides user selectable
icons for navigation.
Allows for commencement
of spot check test.
Provides access to
additional icons and user
configurable options.
Indicates the number of
spot check test credits
remaining on the attached
sensor.
Specifies the size of the
attached sensor (small,
medium or large).
Indicates Wi-Fi and
Bluetooth status.
www.masimo.com 21 Masimo
Pronto-7 Chapter 2: Product Description
Pronto-7 Back Panel
Ref. Feature Description
1 Certification
Label
2 Serial Number Lists unique serial number associated with the
Documents the instrument’s certification marks.
instrument.
www.masimo.com 22 Masimo
Pronto-7 Chapter 2: Product Description
Pronto-7 Bottom Panel
Ref. Feature Description
1 Earphone Jack 3.5mm earphone jack.
2 Micro SD Card
Slot
Micro SD flash memory card slot.
3 Mini USB Port Input port for mini USB cable to computer
4 Power Port Power supply connector for the Pronto-7 specific
connection.
power cable.
Parameter and Measurement Descriptions
Total Hemoglobin (SpHb)
Pulse CO-Oximetry is a continuous and noninvasive method of measuring the
levels of total hemoglobin (SpHb) in arterial blood. It relies on the same
principles of pulse oximetry to make its SpHb measurement. The
measurement is taken by a sensor capable of measuring SpHb, usually on the
fingertip for adult and pediatric patients.
The sensor connects directly to the Pulse CO-Oximeter or with a patient
cable. The sensor collects signal data from the patient and sends it to the
instrument. The instrument displays the calculated data as measurement of
total hemoglobin concentration.
www.masimo.com 23 Masimo
Pronto-7 Chapter 2: Product Description
Arterial Oxygen Saturation (SpO2)
Pulse oximetry is a noninvasive method of measuring the level of functional
arterial oxygen saturation in blood. The instrument displays the calculated
data as a percent value for functional arterial oxygen saturation (SpO2).
A SpO2 reading is associated with correct sensor placement, small
physiological changes during the measurement and acceptable levels of
arterial perfusion at the measurement site.
Pulse Rate (PR)
Pulse rate (PR), measured in beats per minute (BPM) is based on the optical
detection of peripheral flow pulse.
Perfusion Index (PI)
The Perfusion Index (PI) is the ratio of the pulsatile blood flow to the
non-pulsatile or static blood in peripheral tissue. PI thus represents a
noninvasive measure of peripheral perfusion that can be continuously and
noninvasively obtained from a pulse oximeter.
Hematocrit (SpHct)
Hematocrit (SpHct) is the fraction of whole blood volume that consists of red
blood cells. In normal conditions, there is a linear relationship between
hematocrit and the concentration of hemoglobin. An estimated hematocrit as
a percentage may be derived by multiplying the hemoglobin concentration in
g/ dL times three and dropping the units 1,2. The hematocrit measurement is
determined mathematically in the instrument as*:
%SpHct = 3.0422 x SpHb – 0.8059
The measurement is taken by a sensor capable of measuring SpHct, usually
on the fingertip for adults or pediatric patients. The sensor collects signal data
from the patient and sends it to the instrument. The instrument displays the
calculated data as measurement of SpHct. The hematocrit estimation is NOT
reliable in cases of abnormal blood composition and is not indicative of
disease states. Abnormal values of hematocrit that are displayed should
warrant repeat testing by conventional laboratory methods.
*Note: This is the relationship found between tHb and SpHb for 3,226 blood
samples measured on a Coulter LH 500 Hematology Analyzer.
www.masimo.com 24 Masimo
Chapter 3: Setup
Unpacking and Inspection the System
To unpack and inspect the system
1. Remove Pronto-7 and the components from the shipping carton.
2. Examine them for signs of shipping damage.
3. Check all materials against the packing list.
4. Save all packing materials, invoice and bill of lading. These may be
required to process a claim with the carrier.
If anything is missing or damaged, contact the Masimo Technical Service
Department, see Return Procedure on page 7369.
Preparation for Use
Before the Pronto-7 can be used, it needs to be properly setup and fully
charged.
Note: The initial battery charge can take up to 6 hours.
1. Power the instrument on. The touchscreen will illuminate and a
series of audible tones will sound. To change the audible tone
settings, see Sound Icon on page 5249.
2. The time and date need to be set. Enter your region's correct date
and time.
3. Tap the green check mark icon to accept changes.
4. Fully charge the instrument using the included AC power cable.
5. Insert the rainbow 4D DC sensor to the connector port.
www.masimo.com 25 Masimo
Loading...
+ 61 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.