CEM SMART System 5 Operation Manual

600141
Rev.3
CEM
P.O. Box 200
Matthews, North Carolina 28106-0200
(704) 821-7015 FAX: 704-821-7894
Copyright 2000 by CEM Corporation
All Rights Reserved
This manual contains proprietary information which shall not
be reproduced or transferred to other documents or disclosed
to others without prior written permission
of CEM Corporation
SMART System5™ is a trademark of CEM Corporation
CEM is a registered trademark of CEM Corporation
U.S. patent 4,291,775; 4,438,500; 4,457,632;
4,554,132; and 4,753,889
Other U.S. and foreign patents pending
MANUFACTURED IN THE
UNITED STATES OF AMERICA
This instrument complies with United States Code of Federal Regulations (CFR) Title 21, Part 1030 for microwave leakage. A verification report is on file.
This instrument complies with United States Code of Federal Regulations (CFR) Title 47, Federal Communications Commission (FCC) Part 18 – Industrial, Scientific and Medical (ISM) Equipment – emissions requirements. A verification report is on file.
i
NOTICE
The following precautions should be observed to avoid possible exposure to excessive microwave energy:
Do not tamper with the safety interlocks. The SMART System5 is equipped with three safety interlocks which prevent the instrument from producing microwave power if the cover is open.
Do not place any object between the microwave cavity and the cover or allow soil or cleaner residue to accumulate on sealing surfaces.
Do not operate the instrument if it is damaged. It is particularly important that the instrument cover close properly and that there is no damage to the cover (bent), hinges and latch (broken or loosened), or sealing surfaces.
The instrument should be adjusted or repaired only by qualified service personnel.
The following precautions should be observed to avoid instrument-induced electromagnetic interference:
The possibility of instrument-induced electromagnetic interference (EMI) is minimal if precautions outlined above are followed.
The instrument should not be placed close to any electrical device susceptible to EMI. It is suggested that the user post a sign warning pacemaker wearers that a microwave device is in operation.
If the instrument is suspected of inducing EMI, the cover should be carefully inspected. A microwave leakage measurement should be performed as outlined in the Troubleshooting, Maintenance and Service section of this manual. Leakage measured above the legal limit of 5 mW/cm2 should be reported to the CEM Service Department.
This instrument utilizes high voltages and microwave radiation.
Instrument service and repair should be performed only by those trained in repair and maintenance of high voltage and microwave power systems.
ii
To the best of our knowledge, the information contained herein is accurate. However, CEM cannot accept liability of any kind for the accuracy or completeness of the information contained in this manual. The final determination of the suitability and proper use of the instrument described herein, the accuracy of the information and data obtained from such use, and whether such use infringes any patents or the legal safeguards of others are the sole responsibility of the user.
Warnings, Cautions and Notes
Warnings, cautions and notes are included throughout this manual and should be read thoroughly and strictly followed.
WARNING: A warning is inserted for essential information used to emphasize dangerous or hazardous conditions to the operation, cleaning and maintenance of the instrument which may result in personal injury.
CAUTION: A caution is inserted for essential information used to emphasize procedures which, if not strictly followed, may result in damage or destruction to the instrument or improper instrument operation.
NOTE: A note is inserted for emphasis of procedures or conditions which may otherwise be misinterpreted or overlooked and to clarify possible confusing situations.
iii
iv
Table of Contents
Introduction......................................................................................................1
System Installation...........................................................................................3
External Printer ..........................................................................................7
External Balance ........................................................................................7
Personal Computer or Serial Printer Setup ................................................8
Instrument Description ...............................................................................9
Sample Analysis ............................................................................................13
Sampling ..................................................................................................13
Preparation ...............................................................................................14
Weight ......................................................................................................16
Holders .....................................................................................................16
Application ...............................................................................................18
Method Information .................................................................................22
Bias Identification ....................................................................................23
Determining Ash/COH for Processed Meats ...........................................25
Ash Contents of Meats .............................................................................26
Methods Development .............................................................................29
System Operation
Quick Test ................................................................................................31
Constant Weight..................................................................................31
Set Time ..............................................................................................37
Edit/Create Method ..................................................................................43
Edit Method ........................................................................................45
Create Method.....................................................................................47
Moisture/Solids - Standard Test ....................................................47
Moisture/Solids - Wastewater - TSS .............................................53
Moisture/Solids - Wastewater - TSS/TVSS ..................................59
Moisture/Solids - Dilution .............................................................65
Moisture/Solids - Syringe Weigh - Internal Balance ...................73
Moisture/Solids - Syringe Weigh - External Balance ..................79
Moisture/Solids - Ash....................................................................87
Moisture/Fat or Moisture/Fat/Protein - Standard ..........................93
Moisture/Fat or Moisture/Fat/Protein - Dilutions .......................101
Moisture/Fat or Moisture/Fat/Protein - Modified .......................111
Load Method ..........................................................................................121
Setup.......................................................................................................123
System Options .................................................................................123
System Information...........................................................................125
Printer................................................................................................125
Communication Port .........................................................................126
Contrast .............................................................................................126
Clock .................................................................................................127
Calibration.........................................................................................128
Security .............................................................................................129
Company Name ................................................................................133
Delete Method...................................................................................134
Serial Options....................................................................................135
Unit Serial Number ...........................................................................136
v
vi
Table of Contents (Continued)
Print..............................................................................................................137
Statistics .......................................................................................................139
External Components...................................................................................141
Maintenance, Troubleshooting and Service.................................................143
Cleaning Recommendations...................................................................143
Routine Maintenance .............................................................................143
Special Tools and Kits ...........................................................................143
Troubleshooting .....................................................................................144
Error Messages .......................................................................................146
Standard Solution Test ...........................................................................147
Microwave Leakage Measurement ........................................................148
Intelli-Temp Calibration.........................................................................149
Intelli-Temp Verification .......................................................................151
Service and Repair .................................................................................153
Specifications .........................................................................................155
Warranty.................................................................................................156
1
Introduction
The CEM SMART System5™ Moisture/Solids Analyzer is designed to provide a rapid moisture/solids analysis on a broad range of products. The patented weighing technology provides continuous weighing of the sample during the drying process. The system determines the precise time the sample is dry, and the moisture or solids content of the sample is then automatically calculated based on weight loss.
Microwave energy is used to dry the samples. Polar compounds such as water, polar organic solvents such as alcohols or ketones, and ions in the solution will absorb microwave energy and volatilize. Once water and/or other solvents have volatilized, the remaining solid material does not usually absorb microwave energy and will remain relatively cool. Thus, the weight loss is due only to the volatile solvents present in the sample. “Moisture” is used throughout this manual to refer to any volatile substance for which the system can measure weight loss after drying the sample.
The SMART System5 delivers 150 - 300 watts of microwave energy when
programmed for 100 percent power. Voltage of the incoming electrical service is measured and the microwave power is adjusted by the power control system in order to normalize the power and provide repeatable power conditions. Magnetron power automatically adjusted to deliver appropriate wattage to the instrument cavity. The magnetron fan remains on for 15 minutes after the end of the specified drying time to cool the instrument.
System components consist of a microwave drying chamber, an electronic balance, a 1/4” VGA black and white screen, an infrared temperature controller, a microprocessor and an internal impact printer. The Octawave™ cavity is designed to effectively deliver the microwave energy directly to the sample to achieve rapid test times. The 50 gram four-place analytical balance has a ±0.1 mg sensitivity. Infrared temperature measurement permits automatic adjustment of microwave power based on user defined temperature setpoint. The microprocessor controls and monitors system operations, automatically calculates sample results and enables interface with external printers, external balances and/or computers.
The system is equipped with three safety interlock switches to monitor the mating of the cavity cover and floor and one safety interlock switch to monitor the cavity cover latch. A flame detector with an infrared feedback automatically turns off the microwave power if excessive burning is detected.
The SMART System5 software is user friendly and easy to operate. It is capable of storing up to 100 methods and 300 test results, and is multilingual (English, German, Spanish, and French).
2
3
System Installation
1. Following the instructions provided in the packing carton, carefully remove the SMART System5 from its shipping carton and place it on a vibration-free workbench or laboratory table in a location that:
a. provides at least 8 in (20 cm) open space on each side and 6 in. (15 cm)
open space in the rear of the instrument for ventilation.
b. is free from vibration of large equipment and/or excessive walk-through
traffic.
c provides a temperature range of 41 °F (5 °C) to 104 °F (40 °C) and a
humidity range of 10-85 percent relative humidity. d. provides adequate bench space for sample handling. e. allows the instrument to be connected to a dedicated, grounded 120 or
240 VAC outlet. The SMART System5™ should be operated on a
stabilized, constant voltage AC power supply. To operate properly, the
voltage must be within ±10% of the specified level.
2. Retain all packing materials.
3. Inspect the instrument for cracks, dents or warping.
4. Remove the tape from the instrument cover. Inspect the top cover for proper alignment. The cover must open and close freely with no binding or restriction of movement.
WARNING
If damage is noted, do not attempt instrument operation.
AVERTISSEMENT
Ne pas mettre en marche si linstrument est endommagé.
5. If the instrument has been damaged in shipping, contact the freight
carrier to report the damage and to file a damage report. Contact the CEM Service Department or the nearest subsidiary or distributor (page 4) to request service information.
6. Verify that all accessories illustrated and listed in figure 1 are included.
7. Lift the instrument cover. Install the balance stem into the balance stem retainer in the opening of the cavity floor, ensuring that the balance stem is firmly seated. Install the balance pan onto the stem.
Note: If the balance stem requires removal, lift the stem straight up from its installed position. Moving the balance stem from side to side can cause damage to the balance assembly.
4
CEM Corporation
Service Department
P.O. Box 200
3100 Smith Farm Road
Matthews, NC 28104 -5044 USA
Within the continental United States
Telephone: (800) 726-5551
Fax: (704) 821-4368
Outside the United States
Telephone: (704) 821-7015
Fax: (704) 821-4368
Subsidiary Offices
CEM Ltd.
2 Middle Slade
Buckingham Ind. Part
Buckingham MK18 1WA
United Kingdom
Tel: (44) 1-280-822873
Fax: (44) 1-280-822342
CEM GmbH
Carl-Friedrich-Gauss-Str. 9
47454 Kamp-Lintfort
Germany
Tel: (49) 2842-96440
Fax: (49) 2842-964411
CEM S.r.l.
Via Dell’Artigianato, 6/8
24055 Cologno al Serio
Italy
Tel: (39) 35-896224
Fax: (39) 35-891661
5
Figure 1. Accessories
Starter Accessory Kit containing:
• Balance Stem (159600) (2)
Balance Pan (004120) (2)
Air Shield (568055M) (Installed in Instrument)
Standard Solution (302600)
Dispensing Bottle (230230)
Spatula (302001)
Disposable Pipette (302150)
Glass Fiber Sample Pads (200150)
10 g Weight (260111)
Fuses, 10AMP (2) (188340) and 5AMP (2) (188331)
Operation Manual (600140)
Printer Paper Holder (159065)
Printer Ribbon (315122) (Installed in Instrument)
Printer Paper (1 Roll) (315120)
Glass Fiber
Sample Pads
Reorder part # 200150
C
E
M
C
o
r
p
o
r
a
t
io
n
P
.
O
.
B
o
x
2
0
0
M
a
t
t
h
e
w
s
,
N
C
2
8
1
0
6
1
-
8
0
0
-
3
3
4
-
6
3
1
7
(
in
N
C
:
1
-
7
0
4
-
8
2
1
-
7
0
1
5
)
T
e
le
x
:
8
0
2
1
1
8
F
A
X
:
7
0
4
-
8
2
1
-
7
0
1
5
Moisture/Solids Analyzer
Operation Manual
6
8. Using a voltage meter, measure the voltage of the dedicated power outlet to verify voltage (120 VAC, 60 Hz or 220/240 VAC, 50 Hz).
9. Facing the back of the instrument, use a small screwdriver to pry the dual voltage supply from the power module.
10. Install the voltage slide switch in the proper direction for the measured voltage (120V, 60Hz or 220/240V, 50Hz).
11. Based on the voltage selected in step 9, install the proper fuses in the dual voltage supply – 10 AMP (60 Hz) or 5 AMP (50 Hz).
12. Install the dual voltage supply with the appropriate fuses and the properly installed voltage slide switch in the power module.
13. If applicable, install the exhaust hose.
14. Plug the power cord into the power cord receptacle in the rear of the instrument and into a grounded, dedicated electrical outlet. Note: If using external components such as a printer, computer or balance, install these components as outlined in this manual.
15. Position the power switch located in the rear of the instrument above the power cord receptacle in the “on” position.
CAUTION
Permit the SMART System5™ to warm up (power switch on, microwaves not being produced) for at least 45 minutes prior to operation. If the instrument is switched off for an extended period of time (weekend, holidays, etc.), ensure that warmup time is permitted prior to use.
Mise en garde
Permettre une période de 45 minutes de réchauffement du SMART System5 avant lutilisation (interrupteur de courant en marche, aucune micro-onde n’est émise). Lorsque linstrument
est en mode fermé pour un temps prolongé (fin de semaine, vacances, etc. . .) il est important que la période de réchauffement de 45 minutes soit observée.
16. Refer to the “Maintenance, Troubleshooting And Service” section of this manual and perform the Standard Solution Test at least five times.
Note: CEM Corporation recommends that the power cord be connected to an AC power outlet and the power switch remain in the “on” position at all times. After 15 minutes of idle time, the SMART System5 assumes a “sleep” mode.
7
Instrument Description
Display – displays menu choices, instructional messages and analysis results on a 1/4 VGA black and white screen
Keyboard – controls operation of the SMART System5. Temperature, time, or other numeric data may be entered with the numeric keys as requested on the display.
Internal Printer – provides printout of methods, data and results. Cavity – decreases test times due to its patented compact Octawave™ design. Air Shield – prevents air flow within the microwave cavity from affecting weight
measurements.
Balance Pan – holds sample and sample pads during analysis. Interlock Assembly – monitors mating of cavity cover and instrument skirt.
Display
Keyboard
Internal
Printer
Cavity
Air Shield
Balance
Pan
Figure 2. Smart System5 Front View
Interlock
Assembly
Interlock
Assembly
Interlock
Assembly
8
Enter – Press to store entered parameters and data in the computer memory. Exit – Press to exit screen. Ready – Press to initiate current method for a test or to reset method for next test. Start/Pause – Press to begin or pause sample analysis. Microwave heating begins
when the Start/Pause button is pressed unless the cavity cover is open. Operation will continue through the analysis unless the Start/Pause button is pressed to interrupt the analysis or the operation key indicating “Stop Test” is pressed.
Tare – Press to tare or zero the weight of the sample pads prior to sample analysis. 0 - 9 – Press to select items from menus such as the Main Menu, setup screens, etc.
and to select values for parameters such as time and power. Operation KeysPress to select specific operation options such as “Stop
Test,” “Print,” “Main Menu,” Prev. Page,” etc.
Figure 3. Smart System5 Keyboard
9
Fuses – prevent electrical power overload. Power Cord Receptacle – receives the female end of the power cord. Power Switch – turns electrical power to the instrument on and off. Nameplate – lists model and serial number. Dual Voltage Selector Switch - permits electrical voltage for the instrument to be
switched from 110 VAC to 220 VAC. PC Port – permits connection and communication with a personal computer for
data storage, etc.
Printer Port – permits connection and communication with an external printer. External Balance Port – permits connection and communication with an external
balance for specific applications.
Internal
Printer
Power
Switch
Power
Cord
Receptacle
Dual Voltage
Selector
Switch &
Fuses
Printer
Port
External
Balance
Port
PC
Port
Figure 4. Smart System5 Rear View
CE
CEM
Nameplate
10
11
Sample Analysis
Proper sample preparation and testing are critical to obtaining accuracy and precision. Each sample has a specific characteristic composition; therefore, each sample may require an individual method of analysis. However, guidelines can be established for analysis of general sample types.
Sampling – A representative sample is critical to obtain proper analysis.
Preparation – As with any chemical analysis, a homogeneous sample is
critical to obtaining precision and accuracy.
Weight – A consistent sample size helps ensure precision and accuracy.
Most CEM applications specify a sample size of 2-5 grams, emphasizing the importance of a homogeneous sample.
Holders – Absorbent pads to which a sample can be applied are used for
most analyses. Glass fiber pads are recommended because glass fibers do not absorb microwaves. Other holders include Teflon coated glass fiber mesh, a drying basket and a thermapad.
Application – Most samples should be spread smoothly, avoiding lumps,
peaks or any uneven areas. Sample spread must be consistent.
Drying Time – The drying time is determined by the type of sample and its
microwave absorbency. Constant weight drying permits drying of a sample until a constant weight is achieved. Set time drying permits the control of the sample analysis by entering a specified time and power level.
Infrared Temperature – The infrared temperature measurement may be
used to operate at temperatures used in standard or vacuum oven procedures.
Microwave Power – The power necessary to dry a sample is determined by
sample composition. A sample should be tested at the highest possible power without burning or degradation of the sample.
Sampling
Proper sampling of a product or process is critical in achieving proper moisture/ solids analysis. A representative sample is required to ensure that the test results are representative of the entire batch or lot of material.
Many standard procedures such as AOAC, USDA and ASTM methods reference sampling techniques. It is extremely important to follow techniques outlined in these procedures for a particular sample type.
Proper handling and storage of the sample prior to analysis is also critical in achieving accurate and representative results. Samples should not be exposed to the air for long intervals prior to analysis. Proper storage also ensures that no moisture is lost.
12
Preparation
Sample preparation is critical to achieving precise and accurate test results. The sample should be properly prepared to achieve repeatable moisture results between duplicate analyses of the same sample. Sample preparation is either a physical or chemical modification of the sample. Physical modification may be as simple as stirring or shaking the sample to ensure homogeneity or more complex such as particle size reduction, dilution or deaeration.
Chemical modification is the addition of an enzyme to hydrolyze chemically bound fat. The objective is to create a homogeneous sample and to improve the correlation of the test method to standard methods. Some products, such as a beef emulsion or comminuted poultry, may require no additional preparation. Other materials will require particle size reduction, dilution or dearation prior to testing. Samples should be kept cool and in sealed containers to prevent loss of moisture. If repeatable results cannot be obtained on the same sample, re-blend the sample to improve the homogeneity and repeat tests. Sample preparation methods include:
1. Mixing of Sample – Thoroughly stir or shake sample to ensure
homogeneity.
2. Particle Size Reduction – Many products require particle size reduction
prior to being applied to the glass fiber pad. Reduction, based on product sample, can be accomplished with a meat grinder, either an industrial or consumer grade food processor, a coffee mill or a cheese grater. The meat grinder is the best choice for all meat products. USDA procedure specifies three to four passes through a 1/8” or 5/64” blade with mixing between grinds. A USDA inspector or outside laboratory can make recommendations. An industrial grade food processor is a good alternative for grinding meat products. A consumer grade food processor is ideal for grinding frozen potatoes, breaded products, or “dry” snack foods such as cookies, crackers, and chips. A coffee mill can be used to grind or mix “dry” snack foods. A cheese grater can be used to grate small quantities of hard cheese samples or soap.
3. Dilution – Products with a high carbohydrate content are strong
microwave absorbers and have a tendency to burn. These samples may require either a water or salt dilution.
A. Water Dilution – A specified amount of sample is diluted with a
specified amount of water and mixed in a blender. Typical samples include doughs, cheese powders, corn syrups, dried meat bases and sauces. An external balance and a blender are required for this procedure.
(1) Interface an external balance with the SMART System5. (2) Tare the blender cup on the external balance. (3) Add product sample to the blender cup.
13
(4) Add water to the product sample in the blender cup. (5) Blend until a homogeneous mixture is obtained. (6) Use the diluted sample to perform analysis.
B. Salt Dilution – A specified amount of salt is added to the product
sample prior to analysis. Typical samples include frozen potatoes and frozen breaded products. This procedure requires a teflon basket and dry table salt (sodium chloride).
(1) Line a teflon basket (CEM p.n. 200090) with two square glass
fiber pads.
(2) Place the lined basket on the balance of the Moisture/Solids
Analyzer. Press “Tare.”
(3) Place approximately 8g of salt in the lined basket. Press
“Tare.”
(4) Place the product sample on top of the salt. Press “Start” to
record the initial weight. (5) Press “Stop.” (6) Thoroughly mix the salt and sample in the basket. (7) Return the basket to the balance of the Moisture/Solids
Analyzer. Press “Start” to continue the sample analysis.
4. Deaeration – Frozen dairy and imitation dairy products must be
deaerated to remove the excess air in order to provide reproducible test results. This procedure requires a 125 mL Erlenmeyer flask, a 1 liter beaker, a magnetic stirrer, a 11/2” x 3/8 magnetic stirring bar, and a #5 rubber stopper.
(A) Place 100 mL of sample in a 125 mL Erlenmeyer flask. (B) Place a 11/2” x 3/8” magnetic stirring bar in the flask. Using a #5
rubber stopper, loosely cap the flask.
(C) Place flask in a 1 liter beaker containing 250 mL of water which has
been heated to 70°C.
(D) Place beaker with flask on a magnetic stirrer. Mix for 5 - 7 minutes. (E) Cool sample to ambient temperature and thoroughly mix sample
prior to testing.
14
5. Enzymatic Hydrolysis – The fat in processed cheese and dairy products
is often tightly bound to the protein. In order to achieve accurate test results, the fat must be released from the protein prior to the fat extraction process. This can be accomplished by a protease enzymatic hydrolysis of the sample. Note: The sample must be in a liquid or “pourable” state for the enzymatic hydrolysis to be effective.
(A) Thoroughly mix approximately 100g of sample. Heat sample in a
water bath at 60°C (± 5°C) for five minutes.
(B) Add eight (8) drops of CEM Enzyme (p.n. 302630) and mix well. (C) Return sample to water bath for five minutes.
Weight
Sample weight can be very critical to the final test result. As with most analytical procedures, a consistent sample size helps ensure precision and accuracy. Sample sizes can vary, based on type of sample, from 1 gram on the low end (high percent solids level) to 10-15 grams on the high end (very low percent solids level). Test procedures documented by CEM and included in this manual, are for specific sample types and suggested sample size. Sample size should remain within the weight guidelines. CEM has determined that the suggested sample size results in the most consistent drying and test results. Heavier sample weights may cause sample degradation or burning, resulting in high moisture results. Lower sample
weights may result in under-drying of the sample.
Holders
Sample holders should be transparent to microwave energy. CEM Square Sample Pad – The square pad is the most commonly used sample
holder. Typically, two (2) pads are used with the sample “sandwiched” between the pads. For heat sensitive samples, only one pad is recommended to minimize heat buildup. For low solids samples, three or four pads are used for absorption.
CEM Round Sample Pad – The round pad is used as a collection pad for fat extraction and is recommended, but not required, for thermapad applications.
Sample Pads – CEM glass fiber pads are designed to provide optimum conditions for rapid, thorough and repeatable drying. CEM Corporation follows strict specifications in the manufacture of sample pads. Specifications include low moisture content and optimum absorbency, density, porosity, and strength for microwave drying.
Moisture Content – Excessive moisture in sample pads will cause
moisture or solids values to be incorrect. Controlling the moisture ensures that residual moisture is maintained to very close tolerances.
15
Absorbency – The pad material must be absorbent enough to form a wick for a liquid sample and to be a sample holder for more viscous type samples.
Density – The sample medium must have uniform density to guarantee that the moisture will volatilize rapidly and evenly, without causing residual heating.
Porosity – Correct porosity of the pad material ensures maximum surface area for liquid samples in order to increase volatilization of moisture. Correct porosity helps prevent sample degradation or burning.
Strength – The tensile strength of the medium has a direct effect on the ability of the sample pad to withstand the stress imposed during sample application.
Teflon Coated Glass Fiber Mesh – The mesh is used as a holder for fiber samples or as a “spacer” between multiple sheets of paper samples.
Drying Basket – The basket is made of Teflon coated glass fiber mesh. It is typically lined with two square pads. It is used to hold large samples and for salt dilutions.
16
Application
The viscosity of the sample plays an important role in determining how the sample is spread onto the pad for analysis. If a sample is thin, it can be pipetted onto the pad and can be tested at high power levels. A thin sample spread over a large area permits moisture to evaporate rapidly with little residual heat buildup. Glass fiber pads may be stacked for increased absorbency.
High viscosity samples (samples that do not flow well) require a different technique. If the sample can be easily spread across the glass fiber pad, no dilution is required. A thick sample will tend to puddle and form a crust, sometimes trap­ping volatiles within the bubble. Overheating or burning can also occur on thick samples. If repeatable results cannot be obtained by spreading the sample, a dilution may be necessary. The sample must be soluble in the diluting agent, and the diluting agent must be a microwave absorber.
To generate the appropriate amount of heat and achieve optimum moisture results, different types of samples must be spread to different thicknesses over different areas of the glass fiber sample pad. As a general guideline, the sample will generate heat in direct proportion to the thickness of the sample layer on the pad. A sample spread too thickly can cause sample burning; a sample spread too thinly can cause insufficient drying. The prepared sample should be applied to a glass fiber sample pad. Once the sample is properly applied to the sample pad, it should be covered with a second glass fiber pad to ensure heat retention and to eliminate splattering.
Liquid samples should be pipetted onto the pad. Solids samples should be spread onto a pad placed on the edge of a clean, flat surface, such as a countertop. The spatula should be held level with the pad when spreading the sample to ensure a smooth, uniform application of sample to the pad surface.
If the sample is an organic solvent-based liquid, special handling is required. Equipment and method modification may be required. CEM recommends using the SMART System5 Exhaust Kit when analyzing solvent based samples.
Non-polar organic solvents do not absorb microwave energy; therefore, samples containing this type of solvent will not dry unless they are mixed with a diluent which absorbs microwave energy and has a higher boiling point than the non-polar solvent.
If the sample is a polar organic solvent-based liquid, samples will absorb microwave energy and dry as readily as a water-based liquid. If the organic solvent is an alcohol, a ketone, or an ester, samples should be tested with water-based techniques.
17
Some base resins are microwave absorbers. Polyester is a known base resin that is an absorber of microwaves. Titanium dioxide and calcium carbonate have also been found to be microwave absorbent. If these materials are present in a sample, usually enough heat is generated to remove most high boiling non-polar solvents and any other volatiles. Oils become very hot when placed in the microwave field. Some metal based samples can be tested. If carbon black is present in a sample, it will heat very quickly and possibly burn even at low power levels.
It is recommended that CEM Corporation or a local distributor be consulted prior to testing any solvent based sample.
General Sample Application Technique
If the sample is in a paste, semi-solid, or crumb form or a raw or skeletal meat product such as fresh pork, ground beef, or chicken, place the sample on the end of a spatula and spread it across one end of the pad. Then spread the sample to a uniform thickness covering approximately 90% of the pad surface area.
1 Spread Thin Layer Across 90% of Pad
18
If the sample contains bound water such as an all-meat emulsion, cooked all-meat sausage, sausage with extenders, semi-dry sausage, or ham, place the sample on the end of the spatula and apply the sample to the middle of the pad. Then spread the sample around the pad in a circle to generate thermal heat.
2 Spread Layer on Pad
If the sample is a heat sensitive sample such as a water-based or solvent-based liquid, use a pipette to dispense the sample onto the pad(s). Sample size should be 2-5 grams, based on solid content of the sample – 2-3 grams for high solids samples, 4-5 grams for samples containing less than 15 percent solids. Samples containing less than two percent solids may require as much as 10 grams for analysis. Large quantity samples should be pipetted onto the pads in a thin even layer. Additional glass fiber pads may be used for increased absorbency. A sufficient number of pads should be used to ensure that liquid samples do not drip onto the floor of the instrument cavity.
3 Puddle, Spread with Pipette and Cover
19
4 Puddle and Cover
5 Swirl or Zig-Zag Pattern, No Cover
Standard application used for heat sensitive samples if sample burns when using
the puddle and spread technique.
If the sample is a water-based liquid or solvent-based chemical, use a pipette to puddle the sample onto the pad(s). Cover the sample with an additional pad. A sufficient number of pads should be used to ensure that liquid samples do not drip onto the floor of the instrument cavity.
20
Method Information
Most water based samples should achieve constant weight settings in 1 - 3 minutes.
The microwave power and temperature settings should be appropriate to the sample. The standard oven or vacuum oven temperature recommended in a reference method for the sample should be followed. If no reference method is available, a starting temperature of 100 °C should be programmed. Because the power level in the instrument is adjusted based on temperature feedback, most samples can be dried using a 100% power level setting. If a sample contains a strong microwave absorber such as carbohydrates, the power level will likely need to be adjusted. In most instances, a 10% power change will be sufficient to prevent overheating.
During the initial testing of a new sample, the weight display should be monitored to ensure that the sample does not ignite and that a stable weight reading is displayed. Sharp declines in weight indicate excessive sample heating. If ignition occurs, the flame detection sensor will abort the analysis.
Samples that are strong absorbers of microwave energy should be diluted or tested at reduced power levels. These products usually contain simple carbohydrates or tightly held moisture that will not readily escape, causing overheating and/or burning and non-reproducible results. These samples should be tested at a low power. Examples of heat sensitive samples include powders, paper fibers, candy and black liquor.
21
Bias Identification
Most samples can be analyzed for percent moisture content with a CEM drying method, and results will compare precisely with the standard test method. Occasionally, however, all volatiles in a particular sample cannot be removed with microwave power. Different spread techniques can be used to eliminate the need for a moisture bias in most samples. Sometimes browning of the sample in the same manner as in the air oven will also produce test results comparable to the standard test.
If microwave moisture test results are repeatable, but not comparable to the standard test results, a bias can be identified. With most samples, the bias will amount to only a few tenths of a percent. To determine a moisture bias, prepare a sample for testing, ensuring that the sample is homogeneous. Test this sample by both a standard method and the CEM microwave method. Perform at least ten tests by each method to check repeatability, then figure an average of the ten results. Subtract the average of the microwave test results from the average of the standard test results. Enter the resulting number (moisture bias) into the CEM moisture/ solids instrument. Corrected moisture test results will be displayed and printed. If the difference in the average test results is a negative number, the power level should be reduced and the tests should be repeated.
Some meat and poultry products, whether raw or cooked, show a fat bias when analyzed with the CEM instruments. Less fat is extracted by a short solvent extraction than by the longer ether extraction of the traditional Soxhlet technique. The bias will be a constant for a given product and should be considered when generating data.
Cooked sausage (i.e. wieners, bolognas, etc. with extenders) and other cooked products tend to bind fat in the carbohydrate/protein structure of the cooked product and will generally yield a lower fat value when extracted in the CEM instrument than when extracted with the Soxhlet.
For proper determination and use of biases, results from the CEM instrument should be compared with the results from traditional “long” or standard methods used by USDA laboratories for compliance testing. Standard methods consist of an air oven (four hours, 125°C) for moisture and Soxhlet (four hour ether extraction) for fat. USDA recommends correlation of rapid fat and moisture methods with these long methods.
USDA Processed Products Inspection Division encourages the use of rapid methods for fat and moisture analysis to assist in better process control and compliance. USDA recommends periodic back-up analysis (one in every 10 - 20 analyses) by standard long methods when rapid methods are used in approved Total and Partial QC (TQC/PQC) Programs. Backup checks can be performed less frequently as laboratories gain confidence in the established biases. CEM Corporation also recommends a Quality Assurance Program for establishing and maintaining biases and to ensure USDA compliance.
Prior to establishing a fat bias for the Fat Analysis System, a laboratory should be confident of obtaining the best possible results for moisture, since precision of the fat results will depend on the moisture results. Moisture results should have repeatability between duplicate analyses of the same sample and a mean or
22
”average” moisture values that are close to the values being generated on the same samples by standard oven. To obtain the best possible results for moisture, and thus for fat, there are three factors to consider: 1) sample preparation, 2) sample application technique and 3) power and time parameters.
After establishing moisture repeatability and achieving moisture results as close to standard ”long” methods as possible, use the following procedure to identify fat biases for different product categories.
1. Select three or more samples (approximately two pounds each) from different lots of the same product.
2. Prepare each sample according to standard procedures (3 - 4 passes through a 5/64” blade, with mixing between grinds, or equivalent).
3. Divide each ground sample in half and package and label sub-samples in two separate, air-tight containers.
4. Analyze one set of the sub-samples (from step 3) in duplicate by standard ”long” methods – oven for moisture, Soxhlet for fat. If facilities and/or personnel are not available for these tests, use an independent laboratory. Request raw data (individual tests) from the laboratory, not the mean or ”average” value.
5. After obtaining the results from the standard ”long” methods, analyze the other set of subsamples in the CEM Moisture/Solids Analyzer for moisture only. Adjust power and spread technique until acceptable moisture repeatability is achieved.
6. After obtaining acceptable moisture repeatability, use the Fat Analysis System and perform 3 - 5 moisture and fat analyses for each sample. As with moisture, fat results should be repeatable. To obtain optimum fat results, use a sufficient redry time. Two minutes is typical, although some samples may require more time. The sample weight should be stable (<0.2 mg weight change in the last five seconds). An insufficient redry time will yield artificially low fat results since the true dry, extracted weight will not be reached.
7. Subtract values obtained from the SMART System 5 from Soxhlet values. Average the differences to identify the average bias for a specific product or product group.
Sample A B C
Average Soxhlet fat xx xx xx
- Average CEM fat xx xx xx = Bias xx xx xx
(A + B + C)/3 = Average Fat Bias
8. The bias should be relatively less than 2.0% and must be added to the fat results from the SMART System 5 to produce results that correlate closely with official methods. Repeat this process on a regular basis to confirm that the bias is consistent. Raw meats and blends will produce smaller biases than cooked products with extenders.
23
Determining Ash/COH for Processed Meats
For samples that contain salt, sugar, spice, cure or other added components, ash/ carbohydrate (COH) values must be determined by analysis. For purposes of calculating the protein by difference, the specific amount of each component (salt, spice, bone, etc.) is not important, only the total amount.
To determine ash/COH values for finished products or blends with sugar, salt, spice, etc., use the following procedure:
1. Collect and prepare three (3) samples from different lots of the same product. Store the samples in airtight containers until tested.
2. Analyze the 3 samples in duplicate for fat and moisture in the SMART System 5 and Fat Analysis System and for protein using “long” methods (Kjeldahl). If necessary, an independent laboratory can provide this analysis.
3. Average the results of the duplicate analyses for each sample, then total the % Protein + % Fat + % Moisture for each sample.
4. Subtract the sum from 100% to get an Ash/Carbohydrate factor. For example, if the composition of a sample is:
Moisture 52.54% Fat 28.36% Protein 11.25% Total 92.15%
Ash/COH = 100% - 92.15% Ash/COH = 7.85%
5. Calculate an average of the the ash/COH values for each of the 3 samples and program that factor into the SMART System 5.
Check samples periodically to verify the ash/COH values. Variations in the amount of added constituents (salt, sugar, spice, extenders, etc.) can cause the ash/ COH factor to vary.
7.85%
11.25%
8.36%
52.54%
Moisture Fat Protein Ash/COH
Loading...
+ 133 hidden pages